Home Revisiting SFA stent technology: an updated overview on mechanical stent performance
Article
Licensed
Unlicensed Requires Authentication

Revisiting SFA stent technology: an updated overview on mechanical stent performance

  • Wolfram Schmidt EMAIL logo , Christoph Brandt-Wunderlich , Peter Behrens , Christoph Kopetsch , Klaus-Peter Schmitz , Julian Ramin Andresen and Niels Grabow
Published/Copyright: May 15, 2023

Abstract

Objectives

The study investigated mechanical parameters of stent systems indicated for treatment of femoropopliteal (FP) arterial disease to support interpretation of clinical results and the related causalities.

Methods

Eight stent system types of same dimensions were investigated (n=2). Parameters were the profile of stent delivery system (SDS), radiopacity, trackability and pushability, bending stiffness (flexibility) and axial stiffness of expanded stents, length change during expansion, radial force, crush resistance, strut thickness and general surface condition.

Results

The trackability ranged from 0.237 to 0.920 N and the pushability was 47.9–67.6 %. The bending stiffness of SDS was between 108.42 and 412.68 N mm2. The length change during stent release to 5 mm was low, with one exception. The bending stiffness of the expanded stents was 2.73–41.67 N mm2. The normalized radial forces at 5 mm diameter ranged from 0.133 N/mm to 0.503 N/mm. During non-radial compression by 50 %, the forces were 3.07–8.42 N, with one exception (58.7 N). The strut thickness was 153–231 µm.

Conclusions

Large differences occurred for flexibility, radial force and length change during expansion. The data should be used when choosing the proper device for restoring vascular function.


Corresponding author: Wolfram Schmidt, University Medical Center Rostock, Institute for Biomedical Engineering, Friedrich-Barnewitz-Str. 4, 18119 Rostock, Germany, E-mail:

Funding source: German Federal Ministry of Education and Research (BMBF)

Award Identifier / Grant number: RESPONSE

Acknowledgments

The authors thank the staff of the Testing Laboratory for Cardio + Vascular Devices at the Institute for ImplantTechnology and Biomaterials – IIB e.V. for assistance in performing the presented measurements.

  1. Research funding: Financial support by the German Federal Ministry of Education and Research (BMBF) within RESPONSE “Partnership for Innovation in Implant Technology” is gratefully acknowledged. The funding organization played no role in the study design, the collection, analysis and interpretation of data, in the writing of the report, or in the decision to submit the report for publication.

  2. Author contribution: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: Not applicable.

References

1. Sigvant, B, Wiberg-Hedman, K, Bergqvist, D, Rolandsson, O, Andersson, B, Persson, E, et al.. A population-based study of peripheral arterial disease prevalence with special focus on critical limb ischemia and sex differences. J Vasc Surg 2007;45:1185–91. https://doi.org/10.1016/j.jvs.2007.02.004.Search in Google Scholar PubMed

2. Dua, A, Lee, CJ. Epidemiology of peripheral arterial disease and critical limb ischemia. Tech Vasc Intervent Radiol 2016;19:91–5. https://doi.org/10.1053/j.tvir.2016.04.001.Search in Google Scholar PubMed

3. Shammas, NW. An overview of optimal endovascular strategy in treating the femoropopliteal artery: mechanical, biological, and procedural factors. Int J Angiol 2013;22:1–8. https://doi.org/10.1055/s-0032-1331840.Search in Google Scholar PubMed PubMed Central

4. Acin, F, de Haro, J, Bleda, S, Varela, C, Esparza, L. Primary nitinol stenting in femoropopliteal occlusive disease: a meta-analysis of randomized controlled trials. J Endovasc Ther 2012;19:585–95. https://doi.org/10.1583/jevt-12-3898r.1.Search in Google Scholar PubMed

5. Laird, JR, Katzen, BT, Scheinert, D, Lammer, J, Carpenter, J, Buchbinder, M, et al.. Nitinol stent implantation versus balloon angioplasty for lesions in the superficial femoral artery and proximal popliteal artery: twelve-month results from the RESILIENT randomized trial. Circ Cardiovasc Interv 2010;3:267–76. https://doi.org/10.1161/CIRCINTERVENTIONS.109.903468.Search in Google Scholar PubMed

6. Rastan, A, Krankenberg, H, Baumgartner, I, Blessing, E, Müller-Hülsbeck, S, Pilger, E, et al.. Stent placement vs. balloon angioplasty for popliteal artery treatment: two-year results of a prospective, multicenter, randomized trial. J Endovasc Ther 2015;22:22–7. https://doi.org/10.1177/1526602814564386.Search in Google Scholar PubMed

