Home Medicine Quality control in the Netherlands; todays practices and starting points for guidance and future research
Article Open Access

Quality control in the Netherlands; todays practices and starting points for guidance and future research

  • Huub H. van Rossum EMAIL logo , Marith van Schrojenstein Lantman , Michel Severens , Henricus J. Vermeer , Wilhelmine P.H.G. Verboeket-van de Venne , Wytze Oosterhuis ORCID logo and Niels de Jonge
Published/Copyright: April 26, 2024

Abstract

Objectives

Adequate analytical quality of reported results is primarily ensured by performing internal quality control (iQC). Currently, several different iQC practices are in use. As a prelude to the revision of a Dutch guidance document on analytical QC, a questionnaire was sent out to gain insights in the applied practices and the need for guidance.

Methods

A questionnaire, containing 20 multiple-choice questions with possibilities for explanation and comment on iQC practices and aspects was distributed to all clinical chemistry laboratories within the Netherlands. Results were reported descriptively.

Results

Responses were received from 27 clinical laboratories (response 43 %). In 30 % the iQC was based on the analytical characteristics only, while 30 % used a 6-Sigma method, 19 % risk-based beyond 6-Sigma and 22 % used an alternative approach. 89 % of laboratories used a virtual analyzer model for iQC setup within one or more laboratory sites. Practices for determining standard deviation (SD) values included determining SD for each new iQC material (35 %), using historical SD values for new materials (35 %), and incorporating clinical tolerances into the SD value (31 %). Furthermore, 44 % of laboratories used patient moving averages for one or more tests. Daily iQC management was based on either “traffic lights” indicating in or out of control status, and review of all QC charts, often using multiple software systems.

Conclusions

A large heterogeneity of iQC practices in clinical laboratories was observed in the Netherlands. Several starting points for further research and/or guidance were identified, particularly in relation to the determination of SD values, the virtual analyzer model and methods to ensure analyzer equivalence.

Introduction

Ensuring adequate quality of reported test results is one of the primary responsibilities of medical laboratories. Medical laboratories have several tools at their disposal to accomplish this, including internal quality control (iQC) measurements, external quality control measurements, and patient-based real-time quality control (PBRTQC) including manual verification of patient results and patient moving averages [1]. iQC is an essential, and perhaps the most prominent, component of the quality assurance for medical tests. iQC is based on the frequent, e.g. daily or within specified run, analysis of control materials with a predetermined target and standard deviation (SD) value for specified (quantitative) measurement systems. The interpretation of these iQC measures is generally based on a statistical approach using SD-based control rules, also referred to as statistical quality control (SQC).

The iQC practice of setting up SQC and the theoretical concepts of iQC have evolved from standard 2SD rules based solely on analytical characteristics, to sigma metric approaches using clinical tolerance limits based SQC set-up via multi- (Westgard-)rules, to the more recently described, risk-based bracketed SQC that also considers the number of patients at risk [2], [3], [4]. With each new development and conceptual advancement, the complexity of SQC increased significantly. In addition, the use of PBRTQC to support iQC for assuring the analytical quality has also gained momentum [5, 6]. With all of these developments, the general medical laboratory is challenged to select the iQC setup that is most appropriate for the laboratory. This becomes even more important when there are no legal or accreditation/certification-required SQC practices as in the Netherlands. Therefore, each medical laboratory must decide separately how to perform the analytical quality assurance. Finally, consistency and comparability of laboratory results is also increasingly a major priority in The Netherlands due to increasing consolidation of individual laboratories into large lab-chain organizations. This often triggers the desire for harmonized iQC procedures covering a large number of analyzers and multiple laboratory sites.

To have an understanding of the current analytical quality assurance practices and the adoption of the newer SQC and PBRTQC concepts, a survey was conducted among medical laboratories in the Netherlands. A second objective of the survey was to extract relevant information, to be incorporated into a revised version of a Dutch guidance document on Six Sigma analytical quality control.

