Analytical interference by monoclonal immunoglobulins on the direct bilirubin AU Beckman Coulter assay: the benefit of unsuspected diagnosis from spurious results
Abstract
Background: Monoclonal (M) components can interfere with the direct bilirubin (D-Bil) assay on the AU Beckman Coulter instrumentation and produce spurious results, such as D-Bil values greater than total bilirubin (T-Bil) or very low/negative D-Bil values. If properly detected, this interference may uncover undiagnosed patients with monoclonal gammopathy (MG).
Methods: We investigated the interference rate on the D-Bil AU assay in serum samples known to contain M proteins along with their isotype and described the protocol set up in our laboratory to help with the diagnosis of MG based on D-Bil spurious results as first indication.
Results: During a period of 4 years, 15.4% (345 of 2235) of serum samples containing M immunoglobulins produced erroneous D-Bil results, although no clear relationship between the magnitude or isotype of the M component and interference could be found. In total 22 new patients were diagnosed with MG based on the analytical artefact with the D-Bil as first indication.
Conclusions: The D-Bil interference from MG on the Beckman AU analysers needs to be made known to laboratories in order to prevent clinical confusion and/or additional workup to explain the origin of anomalous results. Although this information may not add to the management of existing patients with serum paraproteins, it can benefit patients that have not been diagnosed with MG by triggering follow up testing to determine if M components are present.
Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.
Research funding: None declared.
Employment or leadership: None declared.
Honorarium: None declared.
Competing interests: The funding organisation(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.
References
1. Kroll MH, Elin RJ. Interference with clinical laboratory analyses. Clin Chem 1994;40:1996–2005.10.1093/clinchem/40.11.1996Suche in Google Scholar
2. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105–20.Suche in Google Scholar
3. Bakker AJ, Mücke M. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007;45:1240–3.10.1515/CCLM.2007.254Suche in Google Scholar PubMed
4. Dalal BI, Brigden ML. Factitious biochemical measurements resulting from hematologic conditions. Am J Clin Pathol 2009;131:195–204.10.1309/AJCPY9RP5QYTYFWCSuche in Google Scholar PubMed
5. Pantanowitz L, Horowitz GL, Upalakalin JN, Beckwith BA. Artifactual hyperbilirubinemia due to paraprotein interference. Arch Pathol Lab Med 2003;127:55–9.10.5858/2003-127-55-AHDTPSuche in Google Scholar PubMed
6. Yang Y, Howanitz PJ, Howanitz JH, Gorfajn H, Wong K. Paraproteins are a common cause of interferences with automated chemistry methods. Arch Pathol Lab Med 2008;132:217–23.10.5858/2008-132-217-PAACCOSuche in Google Scholar PubMed
7. Nauti A, Barassi A, Merlini G, d’Eril GV. Paraprotein interference in an assay of conjugated bilirubin. Clin Chem 2005;51:1076–7.10.1373/clinchem.2005.048710Suche in Google Scholar PubMed
8. Song L, Kelly KA, Butch AW. Monoclonal and polyclonal immunoglobulin interference in a conjugated bilirubin assay. Arch Pathol Lab Med 2014;138:950–4.10.5858/arpa.2013-0042-OASuche in Google Scholar PubMed
9. Seimiya M, Suzuki Y, Yoshida T, Sawabe Y, Matsushita K, Nomura F. The abnormal reaction data-detecting function of the automated biochemical analyzer was useful to prevent erroneous total-bilirubin measurement and to identify monoclonal proteins. Clin Chim Acta 2015;441:44–6.10.1016/j.cca.2014.12.013Suche in Google Scholar PubMed
10. Jelinek AG, Bachmann LM. Unexpected test results in a patient with multiple myeloma. Clin Chem 2014;60:1375–8.10.1373/clinchem.2013.213884Suche in Google Scholar PubMed
11. Wadhera RK, Rajkumar SV. Prevalence of monoclonal gammopathy of undetermined significance: a systematic review. Mayo Clin Proc 2010;85:933–42.10.4065/mcp.2010.0337Suche in Google Scholar PubMed PubMed Central
12. Kyle RA, Rajkumar SV. Monoclonal gammopathies of undetermined significance. Best Pract Res Clin Haematol 2005;18:689–707.10.1016/j.beha.2005.01.025Suche in Google Scholar PubMed
13. Monk C, Wallage M, Wassell J, Whiteway A, James J, Beetham R. A monoclonal protein identified by an anomalous lipaemia index. Ann Clin Biochem 2009;46:250–2.10.1258/acb.2008.008192Suche in Google Scholar PubMed
©2016 by De Gruyter
Artikel in diesem Heft
- Frontmatter
- Editorials
- Laboratory analytical quality – the process continues
- Measurement uncertainty – a revised understanding of its calculation and use
- Review
- Substrate-zymography: a still worthwhile method for gelatinases analysis in biological samples
- Opinion Papers
- Is the combination of trueness and precision in one expression meaningful? On the use of total error and uncertainty in clinical chemistry
- The problem with total error models in establishing performance specifications and a simple remedy
- Measurement uncertainty for clinical laboratories – a revision of the concept
- Uncertainty in measurement and total error – are they so incompatible?
