Comparative evaluation of the Aptima HIV-1 Quant Dx assay and COBAS TaqMan HIV-1 v2.0 assay using the Roche High Pure System for the quantification of HIV-1 RNA in plasma

  • Gunnar Schalasta EMAIL logo , Anna Börner , Andrea Speicher and Martin Enders
Published/Copyright: August 18, 2015

Abstract

Background: Quantification of human immunodeficiency virus type 1 (HIV-1) RNA in plasma has become the standard of care in the management of HIV-infected patients. There are several commercially available assays that have been implemented for the detection of HIV-1 RNA in plasma. Here, the new Hologic Aptima® HIV-1 Quant Dx assay (Aptima HIV) was compared to the Roche COBAS® TaqMan® HIV-1 Test v2.0 for use with the High Pure System (HPS/CTM).

Methods: The performance characteristics of the assays were assessed using commercially available HIV reference panels, dilution of the WHO 3rd International HIV-1 RNA International Standard (WHO-IS) and plasma from clinical specimens. Assay performance was determined by linear regression, Deming correlation analysis and Bland-Altman analysis.

Results: Testing of HIV-1 reference panels revealed excellent agreement. The 61 clinical specimens quantified in both assays were linearly associated and strongly correlated.

Conclusions: The Aptima HIV assay offers performance comparable to that of the HPS/CTM assay and, as it is run on a fully automated platform, a significantly improved workflow.


Corresponding author: Gunnar Schalasta, Prof. Gisela Enders and Kollegen MVZ Rosenbergstr. 85, 70193 Stuttgart, Germany, Phone: +49 7116357140, E-mail:

Acknowledgments

This study was supported by Hologic Inc., San Diego, CA. We offer special thanks to the members of the Molecular Biology Department of the Laboratory Prof. Enders in Stuttgart, Germany for excellent technical assistance.

Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

Research funding: None declared.

Employment or leadership: None declared.

Honorarium: None declared.

Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Hughes MD, Johnson VA, Hirsch MS, Bremer JW, Elbeik T, Erice A, et al. Monitoring plasma HIV-1 RNA levels in addition to CD4+ lymphocyte count improves assessment of antiretroviral therapeutic response. Ann Intern Med 1997;126:929–38.10.7326/0003-4819-126-12-199706150-00001Search in Google Scholar PubMed

2. Mellors JW, Muñoz A, Giorgi JV, Margolick JB, Tassoni CJ, Gupta P, et al. Plasma viral load and CD4+ lymphocytes as prognostic markers of HIV-1 infection. Ann Intern Med 1997;126:946–54.10.7326/0003-4819-126-12-199706150-00003Search in Google Scholar PubMed

3. O’Brien WA, Hartigan PM, Daar ES, Simberkoff MS, Hamilton JD. Changes in plasma HIV RNA levels and CD4+ lymphocyte counts predict both response to antiretroviral therapy and therapeutic failure. VA Cooperative Study Group on AIDS. Ann Intern Med 1997;126:939–45.10.7326/0003-4819-126-12-199706150-00002Search in Google Scholar PubMed

4. Asboe D, Aitken C, Boffito M, Booth C, Cane P, Fakoya A, et al., on behalf of the BHIVA Guidelines Subcommittee. British HIV Association guidelines for the routine investigation and monitoring of adult HIV-1-infected individuals 2011. HIV Med 2012;13:1–44.10.1111/j.1468-1293.2011.00971.xSearch in Google Scholar PubMed

5. Website: Department of Health and Human Services. Panel on Antiretroviral Guidelines for Adults and Adolescents. Guidelines for the use of antiretroviral agents in HIV-1-infected adults and adolescents. Available from: http://aidsinfo.nih.gov/ContentFiles/AdultandAdolescentGL.pdf. Accessed 1 December, 2014.Search in Google Scholar

6. Thompson MA, Aberg JA, Hoy JF, Telenti A, Benson C, Cahn P, et al. Antiretroviral treatment of adult HIV infection: 2012 recommendations of the International Antiviral Society-USA panel. J Am Med Assoc 2012;308:387–402.10.1001/jama.2012.7961Search in Google Scholar PubMed

7. World Health Organization. Consolidated guidelines on the use of antiretroviral drugs for treating and preventing HIV infection. Recommendations for a public health approach. Geneva, Switzerland: WHO Press, 2013.Search in Google Scholar

