Agreement between procalcitonin measurements using the new point-of-care testing ichroma™ reader and the automated Kryptor instrument

  • Elisabetta Stenner EMAIL logo , Giulia Barbati , Nicole West , Fabia Del Ben , Francesca Martin und Maurizio Ruscio
Veröffentlicht/Copyright: 15. Mai 2019
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

Background

To evaluate if procalcitonin (PCT) measurements made using the new point-of-care testing (POCT) ichroma™ are interchangeable with those made using Kryptor.

Methods

Serum samples (n = 117) were processed sequentially on Kryptor and ichroma™. Statistical analysis was performed using Passing-Bablok (PB) regression and the Bland-Altman (BA) test. Cohen’s kappa statistic was used to calculate the concordance at the clinically relevant cutoffs.

Results

PB regression did not show a significant deviation from linearity; proportional and constant differences were observed between ichroma™ and Kryptor. The 95% confidence interval (CI) of the mean bias percentage was very large, exceeding the maximum allowable total error (TE) (approximately 20%) and the clinical reference change value (about 60%). However, the concordance between methods at the clinically relevant cutoffs was strong, with the exception of the 0.25 ng/mL cutoff, which was moderate.

Conclusions

Our data suggest that ichroma™ is not interchangeable with Kryptor, so cannot be mixed; one must choose one instrument only and be consistent. However, while the strong concordance at the clinically relevant cutoffs allows us to consider ichroma™ a suitable option to Kryptor to support clinicians’ decision-making, nevertheless the moderate agreement at the 0.25 ng/mL cutoff recommends caution in interpreting the data around this cutoff.

Sepsis is a significant public health problem throughout the world, with more than 31 million cases reported annually and a 17% mortality rate [1]. Early detection of infection is essential: the earlier the diagnosis and the antibiotic therapy, the better the chances of survival [2].

Procalcitonin (PCT) has been used in Europe for many years and was also approved for use in the United States by the US Food and Drug Administration (FDA) as a diagnostic aid for sepsis in 2005. It also gained an FDA indication in 2016 for serial use to assess sepsis progression and 28-day mortality risk [3]. However, due to the suboptimal sensitivity and/or specificity of the PCT test, results should always be interpreted alongside the clinical context [4], [5], [6]. A 24-h availability of PCT determinations is desirable [4], [5], [6]. For this purpose, a point-of-care testing (POCT) could be an attractive solution [7], especially in a local small laboratory, when it is physically separated from the central routine laboratory where the state-of-the-art instruments are usually to be found. The critical point is that PCT is usually trend-evaluated, especially in order to guide/monitor antibiotic therapy [4], [5], [6], [8]. Consequently, if a critical patient should be moved from a local hospital to the central hospital, the interchangeability of the PCT measurements using POCT, and those using the central routine laboratory instrument, is mandatory to evaluate the PCT kinetic changes accurately. The aim of this study was to find out if the PCT measurements using the new POCT ichroma™ are interchangeable with those using Kryptor, the well-accepted gold-standard instrument.

To do this, we investigated samples (n=117), processed within 2 h after blood drawing, selected from pre-existing routine samples with a PCT result (in-house method: Liaison XL, Diasorin, Saluggia, Italy). PCT assays were performed daily using firstly Kryptor (BRAHMS AG, Hennigsdorf/Berlin, Germany) and secondly the ichroma™ reader (Boditech Med Incorporated, Gangwon-do, Korea). We tested serum samples because requests for multiple tests (chemistry and immunoassay) are usually made to a local laboratory where the use of only one sample (i.e. serum) tube is not only practically and economically suitable, but also serves to limit patients’ discomfort.

This study was performed according to the Code of Ethics of the World Medical Association (Declaration of Helsinki) for experiments involving humans. The Hospital Institutional Review Board waived the need for informed consent.

Our range of PCT values was 0.08–60.85 ng/mL using Kryptor and 0.13–66.27 ng/mL using ichroma™. Within-run and between-run analytical imprecisions were consistent with claims made by the manufacturer (<10%) (Table 1). Passing-Bablok (PB) regression (Figure 1A, Table 2) did not show a significant deviation from linearity (all p-values from the Cusum linearity test: >0.10) while proportional and constant differences were observed between Kryptor and ichroma™ except for data under 2 ng/mL (intercept contained 0 and slope contained 1 in the 95% confidence interval [95% CI]). The mean bias% was within the desirable quality specification for total error (TE) (<20%) [10] but the 95% CI was very large, exceeding both the TE specification (20%) and the clinical reference change value (about 60%) (Table 3).

