On Accuracy and Precision of a HPLC Method for Measurement of Urine Porphyrin Concentrations

  • Floris M.J. Zuijderhoudt , Stephan G.A. Koehorst , Willem E. Kluitenberg and Jenneke Dorresteijn-de Bok
Published/Copyright: June 1, 2005
Become an author with De Gruyter Brill
Clinical Chemistry and Laboratory Medicine (CCLM)
From the journal Volume 38 Issue 3

Abstract

We studied the accuracy and precision of a HPLC method for determination of porphyrins in urine. A commercial standard solution appeared to contain less porphyrins than indicated by the manufacturer, since calibration resulted in lower concentrations of uroporphyrin and coproporphyrin: 16% and 8%, respectively. Coefficients of variation for the measurement of uro-, hepta-, copro I and copro III porphyrins in samples of patients with and without porphyria were often much less than 15%. Comparison of measurements with and without calibrated standards revealed differences for uroporphyrin and coproporphyrin of 27% and 5%, respectively. Recovery of added uroporphyrin and coproporphyrin was 99%. The main cause of the variability in test results was apparently the improperly calibrated standard solutions. The precision of porphyrin measurements was not influenced by the type of porphyria.

: