A New Modular Chemiluminescence Immunoassay Analyser Evaluated

  • , , , , , and
Published/Copyright: June 1, 2005
Clinical Chemistry and Laboratory Medicine (CCLM)
From the journal Volume 38 Issue 3

Abstract

Thyrotropin (TSH), free thyroxine (fT4) and testosterone assays have been used as a probe to evaluate the performances of a new modular chemiluminescence (CL) immunoassay analyser, the Abbott Architect™ i2000. The evaluation was run in parallel on other systems that use CL as the detection reaction: DPC Immulite®, Chiron Diagnostics ACS-180 and ACS Centaur (TSH functional sensitivity only). TSH functional sensitivity was 0.0012, 0.009, 0.033 and 0.039 mU/l for the Architect, Immulite, ACS Centaur and ACS-180, respectively. Testosterone functional sensitivity was 0.38, 3.7 and 2.0 nmol/l for Architect, Immulite and ACS-180, respectively. Good correlation was obtained between the ACS-180 and Architect for all assays. The Immulite correlation did not agree well with the Architect or ACS-180 for fT4 and testosterone but was in good agreement for TSH. Regarding fT4 and testosterone, equilibrium dialysis and isotopic dilution gas-chromatography mass-spectrometry (GC-MS) respectively were used as reference methods. For both within- and between-run precision, the Architect showed the best reproducibility for all three analytes (CV < 6%).

: