Startseite Medizin Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes
Artikel
Lizenziert
Nicht lizenziert Erfordert eine Authentifizierung

Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes

  • Gordana Juricic EMAIL logo , Ana Bakliza , Andrea Saracevic , Lara Milevoj Kopcinovic , Sanja Dobrijevic , Sandra Drmic und Ana-Maria Simundic
Veröffentlicht/Copyright: 22. Dezember 2015

Abstract

Background: Delayed sample processing can affect accurate glucose measurement. Our aim was to investigate the stability of glucose in samples collected in serum, sodium fluoride/potassium oxalate (NaF/KOx) and Glucomedics tubes processed according to different controlled pre-centrifugation delays (up to 180 min after venipuncture) in order to simulate prolonged sample transport between venipuncture and centrifugation.

Methods: Samples were collected from healthy volunteers (n=80) into either serum or NaF/KOx and Glucomedics tubes. Glucose concentration was measured in samples centrifuged immediately after venipuncture and compared with tubes processed with a delay of 60, 120 and 180 min prior to centrifugation. Differences between baseline and respective delayed centrifugation glucose value for each tube type were tested using the paired t-test. Mean bias calculated for each tube type and delay protocol was compared to recommended quality specifications for glucose (2.2%).

Results: Glucose concentrations measured in all three delayed tube types were lower in comparison to respective baseline glucose concentrations measured in immediately processed tube (p<0.001). The highest decrease in glucose was observed in serum tubes in all specified time points (p<0.001), while glucose was most stable in Glucomedics tubes (p<0.001). The decrease in glucose observed for serum and NaF/KOx tubes was clinically significant at all specified time points while the bias for Glucomedics tubes did not exceed the criteria even with a centrifugation delay of 180 min.

Conclusions: Glucose stability in un-centrifuged Glucomedics tubes is much superior to serum and NaF/KOx tubes. Glucomedics tubes can be left un-centrifuged for up to 3 h without affecting glucose concentration.


Corresponding author: Gordana Juricic, Department of Laboratory Diagnostics, General Hospital Pula, Aldo Negri 6, Pula, Croatia, Phone: +385 98 908 5506, E-mail:

Acknowledgments

We thank Marija Culjak for her technical skills in performing all venipunctures.

Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

Research funding: This study was supported by the Ministry of Science, Education, and Sports, Republic of Croatia (project number: 134-1340227-0200).

Employment or leadership: None declared.

Honorarium: None declared.

Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Beastall GH. The modernisation of pathology and laboratory medicine in the UK: networking into the future. Clin Biochem Rev 2008;29:3–10.Suche in Google Scholar

2. Lippi G, Simundic AM. Laboratory networking and sample quality: a still relevant issue for patient safety. Clin Chem Lab Med 2012;50:1703–5.10.1515/cclm-2012-0245Suche in Google Scholar

3. Uchida K, Matuse R, Toyoda E, Okuda S, Tomita S. A new method of inhibiting gycolysis in blood samples. Clin Chim Acta 1988;172:1–8.10.1016/0009-8981(88)90125-8Suche in Google Scholar

4. Gambino R, Piscitelli J, Ackattupathil TA, Theriault JL, Andrin RD, Sanfilippo ML, et al. Acidification of blood is superior to sodium fluoride alone as an inhibitor of glycolysis. Clin Chem 2009;55:1019–21.10.1373/clinchem.2008.121707Suche in Google Scholar PubMed

5. del Pino IG, Constanso I, Mourín LV, Safont CB, Vázquez PR. Citric/citrate buffer: an effective antiglycolytic agent. Clin Chem Lab Med 2013;51:1943–9.10.1515/cclm-2012-0735Suche in Google Scholar PubMed

6. Szoke D, Valente C, Panteghini M. Better blood collection tubes for plasma glucose: ready for prime time? Clin Chem Lab Med 2014;52:e87–9.10.1515/cclm-2013-1006Suche in Google Scholar PubMed

7. Fobker M. Stability of glucose in plasma with different anticoagulants. Clin Chem Lab Med 2014;52:1057–60.10.1515/cclm-2013-1049Suche in Google Scholar PubMed

