Therapeutic drug monitoring and drugs of abuse testing on the cobas® 6000 analyzer series: analytical performance under routine-like conditions

  • Gunnar Brandhorst , Hilmar Luthe , Ingrid Domke , Christiane Knoke , Karl-Heinz Rhode , Heike Sauter and Michael Oellerich
Published/Copyright: June 4, 2009

Abstract

Background: The analytical performance of the clinical chemistry module c 501 (cobas® 6000 analyzer series) was evaluated for therapeutic drug monitoring and drugs of abuse testing using a spectrum of representative assays. Particular attention was paid to potential interactions between reagents using a simulated routine workload.

Methods: Within-run and total imprecision were assessed using a selection of representative reagents. Deviation from a consensus mean was tested using samples from a proficiency testing scheme. Method comparison using routine samples was carried out against the MODULAR ANALYTICS SWA and COBAS INTEGRA® 800 analysis systems.

Results: Total coefficients of variation (CV) ranged from 1.9% to 7.8% for individual drugs, and from 3.2% to 8.6% for drugs of abuse testing. Results from proficiency test samples were between 81% and 125% of the consensus mean for therapeutic drugs. Method comparisons (Passing-Bablok regression) showed overall good comparability to MODULAR ANALYTICS SWA and COBAS INTEGRA® 800 systems, with slopes from 0.93 to 1.17 and correlation coefficients r>0.98.

Imprecision in a simulated routine run was tested using a total of 42 methods (10 therapeutic drug monitoring, 9 drugs of abuse testing, 3 enzymes, 12 substrates, 8 specific protein assays). Imprecision in the reference batch run ranged from 0.7% to 5.0% CV for therapeutic drug monitoring assays, except for digoxin (DIG) (7.3%), and from 0.9% to 7.7% for drugs of abuse testing. The CVs of general clinical chemistry and specific protein tests were within the expected limits of 2% and 4%. CV changes in the simulated routine run were within the expected limits for most assays. Negative ΔCVs (≥2%) for DIG, digitoxin (DIGIT), cannabinoids (THC), and phencyclidine (PCP) may indicate improved performance when running these assays in a simulated routine operation. A positive ΔCV (≥3%) was found for amphetamines (AMPHs).

Conclusions: In conclusion, the cobas® c 501 module seems to be well-suited for routine use as consolidated workstation. Except for a potential interaction with AMPH, as indicated by the positive ΔCV, no significant interferences from different reagents could be observed during this study.

Clin Chem Lab Med 2009;47:854–9.


Corresponding author: Gunnar Brandhorst, University Hospital Goettingen, Department of Clinical Chemistry, 37075 Goettingen, Germany Phone: +49 551 39 8062, Fax: +49 551 39 12771,

Received: 2009-2-11
Accepted: 2009-4-6
Published Online: 2009-06-04
Published in Print: 2009-07-01

©2009 by Walter de Gruyter Berlin New York

Articles in the same Issue

  1. Editorials
  2. Change through continuity
  3. EFCC and Labs are Vital announce an Award for Outcomes Research in Laboratory Medicine
  4. Review
  5. The prevalence and risk factors for gallstone disease
  6. Opinion Paper
  7. Measurement uncertainty of test kit results – the ELISA example
  8. Genetics and Molecular Diagnostics
  9. Expression of RhoA and RhoC in colorectal carcinoma and its relations with clinicopathological parameters
  10. High-sensitive microarray substrates specifically designed to improve sensitivity for the identification of fetal paternally inherited sequences in maternal plasma
  11. A new CYP21A2 nonsense mutation causing severe 21-hydroxylase deficiency
  12. The haptoglobin phenotype is associated with the Epstein-Barr virus antibody titer
  13. General Clinical Chemistry and Laboratory Medicine
  14. Evaluation of analytical performance of the Pathfast® cardiac troponin I
  15. Diagnostic performance of the ARCHITECT C-Peptide immunoassay
  16. Clinical significance of anti-cyclic citrullinated peptide antibodies in Egyptian patients with chronic hepatitis C virus genotype IV infection
  17. Cerebrospinal fluid and serum uric acid levels in patients with multiple sclerosis
  18. Therapeutic drug monitoring and drugs of abuse testing on the cobas® 6000 analyzer series: analytical performance under routine-like conditions
  19. Is serum γ-glutamyltransferase an exposure marker of xenobiotics? Empirical evidence with polycylic aromatic hydrocarbon
  20. Reference Values and Biological Variations
  21. Gender difference and determinants of C-reactive protein level in Korean adults
  22. Cancer Diagnostics
  23. Quantitative telomerase activity in circulating human leukocytes: utility of real-time telomeric repeats amplification protocol (RQ-TRAP) in a clinical/epidemiological setting
  24. Intrathecal synthesis of tumor markers is a highly sensitive test in the diagnosis of leptomeningeal metastasis from solid cancers
  25. Letters to the Editor
  26. Hyperleukocytosis: pseudohyperkalaemia and other biochemical abnormalities in hyperleukocytosis
  27. Review of the prevalence of macroprolactinaemia in a South African hospital
  28. A novel aberrant form of e13a2 BCR-ABL1 transcript in chronic myelogenous leukemia undetectable with the standardized real-time quantitative polymerase chain reaction from the Europe Against Cancer Program
  29. Positive correlations between serum and plasma matrix metalloproteinase (MMP)-2 or MMP-9 levels in disease conditions
  30. Doubt on prevention of false-positive results of cardiac troponin I by recentrifugation
  31. Plasmalogens as a marker of elevated systemic oxidative stress in Parkinson's disease
Downloaded on 6.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/CCLM.2009.191/pdf
Scroll to top button