Home Evaluation of different needle gauge blood collection sets (23G/25G) in aged patients
Article Open Access

Evaluation of different needle gauge blood collection sets (23G/25G) in aged patients

  • Adrian Rosada , Katharina Friedrich , Eileen Spieß , Monika Prpic , Ursula Müller-Werdan , Amir Jahić and Kai Kappert ORCID logo EMAIL logo
Published/Copyright: December 9, 2024
Become an author with De Gruyter Brill

Abstract

Objectives

Aged patients are often characterized by difficult blood sampling conditions. Smaller needle gauge (G) may be beneficial for venous access and reduced pain perception, however, potentially at the expense of lower blood quality for laboratory measurements. We systematically compared two blood collection sets with different outer but equal inner diameters; different needle tips, and retract mechanisms in aged patients (Safety-Lok™, 23G, SL vs. UltraTouch™ Push Button, 25G, UT-PB) regarding clinical aspects and laboratory measurements.

Methods

Clinical examination and questionnaires were used in an aged cohort (n=161, average age=81.6 years), to determine characteristics of venipuncture, the phlebotomist’s assessment of blood draw including level of difficulty, and patient’s pain perception with either one or both blood collection sets. Sample quality was evaluated by laboratory analytics considering 13 parameters.

Results

SL, UT-PB, or both were used in 89 (55 %), 72 (45 %) or 36 (22 %) patients. The handling of the blood collection sets was perceived slightly easier for UT-PB compared to SL by the phlebotomist (−30 %, p=0.038). There was no significant difference in other parameters of the phlebotomist’s assessment or patients’ perception of blood collection. There was no clinically relevant difference between both sets in any of the laboratory measurements, including potassium and hemolysis index.

Conclusions

Clinical use of the UT-PB compared with SL in aged patients was associated with slight advantages of UT-PB, e.g. in the handling comfort for the phlebotomist. Sample quality, especially regarding hemolysis, was identical between both blood collecting sets, making its use uncritical in difficult venous conditions commonly seen in elderly patients.

Introduction

Laboratory blood examination is one of the most common diagnostic procedures in clinical medicine. The common way to obtain blood is via venous access with a needle connected to a blood collection set, often by single-use plastic tubes with straight or butterfly needles [1]. Although blood collection with a straight needle is gold standard due to lower risk for hemolysis and better sample quality, the introduction of butterfly needles led to widespread adoption in pediatrics and geriatrics due to greater convenience (better visibility of whether a blood vessel has been struck and the greater freedom of movement of the blood collection set during blood draw) [1].

Difficulties in blood draw can be challenging due to difficult vein condition or non-compliance with blood collection procedure. This is frequently challenging especially in aged patients due to thin or aged veins and frequent cognitive limitations [2]. In addition, difficulties in blood draws quite often result in underfilling of blood sampling tubes. Underfilling the collection tubes increases the risk of hemolysis, which consequently elevates activity levels of lactate dehydrogenase (LD) [3], [4], [5], [6], and hemolysis is one of the major preanalytical errors in laboratory medicine [7]. Since LD is used for diagnostic as well as prognostic monitoring in aged patients [8], 9], this may lead to massive clinical misinterpretations [4].

To ensure a successful and painless blood collection, the needle should not be moved within the vein during blood draw. Especially in cases of multiple tube collection, this can be challenging. Furthermore, at the end of the blood sampling procedure, the tourniquet must be released, the needle carefully withdrawn and at the same time the puncture site covered with gauze. Occasionally, an unintentional premature pull of the needle may occur causing hematoma and pain.

Smaller needle size (measured in gauge, G) can be used to reduce the pain of the procedure [10], 11]. In a detailed trial of Arendt-Nielsen et al. with 30 healthy volunteers who were exposed to automated and standardized needle insertions (40–48 insertions per person) with 27G, 28G, 30G, 31G, and 32G needles, they found a significant and positive correlation between outer needle diameter and frequency of insertion pain [12].

For adults 21G needle size is commonly used, whereas 23G is used for aged patients, because of frequent poorer vein conditions. One standardly used regular 23G blood collection set is the Safety-Lok™ (SL) blood collection set with a standard thin-wall cannula and an inner diameter of 0.015 inches. Smaller needle sizes, like 25G needles, however, might impact blood quality as an important preanalytical requirement for laboratory measurements. Earlier studies reported that 25G needles are associated with increased hemolysis [13].

To facilitate easier blood collection especially regarding the above-mentioned needle draw challenges, a 25G blood sample set called UltraTouch™ Push Button (UT-PB) has been introduced. This set combines an ultra-thin wall needle, which reduces the cannula wall thickness and increases the inner diameter of the cannula to 0.016 inches (RightGauge™), with a specific cut needle tip (PentaPoint™) and automated needle retracing via button (Push Button). An increased rate of successful venipunctures with complex patient populations while statistically significantly reducing insertion force and overall pain during blood collection, without compromising efficiency and blood quality, was shown, potentially alleviating issues of small needle sizes associated with hemolysis [14].

Previous studies suggested that the specific needle tip geometry has impact on needed penetration force and therefore patients’ pain perception in non-blood taking aspects [15], [16], [17]. Comparing the specific 5-bevel design, which is used in the UT-PB set, and a common 3-bevel design, such as in the SL set, the 5-bevel design is supposed to significantly reduce needle insertion force compared to the common design [18].

