Startseite Medizin Quality control and validation in flow cytometry
Artikel Open Access

Quality control and validation in flow cytometry

  • Alexandra Dorn-Beineke EMAIL logo und Ulrich Sack
Veröffentlicht/Copyright: 9. März 2016

Abstract

Diagnostic flow cytometry has been established in laboratory diagnostics as an indispensable, deeply specialized method. Because of the special needs of the analysis in vital cells, the introduction and maintenance of quality management systems stays a challenge. Meanwhile, experience has been collected for quality controls and validation in diagnostic flow cytometry. This will be summarized in this article. The focus will be on the relevant rules and regulations for diagnostic laboratories and the associated requirements for medical diagnostics, as well the guidelines of the German Medical Association, Bundesärztekammer, (RiliBÄK) and in particular the accreditation according to DIN EN ISO 15189. In detail, factors influencing flow cytometry analyses, strategies for the standardization and harmonization in flow cytometry, and possibilities for verification and validation in the flow cytometry laboratory are given. A summary of the most frequent quality problems rounds up this overview.

Reviewed Publication:

Sack U. Conrad K.


Introduction

Methods of diagnostic flow cytometry are commonly used primarily in the fields of immunology and hematology today. The cellular flow cytometric analysis, however, can also be applied in other fields, for example, in hemostaseology, transfusion medicine or pathology. But in most cases, routine diagnosis is the domain of diagnostic laboratories specialized in hematological and/or immunological diagnostics, and this chapter will focus on them. Today, most of these laboratories work with CE-marked test kits and ready-to-use reagents (in-vitro diagnostics). Nevertheless, validation and the correct implementation of quality controls place considerable demands on the laboratory in its clinical-diagnostic routines.

The sample in specialized hematological and immunological laboratories

The requirements for a specialized hematological or immunological laboratory have become very complex as a result of the diversity in the methods that have been established over time. In terms of quality assurance, one cannot only focus on flow cytometry samples, but must also take into account the different requirements.

The following types of tests are typically found in specialized hematological laboratories (in ascending order of specialization):

  • The automated measurement of corpuscular blood components or corpuscular components from various special materials;

  • The methods of cytomorphology and cytochemistry;

  • Flow cytometry;

  • Cytogenetics;

  • The determination of colony forming units (CFUs);

  • The methods of molecular diagnostics; and

  • Next Generation Sequencing (NGS).

For specialized immunological laboratories, one must also add cellular functional assays, which are frequently designed so individually that routine protocols are generally available only rarely. The definition of a specialized hematology laboratory and the minimum requirements that must be put to such a laboratory are outlined in a position paper of the Laboratory Working Group of the German Society of Hematology and Oncology (DGHO) [1].

The spectrum of cytometrically analyzable specimens and their various corpuscular components is broad. For example, particles 0.5–50 μm in size can be analyzed. These include physiological blood cells, tumor cells, chromosomes, as well as algae and protozoa, but they play no role in specialized hematological or immunological laboratories. But of importance are also microvesicles and bead-based assays, which are used in the analysis of cellular functions in the multiplex evaluation of cellular products.

The various sample materials, such as peripheral blood, bone marrow, cerebrospinal fluid, various puncture fluids and suspensions from tissues create different requirements in terms of transport, storage and pre-analysis. The complex technology places special demands on environmental conditions, equipment specifications and the maintenance of equipment. The variety of methods and users will also require a consolidated training policy and careful documentation [1–4].

Given its high degree of complexity, flow cytometry has nowhere near achieved the same degree of standardization as other laboratory methods. However, specific points of flow cytometry have been added to national and international quality-relevant documents in recent years, so that there are already some binding targets in internal and external quality control as well as in the accreditation process.

The requirements for the analytic process can be seen in the respective documents and checklists (see below). The following section will discuss the main quality-relevant points.

Applied quality assurance in the flow cytometry laboratory

Different definitions of the term quality are found in the literature. It is very common to define quality as the degree of agreement between claims or expectations (target) about a product and its characteristics (actual). In the same vein, DIN EN ISO 8402 defines quality as “the totality of characteristics of an entity that bear on its ability to satisfy stated and implied needs.”

The quality standard DIN EN ISO 9000:2000ff goes one step further and describes quality as the ability of a set of inherent characteristics of a product, system or process to fulfill requirements of customers and other interested parties. The quality management system is defined as coordinated activities to direct and control an organization to improve the quality of products manufactured and services provided [5]. This very abstract definition is hard to apply to the requirements of a laboratory. But as a general rule, an accredited quality management serves to attain the following objectives:

  • Meeting the demands of one’s own management;

  • Use for marketing purposes;

  • Improving quality and competence;

  • Increasing customer satisfaction;

  • Increasing legal certainty;

  • Increasing the motivation of employees.

The long-term goal is to increase effectiveness and quality, which will benefit the patients, as well as the staff who implement the accreditation [6].

Regulations for diagnostic laboratories

Quality assurance measures in the laboratory are essential. This applies even to research laboratories, which operate according to the rules of generally accepted scientific practice and must plan, control and document their experiments [7]. In the field of flow cytometry, the MiFlowCyte criteria are especially relevant, compliance with which is now regarded as a prerequisite for acceptance of the tests carried out. These criteria represent recommendations of ISAC (International Society for Analytical Cytometry) and describe the minimum requirements with respect to information that must be provided about a flow cytometry experiment [8].

There are numerous other regulations for diagnostic laboratories that apply to the analyses to be carried out in any given context. Generally, analyses in the context of patient diagnostics in Germany are subject to the guidelines of the German Medical Association (“RiliBÄK”). Where studies and international cooperation are concerned, especially with the US, compliance with Good Laboratory Practice (GLP) is often required. Here it is important to take into account that the GLP criteria of the Food and Drug Administration (FDA) do not correspond with the German approval as a GLP test facility, but rather with accreditation according to DIN EN ISO 15189 [9]. In Germany, however, GLP is regulated by the Chemicals Act and constitutes a quality assurance system that deals with non-clinical health and environmental safety inspections [10].

Where flow cytometric methods are carried out in the context of the manufacture of cell therapy medicinal products or blood products as quality controls, they must meet the conditions of the manufacturing authorization, and they will be checked by the authorizing authority specifically for compliance with Good Manufacturing Practice (GMP) criteria as part of the approval and monitoring process.

