Home Parallel testing of 241 clinical nasopharyngeal swabs for the detection of SARS-CoV-2 virus on the Cepheid Xpert Xpress SARS-CoV-2 and the Roche cobas SARS-CoV-2 assays
Article Publicly Available

Parallel testing of 241 clinical nasopharyngeal swabs for the detection of SARS-CoV-2 virus on the Cepheid Xpert Xpress SARS-CoV-2 and the Roche cobas SARS-CoV-2 assays

  • Jason Wei Ming Tham , Shu Chi Ng , Chean Nee Chai , Siyu Png , Eunice Jia Min Tan , Li Jie Ng , Rui Ping Chua , Musa Sani , Donald Chiang , Ker Xin Tan , Nancy Wen Sim Tee , Roland Jureen , Shaun S. Tan , Gabriel Yan , Benedict Yan and Chun Kiat Lee ORCID logo EMAIL logo
Published/Copyright: November 19, 2020

To the Editor,

The World Health Organization (WHO) has declared the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) outbreak to be a global pandemic. As of the time of this writing, several thousand new confirmed COVID-19 cases have been reported worldwide [1]. To cope with the surging demand for timely diagnosis of COVID-19 cases, many SARS-CoV-2 molecular assays have become commercially available [2]. At our institution (National University Health System, Singapore), the Cepheid Xpert Xpress SARS-CoV-2 (herein referred as the Xpert; Cepheid, CA, USA) and the Roche cobas SARS-CoV-2 (herein referred as the cobas; Roche Molecular Systems, NJ, USA) assays are deployed interchangeably to meet the national testing demand for SARS-CoV-2. The Xpert assay targets the N2 region of the nucleoprotein (N) gene for specific SARS-CoV-2 detection and a conserved region of the structural protein envelope (E) gene for pan-sarbecovirus detection. Likewise, the cobas also targets a conserved region of the E gene for pan-sarbecovirus detection but amplifies the open reading frame 1ab (ORF1ab) nonstructural region for specific SARS-CoV-2 detection. We have had concerns before the integration of two different platforms for SARS-CoV-2 testing. First, it was unclear whether the test performance of the two assays would be comparable in our hands. Second, existing comparative studies on the Xpert and cobas assays relied on retesting of residual positive samples that were initially detected by the cobas assay [3], [4], [5], [6]. This may lead to the exclusion of positive samples near the detection limit of the Xpert assay, which the cobas assay may fail to detect. Third, several reports performed their analyses using only a small number of low positive samples which might not be sufficient to interrogate the detection limit of the assays [3], [4], [5], [6], [7].

To address the concerns, we assayed 241 fresh clinical nasopharyngeal (NP) swab samples in parallel on both the Xpert and the cobas assays to allow for a head-to-head comparative analysis. Ethics approval was granted by the National Healthcare Group Domain specific Review Board (NHG ROAM Reference Number 2020/00337). Statistical analyses were performed using the R statistical language, version 3.6.0, with p values of <0.05 denoting statistical significance. All samples were collected between June and July 2020, transported in universal transport medium (UTM; Copan Diagnostics Inc., Brescia, Italy), and tested within 24 h of collection. A summary of the demographic characteristics of the study samples is shown in Table 1.

Table 1:

Demographic characteristics of study samples.

CharacteristicsAll patientsWardsaEmergency departmentaDormitoriesb
AllSARS-CoV-2 (+)SARS-CoV-2 (−)AllSARS-CoV-2 (+)SARS-CoV-2 (−)AllSARS-CoV-2 (+)SARS-CoV-2 (−)AllSARS-CoV-2 (+)SARS-CoV-2 (−)
Cases
Frequency (%)241 (100%)64 (27%)177 (73%)81 (34%)4 (2%)77 (32%)100 (41%)8 (3%)92 (38%)60 (25%)52 (22%)8 (3%)
Age (year)
Median (IQRc)36.00 (28.00–59.00)32.50 (28.00–38.00)41.00 (28.00–70.00)46.00 (30.00–75.00)38.50 (35.25–41.50)49.00 (30.00–76.00)38.50 (28.00–59.50)32.00 (27.75–37.25)40.00 (28.00–62.25)32.50 (28.00–37.00)32.00 (28.00–36.50)34.38 (30.50–37.00)
Gender
Male (%)195 (81%)63 (26%)132 (55%)58 (24%)4 (2%)54 (22%)77 (32%)7 (3%)70 (29%)60 (25%)52 (22%)8 (3%)
Female (%)46 (19%)1 (0%)45 (19%)23 (10%)0 (0%)23 (10%)23 (10%)1 (0%)22 (9%)0 (0%)0 (0%)0 (0%)
  1. All samples were collected between 1 June and 24 July 2020. aSwab testing of symptomatic patients at the National University Hospital, National University Health System, Singapore. bMass swab testing of symptomatic and asymptomatic migrant workers living in the dormitories. cInterquartile range.

