Endoscopic filter fluorometer for detection of accumulation of Protoporphyrin IX to improve photodynamic diagnostic (PDD)

  • Axel Boese EMAIL logo , Alexander Wagner , Alfredo Illanes , Uwe Bernd Liehr , Johann Jakob Wendler and Michael Friebe
Published/Copyright: September 17, 2020

Abstract

Photodynamic diagnostic (PDD) is an optical enhancement option for the endoscope to support the detection of cancer, for example in the bladder. In real application PDD efficiency suffers due to the complex accumulation of the photosensitizing drug inside the tumor and the associated processes of heme syntheses to create the fluorescent components needed. To optimize the diagnostic outcome of PDD it would be helpful to predict the optimal time for diagnosis based on measurable precursors. In a previous cell study, we proposed a new filter fluorometer to image the accumulation of the precursors Coproporphyrin III (CP-III) and Uroporphyrin III (UP-III) that metabolize to Protoporphyrin IX (PP-IX) later. This accumulation process can be used to predict the optimal time slot for diagnostic imaging. Therefore, a new filter system was designed to distinguish between CP-III and PP-IX. In this work we tested this filter system in combination with a standard PDD endoscopic imaging system. Goal of this study was to prove the technical feasibility in a non-patient setup to prepare a later clinical study.

Introduction

Photodynamic diagnosis (PDD) is an enhancement approach used in endoscopic procedures to increase the detection rate of cancer cells. One application is the detection of non-muscle-invasive cancer (NMIBC) of the urinary bladder. Bladder cancer is the fourth most prevalent cancer for both sexes in Europe and North America [1]. Diagnosis only with withe light cystoscopy (WL) can be challenging. PDD can increase the detection rate of NMIBC [2] due to the visible fluorescence of 5-aminolevulinic acid (5-ALA) induced porphyrins. For PDD in the bladder an intravesical instillation of the ALA based pharmaceutical HEXVIX (Ipsen Pharma, France) is needed 1 h before the diagnostic procedure [3]. Metabolically active cells like cancer cells accumulate this fluorophore and convert it into Protoporphyrin IX (PP-IX). PP-IX emits red fluorescence when exposed to blue light and is used as marker in PDD.

For successful application of PDD in clinical routine timing is a major factor to achieve valuable results [4]. According to the instructions of use of HEXVIX, there is a time window between one and 3 h after instillation to perform the diagnosis [4]. But the accumulation process is individual leading sometimes to insufficient fluorescence intensity with a potentially large effect on specificity [5].

The metabolic process creates two more porphyrins as precursors, Coproporphyrin III (CP-III) and Uroporphyrin III (UP-III) that appear before the highest presence of PP-IX. Both have a fluorescence peak approx. 10 nm below the peak of PP-IX. These precursors can serve as indicators to predict the optimal time slot for PDD. This can lead to a higher efficacy of PDD.

This principle could already be shown on CP-III and PP-IX solutions in a prior work [6] and on cell models in unpublished work. Special optical filters were used during fluorescence imaging to separate the individual fluorescence of the precursors and PP-XI. To achieve that a complex and sensitive filter fluorometer was built [7]. In a next step the filter setup was redesigned following a methodical and user integrated approach [8] to allow application in a clinical setting [9]. This new filter system including two optical bandpass filters mounted into slider (FB620-10, FB630-10 Thorlabs, Germany) was already tested in a laboratory setting using a 405 nm laser light source to induce fluorescence. In this work we now describe the first tests of this new filter system in a clinical environment and in combination with the real PDD equipment as a preparation of a later clinical study on patients.

Methods

Test setup

The new filter system described in [9] can be placed between the endoscopic camera and the endoscopic optic. Standardized C-mount adapters are used to ensure system compatibility. The filter system includes a slider with a 620 nm filter to detect CP-III and UP-III and a second 630 nm filter to detect PP-IX. In the center of the slider an aperture allows the use of standard imaging without additional filter. The filter system is shown in Figure 1.

Figure 1: Filter system including slider with two filters to separate the fluorescence wavelength of Coproporphyrin III (CP-III) and Protoporphyrin IX (PP-IX).
Figure 1:

Filter system including slider with two filters to separate the fluorescence wavelength of Coproporphyrin III (CP-III) and Protoporphyrin IX (PP-IX).

A plastic box was used to create a dark test environment. Inside the box, test samples of CP-III and PP-IX can be placed on a holder. A hole in the side wall allows insertion of the endoscope optics.

This setup was placed inside a standard urology surgery room including an endoscopy and PDD system (Image 1, D-Light, KARL STORZ, Germany). A HYSTERO 4MM 30 DEG cystoscope in combination with an H3-P camera head (both KARL STORZ, Germany) were used for imaging. The test setup is shown in Figure 2.

Figure 2: Test setup–top: overall setup incl. Photodynamic diagnosis (PDD) system and box, down: details of the box setup.
Figure 2:

Test setup–top: overall setup incl. Photodynamic diagnosis (PDD) system and box, down: details of the box setup.

