Startseite Optimised calibration programmes for comparators for instrument transformers
Artikel Open Access

Optimised calibration programmes for comparators for instrument transformers

Using new technologies to the full extent thanks to traceable calibration
  • Christian Mester

    Christian Mester was awarded a PhD for his work at CERN, the European Organization for Nuclear Research, in Geneva, Switzerland, by the University of Bonn, Germany, in 2009. After a postdoc at École Polytechnique Fédérale de Lausanne EPFL, in Lausanne, Switzerland, he joined the Electrical Power and Energy laboratory at the Federal Institute of Metrology METAS in Berne, Switzerland, in 2011. His current research activities include the calibration of conventional and non-conventional instrument transformers, power standards and power quality instruments, including applications in charging stations for electric vehicles, as well as sampling procedures and algorithms.

    ORCID logo EMAIL logo
Veröffentlicht/Copyright: 15. Januar 2021

Abstract

Traditionally, instrument transformers are calibrated using bridges. By definition, bridges use the null method of measurement. The traditional calibration programme for instrument transformer bridges characterise namely this null measurement. Many new commercial comparators for instrument transformer use a very different method. They sample the secondary signals of reference and device under test (dut) transformer independently. Based on the samples, magnitude and phase of both signals are determined. Ratio error and phase displacement are calculated. Consequently, the significance of their calibration using the traditional calibration programme is limited. Moreover, the operating range of modern comparators is much larger than that of bridges. The additional versatility cannot be used without an adapted calibration programme. This article analyses the calibration programmes for both technologies. An experimental study confirms both the suitability of the new calibration programme and the need to chose the calibration programme depending on the technology of the device to be calibrated. The conclusion is very general and applies to all measurement problems where an operating principle is replaced by another – when changing the operating principle, it is important to check the calibration programme and adapt it if necessary.

Zusammenfassung

Traditionell werden bei der Kalibrierung von Messwandlern Messbrücken verwendet. Brücken verwenden definitionsgemäß ein Verfahren mit Nullabgleich. Durch das klassische Kalibrierprogramm wird insbesondere dieser Nullabgleich charakterisiert. Moderne Wandlermessgeräte hingegen tasten die Ausgangssignale von Referenz und Prüfling ab, bestimmen aus den Abtastwerten Effektivwert und Phase der beiden Signale und berechnen Messabweichung und Fehlwinkel hieraus. Folglich ist eine Kalibrierung mit dem klassischen Kalibrierprogramm für Wandlermessbrücken für solche Geräte nur bedingt aussagekräftig. Zudem bieten solche Geräte eine wesentlich größere Flexibilität im Einsatz, die durch klassische Kalibrierprogramme jedoch nicht abgebildet wird. Hier werden verschiedene Kalibrierprogramme für diese beiden Technologien analysiert. Eine experimentelle Studie bestätigt einerseits deren Eignung und andererseits die Notwendigkeit, das Kalibrierprogramm technologieabhängig zu wählen. Die Schlussfolgerung ist sehr grundsätzlich und auf alle Messaufgaben anwendbar, bei denen ein Funktionsprinzip durch ein anderes ersetzt wird – wird das Funktionsprinzip geändert, so muss geprüft werden, ob auch das Kalibrierprogramm angepasst werden muss.

