Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading
-
Cora M. Märtens
Abstract
Objectives
A carrier prototype by Aerocom® (Schwäbisch Gmünd, Germany) for pneumatic tube systems (PTS) is able to transport 9 blood tubes which are automatically fixed by closing the lid. In this study, we examined the influence of the transport on blood sample quality using the carrier prototype comparing to courier transport and a conventional carrier (AD160, Aerocom®).
Methods
Triplicate blood samples sets (1 lithium heparin, 1 EDTA, 1 sodium citrate) of 35 probands were split among the transportation methods: 1. courier, 2. conventional carrier, and 3. carrier prototype. After transport 51 measurands from clinical chemistry, hematology and coagulation were measured and compared.
Results
Overall, 49 of the investigated 51 measurands showed a good concordance among the three transport types, especially between the conventional carrier and the carrier prototype. Focusing on well-known hemolysis sensitive measurands, potassium showed no statistically significant differences. However, between courier and both carrier types lactate dehydrogenase (LDH) and free hemoglobin (fHb) showed statistically significant shifts, whereas the clinical impact of the identified differences was neglectable. The median concentration of fHb, for example, was 0.29 g/L (18 µmol/L), 0.31 g/L (19 µmol/L) and 0.32 g/L (20 µmol/L) for courier transport, conventional carrier and carrier prototype, respectively. These differences cannot be resolved analytically since the minimal difference (MD) for fHb is 0.052 g/L (3.23 µmol/L), at this concentration.
Conclusions
The carrier prototype by Aerocom® is suitable for transportation of diagnostic blood samples. The overall workflow is improved by decreasing hands-on-time on the ward and laboratory while minimizing the risk of incorrectly packed carriers.
Acknowledgments
The study was supported by Aerocom (Schwäbisch Gmünd, Germany).
-
Research funding: This study was in part funded by Aerocom® (Schwäbisch Gmünd, Germany).
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
-
Informed consent: Informed consent was obtained from all individuals included in this study.
-
Ethical approval: The study was approved by the local ethics committee (BB 100/15).
-
Data availability: The datasets generated and analysed during the current study are available from the corresponding author on reasonable request.
References
1. Pupek, A, Matthewson, B, Whitman, E, Fullarton, R, Chen, Y. Comparison of pneumatic tube system with manual transport for routine chemistry, hematology, coagulation and blood gas tests. Clin Chem Lab Med 2017;55:1537–44. https://doi.org/10.1515/cclm-2016-1157.Search in Google Scholar PubMed
2. Le Quellec, S, Paris, M, Nougier, C, Sobas, F, Rugeri, L, Girard, S, et al.. Pre-analytical effects of pneumatic tube system transport on routine haematology and coagulation tests, global coagulation assays and platelet function assays. Thromb Res 2017;153:7–13. https://doi.org/10.1016/j.thromres.2016.12.022.Search in Google Scholar PubMed
3. Subbarayan, D, Choccalingam, C, Lakshmi, CKA. The effects of sample transport by pneumatic tube system on routine hematology and coagulation tests. Adv Hematol 2018;2018:6940152. https://doi.org/10.1155/2018/6940152.Search in Google Scholar PubMed PubMed Central
4. Calmette, L, Ibrahim, F, Gouin, I, Horellou, M-H, Mazoyer, É, Fontenay, M, et al.. Évaluation de l’influence du transport par pneumatique des échantillons pour la réalisation des examens d’hémostase: qualification du réseau pneumatique de l’hôpital Cochin (AP-HP). Ann Biol Clin 2017;75:93–100. https://doi.org/10.1684/abc.2016.1215.Search in Google Scholar PubMed
5. Cakirca, G, Erdal, H. The effect of pneumatic tube systems on the hemolysis of biochemistry blood samples. J Emerg Nurs 2017;43:255–8. https://doi.org/10.1016/j.jen.2016.09.007.Search in Google Scholar PubMed
6. Heireman, L, Stroobants, J, Uyttenbroeck, W, Goossens, Y, Dreezen, C, Luyts, D, et al.. Smartphone application monitoring of acceleration forces during pneumatic tube system transport of emergency department patient samples. Clin Lab 2018;64:1297–304. https://doi.org/10.7754/Clin.Lab.2018.180325.Search in Google Scholar PubMed
7. Gils, C, Broell, F, Vinholt, PJ, Nielsen, C, Nybo, M. Use of clinical data and acceleration profiles to validate pneumatic transportation systems. Clin Chem Lab Med 2020;58:560–8. https://doi.org/10.1515/cclm-2019-0881.Search in Google Scholar PubMed