Materials and methods

Questionnaire design

A questionnaire was designed to cover a number of practical aspects of performing analytical quality assurance. It focused primarily on measurement procedures that produce quantitative results. The final questionnaire consisted of 20 questions that generally addressed the following topics: conceptual basis for setting up iQC; practice in defining SD, target values and SQC; the use of tools other than SQC; Information technology (IT) system used; management of iQC on a daily and long(er) term basis; and need for training or practical guidance. The final questionnaire (in Dutch) is presented in Supplementary Data 1 and English translation in Supplementary Data 2.

Running the survey

The final questionnaire was configured in the application SurveyMonkey (May 2023). Next, an e-mail with the link to the questionnaire was sent via the Dutch Society of Clinical Chemistry and Laboratory Medicine (NVKC) to the medical directors of all medical laboratories in the Netherlands (n=63). They were asked to forward the questionnaire to the person responsible for the analytical quality assurance policy within the laboratory organization and to complete the questionnaire on behalf of the entire laboratory organization. One reminder was sent to the medical directors and the total time the questionnaire was available, was 6 weeks.

Results

Participants

A total of 27 participants completed the questionnaire (43 % response). Three on behalf of university medical centers, 14 on behalf of a single local hospital, 8 on behalf of laboratories supporting multiple local hospitals, 2 on behalf of laboratories focused on specific patient populations (categorical hospital). The majority of the participating laboratories 17/27 (63 %) provided primary care diagnostic services.

General basis of iQC system setup

When asked about the general basis of the iQC system (for definitions see Table 1), 8/27 (30 %) used Six Sigma, 5/27 (19 %) used risk-based iQC that goes beyond Six Sigma and includes additional QA tools based on patient risk, 8/27 (30 %) had iQC based primarily on the analytical characteristics, and 6/27 (22 %) applied an alternative approach. 5/6 of the latter also used biological variation and Six Sigma based approaches, and several used EQA results and performance as a basis for SQC design. EQA in the Netherlands (SKML) has sigma metric based interpretation and reporting. None of the respondents based the iQC on acceptance limits established and provided by suppliers. When asked about (clinical) tolerance limits, 8/27 (30 %) did not take these into account and focused only on the analytical characteristics. 17/27 (63 %) used biological variation as a basis and 2/27 (7 %) used their own clinical knowledge to determine the tolerance. Overall, 30 % of laboratories used the analytical characteristics as the basis for iQC and 70 % used clinical tolerance criteria and patient risk, either based on Six Sigma or alternative methods, as the basis for designing iQC (Table 2).

Table 1:

iQC set-up definitions.

Definitions of basic iQC system Description
Based on analytical characteristics iQC set-up is primarily based on the analytical characteristics of the measurement procedures. No analytical performance criteria or clinical tolerance levels are taken into account
Six Sigma-based iQC set-up is based on Six Sigma quality control; based on the sigma values of a measurement procedures multi (Westgard) – rules are selected
Risk-based extending Six Sigma Basic iQC set-up is Six Sigma quality control; based on the sigma values of a measurement procedures multi (Westgard) – rules are selected. In addition, additional or modifications to the iQC set-up such, as modifications to QC frequency and addition of patient moving averages are applied, based on an additional clinical risk assessment
Table 2:

Overview of laboratories and answers related to SQC iQC set-up.