Testing

In a first test cycle a CP-III sample was placed inside the test box. The cystoscope connected to the imaging system was inserted into the test box from the side hole. The tip of the optic was placed in a distance of 20 mm in front of the test sample. White light images, PDD images and PDD images with filter one (F1) and filter two (F2) were acquired. The slider of the filter system was used to change between the acquisition modes. Then, the test sample was exchanged by a PP-IX sample and the procedure was repeated. All images were directly stored on the endoscopic imaging system.

Since the images of the first test cycle appeared quite dark we placed the camera and light cable directly inside the test box with the sample for the subsequent testing. With that the light intensity could be increased. The light source was placed in a distance of 20 mm from the sample and the camera in a distance on 100 mm. All acquired images were exported form the endoscopic system for later evaluation.

Post processing and evaluation

Following an evaluation approach described in [9] the images were analysed in ImageJ first. Since some of the images appear dark, a change in intensity of the fluorescence as expected was hard to see visually. Contrary to the priory used methods in a laboratory setting, a postprocessing was needed for evaluation. In a first step the Red-Green-Blue (RGB) channels were separated and subsequently only the red channel used for further processing. After adjusting the windowing level, a rectangular region of interest (ROI) of each image was selected and the intensity profile of the ROI was computed. The maximum intensity level was chosen for comparison.

Results

Eight images were acquired in the first test cycle using the cystoscope. The white light image was used for targeting, the PDD mode showed fluorescence of the used samples. The filtered images of the CP-III were dark for both filter types, but the sample showed still an observable fluorescence. At the filtered images of the PP-XI sample the fluorescence structure was not visible. For this reason, the postprocessing was applied as described before. After splitting the RGB channels and separating the red channel the sample structure could be identified in all images. Computation of the intensity profile allows a comparison of the fluorescence. As expected, the intensity was higher when using filter one for CP-III and higher when using filter two for PP-IX (Figures 3 and 4).

Figure 3: First test cycle, images of the CP-III sample with cystoscope. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.
Figure 3:

First test cycle, images of the CP-III sample with cystoscope. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.

Figure 4: First test cycle, images of the PP-IX sample with cystoscope. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.
Figure 4:

First test cycle, images of the PP-IX sample with cystoscope. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.

The filtered images of the second test cycle showed high fluorescence under all conditions. Nevertheless, the post processing was applied to achieve comparable results. Again, the intensity was higher when using filter one for CP-III and higher when using filter two for PP-IX (Figures 5 and 6).

Figure 5: Second test cycle, images of the CP-III sample. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.
Figure 5:

Second test cycle, images of the CP-III sample. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.

Figure 6: Second test cycle, images of the PP-IX sample. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.
Figure 6:

Second test cycle, images of the PP-IX sample. White light mode, PDD mode, filter 1, filter 2, postprocessed and intensity profile.

Discussion

Our tests could demonstrate the feasibility of our filter system to distinguish between CP-III and PP-IX. The light intensity of a standard PDD light source is lower compared to the laser system we have used before [9]. Even without an endoscopic optic the light is still sufficient to detect the differences in fluorescence. When the cystoscope is used, the light intensity is reduced dramatically. This leads to a low fluorescence of the samples. The light has to pass the cystoscope and the fluorescent emission has to pass the way back through the cystoscope again. This leads to a reduction of intensity resulting in dark images. Nevertheless, we were able to detect the sample structure and analyse the intensity by a simple postprocessing. For preparation of a clinical study we now have to prove feasibility in a filled artificial bladder under more real conditions.


Corresponding author: Axel Boese, Otto-von-Guericke University Magdeburg, Medical Faculty, Leipziger Str. 44, Magdeburg, Germany, E-mail:

  1. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  2. Research funding: The author state no funding involved.

  3. Conflict of interest: Authors state no conflict of interest.

  4. Informed consent: Informed consent was obtained from all individuals included in this study.

References

1. GLOBOCAN 2018 online data cancer today. Available from: https://gco.iarc.fr/today/home.Search in Google Scholar

2. Lipiński, MI, Różański, W, Markowski, MP. Photodynamic diagnosis–current tool in diagnosis of carcinoma in situ of the urinary bladder. Contemp Oncol 2015;19:341–2. https://doi.org/10.5114/wo.2015.54391.Search in Google Scholar

3. Karl storz bluelight cystoscopy with cystview. Available from: https://www.karlstorz.com/cps/rde/xbcr/karlstorz_assets/ASSETS/3557592.pdf.Search in Google Scholar

4. Tips and tricks Hexvix®. Available from: https://www.hexvix.com/tips-tricks.Search in Google Scholar

5. McNicholas, K, MacGregor, MN, Gleadle, JM. In order for the light to shine so brightly, the darkness must be present—why do cancers fluoresce with 5-aminolaevulinic acid? Br J Cancer 2019;121:631–9. https://doi.org/10.1038/s41416-019-0516-4.Search in Google Scholar

6. Landes, R, Illanes, A, Goeppner, D, Gollnick, H, Friebe, M. A study of concentration changes of Protoporphyrin IX and Coproporphyrin III in mixed samples mimicking conditions inside cancer cells for photodynamic therapy. PLoS One 2018;13:e0202349. https://doi.org/10.1371/journal.pone.0202349.Search in Google Scholar