Requirements for medical diagnostics

Guidelines of the German Medical Association (“RiliBÄK”)

The currently valid guidelines of the German Medical Association were adopted on 01.08.2008, and last updated in accordance with the decision of the Board of the German Medical Association on 11.04.2014 and 20.06.2014 [2]. The legal basis for the application of RiliBÄK is Section 4a of the Medical Devices Operator Ordinance (MPBetreibV) [11]. RiliBÄK generally assumes the use of in-vitro diagnostics. Flow cytometric analysis methods, however, are often only validated in the laboratory (see “Verification and validation in the flow cytometry laboratory”), with a distinction to be made between qualitative, semiquantitative and quantitative assays [12]. However, only qualitative and quantitative analysis methods are defined under RiliBÄK.

When it comes to CD4-positive T-cells and CD8-positive T-cells, RiliBÄK merely contains specifications about participating in round-robin tests every 6 months. However, this classification contradicts the clinical necessity of quantitative indication of these parameters [13–15]. For the amendment of RiliBÄK, it will be necessary to determine current values for precision and accuracy for the above mentioned parameters in the peripheral blood. Since there is no recognized reference method in the field of immunophenotyping, an evidence-based foundation is to be used for the future definition of coefficients of variation used in internal and external quality control [13].

Other applicable quality-relevant documents

These include, for example, guidelines, standards, checklists and norms. A guideline requires an action or implementation, but is not in itself a formal law. The special position of RiliBÄK was pointed out in Section “Guidelines of the German Medical Association (“RiliBÄK”)”.

Standards with the label “DIN EN ISO” are first prepared and adopted by the International Standardization Organization (ISO), with the participation of various national affiliates. They are then adopted as European Norms (EN), provided that the European standardization bodies and the European Commission agree. Finally, they are taken up by the German Institute for Standardization (DIN) and incorporated into the national standards of Germany. By being incorporated into national standards, an international standard takes effect in a specific state. While the adoption of ISO norms is voluntary, the member states of the European Union are obligated to incorporate European norms into their national rules.

Checklists are very useful tools, such as to prepare for an upcoming accreditation. Evaluators work with checklists that were created by professional societies in cooperation with the accreditation bodies.

Below, a list of some selected documents:

  • DIN EN ISO 15189 checklist for medical laboratories [3]

  • Board resolutions of the sector committee on medical laboratory diagnostics [16]

  • Immunology checklist [4]

  • Immunogenetics checklist [17]

  • Clinical Chemistry and Hematology checklist [18] (currently being revised)

  • College of American Pathologists (CAP): CAP Accreditation Program-Flow Cytometry Checklist [19].

Individual aspects of the above documents are presented below. For the purposes of this paper, it is not possible to discuss in detail test-specific consensus guidelines, such as those of the International Clinical Cytometry Society (ICCS) or CLSI Document Guidelines for Flow Cytometry, or standards for flow cytometric research, such as MIFlowCyt or the guidelines for good scientific practice [7, 8].

Documentation and retention requirements

Documentation plays a very large role in a quality management system. The documentation obligation as well as the duration of the retention requirements of quality-relevant documents are set in multiple locations [2, 3, 11]. RiliBÄK requires a steered documentation system. In addition, according to RiliBÄK, all measured results of quality assurance must be retained for 5 years, unless regulated differently under other provisions that require longer retention periods [2]. MPBetreibV Section 4a(4) defines the retention obligation: The documents relating to the established quality assurance system, the checks performed, as well as the certificates of participation in round-robin tests and inter-laboratory test certificates issued must be kept for a period of 5 years, unless a longer retention period is required under different provisions. The documents must be submitted to the competent authority if and when requested [11]. DIN EN ISO 15189 contains several references to the documentation obligation. According to 4.2.1 “Quality Management System”, the laboratory must set up, document, implement and maintain a QM system and continuously improve upon its effectiveness in compliance with this international norm. Elsewhere, under 4.13 “Control of Records” it is stated: The laboratory must have a documented procedure for identification, collection, directory development, accessibility, storage, maintenance, modification and secure disposal of quality-related and technical records. The laboratory must specify the time period during which records belonging to the QM system must be retained, including pre-analytical measures, test methods and post-analytical measures. The retention period for records may differ [3].

For the daily QM routine of a laboratory, this means: Run proper diagnostics, and document and evaluate everything! The key documents in a quality management system include:

  • Quality management manual;

  • Description of pre-analytical measures;

  • Procedure instructions for the description of processes;

  • Work instructions on test methods;

  • Work instructions on equipment;

  • Form sheets;

  • Lists and notices;

  • Checklists.

The documentation also includes, for example:

  • Archiving of raw data;

  • Batch documentation of test kits, reagents, control preparations used;

  • Documentation for batch testing and batch release of reagents;

  • Training and equipment instructions;

  • Records of internal and external quality controls;

  • Records of external laboratory comparisons;

  • Results of staff comparisons (consensus training);

  • Maintenance records (with date/time);

  • Complaints and their processing;

  • Documentation in the context of risk management.

The documentation in a QM system generally helps in advancing the standardization process. Thus, preparations and measurement protocols are standardized with thorough documentation, which also enables the traceability of batches of reagents and samples.

Accreditation in the flow cytometry laboratory according to DIN EN ISO 15189 and its requirements

While in most European countries the accreditation of medical laboratories is increasingly becoming obligatory, it is voluntary in Germany [20]. Many laboratories in Germany are still certified, as this allows for better comparability, facilitates international cooperation and, moreover, is often required by external clients.

The requirements of DIN EN ISO 15189 are the same for all laboratory areas and therefore also apply to flow cytometry. Special circumstances arise here not only in the context of the analytical process (see below), but also with respect to the general management requirements. The flow cytometry laboratory is often a separate entity in organizational terms. Quality management must take account of the special procedures in such a laboratory. Subcontracts are often associated here with complex interpretations and must be included accordingly when preparing the findings. The inclusion of cytometric raw data and findings in the laboratory’s IT is mostly labor-intensive and requires individual solutions. Ideally, flow cytometry data would be incorporated into a consolidated overall report at the end of the process.

