Abstract

Background: The aim of this study was to compare different analytical methods that are currently in use in the Netherlands for the measurement of whole blood vitamin B6.

Methods: This method comparison study consisted of two separate parts. (1) Four laboratories participated in a pilot study in which the commercial Chromsystems and INstruchemie method, and a laboratory developed liquid chromatography-tandem mass spectrometry (LC-MS/MS) method and HPLC method were compared. Sixty-nine frozen whole blood samples and six lyophilized whole blood samples were used for comparison. (2) In the nationwide part of the study, 49 laboratories participated in the analysis of three identical sets of two whole blood samples of which one set was freshly analyzed, one set was analyzed after storage at −20 °C and one set was analyzed after lyophilization.

Results: In both parts of the study, the HPLC and LC-MS/MS methods showed equivalent results for all sample types tested. The Chromsystems method showed a positive bias of 45% (pilot study) and 30% (nationwide study) towards the LC-MS/MS method when fresh or frozen samples were used. The measurement of lyophilized samples showed no differences between the methods. The results of the INstruchemie method were inconclusive due to the low number of participants.

Conclusions: The different analytical methods for measuring vitamin B6 produce different results when whole blood patient samples are measured. The recognition of a reference method or the development of suitable reference materials and quality control materials might serve as a first step towards improved standardization or harmonization of the whole blood vitamin B6 assay.


Corresponding author: Bertrand D. van Zelst, Department of Clinical Chemistry, Erasmus MC, PO Box 2040, 3000 CA Rotterdam, The Netherlands, Phone: +31107033587, Fax: +31104367894, E-mail:

References

1. Lindner A, Bankson DD, Stehman-Breen C, Mahuren JD, Coburn SP. Vitamin B6 metabolism and homocysteine in end-stage renal disease and chronic renal insufficiency. Am J Kidney Dis 2002;39:134–45.10.1053/ajkd.2002.29904Suche in Google Scholar

2. Lumeng L, Li TK. Vitamin B6 metabolism in chronic alcohol abuse. Pyridoxal phosphate levels in plasma and the effects of acetaldehyde on pyridoxal phosphate synthesis and degradation in human erythrocytes. J Clin Invest 1974;53:693–704.10.1172/JCI107607Suche in Google Scholar

3. Morris MS, Picciano MF, Jacques PF, Selhub J. Plasma pyridoxal 5′-phosphate in the US population: the National Health and Nutrition Examination Survey, 2003-2004. Am J Clin Nutr 2008;87:1446–54.10.1093/ajcn/87.5.1446Suche in Google Scholar

4. Friso S, Girelli D, Martinelli N, Olivieri O, Lotto V, Bozzini C, et al. Low plasma vitamin B-6 concentrations and modulation of coronary artery disease risk. Am J Clin Nutr 2004;79:992–8.10.1093/ajcn/79.6.992Suche in Google Scholar

5. Kelly PJ, Shih VE, Kistler JP, Barron M, Lee H, Mandell R, et al. Low vitamin B6 but not homocyst(e)ine is associated with increased risk of stroke and transient ischemic attack in the era of folic acid grain fortification. Stroke 2003;34:e51–4.10.1161/01.STR.0000071109.23410.ABSuche in Google Scholar

6. Nederlands_Huisartsen_Genootschap. Dementie Samenvattingskaart M21. NHG; 2012; Available from: https://www.nhg.org/standaarden/samenvatting/dementie#Richtlijnendiagnostiek.Suche in Google Scholar

7. van Zelst BD, de Jonge R. A stable isotope dilution LC-ESI-MS/MS method for the quantification of pyridoxal-5′-phosphate in whole blood. J Chromatogr B Analyt Technol Biomed Life Sci 2012;903:134–41.10.1016/j.jchromb.2012.07.007Suche in Google Scholar

8. Talwar D, Quasim T, McMillan DC, Kinsella J, Williamson C, O’Reilly DS. Optimisation and validation of a sensitive high-performance liquid chromatography assay for routine measurement of pyridoxal 5-phosphate in human plasma and red cells using pre-column semicarbazide derivatisation. J Chromatogr B Analyt Technol Biomed Life Sci 2003;792:333–43.10.1016/S1570-0232(03)00320-9Suche in Google Scholar

