Cell counting chamber vs. Sysmex XN-1000 for determining white blood cell count and differentiation for body fluids

  • Ebru Evren ORCID logo EMAIL logo , Ebru Us ORCID logo , Can Hüseyin Hekimoğlu ORCID logo and Zeynep Ceren Karahan ORCID logo
Published/Copyright: September 19, 2022

Abstract

Objectives

Sterile body fluids (BFs) include key information for the diagnosis and monitoring of a variety of diseases. A cornerstone test is the total white blood cell (WBC) count, which comprises the differential of WBC of body fluid analysis. It is important to test automated hematology analyzers that should be verified using patient samples. The aim of this study is to compare both the performance of the Sysmex XN-1000 system’s body fluid module for cell counting and differentiation to the results of a cell counting chamber.

Methods

This study was performed on 200 routinely laboratory sent BFs. Cell counts and differentiation were determined with both bright-lined Neubauer Cell Counting Chamber and Sysmex XN-1000 system body fluid mode.

Results

The correlation coefficients of WBC count by two methods indicated very strong correlation (r≥0.90, p<0.0001) for any specimen except pleural fluid. According to Passing-Bablok regression analysis, Sysmex XN-1000 showed acceptable performance according to bias of the slope criteria (<±20%).

Conclusions

The XN-1000 hematological analyzer’s body fluid mode can rapidly count and identify cells, and it can be used as a simple and rapid screening method for laboratory testing of sterile body fluids, particularly in laboratories with a massive quantity of biological fluids.

Introduction

Biological or sterile body fluids (BFs) provide crucial information for the diagnosis and monitoring of a wide range of disorders, as well as for determining the best therapy options. The total white blood cell (WBC) count, including differential of WBC and malignant cell characterization of body fluid analysis, is regarded a cornerstone test in patients with inflammatory, infectious, and neoplastic disorders. Significant treatment parameters include the amount of white blood cells (WBCs), as well as their differentiation into mononuclear (MN) or polymorphonuclear leukocytes (PMNL) [1, 2].

WBCs play a crucial role as a part of immune system against pathogens. Neutrophils are the first line defense mechanisms especially against bacterial infections. Antigens are recognized by lymphocytes, and monocytes have a variety of activities in the immune system, such as transforming into phagocytic macrophages. Numerous cells can be found in BFs and these cells reflect important informations for inflammatory diseases. Currently, manual microscopy, automated flow cytometry, and automated impedance technology for counting and differentiating WBCs in BFs are used. Conventional manual microscopy is still the gold standard, although automated analyzers have become prevalent in clinical laboratories in the recent decade. The hemocytometer, commonly known as a “counting chamber” is commonly used to determine the total WBC and RBC concentrations in BFs. For determining the MNs and PMNs of WBC a stained slide is prepared and investigated under light microscope [3, 4].

Although conventional manual microscopy is still the gold standard, and commonly performed in medical microbiology laboratories, it has some limitations such as samples showing low cellularity may be highly inaccurate. While the accuracy of the count depends on highly qualified laboratory staff and as a con it is time consuming [4].

For the cytometric analysis of biological fluids, a variety of commercial automated hematological analyzers that can also count sterile body fluids which requires special modules have been on the market in recent years. Automated systems provide two key benefits over manual systems: faster response times and reduced interobserver variability. Additionally, their main advantage is that they can even analyze small sample sizes [5].

The International Standards Council for Hematology (ICSH) pointed out that the performance of an automated hematology analyzers (AHA) should be verified using patient samples [1, 6].

Although studies on cell counting for blood are widely performed, studies evaluating cell count and differentiation especially in body fluids (other than blood and urine) in automated hematology analyzers are limited [7, 8].

The aim of this study is to compare both the performance of the Sysmex XN-1000 system’s body fluid module for cell counting and differentiation to the results of a cell counting chamber.

Materials and methods

Samples

This study was performed on 200 routinely collected sterile body fluid samples from hospitalized- and out-patients. These samples comprised of 93 (46.5%) synovial fluid, 78 (39%) CSF, 22 (11%) peritoneal fluid, 7 (3.5%) pleural fluid samples. CSF samples were collected in sterile tubes while others were collected in tubes coated with K2EDTA (Becton Dickinson, USA). Macroscopically clotted, viscous, mucoid, cloudy, and bloody samples were not enrolled in this study. The BFs were transported to the laboratory at ambient temperature in 1 h. All samples were examined within 1 h after being received to the laboratory.

