Evaluation of the First Automated Thyroglobulin Assay

  • Michael Vogeser , Peter Knesewitsch , Karl Jacob und Dietrich Seidel
Veröffentlicht/Copyright: 1. Juni 2005
Veröffentlichen auch Sie bei De Gruyter Brill
Clinical Chemistry and Laboratory Medicine (CCLM)
Aus der Zeitschrift Band 37 Heft 2

Abstract

The aim of this study was to investigate technical and analytical performance of the first automated thyroglobulin (Tg) assay (DPC-Immulite®; Diagnostic Products Corporation, Los Angeles, USA). In imprecision studies using several human serum pools ranging from 21 to 58 replicates, a coefficient of variation of 9.0 % was obtained at a mean Tg concentration of 0.84 ng/ml and of 6.1 % at a Tg concentration of 62.1 ng/ml. In a method comparison with a non-automated assay (BRAHMS LUMItest Tg®, BRAHMS, Berlin, Germany) using 383 sera of 303 patients with thyroid carcinoma, regression analysis according to Passing and Bablock yielded in the following equation:

Immulite Tg=1.6 × BRAHMS Tg−0.1 ng/ml

(Pearson's r=0.979).

Sera obtained from 59 patients with thyroid carcinoma enabled comparative follow-up studies; in all cases qualitative agreement was found with regard to increase or decrease of serum Tg; in eight cases, however, Tg was detected with the Immulite assay but not with the BRAHMS assay. Further follow-up proved the presence of thyroid tissue in these patients. From these and further methodological data (dilution linearity, interference studies, carry-over study, high-dose hook properties, and short report time) it is concluded that the DPC-Immulite Tg assay meets the requirements of routine diagnostic use.

:
Published Online: 2005-06-01
Published in Print: 1999-02-01

Copyright (c)1999 by Walter de Gruyter GmbH & Co. KG

Artikel in diesem Heft

  1. Evolution or Revolution in Clinical Chemistry
  2. Low Levels of Plasma Proteins: Malnutrition or Inflammation?
  3. Laboratory Medicine: The Need for a Broader View. The “Multiple Bundle” Model of Clinical Laboratory Function
  4. Activated Protein C (APC)-Resistance: Automated Detection of the Point Mutation at Position 1691 in the Factor V Gene
  5. A Rapid and Sensitive Automated Light Scattering Immunoassay for Serum C-Reactive Protein and the Definition of a Reference Range in Healthy Blood Donors
  6. Photodynamic Effects In Vitro in Fresh GynecologicTumors Analyzed with a Bioluminescence Method
  7. An ELISA for the H-Subunit of Human Ferritin which Employs a Combination of Rabbit Poly- and Mice Monoclonal Antibodies and an Enzyme Labeled Anti-Mouse-IgG
  8. The Use of Free Cortisol Index for Laboratory Assessment of Pituitary-Adrenal Function
  9. Allele-Specific Amplification for the Diagnosis of Autosomal Recessive Spinal Muscular Atrophy
  10. Diagnostic Value of Biochemical Markers of Bone Turnover and Postmenopausal Osteoporosis
  11. Isoenzmyes of Class I and II Alcohol Dehydrogenase in Chronic Hepatitis
  12. Screening School Children for Albuminuria, Proteinuria and Occult Blood with Dipsticks
  13. Evaluation of the First Automated Thyroglobulin Assay
  14. Selective Imidazoline Receptor Agonists and Lipid Metabolism
  15. Judgement on Analytical Quality Requirements from Published Clinical Vignette Studies Is Flawed
  16. The Meaning of Good Laboratory Practice (GLP) for the Medical Laboratory
  17. Improved Procedure of Measuring on Radiometer 500 and 600 Series Bloodgas Analysers
  18. Scientific Papers and Presentations by Martha Davis
  19. New Myocardial Marker Proteins in Acute Myocardial Infarction. Quantitative Aspects by J. A. Kragten
  20. Normalwerte by Michael Jakob
  21. Meetings
Heruntergeladen am 11.9.2025 von https://www.degruyterbrill.com/document/doi/10.1515/CCLM.1999.028/html
Button zum nach oben scrollen