9. Bates CJ, Pentieva KD, Matthews N, Macdonald A. A simple, sensitive and reproducible assay for pyridoxal 5′-phosphate and 4-pyridoxic acid in human plasma. Clin Chim Acta 1999;280:101–11.10.1016/S0009-8981(98)00173-9Suche in Google Scholar

10. Bisp MR, Bor MV, Heinsvig EM, Kall MA, Nexo E. Determination of vitamin B6 vitamers and pyridoxic acid in plasma: development and evaluation of a high-performance liquid chromatographic assay. Anal Biochem 2002;305:82–9.10.1006/abio.2002.5638Suche in Google Scholar PubMed

11. Ericson KL, Mahuren JD, Zubovic YM, Coburn SP. Use of chlorite to improve HPLC detection of pyridoxal 5′-phosphate. J Chromatogr B Analyt Technol Biomed Life Sci 2005;823:218–20.10.1016/j.jchromb.2005.07.017Suche in Google Scholar PubMed

12. Gregory JF, 3rd. Determination of pyridoxal 5′-phosphate as the semicarbazone derivative using high-performance liquid chromatography. Anal Biochem 1980;102:374–9.10.1016/0003-2697(80)90170-0Suche in Google Scholar

13. Johansen K, Edlund PO. Determination of water-soluble vitamins in blood and plasma by coupled-column liquid chromatography. J Chromatogr 1990;506:471–9.10.1016/S0021-9673(01)91601-2Suche in Google Scholar

14. Kimura M, Kanehira K, Yokoi K. Highly sensitive and simple liquid chromatographic determination in plasma of B6 vitamers, especially pyridoxal 5′-phosphate. J Chromatogr A 1996;722:295–301.10.1016/0021-9673(95)00354-1Suche in Google Scholar

15. Rybak ME, Pfeiffer CM. Clinical analysis of vitamin B(6): determination of pyridoxal 5′-phosphate and 4-pyridoxic acid in human serum by reversed-phase high-performance liquid chromatography with chlorite postcolumn derivatization. Anal Biochem 2004;333:336–44.10.1016/j.ab.2004.06.036Suche in Google Scholar PubMed

16. Schrijver J, Speek AJ, Schreurs WH. Semi-automated fluorometric determination of pyridoxal-5′-phosphate (vitamin B6) in whole blood by high-performance liquid chromatography (HPLC). Int J Vitam Nutr Res 1981;51:216–22.Suche in Google Scholar

17. Midttun O, Hustad S, Solheim E, Schneede J, Ueland PM. Multianalyte quantification of vitamin B6 and B2 species in the nanomolar range in human plasma by liquid chromatography-tandem mass spectrometry. Clin Chem 2005;51:1206–16.10.1373/clinchem.2005.051169Suche in Google Scholar PubMed

18. van der Ham M, Albersen M, de Koning TJ, Visser G, Middendorp A, Bosma M, et al. Quantification of vitamin B6 vitamers in human cerebrospinal fluid by ultra performance liquid chromatography-tandem mass spectrometry. Anal Chim Acta 2012;712:108–14.10.1016/j.aca.2011.11.018Suche in Google Scholar PubMed

19. Rybak ME, Jain RB, Pfeiffer CM. Clinical vitamin B6 analysis: an interlaboratory comparison of pyridoxal 5′-phosphate measurements in serum. Clin Chem 2005;51:1223–31.10.1373/clinchem.2005.050278Suche in Google Scholar PubMed

20. Hoad KE, Johnson LA, Woollard GA, Walmsley TA, Briscoe S, Jolly LM, et al. Vitamin B1 and B6 method harmonization: comparison of performance between laboratories enrolled in the RCPA Quality Assurance Program. Clin Biochem 2013;46:772–6.10.1016/j.clinbiochem.2013.01.020Suche in Google Scholar PubMed

21. Steen G, Vlasveld LT, Poot CC, van der Slot-Verhoeven AJ, Castel A. Onderzoek naar referentiewaarden van laboratoriumonderzoek in een algemeen ziekenhuis: resultaten en bevindingen. Ned Tijdschr Klin Chem Labgeneesk 2009;34:35–43.Suche in Google Scholar


Supplemental Material:

The online version of this article (DOI: 10.1515/cclm-2015-0385) offers supplementary material, available to authorized users.


