Abstract
Background
Procalcitonin (PCT) measurement is required for intensive care patients with systemic inflammation symptoms, early diagnosis of possible infections, and evaluation of sepsis severity and prognosis.
Objectives
We aimed to determine the analytical performance of PCT measurement in a Roche Modular E170 (ECLIA) analyzer and compare the performance with VIDAS (BRAHMS/ELFA) analyzer findings.
Material and methods
Within-day and between-day precision value, linearity was determined, and two methods were compared with regression and Bland–Altman analysis.
Results
Both ECLIA and ELFA assays indicated excellent precision, where within-day precision varied between 1.18% and 3.97% CV, and between-day precision varied between 1.77% and 3.93% CV. The ECLIA method was linear up to 62.15 ng/mL. The arithmetic mean was 6.02 ng/mL with the ECLIA method and 8.02 ng/mL with the ELFA method. The correlation coefficient was r=0.996 and p=0.001. The correlation was linear between the two methods. Regression equation was found y=0.78x − 0.23. The Bland–Altman figure was revealed the difference between the methods was specifically in lower concentrations (<0.15 ng/mL).
Conclusions
Both methods show good precision and correlation. It was determined that the difference between methods was significant, especially at <0.15 ng/mL concentration.
Introduction
There is an arising demand for PCT tests from all hospital departments, especially emergency and intensive care units, for early diagnosis and monitoring of sepsis cases. The utilization of antibiotics following PCT results in sepsis and the other infections led to a shorter treatment period and antibiotic use [1], [2], [3].
Within-day and between-day assay precision at low and high concentration for Roche Modular E170 and VIDAS.
Roche modular E170 | Within-day | Between-day | ||
---|---|---|---|---|
Concentration | Low concentration (0.377 ng/mL) | High concentration (9.932 ng/mL) | Low concentration (0.411 ng/mL) | High concentration (10.169 ng/mL) |
%CV | 3.97 | 1.18 | 3.16 | 1.77 |
SD | 0.015 | 0.117 | 0.013 | 0.180 |
|
||||
VIDAS | Within-day | Between-day | ||
Concentration | Low concentration (0.151 ng/mL) | High concentration (10.590 ng/mL) | Low concentration (1.698 ng/mL) | High concentration (17.048 ng/mL) |
|
||||
%CV | 3.30 | 3.18 | 2.65 | 3.93 |
SD | 0.005 | 0.337 | 0.045 | 0.67 |
Despite there isn’t recently approved reference method for PCT measurement, the BRAHMS PCT Kryptor method fairly improved from the original BRAHMS luminometric immunoassay [4]. A few diagnostic companies such as BioMerieux, Roche, DiaSorin, and Siemens have since collaborated with BRAHMS for establishing automated PCT assays on their registered devices utilizing other methods. Additionally, a novel amplified luminescent proximity homogeneous assay (Alpha LISA) was established [5].
New methods have been investigated for various reasons, such as increasing total laboratory and hospital quality, providing further medical benefits, increasing precision, accuracy, and specificity in diagnosis, and reducing costs.
The present study aimed to determine the analytical performance of PCT measurement with ECLIA method and compare the performance of the ECLIA method with the ELFA method.
Materials and methods
The 62 serum samples of the patients admitted to Vakıf Gureba Training and Research Hospital Intensive Care Unit and PCT requests were included in the study. All study procedures were approved by Bezmialem Valide Sultan Vakıf Gureba Education and Research Hospital Ethics Committee (25.02.2009-B.02.1.VGM.2.03.01). Thirty minutes after collecting patient blood samples to determine PCT, the samples were centrifuged in a Becton Dickinson vacutainer tube for 10 min at 4,000 rpm on Nuve 1200 centrifuge.
After PCT levels were measured with the ELFA method on the VIDAS autoanalyzer, samples that varied between 0.05 ng/mL and 68.51 ng/mL were stored at −20 °C until the measurements were conducted with the ECLIA method on Roche Modular E 170 autoanalyzer. Original reagents were used for the measurement on both auto analyzers. We divided samples <0.15 ng/mL (n=8), 0.15–2 ng/mL (n=28) and >2 ng/mL (n=26).
Icteric and hemolyzed samples were excluded from the study. The instrument measurement performances were compared with the modified CLSI EP09-c (Clinical Laboratory Standards Institute) protocol [6].