Lab ID Type of lab 6-Sigma applied iQC setup methodology Use of clinical tolerance limits SD value determination QC rules applied Virtual analyzer system Other QA/QC tools Use of “traffic” light
1 MPH No Anal char based No For each new QC material 2× 2SD, 1× 3SD Yes, applied for >1 lab sites No Yes
5 SPH, PC No Anal char based Yes, based on EL For each new QC material 1× 3SD, 10× 1SD Yes, within lab site No Daily QC graph check
20 SPH No Anal char based Yes, based on EL For each new QC material 1 × 2SD, 2 × 2SD, 1 × 3SD, trend Yes, applied for >1 lab sites Yes, PMA Yes
8 MPH No Anal char based No Historical SD or CV(%) 2× 2SD Yes, within lab site No Yes
18 SPH, PC No Anal char based No Historical SD or CV(%) 2 × 2SDa, 1 × 3SD, R 4SDa No Yes, Bhat-anal. Daily QC graph check
10 UMC No Anal char based No SD includes clinical tolerance 2 × 2SD, 1 × 3SD Yes, within lab site Yes, PMA Daily QC graph check
14 SPH, PC No Biol var based Yes, based on BV SD includes clinical tolerance 2 × 2SD, 1 × 3SD Yes, within lab site No Daily QC graph check
2 SPH, MPH, PC No Method based No For each new QC material 1× 2SD, 2× 2SD, 1× 3SD, suppl Yes, within lab site Yes, PMA EQA Daily QC graph check
3 MPH, PC No Method based No Historical SD or CV(%) 2× 2SD, 1× 3SD Yes, within lab site Yes, PMA Daily QC graph check
15 SPH No Other Yes, based on BV SD includes clinical tolerance 2 × 2SD, 1 × 3SD Yes, applied for >1 lab sites Yes, PMA Yes
25 SPH, PC, others Yes 6-Sigma based Yes, based on BV For each new QC material 1 × 2SD, 1 × 3SD, 1× 3.5SD, suppl Yes, within lab site No Yes
9 SPH, PC Yes 6-Sigma based Yes, based on BV Historical SD or CV(%) 2× 2SD, 1 × 2.5SD, 1 × 3SD, R 4SD, 4 × 1SD Yes, within lab site Yes, PMA Yes
12 MPH, PC Yes 6-Sigma based Yes, based on BV Historical SD or CV(%) 2 × 2SD, 1× 2.5SD, 1 × 3SD, 1 × 3.5SD, R 4SD, 4 × 1SD, Suppl. Yes, applied for >1 lab sites No Daily QC graph check
13 UMC Yes 6-Sigma based Yes, based on BV Historical SD or CV(%) 2 × 2SD, 1 × 3SD, 1 × 3.5SD, 4 × 1SD Yes, within lab site Yes Yes
21 SPH, PC Yes 6-Sigma based No Historical SD or CV(%) 1 × 2SD, 2 × 2SD, 1 × 2.5SD, 1 × 3SD, 1 × 3.5SD, R × 4SD Yes, within lab site No Yes
27 MPH Yes 6-Sigma based Yes, based on BV Historical SD or CV(%) 1 × 2SD, 2 × 2SD, 1 × 2.5SD, 1 × 3SD, 1 × 3.5SD, R × 4SD, 10 × 1SD Yes, applied for >1 lab sites Yes, peer-comp. Daily QC graph check
16 SPH, PC Yes 6-Sigma based Yes, based on BV SD includes clinical tolerance 2 × 2SD, 1 × 3SD, 1 × 3.5SD Yes, within lab site No Yes
23 SPH, PC Yes 6-Sigma based Yes, based on BV SD includes clinical tolerance 1 × 2SD, 1 × 3SD, R × 4SD Yes, within lab site Yes, PMA Yes
22 SPH, PC Yes Other (incl. 6-Sigma) Yes, based on BV For each new QC material 1 × 2SD, 2 × 2SD, 1 × 2.5SD, 1 × 3SD, R × 4SD, 10 × 1SD No Yes, EQA Daily QC graph check
19 UMC, PC Yes Other (incl. 6-Sigma) Yes, based on BV For each new QC material 1 × 2SD, trend Yes, within lab site Yes, PMA Daily QC graph check
11 MPH Yes Other (incl. 6-Sigma) Yes, based on BV SD includes clinical tolerance 2 × 2SD, 1 × 3SD, 1 × 3.5SD, R 4SD, 7 × 1SD Yes, applied for >1 lab sites Yes, PMA Daily QC graph check
17 PC Yes Other (incl. 6-Sigma) Yes, based on BV SD includes clinical tolerance 1× 2SD, 2 × 2SD, 1 × 3SD, 1 × 3.5SD, R 4SD, trend Yes, within lab site Yes, PMA Daily QC graph check
7 SPH, PC Yes Risk-based No For each new QC material 2× 2SD, 1× 3SD, R 4SD Yes, within lab site No Daily QC graph check
26 Others Yes Risk-based Yes, based on BV For each new QC material 2 × 2SD, 1 × 2.5SD, 1 × 3SD, R × 4SD Yes, within lab site Yes, PMA Yes
4 MPH, PC Yes Risk-based Yes, based on BV Historical SD or CV(%) 2× 2SD, 1× 3SD, 1× 3.5SD, R 4SD, 10 × 1SD Yes, within lab site Yes, PMA Daily QC graph check
24 SPH Yes Risk-based Yes, based on BV Historical SD or CV(%) 2 × 2SD, 1 × 2.5SD, 1 × 3SD, 1 × 3.5SD, 4 × 1SD Yes, within lab site Yes, PMA Yes
6 SPH, PC Yes Risk-based Yes, based on BV SD includes clinical tolerance 2× 2SD, 1× 3SD, 4× 1SD No Yes, PMA peer-comp. Daily QC graph check
  1. MPH, multiple peripheral hospitals; SPH, single peripheral hospital; PC, primary care; UMC, university medical center; BV, biological variation; EL, expert level; PMA, patient moving average (any kind of algorithm e.g. EWMA, AoN, Bulls (XbarM/B), etc.), EQA, external quality assessment, awithin and between QC materials.