The staff must be trained appropriately. Likewise, a program for ongoing professional development (PD) must be established. This includes regular internal and external PD opportunities, the success of which must be checked. What does this mean in detail and where does one find appropriate specifications? The basic training for specialists or medical-technical laboratory assistants alone is not enough to meet the very specific requirements of a hematological laboratory. Accordingly, the laboratory should have on its staff a hematological medical-technical lab assistant or medical-technical lab assistant with equivalent qualifications. The diagnostic report should be drafted as a synopsis by a staff physician (laboratory physician and/or hematologist-oncologist, pathologist), and should always contain an assessment. Furthermore, the medical and technical staff of a specialist laboratory must have the necessary experience with the methods and procedures employed. The number of tests per month and year and should correspond to a minimum requirement [1]. According to 5.1.2 of DIN EN ISO 15189, the laboratory management must document the qualifications for the entire staff and for each position. The qualifications must reflect the proper education, training, necessary experience and proven skills, and must be appropriate for the tasks to be performed. The staff carrying out technical assessments of tests must have a relevant theoretical and practical background and experience. Regarding training measures and skills assessment, one can read in 5.1.5 and 5.1.6 that the laboratory must provide training measures to the entire staff in the following areas: a) the QM system, b) the work processes and methods assigned, etc. The staff must be under technical supervision during training measures at all times. The effectiveness of the training programs must be assessed regularly. Once the training has been completed, the laboratory will have to assess each employee’s capability of performing the assigned management or technical tasks in accordance with specified criteria.

Follow-up assessments must be carried out at regular intervals. Where necessary, the training will have to be repeated. As for the regularity and documentation of training and PD measures, it is stated in 5.1.8 and 5.1.9 that a program for ongoing professional development must be made available to employees who are involved in management and technical processes, and employees must participate in PD. The effectiveness of the PD program must be checked regularly. Employees must participate regularly in professional development or other activities associated with their profession. Records of relevant training and professional qualifications obtained, professional development and experience, as well as skills assessment of all employees must be maintained [3]. One could thus say that employees “owe” a duty to participate, while the employer has a duty to collect on such “debt”.

The facilities must be of adequate size and properly ventilated – cytometers generate heat, a situation that is compounded further by additional refrigerators, centrifuges, as well as, perhaps, safety workbenches and incubators.

Factors affecting flow cytometric analysis

Each step in the analytical process includes factors affecting the precision and accuracy of measurement results. Tables 13 show a list of the various factors. The factors, variables and corresponding solutions of that list are explained in detail in the following chapters.

Table 1:

Pre-analytical variables influencing flow cytometric tests.

Influencing factorVariableProblem
Pre-analytics
 IndicationsAntibody panelCorrect question regarding present health problem?
Sample collection
 Sample materialSuch as peripheral blood, bone marrow, CSF, puncture fluidsProper and sufficient material collected for the question? Is the material representative?
 AnticoagulantEDTAa, heparin, citrate, ACDbLoss of vitality (EDTA), cell aggregation (heparin), dilution artifacts (ACD).
 Sample transport and sample storageDuration, temperature, vibration, sample preparation prior to analysis, e.g. sample-stabilizing reagentsDecreased vitality of the cells, selective cell loss, increased autofluorescence, cell aggregation.
 Sample preparationSuch as centrifugation, temperature, transferring, processing of solid tissues to cell suspensionsDecreased vitality of the cells, selective cell loss, increased autofluorescence, cell aggregation, cell contamination.

aEDTA, Ethylene diamine tetra-acetic acid; bACD, Acid Citrate Dextrose.

Table 2:

Analytical variables influencing flow cytometric tests.

Influencing factorVariableProblem
Staining
 PreparationFicoll vs. lysis, fixationDifferent scattering properties of lyses, pH instability of ammonium chloride lysis and selective loss of neutrophils, fixating lyses are more stable, single-stage (FACS-lysing solution) and two-stage lyses (VersaLyse), selective accumulation or loss of cells due to Ficoll.
Wash/no washWash: better signal-to-noise ratio, better represented populations, non-specific loss of cells, not well suited for absolute cell counts; No wash: no cell loss, higher fluorescence background, ideal for absolute cell counts, best in combination with pH-stable lysis.
Cell count adjustmentIdentical antibodies-cell ratio, reproducible fluorescence intensities.
 Antibody clonesEpitope, amount of antibodies used, direct or indirect immunofluorescencePolymorphism? Cross-reactivity? Non-specific binding? Saturation? Equilibrium? Signal amplification?
 FluorochromesType and number of fluorochromes used, antibody panelSpectral overlap? The more fluorochromes are used, the higher the spectral overlap of the individual fluorochromes to be noted, using bright fluorochromes (such as PE, APC, PC7) for weak antigens and vice versa, observing the staining indices (caution: not applicable to fluorochrome cocktails).
Choice of conjugatesDegradation in tandem dyes (e.g. generating a PE or APC signal which must then be compensated), increased sensitivity of tandem dyes to light, fixation (formaldehyde) and elevated temperatures.
Measurement
 CytometerSpecifications (lasers, filters), electronic and optical noise, standardization, calibrationSensitivity, linearity, spectral overlap
CompensationMinimization of spectral overlap ensured?
Trigger parameterRight choice of the threshold parameter (e.g. FSC, SSC, fluorescence, CD45)
Flow rate and dead timeLoss of measuring signals
 Quality controlsPopulation controls, internal negative and positive controlsDiscrimination of antibody binding (positive/negative), detection of autofluorescence, detection of non-specific Ab-binding compared to unstained control
External negative and positive controls (e.g. stabilized QC material, cell lines, healthy donors)Suitable process controls available for checking the lysis and/or staining process?
Isotype controlsUnwanted antibody binding via Fc receptors? Fluorochrome-induced binding?
Iso clone controlsFluorochrome-induced binding?
FMO controlsaCorrect determination of LOBb? Proper gating in activation markers?
Permeabilization controlsSuitable process controls for intracytoplasmic staining?
Data analysis
 GatingImmunological gating, ScatterGateClean identification of cell populations to be examined? Correct linking of gates?
Viability testingNon-specific binding, false positives
Duplicate discriminationIncorrect results especially in DNA analysis and in measurement of rare events
 EvaluationProgram usedSufficiently validated? Comparability given when using multiple programs?

aFMO, Fluorescence minus one. This is a gating control, which contains all the markers except the marker of interest. For low-density or “smeared” populations, such as activation markers, a FMO control allows for accurate gating of the positive cell population. FMO control is also useful in a multi-parameter setting. It provides information on the spectral overlap of the fluorochromes in an unstained channel and allows the correct definition of the negative cell population; bLimit of blank.

