At present time, immunoassay systems, using fully automated platform, are the methods of choice in the clinical laboratory, especially for the measurement of particularly complex heterogeneous molecules, which are present in biological fluids in concentration in the range of ng/L or less [1], [2], [3]. It is not surprising, as immunoassays involve the reaction of complex biological reagents (like antibodies) with other complex biological reagents (like some peptides and proteins) in a variable biological matrix, that they are inherently vulnerable to different types of interference. Unfortunately, the frequency with which such interference may occur – and most importantly, the proportion of consequently erroneous results that may significantly and adversely affect clinical management – is rather difficult to assess [2], [3].
It is well known for a long time that drugs can often affect the laboratory test results by interfering with the analytical systems themselves, or by influencing endogenous constituents [4]. Indeed, clinicians are accurately informed, and so they are usually aware on the effects of pharmacological treatment on laboratory test results, like as those of amiodarone on metabolism of thyroid hormones [5], or those of diuretics on plasma electrolytes and volume homeostasis [6], as well as those caused by toxic effects of chemotherapy on several tissues [7]. However, drug interferences often go unrecognized in the laboratory due to lack of relevant patient medication information [8].
The interferences of immunoassay methods affecting the analytical systems themselves are less known by the clinicians, and so they are more clinically insidious. Among these interferences, those altering antibody binding are probably the most frequent, especially the interferences due to heterophile antibodies and human anti-mouse antibodies (HAMA), depending upon the assay and the analyte [1]. Other interferences, affecting the formation of antigen-antibody complex, can be related to streptavidin-biotin interaction.
The streptavidin-biotin interaction provides an efficient and convenient method for accurately separating free from bound antigen in both competitive and noncompetitive immunoassay methods [9], [10], [11]. Indeed, the high-affinity interaction between streptavidin and biotin is not disturbed by multiple washings, and biotinylation typically does not alter biological activity or immunologic specificity when bound to a molecule [9]. This methodology is currently and widely used in several immunoassay systems using fully automated platforms, including: Access, DxI and DxC (Beckman Coulter); Elecsys, Cobas and Modular platforms (Roche Diagnostics); Isys platforms (Immuno Diagnostic System); Ortho Vitros platform (Ortho Clinical Diagnostics); Dimension Vista, Exl, Immulite platforms (Siemens Healthineers) [11]. However, it is important to note that biotin-streptavidin technology is used only by some (but not all) the immunoassay methods, supported by these fully automated platforms [11]. Accordingly, the biotin interference should be evaluated specifically for each immunoassay methods. Some recent studies suggested that the results of immunoassay methods, using streptavidin-biotin interaction methodology, can be affected by the presence of anti-streptavidin antibodies [10] or very high biotin circulating levels due to supplemental therapy [11], [12], [13]. However, the evidences so far reporting about biotin interference on immunoassay methods are prevalently based on case reports, rather than on experimental or clinical studies, as recently reviewed in detail [11].
In this issue of the Journal, Piketty et al. [14] evaluated the assay interference on several immunoassays for thyroid, steroid and protein hormones (i.e. FT3, FT4, PTH, TSH, 25OH-vitamin D, cortisol, FSH, LH, PTH, and C-peptide), using Cobas e411 platform (Roche Diagnostics), in subjects receiving moderate to very high doses of biotin. The aim of this study was to evaluate the relationship between the daily biotin dose, the plasma biotin concentration, and the magnitude of analytical errors. The degree of interference of biotin concentration was estimated before and after adsorption of biotin, present in the sample, to magnetic microparticles coated with streptavidin, and also by measuring the same plasma samples with other two immunoassays, not based on the biotin-streptavidin procedure for the separation of free from bound antigen (i.e. Liaison XL platform by Diasorin for FT4 and PTH assays, and Access platform by Beckman Coulter for TSH assay) [14]. The biotin concentrations ranged from 31.7 to 1160 ng/mL in 23 plasma samples of patients with multiple sclerosis or voluntary subjects supplemented with daily dose up to 300 mg of biotin [14]. After the removal of biotin by adsorption, the vitamin concentration fall below the detection limit of immunoassay in all but two samples. Most hormonal results using the Cobas e411 platform were abnormal in subjects/patient supplemented with biotin, and normalized after vitamin adsorption with magnetic microparticles coated with streptavidin, suggesting a significant presence of interference. On the contrary, all results of alternative methods were normal except two slight PTH elevations with Liaison platform. Furthermore, the percentage change in hormone concentrations before and after adsorption procedure correlated strongly with the biotin concentration before the procedure, strongly suggesting the biotin is responsible of this interference. The results of this study [14] demonstrate that immunoassays, based on the biotin-streptavidin procedure for the separation of free from bound antigen, can be significantly affected by high biotin concentrations, leading to a high risk of misdiagnosis.
