A simple liquid chromatography-tandem mass spectrometry method for urinary free cortisol analysis: suitable for routine purpose
-
Silvia Persichilli
Abstract
Background: The best index of adrenal dysfunction is urinary free cortisol (UFC) measurements performed using a 24-h urine collection. This measurement is also useful in the investigation of Cushing's syndrome. In this paper, we report a simple and selective method for the analysis of UFC by liquid chromatography-tandem mass spectrometry (LC-MS/MS) suitable for use in a high-volume clinical laboratory routine. The results were compared to those obtained using a commercial immunoassay method used in our laboratory.
Methods: Urine samples containing 50 ng of internal standard (Cortisol-9,11,12,12-d4) were deproteinized using centrifugal filters with a molecular weight 10,000 Da cut-off and injected on a reversed phase column. Cortisol was analyzed in highly selective reaction monitoring in positive atmospheric pressure chemical ionization mode, at a resolution of 0.4 amu full width half maximum, and following the transitions related to the precursor 363.2 for cortisol and 367.2 for deuterated cortisol. The method validation included analysis of precision, linearity, sensitivity, recovery and interference from structurally similar steroids. UFC from 230 subjects was measured using LC-MS/MS and electrochemiluminescence immunoassay (ECLIA) methods.
Results: The calibration curves exhibited linearity and reproducibility in the range 7–10,000 nmol/L. Total imprecision was lower than 10%. The limit of detection and limit of quantification were 2 and 7 nmol/L, respectively. Mean recovery was higher than 90%. Structurally similar steroids do not interfere in the proposed method, but cause a significant change in the ECLIA results. Cortisol values obtained using the ECLIA method were always higher than those obtained using the LC-MS/MS method, with the bias directly proportional to cortisol concentrations. The reference values calculated using 180 normal subjects were 11–70 μg/day.
Conclusions: The proposed method is sensitive, simple, free from interferences and reliable for routine use.
Clin Chem Lab Med 2010;48:1433–7.
©2010 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Editorial
- Darwinian evolution or regression? The fate of laboratory professionals
- Review
- Automated reticulocyte counting: state of the art and clinical applications in the evaluation of erythropoiesis
- Minireview
- Laboratory diagnostics in acute poisoning: critical overview
- Opinion Papers
- On the origins of physicians: Darwinian or Lamarckian evolution?
- Observed, unknown distributions of clinical chemical quantities should be considered to be log-normal: a proposal
- Guidelines and Recommendations
- Development and description of GETT: a Genetic testing Evidence Tracking Tool
- Genetics and Molecular Diagnostics
- DNA sequencing errors in molecular diagnostics of filamin myopathy
- High resolution melting for the identification of mutations in the iron responsive element of the ferritin light chain gene
- General Clinical Chemistry and Laboratory Medicine
- Detection of serum free light chains: the problem with antigen excess
- Hepcidin concentrations and iron homeostasis in preeclampsia
- Automated assay for non-transferrin-bound iron in serum samples
- A simple liquid chromatography-tandem mass spectrometry method for urinary free cortisol analysis: suitable for routine purpose
- Evaluation of a multiplex assay for adipokine concentrations in obese children
- The use of serial patient blood gas, electrolyte and glucose results to derive biologic variation: a new tool to assess the acceptability of intensive care unit testing
- Point-of-care determination of neonatal bilirubin with the blood gas analyzer RapidLab 1265
- The effect of pre-analytical variables on light transmittance aggregometry in citrated platelet-rich plasma from healthy subjects
- Cancer Diagnostics
- Clinical significance of serum macrophage-colony stimulating factor (M-CSF) in esophageal cancer patients and its comparison with classical tumor markers
- VKORC1 haplotypes influence the performance characteristics of PIVKAII for screening of hepatocellular carcinoma
