Evaluation of the automated coagulation analyzer CS-5100 and its utility in high throughput laboratories

  • Franz Ratzinger , Klaus G. Schmetterer , Helmuth Haslacher , Thomas Perkmann , Sabine Belik und Peter Quehenberger EMAIL logo
Veröffentlicht/Copyright: 10. April 2014
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

Background: Automated analyzers are an important component of modern laboratories. As a representative of the newest generation of coagulation analyzers, the CS-5100 features several technical refinements including a pre-analytical assessment unit as well as multi-wavelength optical detection units. Therefore, the CS-5100 is supposed to rapidly and accurately perform a broad panel of coagulation tests. In the current study, the CS-5100 was evaluated regarding its precision and practicability in a clinical laboratory setting.

Methods: The CS-5100 was evaluated regarding its intra- and inter-assay precision using commercially available control samples. Results of patient samples, including hemolytic, icteric and lipemic specimens, measured on the CS-5100 were compared to reference analyzers, which are used in our accredited laboratory.

Results: The coefficients of variation, assessed in the intra- and inter-assay precision analyses were below 5% representatively for most parameters. Results, obtained by the CS-5100 showed predominantly a high comparability to used reference analyzers, with correlation coefficients ranging from 0.857 to 0.990. Only minor ranged systemic or proportional differences were found in Passing-Bablok regression between the CS-5100 and reference analyzers regarding most of the tested parameters. Lipemic samples had a tendency to deteriorate correlation coefficients, but an overall effect of the sample’s triglyceride level could be ruled out. In a routine setting, the analyzer reached a sample throughput rate of 160 tests per hour.

Conclusions: The CS-5100 is able to rapidly and precisely measure patient samples. No considerable influence on test comparability was found for elevated levels of free hemoglobin, bilirubin or triglycerides.


Corresponding author: Peter Quehenberger, Department of Laboratory Medicine, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria, Phone: +43 1 40400 5383, Fax: +43 1 40400 5392, E-mail:

Acknowledgments

We are grateful that Siemens provide required reagents for measuring on the CS-5100.

Conflict of interest statement

Authors’ conflict of interest disclosure: The authors stated that there are no conflicts of interest regarding the publication of this article. Research support played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

Research funding: None declared.

Employment or leadership: None declared.

Honorarium: None declared.

References

1. Söderberg J, Jonsson PA, Wallin O, Grankvist K, Hultdin J. Haemolysis index – an estimate of preanalytical quality in primary health care. Clin Chem Lab Med 2009;47:940–4.10.1515/CCLM.2009.227Suche in Google Scholar

2. Lippi G, Blanckaert N, Bonini P, Green S, Kitchen S, Palicka V, et al. Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories. Clin Chem Lab Med 2008;46:764–72.10.1515/CCLM.2008.170Suche in Google Scholar

3. Favaloro EJ, Lippi G, Adcock DM. Preanalytical and postanalytical variables: the leading causes of diagnostic error in hemostasis? Semin Thromb Hemost 2008;34:612–34.10.1055/s-0028-1104540Suche in Google Scholar

4. Forsman RW. Why is the laboratory an afterthought for managed care organizations? Clin Chem 1996;42:813–6.10.1093/clinchem/42.5.813Suche in Google Scholar

5. Passing H, Bablok. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 1983;21:709–20.Suche in Google Scholar

6. Martin Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurment. Lancet 1986;327:307–10.10.1016/S0140-6736(86)90837-8Suche in Google Scholar

7. Hanneman SK. Design, analysis, and interpretation of method comparison studies. AACN Adv Crit Care 2008;19: 223–34.Suche in Google Scholar

8. Pearson ECFHOHES. Tests for rank correlation coefficients. I. Biometrika 1957;44:470–81.10.1093/biomet/44.3-4.470Suche in Google Scholar

9. Richard L. Gorsuch CS. Correlation coefficients: mean bias and confidence interval distortions. J Methods Meas Soc Sci 2010;1:52–65.Suche in Google Scholar

10. Fisher RA. Frequency distribution of the values of the correlation coefficient in samples from an indefinitely large population. Biometrika 1915;10:507–21.10.2307/2331838Suche in Google Scholar

11. Salvagno GL, Lippi G, Bassi A, Poli G, Guidi GC. Prevalence and type of pre-analytical problems for inpatients samples in coagulation laboratory. J Eval Clin Pract 2008;14:351–3.10.1111/j.1365-2753.2007.00875.xSuche in Google Scholar PubMed

12. Green SF. The cost of poor blood specimen quality and errors in preanalytical processes. Clin Biochem 2013;46:1175–9.10.1016/j.clinbiochem.2013.06.001Suche in Google Scholar PubMed

13. Fischer F, Appert-Flory A, Jambou D, Toulon P. Evaluation of the automated coagulation analyzer Sysmex CA-7000. Thromb Res 2006;117:721–9.10.1016/j.thromres.2005.06.012Suche in Google Scholar PubMed

14. Quehenberger P, Kapiotis S, Handler S, Ruzicka K, Speiser W. Evaluation of the automated coagulation analyzer SYSMEX CA 6000. Thromb Res 1999;96:65–71.10.1016/S0049-3848(99)00069-9Suche in Google Scholar

15. Appert-Flory A, Fischer F, Jambou D, Toulon P. Evaluation and performance characteristics of the automated coagulation analyzer ACL TOP. Thromb Res 2007;120:733–43.10.1016/j.thromres.2006.12.002Suche in Google Scholar PubMed

16. Milos M, Herak DC, Zadro R. Discrepancies between APTT results determined with different evaluation modes on automated coagulation analyzers. Int J Lab Hematol 2010;32:33–9.10.1111/j.1751-553X.2008.01111.xSuche in Google Scholar PubMed