7. Krankenberg, H, Schlüter, M, Steinkamp, HJ, Bürgelin, K, Scheinert, D, Schulte, KL, et al.. Nitinol stent implantation versus percutaneous transluminal angioplasty in superficial femoral artery lesions up to 10 cm in length: the femoral artery stenting trial (FAST). Circulation 2007;116:285–92. https://doi.org/10.1161/CIRCULATIONAHA.107.689141.Search in Google Scholar PubMed

8. Burket, MW. SFA disease: deciding what to leave behind. Endovasc Today 2019;18:61–4.Search in Google Scholar

9. Dake, MD, Ansel, GM, Jaff, MR, Ohki, T, Saxon, RR, Smouse, HB, et al.. Durable clinical effectiveness with paclitaxel-eluting stents in the femoropopliteal artery: 5-year results of the zilver PTX randomized trial. Circulation 2016;133:1472–83. https://doi.org/10.1161/CIRCULATIONAHA.115.016900.Search in Google Scholar PubMed PubMed Central

10. Gray, W. Twelve-month results of the IMPERIAL randomized trial. In: Presented at: Transcatheter Cardiovascular Therapeutics (TCT); September 22. San Diego, CA; 2018.Search in Google Scholar

11. Schmidt, W, Wissgott, C, Andresen, R, Behrens, P, Schmitz, KP. Performance characteristics of modern self-expanding nitinol stents indicated for SFA. Röfo 2011;183:818–25. https://doi.org/10.1055/s-0031-1273445.Search in Google Scholar PubMed

12. Rits, J, van Herwaarden, JA, Jahrome, AK, Krievins, D, Moll, FL. The incidence of arterial stent fractures with exclusion of coronary, aortic, and non-arterial settings. Eur J Vasc Endovasc Surg 2008;36:339–45. https://doi.org/10.1016/j.ejvs.2008.05.005.Search in Google Scholar PubMed

13. Matsumoto, T, Matsubara, Y, Aoyagi, Y, Matsuda, D, Okadome, J, Morisaki, K, et al.. Radial force measurement of endovascular stents: influence of stent design and diameter. Vascular 2016;24:171–6. https://doi.org/10.1177/1708538115590040.Search in Google Scholar PubMed

14. Maleckis, K, Deegan, P, Poulson, W, Sievers, C, Desyatova, A, MacTaggart, J, et al.. Comparison of femoropopliteal artery stents under axial and radial compression, axial tension, bending, and torsion deformations. J Mech Behav Biomed Mater 2017;75:160–8. https://doi.org/10.1016/j.jmbbm.2017.07.017.Search in Google Scholar PubMed PubMed Central

15. Berti, F, Wang, P-J, Spagnoli, A, Pennati, G, Migliavacca, F, Edelman, ER, et al.. Nickel-Titanium peripheral stents: which is the best criterion for the multi-axial fatigue strength assessment? J Mech Behav Biomed Mater 2020;113:104142. https://doi.org/10.1016/j.jmbbm.2020.104142.Search in Google Scholar PubMed PubMed Central

16. Stoeckel, D, Bonsignore, C, Duda, S. A survey of stent designs. Minim Invasive Ther Allied Technol 2002;11:137–47. https://doi.org/10.1080/136457002760273340.Search in Google Scholar PubMed

17. Duda, SH, Wiskirchen, J, Tepe, G, Bitzer, M, Kaulich, TW, Stoeckel, D, et al.. Physical properties of endovascular stents: an experimental comparison. J Vasc Intervent Radiol 2000;11:645–54. https://doi.org/10.1016/s1051-0443(07)61620-0.Search in Google Scholar PubMed

18. Gong, XY, Pelton, AR, Duerig, TW, Rebelo, N, Perry, K. Finite element analysis and experimental evaluation of superelastic nitinol stent. In: Duerig, TW, Pelton, AR, editors. Proceedings of SMST-2003; 2003.Search in Google Scholar

19. Colombo, M, Corti, A, Gallo, D, Colombo, A, Antognoli, G, Bernini, M, et al.. Superficial femoral artery stenting: impact of stent design and overlapping on the local hemodynamics. Comput Biol Med 2022;143:105248. https://doi.org/10.1016/j.compbiomed.2022.105248.Search in Google Scholar PubMed

20. Mazzaccaro, D, Berti, F, Antonini, L, Pennati, G, Petrini, L, Migliavacca, F, et al.. Biomechanical interpretation of observed fatigue fractures of peripheral Nitinol stents in the superficial femoral arteries through in silico modelling. Med Hypotheses 2020;142:109771. https://doi.org/10.1016/j.mehy.2020.109771.Search in Google Scholar PubMed

