Abstract
Background:
Cellular analysis in cerebrospinal fluid (CSF) provides important diagnostic information in various medical conditions. The aim of this study was to evaluate the application of Mindray BC-6800 body fluid (BF) mode in cytometric analysis of CSF compared to light microscopy (LM).
Methods:
One hundred and twenty-nine consecutive CSF samples were analyzed by BC-6800-BF mode as well as by LM. The study also included limits of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), carryover and linearity.
Results
White blood cells LoQ was 4.0×106 cells/L. Linearity was good and carryover was negligible. As for the total and white blood cells, the BC-6800-BF parameters vs. LM showed both bias ranged from –10.28 to 0.06×106 cells/L. Polymorphonuclear and mononuclear cells ranged from 6.64 to 10.90%. For white blood cell the diagnostic agreement was 93% at the cut-off >5.0×106 cells/L, and for polymorphonuclear and mononuclear at the cut-off >50% was 91% and 92%, respectively.
Conclusions:
BC-6800-BF offers rapid and accurate counts in clinically relevant concentration ranges, replacing LM for most samples. However, in samples with abnormal cell counts or with abnormal white blood cell differential scattergrams the need to microscopic review for a correct clinical outcome remains.
Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
References
1. Fleming C, Russcher H, Lindemans J, de Jonge R. Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease. Clin Chem Lab Med 2015;53:1689–706.10.1515/cclm-2014-1247Search in Google Scholar PubMed
2. Bignardi GE. Flow cytometry for the microscopy of body fluids in patients with suspected infection. J Clin Pathol 2015;68:870–8.10.1136/jclinpath-2015-203088Search in Google Scholar PubMed
3. Sandhaus LM. Body fluid cell counts by automated methods. Clin Lab Med 2015;35:93–103.10.1016/j.cll.2014.10.003Search in Google Scholar PubMed
4. De Luca GC, Bartleson JD. When and how to investigate the patient with headache. Semin Neurol 2010;30:131–44.10.1055/s-0030-1249221Search in Google Scholar PubMed
5. Fitch MT, Abrahamian FM, Moran GJ, Talan DA. Emergency department management of meningitis and encephalitis. Infect Dis Clin North Am 2008;22:33–52.10.1016/j.idc.2007.10.001Search in Google Scholar PubMed
6. de Graaf MT, de Jongste AH, Kraan J, Boonstra JG, Sillevis Smitt PA, Gratama JW. Flow cytometric characterization of cerebrospinal fluid cells. Cytometry Part B Clin Cytom 2011;80:271–81.10.1002/cyto.b.20603Search in Google Scholar PubMed
7. Clinical and Laboratory Standard Institute. Body fluid analysis for cellular composition; approved guidelines. CLSI document H56-A. Wayne, PA: Clinical and Laboratory Standard Institute, 2006.Search in Google Scholar
8. Zimmermann M, Otto C, Gonzalez JB, Prokop S, Ruprecht K. Cellular origin and diagnostic significance of high-fluorescent cells in cerebrospinal fluid detected by the XE-5000 hematology analyzer. Int J Lab Hematol 2013;35:580–8.10.1111/ijlh.12090Search in Google Scholar PubMed
9. Cho YU, Chi HS, Park SH, Jang S, Kim YJ, Park CJ. Body fluid cellular analysis using the Sysmex XN-2000 automatic hematology analyzer: focusing on malignant samples. Int J Lab Hematol 2015;37:346–56.10.1111/ijlh.12292Search in Google Scholar PubMed
10. Mahieu S, Vertessen F, Van der Planken M. Evaluation of ADVIA 120 CSF assay (Bayer) vs. chamber counting of cerebrospinal fluid specimens. Clin Lab Haematol 2004;26:195–9.10.1111/j.1365-2257.2004.00611.xSearch in Google Scholar PubMed
11. Zimmermann M, Weimann A. Automated vs. manual cerebrospinal fluid cell counts: a work and cost analysis comparing the Sysmex XE-5000 and the Fuchs-Rosenthal manual counting chamber. Int J Lab Hematol 2011;33:629–37.10.1111/j.1751-553X.2011.01339.xSearch in Google Scholar PubMed