8. Ellis, G. An episode of increased hemolysis due to a defective pneumatic air tube delivery system. Clin Biochem 2009;42:1265–9. https://doi.org/10.1016/j.clinbiochem.2009.05.002.Search in Google Scholar PubMed
9. Strubi-Vuillaume, I, Carlier, V, Obeuf, C, Vasseur, F, Maury, J-C, Maboudou, P, et al.. Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays. Ann Clin Biochem 2016;53:295–7. https://doi.org/10.1177/0004563215586600.Search in Google Scholar PubMed
10. Suchsland, J, Winter, T, Greiser, A, Streichert, T, Otto, B, Mayerle, J, et al.. Extending laboratory automation to the wards: effect of an innovative pneumatic tube system on diagnostic samples and transport time. Clin Chem Lab Med 2017;55:225–30. https://doi.org/10.1515/cclm-2016-0380.Search in Google Scholar PubMed
11. Medical laboratories – Requirements for quality and competence. (ISO 15189:2012, Corrected version 2014-08-15); German version EN ISO 15189:2012.Search in Google Scholar
12. Thomas, L. Labor und diagnose: indikation und bewertung von laborbefunden für die medizinische diagnostik, 8th ed. Frankfurt/Main: Th-Books Verl.-Ges; 2012.Search in Google Scholar
13. Mensel, B, Wenzel, U, Roser, M, Lüdemann, J, Nauck, M. Considerably reduced centrifugation time without increased hemolysis: evaluation of the new BD vacutainer SSTTMII advance. Clin Chem 2007;53:794–5. https://doi.org/10.1373/clinchem.2006.079582.Search in Google Scholar PubMed
14. Vogt, W. Revision of the “guideline of the German medical association on quality assurance in medical laboratory examinations – rili-BAEK” (unauthorized translation). LaboratoriumsMedizin 2015;39:34–7. https://doi.org/10.1515/labmed-2014-0046.Search in Google Scholar
15. Benjamini, Y, Hochberg, Y. Controlling the false discovery rate: a practical and powerful approach to multiple testing. J Roy Stat Soc B 1995;57:289–300. https://doi.org/10.1111/J.2517-6161.1995.TB02031.X.Search in Google Scholar
16. Keutmann, S, Zylla, S, Dahl, M, Friedrich, N, Landgraf, R, Heinemann, L, et al.. Measurement uncertainty impacts diagnosis of diabetes mellitus: reliable minimal difference of plasma glucose results. Diabetes Ther 2020;11:293–303. https://doi.org/10.1007/s13300-019-00740-w.Search in Google Scholar PubMed PubMed Central
17. Yücel, D, Dalva, K. Effect of in vitro hemolysis on 25 common biochemical tests. Clin Chem 1992;38:575–7.10.1093/clinchem/38.4.575Search in Google Scholar
18. Lippi, G, Plebani, M, Di Somma, S, Cervellin, G. Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories. Crit Rev Clin Lab Sci 2011;48:143–53. https://doi.org/10.3109/10408363.2011.600228.Search in Google Scholar PubMed
19. Lee, A-J, Suk Suh, H, Jeon, C-H, Kim, S-G. Effects of one directional pneumatic tube system on routine hematology and chemistry parameters; a validation study at a tertiary care hospital. Pract Lab Med 2017;9:12–7. https://doi.org/10.1016/j.plabm.2017.07.002.Search in Google Scholar PubMed PubMed Central
20. Kapoula, GV, Kontou, PI, Bagos, PG. The impact of pneumatic tube system on routine laboratory parameters: a systematic review and meta-analysis. Clin Chem Lab Med 2017;55:1834–44. https://doi.org/10.1515/cclm-2017-0008.Search in Google Scholar PubMed
21. Cui, M, Jing, R, Wang, H. Changes of serum lactate dehydrogenase and potassium levels produced by a pneumatic tube system. Lab Med 2009;40:728–31. https://doi.org/10.1309/LMZNZ2Y1P3KRDQYL.Search in Google Scholar
22. Heireman, L, van Geel, P, Musger, L, Heylen, E, Uyttenbroeck, W, Mahieu, B. Causes, consequences and management of sample hemolysis in the clinical laboratory. Clin Biochem 2017;50:1317–22. https://doi.org/10.1016/j.clinbiochem.2017.09.013.Search in Google Scholar PubMed
23. Kara, H, Bayir, A, Ak, A, Degirmenci, S, Akinci, M, Agacayak, A, et al.. Hemolysis associated with pneumatic tube system transport for blood samples. Pakistan J Med Sci 2014;30:50–8. https://doi.org/10.12669/pjms.301.4228.Search in Google Scholar PubMed PubMed Central
Supplementary Material
The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2022-0193).
© 2022 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorial
- Transdermal measurement of cardiac troponins: the future is now
- Reviews
- Perinatal presepsin assessment: a new sepsis diagnostic tool?
