Startseite Evaluation of BD Vacutainer Eclipse and BD Vacutainer Ultra-Touch butterfly blood collecting sets in laboratory testing
Artikel Open Access

Evaluation of BD Vacutainer Eclipse and BD Vacutainer Ultra-Touch butterfly blood collecting sets in laboratory testing

  • Saadet Kader ORCID logo EMAIL logo , Adnan Kirmit , Turan Akdağ ORCID logo und Mustafa Agah Tekindal ORCID logo
Veröffentlicht/Copyright: 23. August 2021

Abstract

Objectives

Choosing the right device for blood collection is an issue that is frequently encountered in laboratory medicine. The purpose of this study was to compare the quality of the samples collected in terms of hemolysis index, pain intensity and tube filling rates of the samples drawn with BD Vacutainer® EclipseNeedle and BD Vacutainer Ultra-Touch™ Butterfly Blood Collection Set, which is currently used in laboratories.

Methods

Blood samples were drawn from the 44 apparently healthy adult individuals from different arm sites by routine phlebotomy into Citrate, SST and EDTA tubes in Karapınar State Hospital. K, P, AST, LDH tests and hemolysis index were analyzed by Mindray BS-800 chemistry analyzer. Complete blood count (CBC) was determined by Mindray BC 6000.

Results

There was no significant difference in K, P, AST, LDH and CBC values between the devices for hemolysis (p>0.05). No significant difference was found out with respect to visual hemolysis and hemolysis index, tube filling speed and pain intensity between the devices.

Conclusions

Our findings demonstrate that two sets may used for venipuncture blood collection without creating additional hemolysis risk.

Introduction

Most laboratory specialists can ignore the possibility of factors other than laboratory conditions contributing in test results particularly with abnormal values. An important part of clinical and laboratory evidence suggests that most of the laboratory errors occur at the preanalytical phase (70%), emphasizing the need to acquire or apply more accurate methods to solve and classify possible traps in this basic stage of laboratory work [1].

Most of the problems in preanalytical phase is related to the factors (especially hemolysis) are associated with collecting samples immediately. Causes of pre-analytical hemolysis are associated with sample collection, inappropriate transportation methods, extreme temperature, sample handling, delayed processing, and prolonged storage [2], [3], [4].

Hemolysis can notably effect the reliability of test results – especially potassium (K), aspartate aminotransferase (AST), lactate dehydrogenase (LDH), phosphorus (P) and bilirubin etc. – and requires to repeat the tests. Repeating the test can increase patient discomfort, hospital costs, and working hours.

In hemolyzed samples, biochemistry test results may not reflect the patient’s clinical condition, either due to spilling of red blood cell contents into the serum or plasma (in fact changing the plasma concentration of the analytes) or due to red blood cell contents interfering with the testing methodology (i.e., hemoglobin could discolor a solution and effect spectrophotometric absorbance measurements at certain wavelengths). Overall, the analytes which are frequently affected by this primary interference mechanism are K, LDH, and AST with increased concentrations. Many of other analytes can increase or decrease based on the testing methodology; most laboratories examine the degree of hemolysis in the sample and have protocols in compliance with the laboratory manager’s instructions (for example: the hemoglobin (H), icterus (I) and lipaemia (L)/(HIL) index, in turn, corresponds to the approximate concentration of hemoglobin, bilirubin, or lipid present in the sample) to address resulting hemolyzed samples. Hemolyzed samples can significantly affect the delivery and quality of health care [5], [6], [7].

Improper design or use of blood collection devices can adversely affect the accuracy of laboratory test results. Vascular access devices such as catheters and needles may lead to cell lysis by exerting shear forces across the blood stream. Components from blood collection tubes such as stoppers, lubricants, surfactants and separating gels can leak into samples and/or adsorb analytes from the sample; special tube additives can also change analyte stability. Due to these interactions with blood samples, blood collection sets are a potential source of pre-analytical errors in laboratory tests. Therefore, selection of device for blood collection becomes the most important consideration in optimizing the preanalytical phase and achieving consistent results [8]. While the main determinants have preanalytical interchangeability, the blood collection technique has a significant impact on the reliability of laboratory test results. The butterfly collection set, which is a small needle attached to the elastic plastic wings and connected with an extension flexible tube, can be considered a reliable alternative to the classic straight needle for collecting blood in selected patient groups. In fact, an adapter can simply be attached so that it can fit into a vacuum needle holder and a vacuum system. Blood collection with the butterfly blood collection set can be extremely difficult or easier and less painful to relieve the supportive hand (venipuncture in the hand, leg, heel or skull) in newborns, children, small animals and patients with small, hard and atypical venous access. Moreover, the use of a butterfly collection set which is less frightening due to the reduced size of the needle, may make sense in some other situations, especially when approaching nervous or anxious patients. Ease of use for unprofessional or inexperienced laboratory technicians and nurses is an extra advantage of this system because the needle does not need to be held once when in the vein. The disadvantages of the butterfly collection sets are high cost, increased chance of needle damage, and there is a small chance of blood spillage as the needle is retracted.