Cell counts and differentiations were simultaneously determined with both bright lined Neubauer Cell Counting Chamber® (Marienfeld, Germany) and Sysmex XN-1000 (Sysmex, Japan) hematology analyzer using body fluid mode according to the manufacturers’ recommendations. Only the samples with cell counts within the analytical limits (0.03–440.00 103xcell/µL) were included in the study, and no dilutions were applied. For the Sysmex XN-1000 analyzer, three levels of whole blood internal quality control materials provided by the manufacturer were analyzed daily. The analytical coefficients of variations (CVs) for the study period were as follows: 2.31% for level 1 (mean = 3.07 103 × cell/µL), 1.56% for level 2 (mean = 7.36 103 × cell/µL) and 1.19% for level 3 (mean = 16.6 103 × cell/µL) for WBC count. For neutrophil, lymphocyte, monocyte, eosinophil, and basophil differential percentage CVs were under 8%. During the study period Randox Quality Control (Country Antrim, United Kingdom) external quality assurance (EQA) program was used and all bias values within the limits of the EQA program.

Manual microscopic examination was performed with light microscope using an eyepiece set at 10× and a lens set at 40×, for a total magnification of 400×. All samples were evaluated by the same medical microbiology specialist. Differential counting of 200 samples was performed centrifugation at 300 g, for 5 min and stained with Wright-Giemsa staining. The differential count as polymorphonuclear leukocytes (PMNL) was performed at 400× magnification on 100 cells per each sample. The Sysmex XN-1000 was used in body fluid mode both for the counting of cells and percentage of PMNL. Cell count was evaluated in cells per milliliter and cell distribution was expressed as percentage distribution.

Ethical statement

Ethical approval was waived by the local Ethics Committee of Ankara University School of Medicine in view of the retrospective nature of the study (I07-409-22).

Statistical analysis

Shapiro-Wilk test was performed and showed that the distribution of results departed significantly from normality (p<0.001) Therefore, non-parametric tests were used, and the median with 25th and 75th percentile were given for descriptive statistics.

For method comparison, the Spearman correlation coefficient [9] was calculated and Passing-Bablok linear regression analysis was performed. The Bland-Altman plots were also utilized, and absolute mean differences were calculated. Percentage bias of the Passing-Bablok equation slope were used to assess the agreement between the methods and ±20% was used as the agreement limit as suggested by Cho et al. [10]. p-Values < 0.05 were considered significant. The 95% confidence intervals (CIs) of each test were also calculated.

The Passing‐Bablok regression analysis and the Bland‐Altman deviation analysis were performed using Microsoft Excel 2010 (Microsoft Corp., Redmond, Washington, USA), and Jamovi software. Descriptive statistics were calculated using IBM SPSS Statistics for Windows, version 25.0 (IBM Corp., Armonk, NY, USA).

Results

In our study, 200 body fluid samples were counted in the manual method and XN-1000 series body fluid mode. Descriptive statistics of the cell enumerations for each specimen type were given in the Table 1.

Table 1:

Descriptive statistics of the cell count and types for each specimen type.

Manual method Sysmex XN-1000
Min Max Median Percentile Min Max Median Percentile
25th 75th 25th 75th
CSF (n=78) WBC (/µL) 0 1.200 30 3 100 0 1.270 43 4 116
PMNL (%) 0 100 10 0 60 0 100 26 0 63
MNL (%) 0 100 24 0 70 0 100 37 0 74
Synovial fluid (n=92) WBC (/µL) 20 88.800 2.000 280 5.700 23 73.600 1.890 400 6.360
PMNL (%) 0 100 65 20 86 2 98 72 27 89
MNL (%) 0 100 35 14 80 2 98 28 11 70
Peritoneal fluid (n=22) WBC (/µL) 0 88.000 300 166 1.540 2 78.000 321 242 1.720
PMNL (%) 0 98 20 10 34 0 95 20 14 40
MNL (%) 0 100 78 61 90 0 97 76 42 86
Pleural fluid (n=7) WBC (/µL) 125 1.000 150 129 275 133 1.100 230 200 323
PMNL (%) 0 97 1 0 50 6 88 8 7 45
MNL (%) 3 100 99 50 100 12 94 92 55 93
  1. CI, confidence interval; CSF, cerebrospinal fluid; Min, minimum; Max, maximum.