8. Bourlet T, Signori-Schmuck A, Roche L, Icard V, Saoudin H, Trabaud MA, et al. HIV-1 load comparison using four commercial real-time assays. J Clin Microbiol 2011;49:292–7.10.1128/JCM.01688-10Search in Google Scholar PubMed PubMed Central

9. Church D, Gregson D, Lloyd T, Klein M, Beckthold B, Laupland K, et al. Comparison of the RealTime HIV-1, COBAS TaqMan 48 v1.0, Easy Q v1.2 and Versant assays for determination of HIV-1 viral loads in a cohort of Canadian patients with diverse HIV subtype infections. J Clin Microbiol 2011;49:118–24.10.1128/JCM.00685-10Search in Google Scholar PubMed PubMed Central

10. Garcia-Diaz A, Clewley GS, Booth CL, Labett W, McAllister N, Geretti AM. Comparative evaluation of the performance of the Abbott Real-Time Human Immunodeficiency Virus type 1 (HIV-1) assay for the measurement of HIV-1 plasma viral load following automated specimen preparation. J Clin Microbiol 2006;44:1788–91.10.1128/JCM.44.5.1788-1791.2006Search in Google Scholar PubMed PubMed Central

11. Garcia-Diaz A, Labbett W, Clewley GS, Guerrero-Ramos A, Geretti AM. Comparative evaluation of the Artus HIV-1 QS-RGQ assay and the Abbott RealTime HIV-1 assay for the quantification of HIV-1 RNA in plasma. J Clin Virol 2013;57:66–9.10.1016/j.jcv.2013.01.006Search in Google Scholar PubMed

12. Schutten M, Peters D, Back NK, Beld M, Beuselinck K, Foulongne V, et al. Multicenter evaluation of the new Abbott RealTime assays for quantitative detection of human immunodeficiency virus type 1 and hepatitis C virus RNA. J Clin Microbiol 2007;45: 1712–7.10.1128/JCM.02385-06Search in Google Scholar PubMed PubMed Central

13. Perrin L, Pawlotsky JM, Bouvier-Alias M, Sarrazin C, Zeuzem S, Colucci G. Multicenter performance evaluation of a new TaqMan PCR assay for monitoring human immunodeficiency virus RNA load. J Clin Microbiol 2006;44:4371–5.10.1128/JCM.00505-06Search in Google Scholar PubMed PubMed Central

14. Coombs R, Bremer J, Sanchez J, La Rosa A, Fernandes H, Ganoza C, et al. Clinical performance evaluation of the COBAS TaqMan HIV-1 Test for use with the High Pure System v2.0 (HPS/CTM HIV v2) using manual specimen extraction, compared to the COBAS AmpliPrep/COBAS TaqMan HIV-1 Test v2.0 (CAP/CTM HIV Test v2) using automated specimen extraction. J Int AIDS Soc 2012;15(Suppl 3):THPE060.Search in Google Scholar

15. Ferenci P, Fried MW, Shiffman ML, Smith CI, Marinos G, Goncales FL Jr, et al. Predicting sustained virological responses in chronic hepatitis C patients treated with peginterferon alfa-2a (40 KD)/ribavirin. J Hepatol 2005;43:425–33.10.1016/j.jhep.2005.04.009Search in Google Scholar PubMed

16. Wiesmann F, Naeth G, Sarrazin C, Berger A, Kaiser R, Ehret R, et al. Variation analysis of six HCV viral load assays using low viremic HCV samples in the range of the clinical decision points for HCV protease inhibitors. Med Microbiol Immunol 2015;204:515–25.10.1007/s00430-014-0364-zSearch in Google Scholar PubMed PubMed Central

17. Pyne MT, Vest L, Clement J, Lee J, Rosvall JR, Luk K, et al. Comparison of three Roche hepatitis B virus viral load assay formats. J Clin Microbiol 2012;50:2337–42.10.1128/JCM.00746-12Search in Google Scholar PubMed PubMed Central

18. Hill CS. Molecular diagnostic testing for infectious diseases using TMA technology. Expert Rev Mol Diagn 2001;1:445–55.10.1586/14737159.1.4.445Search in Google Scholar PubMed

19. Janssens W, Heyndrickx L, Fransen K, Motte J, Peeters M, Nkengasong JN, et al. Genetic and phylogenetic analysis of env subtypes G and H in central Africa. Aids Res Hum Retroviruses 1994;10:877–9.10.1089/aid.1994.10.877Search in Google Scholar PubMed

