Limits of preservation of samples for urine strip tests and particle counting

  • Timo Kouri , Outi Malminiemi , Joris Penders , Virpi Pelkonen , Lotta Vuotari und Joris Delanghe
Veröffentlicht/Copyright: 10. März 2008
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

Background: Preservation of urine samples is important for centralised laboratory services with automated instruments.

Methods: A multicentre evaluation was carried out to assess preservative tubes from BD Diagnostics-Pre-analytical Systems and from Greiner Bio-One for test strip reading (documented at the level of remission values), for particle counting by flow cytometers (UF-100) and for visual microscopy. Failures were expressed as percentages of originally positive samples beyond a two-fold change (+100% or –50%) from the original values.

Results: The preservative-containing BD Plus C&S plastic, BD Plus UAP and Greiner Stabilur tubes succeeded in preservation of test strip results for 6–24 h (exceptions were glucose and nitrite tests). Greiner boric acid tube showed false negative results in leukocyte, protein and ketone strip tests immediately after adding the preservative. Urine red blood cell counts (with Sysmex UF-100) were preserved for 5 h in BD Plus C&S plastic and Greiner Stabilur tubes (Greiner tubes having clearly larger preservative-related background). Bacteria or white blood cell counting succeeded in BD Plus C&S plastic tubes for 5 or 24 h, respectively, but up to 72 h in Greiner Stabilur tubes. In visual microscopy, the Greiner Stabilur tube was slightly better than the BD Plus C&S plastic tube.

Conclusions: Urine specimens can be transported at +20°C on the day of collection if preserved properly. Longer delays need careful planning with current preservatives. Flow cytometry with UF-100 is sensitive to non-dissolved preservative remnants.

Clin Chem Lab Med 2008;46:703–13.


Corresponding author: Dr. Timo T. Kouri, MD, Docent, Oulu University Hospital, Laboratory Administration, P.O. Box 50, 90029 OYS, Finland Phone: +358-40-7730537 (mobile), +358-8-3154640 (office), Fax: +358-8-315 5541,

Received: 2007-10-30
Accepted: 2008-1-8
Published Online: 2008-03-10
Published in Print: 2008-05-01

©2008 by Walter de Gruyter Berlin New York

Artikel in diesem Heft

  1. Editorial
  2. Comparability of tumor marker immunoassays: still an important issue for clinical diagnostics?
  3. Review
  4. The diagnostic role of autoantibodies in the prediction of organ-specific autoimmune diseases
  5. Validation and Outcome Studies
  6. Alternative antibody for the detection of CA125 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® OV Monitor assay on the UniCel® DxI 800 Immunoassay System
  7. Alternative antibody for the detection of CA19-9 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® GI Monitor assay on the UniCel® DxI 800 Immunoassay System
  8. Alternative antibody for the detection of CA15-3 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® BR Monitor assay on the UniCel® DxI 800 Immunoassay System
  9. Toward metrological traceability in the determination of prostate-specific antigen (PSA): calibrating Beckman Coulter Hybritech Access PSA assays to WHO standards compared with the traditional Hybritech standards
  10. Multicentre evaluation of a new point-of-care test for the determination of CK-MB in whole blood
  11. Down's syndrome screening: population statistic dependency of screening performance
  12. General Clinical Chemistry and Laboratory Medicine
  13. Association between plasma thiols and immune activation marker neopterin in stable coronary heart disease
  14. Role for mitochondrial uncoupling protein-2 (UCP2) in hyperhomocysteinemia and venous thrombosis risk?
  15. Plasma procalcitonin measured by time-resolved amplified cryptate emission (TRACE) in liver transplant patients. A prognosis marker of early infectious and non-infectious postoperative complications
  16. Interactions of lipoprotein(a) with diabetes mellitus, apolipoprotein B and cholesterol enhance the prognostic values for coronary artery disease
  17. Association of thyroid stimulating hormone and coronary lipid risk factors with lipid peroxidation in hypothyroidism
  18. The effect of the mode of delivery on the maternal-neonatal carnitine blood levels and antioxidant status
  19. Time to reconsider the clinical value of immunoglobulin G4 to foods?
  20. Reference Values
  21. Determination of reference intervals for 26 commonly measured biochemical analytes with consideration of long-term within-individual variation
  22. Beverage-specific effects of ethanol consumption on its biological markers
  23. Limits of preservation of samples for urine strip tests and particle counting
  24. Reference ranges and diagnostic thresholds of laboratory markers of cardiac damage and dysfunction in a population of apparently healthy black Africans
  25. Letters to the Editor
  26. External quality assessment schemes for real-time PCR: a statistical procedure to corrective actions
  27. Application of two different microarray-based copy-number detection methodologies – array-comparative genomic hybridization and array-multiplex amplifiable probe hybridization – with identical amplifiable target sequences
  28. Criticism to the article: “Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: I. Analyte definition and proposal of a candidate reference method.” Authors: J.O. Jeppsson et al. Clin Chem Lab Med 2007;45(4):558–562
  29. Standardization of carbohydrate-deficient transferrin: reply to the letter by Tagliaro and Bortolotti
  30. Comment on: Bellini C, Serra G, Risso D, Mazzella M, Bonioli E. Reliability assessment of glucose measurement by HemoCue analyser in a neonatal intensive care unit. Clin Chem Lab Med 2007;45(11):1549–54
  31. Reliability assessment of glucose measurement by HemoCue analyzer in a neonatal intensive care unit: reply to Dr. Joakim Hagvik
Heruntergeladen am 15.9.2025 von https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2008.122/html
Button zum nach oben scrollen