There are very few studies on optimal methods for collecting blood samples from catheters [9]. In the study performed by Lippi et al. [10], the degree of hemolysis assessed by measurement of free plasma hemoglobin, LDH and AST did not appear to be higher between a conventional straight needle and butterfly set.

In this study, we evaluated the quality of samples collected using the BD Vacutainer® Eclipse™ Blood Collection Needle compared to the samples collected using the BD Vacutainer® Ultra-Touch™ Butterfly Blood Collection Set for selected serum chemistry analyte performance, and complete blood count parameters (CBC). In addition, an assessment was made for the potential of hemolysis as measured by comparisons of changes in potassium, aspartate transaminase and lactate dehydrogenase concentration as well as the hemolysis index, pain scale and by visual inspectation.

Materials and methods

The blood samples were collected from the 44 apparently healthy adult subjects by routine phlebotomy into BD Vacutainer® Citrate (Ref. No: 363803, Lot No: 914612, 2.7 mL/13 × 75 mm), BD Vacutainer®SST™ (Ref. No: 367955, Lot No: 9127605, 5.0 mL/13 × 100 mm) and BD Vacutainer® EDTA (Ref. No: 368841, Lot No: 9043866, 2.0 mL/13 × 75 mm) tubes in Karapınar State Hospital. A total of six tubes of blood were collected from each subject as; three tubes using the BD Vacutainer® Eclipse™ Blood Collection Needle (Ref. No: 3290672, Lot No: 9080926, 21Gx1-1/4″) and three tubes using the BD Vacutainer® Ultra-Touch™ Butterfly Blood Collection Set (Ref. No: 367393, Lot No: 8332922, 21Gx3/4″ × 7″).

Venipunctures were performed according to a randomization schedule which accounted for randomization of device and arm. All blood samples were drawn by the same phlebotomist and blood collection was performed from two arms at the same time with needle and butterfly set 10 s between them consecutively. The study tubes were collected in the recommended order of draw BD Citrate (3–4 Tube inversions, 1,500 relative centrifugal force [RCF], 15 min), BD SST™ (5 Tube inversions, 1,300 RCF, 10 min), BD EDTA (8–10 Tube inversions) [11] and handled and processed according to the test plan.

BD SST™ tubes were inverted 5 times immediately after collection and clotted for 30 min in an upright position. Following centrifugation at 1,300 RCF for 10 min, the tubes were evaluated for visual hemolysis. The hemolysis index was also measured on the Mindray BS-800 (Mindray Bio-Medical Electronics Co., Ltd, Shenzhen, China). Serum was tested for 4 selected routine chemistry analytes on the Mindray BS-800 (within 8 h after centrifugation) (Table 1).

Table 1:

Summary statistics.

Mindray BS-800
Analyte/unit Device type n Mean SD Bias, % Range obtained
[min–max]
AST, U/L BD Eclipse 44 19.7 4.5 −0.8 10.5–28.9
BD UTBCS 44 19.5 4.4 10.6–28.5
K, mmol/L BD Eclipse 44 4.2 0.2 −0.9 3.7–4.8
BD UTBCS 44 4.2 0.2 3.7–4.7
LDH, U/L BD Eclipse 44 201.5 41.8 3.8 117.9–285.1
BD UTBCS 44 205.3 39.2 126.91–283.8
P, mmol/L BD Eclipse 44 3.0 0.4 0 2.1–3.9
BD UTBCS 44 3.0 0.4 2.0–4.0
  1. BD UTBCS, BD Vacutainer®-Ultra-Touch™ Butterfly Blood Collection Set; BD Eclipse, BD Eclipse: Blood Collection Needle with Pre-attached Holder; PLT, Platelet; n, number of samples; SD, standard deviation; K, potassium; AST, aspartate aminotransferase; P, phosphor; LDH, lactate dehydrogenase.

