Multicentre evaluation of a new point-of-care test for the determination of CK-MB in whole blood
-
Evangelos Giannitsis
Abstract
Background: The point-of-care (POC) test Roche CARDIAC CK-MB is a new assay which has been developed for the existing Roche Cardiac reader system.
Methods: We performed a multicentre evaluation at six sites to assess the analytical performance of the POC CK-MB assay and to compare it with a quantitative laboratory CK-MB assay.
Results: Within-series coefficients of variation (CV) resulting from 34 ten-fold measurements with patient samples ranged from 4.3% to 16.4%. Using quality control material, the mean CV values for day-to-day imprecision were 6.5% for the low level control and 8.4% for the high level control. Based upon 847 pairs of values, the mean relative bias of three independently calibrated lots of the POC CK-MB assay ranged from –6% to –11% in method comparisons with the lab CK-MB assay. The mean relative lot-to-lot differences of POC CK-MB were between –2% and +1%. No interference was observed with lipaemic blood (triglyceride concentrations up to 8.1 mmol/L), icteric blood (bilirubin concentrations up to 513 μmol/L), haemolytic blood (haemoglobin concentrations up to 0.12 mmol/L), biotin (up to 30 mg/L) and rheumatoid factor (up to 119 IU/mL), or with 53 standard or cardiological drugs even in toxic concentrations. There was no influence on the results by varying haematocrit values in the range from 21% to 54%. A slight interference with human anti-mouse antibodies type 2 was found. No significant influence on the results with POC CK-MB was found by using sample volumes between 135 and 165 μL. High CK-MB concentrations above the measuring range of POC CK-MB (1–40 μg/L) did not lead to false low results due to potential high-dose hook effect. No significant effect of sample age on recovery occurred up to a sample age of 24 h. No cross-reactivity was found between the POC CK-MB assay and either CK-MM or CK-BB. A substudy with healthy individuals confirmed the reference limits of 3.8 μg/L for females and 6.7 μg/L for males.
Conclusions: The POC CK-MB assay showed a very good analytical performance with an excellent concordance with the calibration and reference laboratory method. It should be therefore suitable for its intended use in POC settings.
Clin Chem Lab Med 2008;46:630–8.
©2008 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Editorial
- Comparability of tumor marker immunoassays: still an important issue for clinical diagnostics?
- Review
- The diagnostic role of autoantibodies in the prediction of organ-specific autoimmune diseases
- Validation and Outcome Studies
- Alternative antibody for the detection of CA125 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® OV Monitor assay on the UniCel® DxI 800 Immunoassay System
- Alternative antibody for the detection of CA19-9 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® GI Monitor assay on the UniCel® DxI 800 Immunoassay System
- Alternative antibody for the detection of CA15-3 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® BR Monitor assay on the UniCel® DxI 800 Immunoassay System
- Toward metrological traceability in the determination of prostate-specific antigen (PSA): calibrating Beckman Coulter Hybritech Access PSA assays to WHO standards compared with the traditional Hybritech standards
- Multicentre evaluation of a new point-of-care test for the determination of CK-MB in whole blood
- Down's syndrome screening: population statistic dependency of screening performance
- General Clinical Chemistry and Laboratory Medicine
- Association between plasma thiols and immune activation marker neopterin in stable coronary heart disease
- Role for mitochondrial uncoupling protein-2 (UCP2) in hyperhomocysteinemia and venous thrombosis risk?
- Plasma procalcitonin measured by time-resolved amplified cryptate emission (TRACE) in liver transplant patients. A prognosis marker of early infectious and non-infectious postoperative complications
- Interactions of lipoprotein(a) with diabetes mellitus, apolipoprotein B and cholesterol enhance the prognostic values for coronary artery disease
- Association of thyroid stimulating hormone and coronary lipid risk factors with lipid peroxidation in hypothyroidism
- The effect of the mode of delivery on the maternal-neonatal carnitine blood levels and antioxidant status
- Time to reconsider the clinical value of immunoglobulin G4 to foods?
