Objectives:

We investigated the accuracy of left-to-right shunt detection using transpulmonary ultrasound dilution (TPUD) and compared the agreement between pulmonary over systemic blood flow (Qp/Qs) ratio measured by TPUD [Qp/Qs(tpud)] and ultrasonic flow probes [Qp/Qs(ufp)].

Methods:

Seven newborn lambs under general anesthesia were connected to the TPUD monitor (COstatus™) after insertion of arterial and central venous catheters. A Gore-Tex® shunt, inserted between the descending aorta and left pulmonary artery, was intermittently opened and closed while cardiac output was varied by blood withdrawals. Flow probes were placed around the main pulmonary artery (Qufp) and the descending aorta proximal (Qpre) and distal (Qpost) to the shunt insertion. Qp/Qs(ufp) was calculated as (Qufp+Qpre–Qpost)/Qufp.

Results:

Seventy-two paired measurement sessions were analyzed. Shunts were detected by TPUD with a positive predictive value of 86%, a negative predictive value of 100%, a sensitivity of 100% and a specificity of 83%. The Bland-Altman analysis comparing Qp/Qs(tpud) and Qp/Qs(ufp) showed an overall mean bias (SD) of 0.1 (0.3), limits of agreement (LOA) of ±0.6 and a percentage error of 34.8%.

Conclusions:

The qualitative diagnostic accuracy of TPUD for shunt detection is high. Modification of the algorithm seems required as shunt quantification by TPUD is accurate, but not yet very precise.

Acknowledgments

We would like to thank J.J.M. Menssen, Laboratory of Clinical Physics of the Department of Pediatrics, Radboud University Nijmegen Medical Centre; A.E.J. Hanssen, Animal Laboratory of the Radboud University Nijmegen and A.F.J. de Haan, MSc, Department of Epidemiology, Statistics and Health Technology Assessment of the Radboud University Nijmegen for their outstanding support. Our research group received financial support for the technical realization of this experiment from Transonic Systems Inc., Ithaca, NY, USA and Pulsion Medical Systems, Munich, Germany.

References

[1] Project IotV-OTND. The vermont-oxford trials network: very low birth weight outcomes for 1990. Pediatrics. 1993;91:540–5.10.1542/peds.91.3.540Suche in Google Scholar

[2] Reller MD, Rice MJ, McDonald RW. Review of studies evaluating ductal patency in the premature infant. J Pediatr. 1993;122:S59–62.10.1016/S0022-3476(09)90044-0Suche in Google Scholar

[3] Ichihashi K, Shiraishi H, Endou H, Kuramatsu T, Yano S, Yanagisawa M. Cerebral and abdominal arterial hemodynamics in preterm infants with patent ductus arteriosus. Acta Paediatr Jpn. 1990;32:349–56.10.1111/j.1442-200X.1990.tb00842.xSuche in Google Scholar PubMed

[4] Kluckow M, Evans N. Ductal shunting, high pulmonary blood flow, and pulmonary hemorrhage. J Pediatr. 2000;137:68–72.10.1067/mpd.2000.106569Suche in Google Scholar PubMed

[5] Hamrick SE, Hansmann G. Patent ductus arteriosus of the preterm infant. Pediatrics. 2010;125:1020–30.10.1542/peds.2009-3506Suche in Google Scholar PubMed

[6] Evans N, Kluckow M. Early ductal shunting and intraventricular haemorrhage in ventilated preterm infants. Arch Dis Child Fetal Neonatal Ed. 1996;75:F183–6.10.1136/fn.75.3.F183Suche in Google Scholar

[7] Evans N. Diagnosis of the preterm patent ductus arteriosus: clinical signs, biomarkers, or ultrasound? Semin Perinatol. 2012;36:114–22.10.1053/j.semperi.2011.09.021Suche in Google Scholar PubMed

[8] Cambonie G, Dupuy AM, Combes C, Vincenti M, Mesnage R, Cristol JP. Can a clinical decision rule help ductus arteriosus management in preterm neonates? Acta Paediatr. 2012;101:e213–8.10.1111/j.1651-2227.2011.02469.xSuche in Google Scholar PubMed

[9] Kwinta P, Rudzinski A, Kruczek P, Kordon Z, Pietrzyk JJ. Can early echocardiographic findings predict patent ductus arteriosus? Neonatology. 2009;95:141–8.10.1159/000153098Suche in Google Scholar PubMed

[10] Sehgal A, McNamara PJ. Does echocardiography facilitate determination of hemodynamic significance attributable to the ductus arteriosus? Eur J Pediatr. 2009;168:907–14.10.1007/s00431-009-0983-3Suche in Google Scholar PubMed

[11] Zonnenberg I, de Waal K. The definition of a haemodynamic significant duct in randomized controlled trials: a systematic literature review. Acta Paediatr. 2012;101:247–51.10.1111/j.1651-2227.2011.02468.xSuche in Google Scholar PubMed

[12] Jain A, Shah PS. Diagnosis, evaluation, and management of patent ductus arteriosus in preterm neonates. JAMA Pediatr. 2015;169:863–72.10.1001/jamapediatrics.2015.0987Suche in Google Scholar PubMed

[13] Kluckow M, Seri I, Evans N. Functional echocardiography: an emerging clinical tool for the neonatologist. J Pediatr. 2007;150:125–30.10.1016/j.jpeds.2006.10.056Suche in Google Scholar PubMed

[14] Kluckow M, Evans N. Low superior vena cava flow and intraventricular haemorrhage in preterm infants. Arch Dis Child Fetal Neonatal Ed. 2000;82:F188–94.10.1136/fn.82.3.F188Suche in Google Scholar PubMed PubMed Central