Following testing of 241 NP swabs, 58 tested positive and 172 tested negative with both assays, for an agreement of 95% (95% confidence interval (CI) 91.8–97.6%). The median cycle threshold (CT; Figure 1A) values for the E gene by the Xpert and cobas assays were 35.80 (interquartile range (IQR) 32.65–38.98) and 36.90 (IQR 35.57–37.86), respectively. The median CT values for the Xpert N and cobas ORF1ab gene targets are 38.30 (IQR 36.90–40.70) and 31.29 (IQR 23.85–33.22), respectively. For samples that were positive on both assays, 52/58 (90%) were detected with high E gene CT values ranging between 30.90 and 42.70 on either assay. Bland-Altman analysis (Figure 1B) showed a mean bias of −0.62 ± 2.87 (mean difference: −2.47; 95% limits of agreement: −6.26 to 5.01) on 45 samples that had E gene target detected on both assays. Interestingly, the analysis suggested that a significant proportion of our positive samples harbored SARS-CoV-2 at concentration near the assays’ detection limit, i.e., low viral titer, thus accounting for the large scattering observed exclusively at high mean E gene CT of >34.

Figure 1: (A) Dot and box plots of the pan-sarbecovirus E gene cycle threshold (CT) values obtained with the Cepheid Xpert Xpress SARS-CoV-2 and the Roche cobas SARS-CoV-2 assays for all positive samples. The median CT values by the Xpert and cobas assays were 35.80 and 36.90, respectively. Wilcoxon test showed no significant difference in the CT values obtained with both assays (p=0.44). (B) Bland–Altman analysis on 45 samples that had E gene target detected on both the Cepheid Xpert Xpress SARS-CoV-2 assay (method 1) and the Roche cobas SARS-CoV-2 test (method 2). Bias: −0.62; standard deviation of bias: 2.87; mean of differences/means: −2.47; (95% confidence interval (CI): −1.49 to 0.24); upper limit of agreement: 5.01 (95% CI 3.52–6.50); lower limit of agreement: −6.26 (95% CI −7.75 to −4.78). (C) A dot plot showing the CT values obtained with the Xpert assay (N and E genes) and the cobas assay (ORF1ab and E genes) for 58 low positive samples. Data points above CT value 45 (horizontal dotted line) are negative. The detection rate for the respective gene targets is indicated below the data points. The highest sensitivity for the low positive samples was the Xpert SARS-CoV-2-specific N gene (57/58; 98%) followed by the cobas pan-sarbecovirus E gene (51/58; 88%) and the Xpert pan-sarbecovirus E gene (45/58; 78%). The cobas SARS-CoV-2-specific ORF1ab gene was the least sensitive (9/58; 16%).
Figure 1:

(A) Dot and box plots of the pan-sarbecovirus E gene cycle threshold (CT) values obtained with the Cepheid Xpert Xpress SARS-CoV-2 and the Roche cobas SARS-CoV-2 assays for all positive samples. The median CT values by the Xpert and cobas assays were 35.80 and 36.90, respectively. Wilcoxon test showed no significant difference in the CT values obtained with both assays (p=0.44). (B) Bland–Altman analysis on 45 samples that had E gene target detected on both the Cepheid Xpert Xpress SARS-CoV-2 assay (method 1) and the Roche cobas SARS-CoV-2 test (method 2). Bias: −0.62; standard deviation of bias: 2.87; mean of differences/means: −2.47; (95% confidence interval (CI): −1.49 to 0.24); upper limit of agreement: 5.01 (95% CI 3.52–6.50); lower limit of agreement: −6.26 (95% CI −7.75 to −4.78). (C) A dot plot showing the CT values obtained with the Xpert assay (N and E genes) and the cobas assay (ORF1ab and E genes) for 58 low positive samples. Data points above CT value 45 (horizontal dotted line) are negative. The detection rate for the respective gene targets is indicated below the data points. The highest sensitivity for the low positive samples was the Xpert SARS-CoV-2-specific N gene (57/58; 98%) followed by the cobas pan-sarbecovirus E gene (51/58; 88%) and the Xpert pan-sarbecovirus E gene (45/58; 78%). The cobas SARS-CoV-2-specific ORF1ab gene was the least sensitive (9/58; 16%).