BD EDTA tubes were inverted 8–10 times immediately after collection and visually inspected for clot formation. Specimens were analyzed for platelets (PLT) on the Mindray BC-6000 (Mindray Bio-Medical Electronics Co., Ltd, Shenzhen, China) within the same day as that of collection and evaluated for visual hemolysis.

The presence of hemolysis was detected in serum by K, AST, P and LDH results, which have been shown to be widely accepted indicators of hemolysis [12, 13]. In addition, visual hemolysis was assessed for each tube in the study. In addition, the Universal Pain Assessment Tool (UPAT) [14] was also used to evaluate the level of pain by looking at changes in individuals’ faces, and to assess the level of pain to harmonize them with UPAT. Tube filling rate is assessed for 5 mL SST tubes for each blood collection device by the help of a chronomometer.

The study protocol was approved by the local ethics committee of Scientific Research Committee of Konya Provincial Health Directorate (Approval number: 86737044-806.01.03).

Statistical analysis

For discrete and continuous variables, descriptive statistics (mean, standard deviation, median, minimum value, maximum value, and percentile) were given. In addition, the homogeneity of the variances, which is one of the prerequisites of parametric tests, was checked through Levene’s test. The assumption of normality was tested via the Shapiro–Wilk test. To compare the differences between the two groups, the Student’s t test was used when the parametric test prerequisites were fulfilled, and the Mann–Whitney U test was used when such prerequisites were not fulfilled. If the differences between the two dependent groups meet the pre-conditions of the parametric test, Paired t test; if not, it was evaluated with the Wilcoxon test. The relations among the parametric tests were evaluated by Spearman correlation coefficient. Statistical compliance between measurements was given by two-way random class correlation coefficient and concordance coefficient. The compatibility and bias of the devices were examined with Youden and Bland Altman graphics. p value <0.05 was considered to be statistically significant for every analysis. Statistical package program Syntax Function SPPS 20 (IBM Corp. Released 2011. IBM SPSS Statistics for Windows, Version 20.0, Armonk, NY: IBM Corp.) was used to evaluate the data.

Results

Mean biases are found respectively for platelet count (−0.9%), platelet cluster (−1.6%), AST (−0.8%), K (−0.9), LDH (1.9%), P (0.0%), hemolysis (−4.5%), lipemia (11.8%), icterus (−2.8%). Also these parameters have values as platelet count (10%), AST (10%), K (0.3 mmol/L), LDH (20%), P (0.4 mmol/L) with a predetermined clinical acceptance limit (CAL) [15], which represents an estimate of the maximum allowable difference or change in test results on average. The summary statistics for each tube type and each device type are shown in Tables 1 and 2. The mean and SD are listed to one more decimal place than that reported by the analyzer. We found agreement in all parameter measurements between BD UTBCS and BD Eclipse™ as shown in Table 3 (p<0.05). Bland–Altman plots have showed the 95% agreement interval in the set of differences between samples was always within the current analytical quality specifications for desirable bias derived from biological variation (Figure 1).

Table 2:

Summary statistics for Mindray BC-6000.

Analyte Device type n Mean SD Bias, % Range obtained (min–max)
PLT (×103/µL) BD Eclipse 44 255.4 65.1 −0.9 125–385.7
BD UTBCS 44 253.0 60.4 132.1–373.8
  1. BD UTBCS, BD Vacutainer®-Ultra-Touch™ Butterfly Blood Collection Set; BD Eclipse, BD Eclipse: Blood Collection Needle with Pre-attached Holder; PLT, Platelet; n, number of samples; SD, standard deviation.

Table 3:

ICC values and confidence intervals (95%) of the hemogram parameters measured on both devices.