21. Yang, H, Fortier, A, Horne, K, Mohammad, A, Banerjee, S, Han, HC. Investigation of stent implant mechanics using linear analytical and computational approach. Cardiovasc Eng Technol 2017;8:81–90. https://doi.org/10.1007/s13239-017-0295-0.Search in Google Scholar PubMed

22. Dordoni, E, Meoli, A, Wu, W, Dubini, G, Migliavacca, F, Pennati, G, et al.. Fatigue behaviour of Nitinol peripheral stents: the role of plaque shape studied with computational structural analyses. Med Eng Phys 2014;36:842–9. https://doi.org/10.1016/j.medengphy.2014.03.006.Search in Google Scholar PubMed

23. Ní Ghriallais, R, Bruzzi, M. A computational analysis of the deformation of the femoropopliteal artery with stenting. J Biomech Eng 2014;136:071003-1–10. https://doi.org/10.1115/1.4027329.Search in Google Scholar PubMed

24. Abbott Vascular. Supera peripheral stent system. EL2099533 Rev A; 2015 [Accessed 27 Jan 2017].Search in Google Scholar

25. Cardinal Health (Hg.). S.M.A.R.T. Control - Nitinol stent system (EU521-06/17); 2017 [Accessed 12 Mar 2019].Search in Google Scholar

26. Gore & Associates. GORE TIGRIS vascular stent. Instructions for use - multilingual. Hg. v. Gore & Associates; 2017 [Accessed 12 Mar 2019].Search in Google Scholar

27. Boston Scientific Corporation (Hg.). Innova vascular self-expanding stent system sell sheet. PI-322605-AA JUL2015; 2015 [Accessed 12 Mar 2019].Search in Google Scholar

28. Available from: https://www.veryanmed.com/uk/technology/harnessing-biomimicry [Accessed 25 June 2020].Search in Google Scholar

29. Available from: https://www.cardiovascular.abbott/us/en/hcp/products/peripheral-intervention/absolute-pro-vascular-self-expanding-stent-system.html [Accessed 25 June 2020].Search in Google Scholar

30. Medtronic Inc. (Hg.). Complete SE Vascular Stent System. Instructions for Use (M729425B001_Rev1B) [Accessed 26 Mar 2021].Search in Google Scholar

31. BIOTRONIK (Hg.). (2018) Astron Pulsar. Selbstexpandierender Stent/0,018"/OTW Indiziert für die Behandlung femoraler und infrapoplitealer Arterien. 355967/B/602 [Accessed 26 Mar 2021].Search in Google Scholar

32. Schmidt, W, Wissgott, C, Brandt-Wunderlich, C, Behrens, P, Schmitz, KP, Grabow, N, et al.. Biomechanics and clinical experience of a 3D biomimicking vascular stent. Current Directions in Biomedical Engineering 2018;4:135–9. https://doi.org/10.1515/cdbme-2018-0034.Search in Google Scholar

33. ASTM F640-20. Standard test methods for determining radiopacity for medical use; 2020.Search in Google Scholar

34. DIN 13273-7:2020-12. Catheters for medical use, Part 7 - determination of the x-ray attenuation of catheters, requirements and testing; 2020.Search in Google Scholar

35. Lanzer, P, Schmidt, W. Instrumentation for coronary interventions. In: Lanzer, P, editor PanVascular medicine. Berlin, Heidelberg: Springer Verlag; 2015:1979–2027 pp.10.1007/978-3-642-37078-6_67Search in Google Scholar

36. Wissgott, C, Brandt-Wunderlich, C, Kopetsch, C, Schmidt, W, Andresen, R. Initial clinical results and in vitro testing of the new CGuard MicroNet-covered “One-Size-Fits-All” carotid stent. J Endovasc Ther 2019;26:1526602819849078. https://doi.org/10.1177/1526602819849078.Search in Google Scholar PubMed

37. Brandt-Wunderlich, C, Schmidt, W, Grabow, N, Stiehm, M, Siewert, S, Andresen, R, et al.. Support function of self-expanding nitinol stents – are radial resistive force and crush resistance comparable? Curr Dir Biomed Eng 2019;5:465–7. https://doi.org/10.1515/cdbme-2019-0117.Search in Google Scholar

38. Wissgott, C, Scheer, F, Goermez, M, Andresen, R. Evaluation of safety and effectiveness of the Biomimics 3D stent system in the femoropopliteal arteries of patients with symptomatic peripheral arterial disease: single-centre 24 Months results. J Clin Med Res 2022;4:1–21. https://doi.org/10.37191/Mapsci-2582-4333-4(1)-108.Search in Google Scholar

39. Li, H, Rha, S-W, Choi, BG, Choi, SY, Moon, SK, Jang, WY, et al.. Impact of chronic outward force on arterial responses of proximal and distal of long superficial femoral artery stent. BMC Cardiovasc Disord 2021;21. https://doi.org/10.1186/s12872-021-02141-z.Search in Google Scholar PubMed PubMed Central