Method comparison results for all specimen types are summarized in Table 2. For CSF and synovial fluids, Passing-Bablok and Bland-Altman plots are shown in Figures 1 and 2, respectively. For other fluid types related plots are provided in the Supplementary Material.

Table 2:

Method comparison results for each specimen type.

Spearman’s correlation Passing-Bablok

Regression
Bland-Altman plot
p r (95% CI) Slope (95% CI) Intercept (%95 CI) % Bias of the slope Mean absolute difference
CSF WBC (/µL) <0.0001 0.96 (0.94–0.98) 1.13 (1.05–1.23) 0.38 (0.00–2.65) 13.0 −17.3
PMNL (%) <0.0001 0.75 (0.64–0.84) 1.00 (0.95–1.07) 0.00 (0.00–2.09) 0 −5.1
  MNL (%) <0.0001 0.72 (0.60–0.82) 0.97 (0.92–1.00) 0.00 (0.00–1.17) 3.00 −5.4
Synovial fluid WBC (/µL) <0.0001 0.95 (0.93–0.97) 0.99 (0.94–1.02) 35.2 (7.77–95.25) 1.00 279
PMNL (%) <0.0001 0.95 (0.92–0.97) 0.94 (0.86–0.99) 6.68 (2.96–10.50) 6.00 −4.26
  MNL (%) <0.0001 0.95 (0.92–0.97) 0.93 (0.87–0.98) −0.33 (−1.84–1.71) 7.00 4,70
Peritoneal fluid WBC (/µL) <0.0001 0.92 (0.83–0.97) 0.89 (0.71–1.09) 52.27 (−0.31–108) 11.0 969
PMNL (%) <0.0001 0.90 (0.78–0.96) 1.00 (0.85–1.28) 1.00 (−4.69–4.77) 0 −1.59
  MNL (%) <0.0001 0.91 (0.79–0.96) 0.98 (0.84–1.13) 1.43 (−8.33–11.00) 2.00 1.14
Pleural fluid WBC (/µL) 0.0744 0.71 (0.094–0.95) 1.10 (n/c) 18.6 (n/c) 10.0 −60.9
PMNL (%) 0.0335 0.79 (0.098–0.97) 0.83 (n/c) 6.0 (n/c) 17.0 −1.14
  MNL (%) 0.0335 0.79 (0.098–0.97) 0.82 (n/c) 11.19 (n/c) 18.0 1.14
  1. CI, confidence interval; CSF, cerebrospinal fluid; r, Spearman’s correlation coefficient; n/c, confidence intervals cannot be calculated due to low sample count.

Figure 1: 
Passing-bablok plots for the CSF and synovial fluids. CSF, cerebrospinal fluid; MNL, mononuclear leukocytes; PMNL, polymorphonuclear leukocytes; WBC, white blood cells. Identity lines (y=x) are gray, confidence intervals are dashed red and regression lines are blue.
Figure 1:

Passing-bablok plots for the CSF and synovial fluids. CSF, cerebrospinal fluid; MNL, mononuclear leukocytes; PMNL, polymorphonuclear leukocytes; WBC, white blood cells. Identity lines (y=x) are gray, confidence intervals are dashed red and regression lines are blue.

Figure 2: 
Bland-altman plots for the CSF and synovial fluids. CSF, cerebrospinal fluid; MNL, mononuclear leukocytes; PMNL, polymorphonuclear leukocytes; WBC, white blood cells. Mean absolute differences are blue, 95% limits of agreement are dashed red, confidence intervals are cyan.
Figure 2:

Bland-altman plots for the CSF and synovial fluids. CSF, cerebrospinal fluid; MNL, mononuclear leukocytes; PMNL, polymorphonuclear leukocytes; WBC, white blood cells. Mean absolute differences are blue, 95% limits of agreement are dashed red, confidence intervals are cyan.

