Commutability assessment of candidate reference materials for plasma renin activity measurement: current challenges
-
Zhenni Liu
, Lizi Jin
, Zijia Ma
and Chuanbao Zhang
Abstract
Objectives
This study aims to evaluate the commutability of external quality assessment (EQA) materials and candidate reference materials (RMs) for plasma renin activity (PRA) assay.
Methods
Commutabilities of 16 candidate RMs were measured along with 40 clinical samples by the four different routine PRA assays, including three LC‒MS/MS assays and one chemiluminescence immunoassay. Sixteen candidate RMs included native/spiked human plasma pools (small-scale pools with <50 individuals) and current EQA materials (large-scale pools with >1,000 individuals). Difference in bias approach and linear regression with prediction interval approach were adopted to determine the commutability. Two-way variance analysis was used to estimate the effects of spiked and pool size on the commutability. Stability and homogeneity studies were performed.
Results
Precision and correlation performance of all assays was acceptable. In the difference in bias approach, the commutability results were not satisfactory (noncommutability: 14/16) and significant sample-specific effects were detected in assay pairs using different incubation buffers. For the prediction interval approach, no commutability was observed in the spiked small-scale pools; EQA materials (4/9) had more satisfactory commutability among all assays than the small-scale pools (2/7); RMs of large-scale pools tend to have better commutability no matter spiked or not.
Conclusions
Commutable RMs were obtainable but challenging. Current EQA materials with relatively good commutability, stability, and homogeneity were appropriate RMs. Large-scale pools are tending to be commutable. Spiking in small-scale pools was not suggested to prepare RMs. MPs adopting a uniform incubation buffer would be preferable for further commutability research.
Funding source: National High Level Hospital Clinical Research Funding
Award Identifier / Grant number: BJ-2023-094
Funding source: Beijing Natural Science Foundation
Award Identifier / Grant number: 7212087
Acknowledgments
We would like to thank our colleagues in Beijing Hospital for their kind help in sample collection. We sincerely thank the Zhejiang Disigns Diagnostics Technology Co., Ltd., Shanghai Biotree Biomedical Technology Co., Ltd, (Shanghai, China), and Shenzhen New Industries Biomedical Engineering Co., Ltd, (Shenzhen, China) for the sample testing and technical support.
-
Research ethics: The collection of plasma was approved by the Ethics Committee of Beijing Hospital.
-
Informed consent: This study was approved an exemption from informed consent.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: The authors state no conflict of interest.
-
Research funding: This work was supported by the Beijing Natural Science Foundation (grant no. 7212087) and the National High Level Hospital Clinical Research Funding (grant no. BJ-2023-094).
References
1. Funder, JW, Carey, RM, Mantero, F, Murad, MH, Reincke, M, Shibata, H, et al.. The management of primary aldosteronism: case detection, diagnosis, and treatment: an endocrine society clinical practice guideline. J Clin Endocrinol Metab 2016;101:1889–916. https://doi.org/10.1210/jc.2015-4061.Search in Google Scholar PubMed
2. Chappell, DL, McAvoy, T, Weiss, B, Weiner, R, Laterza, OF. Development and validation of an ultra-sensitive method for the measurement of plasma renin activity in human plasma via LC-MS/MS. Bioanalysis 2012;4:2843–50. https://doi.org/10.4155/bio.12.268.Search in Google Scholar PubMed
3. Chen, F, Cheng, Z, Peng, Y, Wang, Z, Huang, C, Liu, D, et al.. A liquid chromatography-tandem mass spectrometry (LC-MS/MS)-based assay for simultaneous quantification of aldosterone, renin activity, and angiotensin II in human plasma. J Chromatogr B Anal Technol Biomed Life Sci 2021;1179:122740. https://doi.org/10.1016/j.jchromb.2021.122740.Search in Google Scholar PubMed
4. Morganti, A. A comparative study on inter and intralaboratory reproducibility of renin measurement with a conventional enzymatic method and a new chemiluminescent assay of immunoreactive renin. J Hypertens 2010;28:1307–12. https://doi.org/10.1097/hjh.0b013e32833857ad.Search in Google Scholar
5. Carter, S, Owen, LJ, Kerstens, MN, Dullaart, RP, Keevil, BG. A liquid chromatography tandem mass spectrometry assay for plasma renin activity using online solid-phase extraction. Ann Clin Biochem 2012;49:570–9. https://doi.org/10.1258/acb.2012.011186.Search in Google Scholar PubMed
