Automatic stent and catheter marker detection in X-ray fluoroscopy using adaptive thresholding and classification

  • Negar Chabi EMAIL logo , Oliver Beuing , Bernhard Preim und Sylvia Saalfeld
Veröffentlicht/Copyright: 17. September 2020
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

In this study, we propose a method for marker detection in X-ray fluoroscopy sequences based on adaptive thresholding and classification. Adaptive thresholding yields multiple marker candidates. To remove non-marker areas, 24 specific features are extracted from each extracted patch and four supervised classifiers are trained to differentiate non-marker areas from marker areas. Quantitative evaluation was carried out to assess different classifier performance by calculating accuracy, sensitivity, specificity and precision. SVM outperforms other classifiers based on the mean value for accuracy, specificity and precision with 81.56, 91.94 and 84.21%, respectively.

Introduction

Intracranial stenting is one of the minimally-invasive therapies for neurovascular pathologies such as intracranial aneurysms and is accomplished via continuous X-ray imaging. Intracranial stents and flow diverters are small flexible mesh tube devices. Precise real-time device placement depends on the proper intra-operative visualization of the stent. Although radio-opaque stent markers are visible in fluoroscopy, they are as small as possible and their detection is challenging.

Sabrina et al. [1] proposed a technique to detect stent-graft markers to support the physicians with the stent-graft deployment in intra-operative fluoroscopic images. Fast radial symmetry transform following an ellipse or line fitting approach was used to detect a group of markers.

Movassaghi et al. [2] proposed a 3D representation of the deployed stent during a percutaneous transluminal coronary angioplasty procedure. Markers on the balloon catheter were utilized for motion compensation. Here the markers are clearly visible because of the high absorption coefficients and are detected using a scale-space algorithm.

Bismuth et al. [3] presented a technique to improve the stent visualization on an X-ray image sequence. Stent motion is inferred based on the motion of balloon marker balls as well as a guide wire. Potential marker balls are detected by local minima selection and a priori knowledge.

Wang et al. [4] proposed a fully balloon marker pair tracking and guide wire localization to employ 3D motion compensation and stent reconstruction. Marker candidates were detected using a probabilistic boosting tree classifier.

Many studies used blob-like structure detectors to detect markers [4], [5], [6]. Blob analysis actually refers to methods aiming at detecting regions with different characteristics, such as brightness or color, compared to their neighborhood.

Many approaches for stent marker detection are not applicable to intracranial stents because of different properties. Most studies focus on coronary stent enhancement (e.g., balloon-mounted stent with a couple of radio-opaque markers connected by a guide wire [3], [4], [5], [6]). This specific geometry is used to remove false positive candidate markers [6]. However, we are dealing with a type of intracranial stent, e.g., self-expanding stents with a different number of radio-opaque markers, which are not connected by a specific landmark.

In this paper, we focus on the X-ray opaque marker detection in intracranial fluoroscopic images. We have implemented a multi-step marker detection algorithm for fluoroscopy sequences based on an adaptive thresholding technique and binary classification.

We use blob analysis as feeding features to train classifiers. Different classifiers include linear discriminant analysis classifier (LDA), naïve Bayesian classifier (NB), k-nearest neighbors (KNN) and support vector machine (SVM) which will be described in the next section.

Methodology

We develop an algorithm to detect radio-opaque markers located on intracranial stent and catheter.

Data acquisition

Seven clinical X-ray fluoroscopy sequences were used in this study. Images were acquired at the Institute for Neuroradiology, Magdeburg, Germany with an Artis Q (Siemens Healthineers, Forchheim, Germany). The image size is 512×512 pixels. Frame rates are between 8 and 30 fps.

Different types of stents were used for treatment, for example the DERIVO flow diverter (ACANDIS GmbH Co. KG, Pforzheim, Germany), a self-expanding stent with 15–80 mm length in the vessel and applicable for 2.5–6 mm vessel diameter.

Pre-processing and patch extraction

To enhance the visibility of the stent markers, a simple pre-processing technique was employed which is described in our previous work [7]. Afterward, adaptive thresholding [8] was employed to identify marker areas. The sensitivity is a parameter ranged in [0-1]. Higher sensitivity equals a higher number of detected regions as foreground. It was empirically selected in [0.4-0.5] for different frames. Patch extraction includes detecting components connecting in the diagonal, vertical, and horizontal direction. Then, corresponding patches were extracted from the original frame and its corresponding pre-processed frame. Patch size is empirically selected 15×15 to cover even very big markers such as catheter markers. Since other structures, which are not of interest, such as guide wire or catheter, are also detected as foreground, eccentricity and area were used to get rid of them. Detected regions with high-eccentricity were removed using 0.98 as a threshold. This threshold was selected empirically. Regions larger than a threshold (e.g., coil) were removed. The threshold was empirically chosen based on the marker size (bigger than the maximum marker size), estimated radiographic magnification factor and image pixel spacing.