Table 3:

Post-analytical variables influencing flow cytometric tests.

Influencing factorVariableProblem
IT integrationUnidirectional vs. bidirectional, middleware, direct connection, no online connectionHigh error-proneness for manual entry of data, inflexibility, additional costs and high maintenance costs of the data in middleware solution.
Data managementSuch as listmode data, pdf-evaluation data, QC data, report of findingsManagement of data according to QC standards? Fast access to data possible? Storing medical reports in device software, middleware or laboratory software?
Specifying absolute cell counts, percentage data, relative intensities, MFIa, MESFb, ABScDual vs. single-platform, device calibrationError multiplication with dual-platform procedures, pipetting errors (pipetting of beads) with single-platform method, calibration: Comparability of equipment or laboratories given?
Reference intervals, cut-offsAge and gender of patients, pre-analysis (e.g. lysis), equipment specificityUse of correspondingly adapted reference intervals?
Design of the report of findingsPatient details, submission of original dot plots or histograms.Sufficient report of findings? Suitable for follow-up? Inclusion of data in a consolidated overall report of findings possible?

aMFI, Mean or median fluorescence intensity; bMESF, molecular equivalent soluble fluorochrome; cABS, antibody binding sites.

Standardization and harmonization in flow cytometry

When using modern flow cytometers and properly validated antibody panels, highly reproducible measurements are possible. This generally involves two strategies to ensure optimal analyses: standardization and harmonization. Standardization is about the attempt, from the outset, to achieve a high degree of agreement through given antibodies with defined markings [21], while the objective of harmonization is to render the results from many platforms comparable [22]. The best known standardization activity is the Euroflow Initiative [21]. For most laboratories that receive variety of requests and frequently set up new panels, however, harmonization remains the only feasible solution. The flow cytometer requires an instrument that measures multiple parameters simultaneously for the setup and calibration of different types of standards:

  • Reference standards;

  • Compensation standards;

  • Binding standards; and

  • Calibration standards.

Reference standards often consist of microbeads having a defined fluorescence that corresponds to the excitation and emission spectra of cells labeled with antibodies of the corresponding conjugate. Compensation standards are also microbeads that may be used for compensation instead of cells. Binding standards bind a defined quantity of antibodies and thereby allow for a comparison of staining intensity. Calibration standards allow for the comparability of measurements relating to staining intensity and thus receptor density or expression level.

An example of the use of reference or calibration standards is the determination of antigen expression densities with (semi-)quantification. The specification of antigen expression densities includes, among other factors, the problem of dependence on equipment due to the different compensation capabilities of devices and the dependence on the antibody clone as well as the antibody concentrations used. Without (semi-)quantification, there is no comparability between laboratories in multicenter studies. In addition, the calculated standard values can only be used internally. One possible strategy is internal calibration by use of reference antigens (fresh blood or commercial cell products) with a known and relatively stable number of surface antigens (e.g. CD4) and external calibration by use of hard dye beads, such as Cyto-Cal™ Multifluor-Beads or Rainbow™-Beads. In the latter method, the determination of instrument-specific MESF (Molecules of Equivalent Soluble Fluorochrome) values is made possible by different concentrations of the fluorescent dyes, that is, different levels of brightness, via a calibration curve. The number of bound fluorochrome molecules on cell surfaces (antibodies bound per cell, ABC), for example, can be determined by means of QuantiBRITE™-Beads. A prerequisite for both types of standardization, as for all other types of measurements, is the compensation of the spectral overlaps, which is achieved with spectrally matched reference beads (e.g. CaliBRITE™ beads). [23] and [24] provide good overviews of this.

Verification and validation in the flow cytometry laboratory

The terms verification and validation are central to the practice of quality assurance. Nevertheless, there is no uniform definition. According to DIN ISO 9000:2000, verification is the objective evidence that the requirements are met (“familiarization”). Validation under this norm is defined as the objective evidence that the requirements are met for a specific purpose [2]. These terms are also defined in the standard DIN EN ISO 15189 [25]. A more liberal interpretation might be: Verification and validation are both indispensable for process control. Both are independent procedures in the establishment, review, evaluation and documentation in the context of assessing whether a product or system complies with regulatory and/or technical standards. The “validation question” is: “Are you building the right thing?”, and the “verification question” is: “Are you building it right?” (see Table 4).

Table 4:

Verification and validation in the flow cytometry laboratory.

VerificationValidation
Confirmation of specifications of CE-IVD-marked reagents or test kitsEstablishment of IVD from own production (“in-house assays”, “home brewed assays”), modified standard methods, non-standard methods
PrecisionPrecision
Accuracy in at least two different concentration rangesAccuracy in at least two different concentration ranges
Measuring rangeReproducibility, including control of known relevant interference
Verification of the reference range provided by the manufacturerAnalytical and diagnostic specificity
Analytical and diagnostic sensitivity
Measuring range
Detection limits
Linearity

When a laboratory uses ready-to-use test kits and reagents marked IVD, it is possible to use them for diagnostics after a short familiarization phase. In flow cytometry, this is the case only for few, but often-used procedures, since most cases do not only involve prefabricated tests, but also their adaptation to the specific conditions at the laboratory. There are very good guidelines for implementing validation that comprise the most important elements [26]. The parameters that need to be collected, depending on the application, include the following entities relative to the type of test (also see Table 5):

  • Accuracy, secured morphologically in hematology; most of the time, ten replicates are measures, while qualitative methods use 20 [27, 28];

  • Specificity, the analytical specificity is mostly <5% [27]; clinical specificity is also important [28];

  • Sensitivity, also includes detection limit, depends on device settings, measurement conditions and antibody titration [27]; this is especially important in connection with the measurement of CD34 and MRD [28];

  • Precision, can be determined, for example, by way of peripheral blood, cell lines or blood standards and 20 replicates (recommended); for qualitative methods, at least three positive and negative ones each; a device comparison may also be required, which should come out to <10% or, worst case, <20% [27, 28];

  • Linearity, this is particularly relevant for rare events (MRD, fetal erythrocytes) or quantitative measurements (CD64) [28];

  • Stability, pre-analytical and post-processing stability are important to assess the robustness of the method; deviation should be under 20% [28].