While direct comparisons of different blood draw sets have previously been carried out for specific cohorts such as healthy volunteers [14], emergency room patients [19], cancer patients [20], and children [21], here we systematically focused on aged patients.

Materials and methods

Patients

A total of 161 inpatients with an age >60.0 years (range: 61.9–99.3 years with a mean of 81.6 years), treated in the Department of Geriatrics and Medical Gerontology at the Charité–Universitätsmedizin Berlin between July 2020 and May 2022, were enrolled prospectively. All patients reported informed consent. Blood sampling was performed in patients treated under routine clinical conditions. 89 patients underwent their regular blood collection performed with the SL collection set, 72 patients with the UT-PB set. Thirty-six patients underwent blood collection with both sets and on different days (Figure 1). In some cases, multiple rounds of blood collection were performed at different days leading to the final sample size as reported in Table 1.

Figure 1: 
Flow chart with step-by-step process of the study. Graph depicts the step-by-step process of the study after inclusion.
Figure 1:

Flow chart with step-by-step process of the study. Graph depicts the step-by-step process of the study after inclusion.

Table 1:

Assessment of the phlebotomist.

Total cohort Safety-Lok (n=120) UltraTouch (n=120) Difference, % p-Value
Number of blood sample tubes 3.16 ± 0.5 3.29 ± 0.7 4.22 0.05
All planned tubes filled, % 99.17 ± 9.1 100.00 ± 0 0.84 0.32
All tubes filled completely, % 98.33 ± 12.9 97.50 ± 15.7 −0.85 0.65
Blood flow dry up, % 5.83 ± 23.5 5.00 ± 21.9 −14.29 0.78
Need to reposition the cannula to restore blood flow, % 8.33 ± 27.8 5.83 ± 23.5 −30.00 0.45
Immediate reddening (reaction of the skin), % 4.17 ± 20.1 6.67 ± 25.0 60.00 0.39
Hematoma at the injection site after 24 h or 48 h, % 30.83 ± 46.4 23.33 ± 42.5 −24.32 0.19
Size of hematoma, mma 19.30 ± 14.3 15.93 ± 13.9 −17.46 0.30
Severity of blood collection with the blood collection set in place with regard to finding veins and filling tubesc 1.38 ± 1.7 0.96 ± 1.4 −30.30 0.04

Sub-cohort Safety-Lok (n=38) UltraTouch (n=36) Difference, % p-Value

Number of blood sample tubes 3.16 ± 0.5 3.33 ± 0.7 5.56 0.13
All planned tubes filled, % 100.00 ± 0 100.00 ± 0 0 >0.99
All tubes filled completely, % 100.00 ± 0 100.00 ± 0 0 >0.99
Blood flow dry up, % 0.00 ± 0 0.00 ± 0 0 >0.99
Need to reposition the cannula to restore blood flow, % 0.00 ± 0 0.00 ± 0 0 >0.99
Immediate reddening (reaction of the skin), % 0.00 ± 0 5.56 ± 23.2 5.56 0.23
Hematoma at the injection site after 24 h or 48 h, % 26.32 ± 44.6 36.11 ± 48.7 37.22 0.45
Size of hematoma, mmb 17.70 ± 13.9 18.46 ± 16.3 4.30 >0.99
Severity of blood collection with the blood collection set in place with regard to finding veins and filling tubesc 1.26 ± 1.6 0.83 ± 1.4 −34.03 0.20
  1. If not indicated differently, values represent mean ± SD, relative differences in percent, and p-value of Mann–Whitney test for statistical testing. Boolean data is presented as mean ± SD, in percent (%). an=37 (Safety-Lok) and n=28 (UltraTouch). bn=10 (Safety-Lok) and n=13 (UltraTouch). c0=none to 10=biggest conceivable.

Ethic approval

The study complied with all relevant national regulations, institutional policies and was performed in accordance with the Helsinki Declaration as revised in 2013 and approved by the local Ethics Committee of the Charité–Universitätsmedizin Berlin (EA2/231/20).

Materials

Two different blood collection sets were used: The 25G; 19 mm needle set UltraTouch™ Push Button (by Becton, Dickinson and Company, #367391) and the 23G; 19 mm (0.75 inch) needle set Safety-Lok™ (by Becton, Dickinson and Company, #367284), both with 178 mm (7inch) tubing length and an estimated maximum dead volume of 0.3 mL. The inner diameter of UT-PB was 0.016 inches and of the SL was 0.015 inches. For blood collection heparin plasma vacuum tubes from Greiner Bio-One GmbH were used (LH lithium heparin sep, #456083). Randomization was carried out by taking one of the two blood collection systems from a non-transparent box in which the systems were mixed in a 1:1 ratio.

All phlebotomies were conducted by the same physician with no technical differences between the individual experimental phlebotomies of the same blood collection set. Some patients had blood drawn with both sets at different days and were analyzed in a sub-cohort analysis.