According to the results of Spearman’s correlation results for WBC cell enumeration, there were statistically significant (p<0.0001) and very strong correlation (r≥0.90) between the manual chamber counting and the XN-1000 BF mode for all BF types except pleural fluid samples. The correlation for WBC count for pleural fluid is strong although p value is >0.05 due to possible unsufficient sample size (r=0.71, p=0.0744). When Passing-Bablok regression results were evaluated, Sysmex XN-1000 demonstrated acceptable performance according to bias of the slope criteria and all percentage bias of the slope values were under 20%. Moreover, for CSF and peritoneal fluids PMNL and MNL results met the Passing-Bablok method agreement criteria (95% CI for the intercept, covers zero, and CI for the slope, covers one). However, for the synovial fluid WBC count results CIs covered one for the slope but does not cover zero for the intercept.

The Bland-Altman plots show that Sysmex XN-1000 presented slightly higher cell count results compared to the manual method and for the CSF samples. Additionally, there was no apparent increasing or decreasing trend for any BF type in Bland Altman plots.

Discussion

In the current study, the cell count and differentiation of PMNL in significant clinical sterile BFs other than urine and blood were examined using the Sysmex XN-1000 BF mode and a conventional cell counting chamber. Our results showed that Sysmex XN-1000 had comparable results with manual WBC and differential count which is considered as gold standard method for cell enumeration.

This finding is supported by evaluation of the Passing-Bablok regression results and Bland-Altman analysis. For all BF types and test results, % bias of the regression analysis slope values were lower than predefined acceptability criteria (within 20%). Furthermore, Bland-Altman analysis showed relatively low bias for differential counting (1.14–5.4%).

Owing to lack of data in this field, it is critical to assess the performance of various automated cell counting and differential technologies [7]. Lee et al. reported that for cerebrospinal fluid, ascitic fluid, and pleural fluid sample types the XN-350s cell enumeration and differential counting correlated well with manual chamber counting (except for differential counting in CSF samples). For WBC count in CSF, they reported −16 cell/µL absolute bias with the 95% CI of −36 to 2 [5]. These results were similar with our finding (Absolute bias for WBC = −17 cell/µL). Our cell differential count for PMNL and MNL had similar but lower absolute bias compared to their results (−5.1% vs.–16% for PMNL and −5.4 vs. 27 for MNL).

Lohajaroensub et al. has also reported that except for synovial fluid, the Sysmex XT-4000i performed well in terms of basic body fluids assessment [11]. Evren et al. have reported that for macroscopically purulent samples, counting chambers would still be the best choice for WBC counting, however for macroscopically non-purulent samples Sysmex UF-1000i is a good alternative instrument for rapid workflow [12]. Moreover, these systems even enable the detection of malignant cells in serous body fluids. However, it is emphasized that research on this area still needs to be evaluated [13].

Total WBC and differential count by XN-1000 had a high correlation with the conventional manual microscopy approach. Due to the high cellular content of the CSF samples evaluated in this study, drawing inferences about low cellularities was challenging. Cell differentiation may be insufficient in general due to the low cellularity of CSF samples, particularly for samples with low cellular content.

A differential cell count for sterile body fluids indicates the proportion of each type of white blood cell and aids in guiding therapy. For this purpose, modern hematology analyzers combine multiple techniques to evaluate cell counting and subgroups percentage of leucocytes, as well. Identifying the types of cells present in sterile body fluids such as joint fluid, CSF, pleural and peritoneal fluids except urine and blood provides valuable diagnostic information. For this reason, the analysis of sterile body fluids is important for rapid presumptive diagnosis. The laboratory analysis of these samples should include both cell counting and a differential cell count. The gold standard method for cell counting is the cell counting chambers while the differential cell count should be performed on a stained smear [5, 14].

Hematology analyzers were designed to measure cells in whole blood at first. Automated cell counters have led to significant improvements in BF analysis despite the fact that their use is limited and not widely preferred since their inception in clinical laboratories, and these have received increasing acceptance as a crucial alternative to manual microscopy. WBCs in blood samples, on the other hand, are at least 1,000 times higher than in BF samples, making these analyzers unsuitable for BF analysis due to high imprecision in the lower concentration range [15, 16].

However, in recent years there are a lot of manufacturers on the market with BF cell counting tools (hematology and urine), and several manufacturers have increased their technological capabilities by incorporating specific BF applications into their automated cell counters. Sysmex’s latest hematology analyzer, the XN-Series, was launched in 2011. Sysmex analyzers such as XE-5000 and XN-series include separate BF modes and are fully automated systems that measures a variety of BFs and counts. Urinalysis analyzers, such as the Sysmex UF-Series and Iris iQ200, have risen in favor for BF analysis in addition to hematological analyzers. The Sysmex XN-1000 hematology analyzer is a fully automated flow cytometric system with validated body fluid modules for the evaluation of cellular contents of sterile body fluids. In this study, we have evaluated the performance of the body fluid module of Sysmex XN-1000 system in comparison with manual methods.