  • Matrix influences must be taken into account by validating all methods for each matrix used separately [28].

  • Influences of anticoagulants on the stability must be validated, ideally on the basis of at least five samples at different times [28].

Table 5:

Examples of requirements for verification, validation and quality assurance in the flow cytometry laboratory, depending on the test method.

Quantitative determination of lymphocyte subpopulations (immune status)Quantification of CD34 stem cellsLeukemia and lymphoma immunophenotypingFunctional tests
Comparisons with the previous method (min. 10 samples)

Precision based on min. 20 replicates in the diagnostically important ranges considering different sample matrices (EDTA whole blood, BAL, etc.)

Stability testing for different sample matrices (EDTA whole blood, BAL, etc.) and anticoagulants if timely analysis cannot be guaranteed

Verification of the reference ranges provided by the manufacturer

Antibody titration or dilution experiments when departing from the manufacturer’s specifications or if no information exists
The validation must include the requirements of the manufacturer’s license; changes have to be notified to the client and the competent authority

Precision based on min. 20 replicates in the diagnostically important areas considering different sample matrices (cord blood, apheresate, etc.)

Determination of the detection limit, for example by way of dilution experiments and/or fluorescence beads

Determination of the linearity of the measuring range, for example by way of dilution experiments and/or fluorescence beads

Validation of the gating strategy, unless automated program is used

Stability testing for different sample matrices (cord blood, apheresate, etc.) and anticoagulants

Antibody titration or dilution experiments when departing from the manufacturer’s specifications or if no information exists
If necessary, comparisons with the previous method (min. 20 patient samples and different antibody panels); verification of accuracy, including morphological assessment

Stability testing for different sample matrices (EDTA whole blood, aspirates, etc.) and anticoagulants if timely analysis cannot be guaranteed

Antibody titration or dilution experiments when departing from the manufacturer’s specifications or if no information exists
Comparisons with the previous method (min. 10 samples)

Precision based on min. 10 replicates

Proof of sample stability, taking into account time, temperature and anticoagulant

Verification of the reference ranges provided by the manufacturer

Antibody titration or dilution experiments when departing from the manufacturer’s specifications or if no information exists
Daily internal quality controls in at least two concentration ranges for CD4 count, evaluation of manufacturer’s limits

Daily process controls (healthy donors)

Device comparison on a regular basis, if necessary
Daily internal quality controls in at least two concentration ranges, evaluation of manufacturer’s limits

Device comparison on a regular basis, if necessary
Daily internal quality controls if available, evaluation of manufacturer’s limits

Daily process controls (healthy donors) for intracytoplasmic antigens

Device comparison on a regukar basis, if necessary
Daily supply of appropriate process controls; clinical evaluation of controls

Device comparison on a regular basis, if necessary
Regular comparison of operators in the evaluation and clinical evaluation of samplesRegular comparison of operators in the evaluation and clinical evaluation of samples, unless automated program is usedRegular comparison of operators in the evaluation and clinical evaluation of samplesRegular comparison of operators in the evaluation and clinical evaluation of samples
Inter-laboratory tests according to RiliBÄK Part B2 Table B2-2; laboratory comparisons if necessaryParticipation in external quality assessment schemes, RiliBÄK-adapted, at least twice a yearParticipation in external quality assessment schemes, RiliBÄK-adapted, at least twice a yearInter-laboratory comparisons where possible, twice a year, with two samples each where no quality assessment schemes exist

Validation is mandatory each time a new or modified method is introduced at the laboratory. Records should be highly structured and follow the scheme of validation plan, measured data, and validation record with release.

Key factors for valid flow cytometry data

Several factors are crucial to reliable flow cytometry diagnostics. Regardless of the system used, they are always part of it. An overview of relevant quality assurance measures can be found in Nebe et al. [13].

All cytometers undergo a daily system check. In diagnostic equipment, this tends to be software-based. When fluorescence-marked beads are used, the lasers, detectors and correct compensation are checked.

Compensation between detectors today is almost always software-based, because a manual correction multi-color fluorescence is very difficult to implement in day-to-day practice.

Appropriate control materials must be included for the analyses performed, if available. In general, stabilized blood is used. The composition to be expected is known at least for the basic parameters. Each working day, at least two control samples with target values in different concentration ranges are to be analyzed.

The parameters of the immune status and of the determination of CD34 stem and progenitor cells must be evaluated continuously according to the stipulations of RiliBÄK and/or the standard. Given the short batch life of many control preparations (<12 weeks), manufacturer limits generally apply (control preparations, such as CD-Chex®, CD-Chex® Low, Immuno-Trol™ cells, Immuno-Trol™ Low cells, BD Multi-Check™ Control, BD Multi-Check™ CD4 Low Control, CD-Chex®CD34, BD™ Stem Cell Control, Stem Trol™ Control cells).

In hematologic samples for leukemia or lymphoma typing, this is possible only in few cases (CD Chex Plus, CD Chex BC, CD-Chex™ Select). Otherwise, control material will be replaced by plausibility checks to be performed within the analytical process.

The regular maintenance of equipment, compliance with all framework conditions of the analytical process (see below) and the processing by competent staff thus allow for the correct and clinically relevant analysis of the samples. In addition, the equipment used must be compared regularly [16].

The analytical process in the flow cytometry laboratory

The flow cytometry laboratory is typically integrated into the routine laboratory and interacts with the basic rules and structures, as well as work and analytical processes of other departments. The diagnostic process is comparable to that of other departments, but also has some special characteristics that are shown in Tables 13.

The selected anticoagulation matters in pre-analysis, and some tests have specific requirements, such as the calcium-based function test [29]. The samples must reach the laboratory within the time that is specified by the test manufacturers in the specification of their IVD. Generally, 24 h is the time frame that has proved practical for many immunophenotyping tests for pre-analysis purposes. Cell stabilizers may be used if necessary, but this must always be validated in each individual case. Identification of the sample is crucial to the correct allocation of the findings.