A total of 11 discordant samples were noted where 10 were detected by the Xpert assay only (median E gene CT: 38.50 [IQR 38.03–39.38]; median N gene CT: 40.60 [IQR 39.20–41.60]) and one was detected by the cobas assay only (E gene CT: 38.66; ORF1ab CT: not detected). To adjudicate discrepant results, we retrieved each of the 11 samples for automated nucleic acid extraction on the KingFisher Flex instrument (Thermo Fisher Scientific, Massachusetts, USA) and retested them using the TaqPath COVID-19 RT-PCR assay (herein referred as the TaqPath; Thermo Fisher Scientific) on the Applied Biosystems 7500 Fast Real-Time PCR Instrument (Thermo Fisher Scientific), according to the manufacturer’s instructions. The TaqPath assay targets three different regions (ORF1ab, N, and Spike [S] genes) of the SARS-CoV-2 genome and the claimed limit of detection is 10 genomic copy equivalents per reaction with 40 cycles of amplification. The consensus result was defined as the result obtained by at least two of the three assays. After retesting the 11 discordant samples using the TaqPath assay, six of the 10 Xpert-only positive samples (median E gene CT: 38.40 [IQR 37.20–38.60]; median N gene CT: 39.50 [IQR: 38.45–40.40]) agreed with the TaqPath results and were deemed as positives based on consensus. On the other hand, four of the 10 Xpert-only positive samples (median E gene CT: 39.90 [IQR 39.10–41.55]; median N gene CT: 41.60 [IQR 41.60–42.10]) disagreed with the TaqPath results and were reclassified as negatives. Likewise, the single cobas-only positive sample also disagreed with the TaqPath result and was subsequently reclassified as negative. Compared to the consensus results, the Xpert assay had a kappa coefficient of 0.96 (95% CI 0.92–1.00; almost perfect agreement), positive percent agreement (PPA) of 100% (95% CI 94–100%), and negative percent agreement (NPA) of 98% (95% CI 94–99%). For the cobas assay, the kappa coefficient was 0.92 (95% CI 0.87–0.98; almost perfect agreement), PPA was 91% (95% CI 81–96%), and NPA was 99% (95% CI 97–100%).

In summary, we assayed 241 fresh NP swabs in parallel across the Cepheid Xpert Xpress SARS-CoV-2 and the Roche cobas SARS-CoV-2 assays. In our hands, there was 95% agreement between the assays for both negative and positive clinical samples, suggesting comparable diagnostic performance in the detection of SARS-CoV-2 virus from NP swabs. To the best of our knowledge, this is the third head-to-head comparison data generated via parallel testing of samples on the two automated platforms [6], [, 7]. The present work challenged the assays’ detection limit using 58 clinical samples with low viral titers. Figure 1C depicts the test performance of the different gene targets. The Xpert SARS-CoV-2-specific N gene is the most sensitive, detecting 98% (57/58) of the low positives while the Xpert pan-sarbecovirus E gene detects 78% (45/58) of the low positives. On the contrary, the cobas SARS-CoV-2-specific ORF1ab gene exhibits inferior sensitivity, detecting only 16% (9/58) of the low positives. However, the poor diagnostic yield is compensated by the cobas pan-sarbecovirus E gene which detects 88% (51/58) of the low positive samples. Our data corroborates with previous studies that the Xpert and cobas pan-sarbecovirus E gene targets performed similarly [3], [4], [5], [6], [7]. For the SARS-CoV-2-specific N and ORF1ab gene targets, a significant difference in the positive identification rate was observed using the McNemar’s test (p value < 0.05). Notably, the cobas ORF1ab gene has performed poorly when compared to the Xpert N gene which corroborated with Lowe’s data [3]. Altogether, our results show that the Xpert assay has higher sensitivity compared to the cobas assay, with the former detecting six additional cases among the low positives (6/58; 10%). It is unknown whether high CT values (low viral load) has significant diagnostic value due to the lack of correlation with culture results, variables associated with sampling site, and collection methodology. Further studies are required to determine the diagnostic value of using assays with higher analytic sensitivity. However, the authors suggest that retesting of the samples using the Xpert assay should be considered when the SARS-CoV-2 results do not correlate with the patients’ clinical presentations. Considering our findings, we believe that both assays can be deployed interchangeably in experienced molecular diagnostic laboratories for routine diagnostics of COVID-19 and the detection of low positive SARS-CoV-2 samples.