17. Molenaar PJ, Leyte A. Pre-acquisition system assessment of the Sysmex(®) Coagulation System CS-2100i and comparison with end-user verification; a model for the regional introduction of new analysers and methods. Clin Chem Lab Med 2011;49:1479–89.Suche in Google Scholar

18. de Bie P, Schornagel WJ, van den Dool EJ, Bakker B, van Dam W, Heckman M, et al. Laboratory evaluation of the Coasys® Plus C coagulation analyzer. Thromb Res 2013;131:357–62.10.1016/j.thromres.2013.02.004Suche in Google Scholar PubMed

19. Mullier F, Vanpee D, Jamart J, Dubuc E, Bailly N, Douxfils J, et al. Comparison of five D-dimer reagents and application of an age-adjusted cut-off for the diagnosis of venous thromboembolism in emergency department. Blood Coagul Fibrinolysis 2013 Nov 15. [Epub ahead of print].10.1097/MBC.0000000000000020Suche in Google Scholar PubMed

20. Park SJ, Chi HS, Chun SH, Jang S, Park CJ. Evaluation of performance including influence by interfering substances of the innovance D-dimer assay on the Sysmex coagulation analyzer. Ann Clin Lab Sci 2011;41:20–4.Suche in Google Scholar

21. Jennings I, Woods TA, Kitchen DP, Kitchen S, Walker ID. Laboratory D-dimer measurement: improved agreement between methods through calibration. Thromb Haemost 2007;98:1127–35.10.1055/s-0037-1613742Suche in Google Scholar

22. Milos M, Herak D, Kuric L, Horvat I, Zadro R. Evaluation and performance characteristics of the coagulation system: ACL TOP analyzer – HemosIL reagents. Int J Lab Hematol 2009;31:26–35.10.1111/j.1751-553X.2007.00999.xSuche in Google Scholar

23. Flanders MM, Crist R, Safapour S, Rodgers GM. Evaluation and performance characteristics of the STA-R coagulation analyzer. Clin Chem 2002;48:1622–4.10.1093/clinchem/48.9.1622Suche in Google Scholar

24. Magari RT. Bias estimation in method comparison studies. J Biopharm Stat 2004;14:881–92.10.1081/BIP-200035450Suche in Google Scholar PubMed

25. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135–60.10.1177/096228029900800204Suche in Google Scholar PubMed

26. Bilic-Zulle L. Comparison of methods: Passing and Bablok regression. Biochem Med 2011;21:49–52.10.11613/BM.2011.010Suche in Google Scholar

Received: 2013-12-18
Accepted: 2014-3-13
Published Online: 2014-4-10
Published in Print: 2014-8-1

©2014 by Walter de Gruyter Berlin/Boston

Artikel in diesem Heft

  1. Frontmatter
  2. Editorial
  3. Colorectal cancer and screening programs: not only analytical issues
  4. Reviews
  5. Laboratory diagnostics of inherited platelet disorders
  6. Reticulated platelets: analytical aspects and clinical utility
  7. Genetics and Molecular Diagnostics
  8. Advanced tools for BRCA1/2 mutational screening: comparison between two methods for large genomic rearrangements (LGRs) detection
  9. General Clinical Chemistry and Laboratory Medicine
  10. Establishing, harmonizing and analyzing critical values in a large academic health center
  11. Standardization of DiaSorin and Roche automated third generation PTH assays with an International Standard: impact on clinical populations
  12. First fully automated immunoassay for anti-Müllerian hormone
  13. A multicenter evaluation of dysthyroxinemia in a defined patient cohort
  14. New biomarkers in diagnosis of early onset preeclampsia and imminent delivery prognosis
  15. Soluble Fms-like tyrosine kinase-1 to placental growth factor ratio in mid-pregnancy as a predictor of preterm preeclampsia in asymptomatic pregnant women
  16. Development of a new immunoassay for the detection of ethyl glucuronide (EtG) in meconium: validation with authentic specimens analyzed using LC-MS/MS. Preliminary results
  17. Optimizing centrifugation of coagulation samples in laboratory automation
  18. Evaluation of the automated coagulation analyzer CS-5100 and its utility in high throughput laboratories
  19. A new sampling device for faecal immunochemical testing: haemoglobin stability is still an open issue
  20. Reference Values
  21. Faecal haemoglobin concentrations vary with sex and age, but data are not transferable across geography for colorectal cancer screening
  22. Cancer Diagnostics
  23. Enhanced miR-182 transcription is a predictor of poor overall survival in colorectal adenocarcinoma patients
  24. Importance of promoter methylation of GATA4 and TP53 genes in endometrioid carcinoma of endometrium
  25. Frequent methylation of HOXA9 gene in tumor tissues and plasma samples from human hepatocellular carcinomas
  26. Letters to the Editor
  27. Further comments on “Critical review of laboratory investigations in clinical practice guidelines: proposals for the description of investigation”
  28. A questionnaire study among nurses: awareness of blood and urine sample collection procedures
  29. Measurement uncertainty and clinical interpretation of measurement results
  30. Laboratory automation: how will you select the boarding assays?
  31. Improvement and evaluation of a 1,2-dioleoylglycerol method for measuring pancreatic lipase catalytic activity in serum
  32. The novel variant p.Ser465Leu in the PCSK9 gene does not account for the decreased LDLR activity in members of a FH family
  33. 1,5 Anhydroglucitol serum concentration as a biomarker for screening gestational diabetes in early pregnancy
  34. A rare condition: IgE type monoclonal gammopathy of undetermined significance
  35. Laboratory analysis of intraosseous blood: bad to the bone?
Heruntergeladen am 9.11.2025 von https://www.degruyterbrill.com/document/doi/10.1515/cclm-2013-1094/html
Button zum nach oben scrollen