12. Kresie L, Benavides D, Bollinger P, Walters J, Pierson D, Richmond T, et al. Perfomance evaluation of body fluids on the Sysmex XE-2100 series automated hematology analyzer. Lab Hematol 2005;11:24–30.10.1532/LH96.04067Search in Google Scholar PubMed
13. Brown W, Keeney M, Chin-Yee I, Johnson K, Lantis K, Finn W et al. Validation of body fluid analysis on the Coulter LH 750. Lab Hematol 2003;9:155–9.Search in Google Scholar
14. Hoffmann JJ, Jansen WC. Automated counting of cells in cerebrospinal fluid using the Celldyn-4000 haematology analyzer. Clin Chem Lab Med 2002;44:1168–73.10.1515/cclm.2002.204Search in Google Scholar
15. De Smet D, Van Moer G, Martens GA, Nanos N, Smet L, Jochmans K, et al. Use of the Cell-Dyn Sapphire hematology analyzer for automated counting of blood cells in body fluids. Am J Clin Pathol 2010;133:291–99.10.1309/AJCPY7J7OLASZAPCSearch in Google Scholar PubMed
16. Boer K, Deufel T, Reinhofer M. Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid. Clin Biochem 2009;42:684–91.10.1016/j.clinbiochem.2009.01.025Search in Google Scholar PubMed
17. de Jong R, Brower R, de Graaf MT, Luitwieler RL, Fleming C, de Frankrijker-Merkestijn M, et al. Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids. Clin Chem Lab Med 2010;48:665–75.10.1515/CCLM.2010.108Search in Google Scholar PubMed
18. Pernè A, Hainfellner JA, Womastek I, Haushofer A, Szekeres T, Schwarzinger I. Performance evaluation of the Sysmex XE-5000 hematology analyzer for white blood cell analysis in cerebrospinal fluid. Arch Pathol Lab Med 2012;19:194–8.10.5858/arpa.2011-0030-OASearch in Google Scholar PubMed
19. Li A, Gronlund E, Brattsand G. Autoamted white blood cell counts in cerebrospinal fluid using the body fluid mode on the platform Sysmex XE-5000. Scand J Clin Lab Invest 2014;74:673–80.10.3109/00365513.2014.939994Search in Google Scholar PubMed
20. Letho TM, Leskinen P, Hedberg P, Vaskivuo TE. Evaluation of the Sysmex XT-4000i for the automated body fluid analysis. Int J Lab Hematol 2014;36:114–23.10.1111/ijlh.12134Search in Google Scholar
21. Fleming C, Brouwer R, Lindemans J, de Jong R. Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids. Clin Chem Lab Med 2012;50:1791–8.10.1515/cclm-2011-0927Search in Google Scholar PubMed
22. Zur B, Eichhorn L, Albers E, Stoffel-Wagner B. Evaluation of 2 hematology analyzers in body fluid mode versus flow cytometry immunophenotyping of mainly neurosurgical cerebrospinal fluid samples. J Neurol Surg A Cent Eur Neurosurg 2012;73:93–8.10.1055/s-0032-1309067Search in Google Scholar PubMed
23. Strik H, Luthe H, Nagel I, Ehrilch B, Bahr M. Automated cerebrospinal fluid cytology. Limitation and reasonable applications. Anal Quant Cytol Histol 2005;27:167–73.Search in Google Scholar
24. Nanos NE, Delanghe JR. Evaluation of Sysmex UF-1000i for use in cerebrospinal fluid analysis. Clinica Chimica Acta 2008;392:30–3.10.1016/j.cca.2008.02.020Search in Google Scholar PubMed
25. Buoro S, Ottomano C, Appassiti S, Gherardi P, Alessio MG, Crippa A, et al. Analytical and clinical evaluation of Sysmex UF1000i for automated screening of cerebrospinal fluids. J Med Biochem 2013;32:1–6.Search in Google Scholar
26. Buoro S, Apassiti Esposito S, Alessio M, Crippa A, Ottomano C, Lippi G. Automated cerebrospinal fluid cell counts using the new body fluid mode of Sysmex UF-1000i. J Clin Lab Anal 2015; doi: 10.1002/jcla.21866.10.1002/jcla.21866Search in Google Scholar PubMed PubMed Central
27. Glasser L, Murphy CA, Machan JT. The clinical reliability of automated cerebrospinal fluid cell counts on the Beckman-Coulter LH750 and Iris iQ200. Am J Clin Pathol 2009;131:58–63.10.1309/AJCPCHHTZ6UDQ6WFSearch in Google Scholar PubMed