1 Introduction

Traditionally, instrument transformers are calibrated using bridges. By definition, bridges use the null method of measurement (Fig. 1a) [1]. Some practical realisations use a differential method of measurement [1] instead, where the complex difference between the reference and the dut signal is measured directly, without adjustment to zero. Please note that comparators have at least two channels, usually labelled “reference” and “dut”. Therefore, this paper uses the abbreviation dut only in relation to the comparator’s channel labelled dut, even though the device whose calibration is discussed here is the comparator, not the transformer connected to the comparator’s dut channel. The traditional calibration programme for instrument transformer bridges characterises the difference measurement and the normalisation to reference amplitude. During the calibration, the complex error is varied while keeping the reference signal at a constant amplitude. In a second step, the excitation, i. e. the amplitude of the signal normalised to its rated value, is varied while the error is zero. The latter part is very easy to realise, since it is sufficient to apply the same signal to both the reference and the dut channel. To vary the complex error, an error generator is required. The error generator emulates an instrument transformer with adjustable complex error in the dut channel [2]. The complex error is usually specified and displayed in terms of ratio error and phase displacement, as defined in iec 61869-1 [3]. One very convenient property of this is that the settings of the error and the display of the bridge are in terms of the same quantities, excitation, ratio error and phase displacement. A significant disadvantage is that bridges cannot be used when the transformation ratio of reference transformer and dut transformer differ. Commercial bridges avoid this problem with adaptation transformers within the bridge, adding some flexibility at the expense of a more complicated calibration. However, when the dut transformer is a low-power instrument transformer or has a digital output – for instance according to iec 61850-9-2 [4] – bridges are inconvenient, if not impossible to use. When calibrating high-accuracy reference transformers in national metrology institutes such as the Federal Institute of Metrology metas, bridges are still the preferred choice. They can be designed for such transformers with differences below 50×10-6 and reach uncertainties below 0.01×10-6. Most participants in the last European current transformer comparison used bridges [5].

Figure 1 Operating principles. For simplicity, the instrument transformers are replaced by pairs of resistors here.
Figure 1

Operating principles. For simplicity, the instrument transformers are replaced by pairs of resistors here.

Nowadays, most new measuring instruments for the calibration of instrument transformers – even though they are often sold as bridges – are not bridges anymore. To reflect this, the iec introduced the general term comparator [6]. Those comparators using an indirect method of measurement [1] sample both analogue input channels (Fig. 1b). The sampling of the channels is independent, only the triggering is synchronised. It is easily possible to have very different input signals. When calibrating a current transformer, the reference transformer’s secondary signal is usually a current. The dut could be a low-power instrument transformer lpit with a secondary voltage. As long as the synchronisation is maintained, it is possible to move the analogue-to-digital converter into the dut, making the interface digital. Moreover, since there is no difference being directly measured, arbitrary phase displacements are permissible without any impact on the uncertainty. A practical case is the calibration of a Rogowski coil [7] as dut, where the rated phase displacement is 90 °. Another example is the calibration of an inverting dut, where the amplitude of the difference signal would be as large as the sum of the amplitudes of the secondary signals of reference and dut, exceeding the operating range of a well-designed bridge by far.

In principle, comparators using an indirect method of measurement can also be used to calibrate instrument transformers for power quality measurements [8]. While commercial comparators do not include this feature yet, metas set up and calibrated such a comparator along with the appropriate sources.

It is possible and unfortunately very common to maintain the traditional calibration programme when replacing bridges by comparators using an indirect method of measurement without giving it much thought. The lack of a term as concise and established as bridge incites the use of the wrong term bridge by commercial vendors. In this confusion, it is commonly overlooked that the usable operating range becomes strongly limited by the calibration programme.

This article analyses the operating principles of traditional bridges and comparators using an indirect method of measurement theoretically. This article shows that it is possible to calibrate the complete operating range of such a comparator without increasing the number of calibration points. It also shows how the usable operating range would be limited by the traditional calibration programme for bridges. Furthermore, it proposes to replace the traditional, extensive calibration by a short calibration and an extensive type test. The calibration characterises properties, mainly in the analogue domain, that are subject to ageing, wear and tear. The type test characterises properties, mainly due to software, that are equally important, but do not change with time. While the calibration needs to be repeated regularly, the type test does not. An experimental study confirms the assumptions.

2 Traditional comparators: bridges

Figure 2 Vector diagram: Signals in bridges.
Figure 2

Vector diagram: Signals in bridges.