ICC 95% Lower (ICC) 95% Upper (ICC) p
BD Eclipse (Visual Hemolysis)

BD UTBCS (Visual Hemolysis)

X1–Y1
0.6484 0.3556 0.8081 0.001
BD Eclipse (Instrument Hemolysis Index)

BD UTBCS (Instrument Hemolysis Index)

X2–Y2
0.8099 0.6517 0.8963 0.001
BD Eclipse (Visual Lipemia)

BD UTBCS (Visual Lipemia)

X3–Y3
0.7648 0.5689 0.8716 0.001
BD Eclipse (Instrument Lipemia Index)

BD UTBCS (Instrument Lipemia Index)

X4–Y4
0.9722 0.9491 0.9848 0.001
BD Eclipse (Instrument Icterus Index)

BD UTBCS (Instrument Icterus Index)

X6–Y6
0.9385 0.8873 0.9665 0.001
BD Eclipse K (potassium)

BD UTBCS K (potassium)

X7–Y7
0.7664 0.5719 0.8725 0.001
BD Eclipse (LDH (Lactate dehydrogenase)

BD UTBCS (LDH (Lactate dehydrogenase)

X8–Y8
0.8818 0.7834 0.9355 0.001
BD Eclipse AST (Aspartate aminotransferase)

BD UTBCS AST (Aspartate aminotransferase)

X9–Y9
0.9757 0.9555 0.9867 0.001
BD Eclipse (P) Phosphor

BD UTBCS (P) Phosphor

X10–Y10
0.9897 0.9812 0.9944 0.001
BD Eclipse (Instrument platelet cluster)

BD UTBCS (Instrument platelet cluster)

X12–Y12
0.6792 0.4121 0.825 0.001
BD Eclipse (PLT) platelets

BD UTBCS(PLT) platelets

X16–Y16
0.9741 0.9521 0.986 0.001
BD Eclipse Tube filling speed

BD UTBCS Tube filling speed

X17–Y17
0.344 −0.2023 0.642 0.001
BD Eclipse Pain Asking Patient

BD UTBCS Pain Asking Patient

X18–Y18
0.1766 −0.5089 0.5507 0.001
BD Eclipse Universal Pain Assessment Tool

BD UTBCS Universal Pain Assessment Tool

X19–Y19
0.4427 −0.02136 0.6959 0.001
  1. ICC, intraclass correlation coefficient; X, BD Eclipse; Y, BD UTBCS; BD Eclipse, Blood Collection Needle with Pre-attached Holder; BD UTBCS, UltraTouch™ Butterfly Blood Collection Set. Parameters with significant difference between two analyzer in the same samples with ICC<90.

Figure 1: 
Blot–Altman graphics among all parameters.
Figure 1:

Blot–Altman graphics among all parameters.

Discussion

Laboratory testing is an integral part of medical decision and an appropriate phlebotomy procedure is an important factor in collecting sufficient samples for analysis. This can be particularly challenging with the variable skill sets of blood-drawing medical staff, failure to follow appropriate procedures may result in poor sample quality, for example, in vitro hemolysis is the most common problem [8]. Inappropriate samples (hemolyzed, insufficient, activated or clotted) can account for more than 80% of errors occurring in the preanalytic phase due to blood draw problems from patients [1]. As these expensive phlebotomy systems are often used excessively to draw blood for routine laboratory tests, there is a common agreement on the problem of using intravenous catheters and butterfly collection sets. While a high risk of obtaining inappropriate specimens is frequently associated with the material used for venipuncture small catheters [16], there are very few reliable researches in the literature investigating the power of butterfly collection sets on hematological, coagulation and clinical chemistry testing. Basically, laboratory measurements appear to be unaffected by where the sample was withdrawn, whether there was a tourniquet, or the time between blood sampling and analysis [17, 18].