28. Walker TJ, Nelson LD, Dunphy BW, Anderson DM, Kickler TS. Comparative evaluation of the Iris iQ200 body fluid module with manual hemacytometer count. Am J Clin Pathol 2009;131:333–8.10.1309/AJCPSVG4POONQQ1RSearch in Google Scholar PubMed
29. Buoro S, Mecca T, Azzarà G, Seghezzi M, Vavassori M, Crippa A et al. Mindray BC-6800 automated body fluid mode cell counts on pleural and ascitic fluids. Int J Lab Hematol 2016;38:90–101.10.1111/ijlh.12446Search in Google Scholar PubMed
30. Lippi G, Cattabiani C, Benegiamo A, Gennari D, Pavesi F, Caleffi A, et al. Evaluation of white blood cell count in peritoneal fluid with five different hemocytometers. Clin Biochem 2013;46:173–6.10.1016/j.clinbiochem.2012.10.034Search in Google Scholar PubMed
31. Bourner G, De La Salle B, George T, Tabe Y, Baum H, Culp N, et al. ICSH guidelines for the verification and performance of automated cell counters for body fluids. Int J Lab Hematol 2014;36:598–612.10.1111/ijlh.12196Search in Google Scholar PubMed
32. Briggs C, Culp N, Davis B, D’onofrio G, Zini G, Machin SJ. ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting Int. Jnl Lab Hematol 2014;36:613–27.10.1111/ijlh.12201Search in Google Scholar PubMed
33. Clinical and Laboratory Standards Institute. Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation of instrumental Methods; Approved Guideline – second edition. CLSI document H20-A2. Wayne, PA: Clinical and Laboratory Standards Institute, 2010.Search in Google Scholar
34. Clinical and Laboratory Standards Institute. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved guideline – second edition. CLSI document EP17-A2. Wayne, PA: Clinical Laboratory Standards Institute, 2012.Search in Google Scholar
35. Clinical and Laboratory Standard Institute. Evaluation of the linearity of quantitative measurement procedures: a statistical approach; Approved Guideline. CLSI document EP06-A. Wayne, PA: Clinical and Laboratory Standard Institute, 2003.Search in Google Scholar
36. Bremell D, Mattsson N, Wallin F, Henriksson J, Wall M, Blennow K, et al. Automated cerebrospinal fluid cell count-new reference ranges and evaluation of its clinical use in central nervous system infections. Clin Biochem 2014;47:25–30.10.1016/j.clinbiochem.2013.11.010Search in Google Scholar PubMed
37. Sandhaus LM, Ciarlini P, Kidric D, Dillman C, O’Riordan M. Automated cerebrospinal fluid cell counts using the Sysmex XE-5000: is it time for new reference ranges? Am J Clin Pathol 2010;134:734–8.10.1309/AJCPABGQXSIA4SMTSearch in Google Scholar PubMed
38. Omuse G, Makau P. Interference of cerebrospinal fluid white blood cell counts performed on the Sysmex XT-4000i by yeast and bacteria. Int J Lab Hematol 2013;35:e5–7.10.1111/ijlh.12088Search in Google Scholar PubMed
39. Kim HR, Park BR, Lee MK. Effects of bacteria and yeast on WBC counting in three automated hematology counters. Ann Hematol 2008;87:557–62.10.1007/s00277-008-0464-1Search in Google Scholar PubMed
40. Buoro S, Appassiti Esposito S, Vavassori M, Mecca T, Ottomano C, Dominoni P, et al. Reflex testing rules for cell count and differentiation of nucleated elements in pleural and ascitic fluids on Sysmex XE-5000. J Lab Autom 2016;21:297–304.10.1177/2211068215593375Search in Google Scholar PubMed
41. Buoro S, Mecca T, Azzarà G, Seghezzi M, Dominoni P, Crippa A, et al. Cell Population Data and reflex testing rules of cell analysis in pleural and ascitic fluids using body fluid mode on Sysmex XN-9000. Clin Chim Acta 2016;452:92–8.10.1016/j.cca.2015.11.005Search in Google Scholar PubMed
©2016 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorials
- CCLM Award for the Most Cited Paper Recently Published
- Laboratory economics. Risk or opportunity?