Corresponding author: Chun Kiat Lee,Department of Laboratory Medicine, Molecular Diagnosis Centre, National University Health System, 5, Lower Kent Ridge Road, 119074, Singapore, Singapore, Phone: +65 6772 4175, Fax: +65 6772 4407, E-mail:

  1. Research funding: None declared.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Ethical approval: Ethics approval was granted by the National Healthcare Group Domain specific Review Board (NHG ROAM Reference number 2020/00337).

References

1. World Health Organization. Coronavirus disease (COVID-19) pandemic; 2020. Available from: https://www.who.int/emergencies/diseases/novel-coronavirus-2019 [Accessed 19 Aug 2020].Search in Google Scholar

2. Zhen, W, Manji, R, Smith, E, Berry, GJ. Comparison of four molecular in vitro diagnostic assays for the detection of SARS-CoV-2 in nasopharyngeal specimens. J Clin Microbiol 2020;58:e00743–20, https://doi.org/10.1128/jcm.00743-20.Search in Google Scholar PubMed PubMed Central

3. Lowe, CF, Matic, N, Ritchie, G, Lawson, T, Stefanovic, A, Champagne, S, et al.. Detection of low levels of SARS-CoV-2 RNA from nasopharyngeal swabs using three commercial molecular assays. J Clin Virol 2020;128:104387, https://doi.org/10.1016/j.jcv.2020.104387.Search in Google Scholar PubMed PubMed Central

4. Smithgall, MC, Scherberkova, I, Whittier, S, Green, DA. Comparison of Cepheid Xpert Xpress and Abbott ID Now to Roche cobas for the rapid detection of SARS-CoV-2. J Clin Virol 2020;128:104428, https://doi.org/10.1016/j.jcv.2020.104428.Search in Google Scholar PubMed PubMed Central

5. Broder, K, Babiker, A, Myers, C, White, T, Jones, H, Cardella, J, et al.. Test agreement between Roche Cobas 6800 and Cepheid GeneXpert Xpress SARS-CoV-2 assays at high cycle threshold ranges. J Clin Microbiol 2020;58:e01187–2, https://doi.org/10.1128/jcm.01187-20.Search in Google Scholar

6. Lieberman, JA, Pepper, G, Naccache, SN, Huang, ML, Jerome, KR, Greninger, AL. Comparison of commercially available and laboratory developed assays for in vitro detection of SARS-CoV-2 in clinical laboratories. J Clin Microbiol 2020;58:e00821–20, https://doi.org/10.1128/jcm.00821-20.Search in Google Scholar PubMed PubMed Central

7. Moran, A, Beavis, KG, Matushek, SM, Ciaglia, C, Francois, N, Tesic, V, et al.. Detection of SARS-CoV-2 by use of the Cepheid Xpert Xpress SARS-CoV-2 and Roche cobas SARS-CoV-2 assays. J Clin Microbiol 2020;58:e00772–20, https://doi.org/10.1128/jcm.00772-20.Search in Google Scholar

Received: 2020-09-02
Accepted: 2020-11-08
Published Online: 2020-11-19
Published in Print: 2021-02-23