Serum samples were used to develop low and high serum pools. The low serum pool was negative for the levels of procalcitonin, such as blank. CLSI EP10-A3 AMD protocol was applied for within-day and between-day precisions [7] with low and high serum pools. Each one was divided into 20 portions and stored at −20 °C. The serums with low PCT concentration (0.377 ng/mL) and those with high PCT concentration (9.932 ng/mL) were studied 20 times consecutively for within-day precision (Table 1). Low and high concentrations of PCT pool samples were studied for 20 consecutive days for between-day precision. All samples were analyzed in duplicate for linearity and precision assays.
CLSI EP06 protocol was applied for linearity analysis [8]. Serum sample with 62.15 ng/mL PCT concentration was used to determine linearity by mixing 100%, 75%, 50%, 30%, 20%, 10% and 0% serum samples at 0.041 ng/mL PCT concentration in Roche Modular E 170.
SPSS version 12.0 (Statistical Package for Social Sciences) for Windows software was used for the analysis of the study findings. The findings were analyzed within the 95% confidence interval and p<0.05 significance level. The comparison of the quantitative data was based on between-day and within-day precisions and the two methods (regression analysis, correlation coefficient, Deming regression analysis, Bland–Altman graphic).
Results
ECLIA PCT assay was linear between 0.041 and 62.15 ng/mL (r2=0.99).
The arithmetic mean ± SD (min–max) was 6.02 ± 11.8 (0.02–51.6) ng/mL with the ECLIA method and 8.02 ± 15.1 (0.05–68.1) ng/mL with the ELFA method.
95% confidence interval of the difference was 3.00–9.03 ng/mL for ECLIA method and 4.17–11.87 ng/mL for ELFA method.
The correlations between the findings were analyzed with regression analysis and expressed as the correlation coefficient. Correlation coefficient was found 0.996 and p=0.001. The regression equation was y=0.78x − 0.23 (n=62).
The correlation coefficient was high between the two instruments. Curve fit analysis was conducted to investigate the form of the correlation, and it was determined that the correlation was linear (Figure 1).

Linear relation graphic of E170 and VIDAS measurement results. Regression equation was found y=0.78x − 0.23.
Deming regression analysis was also conducted (Figure 2). Deming regression equation was y=0.78x − 0.26; Sy/x=1.14.

Deming regression analyse of E170 and VIDAS measurement results. Deming regression equation was found y=0.78x − 0.26 (Sy/x=1.14).
Bland–Altman figure (the graph of differences between the means) was plotted with the differences of the measurements from the means for the two devices (Figure 3) for three groups.

Bland–Altman graphic for measurement results of E170 and VIDAS.
Discussion
Steinbach et al. [9] conducted a PCT study on a Cryptor analyzer, and they determined that within-day and between-day assay precisions were <5% and <10%, respectively. Another study was carried out by Wolf and et al. on the Roche Modular E170 analyzer. It was determined that within-day and between-day precisions were 3% and 2.8% for low concentration and 1.3% and 6.3% for high concentration, respectively [10]. These results were consistent with our results.
PCT does not have an allowed CV value determined by the CLIA. In such situations, the CV value obtained in studies that investigated the biological variations of analyte could be used. In the present study, the value determined by Barassi et al. [11] on biological variations of PCT in a healthy population was used as the acceptable CV value. The target CV% cited in that study was 8.1%, the present study CV% was quite good.
In our linearity study, PCT measurements were found to be linear between 0.041 and 62.15 ng/mL. In a linearity study conducted by Steinbach et al. [9], it was linear between 0.04 and 50 ng/mL.
The comparison of the two methods was conducted with regression analysis with the method comparison experiment data. The regression equation was y=0.78x − 0.23 and correlation coefficient was r=0.996 (p=0.001, n=62) (Figure 1). Our r value demonstrated an excellent correlation between the two methods.
Since the comparison method adopted in the ELFA method was not a reference method for PCT measurements, Deming regression analysis was also conducted [12]. The disadvantage of our study is that the ELFA method is not a reference method. In addition, our small sample size was a disadvantage of our work. Deming regression equation was y=0.78x − 0.26; Sy/x=1.14 (Figure 2). It was determined that there was a proportional difference and the difference decreased with the increase in concentration. A novel PCT immunoassay kit on HYBIOME AE-180 system compared to VIDAS BRAHMS PCT by Wang et al. They revealed that the new kit has excellent analytical performance, including analytical sensitivity and a wide linear range. But the new kit slightly overestimated the PCT values, which approves that harmonization of different measurement methods of PCT has not yet been achieved [13].