To date, there is a debate concerning whether to utilize automated cell counters or manual microscopy for BF cell counts and evaluate percentage of cellular subgroups. Many of their advantages, such as improved proficiency and workflow productivity, advantage of being able to assess small sample quantities like CSF have been proven in many studies, and they are recommended as acceptable alternative to the conventional microscopy [3, 14, 17], [18], [19], [20], [21].

Despite the benefits of hematology analyzers, considerable quantities of epithelial cells and macrophages in the PMNL count, making automated differentiation unreliable depending on the brand of the analyzer [18, 22, 23]. Additionally, overestimation of PMNL counts with low WBC counts was shown to be associated with certain cell debris or fragments in CSF samples that usually contain very few cells. We shouldn’t evaluate the low counts for CSF in our study since the mean cell count was around 200 cells/μL. In routine practice, these tests can be performed both by clinical biochemistry and microbiology experts. On the biochemistry side, the tests are mostly performed by automated systems such as hematology analyzers, while manual procedures are most often used on the microbiology side. Although manual procedures are mostly preferred, it is well known that these methods are time consuming and labor intensive. The use of automated systems in clinical microbiology laboratories has recently gained attention as an alternative method to microscopic examination.

One limitation of this study is related with evaluation of the calculated biases. In contrast to complete blood count analysis, body fluids do not have well defined analytical performance specifications. This is the most important challenge related to the body fluid method comparison studies. To overcome this problem, we used the predefined criteria recommended by Cho et al. [10]. Other limitation is, in the study body fluid specific control materials was not used and only commercial materials were evaluated. Final limitation of this study was the number of BF samples. Only 7 pleural fluid and 22 peritoneal fluid samples were included in the study which is lower than minimal requirement of 40 samples [23].

As a conclusion, clinical laboratories want to use automated and standardized analysis methods to reduce turnaround time and variability between laboratories. The body fluid mode on the XN-1000 hematological analyzer can count and identify cells immediately and can be used as a simple and rapid screening approach for laboratory testing of sterile body fluids, especially in laboratories with massive quantities of biological fluids. The widespread use of BF modes in automated hematology analyzers can reduce manual methods’ high inter-observer variability, bias, and low reproducibility. For widespread usage, expanded samples including various BFs should be tested and verified for automated hematology analyzers. There is still need further investigations.


Corresponding author: Ebru Evren, Department of Medical Microbiology and İbn-i Sina Education and Research Hospital Central Microbiology Laboratory, Faculty of Medicine, Ankara University, Ankara, Türkiye, E-mail:

Acknowledgments

Thanks to Assist. Prof. Deniz İlhan Topcu from Baskent University, School of Medicine Department of Medical Biochemistry for data evaluation and his great contribution to the study. Thanks to Mehmet Ramazan Ayaş, Flow & Chemistry Business Unit Manager of Sysmex Turkey for his support during the study.

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Conflict of interest: No conflict of interest was declared.

  4. Competing interests: Authors state no conflict of interest.

  5. Informed consent: Not applicable.

  6. Ethical approval: Ethical approval was waived by the local Ethics Committee of Ankara University School of Medicine in view of the retrospective nature of the study (I07-409-22).

References

1. CLSI. Clinical Laboratory Standards Institute. Body fluid analysis for cellular composition; approved guideline CLSI document H56-A. Pennslyvania, USA, Wayne, PA; 2006.Search in Google Scholar

2. Fleming, C, Russcher, H, Lindemans, J, de Jonge, R. Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease. Clin Chem Lab Med 2015;53:1689–706. https://doi.org/10.1515/cclm-2014-1247.Search in Google Scholar PubMed

3. Sandhaus, LM, Ciarlini, P, Kidric, D, Dillman, C, O’Riordan, M. Automated cerebrospinal fluid cell counts using the sysmex XE-5000: is it time for new reference ranges? Am J Clin Pathol 2010;134:734–8. https://doi.org/10.1309/ajcpabgqxsia4smt.Search in Google Scholar