- Reviews
- Molecular diagnosis and precision medicine in allergy management
- The insulin autoimmune syndrome (IAS) as a cause of hypoglycaemia: an update on the pathophysiology, biochemical investigations and diagnosis
- Opinion Paper
- Role of microsatellite instability, immunohistochemistry and mismatch repair germline aberrations in immunosuppressed transplant patients: a phenocopy dilemma in Muir-Torre syndrome
- Genetics and Molecular Diagnostics
- Time and tumor type (primary or metastatic) do not influence the detection of BRAF/NRAS mutations in formalin fixed paraffin embedded samples from melanomas
- False low holotranscobalamin levels in a patient with a novel TCN2 mutation
- General Clinical Chemistry and Laboratory Medicine
- Quality performance of laboratory testing in pharmacies: a collaborative evaluation
- Cost evaluation of clinical laboratory in Taiwan’s National Health System by using activity-based costing
- Impact of sample processing on the measurement of circulating microparticles: storage and centrifugation parameters
- HbA1c: EQA in Germany, Belgium and the Netherlands using fresh whole blood samples with target values assigned with the IFCC reference system
- Iohexol clearance in unstable critically ill patients: a tool to assess glomerular filtration rate
- Characteristics of Chinese patients with antiphospholipid syndrome and the ability of lupus anticoagulant assays to identify them
- Reference Values and Biological Variations
- Distribution of soluble suppression of tumorigenicity 2 (sST2), N-terminal pro-brain natriuretic peptide (NT-proBNP), high sensitive troponin I and high-sensitive troponin T in umbilical cord blood
- Hematology and Coagulation
- Evaluation of Mindray BC-6800 body fluid mode for automated cerebrospinal fluid cell counting
- Cancer Diagnostics
- Increased sialylation and reduced fucosylation of exfoliated cervical cells are potential markers of carcinogenesis in the cervix
- Cardiovascular Diseases
- High incidence of macrotroponin I with a high-sensitivity troponin I assay
- Infectious Diseases
- Prospective evaluation of biomarkers for prediction of quality of life in community-acquired pneumonia
- Letter to the Editor
- The use of extra-analytical phase quality indicators by clinical laboratories: the results of an international survey
- Quality of reporting of diagnostic test accuracy studies in medical laboratory journals
- Impact of under-filled blood collection tubes containing K2EDTA and K3EDTA as anticoagulants on automated complete blood count (CBC) testing
- Stability of plasma albumin depends on measurement method
- Evaluation of screening method for Bence Jones protein analysis
- Reference intervals for the Kryptor second-generation chromogranin A assay
- Evaluation of an automated urinary iodine measurement using AU5800 analyzer with AutoLab Iodine reagent
- Evaluation of the accuracy of complete blood count for insufficient blood samples
- The use of a “gray zone” considering measurement uncertainty in pharmacological tests. The serum growth hormone stimulation test as an example
- Keeping Ebola out of the lab: a practical solution on how to analyze Ebola associated blood anomalies
- Standardized fixation process is crucial to permit molecular analyses in formalin-fixed and paraffin-embedded melanoma samples
- Congress Abstracts
- 48th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)
- 8th Santorini Conference Systems Medicine and Personalised Health and Therapy
- Congress of Laboratory Medicine and Clinical Chemistry
Articles in the same Issue
- Frontmatter
- Editorials
- CCLM Award for the Most Cited Paper Recently Published
- Laboratory economics. Risk or opportunity?