Blood sampling and other preanalytical conditions

Blood was preferably taken from an easily accessible vein (antecubital, other locations or hand veins). At least one heparin plasma tube was collected during the same venipuncture from each patient. A dummy tube was taken prior to tubes used for laboratory analyses (99.2 % of cases). Further tubes (order according the EFLM guideline) were collected respective to individual patient treatment and required blood laboratory parameters. The vacuum tubes were filled to the predetermined volume. Tubes were inverted as recommended and thereafter left at room temperature prior to transportation to the analyzing laboratory, where samples were centrifuged at 3,000×g for 10 min at 20 °C before measuring blood analytes immediately.

Patients’/phlebotomist’s questionnaire

Directly after the blood collection, the patients were interviewed by the phlebotomist with a standardized questionnaire about their subjective pain perception, fear of the next blood sample and general perception of the blood collection measured with a visual analogue scale for pain, a visual analogue scale for anxiety and a visual analogue scale for general perception, all 10-point scales (Supplementary Material questionnaire).

Further, the assessment of the procedure by the phlebotomist’s was done by a structured self-assessment, a questionnaire with the Boolean parameters blood flow dry up, need to reposition the cannula to restore blood flow, immediate reddening of the skin, complete sampling of all planned blood tubes, and complete filling of each blood tube (Supplementary Material questionnaire). In addition, a subjective perception of the severity of the procedure with the used set was assessed on a 10-point scale (0=no difficulty to 10=greatest difficulty imaginable).

Each patient was re-examined 24 h after the blood taking regarding appearance of new hematoma at the blood taking site (Boolean parameter) and, if so, the size of the hematoma in millimeters.

Routine laboratory blood analysis

Thirteen parameters were analyzed per tube under routine conditions in an accredited laboratory (Labor Berlin–Charité Vivantes GmbH): sodium, potassium, chloride, glutamic oxaloacetic transaminase (GOT), lactate dehydrogenase activity (LD), creatinine kinase (CK), creatinine, inorganic phosphorus (PO4), urea, C-reactive protein (CRP), and indices for hemolysis, icterus, and lipemia. Sodium, potassium, and chloride were determined by an indirect ion selective electrode (Roche cobas ISE indirect Na-K-Cl for Gen.2) on a Roche cobas 6000 ISE module. GOT was measured by Roche cobas Aspartate Aminotransferase acc. to IFCC with pyridoxal phosphate activation (ASTLP, #04467493190), LD activity by Roche cobas Lactate Dehydrogenase acc. to IFCC ver.2 (LDHI2, #03004732122), CK by Roche cobas Creatine Kinase (CK, #07190794190), creatinine by Roche cobas Creatinine Jaffé Gen.2 (CREJ2, #04810716190), PO4 by Roche cobas Phosphate (Inorganic) ver.2 (PHOS2, #03183793122), urea by Roche cobas Urea/BUN (UREAL, #04460715190), and CRP by Roche cobas C-Reactive Protein Gen.3 (CRPL3, #04956842190) on Roche cobas c501/502 modules within a cobas 6,000 analyzer configuration. Proprietary check of specimen quality was determined by means of serum indices. Estimated glomerular filtration rate (eGFR) was calculated as described [22].

Statistics

Numerical data is expressed as mean ± standard deviation (SD). Boolean data, with no=0 and yes=1, is expressed as percentage ± SD Data was tested for distribution normality using Kolmogorov-Smirnov test with a significance level of p=0.05 (Supplementary Material Tables 1–4). For assessment of statistical significance Mann–Whitney tests were performed, where a p-value of <0.05 was regarded as significant. For assessment of correlation of data sets (not normal distributed) non-parametric Spearman correlation coefficient r and two-tailed p-value was computed (Supplementary Material Table 1). Statistical analysis was performed using GraphPad Prism 10.

Results

Baseline characteristics

In total n=161 patients were included in the study. Their mean age was 81.6 ± 6.8 years 59.6 % of patients were females and 40.4 % of patients were males. The study cohort included aged inpatients with blood draw always performed by the same phlebotomist. The indication for hospital admission varied, the most common were falls with or without bone fracture, infections of the pulmonary or urinary tract and congestive heart failure (data not shown). Based on laboratory data, n=107 (66.5 %) patients were characterized with anemia and n=8 (5 %) patients with hyperkalemia.

Blood collection

SL or UT-PB were used in 89 (55 %) and 72 (45 %) patients, respectively. For particular patients, multiple blood collections were performed, in sum resulting in 120 blood collections per set (Figure 1). In 36 (22 %) patients both blood collection sets were used. As a particularly interesting group, which provided intraindividual experience with both sets, specific analysis were performed with this sub-cohort.

Patients’ pain perception or fear of blood draw was not associated with age or sex of the patients nor with blood draw area (antecubital, hand or other vein location), as indicated by Spearman’s correlation coefficients r <0.4 for each correlation analysis (Supplementary Material Table 1).

The antecubital vein was the most used vein access in both groups (SL and UT-PB), followed by other sites (cephalic vein) and by hand vein with no significant difference of the vein access site between both sets, resulting in no significant difference in blood draw location between both groups (Mann–Whitney test, p=0.30, data not shown).