Signals in comparators for instrument transformers can be represented as vectors (Fig. 2). The input signal in the reference channel, i. e. the secondary signal of the reference transformer, is labelled N, that in the dut channel X. The difference of the angle of X with respect to N is called phase displacement Δφ, where Δφ is defined to be positive when X is leading with respect to N. In bridges, the difference D:=XN is realised in hardware. The three vectors N, X and D define a triangle. Normalisation yields a congruent triangle from which the ratio error ε can be determined easily. For the following analysis, the difference between comparators using a null method of measurement (bridges) and comparators using a differential method of measurement is irrelevant. Therefore, these two methods are treated as one for simplicity. The vector diagram shows clearly that bridges can only be used if ε and Δφ are suffiently small to avoid overdriving the element measuring D. In the extreme case of X=N, D=2N. Commercial bridges are usually protected against damage resulting from this case. However, they are not designed to measure under such conditions because if they were, their resolution would be insufficient for the small values of ε and Δφ they are intended to measure. Usually, commercial bridges include adaptation transformers to allow a predefined mismatch between N and X, for instance 1:1/3 for voltage transformers or 5:1 for current transformers. Table 1 shows typical specifications of a commercial bridge.

Table 1

Typical specifications of a commercial bridge.

Input voltage3 V to 400 V
Input impedance<1VA at 100 V, corresponding to >10kΩ
Ratio error ε: Ranges19.99% to 19.99 %
1.999% to 1.999 %
0.1999% to 0.1999 %
Phase displacement Δφ: Ranges19.99crad to 19.99crad (687 to 687)
1.999crad to 1.999 crad (68.7 to 68.7)
0.1999crad to 0.1999 crad (6.87 to 6.87)
Tolerance on ε and Δφ50×106, 50µrad (0.17)

A bridge for current transformers (Fig. 3), for instance, consists of two measuring elements, one for the complex current IN in the reference channel and one for the complex current ID in the difference branch. Furthermore, a calculation stage determines the ratio error ε and the phase displacement Δφ. Since the difference signal D depends on both ε and Δφ, the measured values of ε and Δφ are not independent of each other. Old bridges often use approximations. The calibration needs to cover both measuring elements as well as the calculation stage.

Usually, the whole operating range is calibrated on the two axes with ε=0 and with Δφ=0 (Fig. 4). As a check, it is common to calibrate a very limited number of points with ε0 and Δφ0. If the device contains adaptation transformers, these need to be calibrated, too, but since they are not influenced by the other channels’ signal, there is no need to vary ε and Δφ in this part of the calibration. Furthermore, the zero (ε=Δφ=0) is calibrated for all permissible excitations. The latter calibration is very simple, since it is sufficient to apply the same signal to both channels at the same time. This part can be easily repeated by the user as a quick – albeit limited – check showing that the bridge’s characteristics influencing this measurement did not change since the last calibration. However, this check does not replace regular calibrations – the linearity of the difference measurement, for instance, is essential for the performance of the bridge, but is not checked in this way.

Figure 3 Bridge for current transformer calibration.
Figure 3

Bridge for current transformer calibration.

Figure 4 Calibration of a bridge: simulation results. Typical uncertainties (k=2k=2) in real calibrations: 26×10-6 and 0.09′{0.09^{\prime }}.
Figure 4

Calibration of a bridge: simulation results. Typical uncertainties (k=2) in real calibrations: 26×10-6 and 0.09.

3 Comparators using an indirect method of measurement

Comparators using an indirect method of measurement can be used in the same way as bridges, but they are much more versatile. As shown in the block diagram (Fig. 5), they determine amplitude and phase of the reference (N) and the dut (X) signal independently. They do not realise the difference signal in hardware. Therefore, there is no limitation on the difference signal. The worst case for a traditional bridge, X=N, i. e. D=2N, can be measured with the same uncertainty as the best case X=N, i. e. D=0. Thanks to modern electronics, they reach uncertainties that are comparable to those of commercial bridges and exceed the requirements of commercial applications (Tab. 2).

It is possible to use the calibration programme for bridges, which has been used for decades and is generally accepted, since the signal levels used during calibration are covered by the operating range. However, the operating range is much larger. This calibration programme would only cover a very small part of it, thereby limiting the usable operating range significantly. The increased versatility of the comparator could not be used. In addition, the calibration programme would contain many redundant points.

Figure 5 Comparator for voltage transformers using an indirect method of measurement.
Figure 5

Comparator for voltage transformers using an indirect method of measurement.

Table 2

Typical specifications of a commercial comparator using an indirect method of measurement.

