Automated genomic DNA extraction from saliva using the QIAxtractor
-
Henry Keijzer
Abstract
Background: Venipuncture is an invasive procedure to obtain whole blood in order to obtain high quality and sufficient amounts of genomic DNA. Obtaining DNA from non-invasive sources is preferred by patients, medical doctors and researchers. Saliva collected with cotton swabs (Salivette®) is increasingly being used to study chemical compounds, and it can also be a source of DNA. However, extracting DNA from Salivettes is very laborious and time consuming. Therefore, we developed a protocol for automated genomic DNA extraction from saliva collected in Salivette® using the QIAxtractor.
Methods: Saliva (0.1–2.0 mL) was collected by chewing on a Salivette® for 1–2 min. A total of 70 samples, collected from healthy volunteers, were extracted with the QIAxtractor robot and a Qiagen DX reagent pack. Quantity and quality was assessed using UV spectrometry and real-time polymerase chain reaction (PCR) (substitution at position –729 in the CYP1A2 gene).
Results: The average DNA concentration from the saliva samples was 6.0 μg/mL (95% CI 5.4–6.6 μg/mL). In 100% of the saliva samples, PCR products were detected with an average cycle threshold of 23.1 (95% CI 22.6–23.6).
Conclusions: DNA can be extracted in sufficient amounts from Salivette® with a fully automated system with a short turnaround time. Real-time PCR can be performed with these samples.
Clin Chem Lab Med 2010;48:641–3.
©2010 by Walter de Gruyter Berlin New York
Articles in the same Issue
- Editorials
- Laboratory medicine and natural disasters: are we ready for the challenge?
- Reference values: still a critical issue in laboratory medicine
- Review
- Contemporary platelet function testing
- Minireviews
- Disasters: where they find us
- Inter-laboratory variation in cerebrospinal fluid biomarkers for Alzheimer's disease: united we stand, divided we fall
- Opinion Paper
- Glycated hemoglobin (HbA1c): old dogmas, a new perspective?
- Guidelines and Recommendations
- IFCC reference procedures for measurement of the catalytic concentrations of enzymes: corrigendum, notes and useful advice
- Recommendations for the implementation of international standardization of glycated hemoglobin in Italy
- Genetics and Molecular Diagnostics
- UGT2B17 and SULT1A1 gene copy number variation (CNV) detection by LabChip microfluidic technology
- Genetic polymorphisms of CYP1A2, CYP3A4, CYP3A5, pregnane/steroid X receptor and constitutive androstane receptor in 207 healthy Spanish volunteers
- Automated genomic DNA extraction from saliva using the QIAxtractor
- General Clinical Chemistry and Laboratory Medicine
- A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometry
- Evaluation of the Becton-Dickinson rapid serum tube: does it provide a suitable alternative to lithium heparin plasma tubes?
- Immunoglobulin M antibodies to aquaporin-4 in neuromyelitis optica and related disorders
- Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids
- Extracellular haemoglobin, oxidative stress and quality of red blood cells relative to perioperative blood salvage
- Positive predictive value of serological diagnostic measures in celiac disease
- Retinol potentiates the inhibitory effect of ascorbic acid on uric acid assay
- Reference Values and Biological Variations
- Reference intervals for brain natriuretic peptide in healthy newborns and infants measured with an automated immunoassay platform
- Factors influencing serum total tryptase concentrations in a general adult population
- Leptin and adiponectin levels in pubertal children: relationship with anthropometric variables and body composition
- Cardiovascular risk factors and γ-glutamyltransferase fractions in healthy individuals
- Cancer Diagnostics
- Combined analysis of survivin autoantibody and carcinoembryonic antigen biomarkers for improved detection of colorectal cancer
- Oligoclonal bands and immunoglobulin isotype switch during monitoring of patients with multiple myeloma and autologous hematopoietic cell transplantation: a 16-year experience
- Letters to the Editor
- Methodological weakness in using correlation coefficients for assessing the interchangeability of analyte data between samples collected under different sampling conditions – the example of matrix metalloproteinase 9 determined in serum and plasma samples
- Intralipid causes falsely increased glucose concentrations with the Hemocue glucose analyzer
- Serum cholesterol: required and actual bias values
- Abstracts
- ISMD2010 Eighth International Symposium on Molecular Diagnostics
Articles in the same Issue
- Editorials
- Laboratory medicine and natural disasters: are we ready for the challenge?
- Reference values: still a critical issue in laboratory medicine
- Review
- Contemporary platelet function testing
- Minireviews
- Disasters: where they find us
- Inter-laboratory variation in cerebrospinal fluid biomarkers for Alzheimer's disease: united we stand, divided we fall
- Opinion Paper
- Glycated hemoglobin (HbA1c): old dogmas, a new perspective?
- Guidelines and Recommendations
- IFCC reference procedures for measurement of the catalytic concentrations of enzymes: corrigendum, notes and useful advice
- Recommendations for the implementation of international standardization of glycated hemoglobin in Italy
- Genetics and Molecular Diagnostics
- UGT2B17 and SULT1A1 gene copy number variation (CNV) detection by LabChip microfluidic technology
- Genetic polymorphisms of CYP1A2, CYP3A4, CYP3A5, pregnane/steroid X receptor and constitutive androstane receptor in 207 healthy Spanish volunteers
- Automated genomic DNA extraction from saliva using the QIAxtractor
- General Clinical Chemistry and Laboratory Medicine
- A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometry
- Evaluation of the Becton-Dickinson rapid serum tube: does it provide a suitable alternative to lithium heparin plasma tubes?
- Immunoglobulin M antibodies to aquaporin-4 in neuromyelitis optica and related disorders
- Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids
- Extracellular haemoglobin, oxidative stress and quality of red blood cells relative to perioperative blood salvage
- Positive predictive value of serological diagnostic measures in celiac disease
- Retinol potentiates the inhibitory effect of ascorbic acid on uric acid assay
- Reference Values and Biological Variations
- Reference intervals for brain natriuretic peptide in healthy newborns and infants measured with an automated immunoassay platform
- Factors influencing serum total tryptase concentrations in a general adult population
- Leptin and adiponectin levels in pubertal children: relationship with anthropometric variables and body composition
- Cardiovascular risk factors and γ-glutamyltransferase fractions in healthy individuals
- Cancer Diagnostics
- Combined analysis of survivin autoantibody and carcinoembryonic antigen biomarkers for improved detection of colorectal cancer
- Oligoclonal bands and immunoglobulin isotype switch during monitoring of patients with multiple myeloma and autologous hematopoietic cell transplantation: a 16-year experience
- Letters to the Editor
- Methodological weakness in using correlation coefficients for assessing the interchangeability of analyte data between samples collected under different sampling conditions – the example of matrix metalloproteinase 9 determined in serum and plasma samples
- Intralipid causes falsely increased glucose concentrations with the Hemocue glucose analyzer
- Serum cholesterol: required and actual bias values
- Abstracts
- ISMD2010 Eighth International Symposium on Molecular Diagnostics