Table 1:

Main characteristics of the reagents and instruments used in this study.

Methodichroma™ PCTKryptor BRAHMS PCT sensitive
Companyichroma™, Boditech Med, Gangwon-do, KoreaBRAHMS, Hennigsdorf, Germany
Assay principleFluorescence immunoassay (FIA)Time-resolved amplified cryptate emission (TRACE) immunoassay
Within-run imprecision (CV%)aL1 mean: 0.45 ng/mL; CV%: 5.2L1 mean: 0.28 ng/mL; CV%: 4.7
L2 mean: 15.16 ng/mL; CV%: 3.9L2 mean: 9.86 ng/mL; CV%: 2.6
Between-run imprecision (CV%)aL1 mean: 0.47 ng/mL; CV%: 9.4L1 mean: 0.29 ng/mL; CV%: 7.6
L2 mean: 15.2 ng/mL; CV%: 4.8L2 mean: 10.0 ng/mL; CV%: 5.3
Within-run imprecision (CV%)bL1 mean: 0.44 ng/mL; CV%: 5.6L1 mean: 0.43 ng/mL; CV%: 4.5
L2 mean: 1.68 ng/mL; CV%: 4.20L2 mean: 1.46 ng/mL; CV%: 2.9
Between-run imprecision (CV%)bL1 mean: 0.50 ng/mL; CV%: 6.8L1 mean: 0.32 ng/mL; CV%: 7.1
L2 mean: 1.67 ng/mL; CV%: 6.1L2 mean: 1.58 ng/mL; CV%: 5.5
Sample volume150 μL50 μL
Measuring time12 min19 min
Measuring range0.06–100 ng/mL0.02–50 ng/mL
LoD0.06 ng/mL0.02 ng/mL
LoQ0.10 ng/mL0.06 ng/mL
DilutionsManuallyOn-board automatic dilution
Specimen typeSerum, heparinized plasma, whole bloodSerum, EDTA or heparinized plasma
  1. Within-run and between-run imprecision, expressed by the analytical coefficient of variation (CV%), was calculated for all instruments by running five replicates of the same quality control (a) (Brahms PCT sensitive Kryptor QC Level 1 [L1] and Level 2 [L2]) and the same serum samples (b), for 5 times [9] during a period of 2 weeks to compare the CV% at the same concentrations. The acceptance limits for the bias% were defined a priori (1) by the desirable specifications for the allowable total error (TE) based on the biological variation [10], [11], namely <20% and (2) by the clinically well-accepted daily decrease of 50% change values, useful to suggest the adequacy of the antibiotic treatment [12] or by an increase of 88% during the first 24 h, suggesting the patient has turned out to have an infection [12]. LoD, limit of detection; LoQ, limit of quantitation; EDTA, ethylenediaminetetraacetic acid.

Figure 1: Passing-Bablok regression (A) and Bland-Altman plot (B and C) comparing procalcitonin (PCT) measurements using Kryptor and ichroma™.The acceptance limits for the bias% were defined a priori (1) by the desirable specifications for the allowable total error (TE), based on the biological variation [11], [12], namely <20% and (2) by the clinically well-accepted daily decrease of 50% change values, useful to suggest the adequacy of the antibiotic treatment [11] or by an increase of 88% during the first 24 h, suggesting the patient has turned out to have an infection [11]. (A) Red-dashed line: identity, blue-dotted line: Passing-Bablok regression. (B) Central bold-dashed red line represents the difference standardized with respect to the mean ([method A – method B]/mean %), with corresponding red-dashed 95% confidence interval. Blue bold-dashed lines represent limits of agreement with respect to the percentage standardized difference. Blue-dashed lines represent 95% confidence intervals for the values of each limit of agreement. (C) Central bold-dashed red line represents the mean difference with corresponding red-dashed 95% confidence interval. Blue bold-dashed lines represent limits of agreement (mean difference plus and minus 1.96 times the standard deviation of the differences). Blue-dashed lines represent 95% confidence intervals for the values of each limit of agreement.
Figure 1:

Passing-Bablok regression (A) and Bland-Altman plot (B and C) comparing procalcitonin (PCT) measurements using Kryptor and ichroma™.

