Abstract
Background: The immediate impact of blood gas test results on patient care favors the use of blood gas analyzers as point-of-care-testing (POCT) devices. We performed an analytical performance evaluation of four cartridge-type blood gas analyzers for the determination of pH, partial carbon dioxide pressure (pCO2), partial oxygen pressure (pO2), ionized calcium (iCa2+), potassium (K+), glucose, lactate and total hemoglobin (tHb), in comparison with a traditional blood gas analyzer.
Methods: The analyzers included in the study are: RP405, GEM Premier 4000, ABL90 FLEX and Cobas b 123. The ABL700 served as comparator. For each instrument the imprecision was estimated according to the CLSI EP5-A2. Based on the CLSI EP9-A2 evaluation protocol, a method comparison was performed using patient samples. Obtained data were compared against preset quality specifications, based on ABL700 performance and biological variation.
Results: The precision of the RP405 and ABL90 FLEX was remarkably better than the preset criteria based on ABL700 performance. The GEM appears to have the worst precision. The RP405 yielded the best overall performance, with exception for tHb and iCa2+. Noteworthy is the very good performance of the glucose determination on RP405. The ABL90 FLEX showed the best performance for pH, K+ and iCa2+ measurements. For tHb determination the Cobas b123 revealed the best results. Regarding practicability, all instruments were found to be user friendly.
Conclusions: Globally, all four cartridge-type blood gas analyzers produced clinically acceptable results. The analytical performance, together with the ease of use and low maintenance time of the instruments, demonstrate that these analyzers are perfectly suitable for both POCT and laboratory use.
©2012 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Masthead
- Masthead
- Editorial
- Laboratory maternal-fetal medicine: challenges and perspectives
- Reviews
- Candidate biochemical markers for screening of pre-eclampsia in early pregnancy
- Linking preeclampsia and cardiovascular disease later in life
- Mini Reviews
- Fetal nucleic acids in maternal blood: the promises
- Ethical aspects of the expansion of neonatal screening programme due to technological advances
- Prenatal screening for trisomy 21: recent advances and guidelines
- Ultrasound markers for the detection of women at risk of developing pre-eclampsia
- Are biological markers relevant for the diagnosis and the prognosis of preterm premature rupture of membranes (PPROM)?
- Genetics and Molecular Diagnostics
- Study of FTMT and ABCA4 genes in a patient affected by age-related macular degeneration: identification and analysis of new mutations
- Rapid detection of CFH (p.Y402H) and ARMS2 (p.A69S) polymorphisms in age-related macular degeneration using high-resolution melting analysis
- Unlabeled-probe high-resolution melting to detect KRAS codon 12 and 13 mutations in pancreatic adenocarcinoma tissues
- General Clinical Chemistry and Laboratory Medicine
- Volume, rates, source and types of add-on pathology test requests across five hospitals
- Schizophrenia is associated with increased levels of serum Fas and FasL
- Comparison between capillary, venous and arterial levels of protein S100B in patients with severe brain pathology
- Changes of serum adiponectin and soluble intercellular adhesion molecule-1 concentrations after smoking cessation
- Lack of consistency between two commercial ELISAs and against an in-house ELISA for the detection of CD36 in human plasma
- Comparison of a commercial urinary neopterin radioimmunoassay with high performance liquid chromatography
- Evaluation of Prevent ID and Quantum Blue rapid tests for fecal calprotectin
- Analytical performance evaluation of four cartridge-type blood gas analyzers
- Cancer Diagnostics
- Serum free light chains and oligoclonal bands in patients with multiple myeloma and autologous stem cell transplantation
- Cardiovascular Diseases
- Homocysteine as a determinant of left ventricular ejection fraction in patients with diabetes
- Plasma total homocysteine is a determinant of carotid intima-media thickness and circulating endothelial progenitor cells in patients with newly diagnosed hypertension
- Serum MMP-9 and TIMP-1 concentrations and MMP-9 activity during surgery-induced inflammation in humans
- Effects of endothelin-1 on endothelial progenitor cell function
- Investigation of cystatin C and cystatin C based estimated glomerular filtration rate in pregnant patients with heart failure
- Letters to the Editor
- Comments on a performance evaluation of cartridge-type blood gas analyzers
- Comments on a performance evaluation of cartridge-type blood gas analyzers. Reply
- Carryover can be a cause of false-positive results with the Beckman AccuTnI assay
- Method comparison of a 25-hydroxy vitamin D enzyme immunoassay to liquid chromatography tandem mass spectroscopy
- Bone marrow plasma cell separation – validation of separation algorithm
- Sharing our experience in improving postgraduate clinical biochemistry training with laboratory practice articles
Articles in the same Issue
- Masthead
- Masthead
- Editorial
- Laboratory maternal-fetal medicine: challenges and perspectives
- Reviews
- Candidate biochemical markers for screening of pre-eclampsia in early pregnancy
- Linking preeclampsia and cardiovascular disease later in life
- Mini Reviews
- Fetal nucleic acids in maternal blood: the promises
- Ethical aspects of the expansion of neonatal screening programme due to technological advances
- Prenatal screening for trisomy 21: recent advances and guidelines
- Ultrasound markers for the detection of women at risk of developing pre-eclampsia
- Are biological markers relevant for the diagnosis and the prognosis of preterm premature rupture of membranes (PPROM)?
- Genetics and Molecular Diagnostics
- Study of FTMT and ABCA4 genes in a patient affected by age-related macular degeneration: identification and analysis of new mutations
- Rapid detection of CFH (p.Y402H) and ARMS2 (p.A69S) polymorphisms in age-related macular degeneration using high-resolution melting analysis
- Unlabeled-probe high-resolution melting to detect KRAS codon 12 and 13 mutations in pancreatic adenocarcinoma tissues
- General Clinical Chemistry and Laboratory Medicine
- Volume, rates, source and types of add-on pathology test requests across five hospitals
- Schizophrenia is associated with increased levels of serum Fas and FasL
- Comparison between capillary, venous and arterial levels of protein S100B in patients with severe brain pathology
- Changes of serum adiponectin and soluble intercellular adhesion molecule-1 concentrations after smoking cessation
- Lack of consistency between two commercial ELISAs and against an in-house ELISA for the detection of CD36 in human plasma
- Comparison of a commercial urinary neopterin radioimmunoassay with high performance liquid chromatography
- Evaluation of Prevent ID and Quantum Blue rapid tests for fecal calprotectin
- Analytical performance evaluation of four cartridge-type blood gas analyzers
- Cancer Diagnostics
- Serum free light chains and oligoclonal bands in patients with multiple myeloma and autologous stem cell transplantation
- Cardiovascular Diseases
- Homocysteine as a determinant of left ventricular ejection fraction in patients with diabetes
- Plasma total homocysteine is a determinant of carotid intima-media thickness and circulating endothelial progenitor cells in patients with newly diagnosed hypertension
- Serum MMP-9 and TIMP-1 concentrations and MMP-9 activity during surgery-induced inflammation in humans
- Effects of endothelin-1 on endothelial progenitor cell function
- Investigation of cystatin C and cystatin C based estimated glomerular filtration rate in pregnant patients with heart failure
- Letters to the Editor
- Comments on a performance evaluation of cartridge-type blood gas analyzers
- Comments on a performance evaluation of cartridge-type blood gas analyzers. Reply
- Carryover can be a cause of false-positive results with the Beckman AccuTnI assay
- Method comparison of a 25-hydroxy vitamin D enzyme immunoassay to liquid chromatography tandem mass spectroscopy
- Bone marrow plasma cell separation – validation of separation algorithm
- Sharing our experience in improving postgraduate clinical biochemistry training with laboratory practice articles