Assessment of phlebotomist

From the phlebotomist’s assessment of the total cohort as well as the cohort with both needle sets (sub-cohort) it was evident, that all intended blood tubes (three tubes were collected for both sets) could be successfully collected with both blood collection sets. Further, there were no significant differences in the completeness of their fill levels (Figure 2A and B, Table 1). In the total cohort (Figure 2A) applying the UT-PB blood collection set appeared to be 30 % easier (p=0.04) in locating veins and filling tubes. Nonetheless, both sets showed no significant difference regarding further procedural complications like dry up of blood flow or the need to reposition the cannula to restore blood flow during blood collection. However, the UT-PB set showed a trend toward less dry up (−14 %, p=0.78) and less repositioning (−30 %, p=0.45). Further, both sets showed no significant differences in the immediate skin reaction such as redness, appearance, or size of the hematoma after blood collection. Again, the UT-PB set tended to produce fewer (−24 %, p=0.19) and smaller hematoma (−17.5 %, p=0.30), but more immediate skin redness (60 %, p=0.39).

Figure 2: 
Assessment of the phlebotomist. Graph depicts the difference between the two blood collection sets from total cohort (A) and sub-cohort (B) as relative change of UltraTouch set from Safety-Lok set in percent (red bars). Error bars represent standard deviation (SD). Numbers represent p-values of Mann–Whitney test.
Figure 2:

Assessment of the phlebotomist. Graph depicts the difference between the two blood collection sets from total cohort (A) and sub-cohort (B) as relative change of UltraTouch set from Safety-Lok set in percent (red bars). Error bars represent standard deviation (SD). Numbers represent p-values of Mann–Whitney test.

In the sub-cohort, where both blood collection sets were applied (Figure 2B), there were no significant differences in the assessed parameters, with general improvement in blood collection (34 %, p=0.20), increase in skin reddening (6 %, p=0.23), and increase in hematoma appearance (37 %, p=0.45) as well as size (4 %, p>0.99) in those subjected to the UT-PB.

Subjective perception of patients

The assessment of the total cohort as well as the sub-cohort revealed no significant differences between the two sets used in regard to pain perception, fear, or the general assessment of the procedure (Figure 3A and B, Table 2). Nevertheless, there was a trend towards less pain perception using the UT-PB set (−16 %, p=0.30 (total cohort, Figure 3A); −30 %, p=0.39 (sub-cohort with both sets, Figure 3B)), whereas the blood collection was assessed as less pleasant with the UT-PB set (+16 %, p=0.77 (total cohort, Figure 3A); +25 %, p=0.8 (cohort with both, Figure 3B)).

Figure 3: 
Subjective perception of pain. Graph depicts the difference between the two blood collection sets of total cohort (A) and sub-cohort (B) as relative change of UltraTouch set from Safety-Lok set in percent (red bars). Error bars represent standard deviation (SD). Numbers represent p-values of Mann–Whitney test. *Positive values represent an increase, negative values represent a decrease in pain and fear, respectively. **Positive values represent a decrease, negative values represent an increase in pleasantness, respectively.
Figure 3:

Subjective perception of pain. Graph depicts the difference between the two blood collection sets of total cohort (A) and sub-cohort (B) as relative change of UltraTouch set from Safety-Lok set in percent (red bars). Error bars represent standard deviation (SD). Numbers represent p-values of Mann–Whitney test. *Positive values represent an increase, negative values represent a decrease in pain and fear, respectively. **Positive values represent a decrease, negative values represent an increase in pleasantness, respectively.

Table 2:

Subjective patient’s perception.

Total cohort Safety-Lok (n=120) UltraTouch (n=120) Difference, % p-Value
Sensation of pain during blood samplinga 1.35 ± 1.7 1.13 ± 1.6 −16.05 0.30
Fear of the next blood samplea 0.65 ± 1.4 0.68 ± 1.7 2.53 0.35
General assessment of the blood collectionb 1.03 ± 1.4 1.20 ± 1.7 16.13 0.77

Sub-cohort Safety-Lok (n=38) UltraTouch (n=36) Difference, % p-Value

Sensation of pain during blood samplinga 1.26 ± 1.4 0.89 ± 1.1 −29.63 0.39
Fear of the next blood samplea 0.47 ± 1.0 0.31 ± 0.9 −35.49 0.33
General assessment of the blood collectionb 0.87 ± 1.1 1.08 ± 1.6 24.75 0.80
  1. Shown values represent mean ± SD, relative differences in percent, and p-value of Mann–Whitney test for statistical testing. a0=none to 10=maximum; b0=pleasant to 10=unpleasant.

Analysis of blood samples

Concerning laboratory routine analytics, we analyzed in detail the quality (by serum indices) as well as hemolysis-susceptible and hemolysis-non-susceptible analytes to address potential differences of blood draw sets. Measurements of 13 laboratory parameters showed trends but no clinically relevant differences between the two blood collection sets, neither for hemolysis-susceptible nor for non-hemolysis-susceptible analytes in the total cohort (Table 3). Differences in potassium (−2.9 %) and icteric index (−4.1 %) were detectable, but very small and clinically not relevant. For the sub-cohort with blood draw using both needle sets, we focused only on hemolysis-sensitive analytes, since blood draws were performed on different days. We detected equivalent changes–with hemolysis index (HI) exhibiting the only valid parameter for direct comparisons–for values in individuals being directly evaluated by both blood collection sets. Since an HI above 50 is considered the upper limit of this non-analytical quality indicator in the IFCC program [23], we analyzed our cohorts and were able to determine that only two samples from the SL group had an HI greater than 50. Notably, in the sub-cohort the HI was reduced by 47 % (p=0.57) in UT-PB blood samples, comparable to the finding in the total cohort (−28 %, p=0.76) [23].