4. Buoro, S, Peruzzi, B, Fanelli, A, Seghezzi, M, Manenti, B, Lorubbio, M, et al.. Two-site evaluation of the diagnostic performance of the Sysmex XN Body Fluid (BF) module for cell count and differential in Cerebrospinal Fluid. Int J Lab Hematol 2018;40:26–33. https://doi.org/10.1111/ijlh.12723.Search in Google Scholar PubMed

5. Lee, J, Cho, Y, Kim, HS, Kang, HJ, Kim, M, Lee, YK. A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment. Med 2021;100:e24852. https://doi.org/10.1097/md.0000000000024852.Search in Google Scholar

6. Bourner, G, De la Salle, B, George, T, Tabe, Y, Baum, H, Culp, N, et al.. ICSH guidelines for the verification and performance of automated cell counters for body fluids. Int J Lab Hematol 2014;36:598–612. https://doi.org/10.1111/ijlh.12196.Search in Google Scholar PubMed

7. Danise, P, Maconi, M, Rovetti, A, Avino, D, Di Palma, A, Pirofalo, MG, et al.. Cell counting of body fluids: comparison between three automated haematology analysers and the manual microscope method. Int J Lab Hematol 2013;35:608–13. https://doi.org/10.1111/ijlh.12093.Search in Google Scholar PubMed

8. Ilki, A, Ayas, R, Ozsoy, S, Soyletir, G. Cost-effectiveness of a new system in ruling out negative urine cultures on the day of administration. Eur J Clin Microbiol 2017;36:1119–23. https://doi.org/10.1007/s10096-017-2898-7.Search in Google Scholar PubMed

9. Schober, P, Boer, C, Schwarte, LA. Correlation coefficients: appropriate use and interpretation. Anesth Analg 2018;126:1763–8. https://doi.org/10.1213/ane.0000000000002864.Search in Google Scholar PubMed

10. Cho, J, Oh, J, Lee, SG, Lee, YH, Song, J, Kim, JH. Performance evaluation of body fluid cellular analysis using the beckman coulter UniCel DxH 800, sysmex XN-350, and UF-5000 automated cellular analyzers. Ann Lab Med 2020;40:122–30. https://doi.org/10.3343/alm.2020.40.2.122.Search in Google Scholar PubMed PubMed Central

11. Lohajaroensub, S, Sakoonwatanyoo, P, Sakun, S, Pichanun, D. Comparison of body fluid cell counting between automate hematology analyzer sysmex XT-4000i and manual microscopic method. Vajira Med J 2015;59:14.Search in Google Scholar

12. Evren, E, Us, E, Karahan, ZC. Determining white blood cell count in sterile body fluids: cell counting chamber vs. sysmex UF-1000i. Gazi Med J 2020;31:19–21.10.12996/gmj.2020.05Search in Google Scholar

13. Xu, WY, Yu, Q, Xie, LX, Chen, BD, Zhang, L. Evaluation of Sysmex XN-1000 hematology analyzer for cell count and screening of malignant cells of serous cavity effusion. Med 2017;96:e7433. https://doi.org/10.1097/md.0000000000007433.Search in Google Scholar

14. Aguadero, V, Cano-Corres, R, Berlanga, E, Torra, M. Evaluation of biological fluid analysis using the sysmex XN automatic hematology analyzer. Cytometry Part B-Clin Cy 2018;94:680–8. https://doi.org/10.1002/cyto.b.21587.Search in Google Scholar PubMed

15. Yang, DG, Zhou, YX, Chen, BD. Performance evaluation and result comparison of the automated hematology analyzers Abbott CD 3700, Sysmex XE 2100 and Coulter LH 750 for cell counts in serous fluids. Clin Chim Acta 2013;419:113–8. https://doi.org/10.1016/j.cca.2013.01.022.Search in Google Scholar PubMed

16. Brown, W, Keeney, M, Chin-Yee, I, Johnson, K, Lantis, K, Finn, W, et al.. Validation of body fluid analysis on the Coulter LH 750. Lab Hematol 2003;9:155–9.Search in Google Scholar

17. Barnes, PW, Eby, CS, Shimer, G. An evaluation of the utility of performing body fluid counts on the coulter LH 750. Lab Hematol 2004;10:127–31. https://doi.org/10.1532/lh96.04023.Search in Google Scholar