- Reviews
- Molecular diagnosis and precision medicine in allergy management
- The insulin autoimmune syndrome (IAS) as a cause of hypoglycaemia: an update on the pathophysiology, biochemical investigations and diagnosis
- Opinion Paper
- Role of microsatellite instability, immunohistochemistry and mismatch repair germline aberrations in immunosuppressed transplant patients: a phenocopy dilemma in Muir-Torre syndrome
- Genetics and Molecular Diagnostics
- Time and tumor type (primary or metastatic) do not influence the detection of BRAF/NRAS mutations in formalin fixed paraffin embedded samples from melanomas
- False low holotranscobalamin levels in a patient with a novel TCN2 mutation
- General Clinical Chemistry and Laboratory Medicine
- Quality performance of laboratory testing in pharmacies: a collaborative evaluation
- Cost evaluation of clinical laboratory in Taiwan’s National Health System by using activity-based costing
- Impact of sample processing on the measurement of circulating microparticles: storage and centrifugation parameters
- HbA1c: EQA in Germany, Belgium and the Netherlands using fresh whole blood samples with target values assigned with the IFCC reference system
- Iohexol clearance in unstable critically ill patients: a tool to assess glomerular filtration rate
- Characteristics of Chinese patients with antiphospholipid syndrome and the ability of lupus anticoagulant assays to identify them
- Reference Values and Biological Variations
- Distribution of soluble suppression of tumorigenicity 2 (sST2), N-terminal pro-brain natriuretic peptide (NT-proBNP), high sensitive troponin I and high-sensitive troponin T in umbilical cord blood
- Hematology and Coagulation
- Evaluation of Mindray BC-6800 body fluid mode for automated cerebrospinal fluid cell counting
- Cancer Diagnostics
- Increased sialylation and reduced fucosylation of exfoliated cervical cells are potential markers of carcinogenesis in the cervix
- Cardiovascular Diseases
- High incidence of macrotroponin I with a high-sensitivity troponin I assay
- Infectious Diseases
- Prospective evaluation of biomarkers for prediction of quality of life in community-acquired pneumonia
- Letter to the Editor
- The use of extra-analytical phase quality indicators by clinical laboratories: the results of an international survey
- Quality of reporting of diagnostic test accuracy studies in medical laboratory journals
- Impact of under-filled blood collection tubes containing K2EDTA and K3EDTA as anticoagulants on automated complete blood count (CBC) testing
- Stability of plasma albumin depends on measurement method
- Evaluation of screening method for Bence Jones protein analysis
- Reference intervals for the Kryptor second-generation chromogranin A assay
- Evaluation of an automated urinary iodine measurement using AU5800 analyzer with AutoLab Iodine reagent
- Evaluation of the accuracy of complete blood count for insufficient blood samples
- The use of a “gray zone” considering measurement uncertainty in pharmacological tests. The serum growth hormone stimulation test as an example
- Keeping Ebola out of the lab: a practical solution on how to analyze Ebola associated blood anomalies
- Standardized fixation process is crucial to permit molecular analyses in formalin-fixed and paraffin-embedded melanoma samples
- Congress Abstracts
- 48th National Congress of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC – Laboratory Medicine)
- 8th Santorini Conference Systems Medicine and Personalised Health and Therapy
- Congress of Laboratory Medicine and Clinical Chemistry