- General Clinical Chemistry and Laboratory Medicine
- Using the hazard ratio to evaluate allowable total error in predictive measurands
- Performance of electrolyte measurements assessed by a trueness verification program
- Analytical interference by monoclonal immunoglobulins on the direct bilirubin AU Beckman Coulter assay: the benefit of unsuspected diagnosis from spurious results
- Preliminary probe of quality indicators and quality specification in total testing process in 5753 laboratories in China
- Analytical and clinical evaluation of the new Fujirebio Lumipulse®G non-competitive assay for 25(OH)-vitamin D and three immunoassays for 25(OH)D in healthy subjects, osteoporotic patients, third trimester pregnant women, healthy African subjects, hemodialyzed and intensive care patients
- Patient-performed extraction of faecal calprotectin
- Comparison of the clinical utility of the Elia CTD Screen to indirect immunofluorescence on Hep-2 cells
- Reference Values and Biological Variations
- Sex-related differences in the association of ghrelin levels with obesity in adolescents
- Gestation specific reference intervals for thyroid function tests in pregnancy
- Cancer Diagnostics
- SOX17 promoter methylation in plasma circulating tumor DNA of patients with non-small cell lung cancer
- Dopamine concentration in blood platelets is elevated in patients with head and neck paragangliomas
- Letters to the Editor
- Cancer dynamics and the success of cancer screening programs
- Mother’s instinct – a rare case of multiple test interferences due to heterophile antibodies
- Sigma metric or defects per million opportunities (DPMO): the performance of clinical laboratories should be evaluated by the Sigma metrics at decimal level with DPMOs
- A national survey of preanalytical handling of oral glucose tolerance tests in pregnancy
- Updating pregnancy diabetes guidelines: is (y)our laboratory ready?
- Low serum bilirubin values are associated with pulmonary embolism in a case-control study
- Effect of Hb H on HbA1c measurements as measured by IFCC reference method and affinity HPLC
- Adipocytes in venipunctures cause falsely elevated S-100B serum values
- Earlier detection of sepsis by Candida parapsilosis using three-dimensional cytographic anomalies on the Mindray BC-6800 hematological analyzer
- Theranos phenomenon – part 4: Theranos at an International Conference
Artikel in diesem Heft
- Frontmatter
- Editorials
- Laboratory analytical quality – the process continues
- Measurement uncertainty – a revised understanding of its calculation and use
- Review
- Substrate-zymography: a still worthwhile method for gelatinases analysis in biological samples
- Opinion Papers
- Is the combination of trueness and precision in one expression meaningful? On the use of total error and uncertainty in clinical chemistry
- The problem with total error models in establishing performance specifications and a simple remedy
- Measurement uncertainty for clinical laboratories – a revision of the concept
- Uncertainty in measurement and total error – are they so incompatible?
- General Clinical Chemistry and Laboratory Medicine
- Using the hazard ratio to evaluate allowable total error in predictive measurands
- Performance of electrolyte measurements assessed by a trueness verification program
- Analytical interference by monoclonal immunoglobulins on the direct bilirubin AU Beckman Coulter assay: the benefit of unsuspected diagnosis from spurious results
- Preliminary probe of quality indicators and quality specification in total testing process in 5753 laboratories in China
- Analytical and clinical evaluation of the new Fujirebio Lumipulse®G non-competitive assay for 25(OH)-vitamin D and three immunoassays for 25(OH)D in healthy subjects, osteoporotic patients, third trimester pregnant women, healthy African subjects, hemodialyzed and intensive care patients
- Patient-performed extraction of faecal calprotectin
- Comparison of the clinical utility of the Elia CTD Screen to indirect immunofluorescence on Hep-2 cells
- Reference Values and Biological Variations
- Sex-related differences in the association of ghrelin levels with obesity in adolescents
- Gestation specific reference intervals for thyroid function tests in pregnancy
- Cancer Diagnostics
- SOX17 promoter methylation in plasma circulating tumor DNA of patients with non-small cell lung cancer
- Dopamine concentration in blood platelets is elevated in patients with head and neck paragangliomas
- Letters to the Editor
- Cancer dynamics and the success of cancer screening programs
- Mother’s instinct – a rare case of multiple test interferences due to heterophile antibodies
- Sigma metric or defects per million opportunities (DPMO): the performance of clinical laboratories should be evaluated by the Sigma metrics at decimal level with DPMOs
- A national survey of preanalytical handling of oral glucose tolerance tests in pregnancy
- Updating pregnancy diabetes guidelines: is (y)our laboratory ready?
- Low serum bilirubin values are associated with pulmonary embolism in a case-control study
- Effect of Hb H on HbA1c measurements as measured by IFCC reference method and affinity HPLC
- Adipocytes in venipunctures cause falsely elevated S-100B serum values
- Earlier detection of sepsis by Candida parapsilosis using three-dimensional cytographic anomalies on the Mindray BC-6800 hematological analyzer
- Theranos phenomenon – part 4: Theranos at an International Conference