Feature extraction

Due to the blob-like appearance of the desired regions, we employ different blob analysis algorithms to train different classifiers against false positives. For each patch with a size of 15 × 15, we extract 22 features, which include minimum, average, and standard deviation, as well as the following features.

Laplacian of Gaussian (LoG) computes the second derivative of a function and detects abruptly changing regions. The filter size is empirically selected 15 and the standard deviation is the default value. Then, mean and standard deviation of LoG were fed in the classifiers.

Difference of Gaussian (DoG) can be used to detect edges. The mean and standard deviation of DoG were computed. The standard deviation ratio between two Gaussian was selected 2, and the filter sizes were selected 10 times of their standard deviation. These values were empirically chosen.

Determinant of Hessian (DoH) detects blobs in an image. The Hessian matrix contains the second partial derivatives of a function. DoH was calculated in two different ways: one is the second derivative of the input and the other one is to filter the input with the second derivative of a Gaussian filter with standard deviation at its default value [9]. Then, mean and standard deviation of the computed DoH matrix for patches extracted from both pre-processed and original frame are considered as input features.

Classification

Different classifiers including k-nearest neighbor (KNN), naive Bayesian classifier (NB), support vector machine (SVM) and linear discriminant analysis (LDA) classifier were trained and used to differentiate patches with and without markers.

Experiments and results

The framework for our approach is illustrated in Figure 1. For training and test, we used four and three clinical fluoroscopy sequences, respectively. To detect true X-ray markers, classifiers are trained to discriminate marker regions from non-marker regions. Features were extracted from 112 patches of some arbitrary frames of 4 clinical fluoroscopy sequences to train the classifiers. Different classification techniques were employed to differentiate patches. SVM classification was performed by standardizing the features, and sequential minimal optimization (SMO) was chosen as an optimization routine. The radial basis function was selected as a kernel function. For the KNN classifier, the distance metric was Euclidean distance and the number of nearest neighbor (K) was 1. Both LDA and NB classifiers were trained in their default mode in MATLAB (MathWorks, Natick, USA). To evaluate the performance of the approach, 1948 test patches were extracted from various frames of 3 different fluoroscopy sequences. The number of true positive (TP), true negative (TN), false positive (FP) and false negative (FN) patches were calculated for each test frame. TP and TN values indicate the number of patches that were truly classified as patches with markers and patches excluding markers (including other objects other than radio-opaque markers), respectively. Different criteria including sensitivity, specificity and precision were used to assess the performance of different classifiers.

Figure 1: Workflow for the proposed approach.
Figure 1:

Workflow for the proposed approach.

The results are illustrated in Figure 2 for the first test data set with the four classifiers: KNN, NB, SVM and LDA. Figure 2a shows the original input frame, stent and catheter are specified on the image. Figure 2b is a pre-processed frame that is used for thresholding. X-ray opaque markers can be better visualized on the pre-processed image. Figure 2c, shows the candidate regions for markers that are surrounded by a green circle and are shown as white regions within the image. Finally, Figure 2(d–f) shows prediction results for KNN, NB, SVM and LDA classifier, respectively. Regions marked with green and red stars illustrate predicted marker locations and predicted non-marker areas, respectively. In this specific frame, KNN predicts two markers truly, while it misses 5 other markers. NB has the best performance for this specific frame by 6 TP, 1 FN and 1 FP. SVM predicts 3 markers truly while it misclassified 4 other patches (4 FN). LDA also misclassified three marker patches.

Figure 2: The results of classifications for the first test dataset, a. Original input image, b. pre-processed image, c. Detected areas after thresholding as possible marker candidates, Classification result by d. KNN, e. NB, f. SVM, g. LDA.
Figure 2:

The results of classifications for the first test dataset, a. Original input image, b. pre-processed image, c. Detected areas after thresholding as possible marker candidates, Classification result by d. KNN, e. NB, f. SVM, g. LDA.