Received: 2015-4-22
Accepted: 2015-8-26
Published Online: 2015-10-10
Published in Print: 2016-4-1

©2016 by De Gruyter

Artikel in diesem Heft

  1. Frontmatter
  2. Editorial
  3. Macroprolactin: searching for a needle in a haystack?
  4. Review
  5. Laboratory medicine in the new healthcare environment
  6. Opinion Paper
  7. Developing GRADE outcome-based recommendations about diagnostic tests: a key role in laboratory medicine policies
  8. EFLM Survey
  9. Accreditation process in European countries – an EFLM survey
  10. Genetics and Molecular Diagnostics
  11. HB Puerta del Sol [HBA1:c.148A>C], HB Valdecilla [HBA2:c.3G>T], HB Gran Vía [HBA2:c.98T>G], HB Macarena [HBA2:c.358C>T] and HB El Retiro [HBA2:c.364_366dupGTG]: description of five new hemoglobinopathies
  12. General Clinical Chemistry and Laboratory Medicine
  13. A candidate reference method using ICP-MS for sweat chloride quantification
  14. Quantification of hemoglobin A1c by off-line HPLC separation and liquid chromatography-tandem mass spectrometry: a modification of the IFCC reference measurement procedure
  15. Comparison between B·R·A·H·M·S PCT direct, a new sensitive point-of-care testing device for rapid quantification of procalcitonin in emergency department patients and established reference methods – a prospective multinational trial
  16. Evaluation of a hand-held blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting
  17. Use of a faecal immunochemical test for haemoglobin can aid in the investigation of patients with lower abdominal symptoms
  18. A possible cause of the variable detectability of macroprolactin by different immunoassay systems
  19. A multicenter comparison of whole blood vitamin B6 assays
  20. Analytical and clinical performance of the new Fujirebio 25-OH vitamin D assay, a comparison with liquid chromatography-tandem mass spectrometry (LC-MS/MS) and three other automated assays
  21. Development and validation of a 2nd tier test for identification of purine nucleoside phosphorylase deficiency patients during expanded newborn screening by liquid chromatography-tandem mass spectrometry
  22. A cross-sectional study of biomarkers of exposure and effect in smokers and moist snuff consumers
  23. Reference Values and Biological Variations
  24. Pediatric reference value distributions and covariate-stratified reference intervals for 29 endocrine and special chemistry biomarkers on the Beckman Coulter Immunoassay Systems: a CALIPER study of healthy community children
  25. Development of reference intervals for serum alkaline phosphatase among adults in Southern China traced to the new IFCC reference measurement procedure
  26. Birth season predicts the values of red blood cell distribution width (RDW) in adulthood
  27. Cardiovascular Diseases
  28. Cardiovascular risk stratification in hemodialysis patients in the era of highly sensitive troponins: should we choose between hs-troponin I and hs-troponin T?
  29. A standardised FACS assay based on native, receptor transfected cells for the clinical diagnosis and monitoring of β1-adrenergic receptor autoantibodies in human heart disease
  30. Infectious Diseases
  31. The diagnostic value of serum fucosylated fetuin A in hepatitis B virus-related liver diseases
  32. Letters to the Editor
  33. Discriminant indices for distinguishing thalassemia and iron deficiency in patients with microcytic anemia
  34. Discriminant indices for distinguishing thalassemia and iron deficiency in patients with microcytic anemia: a reply
  35. STORE (Sample Tracking: Organization, Refrigeration and Expansion): a project in a clinical laboratory perspective
  36. Control materials for immunochromatographic strip used for α-thalassemia screening
  37. Comparison of a 10- vs. 15-min centrifugation time for chemical and immunochemical assays and impact on turnaround time in a hospital laboratory
  38. Pre-analytical phase in cryoglobulin (CRG) detection: an alternative method for sample transport
  39. The efficacy of an internet-based e-learning system using the CellaVision Competency Software for continuing professional development
  40. Clinical examinations after the Fukushima disaster: a case report of Soma General Hospital
  41. Non-fasting plasma glucose concentration in blood donors
  42. Analytical evaluation of a second generation assay for chromogranin A; a dual-site study
  43. UBE3A, c.1347_1348delGA: a mutation in question
Heruntergeladen am 12.9.2025 von https://www.degruyterbrill.com/document/doi/10.1515/cclm-2015-0385/html
Button zum nach oben scrollen