Also important in this context is the evaluation of samples on the basis of acceptance criteria. As DIN EN ISO 15189 states in 5.4.6, authorized staff must verify the obtained samples to ensure that they meet the relevant acceptance criteria for the required test(s) [3, 25]. According to the Immunology checklist, the identity and integrity of samples must be checked “appropriately”. Another requirement involves a set of written criteria for rejecting unacceptable samples or for the special treatment of suboptimal samples [4].

An adequate quantity of samples, that is, cell number, and a high vitality of cells are necessary for diagnostics. Today’s test procedures mostly involve whole blood without washing steps. Antibodies and fluorochromes are selected as part of the validation and corrected if necessary. Finally, it is the gating strategy that determines the correct measured results. Nowadays, the use of an absolute cell count measurement in the cytometer is recommended, that is, a single-platform technology [13].

Taking advantage of the age-based reference intervals, the flow cytometry findings report is then created in an indication-specific manner, but this requires access to the original plots [4].

External quality controls and laboratory comparisons

The implementation of external quality assurance in the form of external quality assessment schemes is mandatory in Germany, both under RiliBÄK [2] and DIN EN ISO 15189 [3, 25]. external quality assessment schemes or inter-laboratory tests serve as external quality and plausibility checks in comparison to the findings of other laboratories and specialist areas (inter-laboratory comparison). RiliBÄK makes participation in one external quality assessment scheme per quarter mandatory for the parameters in Table B1 a to c (RiliBÄK). If a user has exceeded the permissible deviation (column 5) of Table B1 a to c, he must, according to RiliBÄK, clarify the causes and – where he is capable of doing so – eliminate them. The flow cytometric parameters CD4 T-cells and CD8 T-cells are listed in RiliBÄK Table B 2-2 together with the manual differential blood count under qualitative laboratory tests. External quality assessment schemes are required every 6 months, or every quarter in the case of blood smears. The need for a new classification within RiliBÄK was already emphasized under “Guidelines of the German Medical Association (“RiliBÄK”)” [13–15].

Since there is no recognized reference method in the field of immunophenotyping, an evidence-based foundation is to be used for the future definition of coefficients of variation used in internal and external quality control [10]. In the cited paper, the coefficients of variation (CV) of lymphocyte subpopulations from 14 laboratories were examined. One conclusion of the authors was that the CV for round-robin testing of CD4 could be set to <15% (normal CD4 concentration) or <20% (low CD4 concentration) [13]. Another conclusion of the authors was that the reference institutions would have to be told to establish reference methods and ensure greater transparency in the representation of the quality of external quality assessment schemes over the long term.

National and international proficiency testing (PT) organizations offer external quality assessment schemes for flow cytometric tests. Table 6 provides an overview of external quality controls currently available.

Table 6:

National and international proficiency testing institutions and external quality assessment schemes offered for test type flow cytometry. The table is not exhaustive.

Interlaboratory testTested parametersINSTAND e.VDGKL-RfBaUK-NEQASInter-lab. comparison
Lymphocyte typing whole bloodT-, B, CD4, CD8, NKXXX
Lymphocyte typing cellular productsT-, B, NKX
Immunodeficiency whole bloodT- and B-cell subpopulationsX
Stem and progenitor cells whole bloodCD34XXX
Leukemia and lymphoma immunophenotypingLeukemia and lymphoma cellsXX
PNH diagnostics EDTA bloodGPI-deficient cellsXX
HLA-B27 whole bloodXXX
Reticulocytes whole bloodReticulocytesXXX
FMH-diagnosticsNumber of fetal erythrocytesX
Platelet diagnosticsSuch as platelet antigens, platelet antibodiesX
EMA testSpherocytesX
Cellular functional testsSuch as proliferation, migration, phagocytosis, oxidative burstX

aRfB, Reference Institute of Bioanalysis.

The German PT organizations INSTAND e.V. and the Reference Institute for Bioanalysis are reference institutions named by the German Medical Association. Both institutions have been accredited by the German Accreditation Body (DAkkS) under DIN EN ISO/IEC 17043 as providers of external quality assessment schemes. The current test programs are available on the respective websites [30, 31]. The providers use different test materials (samples), such as fresh blood, stabilized blood, control material, listmode data on a CD and digital image material. This may be quite demanding in terms of logistics. Fresh blood samples for external quality testing are taken from donors the day before, divided into aliquots, labeled, and sent to the scheme participants by overnight courier. The duration of shipment must be documented and be traceable in case of complaints. The inter-laboratory test participants will be informed that the material should be measured as soon as possible upon arrival. A reordering of samples is not possible in such cases. The preparation and quality assurance of external quality assessment samples, as an entire process, is subject to quality control and has to be strictly documented.

As can be seen in Table 6, external quality assessment schemes are nearly covering the entire spectrum of flow cytometric analysis. Here, according to the current board resolutions of the sector committee on medical laboratory diagnostics, external quality control is still mandatory for test methods for which there are no national external quality assessment schemes [16]. This can either be in the form of an inter-laboratory comparison or, alternatively, can be performed as participation in an international external quality assessment trial [16]. It should be emphasized that external quality testing of equipment manufacturers and test providers cannot replace independent external quality assessment schemes. The frequency of conducting inter-laboratory comparisons should be based on the frequency of round-robin tests. If possible, an inter-laboratory comparison involving at least two samples (different cell concentrations, positive/negative, etc.) should be carried out twice a year [3, 16]. Like the results of round-robin tests, the results of inter-laboratory comparisons must be documented and evaluated. In the event of failure, appropriate measures of problem analysis must be taken [3].

Common quality problems

Improvements to the flow cytometer have led in recent years to devices that provide stable and reliable measurements. Every day that they are started up, they require quality checks, and document relevant quality information. Still, audits and quality circles encounter typical problems that can and should be avoided.

The following Tables 79 give an overview of the most common quality problems.

Table 7:

Common quality problems in pre-analysis.