Reference channelRanges500V, 250V, 125V, 60V, 30V, 15V, 7.5V and 3.75V
Input impedance380kΩ, 500pF
DUT channel 1Ranges500V, 250V, 125V, 60V, 30V, 15V, 7.5V and 3.75V
Input impedance380kΩ, 500pF
DUT channel 2Ranges15V, 10V, 5V, 2.5V and 1V
500mV, 250mV, 100mV, 50mV and 25mV
Input impedance>1GΩ, 70pF
Digital channelaccording IEC 61850-9-2
Tolerance on ε and ΔφDUT channel 1200×106, 200µrad (0.69)at 2 < UN/UX
100×106, 100µrad (0.34)at 1.1 < UN/UX ≤ 2
50×106, 50µrad (0.17)at 0.9 ≤ UN/UX ≤ 1.1
100×106, 100µrad (0.34)at 0.5 ≤ UN/UX < 0.9
200×106, 200µrad (0.69)at UN/UX < 0.5
DUT channel 2400×106, 200µrad (0.69)
Digital DUT100×106, 300µrad (1.1)

3.1 Calibration of the ratio error ε

Calibration programmes for bridges contain a large number of points where the ratio error ε is varied in small steps for a constant reference signal N. The specification requires ε to be very small, e. g. ε<2%, hence XN. For comparators using an indirect method of measurement, this means that the linearity of the X channel is analysed in great detail for small-signal excitation around a fixed operating point. If the analogue front-end including the analogue-to-digital converter is designed properly, the deviation of the comparator is constant for all ε of this calibration programme (Fig. 6). In other words, a single calibration point would have given the same information without loss of quality. A useful addition to the calibration programme would be a characterisation of the linearity of the front end across a much larger range, ideally covering the whole operating range. This point is very relevant when reference and dut transformers have different transformation ratios. Especially dangerous is the false conclusion that given the result of the limited calibration for XN and the absence of adaptation transformers, the behaviour of the comparator was the same across the whole operating range. This argumentation overlooks that these comparators usually use multiple different measuring ranges. Within the ranges, the linearity is usually excellent, but between ranges, steps occur (Fig. 7). Therefore, the calibration programme must be extended to include all ranges. Since the comparator calculates the ratio error ε based on the measured amplitudes of N and X, it is meaningful to specify the applied amplitudes N and X in the calibration certificate rather than the – very large – values of ε.

Figure 6 Calibration of a comparator using an indirect method of measurement: simulation results. Typical uncertainties (k=2k=2) in real calibrations: 26×10-6 and 0.09′{0.09^{\prime }}.
Figure 6

Calibration of a comparator using an indirect method of measurement: simulation results. Typical uncertainties (k=2) in real calibrations: 26×10-6 and 0.09.

Figure 7 Calibration of a comparator using an indirect method of measurement: simulation results. Typical uncertainties (k=2k=2) in real calibrations: 26×10-6. Mismatch between channels is negligible, but strong dependency on ranges needs to be accounted for.
Figure 7

Calibration of a comparator using an indirect method of measurement: simulation results. Typical uncertainties (k=2) in real calibrations: 26×10-6. Mismatch between channels is negligible, but strong dependency on ranges needs to be accounted for.

3.2 Calibration of the phase displacement Δφ

Bridges can only be used for small phase displacements Δφ. Consequently, the rated phase offset of the reference and the dut transformer must be equal. Usually, this is a trivial requirement – bridges are usually designed and used for inductive instrument transformers only. Their rated phase offset is zero by definition [3]. Other instrument transformers than inductive transformers can have any rated phase displacement; for Rogowski coils [7], it is 90 °. Since such transformers – the discussion about whether or not those should still be called instrument transformers in the future is still ongoing in iec tc 38, but current standards use this term – are becoming more and more common, iec 61869-6 introduced the terminology required to address this. The symbol Δφ is used for the phase displacement and φe is the phase error [9]. The most recent draft of iec 61869-1 [10] uses this convention for all kinds of instrument transformers. Comparators using an indirect method of measurement can be used equally well for any phase displacement Δφ. The phase error φe is calculated based on the measured phases of N and X. This calculation is done in software. There is no need to check the correctness of this calculation during every calibration.