Table 3:

Laboratory blood analysis.

Total cohort Safety-Lok n UltraTouch n difference, % p-Value
Na+, mmol/L 139.31 ± 3.6 112 139.16 ± 3.9 120 −0.11 0.08
K+, mmol/L 3.96 ± 0.4 112 3.85 ± 0.5 120 −2.86 0.002
Cl, mmol/L 102.12 ± 4.3 107 101.86 ± 4.6 117 −0.25 0.32
GOT/AST, U/L 31.08 ± 27.8 109 28.16 ± 15.0 119 −9.40 0.22
LD, U/L 252.24 ± 86.3 107 257.02 ± 90.8 117 1.89 0.41
CK, U/L 93.96 ± 319.1 107 99.58 ± 322.2 117 5.98 0.82
Creatinine, mg/dL 1.14 ± 0.9 114 1.37 ± 1.3 120 20.19 0.58
Phosphorus (inorg.), mmol/L 1.05 ± 0.2 107 1.04 ± 0.2 117 −0.36 0.61
Urea, mg/dL 48.55 ± 31.3 113 46.58 ± 27.9 120 −4.07 0.83
CRP, mg/dL 22.58 ± 33.2 109 22.79 ± 31.6 118 0.93 0.72
eGFR after CKD-EPI 62.92 ± 23.9 114 58.14 ± 24.7 120 −7.60 0.20
Icteric index (a.u.) 0.88 ± 0.9 115 0.84 ± 0.6 120 −4.17 0.01
Hemolysis index, mg/dL 8.60 ± 15.0 115 6.16 ± 4.3 120 −28.39 0.76
Lipemia index (a.u.) 12.57 ± 7.8 115 12.64 ± 7.7 120 0.54 0.97

Sub-cohort Safety-Lok n UltraTouch n difference, % p-Value

K+, mmol/L 4.04 ± 0.5 35 3.88 ± 0.5 36 −3.88 0.26
GOT/AST, U/L 33.94 ± 32.7 32 27.29 ± 11.2 35 −19.60 0.67
LD, U/L 260.45 ± 107.1 31 255.03 ± 57.0 34 −2.08 0.25
Hemolysis index, mg/dL 12.49 ± 25.1 35 6.67 ± 5.4 36 −46.61 0.57
  1. Shown values represent mean ± SD, number of venipunctures (n), and relative differences in percent as well as p-value of Mann–Whitney test for statistical testing. AST, aspartate transferase; a.u., arbitrary units; CK, creatine kinase; CKD-EPI, Chronic Kidney Disease Epidemiology Collaboration; Cl, chloride; CRP, C-reactive protein; eGFR, estimated glomerular filtration rate; GOT, glutamic oxaloacetic transaminase; K, potassium; LD, lactate dehydrogenase; Na, sodium, inorg., inorganic.

Discussion

In this study, we compared the routinely applied 23G 3-bevel needle with the smaller outer diameter (25G) 5-bevel UT-PB needle in an aged inpatient population addressing both clinical use and laboratory diagnostics. It is important to mention that the two sets differ in regard of needle outer diameter and the bevel of the needle, both characteristics could theoretically lead to lower pain perception in the UT-PB, with a smaller needle diameter and a flatter needle bevel.

We found no significant differences between both sets considering subjective perception of the patient such as pain, fear, and general perception of blood collection. Previous studies comparing different needle size in regard to pain found mixed results, ranging from no significant differences between different sized needles [10], 24], 25] to a significant decrease in pain perception with decreasing outer needle diameter [11], 12]. As stated above, there seems to be a correlation to some extent between needle tip bevel and pain. Hirsch et al. found a significant lower pain perception in 84 diabetes patients using the 5-bevel design compared to the 3-bevel design for repeated subcutaneous injections [16]. Interestingly, in the initial assessment of pain, where the patients were blinded for the used set, there was no difference documented in pain perception. The difference occurred only after repeated use and after informing the patients about the lower puncture force. Similarly, Jaber et al. reviewed five clinical trials comparing 5- to 3-bevel needles and found less pain occurrence with the 5-bevel set [26]. However, in our study we found no difference in pain perception between the 25G 5-bevel UT-PB and the 23G 3-bevel SL set. Our results are in line with a study conducted in 190 cancer patients, who showed no significant difference in pain perception between the UT-PB and the frequently used 23G needle [20]. In contrast, a study with 214 subjects (age ≥18 up to >70; 40 % healthy subjects, 60 % patients) reported significantly less specific pain during needle insertion with the UT-PB set compared to the SL set, but no significant difference in the overall pain assessment [14]. Accordingly, in a study with 420 children, particularly the age group 3–12 reported significantly less pain with the UT-PB set compared to SL [21]. One possible reason for these differing results could be that although pain perception varied between the two sets (−16 % in the total cohort and −30 % in the sub-cohort), the sample size of our study might have been underpowered and thus was too small to reach statistical significance. Another reason could be that sensitivity–in general-to low-intensity pain, as by blood draw, decreases with aging and our study population was the oldest of all studies cited [27], 28]. However, we did not observe a clear correlation of pain with age within our aged cohort (analyses not shown). Further, we found no difference between both sets used when considering fear of the next blood sampling, which is consistent with the results of existing studies [14].

