Abstract
Background
Heterophile antibody (HAb) interferences in immunoassays can cause falsely elevated hCG concentrations leading to incorrect diagnosis and treatments options. When results are not consistent with the clinical findings, hCG HAb interference investigation may be requested by the physician. A retrospective evaluation of the frequency of HAb interference was performed among cases of physician-requested investigations and the effectiveness of commercially available blocking reagents to detect HAb interference in two immunoassay systems was evaluated.
Methods
One hundred and thirteen physician requests for hCG HAb investigation from 2008 to 2017 were reviewed. The primary method used to measure hCG was the Beckman Coulter Access Total βhCG (2008–2010) and the Roche Elecsys HCG+β (2014–2017). HAb investigation included measurement by two immunoassays before and after treatment of samples with heterophile blocking reagents and serial dilution studies.
Results
Five cases of HAb and HAb-like interference were identified. The interference frequency was 6.7% for the Beckman assay and 2.9% for the Roche assay. The presence of HAb was detected using heterophile blocking reagents and an alternative method in three cases. The other two cases were detected due to discrepant results with an alternative method and non-linear serial dilutions (HAb-like).
Conclusions
HAb interference was observed in the Beckman and the Roche assays. The heterophile blocking reagents failed to detect 40% of interference cases. Blocking reagents should not solely be used for these investigations. Multiple strategies including the use of serial dilutions and using an alternative platform are critical when troubleshooting interferences in hCG immunoassays.
Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
References
1. Berger P, Sturgeon C. Pregnancy testing with hCG – future prospects. Trends Endocrinol Metab 2014;25:637–48.10.1016/j.tem.2014.08.004Search in Google Scholar
2. American College of Obstetricians and Gynecologist. Avoiding inappropriate clinical decision based of false positive human chorionic gonadotropin test results. ACOG Committee opinion No 278. Obstet Gynecol 2002;100:1057–9.10.1097/00006250-200211000-00044Search in Google Scholar
3. Stenman U, Alfthan H, Hotakainen K. Human chorionic gonadotropin in cancer. Clin Biochem 2004;37:549–61.10.1016/j.clinbiochem.2004.05.008Search in Google Scholar
4. Emerson JF, Lai KY. Endogenous antibody interferences in immunoassays. Lab Med 2013;44:69–73.10.1309/LMMURCFQHKSB5YECSearch in Google Scholar
5. Garcia-Gonzalez E, Aremendia M, Alvarez-Ballano D, Trincado P, Rello L. Serum sample containing endogenous antibodies interfering with multiple hormone immunoassays. Laboratory strategies to detect interference. Pract Lab Med 2016;4:1–10.10.1016/j.plabm.2015.11.001Search in Google Scholar
6. Klee GG. Interferences in hormone immunoassays. Clin Lab Med 2004;24:1–18.10.1016/j.cll.2004.01.003Search in Google Scholar
7. Levinson SS, Miller JJ. Towards a better understanding of heterophile (and the like) antibody interference with modern immunoassays. Clin Chim Acta 2002;325:1–15.10.1016/S0009-8981(02)00275-9Search in Google Scholar
8. Ward G, Simpson A, Boscato L, Hickman PE. The investigation of interferences in immunoassay. Clin Biochem 2017;50:1306–11.10.1016/j.clinbiochem.2017.08.015Search in Google Scholar PubMed
9. Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem 1999;45:942–56.10.1093/clinchem/45.7.942Search in Google Scholar