- Reference Values
- Determination of reference intervals for 26 commonly measured biochemical analytes with consideration of long-term within-individual variation
- Beverage-specific effects of ethanol consumption on its biological markers
- Limits of preservation of samples for urine strip tests and particle counting
- Reference ranges and diagnostic thresholds of laboratory markers of cardiac damage and dysfunction in a population of apparently healthy black Africans
- Letters to the Editor
- External quality assessment schemes for real-time PCR: a statistical procedure to corrective actions
- Application of two different microarray-based copy-number detection methodologies – array-comparative genomic hybridization and array-multiplex amplifiable probe hybridization – with identical amplifiable target sequences
- Criticism to the article: “Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: I. Analyte definition and proposal of a candidate reference method.” Authors: J.O. Jeppsson et al. Clin Chem Lab Med 2007;45(4):558–562
- Standardization of carbohydrate-deficient transferrin: reply to the letter by Tagliaro and Bortolotti
- Comment on: Bellini C, Serra G, Risso D, Mazzella M, Bonioli E. Reliability assessment of glucose measurement by HemoCue analyser in a neonatal intensive care unit. Clin Chem Lab Med 2007;45(11):1549–54
- Reliability assessment of glucose measurement by HemoCue analyzer in a neonatal intensive care unit: reply to Dr. Joakim Hagvik
Articles in the same Issue
- Editorial
- Comparability of tumor marker immunoassays: still an important issue for clinical diagnostics?
- Review
- The diagnostic role of autoantibodies in the prediction of organ-specific autoimmune diseases
- Validation and Outcome Studies
- Alternative antibody for the detection of CA125 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® OV Monitor assay on the UniCel® DxI 800 Immunoassay System
- Alternative antibody for the detection of CA19-9 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® GI Monitor assay on the UniCel® DxI 800 Immunoassay System
- Alternative antibody for the detection of CA15-3 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access® BR Monitor assay on the UniCel® DxI 800 Immunoassay System
- Toward metrological traceability in the determination of prostate-specific antigen (PSA): calibrating Beckman Coulter Hybritech Access PSA assays to WHO standards compared with the traditional Hybritech standards
- Multicentre evaluation of a new point-of-care test for the determination of CK-MB in whole blood
- Down's syndrome screening: population statistic dependency of screening performance
- General Clinical Chemistry and Laboratory Medicine
- Association between plasma thiols and immune activation marker neopterin in stable coronary heart disease
- Role for mitochondrial uncoupling protein-2 (UCP2) in hyperhomocysteinemia and venous thrombosis risk?
- Plasma procalcitonin measured by time-resolved amplified cryptate emission (TRACE) in liver transplant patients. A prognosis marker of early infectious and non-infectious postoperative complications
- Interactions of lipoprotein(a) with diabetes mellitus, apolipoprotein B and cholesterol enhance the prognostic values for coronary artery disease
- Association of thyroid stimulating hormone and coronary lipid risk factors with lipid peroxidation in hypothyroidism
- The effect of the mode of delivery on the maternal-neonatal carnitine blood levels and antioxidant status
- Time to reconsider the clinical value of immunoglobulin G4 to foods?
- Reference Values
- Determination of reference intervals for 26 commonly measured biochemical analytes with consideration of long-term within-individual variation
- Beverage-specific effects of ethanol consumption on its biological markers
- Limits of preservation of samples for urine strip tests and particle counting
- Reference ranges and diagnostic thresholds of laboratory markers of cardiac damage and dysfunction in a population of apparently healthy black Africans
- Letters to the Editor
- External quality assessment schemes for real-time PCR: a statistical procedure to corrective actions
- Application of two different microarray-based copy-number detection methodologies – array-comparative genomic hybridization and array-multiplex amplifiable probe hybridization – with identical amplifiable target sequences
- Criticism to the article: “Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: I. Analyte definition and proposal of a candidate reference method.” Authors: J.O. Jeppsson et al. Clin Chem Lab Med 2007;45(4):558–562
- Standardization of carbohydrate-deficient transferrin: reply to the letter by Tagliaro and Bortolotti
- Comment on: Bellini C, Serra G, Risso D, Mazzella M, Bonioli E. Reliability assessment of glucose measurement by HemoCue analyser in a neonatal intensive care unit. Clin Chem Lab Med 2007;45(11):1549–54
- Reliability assessment of glucose measurement by HemoCue analyzer in a neonatal intensive care unit: reply to Dr. Joakim Hagvik