40. Beier, S, Ormiston, J, Webster, M, Cater, J, Norris, S, Medrano-Gracia, P, et al.. Hemodynamics in idealized stented coronary arteries: important stent design considerations. Ann Biomed Eng 2016;84:315–29. https://doi.org/10.1007/s10439-015-1387-3.Search in Google Scholar PubMed PubMed Central

41. Nikanorov, A, Schillinger, M, Zhao, H, Minar, E, Schwartz, LB. Assessment of self-expanding nitinol stent deformation after chronic implantation into the femoropopliteal arteries. EuroIntervention 2013;9:730–7. https://doi.org/10.4244/EIJV9I6A117.Search in Google Scholar PubMed

42. Smouse, HB, Nikanorov, A, LaFlash, D. Biomechanical forces in the femoropopliteal arterial segment. Endovasc Today 2005;60–6. In this issue.Search in Google Scholar

43. Yang, M, Shi, B, Ma, L, Yu, C, Zhang, X, Li, T, et al.. Treatment of atherosclerotic femoropopliteal artery disease with Supera interwoven nitinol stent: a real-world study in China. Ann Vasc Surg 2022;85:183–9. https://doi.org/10.1016/j.avsg.2022.02.008.Search in Google Scholar PubMed

44. Tao, MJ, Gotra, A, Tan, KT, Eisenberg, N, Roche-Nagle, G, Mafeld, S. SUPERA stenting in the common femoral artery: early experience and practical considerations. Vasc Endovascular Surg 2022;56:357–68. https://doi.org/10.1177/15385744211068648.Search in Google Scholar PubMed PubMed Central

45. Guzzardi, G, Spinazzola, A, Cangiano, G, Natrella, M, Paladini, A, Porta, C, et al.. Endovascular treatment of femoro-popliteal disease with the Supera stent: results of a multicenter study. J Public Health Res 2021;11. https://doi.org/10.4081/jphr.2021.2360.Search in Google Scholar PubMed PubMed Central

46. Bunte, MC, Cohen, DJ, Jaff, MR, Gray, WA, Magnuson, EA, Li, H, et al.. Long-term clinical and quality of life outcomes after stenting of femoropopliteal artery stenosis: 3-year results from the STROLL study. Catheter Cardiovasc Interv 2018;92:106–14. https://doi.org/10.1002/ccd.27569.Search in Google Scholar PubMed

47. Lichtenberg, M, Zeller, T, Gaines, P, Piorkowski, M. BioMimics 3D vascular stent system for femoropopliteal interventions. Vasa 2022;51:5–12. https://doi.org/10.1024/0301-1526/a000980.Search in Google Scholar PubMed

48. Lucatelli, P, Cini, M, Tommasino, G, Benvenuti, A, Guaccio, G, Bascetta, S, et al.. Use of the gore Tigris vascular stent in advanced femoropopliteal peripheral arterial disease. J Vasc Intervent Radiol 2018;29:614–22. https://doi.org/10.1016/j.jvir.2017.12.006.Search in Google Scholar PubMed

49. Han, D, Kang, S-H, Kim, S-H, Yoon, CH, Chae, IH. Comparison of patency between two different peripheral self-expandable stents, absolute Pro® versus complete SE® in femoropopliteal occlusive disease. Int Angiol 2019;38:305–11. https://doi.org/10.23736/S0392-9590.19.04102-6.Search in Google Scholar PubMed

50. Karashima, E, Takahara, M, Hozawa, K, Yamauchi, Y, Suzuki, K, Suematsu, N, et al.. Three-Year clinical outcomes of the Innova™ self-expanding nitinol stent for the treatment of femoropopliteal lesions. Cardiovasc Intervent Radiol 2021;44:1722–7. https://doi.org/10.1007/s00270-021-02960-9.Search in Google Scholar PubMed

51. Sarkadi, H, Bérczi, V, Kollár, A, Kiss, D, Jakabfi, P, Végh, E, et al.. Safety, clinical outcome, and fracture rate of femoropopliteal stenting using a 4F compatible delivery system. Eur J Vasc Endovasc Surg 2015;49:199–204. https://doi.org/10.1016/j.ejvs.2014.12.004.Search in Google Scholar PubMed

Received: 2022-10-21
Accepted: 2023-04-28
Published Online: 2023-05-15
Published in Print: 2023-10-26

© 2023 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 8.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/bmt-2022-0412/html?lang=en&srsltid=AfmBOoobh9t48NYMERsfO4XqKXwptMsIc2RklsJfYMG07v3NloceRLAC