Common quality problems in pre-analysisEffects/limitationsReference
There are processed samples for which there is no clinical information.The diagnosis concludes with a finding based on a suspected diagnosis.[2, 3]
No procedures for safe handling and proper transport/shipping were createdThe sample is inadequately packaged and/or has not been released or sent as a Category B sample.[3, 4]
There is no incoming inspection of the samples. Criteria for rejection of unacceptable samples are not defined.Creating invalid reports of findings by processing unacceptable sample materials.[2–4]
The processing of sub-optimal samples has not been clarified.The sample quality is not documented anywhere. The competent reporting physician cannot evaluate the result accordingly and/or restrict the validity of the findings.[2–4]
There is no information about the collection and transport conditions on the dispatch noteVitality of sample material is not determined, which may be necessary.[2–4]
Information on sampling time is missingFor sample material older than 24 h old or not meeting the time-based acceptance criteria of a specific test, the validity of the findings is not restricted by the physician.[2–4]
Table 8:

Common quality problems in analysis.

Common quality problems in analysisEffects/limitationsReference
Validation data missing for analysis performed.Not all valid results are generated.[2–4]
The vitality tests (e.g. 7-AAD, PI, trypan blue) are missing for materials older than 24 hours or not meeting the specific timing requirements for a particular test.Not all valid results are generated.[4]
No (suitable) control material is used. Control material from healthy donor as process control is not present.The method may be out of range. The results are not valid. Due to the mostly stabilized control materials, not all processes can be checked.[4]
There is no control available for intracytoplasmic staining based on which permeabilization would be checked.Permeabilization may have been unsuccessful; the results obtained are wrong.[4]
The gating strategy is not, or inadequately, documented in the SOPs.The results obtained depend on the operator.[4]
There is no uniform set of rules for setting the gate of the negative control and/or there are no FMO controlsa for activation markers.The results obtained depend on the operator.[4]
There are no staff comparisons regarding the evaluation modalities (gating, etc.).The results obtained depend on the operator.[4]
Round-robin tests or, alternatively, laboratory comparisons are not performed for all measured parameters.The results of the laboratory may not be comparable with those of other laboratories.[2–4]
Backup devices are not synchronized with each other.The results from different devices are not comparable. Follow-up checks on a patient are not valid.[4]

aFMO, Fluorescence minus one.

Table 9:

Common quality problems in post-analysis.

Common quality problems in post-analysisEffects/limitationsReference
No information on whether reference intervals used do apply to one’s own patients or not. There are no age-adjusted norm value ranges.Report of findings may be subject to wrong assessment.[3, 4]
Findings do not include information on the time when material was collected.Findings may be misinterpreted for time-critical materials.[2–4]

Correspondence: Alexandra Dorn-Beineke, Institut für Labordiagnostik und Hygiene, Helios Dr. Horst Schmidt Kliniken Wiesbaden, Ludwig-Erhard-Str. 100, 65199 Wiesbaden, Germany

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Haferlach T, Nebe CT. Thesenpapier des Arbeitskreises Laboratorium der DGHO – Spezielle Hämatologische Diagnostik. Mitgliederrundschreiben der DGHO Ausgabe 2014;3:8–9. Available at: http://www.dgho.de/Informationenrundschreiben. Accessed 25 August 2014.Suche in Google Scholar

2. Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Available at: http://www.bundesaerztekammer.de/fileadmin/user_upload/downloads/Rili-BAEK-Laboratoriumsmedizin.pdf. Accessed: 5 August 2015.Suche in Google Scholar

3. Checkliste zur DIN EN ISO 15189:2014 für medizinische Laboratorien. Available at: http://www.dakks.de/content/checklistezur-din-en-iso-151892014-f%C3%BCr-medizinische-laboratorien,Dokument 72 CL 001.3_15189–2014/Rev.1.2/4.12.2014.doc. Accessed: 5 August 2015.Suche in Google Scholar

4. Checkliste Immunologie. Available at: http://www.dakks.de/content/checkliste-immunologie,Dokument75CL3001.pdf. Accessed: 21 January 2014.Suche in Google Scholar

5. Qualitätsmanagement DIN EN ISO 9000ff. Available at: http://www.beuth.de/cn/J-CGUQCWFVJVWGJODNHO3N6WOA.3/d29ya2Zsb3duYW1lPWV4YUJhc2ljU2VhcmNoJnJlZj10cGwtaG-9tZSZsYW5ndWFnZWlkPWRl.html. Accessed: 19 January 2014.Suche in Google Scholar

6. Schüttelpelz K. Auswirkung der Akkreditierung von Prüflaboratorien und Zertifizierungsstellen von Produkten in Deutschland. Dissertation, 2003. Available at: http://opus4.kobv.de/opus4-tuberlin/solrsearch/index/search/searchtype/simple/start/0/rows/10/query/Katrin+Sch%C3%BCttpelz/sortfield/score/sortorder/desc,DokumentSchuettpelz_Katrin.pdf. Accessed: 19 January 2014.Suche in Google Scholar

7. DFG: Sicherung guter wissenschaftlicher Praxis. Available at: http://www.dfg.de/download/pdf/dfg_im_profil/reden_stellungnahmen/download/empfehlung_wiss_praxis_1310.pdf. Accessed: 22 August 2014.Suche in Google Scholar

8. Minimum information about a flow cytometry experiment (MIFlowCyt) checklist (Numbered in accordance with MIFlowCyt 1.0 document). Cytometry A 2010;77:813.10.1002/cyto.a.20941Suche in Google Scholar

9. Kramer DB, Xu S, Kesselheim AS. Regulation of medical devices in the United States and European Union. N Engl J Med 2012;366:848–55.10.1056/NEJMhle1113918Suche in Google Scholar

10. Bundesamt für Risikobewertung: Gute Laborpraxis (GLP). Available at: http://www.bfr.bund.de/de/gute_laborpraxis__glp_-258.html. Accessed: 22 August 2014.Suche in Google Scholar

11. Verordnung über das Errichten, Betreiben und Anwenden von Medizinprodukten (MPBetreibV). Available at: http://www.gesetze-im-internet.de/mpbetreibv/. Accessed: 13 January 2014.Suche in Google Scholar

12. Barnett D, Louzao R, Gambell P, De J, Oldaker T, Hanson CA; on behalf of ICSH/ICCS Working Group. Validation of cell-based fluorescence assays: practice guidelines from the ICSH and ICCS – Part IV – postanalytical considerations. Cytometry Part B 2013;84B:309–14.10.1002/cyto.b.21107Suche in Google Scholar