For the correct determination of the rms values, it is not necessary that both signals have the same frequency. In practice, the frequency of the secondary signal of the reference and the dut transformer is identical by design of the calibration set-up since the primary signal of both transformers is identical. However, differing internal clock frequencies in the measurement elements of the comparator or an unfortunate choice of internal algorithms [11] can cause false frequency differences. This impacts the measurement of the phase displacement Δφ, but not the ratio error ε. The larger the phase displacement, the larger the impact of this effect. Therefore, a characterisation with large phase displacements is necessary once. For small phase displacements, the effect is much smaller than the measurement uncertainty and hence invisible.

Other effects impacting the measurement phase displacement Δφ are the phase shift in the analogue front-end of the individual channels and delay mismatch in the triggering paths of the analogue-to-digital converters. The phase shift in the analogue front-end can be made constant by careful choice of the components within every range, i. e. independent of the signal amplitude as long as the range is not changed. The delay mismatch in the triggering paths is constant as long as the same analogue-to-digital converters are used. Both effects are independent of the ratio error ε and the phase displacement Δφ. There is no need for fine steps of ε and Δφ in the calibration of comparators using an indirect method of measurement.

The measurement of the phase displacement Δφ depends on the linearity of the measuring elements. The phase of a signal can, for instance, be determined from the timing of the first sample after the zero crossing. However, a very high sampling rate is necessary to satisfy the uncertainty requirements. A more appropriate solution is to decompose the sinusoidal signal using the equation 2Nsin2πf0t+φ=N1sin2πf0t+N2cos2πf0t and to determine the phase from the ratio N2/N1. The impact of the linearity of the comparator on the phase difference Δφ is easily visible when reading Δφ during the linearity analysis of the ratio error ε. Additional calibration points are not necessary.

For comparators using an indirect method of measurement with analogue inputs only, it is sufficient to know the phase displacement between the two analogue input signals. If only one signal, usually the secondary signal of the reference transformer, is analogue and the other one digital [12], [13], the phase measurement of the comparator must use the same timing reference as the dut transformer. Usually, this is a pulse-per-second (pps) signal derived from a satellite navigation system. Since both the analogue signal and the external synchronisation signal are delayed in the comparator, separating the contributions of the analogue and the digital path is not trivial [14], [15], [16], [17], [18].

4 Type testing

A calibration is necessarily limited to a finite number of calibration points, but the operating range of a comparator – and, in general, any real measuring instrument – contains an infinite number of possible operating points. Knowledge about the operating principle of the device to be calibrated allows the expert to define a practical calibration programme. Imagine a device to be calibrated that is expected to be linear over its operating range. A calibration with two calibration points at the extrema is sufficient. An additional point between the extrema allows for a simple check of the assumption of linearity. So an expert can calibrate this device with two points; an additional point gives confidence in their claim that two points are sufficient. In principle, the assumption of linearity needs to be checked only once, but since the effort required for a single extra calibration point is very small, it is usually repeated at every calibration. A non-expert, however, would need a calibration programme with a very large number of points. The expert’s advantage is even more pronounced when the device to be calibrated has multiple input quantities – like a comparator for instrument transformers. Since the device is subject to ageing as well as wear and tear, the calibration needs to be repeated at intervals.

This article postulates some properties of comparators depending on their internal implementation. This is of some interest to prospective users choosing a comparator to buy. A practical user of a given comparator is less interested in generalities about possible implementations of comparators in general. They need to know as much as possible about their comparator and convince auditors of this knowledge and its correctness. The operating principle does not age and does not wear out. The knowledge about it is valid for the lifetime of the comparator. The intended properties are usually described in the operating manual with sufficient detail, but unintended properties are not. So the user needs a way to determine the relevant properties by observation rather than by declaration of the manufacturer. An efficient solution is the combination of calibrations at regular intervals and a single type test carried out once. The knowledge acquired or confirmed during the type test allows for a reduction of the calibration programme without loss of information. This procedure is well-established in legal metrology, where type approvals are issued based on extensive type tests [19] of a small number of devices – which must be representative for its type – and every instrument is verified with a very small number of verification points [20].

Table 3 gives an overview of properties of comparators using an indirect method of measurement.