18. de Jonge, R, Brouwer, R, de Graaf, MT, Luitwieler, RL, Fleming, C, de Frankrijker-Merkestijn, M, et al.. Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids. Clin Chem Lab Med 2010;48:665–75. https://doi.org/10.1515/CCLM.2010.108.Search in Google Scholar PubMed

19. Mahieu, S, Vertessen, F, Van der Planken, M. Evaluation of ADVIA 120 CSF assay (Bayer) vs. chamber counting of cerebrospinal fluid specimens. Clin Lab Haematol 2004;26:195–9. https://doi.org/10.1111/j.1365-2257.2004.00611.x.Search in Google Scholar PubMed

20. Strik, H, Luthe, H, Nagel, I, Ehrlich, B, Bahr, M. Automated cerebrospinal fluid cytology: limitations and reasonable applications. Anal Quant Cytol Histol 2005;27:167–73.Search in Google Scholar

21. Tanada, H, Ikemoto, T, Masutani, R, Tanaka, H, Takubo, T. Evaluation of the automated hematology analyzer ADVIA(R) 120 for cerebrospinal fluid analysis and usage of unique hemolysis reagent. Int J Lab Hematol 2014;36:83–91. https://doi.org/10.1111/ijlh.12130.Search in Google Scholar PubMed

22. Keuren, JFW, Hoffmann, JJML, Leers, MPG. Analysis of serous body fluids using the CELL-DYN Sapphire hematology analyzer. Clin Chem Lab Med 2013;51:1285–90. https://doi.org/10.1515/cclm-2012-0549.Search in Google Scholar PubMed

23. De Smet, D, Van Moer, G, Martens, GA, Nanos, N, Smet, L, Jochmans, K, et al.. Use of the cell-dyn sapphire hematology analyzer for automated counting of blood cells in body fluids. Am J Clin Pathol 2010;133:291–9. https://doi.org/10.1309/ajcpy7j7olaszapc.Search in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/tjb-2022-0179).


Received: 2022-06-19
Accepted: 2022-09-01
Published Online: 2022-09-19

© 2022 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Opinion Paper
  3. Ischemia – modified albumin by albumin cobalt binding test: a false myth or reality
  4. Research Articles
  5. Machine learning models can predict the presence of variants in hemoglobin: artificial neural network-based recognition of human hemoglobin variants by HPLC
  6. Investigation of the preanalytical process practices in primary care in Istanbul regarding the newborn screening tests
  7. Cell counting chamber vs. Sysmex XN-1000 for determining white blood cell count and differentiation for body fluids
  8. Synthesis of PEITC-loaded gold nanoparticles and evaluation of the hepatoprotective effect on CCl4-induced damage through Nrf2 pathway
  9. Concentrations of B cell-activating factor, aquaporin-4 antibody and brain-derived neurotrophic factor in neuromyelitis optica spectrum disorder
  10. Macula, choroid and retinal nerve fiber layer optical coherence tomography biomarkers in liver dysfunction
  11. Direct-acting antiviral therapy may help restore HCV-induced impaired redox balance and liver fibrosis process
  12. Are VEGF and SCUBE1 gene expressions increased in diabetic retinopathy?
  13. Biochemical analysis of microbiotas obtained from healthy, prediabetic, type 2 diabetes, and obese individuals
  14. Inflammation parameters, monocyte subgroups and toll-like receptor expression before and after dialysis in patients with chronic kidney disease
  15. Matrix metalloproteinase 9 gene-MMP9-DNA methylation status in Turkish schizophrenia patients
  16. HLA DRB1 alleles, IFN-γ and TGF-β Gene Variants in childhood ALL patients
  17. Epithelial-mesenchymal transition as a potential route for DAPT resistance in breast cancer cells
  18. Comparison with molecular effects of ukrain, tamoxifen, and docetaxel on human breast cancer cell lines
  19. In vitro evaluation of 2-pyrazoline derivatives as DPP-4 inhibitors
  20. Effects of polyphenolic-rich extracts from Citrus hystrix on proliferation and oxidative stress in breast and colorectal cancer
  21. The effects of sodium benzoate exposure on learning and neurobehavior during the prepubertal period in rats
  22. The association between βeta 2-microglobulin and bronchopulmonary dysplasia