Adaption of the virtual analyzer methodology

All responding laboratories had measurement procedures available on more than one analyzer system. A minority of 3/27 (11 %) configured the iQC and especially the SD individually for each analyzer system; 18/27 (67 %) of the laboratories configured the iQC of identical measurement procedures on multiple analyzers within a single laboratory location identically, and 6/27 (22 %) of the laboratory organizations had identical measurement procedures in operation at different laboratory locations and use an identical iQC configuration for all of them. Overall, 24/27 (89 %) of the respondents used some type of “virtual analyzer model” (see discussion for definition), as the basis for the iQC configuration.

Determination of SD and target values

Of the respondents 9/26 (35 %) establish new and realistic SD values for each new QC material on one or more analyzers, 9/26 (35 %) of respondents base the SD value of new QC materials on historically obtained SD (or CV%) values. Here, SD may include some longer-term effects such as assay drift and variance introduced by method calibration. SD values based not only on the analytical characteristics, but also on some sort of an estimate of the clinical tolerance were used by 8/26 (31 %). By using this approach, the SD values can be set wider and do not necessarily reflect the true operational method imprecision.

Use of additional QA tools for iQC other than SQC

9/27 (33 %) of the respondents indicated that they do not use other QA systems to assure the analytical quality. 18/27 (67 %) of the respondents indicate the use of additional tools, with 11/27 (41 %) indicating the use of some sort of patient moving averages (including XbarM of Bulls algorithm for hemocytometry analysis) for one or more measurement procedures [7]. Other additional tools include the use of EQA and its assessment, peer-to-peer comparison of iQC results provided by vendors, and a periodic population assessment using Bhattacharya analysis. In addition, 18/27 (67 %) indicate that they do not use measurement uncertainty to design iQC procedures, while 9/27 (33 %) do use the measurement uncertainty to design iQC procedures.

Daily management of iQC

Table 3 shows the type of software systems used for iQC management. In the Netherlands the majority of the laboratories use their LIS system (GLIMS, Labosys or MOLIS). Furthermore, the data show that 13/27 (48 %) laboratories use several separate systems to manage SQC. When asked how routinely performed iQC measures are evaluated, 12/27 (44 %) use some sort of “traffic light” that only displays or alerts for those iQC or assays for which SQC control rules have exceeded. 15/27 (56 %) stated that all QC are manually checked using the QC graph and SQC control rules are used for alarms. 0/27 (0 %) of the respondents indicated the practice of interpreting the QC graph on a daily basis without using any SQC control rule. Finally, when iQC has identified a (critical) error that requires a recall procedure, it was asked whether consultation of a supervisor or director is required. 18/27 (67 %) of the respondents indicated that this procedure is initiated only after consultation of a supervisor or director and 9/27 (33 %) indicated that this is a strict protocol and consultation of a supervisor or director to start the procedure is not required.