Table 3

Expected properties of comparators using an indirect method of measurement and corresponding checks.

PropertyPhase error of one channel independent of phase and amplitude of the other channel
Phase error function of range but constant within range
CheckVary phase of signal in other channel while keeping phase in first channel constant
Vary amplitude
ExampleVary phase of X signal while keeping phase of N signal constant
Repeat for different amplitudes of N and X
Type testValid for the lifetime of the comparator
PropertyActual value phase offset
CheckApply signal to channel, measure phase error
Repeat for all ranges
ExampleSet one amplitude of N signal for each range while keeping X signal constant
Set one amplitude of X signal for each range while keeping N signal constant
ExampleSet one amplitude of N signal for each range while synchronising to PPS signal
Set one amplitude of X signal for each range while synchronising to PPS signal
ExampleSet one amplitude of N and X signals at the same time for each range while synchronising to PPS signal
CalibrationTo be repeated at regular intervals
PropertyAmplitude error independent of amplitude of the other channel
Amplitude measurement linear within ranges
CheckVary amplitude in small steps within range, measure amplitude
Repeat for all ranges
Repeat for all channels
ExampleVary amplitude of N signal while keeping X signal constant
Vary amplitude of X signal while keeping N signal constant
Type testValid for the lifetime of the comparator
PropertyActual value of amplitude (offset and gain error)
CheckApply minimum, average and maximum amplitude within range, measure amplitude
Repeat for all ranges
ExampleVary amplitude of N and X signals at the same time
CalibrationTo be repeated at regular intervals
PropertyCalculation of ratio error exact within numeric resolution
CheckDuring all previous tests, readout amplitudes of N and X as well as ratio error from comparator
Calculate ratio error based on amplitudes of N and X
Compare calculated ratio error with ratio error readout from the comparator
Type testValid for the lifetime of the comparator’s firmware
PropertyCalculation of phase displacement exact within numeric resolution
CheckDuring all previous tests, readout phases of N and X as well as phase displacement from comparator
Calculate phase displacement based on phases of N and X
Compare calculated phase displacement with phase displacement readout from the comparator
Type testValid for the lifetime of the comparator’s firmware
PropertyCalculation of phase error exact within numeric resolution
CheckSet rated phase offset and rated delay time [9] to random values
During all previous tests, readout phases displacement as well as phase error from comparator
Calculate phase error based on phase displacement
Compare calculated phase error with phase error readout from the comparator
Type testValid for the lifetime of the comparator’s firmware

5 Experimental results

The theoretical analysis discussed above was confirmed experimentally using comparators from different manufacturers. Table 4 gives a summary. Since the extreme values of the deviations are to be found at the extrema of ε and Δφ, the extreme values are given (rows “max” and “min”) together with the values at ε=Δφ=0 (rows “0”). Bridges are easily identified by the influence of Δφ on ε as seen in the column ε(Δφ) and ε on Δφ as seen in the column Δφ(ε).

Table 4

Calibration of different comparator models. The rows labelled “max” show the values for the maximum of ε and Δφ, respectively, in the calibration programme, the rows labelled “min” those for the minimum. The rows labelled “0” show the values for ε=Δφ=0. Models A and D are bridges. The other models use an indirect method of measurement.

ModelDeviation with respect to reference value for
εε106ΔφεεΔφ106ΔφΔφ
Amax20.09−450.31
020.0020.00
min−97−0.1414−0.29
Bmax−50.01−20.02
0−20.02−20.02
min00.02−20.01
Cmax−20.01−20.02
0−10.02−10.02
min−10.01−20.01
Dmax180.10−450.11
040.0040.00
min−2−0.1210−0.20
Emax00.05−40.05
0−1−0.040−0.04
min−4−0.04−4−0.04
Fmax−10.0120.01
000.0000.00
min00.0110.02
Gmax00.1500.15
000.1410.14
min−50.1400.14
Uncertainty260.09260.09

Table 5

Calibration of a comparator using an indirect method of measurement: Linearity within a range.