- Serum soluble tumour necrosis factor receptor type I concentrations independently predict prognosis in patients with breast cancer
- Variations in systemic biomarkers of oxidative/nitrosative stress and DNA damage before and during the consequent two cycles of chemotherapy in breast cancer patients
- Spurious increase in serum chromogranin A: the role of heterophilic antibodies
- Infectious Diseases
- Identification and evaluation of a new nucleic acid amplification test target for specific detection of Mycobacterium tuberculosis
- Application of an oligonucleotide microarray-based nano-amplification technique for the detection of fungal pathogens
- Liquid chromatography-tandem mass spectrometry method for simultaneous quantification of four triazole antifungal agents in human plasma
- Evaluation of two sirolimus assays using the ARCHITECT-i1000® CMIA or RxL® ACMIA methods in comparison with the IMx® MEIA method
- Widespread use of point-of-care testing is irreconcilable with the present-day quest for safety
- Congress Abstracts
- 3rd Annual Congress of the Austrian Society for Laboratory Medicine and Clinical Chemistry Salzburg, Austria, 27-30 October 2010
Articles in the same Issue
- Editorial
- Darwinian evolution or regression? The fate of laboratory professionals
- Review
- Automated reticulocyte counting: state of the art and clinical applications in the evaluation of erythropoiesis
- Minireview
- Laboratory diagnostics in acute poisoning: critical overview
- Opinion Papers
- On the origins of physicians: Darwinian or Lamarckian evolution?
- Observed, unknown distributions of clinical chemical quantities should be considered to be log-normal: a proposal
- Guidelines and Recommendations
- Development and description of GETT: a Genetic testing Evidence Tracking Tool
- Genetics and Molecular Diagnostics
- DNA sequencing errors in molecular diagnostics of filamin myopathy
- High resolution melting for the identification of mutations in the iron responsive element of the ferritin light chain gene
- General Clinical Chemistry and Laboratory Medicine
- Detection of serum free light chains: the problem with antigen excess
- Hepcidin concentrations and iron homeostasis in preeclampsia
- Automated assay for non-transferrin-bound iron in serum samples
- A simple liquid chromatography-tandem mass spectrometry method for urinary free cortisol analysis: suitable for routine purpose
- Evaluation of a multiplex assay for adipokine concentrations in obese children
- The use of serial patient blood gas, electrolyte and glucose results to derive biologic variation: a new tool to assess the acceptability of intensive care unit testing
- Point-of-care determination of neonatal bilirubin with the blood gas analyzer RapidLab 1265
- The effect of pre-analytical variables on light transmittance aggregometry in citrated platelet-rich plasma from healthy subjects
- Cancer Diagnostics
- Clinical significance of serum macrophage-colony stimulating factor (M-CSF) in esophageal cancer patients and its comparison with classical tumor markers
- VKORC1 haplotypes influence the performance characteristics of PIVKAII for screening of hepatocellular carcinoma
- Serum soluble tumour necrosis factor receptor type I concentrations independently predict prognosis in patients with breast cancer
- Variations in systemic biomarkers of oxidative/nitrosative stress and DNA damage before and during the consequent two cycles of chemotherapy in breast cancer patients
- Spurious increase in serum chromogranin A: the role of heterophilic antibodies
- Infectious Diseases
- Identification and evaluation of a new nucleic acid amplification test target for specific detection of Mycobacterium tuberculosis
- Application of an oligonucleotide microarray-based nano-amplification technique for the detection of fungal pathogens
- Liquid chromatography-tandem mass spectrometry method for simultaneous quantification of four triazole antifungal agents in human plasma
- Evaluation of two sirolimus assays using the ARCHITECT-i1000® CMIA or RxL® ACMIA methods in comparison with the IMx® MEIA method
- Widespread use of point-of-care testing is irreconcilable with the present-day quest for safety
- Congress Abstracts
- 3rd Annual Congress of the Austrian Society for Laboratory Medicine and Clinical Chemistry Salzburg, Austria, 27-30 October 2010