On the other hand, studies suggest that blood sampling from intravenous catheters, using butterfly collection sets and other similar collection sets often leads to hemolysis resulting in higher test cancellation than using a conservative straight needle [16, 19], [20], [21]. This has influenced the development of relevant guidelines and recommendations for collecting specimens from indwelling catheters or cannulas by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) [22, 23]. For example, sufficiently large shear stress and mechanical stress caused by the use of PVC, polyurethane and teflon catheters can affect membrane functions and integrity, causing changes in shape, cell activation, damage, and flow of intracellular components into the serum [20]. The potential alteration of cell integrity, especially during blood collection and hemolysis, can lead to unintended consequences requiring repeated blood collections, especially for hematological (PLT), enzymatic (AST, LDH, P) and electrolytic (K) assays [16]. Previous studies have found that blood sampling from peripheral catheters could be clinically permitted for hemoglobin and for most of the clinical chemistry analysis and clotting parameters [24], [25], [26], [27], [28]. Although PVC and other plastic materials are widely used for disposable medical devices, they can react negatively when they contact with body tissues and fluids (synovial, pleural fluids etc.). While the surface hydrophobicity of several artificial surfaces and the effect of blood cells on the tubing walls are possible causes of in vitro hemostatic activation, it may be considered that collecting the blood sample in butterfly collection sets may cause variations compared to blood collection directly to the vacuum tubes with straight needles. Although the risk of platelet activation and hemolysis increases in the long plastic tubes of butterfly collection sets, platelet counts, clotting parameters and potassium measurements are considered to be more vulnerable to changes caused by blood collection technique. A study performed by Sonntag et al. [29] declared that the concentrations of plasma K and LDH are considerably affected by hemoglobin concentrations 0.2 g/L. Another study has been conducted by Lippi and colleagues [30], in which they did not observe substantial differences in routine hematological and clinical chemistry testing for samples collected with the butterfly collection sets vs. straight needle. In the literature, published researches studying the association between BD Vacutainer® UltraTouch™ Push Button Blood Collection Set and the BD Vacutainer® Eclipse™ Blood Collection Needle are limited.

In our study, there are two different comparisons that can be relevant to clinical practice based on utilizing the new BD Vacutainer® UltraTouch™ Push Button Blood Collection Set. We made up clinical equivalence for the selected hematology, clinical chemistry and coagulation parameters in each tube collected using the BD Vacutainer® Eclipse Blood Collection Needle and BD Vacutainer® UltraTouch™ Push Button Collection Set. There was no significant difference in gauge comparison of the BD Vacutainer® UTBCs vs. current BD Vacutainer® Eclipse™ for hemolysis index. Furthermore, both blood collection sets have the acceptance limit of 20 hemolysis index units, which had been set for the mean and 95% confidence intervals. In addition, no significant difference in visual hemolysis or the hemolysis index was noted between the blood collection sets.

Activation of clotting cascade or clotting was assessed by measuring CBC (specifically platelet count/clumping) and by visual clot formation inspectations. All the CBC parameters demonstrated equivalence and no clots were detected in any BD Citrate or BD EDTA tubes collected with both device (Tables 1 and 2), no evidence of increased activation of clotting cascade or clotting was observed in anticoagulated specimens collected with both device.

The correlation between the two analytes is based on the intraclass correlation coefficient (ICC). In some publications, the correlation coefficient above 0.90 was interpreted as being very well [31], while others were found to be moderate compliance between 0.90 and 0.95 [32]. In the ICC analysis we conducted to examine the compatibility of the two blood collection sets in by dividing into two classes. One of them is visual and individual evaluation parameters. The other is coagulation parameters and K. In our study, visual hemolysis, instrument hemolysis index, visual lipemia, instrument lipemia index, instrument icterus index, LDH, AST, P, instrument platelet cluster, PLT, tube filling speed, Pain Asking Patient and Universal Pain Assessment tools are found the ICC=0.81, 0.90, 0.87, 0.98, 0.97, 0.94, 0.99, 0.99, 0.82, 0.99, 0.64, 0.55, 0.69, respectively. K is found to be ICC 0.87. Accordingly, instrument lipemia index, instrument hemolysis index, instrument icterus index, LDH, AST, P and PLT parameters compatibility between the two blood collection sets appears to be very well (ICC is above 0.90). However, for visual hemolysis, visual lipemia, K, instrument platelet cluster, tube filling speed, pain asking patient and universal pain assessment tool are found to be ICC values 0.81, 0.87, 0.87, 0.82, 0.64, 0,5 and 0.70, respectively. From these results, we supposed that compatibility between two blood collection sets are incompatible for these parameters according to ICC values (Table 3, Figure 1). However, although it has been reached to statistical significance, it should be pointed out that the overall discrepancy between the two blood collection devices was limited and mostly within the relative critical differences and current analytical quality specifications for desirable bias [33]. Based on these results, although statistically significant, the differences did not reach clinical relevance in any case for both blood collection devices. We suppose that these results could be attributed to the differences in technologies between the two blood collection sets or depend on the different lab technicians’s skills, working and knowledge. Phlebotomy equipment will keep advancing day by day; selecting the most appropriate blood collection system mainly depends on considerations of convenience, reliability, safety and cost [22].