7. Landes, R, Illanes, A, van Oepen, A, Goeppner, D, Gollnick, H, Friebe, M. Fiber-optic filter fluorometer for emission detection of Protoporphyrin IX and its direct precursors–a preliminary study for improved photodynamic therapy applications. Results in Physics 2018;8:1232–3. https://doi.org/10.1016/j.rinp.2018.01.059.Search in Google Scholar

8. Boese, A. Lösungsfindung mit dem endnutzer, ein neuer ansatz in der methodischen produktentwicklung am beispiel der medizintechnik. Aachen, Germany: Shaker; 2016. ISBN 978-3-8440-4947-3.Search in Google Scholar

9. Boese, A, Wagner, A, Liehr, UB, Wendler, J, Friebe, M. A new filter system for endoscopic fluorescence detection of Protoporphyrin IX and its direct precursors in PDT and PDD. Leipzig, Germany: 54th BMT 2020; 2020.10.1515/cdbme-2020-3150Search in Google Scholar

Published Online: 2020-09-17

© 2020 Axel Boese et al., published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Proceedings Papers
  2. 4D spatio-temporal convolutional networks for object position estimation in OCT volumes
  3. A convolutional neural network with a two-stage LSTM model for tool presence detection in laparoscopic videos
  4. A novel calibration phantom for combining echocardiography with electromagnetic tracking
  5. Domain gap in adapting self-supervised depth estimation methods for stereo-endoscopy
  6. Automatic generation of checklists from business process model and notation (BPMN) models for surgical assist systems
  7. Automatic stent and catheter marker detection in X-ray fluoroscopy using adaptive thresholding and classification
  8. Autonomous guidewire navigation in a two dimensional vascular phantom
  9. Cardiac radiomics: an interactive approach for 4D data exploration
  10. Catalogue of hazards: a fundamental part for the safe design of surgical robots
  11. Catheter pose-dependent virtual angioscopy images for endovascular aortic repair: validation with a video graphics array (VGA) camera
  12. Cinemanography: fusing manometric and cinematographic data to facilitate diagnostics of dysphagia
  13. Comparison of spectral characteristics in human and pig biliary system with hyperspectral imaging (HSI)
  14. COMPASS: localization in laparoscopic visceral surgery
  15. Conceptual design of force reflection control for teleoperated bone surgery
  16. Data augmentation for computed tomography angiography via synthetic image generation and neural domain adaptation
  17. Deep learning for semantic segmentation of organs and tissues in laparoscopic surgery
  18. DL-based segmentation of endoscopic scenes for mitral valve repair
  19. Endoscopic filter fluorometer for detection of accumulation of Protoporphyrin IX to improve photodynamic diagnostic (PDD)
  20. EyeRobot: enabling telemedicine using a robot arm and a head-mounted display
  21. Fluoroscopy-guided robotic biopsy intervention system
  22. Force effects on anatomical structures in transoral surgery − videolaryngoscopic prototype vs. conventional direct microlaryngoscopy
  23. Force estimation from 4D OCT data in a human tumor xenograft mouse model
  24. Frequency and average gray-level information for thermal ablation status in ultrasound B-Mode sequences
  25. Generalization of spatio-temporal deep learning for vision-based force estimation
  26. Guided capture of 3-D Ultrasound data and semiautomatic navigation using a mechatronic support arm system
  27. Improving endoscopic smoke detection with semi-supervised noisy student models
  28. Infrared marker tracking with the HoloLens for neurosurgical interventions
  29. Intraventricular flow features and cardiac mechano-energetics after mitral valve interventions – feasibility of an isolated heart model
  30. Localization of endovascular tools in X-ray images using a motorized C-arm: visualization on HoloLens
  31. Multicriterial CNN based beam generation for robotic radiosurgery of the prostate
  32. Needle placement accuracy in CT-guided robotic post mortem biopsy
  33. New insights in diagnostic laparoscopy
  34. Robotized ultrasound imaging of the peripheral arteries – a phantom study
  35. Segmentation of the distal femur in ultrasound images
  36. Shrinking tube mesh: combined mesh generation and smoothing for pathologic vessels
  37. Surgical audio information as base for haptic feedback in robotic-assisted procedures
  38. Surgical phase recognition by learning phase transitions
  39. Target tracking accuracy and latency with different 4D ultrasound systems – a robotic phantom study
  40. Towards automated correction of brain shift using deep deformable magnetic resonance imaging-intraoperative ultrasound (MRI-iUS) registration
  41. Training of patient handover in virtual reality
  42. Using formal ontology for the representation of morphological properties of anatomical structures in endoscopic surgery
  43. Using position-based dynamics to simulate deformation in aortic valve replacement procedure
  44. VertiGo – a pilot project in nystagmus detection via webcam
  45. Visual guidance for auditory brainstem implantation with modular software design
  46. Wall enhancement segmentation for intracranial aneurysm
Downloaded on 9.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2020-0018/html
Scroll to top button