6. Liu, Z, Jin, L, Zeng, J, Zhang, T, Zhang, J, Zhou, W, et al.. Poor comparability of plasma renin activity measurement in determining patient samples: the status quo and recommendations for harmonization. Clin Chem Lab Med 2023;61:1170–9. https://doi.org/10.1515/cclm-2023-0117.Search in Google Scholar PubMed
7. Plebani, M, Lippi, G. Standardization and harmonization in laboratory medicine: not only for clinical chemistry measurands. Clin Chem Lab Med 2023;61:185–7. https://doi.org/10.1515/cclm-2022-1122.Search in Google Scholar PubMed
8. Liu, Z, Jin, L, Zhang, J, Zhang, T, Zeng, J, Zhou, W, et al.. Development of a designed comparison method based on isotope dilution liquid chromatography-tandem mass spectrometry for determining plasma renin activity and its clinical assessment of renin activity stability in plasma. Anal Methods 2023;15:492–501. https://doi.org/10.1039/d2ay01646j.Search in Google Scholar PubMed
9. Available High-Quality Standard Materials SRM 998 Angiotensin I: The National Institute of Standards and Technology; 2022. Available from: https://www.nist.gov/mml/biomolecular-measurement/reference-materials.Search in Google Scholar
10. JCGM. International vocabulary of metrology— basic and general concepts and associated terms (VIM), 3rd ed. Sevres, France: International Bureau of Weights and Measures; 2012.Search in Google Scholar
11. Braga, F, Panteghini, M. Commutability of reference and control materials: an essential factor for assuring the quality of measurements in Laboratory Medicine. Clin Chem Lab Med 2019;57:967–73. https://doi.org/10.1515/cclm-2019-0154.Search in Google Scholar PubMed
12. Badrick, T, Stavelin, A. Harmonising EQA schemes the next frontier: challenging the status quo. Clin Chem Lab Med 2020;58:1795–7. https://doi.org/10.1515/cclm-2020-0343.Search in Google Scholar PubMed
13. Badrick, T, Miller, WG, Panteghini, M, Delatour, V, Berghall, H, MacKenzie, F, et al.. Interpreting EQA-understanding why commutability of materials matters. Clin Chem 2022;68:494–500. https://doi.org/10.1093/clinchem/hvac002.Search in Google Scholar PubMed
14. Miller, WG, Myers, G, Cobbaert, CM, Young, IS, Theodorsson, E, Wielgosz, RI, et al.. Overcoming challenges regarding reference materials and regulations that influence global standardization of medical laboratory testing results. Clin Chem Lab Med 2023;61:48–54. https://doi.org/10.1515/cclm-2022-0943.Search in Google Scholar PubMed
15. Danilenko, U, Vesper, HW, Myers, GL, Clapshaw, PA, Camara, JE, Miller, WG. An updated protocol based on CLSI document C37 for preparation of off-the-clot serum from individual units for use alone or to prepare commutable pooled serum reference materials. Clin Chem Lab Med 2020;58:368–74. https://doi.org/10.1515/cclm-2019-0732.Search in Google Scholar PubMed PubMed Central
16. Deprez, L, Toussaint, B, Zegers, I, Schimmel, H, Grote-Koska, D, Klauke, R, et al.. Commutability assessment of candidate reference materials for pancreatic α-amylase. Clin Chem 2018;64:1193–202. https://doi.org/10.1373/clinchem.2018.289744.Search in Google Scholar PubMed
17. Long, Q, Qi, T, Zhang, T, Wang, J, Zeng, J, Yan, Y, et al.. Commutability assessment of candidate external quality assessment materials for aminotransferase activity measurements based on different approaches in China. Ann Lab Med 2021;41:68–76. https://doi.org/10.3343/alm.2021.41.1.68.Search in Google Scholar PubMed PubMed Central
18. CLSI. EP15-A3: user verification of precision and estimation of bias; approved guideline, 3rd ed. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.Search in Google Scholar
19. Miller, WG, Schimmel, H, Rej, R, Greenberg, N, Ceriotti, F, Burns, C, et al.. IFCC working group recommendations for assessing commutability part 1: general experimental design. Clin Chem 2018;64:447–54. https://doi.org/10.1373/clinchem.2017.277525.Search in Google Scholar PubMed PubMed Central
20. Nilsson, G, Budd, JR, Greenberg, N, Delatour, V, Rej, R, Panteghini, M, et al.. IFCC working group recommendations for assessing commutability Part 2: using the difference in bias between a reference material and clinical samples. Clin Chem 2018;64:455–64. https://doi.org/10.1373/clinchem.2017.277541.Search in Google Scholar PubMed PubMed Central
21. CLSI. EP30-A: characterization and qualification of commutable reference materials for laboratory medicine; approved guideline. Wayne, PA: Clinical and Laboratory Standards Institute; 2010.Search in Google Scholar
