Home Influence of physical properties of cuvette surface on measurement of serum lipase
Article
Licensed
Unlicensed Requires Authentication

Influence of physical properties of cuvette surface on measurement of serum lipase

  • Lien Cattoir , Valerie Pede , An-Sofie Decavele , Marc L. De Buyzere , Marijn M. Speeckaert , David Seccombe , Veerle Boterberg , Peter Dubruel , Frank de Bisschop and Joris R. Delanghe EMAIL logo
Published/Copyright: July 10, 2013

Abstract

Background: Despite striking similarities among colorimetric lipase assay recipes, marked intervendor differences are noted in the reported lipase values. In the present study, the effect of physical properties of the cuvette surface on measurement of serum lipase was investigated.

Methods: Lipase activity was measured concomitantly in cuvettes from three different analyzers: Vista (Siemens), Modular (Roche), and Synchron (Beckman Coulter). The surface/volume ratio of the cuvettes and the contact angle of the cuvette polymers were determined. The effects of various characteristics of serum (biochemical parameters, surface tension) were also examined.

Results: Serum lipase activities based on the colorimetric methylresorufin assay differed markedly according to the cuvettes used. More specifically, in the lower activity rate, marked differences were reported. The physical properties of the various cuvettes showed remarkable differences, especially in the contact angles. Other biochemical parameters (bilirubin, alkaline phosphatase, total cholesterol, high-density lipoprotein cholesterol, low-density lipoprotein cholesterol, and triglycerides) and serum surface tension did not affect the results.

Conclusions: Serum lipase activity is affected by the physical properties of the cuvette surface.


Corresponding author: Joris R. Delanghe, Department of Clinical Chemistry, De Pintelaan 185, 9000 Gent, Belgium, Phone: +32-9-332-2956, Fax: +32-9-332-4985, E-mail:

References

1. Lott JA, Patel ST, Sawhney AK, Kazmierczak SC, Love JE Jr. Assays of serum lipase: analytical and chemical considerations. Clin Chem 1986;32:1290–302.10.1093/clinchem/32.7.1290Search in Google Scholar

2. Lott JA, Lu CJ. Lipase isoforms and amylase isoenzymes: assays and application in the diagnosis of acute pancreatitis. Clin Chem 1991;37:361–8.10.1093/clinchem/37.3.361Search in Google Scholar

3. Tietz NW, Shuey DF. Lipase in serum – the elusive enzyme: an overview. Clin Chem 1993;39:746–56.10.1093/clinchem/39.5.746Search in Google Scholar

4. Gilham D, Lehner R. Techniques to measure lipase and esterase activity in vitro. Methods 2005;36:139–47.10.1016/j.ymeth.2004.11.003Search in Google Scholar

5. Sarda L, Desnuelle P. Action de la lipase pancréatique sur les esters en émulsion. Biochim Biophys Acta 1958;30:513–21.10.1016/0006-3002(58)90097-0Search in Google Scholar

6. Predotova M, Kretschmann R, Gebauer J, Buerkert A. Effects of cuvette surface material on ammonia-, nitrous oxide-, carbon dioxide-, and methane-concentration measurements. J Plant Nutr Soil Sci 2011;174:347–9.http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcApp=PARTNER_APP&SrcAuth=LinksAMR&KeyUT=000291361500001&DestLinkType=FullRecord&DestApp=ALL_WOS&UsrCustomerID=b7bc2757938ac7a7a821505f8243d9f310.1002/jpln.201000420Search in Google Scholar

7. Available from www.google.com/patents/US4988497. Accessed May 15, 2013.Search in Google Scholar

8. Friedewald WT, Levy RI, Frederickson DS. Estimation of the concentration of low-density lipoprotein cholesterol in plasma, without use of the preparative ultracentrifuge. Clin Chem 1972;18:499–502.10.1093/clinchem/18.6.499Search in Google Scholar

9. De Bisschop FR, Rigole WJ. A physical model for liquid capillary bridges between adsorptive solid spheres: the nodoid of plateau. J Colloid Interf Sci 1982;88:117–28.10.1016/0021-9797(82)90161-8Search in Google Scholar