In our study, the use of either sets for drawing blood, BD Vacutainer® UltraTouch™ Push Button Blood Collection Set or BD Vacutainer® Eclipse™ Blood Collection Needle did not affect the quality of the samples collected. So, using butterfly collection sets had little or no clinical effects. They may be a reliable alternative to the ordinary straight needle collection sets when indicated and within definite limitations for collecting samples for routine clinical chemistry laboratory testing.

The main limitation of the study is the number of participants (n=44). Another limitation is that blood was not drawn from the same arms for each healthy adult. From the literature, there are limited studies investigated the butterfly collection sets on hematological, coagulation and clinical chemistry testing. We also did not standardize the needle diameter (same gauge for both straight needle and butterfly collection sets). Pain scale measurements can change depending on the patients.

It is better to conduct further comparative studies. In addition, there may have been a bias in the selection of the subjects, as those who are severely needle phobic would not likely to agree about including in the study. In addition, the data obtained and the information in the study are limited to the questions posed concerning the 44 healthy adults in the Karapınar State Hospital and are not applicable to general population.


Corresponding author: Saadet Kader, Biochemistry Laboratory, Karapınar State Hospital, Konya, Turkey, E-mail:

Acknowledgments

We are grateful to laboratory technicians and nurses of Karapınar State Hospital who participated to this study.

  1. Research funding: During this study, any pharmaceutical company that is directly related with the research objective, no material and /or moral support has been received from a firm or any commercial institution that provides and /or manufactures medical instruments, equipment and materials, which may adversely affect the decision to be made during the evaluation of the study.

  2. Conflict of interest: The authors declare no conflict of interest and the authors have no competing interests.

References

1. Bonini, PA, Plebani, M, Ceriotti, F, Francesca Rubboli, F. Errors in laboratory medicine. Clin Chem 2002;48:691–8. https://doi.org/10.1093/clinchem/48.5.691.Suche in Google Scholar

2. Tolan, NV, Kaleta, EJ, Fang, JL, Colby, CE, Carey, WA, Karon, BS, et al.. Neonatal intensive care unit quality initiative: identifying preanalytical variables contributing to specimen hemolysis and measuring the impact of evidence-based practice interventions. Am J Clin Pathol 2016;146:113–8. https://doi.org/10.1093/ajcp/aqw086.Suche in Google Scholar PubMed

3. Bush, V, Mangan, L. The hemolyzed specimen: causes, effects, and reduction. BD Vacutainer Syst Preanalytical Solut 2003;2003:1–8.Suche in Google Scholar

4. Streichert, T, Otto, B, Schnabel, C, Nordholt, G, Haddad, M, Maric, M, et al.. Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems. Clin Chem 2011;57:1390–7. https://doi.org/10.1373/clinchem.2011.167932.Suche in Google Scholar PubMed

5. National Hospital Ambulatory Medical Care Survey: 2011 emergency department summary tables. Centers for Disease Control and Prevention Website; 2011. www.cdc.gov/nchs/data/ahcd/nhamcs_emergency/2011_ed_web_tables.pdf [Accessed 25 Apr 2016].Suche in Google Scholar

6. Phelan, MP, Reineks, EZ, Schold, JD, Kovach, A, Venkatesh, A. Estimated national volume of laboratory results affected by hemolyzed specimens from emergency departments. Arch Pathol Lab Med 2016;140:621. https://doi.org/10.5858/arpa.2015-0434-le.Suche in Google Scholar

7. Phelan, MP, Schold, J, Reineks, E, Meldon, S, Cornish, N, Procop, G. Projecting the potential impact of reductions in hemolysis among US emergency departments. Abstract and poster presented at: Annual Meeting of the American College of Medical Quality, Alexandria, Virginia, March 25–28, 2015.Suche in Google Scholar

8. Bowen, AR, Remaley, AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med 2014;24:31–44. https://doi.org/10.11613/bm.2014.006.Suche in Google Scholar PubMed PubMed Central

