Analytical performance of eight enzymatic assays for ethanol in serum evaluated by data from the Belgian external quality assessment scheme
-
Wim Coucke
, Corinne Charlier
, Jan Tytgat
, Sarah Wille
Abstract
Objectives
Fast and reliable ethanol assays analysis are used in a clinical context for patients suspected of ethanol intoxication. Mostly, automated systems using an enzymatic reaction based on ethanol dehydrogenase are used. The manuscript focusses on the evaluation of the performance of these assays.
Methods
Data included 30 serum samples used in the Belgian EQA scheme from 2019 to 2021 and concentrations ranged from 0.13 to 3.70 g/L. A regression line between target concentrations and reported values was calculated to evaluate outliers, bias, variability and measurement uncertainty.
Results
A total of 1,611 results were taken into account. Bias was the highest for Alinity c over the whole concentration range and the lowest for Vitros for low concentrations and Cobas 8000 using the c702 module for high concentrations. The Architect and Cobas c501/c502 systems showed the lowest variability over the whole concentration range. Highest variability was observed for Cobas 8000 using the 702 module, Thermo Scientific and Alinity c. Cobas 8000 using the c702 module showed the highest measurement uncertainty for lower concentrations. For higher concentrations, Alinity c, Thermo Scientific and Vitros were the methods with the highest measurement uncertainty.
Conclusions
The bias of the enzymatic techniques is nearly negligible for all methods except Alinity c. Variability differs strongly between measurement procedures. This study shows that the Alinity c has a worse measurement uncertainty than other systems for concentrations above 0.5 g/L. Overall, we found the differences in measurement uncertainty to be mainly influenced by the differences in variability.
-
Research funding: None declared.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
-
Informed consent: Not applicable.
-
Ethical approval: Not applicable.
References
1. Ishmayana, S, Fadhlillah, M, Kristia, Y, Budiman, H. Validation of a modified alcohol dehydrogenase assay for ethanol determination. Curr Chem Lett 2015;4:77–84. https://doi.org/10.5267/j.ccl.2015.1.001.Search in Google Scholar
2. Dubowski, KM. Alcohol determination in the clinical laboratory. Am J Clin Pathol 1980;74:747–50. https://doi.org/10.1093/ajcp/74.5.747.Search in Google Scholar PubMed
3. Krasowski, MD, Wilcoxon, RM, Miron, J. A retrospective analysis of glycol and toxic alcohol ingestion: utility of anion and osmolal gaps. BMC Clin Pathol 2012;12:1. https://doi.org/10.1186/1472-6890-12-1.Search in Google Scholar PubMed PubMed Central
4. Karch, SB. Drug abuse handbook. Boca Raton, FL, USA: CRC Press; 2019:1156 p.Search in Google Scholar
5. Handbook for calculation of measurement uncertainty in environmental laboratories (NT TR 537 - Edition 4). [Internet]. Nordtest; 2017. Available from: https://www.nordtest.info/wp/2017/11/29/handbook-for-calculation-of-measurement-uncertainty-in-environmental-laboratories-nt-tr-537-edition-4/ [Accessed 17 Jan 2022].Search in Google Scholar
6. Coucke, W, Tanasković, JV, Bouacida, L, Broeders, S, China, B, Demarteau, M, et al.. Alternative sample-homogeneity test for quantitative and qualitative proficiency testing schemes. Anal Chem 2019;91:1847–54. https://doi.org/10.1021/acs.analchem.8b03313.Search in Google Scholar PubMed
7. Coucke, W, Blerk, MV, Libeer, J-C, Campenhout, CV, Albert, A. A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometry. Clin Chem Lab Med 2010;48:645–50. https://doi.org/10.1515/CCLM.2010.122.Search in Google Scholar PubMed
8. Pinheiro, J, Bates, D. Mixed-effects models in S and S-PLUS. New York, USA: Springer Science & Business Media; 2006.Search in Google Scholar
9. Coucke, W, Charlier, C, Lambert, W, Martens, F, Neels, H, Tytgat, J, et al.. Application of the characteristic function to evaluate and compare analytical variability in an external quality assessment scheme for serum ethanol. Clin Chem 2015;61:948–54. https://doi.org/10.1373/clinchem.2015.240176.Search in Google Scholar PubMed
10. Rilibak - German guidelines for quality - Westgard. [Internet]. Available from: https://www.westgard.com/rilibak.htm [Accessed 21 Mar 2022].Search in Google Scholar
11. Dubois, N, Sqalli, G, Gilson, M, Charlier, C. Analytical validation of a quantitative method for therapeutic drug monitoring on the Alinity®c Abbott. Ann Biol Clin 2020;78:147–55. https://doi.org/10.1684/abc.2020.1535.Search in Google Scholar PubMed
12. Krintus, M, Fernandez, JA, Chesters, C, Colla, R, Ford, C, Frattolillo, D, et al.. Analytical performance of 10 high-volume clinical chemistry assays on the alinity c system. Lab Med 2019;50:e1–8. https://doi.org/10.1093/labmed/lmy053.Search in Google Scholar PubMed
Supplementary Material
The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0285).
© 2022 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorial
- Transdermal measurement of cardiac troponins: the future is now
- Reviews
- Perinatal presepsin assessment: a new sepsis diagnostic tool?