In conclusion, the present study demonstrated that the analytical performance of the ECLIA method was quite good based on the precision and linearity findings. However, in comparing the methods, the analysis of the Deming regression figure revealed the proportional difference between the two methods. We found that y values corresponded to x values in increasing concentrations starting from 0.15 ng/mL in the equation Y=0.78x − 0.26. It was determined that the difference decreased with the increase in concentration. Bland–Altman Figure 3 supported these findings. All findings suggested that the proportional difference between the two instruments should be revised for the cut-off values specified in the ECLIA method. Also, it was determined in the plot that the difference between the methods was specifically in lower concentrations.
-
Research funding: None declared.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
-
Ethical approval: All study procedures were approved by Bezmialem Valide Sultan Vakıf Gureba Education and Research B.02.1.VGM.2.03.01).
References
1. Scuhuetz, P, Bretscher, C, Bernasconi, L, Mueller, B. Overview of procalcitonin and procalcitonin-guided protocols for the management of patients with infections and sepsis. Expert Rev Mol Diagn 2017;17:593–601. https://doi.org/10.1080/14737159.2017.1324299.Search in Google Scholar PubMed
2. Scuhuetz, P, Mueller, B, Christ-Crain, M, Stolz, D, Tamm, M, Bouadma, L, et al.. Procalcitonin to initiate or discontinue antibiotics in acute respiratory tract infections. Cohrane Database Syst Rev UK 2017;10:CD007498. https://doi.org/10.1002/14651858.CD007498.pub3.Search in Google Scholar PubMed PubMed Central
3. Scuhuetz, P, Briel, M, Christ-Crain, M, Stolz, D, Bouadma, L, Wolff, M, et al.. Procalcitonin to guide initiation and duration of antibiotic treatment in acute respiratory infections: an individual patient data meta-analysis. Clin Infect Dis 2012;55:651–62. https://doi.org/10.1093/cid/cis464.Search in Google Scholar PubMed PubMed Central
4. Soni, NJ, Samson, DJ, Galaydick, JL, Vats, V, Pitrak, DL, Aronson, N. Procalcitonin-guided antibiotic therapy. Rockville (MD): Agency for Healthcare Researchand Quality (US); 2012.Search in Google Scholar
5. Li, P, Chen, Z, Liu, B, Li, K, Wang, H, Lin, L, et al.. Establishment of a novel homogeneous nanoparticle-based assay for sensitive procalcitonin detection of ultra low-volume serum samples. Int J Nanomed 2018;13:5395–404. https://doi.org/10.2147/IJN.S173776.Search in Google Scholar PubMed PubMed Central
6. Measurement procedure comparison and bias estimation using patient samples. CLSI guideline EP09c, 3rd ed. Wayne, PA, USA: Clinical and Laboratory Standards Institute; 2018.Search in Google Scholar
7. Preliminary evaluation of quantitative clinical laboratory measurement procedures; approved guideline. CLSI document EP10-A3-AMD, 3rd ed. Wayne, PA, USA: Clinical and Laboratory Standards Institute; 2014.Search in Google Scholar
8. Evaluation of linearity of quantitative measurement procedures. CLSI guideline EP06, 2nd ed. USA: Clinical and Laboratory Standards Institute; 2020.Search in Google Scholar
9. Steinbach, G, Rau, B, Debard, AL, Javourez, JF, Bienvenu, J, Ponzio, A, et al.. Multicenter evaluation of a new immunoassay for procalcitonin measurement on the Kryptor System. Clin Chem Lab Med 2004;42:440–9. https://doi.org/10.1515/CCLM.2004.077.Search in Google Scholar PubMed
10. de Wolf, HK, Gunnewiek, JK, Berk, Y, van den Ouweland, J, de Metz, M. Comparison of a new procalcitonin assay from the Roche with the established method on the Brahms Kryptor. Clin Chem 2009;55:1043–4. https://doi.org/10.1373/clinchem.2008.117655.Search in Google Scholar PubMed
11. Barassi, A, Pallotti, F, Melzid’Eril, G. Biological variation of procalcitonin in healthy individuals clin. Inside Chem 2004;50:1878. https://doi.org/10.1373/clinchem.2004.037275.Search in Google Scholar PubMed
12. Saraçlı, S, Doğan, İ, Doğan, N. Medikal metot Karşılaştırma Çalışmalarında deming Regresyon Tekniği. Türkiye Klinikleri J Biostat 2009;1:9–15.Search in Google Scholar
13. Wang, G, Wan, Y, Lin, G, Li, Z, Dong, Z, Liu, T. Development of a novel chemiluminescence immunoassay for the detection of procalcitonin. J Immunol Methods 2020;484–485:112829. https://doi.org/10.1016/j.jim.2020.112829.Search in Google Scholar PubMed
© 2021 Burçin Erdem Kinas et al., published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.
