Comparison of the 99th percentiles of three troponin I assays in a large reference population
-
Astrid Petersmann
, Till Ittermann
, Cornelia Fries , Norbert Lubenow , Thomas Kohlmann , Anders Kallner , Andreas Greinacher and Matthias Nauck
Abstract
Background: The IFCC Task Force on Clinical Applications of Cardiac Biomarkers suggests comparing several contemporary sensitive troponin assays in the same, sufficiently large reference population.
Methods: Three contemporary sensitive assays were used to measure troponin I concentration in samples from a uniquely large healthy population (2404 individuals) and in a sub-group with tighter inclusion criteria of 908 individuals. The 99th percentiles were calculated using quantile regression which takes the entire population into account.
Results: The 99th percentile for the ARCHITECT STAT Troponin I assay was 21 ng/L, 31 ng/L for the ADVIA Centaur Troponin I-Ultra assay and 28 ng/L for the Dimension Vista cTnI assay. Significantly higher values were found in males than in women only in the Dimension Vista cTnI assay and in the subgroup for the ARCHITECT STAT Troponin I assay.
Conclusions: Quantile regression provides a tool to accurately estimate the 99th percentile and establish a continuous function of the relation between the 99th percentile and the age and gender. There was no age dependency demonstrated. A gender difference was found in one assay but inconclusive in another and not demonstrated in a third.
References
1. Jaffe AS, Ravkilde J, Roberts R, Naslund U, Apple FS, Galvani M, et al. It’s time for a change to a troponin standard. Circulation 2000;102:1216–20.10.1161/01.CIR.102.11.1216Search in Google Scholar PubMed
2. Collinson PO, Heung YM, Gsze D, Boa F, Senior R, Christenson R, et al. Influence of population selection on the 99th percentile reference value for cardiac troponin assays. Clin Chem 2012;58:219–25.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000299052700035&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1373/clinchem.2011.171082Search in Google Scholar PubMed
3. Keller T, Ojeda F, Zeller T, Wild PS, Tzikas S, Sinning CR. Defining a reference population to determine the 99th percentile of a contemporary sensitive cardiac troponin I assay. Int J Cardiol 2012 May 4. DOI: http://dx.doi.org/10.1016/j.ijcard.2012.04.063.10.1016/j.ijcard.2012.04.063http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000323566800067&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Search in Google Scholar PubMed
4. Cardinaels EP, Mingels AM, Jacobs LH, Meex SJ, Bekers O, van Dieijen-Visser MP. A comprehensive review of upper reference limits reported for (high-)sensitivity cardiac troponin assays: the challenges that lie ahead. Clin Chem Lab Med 2012;50:791–806.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000304434200005&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Search in Google Scholar
5. Apple FS, Collinson PO. Analytical characteristics of high-sensitivity cardiac troponin assays. Clin Chem 2012; 58:54–61.10.1373/clinchem.2011.165795Search in Google Scholar PubMed
6. Apple FS, Ler R, Murakami MA. Determination of 19 cardiac troponin I and T assay 99th percentile values from a common presumably healthy population. Clin Chem 2012;58:1574–81.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000310489300013&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1373/clinchem.2012.192716Search in Google Scholar PubMed
7. Koenker R. Quantile regression econometric society monograph series. New York: Cambridge University Press, 2005.10.1017/CBO9780511754098Search in Google Scholar
8. Royston P, Sauerbrei W. Multivariable model-building: a pragmatic approach to regression analysis based on fractional polynomials for modelling continuous variables. New York: John Wiley & Sons, 2008.10.1002/9780470770771Search in Google Scholar
9. CLSI. Defining, establishing, and verifying reference intervals in the clinical laboratory; approved guideline, 3rd ed. CLSI document C28-A3. Wayne, PA: CLSI, 2008.Search in Google Scholar
10. Solberg HE, PetitClerc C. Approved recommendation (1988) on the theory of reverence values. Part 3. Preparation of individuals and collection of specimens for the production of reference values. Clin Chim Acta 1988;177:S1–12.10.1016/0009-8981(88)90074-5Search in Google Scholar
