Home Folate and vitamin B12 assays after recalibration to the WHO International Standard 03/178: making the interpretation as simple as possible, but not simpler
Article Publicly Available

Folate and vitamin B12 assays after recalibration to the WHO International Standard 03/178: making the interpretation as simple as possible, but not simpler

  • Simona Ferraro EMAIL logo and Mauro Panteghini
Published/Copyright: February 2, 2019

The recent alignment of serum folate and vitamin B12 (B12) assays to the WHO International Standard (IS) code 03/178 has produced a significant shift in the average vitamin measured values and triggered studies on the re-evaluation of reference intervals (RI) for both biomarkers [1], [2], [3], [4]. This also yielded the opportunity to update recommendations about the cost-effective application of both tests and the interpretation of marker concentrations in serum [5], [6]. Authoritative guidelines, which define the risk for vitamin deficiency according to B12 and folate testing, indeed reported as interpretative criteria, RI estimated by using assays not traceable to the WHO IS 03/178 [7]. For both tests, cost-effectiveness is maximized when the request is applied to subjects at risk for deficiency and when the deficiency itself may represent a life-threatening condition or invalidate the treatment effect (e.g. in pregnant women or hemodialysis patients) [5], [6]. On the contrary, tests should not be ordered to monitor or evaluate the effect of vitamin supplementation. By applying these indications for an appropriate testing strategy, it was demonstrated an ~50% saving in laboratory costs, without compromising the patient safety [5], [6].

In general, for serum B12 the capability of the test to rule out the deficiency at the individual level is low when a single threshold value, such as the lower reference limit (LRL), is adopted. Multiple thresholds have been validated by clinical studies establishing the relationship between marker concentrations, the probability of deficiency and the need for supplementation [5]. Accordingly, the establishment of RI is not helpful for interpreting B12 concentrations in serum. Furthermore, the correct RI derivation by individual laboratories is very challenging because serum B12 levels are strongly conditioned by the selection criteria used to enrol the “apparently healthy” population, by some pre-analytical factors (e.g. time of sample drawing) and can be affected by many variables, such as dietary habit, vitamin supplements, use of the contraceptive pill, drugs, etc., which are sometimes difficult to control [5]. Thus, the experimental establishment of serum B12 RI by Solé-Enrech et al. [3], performed with the aim of improving the interpretation of vitamin concentrations in individuals suspected for deficiency, is of limited value.

On the contrary, the evaluation of the inter-assay variability of B12 measurement at serum concentrations around the established thresholds for the risk of deficiency may be useful in indicating if these cut-offs may be generalized or require an adjustment according to the employed method. The more recently published data about the comparability among current B12 assays agree in showing a good concordance among the most widely used measuring systems, except for results by Beckman Coulter DxI Unicel that were negatively biased [8], [9]. Results from external quality assessment schemes (EQAS) can also aid to assess the current status of harmonization of B12 measurements. Data from the national Qualimedlab EQAS (www.qualimedlab.it) (2016–2018 exercises) for the four most popular assays (Abbott Architect, Beckman Coulter Access/Dx, Roche Modular/Elecsys/cobas and Siemens Advia Centaur) confirm the marked negative bias for Beckman Coulter systems (–22% in average), with other methods fulfilling the quality goal for desirable inter-assay bias (±17.7%) based on biological variation of B12 [9]. According to all these data, we can conclude that the harmonization of B12 results obtained by the commercially available measuring systems is currently acceptable enough, with few exceptions.