Biotin supplementation has progressively expanded over the last years, due to both medically prescribed therapies and vitamin complex preparations purchased for personal dietary supplements [11]. Recently, high doses of biotin were fond to be a therapeutic option in biotin responsive basal ganglia disease, an orphan neuro-metabolic disease caused by mutation in the SLC19A2 gene coding for a thiamine transporter [15]. Moreover, high doses of biotin, ranging from 100 mg to 300 mg per day, which are 10,000 times the daily recommended intake in general population, are being investigated as a treatment for progressive multiple sclerosis [16]. As far as dietary supplementations are considered, the number of subjects involved, and especially the doses taken by subjects/patients are unknown. Indeed, because biotin is included in several very popular poly-vitamin complex preparations, individuals are not aware to take this vitamin and biotin intake is often not reported in medical history. Consequently, detection of interference from biotin ingestion requires early suspicion through collaboration between laboratory and clinical staff and an understanding of local prescribing (and patient self-prescribing) practice.
Piketty et al. [14] reported that susceptibility of immunoassays to the biotin interference was highly variable: PTH assay was least affected, while 25OH-vitamin D assay was most affected. Clinically misleading results were observed for 25OH-vitamin D and PTH (biotin concentration ≥169 ng/mL), FSH, LH and TSH (≥180 ng/mL), fT4 (≥233 ng/mL), fT3 and cortisol (≥363 ng/mL), and C-peptide (≥487 ng/mL). Considering these results, biotin shows a significant interference (>10%) when plasma concentrations is >30 ng/mL, and the concentrations measured in healthy subjects without vitamin supplementation is <5 ng/mL (which is also the limit of detection of the liquid chromatography-mass spectrometry procedure used by the Authors) [14]. Piketty et al. [14] reported that healthy volunteers taken 15–30 mg/daily dose of biotin can actually show concentrations >30 ng/mL, and so should present some significant interferences in immunoassays, based on biotin-streptavidin technology, and using several fully automated platforms by Roche Diagnostics.
Piketty et al. [14] should be praised not only for demonstrating that biotin can significantly interfere some popular immunoassay methods for peptide, steroid and thyroid hormones, but also for reporting a specific procedure able to confirm that biotin itself is responsible of this interference. It is important to note that another group independently reported a similar procedure to test the presence of biotin interference in clinical samples, based on the adsorption of biotin with magnetic microparticles coated with streptavidin [17]. The results of these two studies [14], [17] demonstrate that the adsorption of biotin with magnetic microparticles coated with streptavidin should be considered to be a simple and accurate laboratory test to evaluate the biotin interference in clinical samples. Of course, the results of these studies [14], [17] should be confirmed for other automated platforms using immunoassays, based on biotin-streptavidin technology for separation of bound to free antigen.
From a clinical point of view, it is important that immunoassay methods other than hormones should be tested for biotin interference, in particular those for cardiac biomarkers. It is conceivable that highly sensitive immunoassays for cardiac troponins, using biotin-streptavidin technology, should be vey sensitive to this type of interference, due to very low concentrations of biomarker, which should be measured. Furthermore, it should be considered from a clinical point of view that even very low interference at the level of decision cut-off value (i.e. 99th percentile of reference population) may induce misdiagnosis.
Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
References
1. Tate J, Ward G. Interferences in immunoassay. Clin Rev Biochem 2004;25:105–20.Search in Google Scholar
2. Plebani M. Errors in clinical laboratories or errors in laboratory medicine? Clin Chem Lab Med 2006;44:750–9.10.1515/CCLM.2006.123Search in Google Scholar
3. Sturgeon C, Vijoen A. Analytical error and interference in immunoassay: minimizing risk. Ann Clin Biochem 2011;48: 418–32.10.1258/acb.2011.011073Search in Google Scholar
4. Siest G, Dawkins SJ, Galteau MM. Drug effects on clinical laboratory tests. J Pharm Biomed Anal 1983;1:247–57.10.1016/0731-7085(83)80038-7Search in Google Scholar
5. Iervasi G, Clerico A, Bonini R, Manfredi C, Berti S, Ravani M, et al. Acute effects of amiodarone administration on thyroid function in patients with cardiac arrhythmia. J Clin Endocrinol Metab 1997;82:275–80.10.1210/jcem.82.1.3675Search in Google Scholar PubMed
6. Tarazi RC, Dustan HP, Frohlic ED. Long-term thiazide therapy in essential hypertension. Evidence for persistent alteration in plasma volume and renin activity. Circulation 1970;41:709–17.10.1161/01.CIR.41.4.709Search in Google Scholar
7. Glassman AB. Hemostatic abnormalities associated with cancer and its therapy. Ann Clin Lab Sci 1997;27:391–5.Search in Google Scholar
8. Dimeski G. Interference testing. Clin Biochem Rev 2008;29(suppl 1):S43–8.Search in Google Scholar
9. Diamandis EP, Christopoulos TK. The biotin-(strept)avidin system: principles and applications in biotechnology. Clin Chem 1991;37:625–36.10.1093/clinchem/37.5.625Search in Google Scholar
10. Rulander NJ, Cardamone D, Senior M, Snyder PJ, Master SR. Interference from anti-streptavidin antibody. Arch Pathol Lab Med 2013;137:1141–6.10.5858/arpa.2012-0270-CRSearch in Google Scholar PubMed
11. Piketty M-L, Polak M, Flechtner I, Gonzales-Briceño L, Souberbielle J-C. False biochemical diagnosis of hyperthyroidism in streptavidin-biotin-based immunoassays: the problem of biotin intake and related interferences. Clin Chem Lab Med 2017;55:780–8.10.1515/cclm-2016-0606Search in Google Scholar PubMed
12. Minkovsky A, Lee MN, Dowlatshahi M, Angell TE, Mahrokhian LS, Petrides AK, et al. High-dose biotin treatment for secondary progressive multiple sclerosis may interfere with thyroid assays. AACE Clin Case Rep 2016;2:e370–3.10.4158/EP161261.CRSearch in Google Scholar PubMed PubMed Central
13. Batista MC, Ferreira CE, Faulhaber AC, Hidal JT, Lottenberg SA, Mangueira CL. Biotin interference in immunoassays mimicking subclinical Graves’ disease and hyperestrogenism: a case series. Clin Chem Lab Med 2017;55:e99–103.10.1515/cclm-2016-0628Search in Google Scholar PubMed
14. Piketty M-L, Prie D, Sedel F, Bernard D, Hercend C, Chanson P, et al. High-dose biotin therapy leading to false biochemical endocrine profiles: validation of a simple method to overcome biotin interference. Clin Chem Lab Med 2017;55:817–25.10.1515/cclm-2016-1183Search in Google Scholar PubMed
15. Tabarki B, Al-Shafi S, Al-Shahwan S, Azmat Z, Al-Hashem A, Al-Adwani N, et al. Biotin-responsive basal ganglia disease revisited: clinical, radiologic, and genetic findings. Neurology 2013;80:261–7.10.1212/WNL.0b013e31827deb4cSearch in Google Scholar PubMed
16. Sedel F, Papeix C, Bellanger A, Touitou V, Lebrun-Frenay C, Galanaud D, et al. High doses of biotin in chronic progressive multiple sclerosis: a pilot study. Mult Scler Relat Disord 2015;4:159–69.10.1016/j.msard.2015.01.005Search in Google Scholar PubMed
17. Lam L, Kyle CV. A simple method to detect biotin interference on immunoassays. Clin Chem Lab Med 2017;55:e104–6.10.1515/cclm-2017-0059Search in Google Scholar PubMed
©2017 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorial
- Biotin interference on immunoassay methods: sporadic cases or hidden epidemic?
- Reviews
- False biochemical diagnosis of hyperthyroidism in streptavidin-biotin-based immunoassays: the problem of biotin intake and related interferences
- Vitamin K plasma levels determination in human health
- Mini Review
- Point-of-care testing INR: an overview
- Opinion Paper
- Could accreditation bodies facilitate the implementation of medical guidelines in laboratories?