11. Hickman PE, Tony Badrick T, Wilson SR, McGill D. Reporting of cardiac troponin – problems with the 99th population percentile. Clin Chim Acta 2007;381:182–3.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000247513700016&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1016/j.cca.2007.03.012Search in Google Scholar PubMed
12. Solberg HE. Approved recommendation (1987) on the theory of reverence values. Part 5. Statistical treatment of collected reference values. Determination of reference limits. Clin Chim Acta 1987;170:S13–32.10.1016/0009-8981(87)90151-3Search in Google Scholar
13. Arrebola MM, Lillo JA, Diez De Los Ríos MJ, Rodríguez M, Dayaldasani A, Yahyaoui R, et al. Analytical performance of a sensitive assay for cardiac troponin I with loci™ technology. Clin Biochem 2010;43:998–1002.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000280026900011&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1016/j.clinbiochem.2010.04.073Search in Google Scholar PubMed
14. Clerico A, Fortunato A, Ripoli A, Prontera C, Zucchelli CG, Emdin M. Distribution of plasma cardiac troponin I values in healthy subjects: pathophysiological considerations. Clin Chem Lab Med 2008;46:804–8.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000257542800008&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Search in Google Scholar
15. Ittermann T, Roser M, Wood G, Preez H, Lüdemann J, Völlzke H, et al. Reference intervals for eight measurands of the blood count in a large population based study. Clin Lab 2010; 56:9–19.Search in Google Scholar
16. Wu AH, Shea E, Lu QT, Minyard J, Bui K, Hsu JC, et al. Short- and long-term cardiac troponin I analyte stability in plasma and serum from healthy volunteers by use of an ultrasensitive, single-molecule counting assay. Clin Chem 2009;55:2058–9.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000271269700024&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f3Search in Google Scholar
©2013 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- Letter to the Editor
- Elevated level of cell-free plasma DNA is associated with advanced-stage breast cancer and metastasis
- Is procalcitonin a reliable marker of sepsis in critically ill septic patients undergoing continuous veno-venous hemodiafiltration with “high cut-off” membranes (HCO-CVVHDF)?
- Challenging our serological algorithm for celiac disease (CD) diagnosis by the ESPGHAN guidelines
- Preanalytical quality control in a university hospital in China
- Racial differences and relationships between gestational thyrotropin and free thyroxine in a multiracial Asian population
- Analytical performance and method comparison study of the total homocysteine immunoassay on the AIA 600II analyser
- Easy verification of clinical chemistry reference intervals
- The Elecsys® Vitamin B12 assay is not affected by anti-intrinsic factor antibodies
- Chemiluminescence-based cobalamin assay errors: background and perspectives
- Congress Abstracts
- Abstracts IV Italian Great Network Congress Rome, 14th–18th October 2013*)
- Masthead
- Masthead
- Editorial
- Making colorectal cancer screening FITTER for purpose with quantitative faecal immunochemical tests for haemoglobin (FIT)
- From “panic” to “critical” values: which path toward harmonization?
- Review
- Cerebrospinal fluid analyses for the diagnosis of subarachnoid haemorrhage and experience from a Swedish study. What method is preferable when diagnosing a subarachnoid haemorrhage?
- Opinion Paper
- False myths and legends in laboratory diagnostics
- General Clinical Chemistry and Laboratory Medicine
- National survey on critical values notification of 599 institutions in China
- Influence of physical properties of cuvette surface on measurement of serum lipase
- Red cell indices: differentiation between β-thalassemia trait and iron deficiency anemia and application to sickle-cell disease and sickle-cell thalassemia
- Measurement of immature platelets with Abbott CD-Sapphire and Sysmex XE-5000 in haematology and oncology patients
- Performance characteristics of consensus approaches for small and minor paroxysmal nocturnal hemoglobinuria clone determination by flow cytometry
- Comparison of PR3-ANCA specific assay performance for the diagnosis of granulomatosis with polyangiitis (Wegener’s)
- The integration of the detection of systemic sclerosis-associated antibodies in a routine laboratory setting: comparison of different strategies
- Reference Values and Biological Variations
- Reference interval studies: what is the maximum number of samples recommended?