Table 3:

Software used for iQC management.

Lab ID LIS Middleware Analyzer 3rd party software
1 X X
2 X
3 X X
4 X X X X
5 X X
6 X
7 X
8 X X
9 X
10 X X X X
11 X
12 X
13 X X X X
14 X
15 X X
16 X X
17 X X
18 X
19 X X
20 X X X
21 X X X
22 X
23 X
24 X
25 X
26 X
27 X

Long-term management of iQC

Figure 1 shows the results of 3 questions regarding the long-term management of iQC; who evaluates and manages the long-term iQC configuration, is there a separate dedicated evaluation of the iQC settings, and how are between-analyzer differences of the same measurement procedure controlled. For the latter, several laboratories mentioned that EQA analysis is also used for this purpose.

Figure 1: 
Questions related to the long(er)-term iQC management.
Figure 1:

Questions related to the long(er)-term iQC management.

Training and guidance on iQC

The final questions of the questionnaire were related to the knowledge of new developments in technical quality assurance and control and the need for further training and education. 16/27 (59 %) indicated that they were not familiar with the newer iQC developments such as patient-based real-time QC [5], risk-based QC [3] and the more recently introduced precision QC [8]. In addition, 20/27 (77 %) indicated a need for additional training related to the new developments in the field of analytical QC. The final question was an open-ended question about the need to include specific topics in guidance documents related to analytical QC. 11/27 (41 %) indicated one or more topics that they would like to see included in such guidance documents. Suggested topics included: the need for practical tools and a pragmatic approach, the relationship with ISO 15189 and the formulation of minimum requirements, the relationship with EQA, the “freedom” to use the expertise of staff and specialists in laboratory medicine to differentiate from guidance documents and the calculation method for measurement uncertainty. One participant further suggested to share best practices among the laboratory medicine community.

Discussion

The questionnaire provided relevant insights into the iQC practices performed in the Netherlands. In particular the great heterogeneity of the iQC setup applied in the medical laboratories was considered remarkable. The obtained answers provided some key aspects that are relevant to provide guidance. Furthermore, for some practices performed by the laboratories, such as the determination of SD values, the virtual analyzer methodology and the assurance of between analyzer comparisons, there seems to be a need for additional research to provide the necessary additional insights into best practices.

Recently, several other surveys on the international internal QC practices have been conducted and published. These include a global survey initiated by the IFCC (n=46) [9], a survey amongst academic medical centers in the US (n=21) [10], and by far the largest survey conducted by Westgard QC and published on their website (n=619) [11]. In general, these surveys compared iQC practices around the world and focused on more general aspects, such as, the source of QC materials, the use of Westgard-rules, and QC frequency without providing a methodological basis thereof. The general findings of the IFCC and Westgard surveys were also a large heterogeneity of iQC practices. Both the US-based survey and the Westgard survey showed that within the US, 2SD control used were overwhelmingly used as single iQC rule, although the QC frequency could vary widely [10, 11]. A limitation of these studies was that no information was provided on the methodology and system used to design the SQC. In addition, also more practical details were missing, such as whether a virtual analyzer methodology was used, and what SD values represent and how they are obtained. As this questionnaire was primarily designed to address such practical issues that may be addressed in the guidance document, the questionnaire focused significantly on such more detailed practices. In comparison to the 2SD control rule observation in other surveys, in this survey 14 laboratories reported the use of 1 × 2SD rules; 13 of these 14 also used other SD rules and only a single laboratory responded 1 × 2SD to be the single SQC rule used.

The questionnaire was sent to all medical laboratories in the Netherlands and only a (43 %) minority of the laboratories responded to the request and completed the questionnaire. The responses may be biased in the sense that the laboratories that responded may have a higher than average interest in iQC and have a higher than average iQC standard and use more advanced iQC methods. This is considered a limitation of this study and therefore the results may not accurately represent the true general iQC practice as performed in the Netherlands. There was a representation of different types of medical laboratories, such as those supporting university medical centers, general hospitals and primary care. In addition, laboratories with both single and multiple laboratory sites were included. Therefore, although the results may overestimate the use of advanced iQC methods, the diversity in respondents and the answers to the questionnaire reveal large heterogeneity in the SQC design.