8. Dimeski G, Yow KS, Brown NN. What is the most suitable blood collection tube for glucose estimation? Ann Clin Biochem 2014;52:270–5.10.1177/0004563214544708Suche in Google Scholar PubMed

9. Yagmur E, van Helden J, Koch A, Jadem J, Tacke F, Trautwein C. Effective inhibition of glycolysis in venous whole blood and plasma samples. Available at: http://dx.doi.org/10.1515/labmed-2013-0028. Accessed 1 March 2015.10.1515/labmed-2013-0028Suche in Google Scholar

10. van den Berg SA, Thelen MH, Salden LP, van Thiel SW, Boonen KJ. It takes acid, rather than ice, to freeze glucose. Scientific Reports 5, Article number: 8875, doi:10.1038/srep08875. Accessed 1 March 2015.10.1038/srep08875Suche in Google Scholar PubMed PubMed Central

11. Juricic G, Milevoj Kopcinovic L, Saracevic A, Bakliza A, Simundic AM. Liquid citrate acidification introduces significant glucose bias and leads to misclassification of patients with diabetes. Available at: http://dx.doi.org/10.1515/cclm-2015-0358. Accessed 3 July 2015.10.1515/cclm-2015-0358Suche in Google Scholar PubMed

12. Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Diabetes Care 2011;34:e61–99.10.2337/dc11-9998Suche in Google Scholar PubMed PubMed Central

13. Chan AY, Swaminathan R, Cockram KS. Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem 1989;35:315–7.10.1093/clinchem/35.2.315Suche in Google Scholar

14. Waring WS, Evans LE, Kirkpatrick CT. Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus. J Clin Pathol 2007;60:820–3.10.1136/jcp.2006.039925Suche in Google Scholar PubMed PubMed Central

15. Turchiano M, Nguyen C, Fierman A, Lifshitz M, Convit A. Impact of blood sample collection and processing methods on glucose levels in community outreach studies. Available at: http://dx.doi.org/10.1155/2013/256151. Accessed 29 August 2014.10.1155/2013/256151Suche in Google Scholar PubMed PubMed Central

16. Chan H, Lunt H, Thompson H, Heenan HF, Frampton CM, Florkowski CM. Plasma glucose measurement in diabetes: impact and implications of variations in sample collection procedures with a focus on the first hour after sample collection. Clin Chem Lab Med 2014;52:1061–8.10.1515/cclm-2013-1059Suche in Google Scholar PubMed

17. Narayan S. The preanalytic phase. An important component of laboratory medicine. Am J Clin Pathol 2000;113:429–52.Suche in Google Scholar

18. Fernandez L, Jee P, Klein MJ, Fisher P, Perkins SL, Brooks SP. A comparison of glucose concentrations in paired specimens collected in serum separator and fluoride/potassium oxalate blood collections tubes under survey ‘field’ conditions. Clin Biochem 2013;46:285–8.10.1016/j.clinbiochem.2012.11.027Suche in Google Scholar PubMed

19. Nikolac N, Supak-Smolcic V, Simundic AM, Celap I. Croatian Society of Medical Biochemistry and Laboratory Medicine: national recommendations for venous blood sampling. Biochem Med 2013;23:242–54.10.11613/BM.2013.031Suche in Google Scholar PubMed PubMed Central

20. Evacuated blood collection system for in vitro diagnostic use. Greiner Bio-One, Kremsmünster, Austria. 980200 Rev.15, 12-2013.Suche in Google Scholar

21. Harboe M. A method for determination of hemoglobin in plasma by near-ultraviolet spectrophotometry. Scand J Clin Lab Invest 1959;11:66–70.10.3109/00365515909060410Suche in Google Scholar PubMed

22. Noe DA, Weedn V, Bell WR. Direct spectrophotometry of serum hemoglobin: an Allen correction compared with a three-wavelength polychromatic analysis. Clin Chem 1984;30: 627–30.10.1093/clinchem/30.5.627Suche in Google Scholar

23. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochemia Medica 2014;24:31–44.Suche in Google Scholar

24. Package insert for Glucose on the Architect c Systems and the Aeroset system; Abbott Laboratories; January 2010; REF 3L82-21 and 3L82-41.Suche in Google Scholar

Received: 2015-7-31
Accepted: 2015-11-19
Published Online: 2015-12-22
Published in Print: 2016-3-1

©2016 by De Gruyter

Artikel in diesem Heft

  1. Frontmatter
  2. Editorial
  3. The way of prostate cancer diagnostics
  4. Review
  5. Statistical methods used in the calculation of geriatric reference intervals: a systematic review
  6. Opinion Paper
  7. The side effects of translational omics: overtesting, overdiagnosis, overtreatment
  8. Genetics and Molecular Diagnostics
  9. Rapid detection of non-deletional mutations causing α-thalassemia by multicolor melting curve analysis
  10. General Clinical Chemistry and Laboratory Medicine
  11. Patient pools and the use of “patient means” are valuable tools in quality control illustrated by a bone-specific alkaline phosphatase assay
  12. Long-term stability of glucose: 96-h study using Terumo Glycaemia tubes
  13. Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes
  14. Croatian laboratories have a good knowledge of the proper detection and management of hemolyzed, icteric and lipemic samples
  15. Fetal exposure to ethanol: relationship between ethyl glucuronide in maternal hair during pregnancy and ethyl glucuronide in neonatal meconium
  16. Comparing the effect of isotopically labeled or structural analog internal standards on the performance of a LC-MS/MS method to determine ciclosporin A, everolimus, sirolimus and tacrolimus in whole blood
  17. Relationship between matrix metalloproteinase-9 and oxidative stress in drug-free male schizophrenia: a case control study
  18. Comparison of functional fibrinogen (FF/CFF) and FIBTEM in surgical patients – a retrospective study
  19. Reference Values and Biological Variations
  20. Assessment of serum free light chain levels in healthy adults immediately after marathon running
  21. Pharmacokinetics of a novel dosing regimen of oral melatonin in critically ill patients
  22. Cancer Diagnostics
  23. An epidemiology-based model as a tool to monitor the outbreak of inappropriateness in tumor marker requests: a national scale study
  24. Comparative analysis of prostate cancer specific biomarkers PCA3 and ERG in whole urine, urinary sediments and exosomes
  25. Infectious Diseases
  26. Comparative evaluation of the Aptima HIV-1 Quant Dx assay and COBAS TaqMan HIV-1 v2.0 assay using the Roche High Pure System for the quantification of HIV-1 RNA in plasma
  27. Evaluating the use of procalcitonin in an asymptomatic, HIV-infected antiretroviral therapy-naïve, South African cohort
  28. Diabetes
  29. Early prediction of gestational diabetes: a practical model combining clinical and biochemical markers
  30. Letters to the Editors
  31. Highly-trained dogs’ olfactory system for detecting biochemical recurrence following radical prostatectomy
  32. More on the accuracy of the Architect enzymatic assay for hemoglobin A1c and its traceability to the IFCC reference system
  33. Assessing quality from an accuracy-based HbA1c proficiency survey
  34. HbG-Honolulu interferes with some cation-exchange HPLC HbA1c assays
  35. Between analyser differences in chloride measurements and thus anion gap cause different interpretations of the acid-base balance
  36. Non-albumininuric proteinuria: a urinary tubular marker in the diagnosis of diabetic kidney disease
  37. International Normalized Ratio (INR) testing: analytical and clinical performance of four point-of-care devices versus central laboratory instrumentation analysis
  38. Genetic screening of the makorin ring finger 3 gene in girls with idiopathic central precocious puberty
  39. Practical approach for medical validation of therapeutic drug monitoring results
  40. Marked elevation of procalcitonin in a patient with a drug related infusion reaction to rituximab
Heruntergeladen am 23.1.2026 von https://www.degruyterbrill.com/document/doi/10.1515/cclm-2015-0746/html?lang=de
Button zum nach oben scrollen