- Short-term estimation and application of biological variation of small dense low-density lipoproteins in healthy individuals
- Cancer Diagnostics
- Diagnostic inconsistency of faecal immunochemical tests for haemoglobin in population screening of colorectal cancer
- Cardiovascular Diseases
- Comparison of the 99th percentiles of three troponin I assays in a large reference population
- Assessment of plasma aminothiol levels and the association with recurrent atherothrombotic events in patients hospitalized for an acute coronary syndrome: a prospective study
- Diabetes
- The relationship between estimated average glucose and fasting plasma glucose
- Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay
Articles in the same Issue
- Letter to the Editor
- Elevated level of cell-free plasma DNA is associated with advanced-stage breast cancer and metastasis
- Is procalcitonin a reliable marker of sepsis in critically ill septic patients undergoing continuous veno-venous hemodiafiltration with “high cut-off” membranes (HCO-CVVHDF)?
- Challenging our serological algorithm for celiac disease (CD) diagnosis by the ESPGHAN guidelines
- Preanalytical quality control in a university hospital in China
- Racial differences and relationships between gestational thyrotropin and free thyroxine in a multiracial Asian population
- Analytical performance and method comparison study of the total homocysteine immunoassay on the AIA 600II analyser
- Easy verification of clinical chemistry reference intervals
- The Elecsys® Vitamin B12 assay is not affected by anti-intrinsic factor antibodies
- Chemiluminescence-based cobalamin assay errors: background and perspectives
- Congress Abstracts
- Abstracts IV Italian Great Network Congress Rome, 14th–18th October 2013*)
- Masthead
- Masthead
- Editorial
- Making colorectal cancer screening FITTER for purpose with quantitative faecal immunochemical tests for haemoglobin (FIT)
- From “panic” to “critical” values: which path toward harmonization?
- Review
- Cerebrospinal fluid analyses for the diagnosis of subarachnoid haemorrhage and experience from a Swedish study. What method is preferable when diagnosing a subarachnoid haemorrhage?
- Opinion Paper
- False myths and legends in laboratory diagnostics
- General Clinical Chemistry and Laboratory Medicine
- National survey on critical values notification of 599 institutions in China
- Influence of physical properties of cuvette surface on measurement of serum lipase
- Red cell indices: differentiation between β-thalassemia trait and iron deficiency anemia and application to sickle-cell disease and sickle-cell thalassemia
- Measurement of immature platelets with Abbott CD-Sapphire and Sysmex XE-5000 in haematology and oncology patients
- Performance characteristics of consensus approaches for small and minor paroxysmal nocturnal hemoglobinuria clone determination by flow cytometry
- Comparison of PR3-ANCA specific assay performance for the diagnosis of granulomatosis with polyangiitis (Wegener’s)
- The integration of the detection of systemic sclerosis-associated antibodies in a routine laboratory setting: comparison of different strategies
- Reference Values and Biological Variations
- Reference interval studies: what is the maximum number of samples recommended?
- Short-term estimation and application of biological variation of small dense low-density lipoproteins in healthy individuals
- Cancer Diagnostics
- Diagnostic inconsistency of faecal immunochemical tests for haemoglobin in population screening of colorectal cancer
- Cardiovascular Diseases
- Comparison of the 99th percentiles of three troponin I assays in a large reference population
- Assessment of plasma aminothiol levels and the association with recurrent atherothrombotic events in patients hospitalized for an acute coronary syndrome: a prospective study
- Diabetes
- The relationship between estimated average glucose and fasting plasma glucose
- Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay