Why specifications for allowable glucose meter errors should include 100% of the data
-
Jan S. Krouwer
The recently published ISO 15197 guideline [1], and the CLSI POCT12-A3 guideline [2] are two new glucose meter performance guidelines. Unfortunately, each of these guidelines fails to include error limits for 100% of the data. For the CLSI guideline, specifications are given for 98% of the results and for the ISO guideline, 99% of the results. The purpose of this article is to discuss why a specification for 100% of the results is appropriate.
I have previously commented [3] that in clinical chemistry, errors are often thought of as a distribution of continuous variables and not as discrete error events. Thus, these glucose meter guidelines provide one set of error limits for 95% of the data and a wider set of limits for either 98% or 99% of the data. There is nothing wrong with this construct. However, one can also consider glucose meter errors as discrete failure events. For example, if a glucose sample reference result was 1.67 mmol/L (30 mg/dL), a serious glucose meter failure event would be if the meter result for that sample was 16.65 mmol/L (300 mg/dL) (e.g., within the E zone of a Parkes error grid [4]). And for each sample, this failure event can either occur or not occur. It makes sense to specify a goal of zero percent of data in the E zone of the Parkes error grid. The ISO and CLSI goals can be thought of as specifying the percentage of results that fall within the A and perhaps B region of an error grid. Thus, the ISO and CLSI goals are based on a distribution of a continuous variable and goals for the higher zones in an error grid, which are missing in the ISO and CLSI guidelines, are based on discrete failure events. The balance of the data (2% for CLSI and 1% for ISO) is allowed to fall in intermediate but not the highest error zones. Note that in other fields, where people think in terms of discrete failure events, one would never set goals in the way that they are done for glucose meters. Thus, one would not specify that nuclear power plants should run without meltdowns 99% of the time nor that correct site surgery should occur 99% of the time. Perhaps a difficulty arises because one can never (statistically) prove the occurrence of zero failures but the goal itself is valid.
Manufacturers worry about the approval process and in performing method comparisons, the possibility of an outlier is perhaps a rationale for the ISO and CLSI glucose meter specifications. However, it is the wrong way to think about the problem. If there were one E zone result in a 125 sample method comparison, this error rate of 0.8% would be acceptable according to either glucose meter standard. On the one hand, if this glucose meter had 1% of the glucose meter market (total market=7.9 billion results per year in the US), this implies over 632,000 dangerous glucose results per year in the US [5]. On the other hand, if this glucose meter company ran 10 million samples with no results in the E zone, this only proves (with 95% confidence) that no more than 29 dangerous results would occur [6]. Thus, proving that rare events do not occur by running method comparisons is not practical.
What is done both in other fields and in clinical chemistry is to perform Failure Mode Effects Analysis (FMEA) and fault tree analysis to enumerate the sources of possible failure events and to implement mitigations where appropriate. These techniques also do not prove zero failures but complement method comparisons which need to be performed anyway.
If an E zone event does occur in a method comparison, this by itself is not a reason to consider the meter to be unacceptable. The most important question is the cause of the outlier and it is possible that no cause will be found. There is a financial constraint to collecting data to decide on meter approval and the use of post-market data, the size of which dwarfs method comparison studies should also be used as a way to monitor meter performance.
Conflict of interest statement
Author’s conflict of interest disclosure: The author stated that there are no conflicts of interest regarding the publication of this article.
Research funding: None declared.
Employment or leadership: None declared.
References
1. ISO 15197:2013 In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization, 2013.Suche in Google Scholar
2. CLSI POCT12-A3 point-of-care blood glucose testing in acute and chronic care facilities; approved guideline, 3rd ed. Wayne, PA: CLSI, 2013.Suche in Google Scholar
3. Krouwer JS. Recommendation to treat continuous variable errors like attribute errors. Clin Chem Lab Med 2006;44:797–8.10.1515/CCLM.2006.152Suche in Google Scholar PubMed
4. Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000;23:1143–8.10.2337/diacare.23.8.1143Suche in Google Scholar PubMed
5. Assumes that patients who take insulin test 3 times per day – FDA public meeting clinical accuracy requirements for point of care blood glucose meters March 16, 2010. Available from: http://www.fda.gov/downloads/MedicalDevices/NewsEvents/WorkshopsConferences/UCM208598.pdf. Accessed on 23 May, 2013.Suche in Google Scholar
6. Hahn GJ, Meeker WQ. Statistical intervals. A guide for practitioners. New York: Wiley, 1991:104.10.1002/9780470316771Suche in Google Scholar
©2013 by Walter de Gruyter Berlin Boston
Artikel in diesem Heft
- Letters to the Editors
- Clinical utility of serum tumor markers and cytokines in cervical cancer and neoplasia
- Establishing reference intervals for LDL subfractions in a Korean population using the Lipoprint LDL system
- Measurement imprecision of common urinary biochemical analytes on the Roche Cobas 6000 system
- A comparison between turbidimetric inhibition immunoassay and capillary electrophoresis in glycated hemoglobin (HbA1c) measurement
- Commutability: a peculiar property of calibration and control materials. Definition and evaluation
- Diagnostic sensitivity of a panel of tests to detect monoclonal protein in Korean multiple myeloma patients
- Commutability of proficiency testing (PT): status of the matrix-related bias in general clinical chemistry
- Masthead
- Masthead
- Editorial
- Why specifications for allowable glucose meter errors should include 100% of the data
- Reviews
- ABO blood group: old dogma, new perspectives
- Trace elements and bone health
- Mini Review
- The role of transcription factors in laboratory medicine
- Opinion Papers
- Nobelitis: a common disease among Nobel laureates?