The performance of classifiers on 3 test sequences was compared using different criteria (See Table 1). Based on the accuracy results, the NB classifier performs the best on the first test sequence with 72.29%. In the second test sequence, SVM has the highest accuracy of 85.83%. In the last sequence, LDA and SVM have almost the same accuracy value with 97.61 and 96.97%, respectively. Based on the sensitivity value for different classifiers, it concluded that the NB classifier has a lower number of false negatives. NB has the highest mean sensitivity value with 87.76% and in general based on accuracy, specificity and precision SVM works the best and its average values are 81.56, 91.94 and 84.21%, respectively.

Table 1:

Comparative results of accuracy, sensitivity, specificity and precision values for KNN, LDA, NB and SVM classifier on three different clinical datasets, width = 1.

DatasetAccuracySensitivitySpecificityPrecision
Test_1Test_2Test_3MeanTest_1Test_2Test_3MeanTest_1Test_2Test_3MeanTest_1Test_2Test_3Mean
KNN56.70%74.42%76.40%69.17%39.86%73.58%83.33%65.59%86.74%74.84%76.07%79.21%84.28%59.09%14.45%52.60%
NB72.29%79.26%69.08%73.54%72.97%92%98.33%87.76%71.08%73.16%67.61%70.61%81.81%62.16%13.22%52.39%
SVM61.90%85.83%96.97%81.56%47.97%77.84%53.33%59.71%86.74%89.93%99.16%91.94%86.57%89.87%76.19%84.21%
LDA63.63%77.87%97.61%79.70%55.03%73.85%60%62.96%79.26%79.87%99.49%84.10%82.82%64.57%85.71%77.7%
  1. The highest mean value for accuracy, sensitivity, specificity and precision is highlighted in bold.

Conclusion

Markers are very small and hardly visible in 2D X-ray fluoroscopy and our proposed approach can be used to assist clinicians during the intervention with the stent location in 2D by highlighting the detected markers on X-ray images. This eventually leads to faster procedures with less dose exposure to both patients and clinicians. Furthermore, detected markers may be helpful in improving stent visibility in 2D. The adaptive thresholding method used in this study can detect even very small and hardly visible markers, however, to detect all possible candidates, it still depends on the specified sensitivity. Increasing the sensitivity will increase the number of false positives that have to be ruled out in the classification step. This emphasizes the importance of the classification step. Different classifiers including KNN, NB, SVM and LDA were used as a binary classifier to discriminate between non-marker areas and marker regions. Based on the evaluations, SVM outperforms other classifiers in terms of accuracy, specificity and precision. The highest specificity for SVM with 91.94% was obtained at the cost of low sensitivity. Furthermore, based on the obtained accuracy values for different classifiers, it can be concluded that the performance of the classifiers depends on the dataset. Possible improvements include increasing the number of training patches and increasing features and employing cross-validation to improve the classifiers’ generalizability. Since stent and catheter markers are different in terms of size and shape, employing multi-class classification and considering different types of markers may improve the performance of the classifiers.


Corresponding author: Sylvia Saalfeld, Department of Simulation and Graphics, Forschungscampus STIMULATE, OvG-University, Magdeburg, Germany, E-mail:

  1. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  2. Research funding: This work is partly funded by the German Research Foundation (SA3461/2-1) and the Federal Ministry of Education and Research within the Forschungscampus STIMULATE (13GW0095A) as well as partly conducted within the International Graduate School MEMoRIAL at Otto von Guericke University (OVGU) Magdeburg, Germany, which is kindly supported by the European Structural and Investment Funds (ESF) under the programme ”Sachsen-Anhalt WISSENSCHAFT Internationalisierung” (project no. ZS/2016/08/80646).

  3. Conflict of interest: Authors state no conflict of interest.

  4. Informed consent: Informed consent is not applicable.

  5. Ethical approval: The conducted research is not related to either human or animals use.

References

1. Reiml, S, Pfister, M, Toth, D, Maier, A, Hoffmann, M, Kowarschik, M, et al. Automatic detection of stent graft markers in 2d fluoroscopy images. In: Joint MICCAI workshops on CVII-STENT; 2015.Suche in Google Scholar

2. Movassaghi, B, Schaefer, D, Grass, M, Rasche, V, Wink, O, Garcia, JA, et al. 3d reconstruction of coronary stents in vivo based on motion compensated x-ray angiograms. In: International conference on medical image computing and computer-assisted intervention: Springer; 2006:177–84 pp.10.1007/11866763_22Suche in Google Scholar PubMed