For serum folate testing, a single threshold, set at LRL, is usually employed to screen for vitamin deficiency in non-fortified countries. Consequently, the prevalence of folate deficiency is strictly dependent from the correctness of cut-off establishment and from the accuracy of folate assays [6]. Data for the US National Health and Nutrition Examination Survey (NHANES) first showed that the actual prevalence of folate deficiency in the population may significantly shift with the reformulation of assays [10]. If tracing back calibration of commercial assays to a new reference material changes results, this may alter their relation to existing RI, with a consequent misleading test interpretation. This is exactly what happened with the introduction of WHO IS 03/178 as a calibrator of commercial systems, which caused a significant negative shift in the folate measured values [6]. Using the reformulated Roche assay, at serum folate concentrations around the LRL, we experienced a difference of ~50% vs. the old Roche assay [6]. Therefore, we observed marked changes in the distribution of folate results of the population tested in our laboratory after the adoption of the recalibrated assay, with a wide increase of test results suggesting a vitamin deficiency. Similar results were obtained by other authors [2]. New experimental data from reference individuals should, therefore, be urgently obtained with the recalibrated assays in order to accurately redefine the reference distribution and estimate proper RI.

Table 1 lists the design and results of available studies dealing with the re-evaluation of folate RI [1], [2], [3], [4]. All studies employed the Elecsys Folate III (Roche Diagnostics) reagents traceable to WHO IS 03/178 applied on cobas e601/e801 platforms. Comparing the RI provided by these studies, there are relevant differences that may be ascribed to the study design, sample size, definition of the reference population, type of enrolment and criteria for subject recruitment, pre-analytical factors (fasting, time of blood drawing, detection of sample hemolysis), and measurement protocols. As previously mentioned, the cut-off reported by the running guidelines to define the state of folate deficiency, i.e. <3 μg/L (or <4 μg/L when using homocysteine concentrations as the metabolic indicator) [7], dated back to the use of assays not yet standardized to WHO IS 03/178 and, in contrast to what was claimed by Cluitmans and van den Ouweland [4], cannot be used as comparators to judge the reliability of RI obtained in the studies listed in Table 1.

Table 1:

Characteristics of published studies estimating the reference interval for serum folate measured with Roche Folate III assay, calibrated against WHO International Standard 03/178.

Authors (ref.)Sample size (n)Studied populationExclusion criteriaMeasurement protocolEstimated RI (2.5th–97.5th percentile limits), μg/L
Ferraro et al. [1]322Blood donors, median age 45.4 yearsHb and MCV≠RI, supplementation with folic acid, HI ≥25 (Abbott Architect)Serum stored at −80<°C, measurements performed in four runs using a single reagent lot1.3–9.8
Hepburn et al. [2]9797Outpatients, median age 54 years<18 or >90 years of age, folate concentrations <2 or >20 μg/L, Hb, MCV and B12≠RIaSamples analysed on the day of collection2.4–17.5
Solé-Enrech et al. [3]120Mainly healthy hospital workers and residentsLow dietary consumption, malabsorption, increased requirements or losses of vitaminsaSamples analysed on the day of collection2.0–13.9, <45 years
1262.9–16.6, ≥45 years
Cluitmans and van den Ouweland [4]130Outpatients, median age 55 yearsPregnant or lactating women, supplementation with folic acid, homocysteine >15 μmol/L, HI ≥50 (Roche cobas)Serum stored at −80<°C until measured3.0–19.0
  1. RI, reference interval; Hb, haemoglobin; MCV, erythrocyte mean corpuscular volume; HI; hemolytic index; B12, vitamin B12. aHI not evaluated.

In 2016, Kristensen et al. [9] showed that differences among folate results obtained from different manufacturers were within the limit for desirable bias (±19.2%), which is, however, a relatively large acceptance criterion, as a result of the wide biological variation of the analyte. At that time, only Abbott reported traceability to WHO IS 03/178, while the other assays claimed traceability to different US pharmacopeia (USP) convention or in-house manufacturer calibrators. Since then, to the best of our knowledge no updated comparative data have been published. EQAS could help us in understanding the current status of folate measurements, but non-commutability of control materials may be a very limiting issue, at least for some commercial assays [9].

A final important aspect relates to the general potential inconsistency of the use of LRL (or another single cut-off rule) for detecting folate deficiency. According to the estimation of biological variation of folate in serum, the lower within-person variation with respect to the inter-individual variation implies a low index of individuality (i.e. ratio of within- to between-subject variances) [6]. Under this condition, an isolated dichotomized interpretation of folate results, resorting to a decision limit for deficiency, can be misleading and the longitudinal monitoring of serial folate changes be more effective in classifying individuals with relation to their vitamin status [6].