Approximately 30 % of the laboratories base their iQC setup solely on the analytical characteristics of assays and do not apply clinical tolerance limits. This indicates that the use of clinical tolerance limits and the incorporation of patient risk into SQC design has been adopted by the majority of the participating clinical laboratories. Analytical quality control is an explicit requirement for ISO 15189 accreditation. The prevailing ISO 15189:2012 during the questionnaire does not specify into great detail the way iQC is performed. However, the recently published version (ISO 15189:2022) explicitly requires consideration of the intended clinical application of the examination, as the performance specifications can differ in different clinical settings [12]. The 30 % laboratories which base their iQC setup solely on the analytical characteristics of assays will have to change their iQC setup to fulfill the requirements of the revised ISO 15189.

Of particular interest to the authors was the high adoption of any type of virtual analyzer system methodology for establishing and using SQC settings by 89 % of the laboratories. The “virtual analyzer” was considered as a theoretical analyzer that is comprised of the individual analyzers that operate within it. Within the virtual analyzer model, both the uncertainty of the individual analyzers employed and the variability among these instruments are used to define the uncertainty of the virtual analyzer. Unfortunately, no additional in-depth questions about these practices were included in the questionnaire.

When asked about the method used to determine SD and target values, a variety of practices was used. Unfortunately, there is little research and there are no prospective user studies available on how to set, obtain and manage SD and target values on individual analyzers when using a virtual analyzer setup. However, some have recently started to provide more advanced work on this topic [13, 14]. It seems relevant to explore the pros and cons of different approaches and to share best practices. Another interesting observation was the high percentage of alternative systems to support the SQC. In particularly the adoption of patient moving averages or PBRTQC for one or more analytes in 41 % of the clinical laboratories was much higher than the previously reported adoption rate of 10 % in US academic medical centers [10]. In addition, the adaption of peer-to-peer review of iQC results from different vendors has gained momentum. However, how laboratories use these results in the context of EQA to determine true error or bias seems relevant, as peer-to-peer review can allow for much more frequent peer comparison than EQA.

When reviewing the management aspects of iQC, there is also a wide variation in practices. Some use a traffic light system to identify out-of-control situations, while others manually review all QC charts. Furthermore, although most use their LIS to manage iQC, about half of the clinical laboratories use more than one software system to manage QC results. Some even use all of the software systems mentioned (LIS, middleware, analyzer software and 3rd party software) to manage iQC. The latter primarily seems to reflect practices at the UMCs and may be caused by greater specialization and separation of the various, more independent, specialized laboratories as part of the overall clinical laboratory. Also, the management of longer term settings and the evaluation of the settings are highly variable and seem to depend largely on the daily management set-up. Finally, the methods used to ensure the comparability of analyzers also appear to be variable, with some using iQC measures and others using a patient sample to ensure the use of commutable materials for this purpose. Some report using moving averages.

In conclusion, the questionnaire provided many relevant insights into the QC practices applied in clinical laboratories in the Netherlands. In particular, it has provided some relevant starting points for further research and guidance, as for some of these practices there is little or no evidence and guidance available. The most relevant points according to the authors are listed in Table 4. This overview should provide some guidance for the necessary research and perhaps the sharing of some best practices.

Table 4:

Topics for future research and guidance.

No. Topic Research statusa Guidancea
1 Practices to determine SD and target values for individual and/or virtual analyzer systems N.A. CLSI documents. However not for virtual analyzer methodology and only when SD is based solely on analytical characteristics
2 Set-up and management of virtual analyzer systems on single and multiple laboratory sites Some casus, not prospectively validated Not available, only shared best practices
3 Additional requirements when SD are not solely based on analytical characteristics Not available No guidance available for such practice
4 Practices to ensure analyzer comparability Some casus Not available, only shared best practices
5 Use of peer-to-peer comparison tools and status with respect to EQA Not available No guidance available
  1. aTo the best knowledge of the authors.