10. Rosina J, Kvasnak E, Suta D, Kolarova H, Malek J, Krajci L. Temperature dependence of blood surface tension. Physiol Res 2007;56:S93–8.10.33549/physiolres.931306Search in Google Scholar

11. Hrncír E, Rosina J. Surface tension of blood. Physiol Res 1997;46:319–21.Search in Google Scholar

12. Suttiprasit P, Krisdhasima V, McGuire J. The surface activity of α-lactalbumin, β-lactoglobulin, and bovine serum albumin: I. Surface tension measurements with single-component and mixed solutions. J Colloid Interf Sci 1992;154:316–26.10.1016/0021-9797(92)90146-DSearch in Google Scholar

13. Gutnick DL, Bach H. Biosurfactants. Comprehensive biotechnology, 2nd ed. 2011;3:699–715.10.1016/B978-0-08-088504-9.00237-3Search in Google Scholar

Received: 2013-05-16
Accepted: 2013-06-13
Published Online: 2013-07-10
Published in Print: 2013-11-01

©2013 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Letter to the Editor
  2. Elevated level of cell-free plasma DNA is associated with advanced-stage breast cancer and metastasis
  3. Is procalcitonin a reliable marker of sepsis in critically ill septic patients undergoing continuous veno-venous hemodiafiltration with “high cut-off” membranes (HCO-CVVHDF)?
  4. Challenging our serological algorithm for celiac disease (CD) diagnosis by the ESPGHAN guidelines
  5. Preanalytical quality control in a university hospital in China
  6. Racial differences and relationships between gestational thyrotropin and free thyroxine in a multiracial Asian population
  7. Analytical performance and method comparison study of the total homocysteine immunoassay on the AIA 600II analyser
  8. Easy verification of clinical chemistry reference intervals
  9. The Elecsys® Vitamin B12 assay is not affected by anti-intrinsic factor antibodies
  10. Chemiluminescence-based cobalamin assay errors: background and perspectives
  11. Congress Abstracts
  12. Abstracts IV Italian Great Network Congress Rome, 14th–18th October 2013*)
  13. Masthead
  14. Masthead
  15. Editorial
  16. Making colorectal cancer screening FITTER for purpose with quantitative faecal immunochemical tests for haemoglobin (FIT)
  17. From “panic” to “critical” values: which path toward harmonization?
  18. Review
  19. Cerebrospinal fluid analyses for the diagnosis of subarachnoid haemorrhage and experience from a Swedish study. What method is preferable when diagnosing a subarachnoid haemorrhage?
  20. Opinion Paper
  21. False myths and legends in laboratory diagnostics
  22. General Clinical Chemistry and Laboratory Medicine
  23. National survey on critical values notification of 599 institutions in China
  24. Influence of physical properties of cuvette surface on measurement of serum lipase
  25. Red cell indices: differentiation between β-thalassemia trait and iron deficiency anemia and application to sickle-cell disease and sickle-cell thalassemia
  26. Measurement of immature platelets with Abbott CD-Sapphire and Sysmex XE-5000 in haematology and oncology patients
  27. Performance characteristics of consensus approaches for small and minor paroxysmal nocturnal hemoglobinuria clone determination by flow cytometry
  28. Comparison of PR3-ANCA specific assay performance for the diagnosis of granulomatosis with polyangiitis (Wegener’s)
  29. The integration of the detection of systemic sclerosis-associated antibodies in a routine laboratory setting: comparison of different strategies
  30. Reference Values and Biological Variations
  31. Reference interval studies: what is the maximum number of samples recommended?
  32. Short-term estimation and application of biological variation of small dense low-density lipoproteins in healthy individuals
  33. Cancer Diagnostics
  34. Diagnostic inconsistency of faecal immunochemical tests for haemoglobin in population screening of colorectal cancer
  35. Cardiovascular Diseases
  36. Comparison of the 99th percentiles of three troponin I assays in a large reference population
  37. Assessment of plasma aminothiol levels and the association with recurrent atherothrombotic events in patients hospitalized for an acute coronary syndrome: a prospective study
  38. Diabetes
  39. The relationship between estimated average glucose and fasting plasma glucose
  40. Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay
Downloaded on 27.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2013-0369/html
Scroll to top button