22. Andreasson, U, Gobom, J, Delatour, V, Auclair, G, Noam, Y, Lee, S, et al.. Assessing the commutability of candidate reference materials for the harmonization of neurofilament light measurements in blood. Clin Chem Lab Med 2023;61:1245–54. https://doi.org/10.1515/cclm-2022-1181.Search in Google Scholar PubMed
23. CNAS. GL03: guidance on evaluating the homogeneity and stability of samples used for proficiency testing. Beijing, China: China National Accreditation Service for Conformity Assessment; 2006.Search in Google Scholar
24. Sealey, JE, Laragh, JH. Radioimmunoassay of plasma renin activity. Semin Nucl Med 1975;5:189–202. https://doi.org/10.1016/s0001-2998(75)80033-x.Search in Google Scholar PubMed
Supplementary Material
This article contains supplementary material (https://doi.org/10.1515/cclm-2023-0698).
© 2023 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorial
- Biological management of diabetes mellitus, the laboratory medicine specialist and the patient
- Reviews
- Remote HbA1c testing via microsampling: fit for purpose?
- The effect of hormonal contraceptive therapy on clinical laboratory parameters: a literature review
- Opinion Paper
- Continuous glucose monitoring has an increasing role in pre-symptomatic type 1 diabetes: advantages, limitations, and comparisons with laboratory-based testing
- IFCC Paper
- Comparison and commutability study among four faecal immunochemical tests (FIT) systems
- Guidelines and Recommendations
- Evidence-based procedures to improve the reliability of circulating miRNA biomarker assays
- General Clinical Chemistry and Laboratory Medicine
- Commutability assessment of candidate reference materials for plasma renin activity measurement: current challenges
- Aggregated data from the same laboratories participating in two glucose external quality assessment schemes show that commutability and transfers of values to control materials are decisive for the biases found
- Methodological considerations in determining sex steroids in children: comparison of conventional immunoassays with liquid chromatography-tandem mass spectrometry
- From “wet” matrices to “dry” blood spot sampling strategy: a versatile LC-MS/MS assay for simultaneous monitoring caffeine and its three primary metabolites in preterm infants
- A sensitive LC-MS/MS methotrexate assay capable of assessing adherence to methotrexate therapy in rheumatoid arthritis
- Determination of cortisone and cortisol in human scalp hair using an improved LC-MS/MS-based method
- Mild hypophosphatasia may be twice as prevalent as previously estimated: an effective clinical algorithm to detect undiagnosed cases
- Combined deficient response to polysaccharide-based and protein-based vaccines predicts a severe clinical phenotype
- Reference Values and Biological Variations
- A sex-specific association of leukocyte telomere length with thigh muscle mass
- Cancer Diagnostics
- Optimized procedure for high-throughput transcriptome profiling of small extracellular vesicles isolated from low volume serum samples
- Ultra-short cell-free DNA fragments enhance cancer early detection in a multi-analyte blood test combining mutation, protein and fragmentomics
- Clinical relevance of clonal hematopoiesis and its interference in cell-free DNA profiling of patients with gastric cancer
- Development and analytical validation of a novel nasopharynx swab-based Epstein-Barr virus C promoter methylation quantitative assay for nasopharyngeal carcinoma detection
- Infectious Diseases
- (Pre)analytical considerations concerning the analysis of synovial calprotectin
- Corrigendum
- Thermal and chronological stability of monocyte distribution width (MDW), the new biomarker for sepsis
- Acknowledgment
- Acknowledgment
- Letters to the Editor
- In-house diagnostic devices under the EU IVDR and unwanted side-effects of intentional transparency
- Use of thyroid function tests in urine: a position statement of the Belgian Thyroid Club
- State of the art of measurement uncertainty of serum ferritin
- Analytical performance evaluation and consumable waste reduction strategies using a tube-based quality control material
- Stability of fecal calprotectin extracts using the Diasorin® kit
- Reverse cascade testing from newborn screening: the opportunity to improve family healthcare outcomes
- An indirect data-mining approach to standardise paediatric serum phosphate reference intervals in Wales
- Sodium and risk assessment of osmotic demyelination syndrome: the method matters!