Corresponding author: Huub H. van Rossum, Department of Laboratory Medicine, Netherlands Cancer Institute, Plesmanlaan 121, Amsterdam, The Netherlands, E-mail:

  1. Research ethics: Not applicable.

  2. Informed consent: Not applicable.

  3. Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  4. Competing interests: H. van Rossum is owner and director of Huvaros BV. All other authors state no conflict of interest.

  5. Research funding: None declared.

  6. Data availability: Not applicable.

References

1. van Rossum, HH. Technical quality assurance and quality control for medical laboratories: a review and proposal of a new concept to obtain integrated and validated QA/QC plans. Crit Rev Clin Lab Sci 2022;59:586–600. https://doi.org/10.1080/10408363.2022.2088685.Search in Google Scholar PubMed

2. Westgard, JO. Six sigma quality design & control. Madison WI: Westgard QC, Inc.; 2006.Search in Google Scholar

3. Westgard, JO, Bayat, H, Westgard, SA. Planning risk-based SQC schedules for bracketed operation of continuous production analyzers. Clin Chem 2018;64:289–96. https://doi.org/10.1373/clinchem.2017.278291.Search in Google Scholar PubMed

4. Bayat, H. Selecting multi-rule quality control procedures based on patient risk. Clin Chem Lab Med 2017;55:1702–8. https://doi.org/10.1515/cclm-2016-1077.Search in Google Scholar PubMed

5. Badrick, T, Bietenbeck, A, Cervinski, MA, Katayev, A, van Rossum, HH, Loh, TP. Patient-based real-time quality control: review and recommendations. Clin Chem 2019;65:962–71. https://doi.org/10.1373/clinchem.2019.305482.Search in Google Scholar PubMed

6. van Rossum, HH, Bietenbeck, A, Cervinski, MA, Katayev, A, Loh, TP, Badrick, TC. Benefits, limitations, and controversies on patient-based real-time quality control (PBRTQC) and the evidence behind the practice. Clin Chem Lab Med 2021;59:1213–20. https://doi.org/10.1515/cclm-2021-0072.Search in Google Scholar PubMed

7. Bull, BS, Elashoff, RM, Heilbron, DC, Couperus, J. A study of various estimators for the derivation of quality control procedures from patient erythrocyte indices. Am J Clin Pathol 1974;61:473–81. https://doi.org/10.1093/ajcp/61.4.473.Search in Google Scholar PubMed

8. Schmidt, RL, Moore, RA, Walker, BS, Rudolf, JW. Precision quality control: a dynamic model for risk-based analysis of analytical quality. Clin Chem Lab Med 2023;61:679–87. https://doi.org/10.1515/cclm-2022-1094.Search in Google Scholar PubMed

9. Wheeler, SE, Blasutig, IM, Dabla, PK, Giannoli, JM, Vassault, A, Lin, J, et al.. Quality standards and internal quality control practices in medical laboratories: an IFCC global survey of member societies. Clin Chem Lab Med 2023;61:2094–101. https://doi.org/10.1515/cclm-2023-0492.Search in Google Scholar PubMed

10. Rosenbaum, MW, Flood, JG, Melanson, SEF, Baumann, NA, Marzinke, MA, Rai, AJ, et al.. Quality control practices for chemistry and immunochemistry in a cohort of 21 large academic medical centers. Am J Clin Pathol 2018;150:96–104. https://doi.org/10.1093/ajcp/aqy033.Search in Google Scholar PubMed

11. Westgard, S. The 2021 great global QC survay results; 2023. https://www.westgard.com/2021-global-qc-survey-results.htm [Accessed 30 Nov 2023].Search in Google Scholar

12. Internation Organization for Standardization. ISO 15189:2022 medical laboratories – requirements for quality and competence. Geneva, Switzerland: ISO; 2022.Search in Google Scholar

13. Badrick, T, Graham, P. Can a combination of average of normals and “real time” external quality assurance replace internal quality control? Clin Chem Lab Med 2018;56:549–53. https://doi.org/10.1515/cclm-2017-0115.Search in Google Scholar PubMed

14. Giannoli, JM, Bernard, M, L’Hirondel, J, Heim, A, Badrick, T. A model for managing quality control for a network of clinical chemistry instruments measuring the same analyte. Clin Chem Lab Med 2024;63:853–60. https://doi.org/10.1515/cclm-2023-0965.Search in Google Scholar PubMed


Supplementary Material

This article contains supplementary material (https://doi.org/10.1515/cclm-2024-0316).