In the total cohort, we observed less and smaller hematoma development in the study group punctured with UT-PB (−16 % and −22 %, respectively) compared to SL. However, the differences were not statistically significant and could not be reproduced in the sub-cohort when both sets were used (+37 % and +4 %, respectively). In comparison to our results, Imbrahim et al. found in 50 volunteers, who underwent arterial puncture with either an 29G or a 23G needle, significantly less complications like swelling or bruising in the smaller needle group [29]. Another study reported no difference in hematoma size and pain of injection between 30G and 26G subcutaneous needles in 64 patients [24]. It is noticeable that studies with larger size needles rather show differences in hematoma development in contrast to studies with smaller size needles, which show no differences. It is conceivable that the effect is linked to larger sizes and diminishes with small sized needles, which, however, questions the major impact on a patient individual level.

In regard to laboratory measurements particularly focusing on hemolysis susceptibility, we did not find clinically relevant differences in any parameter between blood sample sets used, a result which is also consistent with other studies [13], 14], 30]. Importantly, direct comparisons, e.g., concerning non-hemolysis-susceptible analytes, should be regarded with caution since blood collection and laboratory analysis were not performed on the same day. Thus, changes within the time of hospitalization could impact measured values. The hemolysis index, indicating the grade of erythrocytes damage and thus the level of free blood hemoglobin, was reduced about 28 % in the total cohort and 47 % in the sub-cohort, although statistically not significant. Here, repeating the study with a larger cohort would possibly unravel, if indeed the UT-PB or comparable sets would lead to less hemolysis. However, ranges of hemolysis indices detected here, are still below relevant interference with laboratory analysis methods and without clinical impact for hemolysis-sensitive parameters.

From the phlebotomist’s subjective assessment, we found a significant difference in favor of the UT-PB set compared to the SL set regarding general severity of blood collection (30 %). This is comparable to the results of Hirsch et al., where diabetes patients favored the 5-bevel set in respect to comfort and ease of insertion [16] and to the results of Padoan et al., where phlebotomists rated the UT-PB set used in children associated with lower difficulty in general compared to the SL set [21]. However, it must be emphasized that this represents a subjective parameter that cannot be blinded due to the technical implementation and could therefore be biased.

Based on previous observations in a pediatric cohort (n=230 children) analyzing pain perception at the age group of 3–12 years [21], we included a comparable cohort size in our study. Although in general not underpowered, a major limitation of the study is the small size of the sub-cohort in which the same patients had blood drawn with both devices. The current sub-cohort, however, did not provide relevant differences in parameters, e.g., pain perception, between both systems. Furthermore, as the UT-PB-system combined a flatter needle bevel and a small needle diameter, the study cannot distinguish the influence of these features on pain perception. Another limitation is that blood draw was performed only by a single phlebotomist. Further, information on current pain medication treatment was not collected, leading to uncertainty about difference in such treatment between groups. However, we specifically aimed to conduct our study in a realistic non-homogenous cohort and therefore, other than age, did not introduce any exclusion criteria, like medication and comorbidities, in our study.

Taken together, there is little evidence from the results of our study that a 25G 5-bevel blood collecting set such as the UT-PB is associated with improved, significantly better handling compared to a commonly and routinely used set in an aged population. New medical devices such as the UT-PB could presumably help in treatment and reduce complications particularly at an individual level in aged patients, when more difficult blood access is evident.


Corresponding Author: Prof. Dr. Kai Kappert, Institute of Diagnostic Laboratory Medicine, Clinical Chemistry and Pathobiochemistry, Charité – Universitätsmedizin Berlin, Augustenburger Platz 1, 13353 Berlin, Germany; and Labor Berlin – Charité Vivantes GmbH, Berlin, Germany, E-mail:

Acknowledgements

We thank the participating patients and further the technical staff of Labor Berlin–Charité Vivantes GmbH for expertise measurement of clinical chemistry analytes.

  1. Research ethics: The study complied with all relevant national regulations, institutional policies and was performed in accordance with the Helsinki Declaration as revised in 2013 and approved by the local Ethics Committee of the Charité–Universitätsmedizin Berlin (EA2/231/20).

  2. Informed consent: All patients reported informed consent.

  3. Author contributions: AR, UMW, and KK contributed to the conception and design of the study. ES and MP contributed to the acquisition of data. AR, MP, KF, and KK contributed to the analysis and interpretation of data. AR, KF and KK drafted the manuscript, which was edited by AJ, UMK. AR, KF, MP and KK had full access to the data in the study. All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  4. Use of Large Language Models, AI and Machine Learning Tools: None declared.

  5. Conflict of interest: The authors state no financial or personal conflict of interest.

  6. Research funding: This work was financed by internal board funds of the Department of Geriatrics and Medical Gerontology, and of the Institute of Diagnostic Laboratory Medicine, Clinical Chemistry and Pathobiochemistry, Charité-Universitätsmedizin Berlin, Berlin, Germany.