- Genetics and Molecular Diagnostics
- QMPSF is sensitive and specific in the detection of NPHP1 heterozygous deletions
- General Clinical Chemistry and Laboratory Medicine
- High-dose biotin therapy leading to false biochemical endocrine profiles: validation of a simple method to overcome biotin interference
- Multicenter performance evaluation of a second generation cortisol assay
- A rapid UPLC-MS/MS assay for the simultaneous measurement of fluconazole, voriconazole, posaconazole, itraconazole, and hydroxyitraconazole concentrations in serum
- A microplate assay to measure classical and alternative complement activity
- Serological diagnosis and prognosis of severe acute pancreatitis by analysis of serum glycoprotein 2
- Retrospective evaluation of the clinical utility of serological biomarkers in Chinese patients with inflammatory bowel disease: 2-year clinical experience
- Infrared analysis of lipoproteins in the detection of alcohol biomarkers
- Coexistence of anti-β2-glycoprotein I domain I and anti-phosphatidylserine/prothrombin antibodies suggests strong thrombotic risk
- Antiphosphatidylserine/prothrombin antibodies as biomarkers to identify severe primary antiphospholipid syndrome
- Cardiovascular Diseases
- The impact of admission neutrophil-to-platelet ratio on in-hospital and long-term mortality in patients with infective endocarditis
- Letters to the Editor
- Biotin interference in immunoassays mimicking subclinical Graves’ disease and hyperestrogenism: a case series
- A simple method to detect biotin interference on immunoassays
- Proposed classification for a variant of Kounis syndrome
- Lyophilized hemoglobin E control material for the dichlorophenol-indophenol (DCIP) test
- Procalcitonin variation before and after 100-km ultramarathon
- Preliminary study in specific activity of molecular components in allergy: implications for diagnostics and relationship with disease severity
- A 2-min at 4500 g rather than a 15-min at 2200 g centrifugation does not impact the reliability of 10 critical coagulation assays
- Immunoassay interference caused by heterophilic antibodies interacting with biotin
- Red blood cell distribution width and mean platelet volume are potential prognostic indices for patients with primary biliary cirrhosis
- Neutrophil CD64 molecule expression can predict bloodstream infection in septic shock patients
Articles in the same Issue
- Frontmatter
- Editorial
- Biotin interference on immunoassay methods: sporadic cases or hidden epidemic?
- Reviews
- False biochemical diagnosis of hyperthyroidism in streptavidin-biotin-based immunoassays: the problem of biotin intake and related interferences
- Vitamin K plasma levels determination in human health
- Mini Review
- Point-of-care testing INR: an overview
- Opinion Paper
- Could accreditation bodies facilitate the implementation of medical guidelines in laboratories?
- Genetics and Molecular Diagnostics
- QMPSF is sensitive and specific in the detection of NPHP1 heterozygous deletions
- General Clinical Chemistry and Laboratory Medicine
- High-dose biotin therapy leading to false biochemical endocrine profiles: validation of a simple method to overcome biotin interference
- Multicenter performance evaluation of a second generation cortisol assay
- A rapid UPLC-MS/MS assay for the simultaneous measurement of fluconazole, voriconazole, posaconazole, itraconazole, and hydroxyitraconazole concentrations in serum
- A microplate assay to measure classical and alternative complement activity
- Serological diagnosis and prognosis of severe acute pancreatitis by analysis of serum glycoprotein 2
- Retrospective evaluation of the clinical utility of serological biomarkers in Chinese patients with inflammatory bowel disease: 2-year clinical experience
- Infrared analysis of lipoproteins in the detection of alcohol biomarkers
- Coexistence of anti-β2-glycoprotein I domain I and anti-phosphatidylserine/prothrombin antibodies suggests strong thrombotic risk
- Antiphosphatidylserine/prothrombin antibodies as biomarkers to identify severe primary antiphospholipid syndrome
- Cardiovascular Diseases
- The impact of admission neutrophil-to-platelet ratio on in-hospital and long-term mortality in patients with infective endocarditis
- Letters to the Editor
- Biotin interference in immunoassays mimicking subclinical Graves’ disease and hyperestrogenism: a case series
- A simple method to detect biotin interference on immunoassays
- Proposed classification for a variant of Kounis syndrome
- Lyophilized hemoglobin E control material for the dichlorophenol-indophenol (DCIP) test
- Procalcitonin variation before and after 100-km ultramarathon
- Preliminary study in specific activity of molecular components in allergy: implications for diagnostics and relationship with disease severity
- A 2-min at 4500 g rather than a 15-min at 2200 g centrifugation does not impact the reliability of 10 critical coagulation assays
- Immunoassay interference caused by heterophilic antibodies interacting with biotin
- Red blood cell distribution width and mean platelet volume are potential prognostic indices for patients with primary biliary cirrhosis
- Neutrophil CD64 molecule expression can predict bloodstream infection in septic shock patients