Articles in the same Issue
- Frontmatter
- Editorial
- Biological management of diabetes mellitus, the laboratory medicine specialist and the patient
- Reviews
- Remote HbA1c testing via microsampling: fit for purpose?
- The effect of hormonal contraceptive therapy on clinical laboratory parameters: a literature review
- Opinion Paper
- Continuous glucose monitoring has an increasing role in pre-symptomatic type 1 diabetes: advantages, limitations, and comparisons with laboratory-based testing
- IFCC Paper
- Comparison and commutability study among four faecal immunochemical tests (FIT) systems
- Guidelines and Recommendations
- Evidence-based procedures to improve the reliability of circulating miRNA biomarker assays
- General Clinical Chemistry and Laboratory Medicine
- Commutability assessment of candidate reference materials for plasma renin activity measurement: current challenges
- Aggregated data from the same laboratories participating in two glucose external quality assessment schemes show that commutability and transfers of values to control materials are decisive for the biases found
- Methodological considerations in determining sex steroids in children: comparison of conventional immunoassays with liquid chromatography-tandem mass spectrometry
- From “wet” matrices to “dry” blood spot sampling strategy: a versatile LC-MS/MS assay for simultaneous monitoring caffeine and its three primary metabolites in preterm infants
- A sensitive LC-MS/MS methotrexate assay capable of assessing adherence to methotrexate therapy in rheumatoid arthritis
- Determination of cortisone and cortisol in human scalp hair using an improved LC-MS/MS-based method
- Mild hypophosphatasia may be twice as prevalent as previously estimated: an effective clinical algorithm to detect undiagnosed cases
- Combined deficient response to polysaccharide-based and protein-based vaccines predicts a severe clinical phenotype
- Reference Values and Biological Variations
- A sex-specific association of leukocyte telomere length with thigh muscle mass
- Cancer Diagnostics
- Optimized procedure for high-throughput transcriptome profiling of small extracellular vesicles isolated from low volume serum samples
- Ultra-short cell-free DNA fragments enhance cancer early detection in a multi-analyte blood test combining mutation, protein and fragmentomics
- Clinical relevance of clonal hematopoiesis and its interference in cell-free DNA profiling of patients with gastric cancer
- Development and analytical validation of a novel nasopharynx swab-based Epstein-Barr virus C promoter methylation quantitative assay for nasopharyngeal carcinoma detection
- Infectious Diseases
- (Pre)analytical considerations concerning the analysis of synovial calprotectin
- Corrigendum
- Thermal and chronological stability of monocyte distribution width (MDW), the new biomarker for sepsis
- Acknowledgment
- Acknowledgment
- Letters to the Editor
- In-house diagnostic devices under the EU IVDR and unwanted side-effects of intentional transparency
- Use of thyroid function tests in urine: a position statement of the Belgian Thyroid Club
- State of the art of measurement uncertainty of serum ferritin
- Analytical performance evaluation and consumable waste reduction strategies using a tube-based quality control material
- Stability of fecal calprotectin extracts using the Diasorin® kit
- Reverse cascade testing from newborn screening: the opportunity to improve family healthcare outcomes
- An indirect data-mining approach to standardise paediatric serum phosphate reference intervals in Wales
- Sodium and risk assessment of osmotic demyelination syndrome: the method matters!