In conclusion, the meritorious work for reducing the inter-assay disagreement between folate and B12 results, started with the availability of the WHO IS 03/178 and the implementation of traceability to it, must be continued. The release and adoption of other suitable reference materials, like the Joint Committee on Traceability in Laboratory Medicine (JCTLM)-listed SRM 1955 from the National Institute of Standards and Technology (NIST), should be encouraged, together with the regular monitoring of the performance of commercial measuring systems through the participation to EQAS using proper commutable materials, aiming to improve the accuracy of results. This will make the interpretation of folate and B12 values more consistent, but not simpler, giving the complexity of issues around diagnosis and treatment of their deficiency.

Acknowledgments

The authors are grateful to Giancarlo Zucchelli and Rudina Ndreu (CNR, Istituto di Fisiologia Clinica & QualiMedLab srl, Pisa) for providing data of the EQA “Immunocheck”.

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

References

1. Ferraro S, Panzeri A, Borille S, Szoke D, Panteghini M. Estimation of the reference interval for serum folate measured with assays traceable to the WHO International Standard. Clin Chem Lab Med 2017;55:e195–6.10.1515/cclm-2016-0941Search in Google Scholar PubMed

2. Hepburn S, Likhari T, Twomey PJ. Roche serum folate assay restandardization: an estimate of the new reference interval. Ann Clin Biochem 2019;56:183–4.10.1177/0004563218793159Search in Google Scholar PubMed

3. Solé-Enrech G, San-José P, Aliste-Fernández M, Cano-Corres R, Claver-Belver N, Palencia-Domínguez J, et al. Vitamin B12 and folate levels in a healthy population: establishing reference intervals. Clin Chem Lab Med 2019;57:e173–5.10.1515/cclm-2018-1080Search in Google Scholar PubMed

4. Cluitmans JC, van den Ouweland JM. Reference values of a new serum folate assay traceable to the WHO International Standard. Clin Chem Lab Med 2019;57:e176–8.10.1515/cclm-2018-1229Search in Google Scholar PubMed

5. Ferraro S, Mozzi R, Panteghini M. Tracing a roadmap for vitamin B12 testing using the health technology assessment approach. Clin Chem Lab Med 2014;52:767–77.10.1515/cclm-2013-0738Search in Google Scholar PubMed

6. Ferraro S, Panzeri A, Panteghini M. Tackling serum folate test in European countries within the health technology assessment paradigm: request appropriateness, assays, and health outcomes. Clin Chem Lab Med 2017;55:1262–75.10.1515/cclm-2016-0804Search in Google Scholar PubMed

7. WHO. Serum and red blood cell folate concentrations for assessing folate status in populations. Vitamin and Mineral Nutrition Information System. Geneva: World Health Organization, 2012. http://apps.who.int/iris/bitstream/10665/75584/1/WHO_NMH_NHD_EPG_12.1_eng.pdf. Accessed: Dec 2018.Search in Google Scholar

8. İspir E, Serdar MA, Ozgurtas T, Gulbahar O, Akın KO, Yesildal F, et al. Comparison of four automated serum vitamin B12 assays. Clin Chem Lab Med 2015;53:1205–13.10.1515/cclm-2014-0843Search in Google Scholar PubMed

9. Kristensen GB, Rustad P, Berg JP, Aakre KM. Analytical bias exceeding desirable quality goal in 4 out of 5 common immunoassays: results of a native single serum sample external quality assessment program for cobalamin, folate, ferritin, thyroid-stimulating hormone, and free T4 analyses. Clin Chem 2016;62:1255–63.10.1373/clinchem.2016.258962Search in Google Scholar PubMed

10. Raiten DJ, Fisher KD. Assessment of folate methodology used in the Third National Health and Nutrition Examination Survey (NHANESIII, 1988–1994). J Nutr 1995;125:1371S–98S.10.1093/jn/125.5.1371SSearch in Google Scholar PubMed