9. Young, DS. Conveying the importance of the preanalytical phase. Clin Chem Lab Med 2003;41:884–7. https://doi.org/10.1515/CCLM.2003.133.Suche in Google Scholar PubMed

10. Lippi, G, Cervellin, G, Favaloro, E, Plebani, M. Vitro and in vivo hemolysis. Berlin: Walter De Gruyter GmbH & Co; 2012:28 p.10.1515/9783110246148Suche in Google Scholar

11. Lima-Oliveira, D, Lippi, G, Salvagno, GL, Picheth, G, Guidi, GC. Laboratory diagnostics and quality of blood collection. J Med Biochem 2015;34:288–94. https://doi.org/10.2478/jomb-2014-0043.Suche in Google Scholar PubMed PubMed Central

12. Mensel, B, Wenzel, U, Roser, M, Ludemann, J, Nauck, M. Considerably reduced centrifugation time without increased hemolysis: evaluation of the new BD Vacutainer® SST™ II Advance. Clin Chem 2007;53:794–5https://doi.org/10.1373/clinchem.2006.079582.Suche in Google Scholar PubMed

13. Milutinovic, D, Andrijevic, I, Licina, M, Andrijevic, L. Confidence level in venipuncture and knowledge on causes of in vitro hemolysis among healthcare professionals. Biochem Med 2015;25:401–9.10.11613/BM.2015.040Suche in Google Scholar PubMed PubMed Central

14. Ohrbach, R, Fillingim, RB, Mulkey, F, Gonzalez, Y, Gordon, S, Gremillion, H. Clinical findings and pain symptoms as potential risk factors for chronic TMD: descriptive data and empirically identified domains from the OPPERA case-control study. J Pain 2011;12:27–45. https://doi.org/10.1016/j.jpain.2011.09.001.Suche in Google Scholar

15. https://www.westgard.com/clia-final-rule.htm.Suche in Google Scholar

16. Raisky, F, Gauthier, C, Marchal, A, Blum, D. Haemolyzed samples: responsibility of short catheters. Ann Biol Clin (Paris) 1994;52:523–7. https://doi.org/10.1002/jcu.1870220814.Suche in Google Scholar

17. Hill, AB, Nahrwold, ML, Noonan, D, Northrop, P. A comparison of methods of blood withdrawal and sample preparation for potassium measurements. Anesthesiology 1980;53:60–3. https://doi.org/10.1097/00000542-198007000-00012.Suche in Google Scholar

18. Lippi, G, Guidi, GC. Effect of specimen collection on routine coagulation assays and D-dimer measurement. Clin Chem 2004;50:2150–2. https://doi.org/10.1373/clinchem.2004.036244.Suche in Google Scholar

19. Kennedy, C, Angermuller, S, King, R, Noviello, S, WalkerJ, Warden, J, et al.. A comparison of hemolysis ratesusing intravenous catheters versus venepuncture tubes for obtaining blood samples. J Emerg Nurs 1996;22:566–9. https://doi.org/10.1016/s0099-1767(96)80213-3.Suche in Google Scholar

20. Sharp, MK, Mohammad, SF. Scaling of hemolysis in needles and catheters. Ann Biomed Eng 1998;26:788–97. https://doi.org/10.1114/1.65.Suche in Google Scholar PubMed

21. Grant, MS. The effect of blood drawing techniques and equipment on the hemolysis of ED laboratory blood samples. J Emerg Nurs 2003;29:116–21. https://doi.org/10.1067/men.2003.66.Suche in Google Scholar PubMed

22. Frey, AM. Drawing blood samples from vascular access devices: evidence-based practice. J Infus Nurs 2003;26:285–93. https://doi.org/10.1097/00129804-200309000-00004.Suche in Google Scholar PubMed

23. Burnett, RW, Covington, AK, Fogh-Andersen, N, Kulpman, WR, Maas, AH, Muller-Plathe, O, et al.. Recommendations on whole blood sampling, transport, and storage for simultaneous determination of pH, blood gases, and electrolytes. International Federation of Clinical Chemistry Scientific Division. J Int Fed Clin Chem 1994;6:115–20.Suche in Google Scholar

24. Mohler, M, Sato, Y, Bobick, K, Wise, LC. The reliability of blood sampling from peripheral intravenous infusion lines. Complete blood cell counts, electrolyte panels, and survey panels. J Intraven Nurs 1998;21:209–14.Suche in Google Scholar

