Status of serum-calcium and -albumin measurement in Argentina assessed in 300 representative laboratories with 20 fresh frozen single donation sera

  • Hedwig C.M. Stepman , Dietmar Stöckl , Rosana Acheme , Sandra Sesini , Daniel Mazziotta and Linda M. Thienpont
Published/Copyright: August 30, 2011

Abstract

Background: The Fundación Bioquímica Argentina (FBA) performs external quality assessment (EQA) of >3200 laboratories. However, FBA realizes that sample non-commutability and predominant use of heterogeneous systems may bias the estimated performance and standardization status. To eliminate these confounding factors, a study using frozen single donation sera was undertaken with the focus on serum-calcium and -albumin measurement.

Methods: Target values were established from the results produced with homogeneous systems. In groups of n=7, system effects were investigated. Laboratory performance was evaluated from the correlation coefficient r between the measurement results for all sera and the target values. This allowed ranking of the laboratories and judgment of the deviation for individual samples (total error) against a 10% limit. The total error specification was a deviation for ≥5 samples exceeding 10% and/or causing a result outside the laboratory's reference interval.

Results: For calcium (n=303) (range: 2.06–2.42 mmol/L), 81 laboratories had an r-value <0.6, 43 even <0.4; the total error was relevant for 97 (10% limit) and 111 (reference interval) laboratories. For albumin (n=311) (range: 34.7–45.7 g/L) r was <0.7 (<0.4) in 44 (16) laboratories; 83 and 36 laboratories exceeded the total error criteria. Laboratories using homogeneous systems were generally ranked higher by correlation. System effects were moderate for calcium, but significant for albumin.

Conclusions: The study demonstrated the need to improve the quality and harmonization of calcium and albumin testing in the investigated laboratories. To achieve this objective, we promote co-operation between laboratories, EQA provider and manufacturers.


Corresponding author: Linda M. Thienpont, Laboratory for Analytical Chemistry, Faculty of Pharmaceutical Sciences, Ghent University, Harelbekestraat 72, 9000 Ghent, Belgium Phone: +32 9 264 81 04, Fax: +32 9 264 81 98,

Received: 2011-3-1
Accepted: 2011-7-24
Published Online: 2011-08-30
Published in Print: 2011-11-01

©2011 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Editorial
  2. Thank you, indeed!
  3. Reviews
  4. Glomerular filtration rate estimation: performance of serum cystatin C-based prediction equations
  5. Oxidative stress and diabetes mellitus
  6. Human papilloma virus (HPV) molecular diagnostics
  7. Opinion Papers
  8. Urine immunofixation electrophoresis remains important and is complementary to serum free light chain
  9. Permissible performance limits of regression analyses in method comparisons
  10. General Clinical Chemistry and Laboratory Medicine
  11. The value of serum ischemia-modified albumin for assessing liver function in patients with chronic liver disease
  12. Relationship between bilirubin and C-reactive protein
  13. Status of serum-calcium and -albumin measurement in Argentina assessed in 300 representative laboratories with 20 fresh frozen single donation sera
  14. Practical application of Sigma Metrics QC procedures in clinical chemistry
  15. High-throughput analysis of testosterone in serum samples by on-line solid phase extraction liquid chromatography-tandem mass spectrometry
  16. Monitoring of mycophenolic acid and kidney function during combined immunosuppressive therapy
  17. Phenotype of Gc-globulin influences the macrophage activating factor (MAF) levels in serum
  18. Combination of cellular population data and CytoDiff™ analyses for the diagnosis of lymphocytosis
  19. Eosinophiluria is common among patients after ileal conduit surgery
  20. Soluble urokinase plasminogen activator receptor (suPAR) levels in healthy pregnancy and preeclampsia
  21. Uptake of cobalamin and markers of cobalamin status: a longitudinal study of healthy pregnant women
  22. Vitamin B12 absorption judged by measurement of holotranscobalamin, active vitamin B12: evaluation of a commercially available EIA kit
  23. Salivary markers of oxidative stress and Salivette collection systems
  24. Infectious Diseases
  25. Development of a microwell adapted immunoblot system with recombinant antigens for distinguishing human herpesvirus (HHV)6A and HHV6B and detection of human cytomegalovirus
  26. Cardiovascular Disease
  27. High-sensitive troponin T in chronic heart failure correlates with severity of symptoms, left ventricular dysfunction and prognosis independently from N-terminal pro-b-type natriuretic peptide
  28. Plasma symmetric dimethylarginine reference limits from the Framingham offspring cohort
  29. Letters to the Editor
  30. Iron deficiency prevalence is increased among Helicobacter pylori infected Brazilian patients
  31. Helicobacter pylori infection and iron deficiency in a group of Brazilian patients
  32. Multiplex protein assay performance/evaluation and the requirement for precision and correlation to clinical assays
  33. Functional assay or antigen test for protein C and protein S in ischemic stroke: which shows the greatest change?
Downloaded on 29.10.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm.2011.681/html
Scroll to top button