20. Paba P, Fabeni L, Ciccozzi M, Perno CF, Ciotti M. Performance evaluation of the COBAS/TaqMan HIV-1 v2.0 in HIV-1 positive patients with low viral load: a comparative study. J Virol Methods 2011;173:399–402.10.1016/j.jviromet.2011.03.014Search in Google Scholar PubMed

21. Ruelle J, Debaisieux L, Vancutsem E, De Bel A, Delforge ML, Piérard D, et al. HIV-1 low-level viraemia assessed with 3 commercial real-time PCR assays show high variability. BMC Infect Dis 2012;12:100–5.10.1186/1471-2334-12-100Search in Google Scholar PubMed PubMed Central

22. Yan CS, Hanafi I, Kelleher AD, Carr AD, Amin J, McNally LP, et al. Lack of correlation between three commercial platforms for the evaluation of human immunodeficiency virus type 1 (HIV-1) viral load at the clinically critical lower limit of quantification. J Clin Virol 2010;49:249–53.10.1016/j.jcv.2010.08.016Search in Google Scholar PubMed

Received: 2015-6-5
Accepted: 2015-7-22
Published Online: 2015-8-18
Published in Print: 2016-3-1

©2016 by De Gruyter

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. The way of prostate cancer diagnostics
  4. Review
  5. Statistical methods used in the calculation of geriatric reference intervals: a systematic review
  6. Opinion Paper
  7. The side effects of translational omics: overtesting, overdiagnosis, overtreatment
  8. Genetics and Molecular Diagnostics
  9. Rapid detection of non-deletional mutations causing α-thalassemia by multicolor melting curve analysis
  10. General Clinical Chemistry and Laboratory Medicine
  11. Patient pools and the use of “patient means” are valuable tools in quality control illustrated by a bone-specific alkaline phosphatase assay
  12. Long-term stability of glucose: 96-h study using Terumo Glycaemia tubes
  13. Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes
  14. Croatian laboratories have a good knowledge of the proper detection and management of hemolyzed, icteric and lipemic samples
  15. Fetal exposure to ethanol: relationship between ethyl glucuronide in maternal hair during pregnancy and ethyl glucuronide in neonatal meconium
  16. Comparing the effect of isotopically labeled or structural analog internal standards on the performance of a LC-MS/MS method to determine ciclosporin A, everolimus, sirolimus and tacrolimus in whole blood
  17. Relationship between matrix metalloproteinase-9 and oxidative stress in drug-free male schizophrenia: a case control study
  18. Comparison of functional fibrinogen (FF/CFF) and FIBTEM in surgical patients – a retrospective study
  19. Reference Values and Biological Variations
  20. Assessment of serum free light chain levels in healthy adults immediately after marathon running
  21. Pharmacokinetics of a novel dosing regimen of oral melatonin in critically ill patients
  22. Cancer Diagnostics
  23. An epidemiology-based model as a tool to monitor the outbreak of inappropriateness in tumor marker requests: a national scale study
  24. Comparative analysis of prostate cancer specific biomarkers PCA3 and ERG in whole urine, urinary sediments and exosomes
  25. Infectious Diseases
  26. Comparative evaluation of the Aptima HIV-1 Quant Dx assay and COBAS TaqMan HIV-1 v2.0 assay using the Roche High Pure System for the quantification of HIV-1 RNA in plasma
  27. Evaluating the use of procalcitonin in an asymptomatic, HIV-infected antiretroviral therapy-naïve, South African cohort
  28. Diabetes
  29. Early prediction of gestational diabetes: a practical model combining clinical and biochemical markers
  30. Letters to the Editors
  31. Highly-trained dogs’ olfactory system for detecting biochemical recurrence following radical prostatectomy
  32. More on the accuracy of the Architect enzymatic assay for hemoglobin A1c and its traceability to the IFCC reference system
  33. Assessing quality from an accuracy-based HbA1c proficiency survey
  34. HbG-Honolulu interferes with some cation-exchange HPLC HbA1c assays
  35. Between analyser differences in chloride measurements and thus anion gap cause different interpretations of the acid-base balance
  36. Non-albumininuric proteinuria: a urinary tubular marker in the diagnosis of diabetic kidney disease
  37. International Normalized Ratio (INR) testing: analytical and clinical performance of four point-of-care devices versus central laboratory instrumentation analysis
  38. Genetic screening of the makorin ring finger 3 gene in girls with idiopathic central precocious puberty
  39. Practical approach for medical validation of therapeutic drug monitoring results
  40. Marked elevation of procalcitonin in a patient with a drug related infusion reaction to rituximab
Downloaded on 23.1.2026 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2015-0522/html?lang=en
Scroll to top button