The acceptance limits for the bias% were defined a priori (1) by the desirable specifications for the allowable total error (TE), based on the biological variation [11], [12], namely <20% and (2) by the clinically well-accepted daily decrease of 50% change values, useful to suggest the adequacy of the antibiotic treatment [11] or by an increase of 88% during the first 24 h, suggesting the patient has turned out to have an infection [11]. (A) Red-dashed line: identity, blue-dotted line: Passing-Bablok regression. (B) Central bold-dashed red line represents the difference standardized with respect to the mean ([method A – method B]/mean %), with corresponding red-dashed 95% confidence interval. Blue bold-dashed lines represent limits of agreement with respect to the percentage standardized difference. Blue-dashed lines represent 95% confidence intervals for the values of each limit of agreement. (C) Central bold-dashed red line represents the mean difference with corresponding red-dashed 95% confidence interval. Blue bold-dashed lines represent limits of agreement (mean difference plus and minus 1.96 times the standard deviation of the differences). Blue-dashed lines represent 95% confidence intervals for the values of each limit of agreement.

Table 2:

Passing-Bablok regression values for procalcitonin measurements using ichroma™ in contrast to Kryptor.

Intercept (95% CI)Slope (95% CI)
Total (n=117)0.12 (0.07–0.18)0.87 (0.80–0.93)
<10 ng/mL (n=79)0.06 (0.01–0.13)0.98 (0.89–1.03)
<2 ng/mL (n=54)0.04 (−0.02–0.12)1.01 (0.86–1.14)
  1. Data analysis (R statistical package, libraries “BlandAltmanLeh” and “mcr” and MedCalc Statistical Software) was performed using Passing-Bablok regression to test the linear relationship between the measurements [13]. The statistical analysis was made of all the results and of two different subgroups (results below 2 ng/mL and below 10 ng/mL).

Table 3:

Bland-Altman percentage differences (bias%), absolute bias, limits of agreements and their confidence intervals for procalcitonin measurements using Kryptor vs. ichroma™.

Bias% (95% CI)Lower limit (95% CI)Upper limit (95% CI)
Absolute bias (95% CI)Absolute lower limit (95% CI)Absolute upper limit (95% CI)
Total (n=117)−2.21% (−8.42–3.98)−68.62% (−79.25 to −57.98)64.18% (53.55–74.81)
1.31 ng/mL (0.50–2.13)−7.41 ng/mL (−8.83 to −6.00)10.05 ng/mL (8.63–11.46)
<10 ng/mL (n=79)−10.40% (−18.06 to −2.75)−78.64% (−91.89 to −65.40)57.82 % (44.58–71.08)
−0.02 ng/mL (−0.13–0.09)−1.04 ng/mL (−1.24 to −0.84)0.99 ng/mL (0.80–1.20)
<2 ng/mL (n=54)−16.42% (−26.84 to −6.00)−93.38% (−111.43 to −75.34)60.53% (42.49–78.58)
−0.07 ng/mL (−0.13 to −0.01)−0.50 ng/mL (−0.60 to −0.40)0.35 ng/mL (0.25–0.45)
  1. Data analysis (R statistical package, libraries “BlandAltmanLeh” and “mcr” and MedCalc Statistical Software) was performed using Bland-Altman to estimate the consistency of the methods [13]. The statistical analysis was made of all the results and of two different subgroups (results below 2 ng/mL and below 10 ng/mL).

The data distribution shows that the bias% and even more the absolute bias could depend on the concentration (Figure 1B and C). However, even if the data analysis is evaluated in a smaller range (under 10 ng/mL and under 2 ng/mL), the bias% is still not acceptable. Therefore, ichroma™ may not be considered suitable to be used interchangeably with Kryptor for monitoring patients.

However, Cohen’s kappa statistic [14], used to calculate the concordance at clinically relevant cutoffs for bacterial infections, showed that even though the agreement between ichroma™ and Kryptor values at the clinical cutoff of 0.25 ng/mL was moderate [14] (κ=0.725; 95% CI: 0.532–0.917), suggesting caution in the interpretation of the data around this value, the agreement at 0.50 ng/mL (κ=0.878; 95% CI: 0.774–0.982), 2.0 ng/mL (κ=0.983; 95% CI: 0.949–1.016) and 10 ng/mL (κ=0.938; 95% CI: 0.869–1.007) was strong, suggesting that the ichroma™ can be used to support clinicians’ decision-making. Basically the same results were observed for Liaison XL with respect to the ichroma™ [15] instrument.

In conclusion, our data suggest that ichroma™ is not interchangeable with Kryptor, so cannot be mixed with it for monitoring patients; one must choose one instrument only and be consistent. However, while the strong concordance at the clinically relevant cutoffs allows us to consider ichroma™ a suitable option to Kryptor to support clinicians’ decision-making, nevertheless the moderate agreement at the 0.25 ng/mL cutoff recommends caution in interpreting the data around this cutoff.