  7. Data availability: The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

References

1. Simundic, AM, Bolenius, K, Cadamuro, J, Church, S, Cornes, MP, van Dongen-Lases, EC, et al.. Joint EFLM-COLABIOCLI Recommendation for venous blood sampling. Clin Chem Lab Med 2018;56:2015–38. https://doi.org/10.1515/cclm-2018-0602.Search in Google Scholar PubMed

2. Gabriel, J. Understanding the challenges to vascular access in an ageing population. Br J Nurs 2017;26:S15–23. https://doi.org/10.12968/bjon.2017.26.14.s15.Search in Google Scholar PubMed

3. Neuwinger, N, Meyer Zum Buschenfelde, D, Tauber, R, Kappert, K. Underfilling of vacuum blood collection tubes leads to increased lactate dehydrogenase activity in serum and heparin plasma samples. Clin Chem Lab Med 2020;58:213–21. https://doi.org/10.1515/cclm-2019-0616.Search in Google Scholar PubMed

4. Rosada, A, Prpic, M, Spiess, E, Friedrich, K, Neuwinger, N, Muller-Werdan, U, et al.. Underfilled blood collection tubes as pathologizing factor for measured laboratory parameters in older patients. J Am Geriatr Soc 2024;72:1553–6. https://doi.org/10.1111/jgs.18790.Search in Google Scholar PubMed

5. Raizman, JE, Goudreau, BL, Fuzery, AK, Cembrowski, GS. Barricor blood collection tubes are equivalent to PST for a variety of chemistry and immunoassay analytes except for lactate dehydrogenase. Clin Chim Acta 2019;496:18–24. https://doi.org/10.1016/j.cca.2019.06.013.Search in Google Scholar PubMed

6. Huyghe, T, Buntinx, F, Bruyninckx, R, Besard, V, Vunckx, J, Church, S, et al.. Studies on the use of BD Vacutainer(R) SST II and RST in general practice: investigation of artefactual hyperkalaemia. Ann Clin Biochem 2014;51:30–7. https://doi.org/10.1177/0004563213488758.Search in Google Scholar PubMed

7. Plebani, M. Errors in clinical laboratories or errors in laboratory medicine? Clin Chem Lab Med 2006;44:750–9. https://doi.org/10.1515/cclm.2006.123.Search in Google Scholar PubMed

8. Gu, S, Yang, C. Serum lactate dehydrogenase level predicts the prognosis in bladder cancer patients. BMC Urol 2023;23:65. https://doi.org/10.1186/s12894-023-01239-0.Search in Google Scholar PubMed PubMed Central

9. Saito, S, Sakamoto, S, Higuchi, K, Sato, K, Zhao, X, Wakai, K, et al.. Machine-learning predicts time-series prognosis factors in metastatic prostate cancer patients treated with androgen deprivation therapy. Sci Rep 2023;13:6325. https://doi.org/10.1038/s41598-023-32987-6.Search in Google Scholar PubMed PubMed Central

10. Patout, M, Lamia, B, Lhuillier, E, Molano, LC, Viacroze, C, Benhamou, D, et al.. A randomized controlled trial on the effect of needle gauge on the pain and anxiety experienced during radial arterial puncture. PLoS One 2015;10:e0139432. https://doi.org/10.1371/journal.pone.0139432.Search in Google Scholar PubMed PubMed Central

11. Gill, HS, Prausnitz, MR. Does needle size matter? J Diabetes Sci Technol 2007;1:725–9. https://doi.org/10.1177/193229680700100517.Search in Google Scholar PubMed PubMed Central

12. Arendt-Nielsen, L, Egekvist, H, Bjerring, P. Pain following controlled cutaneous insertion of needles with different diameters. Somatosens Mot Res 2006;23:37–43. https://doi.org/10.1080/08990220600700925.Search in Google Scholar PubMed

13. Lippi, G, Salvagno, GL, Montagnana, M, Brocco, G, Cesare, GG. Influence of the needle bore size used for collecting venous blood samples on routine clinical chemistry testing. Clin Chem Lab Med 2006;44:1009–14. https://doi.org/10.1515/cclm.2006.172.Search in Google Scholar PubMed

14. Mouser, A, Uettwiller-Geiger, D, Plokhoy, E, Berube, J, Ahuja, AJ, Stankovic, AK. Evaluation of pain and specimen quality by use of a novel 25-gauge blood collection set with ultra-thin wall cannula and 5-bevel tip design. J Appl Lab Med 2017;2:201–10. https://doi.org/10.1373/jalm.2017.023564.Search in Google Scholar PubMed

15. Dau, M, Buttchereit, I, Ganz, C, Frerich, B, Anisimova, EN, Daublander, M, et al.. Influence of needle bevel design on injection pain and needle deformation in dental local infiltration anaesthesia - randomized clinical trial. Int J Oral Maxillofac Surg 2017;46:1484–9. https://doi.org/10.1016/j.ijom.2017.06.013.Search in Google Scholar PubMed

16. Hirsch, L, Gibney, M, Berube, J, Manocchio, J. Impact of a modified needle tip geometry on penetration force as well as acceptability, preference, and perceived pain in subjects with diabetes. J Diabetes Sci Technol 2012;6:328–35. https://doi.org/10.1177/193229681200600216.Search in Google Scholar PubMed PubMed Central