13. Nebe CT, Dorn-Beineke A, Braun P, Daniel V, Ilieva Z, Kuling G, et al. Messunsicherheit und Qualitätssicherung im Bereich der Immunphänotypisierung der Lymphozytensubpopulationen im peripheren Blut. J Lab Med 2013;37:233–50.10.1515/labmed-2013-0052Suche in Google Scholar

14. Grantama JW, Kraan J, Keeney M, Sutherland DR, Wood BL. Enumeration of immunological defined cell populations by flow cytometry; approved guideline, document H42–A2, 2nd ed. Wayne, NJ: Clinical and Laboratory Standards Institute, 2007.Suche in Google Scholar

15. Nicholson JK, Hearn TL, Cross GD, White MD. 1997 revised guidelines for performing CD4+ T-cell determinations in persons infected with human immunodeficiency virus (HIV). Recommendation and reports. Centers for Disease Control and Prevention, Atlanta, USA, Vol. 46, No. RR-2, 1997.Suche in Google Scholar

16. Gremienbeschlüsse für den Bereich Medizinische Laboratoriumsdiagnostik. Available at: http://www.dakks.de/sites/default/files/dokumente/71_sd_3_002_gremienbeschlüsse_medizin_20150828_v1.4.pdf. & Beschlüsse des Sektorkomitees Medizinische Laboratorien zu Anforderungen der DIN EN ISO 15189 an die Qualität und Kompetenz von Medizinischen Laboratorien. Available at: http://www.dakks.de/sites/default/files/dokumente/71_sd_3_025_beschluesse_sk_medlab_20150828_v1.2.pdf. Accessed: 28 August 2015.Suche in Google Scholar

17. Checkliste Immungenetik. Available at: http://www.dakks.de/content/checkliste-immungenetik,Dokument75CL3002.pdf. Accessed: 21 January 2014.Suche in Google Scholar

18. Checkliste Klinische Chemie und Hämatologie. Available at: https://www.zlg.de/medizinprodukte/dokumente/checklistenfuer-laboratorien.html,Dokument212_CL07_121030Stand10/2012. Accessed: 27 Oktober 2015.Suche in Google Scholar

19. College of American Pathologists (CAP). CAP Accreditation Program-Flow Cytometry Checklist. Available at: http://www.cap.org/apps/docs/laboratory_accreditation/checklists/new/flow_cytometry.pdf. Accessed 21 January 2014.Suche in Google Scholar

20. Sack U, Barnett D, Demirel GY, Fossat C, Fricke S, Kafassi N, et al. Accreditation of flow cytometry in Europe. Cytometry B Clin Cytom 2013;84:135–42.10.1002/cyto.b.21079Suche in Google Scholar

21. Kalina T, Flores-Montero J, van der Velden VH, Martin-Ayuso M, Bottcher S, Ritgen M, et al. EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocols. Leukemia 2012;26:1986–2010.10.1038/leu.2012.122Suche in Google Scholar

22. Sack U, Tarnok A. Harmonization of cytometry instrumentation and technologies. Cytometry A 2013;83:1055–6.10.1002/cyto.a.22415Suche in Google Scholar

23. Schwartz A, Marti GE, Poon R, Gratama JW, Fernandéz-Repollet E. Standardization in flow cytometry: a classification system of fluorescence standards used for flow cytometry. Cytometry 1998;33:106–14.10.1002/(SICI)1097-0320(19981001)33:2<106::AID-CYTO4>3.0.CO;2-HSuche in Google Scholar

24. Tárnok A, Lenz D, Mittag A. Qualitätskontrolle und Standardisierung in: Zelluläre Diagnostik. Sack U, Tárnok A, Rothe G (Hrsg.), Karger-Verlag, 1. Auflage, 2007, S. 157–77.10.1159/000097613Suche in Google Scholar

25. ISO 15189 Available at: http://www.beuth.de/.Suche in Google Scholar

26. Validation of cell-based fluorescence assays: practice guidelines from the International Council for Standardization of Haematology and International Clinical Cytometry Society. Cytometry B Clin Cytom 2013;84:281.10.1002/cyto.b.21103Suche in Google Scholar

27. Owens MA, Vall HG, Hurley AA, Wormsley SB. Validation and quality control of immunophenotyping in clinical flow cytometry. J Immunol Methods 2000;243:33–50.10.1016/S0022-1759(00)00226-XSuche in Google Scholar

28. Wood B, Jevremovic D, Bene MC, Yan M, Jacobs P, Litwin V. Validation of cell-based fluorescence assays: practice guidelines from the ICSH and I. Cytometry B Clin Cytom 2013;84:315–23.10.1002/cyto.b.21108Suche in Google Scholar PubMed

29. Davis BH, Dasgupta A, Kussick S, Han JY, Estrellado A. Validation of cell-based fluorescence assays: practice guidelines from the ICSH and I. Cytometry B Clin Cytom 2013;84:286–90.10.1002/cyto.b.21105Suche in Google Scholar PubMed

30. Ringversuchsprogramm INSTAND e.V. Available at: http://www.instandev.de/ringversuche/programm.html,DokumentProspekt_2015_de.pdf. Accessed: 28 August 2015.Suche in Google Scholar

31. Ringversuchsprogramm Referenzinstitut für Bioanalytik. Available at: http://rfb.bio/pdf/Terminplan_2015_de.pdf. Accessed 28 August 2015.Suche in Google Scholar


Article note:

This article is based on the second chapter in the book “Zelluläre Diagnostik und Therapie” (Cellular Diagnostics and Therapy), edited by Ulrich Sack, Walter de Gruyter 2015 ISBN 978-3-11-034406-6.

Original German online version at: http://www.degruyter.com/view/j/labm.2016.40.issue-2/labmed-2015-0083/labmed-2015-0083.xml?format=INT. The German article was translated by Compuscript Ltd. and authorized by the authors.


Received: 2016-2-9
Accepted: 2016-2-9
Published Online: 2016-3-9

©2016 by De Gruyter

This article is distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Heruntergeladen am 31.12.2025 von https://www.degruyterbrill.com/document/doi/10.1515/labmed-2016-0016/html
Button zum nach oben scrollen