- The syndrome of the “obsessive-compulsory scientist”: a new mental disorder?
- Can current analytical quality performance of UK clinical laboratories support evidence-based guidelines for diabetes and ischaemic heart disease? – A pilot study and a proposal
- Guidelines and Recommendations
- Survey of national guidelines, education and training on phlebotomy in 28 European countries: an original report by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) working group for the preanalytical phase (WG-PA)
- General Clinical Chemistry and Laboratory Medicine
- Red cell indices: differentiation between β-thalassemia trait and iron deficiency anemia and application to sickle cell disease and sickle cell thalassemia
- Problems in determining thalassemia carrier status in a program for prevention and control of severe thalassemia syndromes: a lesson from Thailand
- An enzyme linked immunosorbent assay (ELISA) for the determination of the human haptoglobin phenotype
- Blood loss from laboratory diagnostic tests in children
- Hematocrit correction does not improve glucose monitor accuracy in the assessment of neonatal hypoglycemia
- Evaluation of a mobile clinical pathology laboratory developed for the home care of pediatric patients following transplantation of peripheral blood precursor cells
- Folic acid supplementation does not reduce intracellular homocysteine, and may disturb intracellular one-carbon metabolism
- Influence of spurious hemolysis on blood gas analysis
- Serum procalcitonin predicts development of acute kidney injury in patients with suspected infection
- Reference Values and Biological Variations
- Nationwide multicenter study aimed at the establishment of common reference intervals for standardized clinical laboratory tests in Japan
- Cancer Diagnostics
- Application of BRAF, NRAS, KRAS mutations as markers for the detection of papillary thyroid cancer from FNAB specimens by pyrosequencing analysis
- Comparative evaluation of the My5-FU™ immunoassay and LC-MS/MS in monitoring the 5-fluorouracil plasma levels in cancer patients
Artikel in diesem Heft
- Letters to the Editors
- Clinical utility of serum tumor markers and cytokines in cervical cancer and neoplasia
- Establishing reference intervals for LDL subfractions in a Korean population using the Lipoprint LDL system
- Measurement imprecision of common urinary biochemical analytes on the Roche Cobas 6000 system
- A comparison between turbidimetric inhibition immunoassay and capillary electrophoresis in glycated hemoglobin (HbA1c) measurement
- Commutability: a peculiar property of calibration and control materials. Definition and evaluation
- Diagnostic sensitivity of a panel of tests to detect monoclonal protein in Korean multiple myeloma patients
- Commutability of proficiency testing (PT): status of the matrix-related bias in general clinical chemistry
- Masthead
- Masthead
- Editorial
- Why specifications for allowable glucose meter errors should include 100% of the data
- Reviews
- ABO blood group: old dogma, new perspectives
- Trace elements and bone health
- Mini Review
- The role of transcription factors in laboratory medicine
- Opinion Papers
- Nobelitis: a common disease among Nobel laureates?
- The syndrome of the “obsessive-compulsory scientist”: a new mental disorder?
- Can current analytical quality performance of UK clinical laboratories support evidence-based guidelines for diabetes and ischaemic heart disease? – A pilot study and a proposal
- Guidelines and Recommendations
- Survey of national guidelines, education and training on phlebotomy in 28 European countries: an original report by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) working group for the preanalytical phase (WG-PA)
- General Clinical Chemistry and Laboratory Medicine
- Red cell indices: differentiation between β-thalassemia trait and iron deficiency anemia and application to sickle cell disease and sickle cell thalassemia
- Problems in determining thalassemia carrier status in a program for prevention and control of severe thalassemia syndromes: a lesson from Thailand
- An enzyme linked immunosorbent assay (ELISA) for the determination of the human haptoglobin phenotype
- Blood loss from laboratory diagnostic tests in children
- Hematocrit correction does not improve glucose monitor accuracy in the assessment of neonatal hypoglycemia
- Evaluation of a mobile clinical pathology laboratory developed for the home care of pediatric patients following transplantation of peripheral blood precursor cells
- Folic acid supplementation does not reduce intracellular homocysteine, and may disturb intracellular one-carbon metabolism
- Influence of spurious hemolysis on blood gas analysis
- Serum procalcitonin predicts development of acute kidney injury in patients with suspected infection
- Reference Values and Biological Variations
- Nationwide multicenter study aimed at the establishment of common reference intervals for standardized clinical laboratory tests in Japan
- Cancer Diagnostics
- Application of BRAF, NRAS, KRAS mutations as markers for the detection of papillary thyroid cancer from FNAB specimens by pyrosequencing analysis
- Comparative evaluation of the My5-FU™ immunoassay and LC-MS/MS in monitoring the 5-fluorouracil plasma levels in cancer patients