Moreover, our study has some limitations that need to be considered: (a) we tested the interchangeability between methods in only one laboratory and (b) we used only serum samples. Whole blood and plasma samples need to be investigated in future, possibly in a multicenter study.

Acknowledgments

The authors thank Dr. Nicholas Carter for proof-reading this article.

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Fleischmann C, Scherag A, Adhikari NK, Hartog CS, Tsaganos T, Schlattmann P, et al. Assessment of global incidence and mortality of hospital-treated sepsis. Current estimates and limitations. Am J Respir Crit Care Med 2016;193:259–72.10.1164/rccm.201504-0781OCSuche in Google Scholar PubMed

2. Rhodes A, Evans LE, Alhazzani W, Levy MM, Antonelli M, Ferrer R, et al. Surviving sepsis campaign: International Guidelines for Management of Sepsis and Septic Shock: 2016. Intensive Care Med 2017;43:304–77.10.1007/s00134-017-4683-6Suche in Google Scholar PubMed

3. Rhee C. Using procalcitonin to guide antibiotic therapy. Open Forum Infect Dis 2016;4:ofw249.10.1093/ofid/ofw249Suche in Google Scholar PubMed PubMed Central

4. Sager R, Kutz A, Mueller B, Schuetz P. Procalcitonin-guided diagnosis and antibiotic stewardship revised. BMC Med 2017;15:15.10.1186/s12916-017-0795-7Suche in Google Scholar PubMed PubMed Central

5. Hey J, Thompson-Leduc P, Kirson NJ, Zimmer L, Wilkins D, Rice B, et al. Procalcitonin guidance in patients with lower respiratory tract infections: a systematic review and meta-analysis. Clin Chem Lab Med 2018;56:1200–9.10.1515/cclm-2018-0126Suche in Google Scholar PubMed

6. Bartoletti M, Antonelli M, Blasi FA, Casagranda I, Chieregato A, Fumagalli R, et al. Procalcitonin-guided antibiotic therapy: an expert consensus. Clin Chem Lab Med 2018;56:1223–9.10.1515/cclm-2018-0259Suche in Google Scholar PubMed

7. Singh M, Anand L. Bedside procalcitonin and acute care. Int J Crit Illn Inj Sci 2014;4:233–7.10.4103/2229-5151.141437Suche in Google Scholar PubMed PubMed Central

8. Fan SL, Miller NS, Remick DG. Diagnosing sepsis. The role of laboratory medicine. Clin Chim Acta 2016;460:203–10.10.1016/j.cca.2016.07.002Suche in Google Scholar PubMed PubMed Central

9. Clinical and Laboratory Standards Institute. User verification of precision and estimation of bias; approved guideline – Third Edition. CLSI document EP15-A3. Wayne, PA, USA: CLSI; 2014. https://clsi.org/standards/products/method-evaluation/documents/ep15/.Suche in Google Scholar

10. Barassi A, Pallotti F, Melzi d’Eril GV. Biological variation of procalcitonin in healthy individuals. Clin Chem 2004;50:1878.10.1373/clinchem.2004.037275Suche in Google Scholar PubMed

11. Jay DW. Method comparison: where do we draw the line? Clin Chem Lab Med 2011;49:1089–90.10.1515/CCLM.2011.632Suche in Google Scholar PubMed

12. Tràsy D, Molnàr Z. Procalcitonin-assisted antibiotic strategy in sepsis. eJIFCC 2017;28:104–13.Suche in Google Scholar

13. Clinical and Laboratory Standards Institute (CLSI). Measurement procedure comparison and bias estimation using patient samples. Approved guideline – Third Edition. CLSI document EP09-A3. Wayne, PA, USA, 2013. http://www.labac.eu/telechargements_labac/2016/07/CLSI-EP09A3E.pdf.Suche in Google Scholar

14. McHugh ML. Interrater reliability: the kappa statistic. Biochem Med 2012;22:276–82.10.11613/BM.2012.031Suche in Google Scholar

15. Stenner E, Barbati G, West N, Del Ben F, Martin F, Ruscio M. Interchangeability of procalcitonin measurements using the point-of-care testing i-CHROMA™ Reader and the automated Liaison XL. Clin Lab 2018;64:1097–100.10.7754/Clin.Lab.2018.180133Suche in Google Scholar PubMed

Received: 2018-11-14
Accepted: 2019-04-08
Published Online: 2019-05-15
Published in Print: 2019-06-26