Published Online: 2019-02-02
Published in Print: 2019-07-26

©2019 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorials
  3. CCLM Award for the Most Cited Paper
  4. Folate and vitamin B12 assays after recalibration to the WHO International Standard 03/178: making the interpretation as simple as possible, but not simpler
  5. Reviews
  6. Blood contamination in salivary diagnostics: current methods and their limitations
  7. Central adrenal insufficiency: open issues regarding diagnosis and glucocorticoid treatment
  8. Genetics and Molecular Diagnostics
  9. Measuring the chronology of the translational process of molecular genetic discoveries
  10. Development and interlaboratory evaluation of a NIST Reference Material RM 8366 for EGFR and MET gene copy number measurements
  11. General Clinical Chemistry and Laboratory Medicine
  12. Post-translational modification-derived products are associated with frailty status in elderly subjects
  13. Urine chloride self-measurement to monitor sodium chloride intake in patients with chronic kidney disease
  14. Estimated urinary osmolality based on combined urinalysis parameters: a critical evaluation
  15. Measurement of S100B protein: evaluation of a new prototype on a bioMérieux Vidas® 3 analyzer
  16. Measuring thyroglobulin in patients with thyroglobulin autoantibodies: evaluation of the clinical impact of BRAHMS Kryptor® Tg-minirecovery test in a large series of patients with differentiated thyroid carcinoma
  17. Human chorionic gonadotropin suspected heterophile interference investigations in immunoassays: a recommended approach
  18. Certified reference material against PR3 ANCA IgG autoantibodies. From development to certification
  19. Diagnostic accuracy of a fully automated multiplex celiac disease antibody panel for serum and plasma
  20. Fasting serum bile acids concentration is associated with insulin resistance independently of diabetes status
  21. Hematology and Coagulation
  22. The association between activated protein C ratio and Factor V Leiden are gender-dependent
  23. Reference Values and Biological Variations
  24. Determination of sigma score based on biological variation for haemostasis assays: fit-for-purpose for daily practice?
  25. Calcitonin measurement in pediatrics: reference ranges are gender-dependent, validation in medullary thyroid cancer and thyroid diseases
  26. Cancer Diagnostics
  27. Uncovering the clinical impact of kallikrein-related peptidase 5 (KLK5) mRNA expression in the colorectal adenoma-carcinoma sequence
  28. Cardiovascular Diseases
  29. Performance of a novel high sensitivity cardiac troponin I assay in asymptomatic hemodialysis patients – evidence for sex-specific differences
  30. Infectious Diseases
  31. Rapid susceptibility testing of multi-drug resistant Escherichia coli and Klebsiella by glucose metabolization monitoring
  32. Letters to the Editor
  33. Vitamin B12 and folate levels in a healthy population: establishing reference intervals
  34. Reference values of a new serum folate assay traceable to the WHO International Standard
  35. Serum protein electrophoresis and complement deficiencies: a veteran but very versatile test in clinical laboratories
  36. Introduction of a novel ELISA assay for serum AMH determination
  37. Bone alkaline phosphatase on the IDS-iSYS automated analyser; cross-reactivity with intestinal ALP
  38. Evaluation of the MULTISURE HIV Rapid Test in a Korean population with low human immunodeficiency virus prevalence
  39. Vancomycin immunoassay: does the Advia Centaur XPT underestimate the exposure of patients? A method comparison study
  40. Hb Hunan and Hb Hengyang: Two unexpected discoveries during HbA1c measurements
  41. Analytical performance of a CE-marked immunoassay to quantify phosphorylated neurofilament heavy chains
  42. Implementation of an automated method for direct quantification of urinary ammonium
  43. Congress Abstracts
  44. Proceedings of ACBI 2018 41ST Annual Conference Association of Clinical Biochemists in Ireland
Downloaded on 27.10.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2019-0050/html
Scroll to top button