25. Holmes, KR. Comparison of push-pull versus discard method from central venous catheters for blood testing. J Intraven Nurs 1998;21:282–5.Suche in Google Scholar

26. Seemann, S, Reinhardt, A. Blood sample collection from a peripheral catheter system compared with phlebotomy. J Intraven Nurs 2000;23:290–7.Suche in Google Scholar

27. Himberger, JR, Himberger, LC. Accuracy of drawing blood through infusing intravenous lines. Heart Lung 2001;30:66–73. https://doi.org/10.1067/mhl.2001.110535.Suche in Google Scholar PubMed

28. Zlotowski, SJ, Kupas, DF, Wood, GC. Comparison of laboratory values obtained by means of routine venepuncture versus peripheral intravenous catheter after a normal saline solution bolus. Ann Emerg Med 2001;38:497–504. https://doi.org/10.1067/mem.2001.118015.Suche in Google Scholar PubMed

29. Sonntag, G. Hemolysis as an interference factor in clinical chemistry. J Clin Chem Clin Biochem 1986;24:127–39. https://doi.org/10.1128/jcm.24.3.506-506.1986.Suche in Google Scholar

30. Lippi, G, Salvagno, G, Brocco, G, Guidi, G. Preanalytical variability in laboratory testing: influence of the blood drawing technique. Clin Chem Lab Med 2005;43:319–25. https://doi.org/10.1515/CCLM.2005.055.Suche in Google Scholar PubMed

31. Fleiss, JL. The design and analysis of clinical experiments. New York: John Wiley and Sons; 1986: 1–32 pp.Suche in Google Scholar

32. Lawrence, I, Lin, K. A concordance correlation coefficient to evaluate reproducibility. Biometrics 1989;45:255–68.10.2307/2532051Suche in Google Scholar

33. Ricos, C, Alvarez, V, Cava, F, Garcia-Lario, JV, Hernandez, A, Jimenez, CV, et al.. Current databases on biologic variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491–500. https://doi.org/10.1080/00365519950185229.Suche in Google Scholar PubMed


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/tjb-2020-0626).


Received: 2020-05-29
Accepted: 2021-06-22
Published Online: 2021-08-23

© 2021 Saadet Kader et al., published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Artikel in diesem Heft

  1. Frontmatter
  2. Review Article
  3. Overview of COVID-19’s relationship with thrombophilia proteins
  4. Research Articles
  5. Is it possible to determine antibiotic resistance of E. coli by analyzing laboratory data with machine learning?
  6. Detection of circulating prostate cancer cells via prostate specific membrane antigen by chronoimpedimetric aptasensor
  7. Two approaches for measurement uncertainty estimation: which role for bias? Complete blood count experience
  8. Is there any relationship between C-reactive protein/albumin ratio and clinical severity of childhood community-acquired pneumonia
  9. Comparison of nitric oxide and adrenomedullin levels of children with attention deficit hyperactivity disorder and anxiety disorder
  10. Performance evaluation of internal quality control rules, EWMA, CUSUM, and the novel machine learning model
  11. Are serum molecular markers more effective than the invasive methods used in the diagnosis of breast cancers?
  12. HIF-1 inhibitors: differential effects of Acriflavine and Echinomycin on tumor associated CA-IX enzyme and VEGF in melanoma
  13. Evaluation of BD Vacutainer Eclipse and BD Vacutainer Ultra-Touch butterfly blood collecting sets in laboratory testing
  14. The effects of various strength training intensities on blood cardiovascular risk markers in healthy men
  15. Is there any relationship between LGALS3 gene variations and histopathological criteria in laryngeal squamous cell carcinoma (LSCC)?
  16. Expression levels of BAP1, OGT, and YY1 genes in patients with eyelid tumors
  17. Association of bitter and sweet taste gene receptor polymorphisms with dental caries formation
  18. Case Report
  19. Glucose-6-phosphate dehydrogenase gene Ala365Thr mutation in an Iraqi family with confusing clinical differences
  20. Acknowledgment
  21. Acknowledgment
Heruntergeladen am 9.11.2025 von https://www.degruyterbrill.com/document/doi/10.1515/tjb-2020-0626/html?lang=de
Button zum nach oben scrollen