17. Ponsiglione, AM, Ricciardi, C, Bonora, E, Amato, F, Romano, M. Impact of the number of needle tip bevels on the exerted forces and energy in insulin pen injections. Sensors (Basel) 2023;23. https://doi.org/10.3390/s23198043.Search in Google Scholar PubMed PubMed Central

18. BD White Paper VS9248; Comparison of penetration force for the BD Vacutainer® UltraTouch™ push button blood collection set with PentaPoint™ comfort bevel and RightGauge™ cannula to the current BD Vacutainer® push button blood collection set with thin wall 3-bevel cannula. Franklin Lakes, NJ: BD White Paper VS9248; 2016.Search in Google Scholar

19. Phelan, MP, Reineks, EZ, Schold, JD, Hustey, FM, Chamberlin, J, Procop, GW. Preanalytic factors associated with hemolysis in emergency department blood samples. Arch Pathol Lab Med 2018;142:229–35. https://doi.org/10.5858/arpa.2016-0400-oa.Search in Google Scholar PubMed

20. Kirchhoff, D, Baser, R, Kwong, D, Ramanathan, L, McCash, SI. Comparison of a novel thin-walled 25-gauge needle Push button blood collection set to a standard 23-gauge needle in a cancer patient population. J Appl Lab Med 2023;8:264–71. https://doi.org/10.1093/jalm/jfac129.Search in Google Scholar PubMed PubMed Central

21. Padoan, A, Sirini, S, Mazzone, R, Mesiti, C, Grillo, C, Meyer, B, et al.. Evaluation of an improved small gauge needle for venipuncture in children with difficult venous access: impact on sample quality, phlebotomist satisfaction and patient pain perception. Clin Chim Acta 2020;500:213–9. https://doi.org/10.1016/j.cca.2019.10.019.Search in Google Scholar PubMed

22. Levey, AS, Stevens, LA, Schmid, CH, Zhang, YL, Castro, AF3rd, Feldman, HI, et al.. A new equation to estimate glomerular filtration rate. Ann Intern Med 2009;150:604–12. https://doi.org/10.7326/0003-4819-150-9-200905050-00006.Search in Google Scholar PubMed PubMed Central

23. Sciacovelli, L, Padoan, A, Aita, A, Basso, D, Plebani, M. Quality indicators in laboratory medicine: state-of-the-art, quality specifications and future strategies. Clin Chem Lab Med 2023;61:688–95. https://doi.org/10.1515/cclm-2022-1143.Search in Google Scholar PubMed

24. Robb, DM, Kanji, Z. Comparison of two needle sizes for subcutaneous administration of enoxaparin: effects on size of hematomas and pain on injection. Pharmacotherapy 2002;22:1105–9. https://doi.org/10.1592/phco.22.13.1105.33510.Search in Google Scholar PubMed

25. Yee, K, Shetty, AL, Lai, K. ABG needle study: a randomised control study comparing 23G versus 25G needle success and pain scores. Emerg Med J 2015;32:343–7. https://doi.org/10.1136/emermed-2014-203600.Search in Google Scholar PubMed PubMed Central

26. Jaber, A, Bozzato, GB, Vedrine, L, Prais, WA, Berube, J, Laurent, PE. A novel needle for subcutaneous injection of interferon beta-1a: effect on pain in volunteers and satisfaction in patients with multiple sclerosis. BMC Neurol 2008;8:38. https://doi.org/10.1186/1471-2377-8-38.Search in Google Scholar PubMed PubMed Central

27. Lautenbacher, S, Peters, JH, Heesen, M, Scheel, J, Kunz, M. Age changes in pain perception: a systematic-review and meta-analysis of age effects on pain and tolerance thresholds. Neurosci Biobehav Rev 2017;75:104–13. https://doi.org/10.1016/j.neubiorev.2017.01.039.Search in Google Scholar PubMed

28. El Tumi, H, Johnson, MI, Dantas, PBF, Maynard, MJ, Tashani, OA. Age-related changes in pain sensitivity in healthy humans: a systematic review with meta-analysis. Eur J Pain 2017;21:955–64. https://doi.org/10.1002/ejp.1011.Search in Google Scholar PubMed

29. Ibrahim, I, Yau, YW, Ong, L, Chan, YH, Kuan, WS. Arterial puncture using insulin needle is less painful than with standard needle: a randomized crossover study. Acad Emerg Med 2015;22:315–20. https://doi.org/10.1111/acem.12601.Search in Google Scholar PubMed

30. BD White Paper VS9250: Performance evaluation of BD vacutainer UltraTouch Push button blood collection set and comparison with the current BD vacutainer Push button blood collection set for visual and analytical inidcators of hemolysis. Franklin Lakes, NJ: BD White Paper VS9250; 2015.Search in Google Scholar


Supplementary Material

This article contains supplementary material (https://doi.org/10.1515/labmed-2024-0097).


Received: 2024-05-30
Accepted: 2024-11-08
Published Online: 2024-12-09
Published in Print: 2025-02-25

© 2024 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Downloaded on 5.11.2025 from https://www.degruyterbrill.com/document/doi/10.1515/labmed-2024-0097/html?lang=en
Scroll to top button