[15] Krivitski NM, Kislukhin VV, Thuramalla NV. Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients. Pediatr Crit Care Med. 2008;9:423–8.10.1097/01.PCC.0b013e31816c71bcSuche in Google Scholar PubMed PubMed Central

[16] Eremenko AA, Safarov PN. Flow-regulated extracorporeal arteriovenous tubing loop for cardiac output measurements by ultrasound velocity dilution: validation in post-cardiac surgery intensive care unit patients. ASAIO J. 2010;56:522–6.10.1097/MAT.0b013e3181effdf8Suche in Google Scholar PubMed

[17] Tsutsui M, Matsuoka N, Ikeda T, Sanjo Y, Kazama T. Comparison of a new cardiac output ultrasound dilution method with thermodilution technique in adult patients under general anesthesia. J Cardiothorac Vasc Anesth. 2009;23:835–40.10.1053/j.jvca.2009.03.007Suche in Google Scholar PubMed

[18] Gleed RD, Callahan M, Smith T. Validation of novel ultrasound dilution cardiac output method for pediatric and neonatal patients. Eur Soc Intensive Care Med. 2006;32:S172.Suche in Google Scholar

[19] Crittendon I, 3rd, Dreyer WJ, Decker JA, Kim JJ: Ultrasound dilution: an accurate means of determining cardiac output in children. Pediatr Crit Care Med. 2012;13:42–6.10.1097/PCC.0b013e3182196804Suche in Google Scholar PubMed PubMed Central

[20] Boehne M, Baustert M, Paetzel V, Koditz H, Schoof S, Happel CM, et al. Determination of cardiac output by ultrasound dilution technique in infants and children: a validation study against direct fick principle. Br J Anaesth. 2014;112:469–76.10.1093/bja/aet382Suche in Google Scholar PubMed

[21] de Boode WP, van Heijst AF, Hopman JC, Tanke RB, van der Hoeven HG, Liem KD. Cardiac measurement using an ultrasound dilution method: a validation study in ventilated piglets. Pediatr Crit Care Med. 2010;11:103–8.10.1097/PCC.0b013e3181b064eaSuche in Google Scholar PubMed

[22] Vrancken SL, de Boode WP, Hopman JC, Singh SK, Liem KD, van Heijst AF. Cardiac output measurement with transpulmonary ultrasound dilution is feasible in the presence of a left-to-right shunt: a validation study in lambs. Br J Anaesth. 2012;108:409–16.10.1093/bja/aer401Suche in Google Scholar PubMed

[23] Vrancken SL, de Boode WP, Hopman JC, Looijen-Salamon MG, Liem KD, van Heijst AF. Influence of lung injury on cardiac output measurement using transpulmonary ultrasound dilution: a validation study in neonatal lambs. Br J Anaesth. 2012;109:870–8.10.1093/bja/aes297Suche in Google Scholar PubMed

[24] Lindberg L, Johansson S, Perez-de-Sa V. Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children. Pediatr Crit Care Med. 2014;15:139–47.10.1097/PCC.0000000000000053Suche in Google Scholar PubMed

[25] Saxena R, Krivitski N, Peacock K, Durward A, Simpson JM, Tibby SM. Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts. J Clin Monit Comput. 2015;29:407–14.10.1007/s10877-014-9618-ySuche in Google Scholar PubMed

[26] Mook GA, Zijlstra WG. Quantitative evaluation of intracardiac shunts from arterial dye dilution curves. Demonstration of very small shunts. Acta Med Scand. 1961;170:703–15.10.1111/j.0954-6820.1961.tb00289.xSuche in Google Scholar PubMed

[27] Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.10.1080/10543400701329422Suche in Google Scholar PubMed

[28] Carter SA, Bajec DF, Yannicelli E, Wood EH. Estimation of left-to-right shunt from arterial dilution curves. J Lab Clin Med. 1960;55:77–88.Suche in Google Scholar

[29] Morrow AG, Oldham HN, Callard GM, Braunwal E. Assessment of operative results in congenital heart disease by intraoperative indicator-dilution curves. Circulation. 1966;33:263–9.10.1161/01.CIR.33.2.263Suche in Google Scholar PubMed

[30] El Hajjar M, Vaksmann G, Rakza T, Kongolo G, Storme L. Severity of the ductal shunt: a comparison of different markers. Arch Dis Child Fetal Neonatal Ed. 2005;90:F419–22.10.1136/adc.2003.027698Suche in Google Scholar PubMed PubMed Central

[31] Shih A, Maisenbacher III H, Vigani A, Estrada A, Pogue B, Berry C, et al. Ultrasound dilution technique as a minimally invasive way to detect intracardiac shunt. Crit Care Med. 2011;39:61. [abstract 0683].Suche in Google Scholar

[32] Boehne M, Baustert M, Happel CM, Schoff S, Hohmann D, Paetzel V, et al. Novel ultrasound dilution technique detects left-to-right shunts with high accuracy in children. Pediatr Res. 2010;68:117. [abstract 224].10.1203/00006450-201011001-00224Suche in Google Scholar

[33] Boehne M, Schmidt F, Witt L, Koditz H, Sasse M, Sumpelmann R, et al. Comparison of transpulmonary thermodilution and ultrasound dilution technique: movel insights into volumetric parameters from an animal model. Pediatr Cardiol. 2012;33:625–32.10.1007/s00246-012-0192-0Suche in Google Scholar PubMed

  1. The authors stated that there are no conflicts of interest regarding the publication of this article.

Received: 2015-9-6
Accepted: 2016-3-31
Published Online: 2016-5-9
Published in Print: 2016-11-1