- Hypertriglyceridemia, a causal risk factor for atherosclerosis, and its laboratory assessment
- Opinion Paper
- The novelties of the regulation on health technology assessment, a key achievement for the European union health policies
- General Clinical Chemistry and Laboratory Medicine
- Performance of four regression frameworks with varying precision profiles in simulated reference material commutability assessment
- Comparison of six regression-based lot-to-lot verification approaches
- Failure Mode and Effects Analysis (FMEA) at the preanalytical phase for POCT blood gas analysis: proposal for a shared proactive risk analysis model
- Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading
- Analytical performance of eight enzymatic assays for ethanol in serum evaluated by data from the Belgian external quality assessment scheme
- Vitamin D metabolism in living kidney donors before and after organ donation
- Validation of steroid ratios for random urine by mass spectrometry to detect 5α-reductase deficiency in Vietnamese children
- Evaluation of serum neurofilament light in the early management of mTBI patients
- Assessment of urine sample quality by the simultaneous measurement of urinary γ-glutamyltransferase and lactate dehydrogenase enzyme activities: possible application to unravel cheating in drugs of abuse testing
- Reference Values and Biological Variations
- Age and sex specific reference intervals of 13 hematological analytes in Chinese children and adolescents aged from 28 days up to 20 years: the PRINCE study
- Cancer Diagnostics
- Prostate health index (PHI) as a reliable biomarker for prostate cancer: a systematic review and meta-analysis
- A comparison of the faecal haemoglobin concentrations and diagnostic accuracy in patients suspected with colorectal cancer and serious bowel disease as reported on four different faecal immunochemical test systems
- Circulating cell-free DNA undergoes significant decline in yield after prolonged storage time in both plasma and purified form
- Cardiovascular Diseases
- Analytical and clinical performance evaluation of a new high-sensitivity cardiac troponin I assay
- Infectious Diseases
- Results of a SARS-CoV-2 virus genome detection external quality assessment round focusing on sensitivity of assays and pooling of samples
- Letters to the Editors
- Improving D-dimer testing appropriateness by controlling periodicity of retesting: prevention is better than cure
- Biological variation of serum cholinesterase catalytic concentrations
- Three-month ad interim analysis of total anti-SARS-CoV-2 antibodies in healthy recipient of a single BNT162b2 vaccine booster
- Fibrin strands in peripheral blood smear: the COVID-19 era
- Fragments of alpha-1-antitrypsin in patients with severe COVID-19 and bacterial pulmonary sepsis
- Comparison of thyroid stimulating hormone, free thyroxine, total triiodothyronine, thyroglobulin and peroxidase antibodies measurements by two different platforms
- Effect of different incubation times on the detection of factor VIII inhibitor in acquired hemophilia A
Articles in the same Issue
- Frontmatter
- Editorial
- Transdermal measurement of cardiac troponins: the future is now
- Reviews
- Perinatal presepsin assessment: a new sepsis diagnostic tool?
- Hypertriglyceridemia, a causal risk factor for atherosclerosis, and its laboratory assessment
- Opinion Paper
- The novelties of the regulation on health technology assessment, a key achievement for the European union health policies
- General Clinical Chemistry and Laboratory Medicine
- Performance of four regression frameworks with varying precision profiles in simulated reference material commutability assessment
- Comparison of six regression-based lot-to-lot verification approaches
- Failure Mode and Effects Analysis (FMEA) at the preanalytical phase for POCT blood gas analysis: proposal for a shared proactive risk analysis model
- Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading
- Analytical performance of eight enzymatic assays for ethanol in serum evaluated by data from the Belgian external quality assessment scheme
- Vitamin D metabolism in living kidney donors before and after organ donation
- Validation of steroid ratios for random urine by mass spectrometry to detect 5α-reductase deficiency in Vietnamese children
- Evaluation of serum neurofilament light in the early management of mTBI patients
- Assessment of urine sample quality by the simultaneous measurement of urinary γ-glutamyltransferase and lactate dehydrogenase enzyme activities: possible application to unravel cheating in drugs of abuse testing
- Reference Values and Biological Variations
- Age and sex specific reference intervals of 13 hematological analytes in Chinese children and adolescents aged from 28 days up to 20 years: the PRINCE study
- Cancer Diagnostics
- Prostate health index (PHI) as a reliable biomarker for prostate cancer: a systematic review and meta-analysis
- A comparison of the faecal haemoglobin concentrations and diagnostic accuracy in patients suspected with colorectal cancer and serious bowel disease as reported on four different faecal immunochemical test systems
- Circulating cell-free DNA undergoes significant decline in yield after prolonged storage time in both plasma and purified form
- Cardiovascular Diseases
- Analytical and clinical performance evaluation of a new high-sensitivity cardiac troponin I assay
- Infectious Diseases
- Results of a SARS-CoV-2 virus genome detection external quality assessment round focusing on sensitivity of assays and pooling of samples
- Letters to the Editors
- Improving D-dimer testing appropriateness by controlling periodicity of retesting: prevention is better than cure
- Biological variation of serum cholinesterase catalytic concentrations
- Three-month ad interim analysis of total anti-SARS-CoV-2 antibodies in healthy recipient of a single BNT162b2 vaccine booster
- Fibrin strands in peripheral blood smear: the COVID-19 era
- Fragments of alpha-1-antitrypsin in patients with severe COVID-19 and bacterial pulmonary sepsis
- Comparison of thyroid stimulating hormone, free thyroxine, total triiodothyronine, thyroglobulin and peroxidase antibodies measurements by two different platforms
- Effect of different incubation times on the detection of factor VIII inhibitor in acquired hemophilia A