Articles in the same Issue
- Frontmatter
- Review Article
- Is carnosine effective to alleviate lung injury: a systematic review
- Research Article
- Neutralizing antibody response and associated factors in Coronavirus-19 disease (COVID-19) up to one month
- Technical Note
- Comparison of enzyme-linked fluorescent assay and electrochemiluminescence immune assay in procalcitonin measurement
- Research Articles
- Pretreatment of diabetic aged rats with combination of ginsenoside-Mc1 and silibinin protects liver from ischemia-reperfusion injury through an AMPK-dependent mechanism
- Effects of exercise and quercetin on muscle energy charge in metabolic syndrome model of rats
- Effects of peroxisome proliferator activated receptor gamma (PPARγ) agonist on fasting model applied neuron cultures
- Combined effect of midostaurin and sphingosine kinase-1 inhibitor on FMS-like tyrosine kinase 3 (FLT3) wild type acute myeloid leukemia cells
- Association between systemic immune inflammation index and newly diagnosed adult celiac disease
- Total cholesterol/HDL cholesterol ratio and monocyte/HDL cholesterol ratio are related with subclinical hypothyroidism in polycystic ovary syndrome
- Effects of different exercise loads on serum betatrophin (ANGPTL-8/lipasin) and cartonectin (CTRP-3) levels in metabolic syndrome
- Impact of serum 25 hydroxyvitamin D deficiency on lipid biomarkers in established coronary artery disease
- The effect of temozolomide on Hsp60 and Hsp70 expression in extracellular vesicles derived from U87MG glioma cells
- Investigation of RASSF4 gene in head and neck cancers
- The role of A268V exon-7 polymorphism of PPARA in development of axial spondyloarthritis
- Analysis of beta globin gene mutations in Diyarbakir
- Education Section
- Mobile learning in a flipped classroom: findings from a “5-lecture-5” blended learning design for large classes
- Flipped learning in faculty development programs: opportunities for greater faculty engagement, self-learning, collaboration and discussion
- Medical interns’ attitudes towards One Health approach
- Coexisting of interprofessional education and organizational culture
- A snapshot of the coaching practices in undergraduate nursing education: evaluation of stakeholders’ perceptions and program costs
Articles in the same Issue
- Frontmatter
- Review Article
- Is carnosine effective to alleviate lung injury: a systematic review
- Research Article
- Neutralizing antibody response and associated factors in Coronavirus-19 disease (COVID-19) up to one month
- Technical Note
- Comparison of enzyme-linked fluorescent assay and electrochemiluminescence immune assay in procalcitonin measurement
- Research Articles
- Pretreatment of diabetic aged rats with combination of ginsenoside-Mc1 and silibinin protects liver from ischemia-reperfusion injury through an AMPK-dependent mechanism
- Effects of exercise and quercetin on muscle energy charge in metabolic syndrome model of rats
- Effects of peroxisome proliferator activated receptor gamma (PPARγ) agonist on fasting model applied neuron cultures
- Combined effect of midostaurin and sphingosine kinase-1 inhibitor on FMS-like tyrosine kinase 3 (FLT3) wild type acute myeloid leukemia cells
- Association between systemic immune inflammation index and newly diagnosed adult celiac disease
- Total cholesterol/HDL cholesterol ratio and monocyte/HDL cholesterol ratio are related with subclinical hypothyroidism in polycystic ovary syndrome
- Effects of different exercise loads on serum betatrophin (ANGPTL-8/lipasin) and cartonectin (CTRP-3) levels in metabolic syndrome
- Impact of serum 25 hydroxyvitamin D deficiency on lipid biomarkers in established coronary artery disease
- The effect of temozolomide on Hsp60 and Hsp70 expression in extracellular vesicles derived from U87MG glioma cells
- Investigation of RASSF4 gene in head and neck cancers
- The role of A268V exon-7 polymorphism of PPARA in development of axial spondyloarthritis
- Analysis of beta globin gene mutations in Diyarbakir
- Education Section
- Mobile learning in a flipped classroom: findings from a “5-lecture-5” blended learning design for large classes
- Flipped learning in faculty development programs: opportunities for greater faculty engagement, self-learning, collaboration and discussion
- Medical interns’ attitudes towards One Health approach
- Coexisting of interprofessional education and organizational culture
- A snapshot of the coaching practices in undergraduate nursing education: evaluation of stakeholders’ perceptions and program costs