Deviation
Settingswith respect to reference value for
UNVUXVε106Δφ
125100100.00
100100−10.02
50100−16−0.02
Uncertainty260.09

A comparator using an indirect method of measurement was calibrated for different transformation ratios of reference and dut transformer. Table 5 shows the deviations of ε and Δφ from their reference values as a function of the voltages in the channels N and X. Variations of 1%ε1% and 100Δφ100 were shown not to have any effect on their deviations from their reference values, as expected for this comparator.

Table 6

Calibration of a bridge: Effect of the adaptation transformer as a function of the ratio error ε with Δφ=0.

Variation of ε/106Variation of Δφ/
Settingwith respect to UNwith respect to UN
ε/%50 V125 V50 V125 V
1.003−2−0.120.00
0.105−2−0.130.02
0.016−1−0.130.02
0.003−7−0.120.02
−0.014−3−0.130.01
−0.106−3−0.130.02
−1.00154−0.110.02

Table 7

Calibration of a bridge: Effect of the adaptation transformer as a function of the phase displacement Δφ with ε=0.

Variation of ε/106Variation of Δφ/
Settingwith respect to UNwith respect to UN
Δφ/50 V125 V50 V125 V
10052−0.200.00
104−3−0.130.01
12−4−0.150.00
03−7−0.120.02
−12−4−0.130.01
−105−4−0.120.02
−1002−9−0.100.10

A bridge was calibrated the same way. The deviations of ε and Δφ from their reference values depend on both ε and Δφ, as discussed above. The calibration at ε=Δφ=0 shows that a well-designed adaptation transformer yields better results than the analogue front-end and the analogue-to-digital converter of the comparator using an indirect method of measurement. Contrary to the comparator using an indirect method of measurement, the bridge shows a clear effect when varying 1%ε1% and 100Δφ100. However, the adaptation transformer is only subject to one signal and cannot be influenced by the other signal. Tables 6 and 7 show the differences in the dependency on 1%ε1% and 100Δφ100 due to the adaptation transformer. They are smaller than the measurement uncertainty.

6 Conclusion

It is always important to take the operation principle into account when defining calibration programmes. The example discussed here is comparators for instrument transformers. Some are bridges, but most new commercial comparators use an indirect method of measurement. An adapted calibration programme gives traceability to the full operating range and not just to the part that is in common with traditional bridges. Since many properties are not subject to ageing, this article proposed defining a type test programme – to be carried out once – and a calibration programme – to be repeated regularly. The programmes were derived from purely theoretical knowledge about the instrument. Experiments with real instruments confirmed the assumptions.

Award Identifier / Grant number: 16ENG04

Funding statement: This research was developed in the framework of the 16eng04 myrails project [22]. The latter received funding from the empir programme co-financed by the Participating States and the European Union’s Horizon 2020 research and innovation programme.

About the author

Christian Mester

Christian Mester was awarded a PhD for his work at CERN, the European Organization for Nuclear Research, in Geneva, Switzerland, by the University of Bonn, Germany, in 2009. After a postdoc at École Polytechnique Fédérale de Lausanne EPFL, in Lausanne, Switzerland, he joined the Electrical Power and Energy laboratory at the Federal Institute of Metrology METAS in Berne, Switzerland, in 2011. His current research activities include the calibration of conventional and non-conventional instrument transformers, power standards and power quality instruments, including applications in charging stations for electric vehicles, as well as sampling procedures and algorithms.

Acknowledgment

This research and the related metrology services are part of the European Metrology Network for Smart Electricity Grids [21], the single point of contact across Europe that provides stakeholder support for measurement challenges in the realisation of smart electricity grids.

References

1. IEC 60050: International Electrotechnical Vocabulary, 2020-11-09.Suche in Google Scholar

2. S. Siegenthaler and C. Mester. A computer-controlled calibrator for instrument transformer test sets. IEEE Transactions on Instrumentation and Measurement, vol. 66 no. 6, June 2017, pp. 1184–1190.10.1109/TIM.2017.2659858Suche in Google Scholar

3. IEC 61869-1:2007: Instrument transformers – Part 1: General requirements.Suche in Google Scholar

4. IEC 61850-9-2:2011: Communication networks and systems for power utility automation – Part 9-2: Specific communication service mapping (SCSM) – Sampled values over ISO/IEC 8802-3.Suche in Google Scholar