10. Sturgeon CM, Berger P, Bidart JM, Birken S, Burns C, Norman RJ, et al. Differences in recognition of the 1st WHO international reference reagents for hCG-related isoforms by diagnostic immunoassays for human chorionic gonadotropin. Clin Chem 2009;55:1484–91.10.1373/clinchem.2009.124578Search in Google Scholar PubMed
11. Soares DG, Millot F, Lacroix I, Lotz JP. Heterophile antibody interference led to unneeded chemotherapy in a testicular cancer patient. Urol Case Rep 2016;9:1–3.10.1016/j.eucr.2016.07.006Search in Google Scholar PubMed PubMed Central
12. Bolstad N, Warren DJ, Nustad K. Heterophilic antibody interference in immunometric assays. Best Pract Res Clin Endocrinol Metab 2013;27:647–61.10.1016/j.beem.2013.05.011Search in Google Scholar PubMed
13. Preissner CM, Dodge LA, O’Kane DJ, Singh RJ, Grebe SK. Prevalence of heterophilic antibody interference in eight automated tumor marker immunoassays. Clin Chem 2005;51:208–10.10.1373/clinchem.2004.040501Search in Google Scholar PubMed
14. Butler SA, Cole LA. Use of heterophilic antibody blocking agent (HBT) in reducing false-positive hCG results. Clin Chem 2001;47:1332–3.10.1093/clinchem/47.7.1332Search in Google Scholar
15. Ismail AA, Walker PL, Barth JH, Lewandowski KC, Jones R, Burr WA. Wrong biochemistry results: two case reports and observational study in 5310 patients on potentially misleading thyroid-stimulating hormone and gonadotropin immunoassay results. Clin Chem 2002;48:2023–9.10.1093/clinchem/48.11.2023Search in Google Scholar
16. Cole LA. Human chorionic gonadotropin (hCG) and hyperglycosylated hCG, seven semi-independent critical molecules: a review. J Mol Oncol Res 2017;1:22–44.Search in Google Scholar
©2019 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorials
- CCLM Award for the Most Cited Paper
- Folate and vitamin B12 assays after recalibration to the WHO International Standard 03/178: making the interpretation as simple as possible, but not simpler
- Reviews
- Blood contamination in salivary diagnostics: current methods and their limitations
- Central adrenal insufficiency: open issues regarding diagnosis and glucocorticoid treatment
- Genetics and Molecular Diagnostics
- Measuring the chronology of the translational process of molecular genetic discoveries
- Development and interlaboratory evaluation of a NIST Reference Material RM 8366 for EGFR and MET gene copy number measurements
- General Clinical Chemistry and Laboratory Medicine
- Post-translational modification-derived products are associated with frailty status in elderly subjects
- Urine chloride self-measurement to monitor sodium chloride intake in patients with chronic kidney disease
- Estimated urinary osmolality based on combined urinalysis parameters: a critical evaluation
- Measurement of S100B protein: evaluation of a new prototype on a bioMérieux Vidas® 3 analyzer
- Measuring thyroglobulin in patients with thyroglobulin autoantibodies: evaluation of the clinical impact of BRAHMS Kryptor® Tg-minirecovery test in a large series of patients with differentiated thyroid carcinoma
- Human chorionic gonadotropin suspected heterophile interference investigations in immunoassays: a recommended approach
- Certified reference material against PR3 ANCA IgG autoantibodies. From development to certification
- Diagnostic accuracy of a fully automated multiplex celiac disease antibody panel for serum and plasma
- Fasting serum bile acids concentration is associated with insulin resistance independently of diabetes status
- Hematology and Coagulation
- The association between activated protein C ratio and Factor V Leiden are gender-dependent
- Reference Values and Biological Variations
- Determination of sigma score based on biological variation for haemostasis assays: fit-for-purpose for daily practice?
- Calcitonin measurement in pediatrics: reference ranges are gender-dependent, validation in medullary thyroid cancer and thyroid diseases
- Cancer Diagnostics
- Uncovering the clinical impact of kallikrein-related peptidase 5 (KLK5) mRNA expression in the colorectal adenoma-carcinoma sequence
- Cardiovascular Diseases
- Performance of a novel high sensitivity cardiac troponin I assay in asymptomatic hemodialysis patients – evidence for sex-specific differences
- Infectious Diseases
- Rapid susceptibility testing of multi-drug resistant Escherichia coli and Klebsiella by glucose metabolization monitoring
- Letters to the Editor
- Vitamin B12 and folate levels in a healthy population: establishing reference intervals
- Reference values of a new serum folate assay traceable to the WHO International Standard
- Serum protein electrophoresis and complement deficiencies: a veteran but very versatile test in clinical laboratories
- Introduction of a novel ELISA assay for serum AMH determination
- Bone alkaline phosphatase on the IDS-iSYS automated analyser; cross-reactivity with intestinal ALP
- Evaluation of the MULTISURE HIV Rapid Test in a Korean population with low human immunodeficiency virus prevalence
- Vancomycin immunoassay: does the Advia Centaur XPT underestimate the exposure of patients? A method comparison study
- Hb Hunan and Hb Hengyang: Two unexpected discoveries during HbA1c measurements
- Analytical performance of a CE-marked immunoassay to quantify phosphorylated neurofilament heavy chains
- Implementation of an automated method for direct quantification of urinary ammonium
- Congress Abstracts
- Proceedings of ACBI 2018 41ST Annual Conference Association of Clinical Biochemists in Ireland
Articles in the same Issue
- Frontmatter
- Editorials
- CCLM Award for the Most Cited Paper
- Folate and vitamin B12 assays after recalibration to the WHO International Standard 03/178: making the interpretation as simple as possible, but not simpler
- Reviews
- Blood contamination in salivary diagnostics: current methods and their limitations
- Central adrenal insufficiency: open issues regarding diagnosis and glucocorticoid treatment
- Genetics and Molecular Diagnostics
- Measuring the chronology of the translational process of molecular genetic discoveries
- Development and interlaboratory evaluation of a NIST Reference Material RM 8366 for EGFR and MET gene copy number measurements
- General Clinical Chemistry and Laboratory Medicine
- Post-translational modification-derived products are associated with frailty status in elderly subjects
- Urine chloride self-measurement to monitor sodium chloride intake in patients with chronic kidney disease
- Estimated urinary osmolality based on combined urinalysis parameters: a critical evaluation
- Measurement of S100B protein: evaluation of a new prototype on a bioMérieux Vidas® 3 analyzer
- Measuring thyroglobulin in patients with thyroglobulin autoantibodies: evaluation of the clinical impact of BRAHMS Kryptor® Tg-minirecovery test in a large series of patients with differentiated thyroid carcinoma
- Human chorionic gonadotropin suspected heterophile interference investigations in immunoassays: a recommended approach
- Certified reference material against PR3 ANCA IgG autoantibodies. From development to certification
- Diagnostic accuracy of a fully automated multiplex celiac disease antibody panel for serum and plasma
- Fasting serum bile acids concentration is associated with insulin resistance independently of diabetes status
- Hematology and Coagulation
- The association between activated protein C ratio and Factor V Leiden are gender-dependent
- Reference Values and Biological Variations
- Determination of sigma score based on biological variation for haemostasis assays: fit-for-purpose for daily practice?
- Calcitonin measurement in pediatrics: reference ranges are gender-dependent, validation in medullary thyroid cancer and thyroid diseases
- Cancer Diagnostics
- Uncovering the clinical impact of kallikrein-related peptidase 5 (KLK5) mRNA expression in the colorectal adenoma-carcinoma sequence
- Cardiovascular Diseases
- Performance of a novel high sensitivity cardiac troponin I assay in asymptomatic hemodialysis patients – evidence for sex-specific differences
- Infectious Diseases
- Rapid susceptibility testing of multi-drug resistant Escherichia coli and Klebsiella by glucose metabolization monitoring
- Letters to the Editor
- Vitamin B12 and folate levels in a healthy population: establishing reference intervals
- Reference values of a new serum folate assay traceable to the WHO International Standard
- Serum protein electrophoresis and complement deficiencies: a veteran but very versatile test in clinical laboratories
- Introduction of a novel ELISA assay for serum AMH determination
- Bone alkaline phosphatase on the IDS-iSYS automated analyser; cross-reactivity with intestinal ALP
- Evaluation of the MULTISURE HIV Rapid Test in a Korean population with low human immunodeficiency virus prevalence
- Vancomycin immunoassay: does the Advia Centaur XPT underestimate the exposure of patients? A method comparison study
- Hb Hunan and Hb Hengyang: Two unexpected discoveries during HbA1c measurements
- Analytical performance of a CE-marked immunoassay to quantify phosphorylated neurofilament heavy chains
- Implementation of an automated method for direct quantification of urinary ammonium
- Congress Abstracts
- Proceedings of ACBI 2018 41ST Annual Conference Association of Clinical Biochemists in Ireland