© 2020 Jason Wei Ming Tham et al., published by De Gruyter, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. Assuring the quality of examinations using faecal immunochemical tests for haemoglobin (FIT)
  4. Review
  5. Capillary electrophoresis based on nucleic acid analysis for diagnosing inherited diseases
  6. Mini Reviews
  7. Negative hair test result after long-term drug use. About a case involving morphine and literature review
  8. Prorenin and active renin levels in paediatrics: a bioanalytical review
  9. Opinion Paper
  10. From therapeutic drug monitoring to total drug monitoring and drug-omics
  11. Perspectives
  12. The internal quality control in the traceability era
  13. Genetics and Molecular Diagnostics
  14. External quality assessment (EQA) and alternative assessment procedures (AAPs) in molecular diagnostics: findings of an international survey
  15. General Clinical Chemistry and Laboratory Medicine
  16. An evaluation of ten external quality assurance scheme (EQAS) materials for the faecal immunochemical test (FIT) for haemoglobin
  17. Optimizing hepcidin measurement with a proficiency test framework and standardization improvement
  18. Development of a certified reference material for anti-β2-glycoprotein I IgG – commutability studies
  19. Influence of patients’ clinical features at intensive care unit admission on performance of cell cycle arrest biomarkers in predicting acute kidney injury
  20. Efficacy of weekly administration of cholecalciferol on parathyroid hormone in stable kidney-transplanted patients with CKD stage 1–3
  21. Plasma metanephrines and prospective prediction of tumor location, size and mutation type in patients with pheochromocytoma and paraganglioma
  22. Emicizumab, the factor VIII mimetic bi-specific monoclonal antibody and its measurement in plasma
  23. Reference Values and Biological Variations
  24. Age appropriate reference intervals for eight kidney function and injury markers in infants, children and adolescents
  25. Cardiovascular Diseases
  26. Comparison of acetylsalicylic acid and clopidogrel non-responsiveness assessed by light transmittance aggregometry and PFA-100® in patients undergoing neuroendovascular procedures
  27. Trimethylamine-N-oxide (TMAO) predicts short- and long-term mortality and poor neurological outcome in out-of-hospital cardiac arrest patients
  28. Non-lactate strong ion difference and cardiovascular, cancer and all-cause mortality
  29. Infectious Diseases
  30. Performance of three automated SARS-CoV-2 antibody assays and relevance of orthogonal testing algorithms
  31. Development, evaluation, and validation of machine learning models for COVID-19 detection based on routine blood tests
  32. Searching for a role of procalcitonin determination in COVID-19: a study on a selected cohort of hospitalized patients
  33. Association of kidney function with effectiveness of procalcitonin-guided antibiotic treatment: a patient-level meta-analysis from randomized controlled trials
  34. Erratum
  35. Corrigendum to: The role of hyperhomocysteinemia as well as folate, vitamin B6 and B12 deficiencies in osteoporosis – a systematic review
  36. Letter to the Editors
  37. Harmonization of antineutrophil cytoplasmic antibodies (ANCA) testing by reporting test result-specific likelihood ratios: position paper
  38. Independent internal quality control (IQC) for faecal immunochemical tests (FIT) for haemoglobin: use of FIT manufacturers’ IQC for other FIT systems
  39. Parallel testing of 241 clinical nasopharyngeal swabs for the detection of SARS-CoV-2 virus on the Cepheid Xpert Xpress SARS-CoV-2 and the Roche cobas SARS-CoV-2 assays
  40. Sustained SARS-CoV-2 nucleocapsid antibody levels in nonsevere COVID-19: a population-based study
  41. Evidence for increased circulating procoagulant phospholipids in patients with COVID-19 pneumonia and their prognostic role
  42. Complete blood counts and cell population data from Sysmex XN analyser in the detection of SARS-CoV-2 infection
  43. Macro creatine kinase in an asymptomatic patient: a case report
  44. Implementation of pre-labelled barcode tubes and the GeT-System in a general hospital for the exact documentation of the time of venous blood sample and improvement of sample quality
  45. Determination of free light chains: assay-dependent differences in interpretation
  46. Association between SOX17, Wif-1 and RASSF1A promoter methylation status and response to chemotherapy in patients with metastatic gastric cancer
  47. Effect of two organizational interventions on the frequency of haemoglobin A1c and erythrocyte sedimentation rate testing
  48. Haemoglobin A1c determination from dried blood spots prepared with HemaSpot™ blood collection devices: comparison with fresh capillary blood
  49. Stability of catecholamines in whole blood: influence of time between collection and centrifugation
Downloaded on 28.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2020-1338/html
Scroll to top button