Home Practical application of Sigma Metrics QC procedures in clinical chemistry
Article
Licensed
Unlicensed Requires Authentication

Practical application of Sigma Metrics QC procedures in clinical chemistry

  • Christian H.H. Schoenmakers , Andre J.M. Naus , Henricus J. Vermeer , Douwe van Loon and Gerard Steen
Published/Copyright: July 29, 2011

Abstract

Background: Many labs have not yet selected the most appropriate Westgard Quality Control (QC) rule for each test. This is mainly due to the apparent complexity of the matter.

Methods: From the Westgard OPSpecs Charts QC planning tool and the Sigma Metrics formula's it was deduced that every Westgard rule has its own Sigma value. This was converted to an easy three-step road map to optimal Westgard QC rules.

Results: The road map provided is based on Sigma Metrics that hold a definition of “world class quality”, at which no further effort to increase quality needs to be taken. Furthermore, it is shown that clinical chemical tests can be classified as “good”: quality at or above world class, “bad”: quality below world class but controllable with Westgard QC rules and “ugly”: quality not controllable with Westgard QC rules alone. Finally, practical tips of how to deal with this and related aspects are given.

Conclusions: The use of the road map based on Sigma Metrics leads to fast and easy implementation of optimal Westgard QC rules.


Corresponding author: Dr. Christian H.H. Schoenmakers, AKL, Elkerliek Ziekenhuis, PO Box 98, 5700 AB Helmond, The Netherlands Phone: +31 492 595052, Fax: +31 492 595059,

Received: 2010-12-15
Accepted: 2011-6-27
Published Online: 2011-07-29
Published in Print: 2011-11-01

©2011 by Walter de Gruyter Berlin Boston

Articles in the same Issue

  1. Editorial
  2. Thank you, indeed!
  3. Reviews
  4. Glomerular filtration rate estimation: performance of serum cystatin C-based prediction equations
  5. Oxidative stress and diabetes mellitus
  6. Human papilloma virus (HPV) molecular diagnostics
  7. Opinion Papers
  8. Urine immunofixation electrophoresis remains important and is complementary to serum free light chain
  9. Permissible performance limits of regression analyses in method comparisons
  10. General Clinical Chemistry and Laboratory Medicine
  11. The value of serum ischemia-modified albumin for assessing liver function in patients with chronic liver disease
  12. Relationship between bilirubin and C-reactive protein
  13. Status of serum-calcium and -albumin measurement in Argentina assessed in 300 representative laboratories with 20 fresh frozen single donation sera
  14. Practical application of Sigma Metrics QC procedures in clinical chemistry
  15. High-throughput analysis of testosterone in serum samples by on-line solid phase extraction liquid chromatography-tandem mass spectrometry
  16. Monitoring of mycophenolic acid and kidney function during combined immunosuppressive therapy
  17. Phenotype of Gc-globulin influences the macrophage activating factor (MAF) levels in serum
  18. Combination of cellular population data and CytoDiff™ analyses for the diagnosis of lymphocytosis
  19. Eosinophiluria is common among patients after ileal conduit surgery
  20. Soluble urokinase plasminogen activator receptor (suPAR) levels in healthy pregnancy and preeclampsia
  21. Uptake of cobalamin and markers of cobalamin status: a longitudinal study of healthy pregnant women
  22. Vitamin B12 absorption judged by measurement of holotranscobalamin, active vitamin B12: evaluation of a commercially available EIA kit
  23. Salivary markers of oxidative stress and Salivette collection systems
  24. Infectious Diseases
  25. Development of a microwell adapted immunoblot system with recombinant antigens for distinguishing human herpesvirus (HHV)6A and HHV6B and detection of human cytomegalovirus
  26. Cardiovascular Disease
  27. High-sensitive troponin T in chronic heart failure correlates with severity of symptoms, left ventricular dysfunction and prognosis independently from N-terminal pro-b-type natriuretic peptide
  28. Plasma symmetric dimethylarginine reference limits from the Framingham offspring cohort
  29. Letters to the Editor
  30. Iron deficiency prevalence is increased among Helicobacter pylori infected Brazilian patients
  31. Helicobacter pylori infection and iron deficiency in a group of Brazilian patients
  32. Multiplex protein assay performance/evaluation and the requirement for precision and correlation to clinical assays
  33. Functional assay or antigen test for protein C and protein S in ischemic stroke: which shows the greatest change?
Downloaded on 14.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2011.249/html
Scroll to top button