5. K. Draxler et al. Results of an International Comparison of Instrument Current Transformers up to 10 kA and 50 Hz Frequency. In: Conference on Precision Electromagnetic Measurements CPEM, July 2018.10.1109/CPEM.2018.8500821Suche in Google Scholar

6. IEC/IEEE TS CD 61869-105:2018 (38/567/CD): Instrument transformers – Part 105: Uncertainty evaluation in the calibration of instrument transformers.Suche in Google Scholar

7. A. P. Chattock. On a magnetic potentiometer. Philosophical Magazine and Journal of Science, vol. XXIV, no. 5, 1887, pp. 94–96.10.1088/1478-7814/9/1/305Suche in Google Scholar

8. C. Mester et al. Einführung in die rückführbare Messung von Power Quality. tm – Technisches Messen, Band 85, Heft 12, 2018, pp. 738–745.10.1515/teme-2018-0021Suche in Google Scholar

9. IEC 61869-6:2016: Instrument transformers – Part 6: Additional general requirements for low-power instrument transformers.Suche in Google Scholar

10. IEC CD 61869-1:2020 (38/631/CD): Instrument transformers –- Part 1: General requirements.Suche in Google Scholar

11. L. Hentgen and C. Mester. Sampling AC signals: Comparison of fitting algorithms and FFT. In: Conference on Precision Electromagnetic Measurements CPEM, July 2018.10.1109/CPEM.2018.8500885Suche in Google Scholar

12. IEC 61869-9:2016: Instrument transformers – Part 9: Digital interface for instrument transformers.Suche in Google Scholar

13. IEC FDIS 61869-13:2020 (38/634/FDIS): Instrument Transformers -– Part 13: Standalone Merging Unit.Suche in Google Scholar

14. C. Mester. Timestamping type 3458A multimeter samples. In: Conference on Precision Electromagnetic Measurements CPEM, July 2018.10.1109/CPEM.2018.8501165Suche in Google Scholar

15. J. Feng et al. Time Latency of DC Input Path for 3458A Sampling Multimeter. In: Conference on Precision Electromagnetic Measurements CPEM, July 2018.10.1109/CPEM.2018.8501082Suche in Google Scholar

16. G. Crotti et al. A method for the measurement of digitizers’ absolute phase error. Journal of Physics: Conference Series, vol. 1065, 2018.10.1088/1742-6596/1065/5/052035Suche in Google Scholar

17. G. Crotti et al. Measurement of the absolute phase error of digitizers. IEEE Transactions on Instrumentation and Measurement, vol. 68, no. 6, June 2019, pp. 1724–1731.10.1109/CPEM.2018.8501177Suche in Google Scholar

18. J.-P. Braun. Measure of the Absolute Phase Angle of a Power Frequency Sinewave with Respect to UTC. In: Conference on Precision Electromagnetic Measurements CPEM, July 2018.10.1109/CPEM.2018.8501065Suche in Google Scholar

19. EN 50470: Electricity metering equipment (a. c.). General requirements, tests and test conditions. Metering equipment (class indexes A, B and C).Suche in Google Scholar

20. Verordnung des EJPD vom 26. August 2015 über Messmittel für elektrische Energie und Leistung (EMmV, SR 941.251).Suche in Google Scholar

21. European Metrology Network for Smart Electricity Grids. www.euramet.org/european-metrology-networks/smart-electricity-grids/, 2020-11-13.Suche in Google Scholar

22. D. Giordano et al. Accurate Measurement of Energy, Efficiency and Power Quality in the Electric Railway System. In: Conference on Precision Electromagnetic Measurements CPEM, July 2018.10.1109/CPEM.2018.8500811Suche in Google Scholar

Received: 2020-11-14
Accepted: 2020-12-21
Published Online: 2021-01-15
Published in Print: 2021-02-26

© 2021 Mester, published by De Gruyter

This work is licensed under the Creative Commons Attribution 4.0 International License.

Heruntergeladen am 3.10.2025 von https://www.degruyterbrill.com/document/doi/10.1515/teme-2020-0086/html?lang=de
Button zum nach oben scrollen