Received: 2024-03-08
Accepted: 2024-04-12
Published Online: 2024-04-26
Published in Print: 2024-10-28

© 2024 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. Circulating tumor DNA measurement: a new pillar of medical oncology?
  4. Reviews
  5. Circulating tumor DNA: current implementation issues and future challenges for clinical utility
  6. Circulating tumor DNA methylation: a promising clinical tool for cancer diagnosis and management
  7. Opinion Papers
  8. The final part of the CRESS trilogy – how to evaluate the quality of stability studies
  9. The impact of physiological variations on personalized reference intervals and decision limits: an in-depth analysis
  10. Computational pathology: an evolving concept
  11. Perspectives
  12. Dynamic mirroring: unveiling the role of digital twins, artificial intelligence and synthetic data for personalized medicine in laboratory medicine
  13. General Clinical Chemistry and Laboratory Medicine
  14. Macroprolactin in mothers and their babies: what is its origin?
  15. The influence of undetected hemolysis on POCT potassium results in the emergency department
  16. Quality control in the Netherlands; todays practices and starting points for guidance and future research
  17. QC Constellation: a cutting-edge solution for risk and patient-based quality control in clinical laboratories
  18. OILVEQ: an Italian external quality control scheme for cannabinoids analysis in galenic preparations of cannabis oil
  19. Using Bland-Altman plot-based harmonization algorithm to optimize the harmonization for immunoassays
  20. Comparison of a two-step Tempus600 hub solution single-tube vs. container-based, one-step pneumatic transport system
  21. Evaluating the HYDRASHIFT 2/4 Daratumumab assay: a powerful approach to assess treatment response in multiple myeloma
  22. Insight into the status of plasma renin and aldosterone measurement: findings from 526 clinical laboratories in China
  23. Reference Values and Biological Variations
  24. Reference values for plasma and urine trace elements in a Swiss population-based cohort
  25. Stimulating thyrotropin receptor antibodies in early pregnancy
  26. Within- and between-subject biological variation estimates for the enumeration of lymphocyte deep immunophenotyping and monocyte subsets
  27. Diurnal and day-to-day biological variation of salivary cortisol and cortisone
  28. Web-accessible critical limits and critical values for urgent clinician notification
  29. Cancer Diagnostics
  30. Thyroglobulin measurement is the most powerful outcome predictor in differentiated thyroid cancer: a decision tree analysis in a European multicenter series
  31. Cardiovascular Diseases
  32. Interaction of heparin with human cardiac troponin complex and its influence on the immunodetection of troponins in human blood samples
  33. Diagnostic performance of a point of care high-sensitivity cardiac troponin I assay and single measurement evaluation to rule out and rule in acute coronary syndrome
  34. Corrigendum
  35. Reference intervals of 24 trace elements in blood, plasma and erythrocytes for the Slovenian adult population
  36. Letters to the Editor
  37. Disturbances of calcium, magnesium, and phosphate homeostasis: incidence, probable causes, and outcome
  38. Validation of the enhanced liver fibrosis (ELF)-test in heparinized and EDTA plasma for use in reflex testing algorithms for metabolic dysfunction-associated steatotic liver disease (MASLD)
  39. Detection of urinary foam cells diagnosing the XGP with thrombopenia preoperatively: a case report
  40. Methemoglobinemia after sodium nitrite poisoning: what blood gas analysis tells us (and what it might not)
  41. Novel thiopurine S-methyltransferase (TPMT) variant identified in Malay individuals
  42. Congress Abstracts
  43. 56th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)
Downloaded on 4.2.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2024-0316/html
Scroll to top button