Abstract
Background: Since the early 1960s, quantitative amino acid analysis (AAA) has traditionally been performed using ion-exchange chromatography with post-column ninhydrin derivatization. This established technology has many advantages, including relatively low cost of operation and ease of use. However, analysis times of 120+ min and high maintenance requirements make this technology unsuitable for the modern clinical laboratories with a requirement for rapid turnaround times. The work described here is a summary of our experiences with a new approach to full profile analysis of physiological amino acids.
Methods: Amino acids were derivatized in batches with a proprietary reagent, AccQTag®, which reacts with primary and secondary amines. The derivatized amino acids were separated using Ultraperformance® Liquid Chromatography (UPLC). In a prospective method comparison, quantitative plasma amino acid data obtained from approximately 170 patient samples using both the UPLC method and a traditional ion-exchange chromatography amino acid analyzer were evaluated.
Results: The data obtained from the two methods were found to agree well. Correlation coefficients for the most important amino acids seen in inborn errors of metabolism, such as phenylalanine, tyrosine and branched chain amino acids varied from 0.8658 to 0.9932 with minor slope biases. This approach also reduced the run time from 120 to 45 min per sample using a sample volume of 0.1 mL, compared to the 0.5 mL volume required for ion-exchange chromatography.
Conclusions: This new approach for the full profile analysis of physiological amino acids has been shown to be a viable substitute for current ion-exchange methodologies. It provides substantial benefits including significant reductions in runtime and necessary sample volume for the investigation and monitoring of patients with metabolic disorders and for nutritional management of a variety of patients.
©2011 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Editorial
- Method comparison: where do we draw the line?
- Review
- Current role of liquid chromatography coupled to mass spectrometry in clinical toxicology screening methods
- Minireview
- Molecular diagnostics for pharmacogenomic testing of fluoropyrimidine based-therapy: costs, methods and applications
- Opinion Paper
- Preanalytical quality improvement: from dream to reality
- Point
- Towards more complete specifications for acceptable analytical performance – a plea for error grid analysis
- Counterpoint
- Closing the brain-to-brain loop in laboratory testing
- General Clinical Chemistry and Laboratory Medicine
- Harmonization of growth hormone measurements with different immunoassays by data adjustment
- Comparison of the results from two different External Quality Assessment Schemes supports the utility of robust quality specifications
- High-throughput LC-MS/MS method for monitoring sirolimus and everolimus in the routine clinical laboratory
- Performance of the Roche Total Mycophenolic Acid® assay on the Cobas Integra 400®, Cobas 6000® and comparison to LC-MS/MS in liver transplant patients
- Diagnostic efficiency of truncated area under the curve from 0 to 2 h (AUC0–2) of mycophenolic acid in kidney transplant recipients receiving mycophenolate mofetil and concomitant tacrolimus
- Good performance of an immunoassay based method for nevirapine measurements in human breast milk
- Measurement of plasma amino acids by Ultraperformance® Liquid Chromatography
- Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
- Granularity Index of the SYSMEX XE-5000 hematology analyzer as a replacement for manual microscopy of toxic granulation neutrophils in patients with inflammatory diseases
- Direct chromogenic substrate immuno-capture activity assay for testing of factor VII-activating protease
- The antinuclear antibody assay: developing criteria for reflexive anti-dsDNA antibody testing in a laboratory setting
- A simple liquid chromatography-tandem mass spectrometry method for measuring metanephrine and normetanephrine in urine
- Monoclonal gammopathy missed by capillary zone electrophoresis
- Adaptation of the Diazyme Direct Enzymatic HbA1c Assay for a microplate reader at room temperature
- Validation and Outcome Studies
- Laboratory assessment of iron status in pregnancy
- Cardiovascular Diseases
- Blood cells characteristics as determinants of acute myocardial infarction
- No evidence for an association between the rs2824292 variant at chromosome 21q21 and ventricular fibrillation during acute myocardial infarction in a German population
- Letters to the Editor
- Letter to the Editor Reply: Statistical methods for assessment of added usefulness of new biomarkers
- Reply to Vavrova et al. Clin Chem Lab Med 2011;49:89–92
Articles in the same Issue
- Editorial
- Method comparison: where do we draw the line?
- Review
- Current role of liquid chromatography coupled to mass spectrometry in clinical toxicology screening methods
- Minireview
- Molecular diagnostics for pharmacogenomic testing of fluoropyrimidine based-therapy: costs, methods and applications
- Opinion Paper
- Preanalytical quality improvement: from dream to reality
- Point
- Towards more complete specifications for acceptable analytical performance – a plea for error grid analysis
- Counterpoint
- Closing the brain-to-brain loop in laboratory testing
- General Clinical Chemistry and Laboratory Medicine
- Harmonization of growth hormone measurements with different immunoassays by data adjustment
- Comparison of the results from two different External Quality Assessment Schemes supports the utility of robust quality specifications
- High-throughput LC-MS/MS method for monitoring sirolimus and everolimus in the routine clinical laboratory
- Performance of the Roche Total Mycophenolic Acid® assay on the Cobas Integra 400®, Cobas 6000® and comparison to LC-MS/MS in liver transplant patients
- Diagnostic efficiency of truncated area under the curve from 0 to 2 h (AUC0–2) of mycophenolic acid in kidney transplant recipients receiving mycophenolate mofetil and concomitant tacrolimus
- Good performance of an immunoassay based method for nevirapine measurements in human breast milk
- Measurement of plasma amino acids by Ultraperformance® Liquid Chromatography
- Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
- Granularity Index of the SYSMEX XE-5000 hematology analyzer as a replacement for manual microscopy of toxic granulation neutrophils in patients with inflammatory diseases
- Direct chromogenic substrate immuno-capture activity assay for testing of factor VII-activating protease
- The antinuclear antibody assay: developing criteria for reflexive anti-dsDNA antibody testing in a laboratory setting
- A simple liquid chromatography-tandem mass spectrometry method for measuring metanephrine and normetanephrine in urine
- Monoclonal gammopathy missed by capillary zone electrophoresis
- Adaptation of the Diazyme Direct Enzymatic HbA1c Assay for a microplate reader at room temperature
- Validation and Outcome Studies
- Laboratory assessment of iron status in pregnancy
- Cardiovascular Diseases
- Blood cells characteristics as determinants of acute myocardial infarction
- No evidence for an association between the rs2824292 variant at chromosome 21q21 and ventricular fibrillation during acute myocardial infarction in a German population
- Letters to the Editor
- Letter to the Editor Reply: Statistical methods for assessment of added usefulness of new biomarkers
- Reply to Vavrova et al. Clin Chem Lab Med 2011;49:89–92