- Hypertriglyceridemia, a causal risk factor for atherosclerosis, and its laboratory assessment
- Opinion Paper
- The novelties of the regulation on health technology assessment, a key achievement for the European union health policies
- General Clinical Chemistry and Laboratory Medicine
- Performance of four regression frameworks with varying precision profiles in simulated reference material commutability assessment
- Comparison of six regression-based lot-to-lot verification approaches
- Failure Mode and Effects Analysis (FMEA) at the preanalytical phase for POCT blood gas analysis: proposal for a shared proactive risk analysis model
- Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading
- Analytical performance of eight enzymatic assays for ethanol in serum evaluated by data from the Belgian external quality assessment scheme
- Vitamin D metabolism in living kidney donors before and after organ donation
- Validation of steroid ratios for random urine by mass spectrometry to detect 5α-reductase deficiency in Vietnamese children
- Evaluation of serum neurofilament light in the early management of mTBI patients
- Assessment of urine sample quality by the simultaneous measurement of urinary γ-glutamyltransferase and lactate dehydrogenase enzyme activities: possible application to unravel cheating in drugs of abuse testing
- Reference Values and Biological Variations
- Age and sex specific reference intervals of 13 hematological analytes in Chinese children and adolescents aged from 28 days up to 20 years: the PRINCE study
- Cancer Diagnostics
- Prostate health index (PHI) as a reliable biomarker for prostate cancer: a systematic review and meta-analysis
- A comparison of the faecal haemoglobin concentrations and diagnostic accuracy in patients suspected with colorectal cancer and serious bowel disease as reported on four different faecal immunochemical test systems
- Circulating cell-free DNA undergoes significant decline in yield after prolonged storage time in both plasma and purified form
- Cardiovascular Diseases
- Analytical and clinical performance evaluation of a new high-sensitivity cardiac troponin I assay
- Infectious Diseases
- Results of a SARS-CoV-2 virus genome detection external quality assessment round focusing on sensitivity of assays and pooling of samples
- Letters to the Editors
- Improving D-dimer testing appropriateness by controlling periodicity of retesting: prevention is better than cure
- Biological variation of serum cholinesterase catalytic concentrations
- Three-month ad interim analysis of total anti-SARS-CoV-2 antibodies in healthy recipient of a single BNT162b2 vaccine booster
- Fibrin strands in peripheral blood smear: the COVID-19 era
- Fragments of alpha-1-antitrypsin in patients with severe COVID-19 and bacterial pulmonary sepsis
- Comparison of thyroid stimulating hormone, free thyroxine, total triiodothyronine, thyroglobulin and peroxidase antibodies measurements by two different platforms
- Effect of different incubation times on the detection of factor VIII inhibitor in acquired hemophilia A
Articles in the same Issue
- Frontmatter
- Editorial
- Transdermal measurement of cardiac troponins: the future is now
- Reviews
- Perinatal presepsin assessment: a new sepsis diagnostic tool?
- Hypertriglyceridemia, a causal risk factor for atherosclerosis, and its laboratory assessment
- Opinion Paper
- The novelties of the regulation on health technology assessment, a key achievement for the European union health policies
- General Clinical Chemistry and Laboratory Medicine
- Performance of four regression frameworks with varying precision profiles in simulated reference material commutability assessment
- Comparison of six regression-based lot-to-lot verification approaches
- Failure Mode and Effects Analysis (FMEA) at the preanalytical phase for POCT blood gas analysis: proposal for a shared proactive risk analysis model
- Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading
- Analytical performance of eight enzymatic assays for ethanol in serum evaluated by data from the Belgian external quality assessment scheme
- Vitamin D metabolism in living kidney donors before and after organ donation
- Validation of steroid ratios for random urine by mass spectrometry to detect 5α-reductase deficiency in Vietnamese children
- Evaluation of serum neurofilament light in the early management of mTBI patients
- Assessment of urine sample quality by the simultaneous measurement of urinary γ-glutamyltransferase and lactate dehydrogenase enzyme activities: possible application to unravel cheating in drugs of abuse testing
- Reference Values and Biological Variations
- Age and sex specific reference intervals of 13 hematological analytes in Chinese children and adolescents aged from 28 days up to 20 years: the PRINCE study
- Cancer Diagnostics
- Prostate health index (PHI) as a reliable biomarker for prostate cancer: a systematic review and meta-analysis
- A comparison of the faecal haemoglobin concentrations and diagnostic accuracy in patients suspected with colorectal cancer and serious bowel disease as reported on four different faecal immunochemical test systems
- Circulating cell-free DNA undergoes significant decline in yield after prolonged storage time in both plasma and purified form
- Cardiovascular Diseases
- Analytical and clinical performance evaluation of a new high-sensitivity cardiac troponin I assay
- Infectious Diseases
- Results of a SARS-CoV-2 virus genome detection external quality assessment round focusing on sensitivity of assays and pooling of samples
- Letters to the Editors
- Improving D-dimer testing appropriateness by controlling periodicity of retesting: prevention is better than cure
- Biological variation of serum cholinesterase catalytic concentrations
- Three-month ad interim analysis of total anti-SARS-CoV-2 antibodies in healthy recipient of a single BNT162b2 vaccine booster
- Fibrin strands in peripheral blood smear: the COVID-19 era
- Fragments of alpha-1-antitrypsin in patients with severe COVID-19 and bacterial pulmonary sepsis
- Comparison of thyroid stimulating hormone, free thyroxine, total triiodothyronine, thyroglobulin and peroxidase antibodies measurements by two different platforms
- Effect of different incubation times on the detection of factor VIII inhibitor in acquired hemophilia A