3. Bismuth, V, Vaillant, R, Funck, F, Guillard, N, Najman, L. A comprehensive study of stent visualization enhancement in x-ray images by image processing means. Med Image Anal 2011;15:565–76. https://doi.org/10.1016/j.media.2011.03.002.Suche in Google Scholar

4. Wang, Y, Chen, T, Wang, P, Rohkohl, C, Comaniciu, D. Automatic localization of balloon markers and guidewire in rotational fluoroscopy with application to 3d stent reconstruction. In: European conference on computer vision: Springer; 2012:428–41 pp.10.1007/978-3-642-33783-3_31Suche in Google Scholar

5. Schoonenberg, G, Lelong, P, Florent, R, Wink, O, ter Haar Romeny, B. The effect of automated marker detection on in vivo volumetric stent reconstruction. In: International conference on medical image computing and computer-assisted intervention: Springer; 2008:87–94 pp.10.1007/978-3-540-85990-1_11Suche in Google Scholar PubMed

6. Lu, X, Chen, T, Comaniciu, D. Robust discriminative wire structure modeling with application to stent enhancement in fluoroscopy. In: CVPR 2011: IEEE; 2011:1121–7 pp.10.1109/CVPR.2011.5995714Suche in Google Scholar

7. Chabi, N, Beuing, O, Preim, B, Saalfeld, S. Comparison of background removal approaches in x-ray fluoroscopy for detection of cerebral stent markers. In: 18. Jahrestagung der Deutschen Gesellschaft für Computer- und Roboterassistierte Chirurgie e.V; 2019:122–7 pp.Suche in Google Scholar

8. Bradley, D, Roth, G. Adaptive thresholding using the integral image. J Graph Tool 2007;12:13–21. https://doi.org/10.1080/2151237x.2007.10129236.Suche in Google Scholar

9. Kroon, DJ, Schrijver, M. Hessian based frangi vesselness filter: MATLAB File Exchange; 2009:24409.Suche in Google Scholar

Published Online: 2020-09-17

© 2020 Negar Chabi et al., published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Artikel in diesem Heft

  1. Proceedings Papers
  2. 4D spatio-temporal convolutional networks for object position estimation in OCT volumes
  3. A convolutional neural network with a two-stage LSTM model for tool presence detection in laparoscopic videos
  4. A novel calibration phantom for combining echocardiography with electromagnetic tracking
  5. Domain gap in adapting self-supervised depth estimation methods for stereo-endoscopy
  6. Automatic generation of checklists from business process model and notation (BPMN) models for surgical assist systems
  7. Automatic stent and catheter marker detection in X-ray fluoroscopy using adaptive thresholding and classification
  8. Autonomous guidewire navigation in a two dimensional vascular phantom
  9. Cardiac radiomics: an interactive approach for 4D data exploration
  10. Catalogue of hazards: a fundamental part for the safe design of surgical robots
  11. Catheter pose-dependent virtual angioscopy images for endovascular aortic repair: validation with a video graphics array (VGA) camera
  12. Cinemanography: fusing manometric and cinematographic data to facilitate diagnostics of dysphagia
  13. Comparison of spectral characteristics in human and pig biliary system with hyperspectral imaging (HSI)
  14. COMPASS: localization in laparoscopic visceral surgery
  15. Conceptual design of force reflection control for teleoperated bone surgery
  16. Data augmentation for computed tomography angiography via synthetic image generation and neural domain adaptation
  17. Deep learning for semantic segmentation of organs and tissues in laparoscopic surgery
  18. DL-based segmentation of endoscopic scenes for mitral valve repair
  19. Endoscopic filter fluorometer for detection of accumulation of Protoporphyrin IX to improve photodynamic diagnostic (PDD)
  20. EyeRobot: enabling telemedicine using a robot arm and a head-mounted display
  21. Fluoroscopy-guided robotic biopsy intervention system
  22. Force effects on anatomical structures in transoral surgery − videolaryngoscopic prototype vs. conventional direct microlaryngoscopy
  23. Force estimation from 4D OCT data in a human tumor xenograft mouse model
  24. Frequency and average gray-level information for thermal ablation status in ultrasound B-Mode sequences
  25. Generalization of spatio-temporal deep learning for vision-based force estimation
  26. Guided capture of 3-D Ultrasound data and semiautomatic navigation using a mechatronic support arm system
  27. Improving endoscopic smoke detection with semi-supervised noisy student models
  28. Infrared marker tracking with the HoloLens for neurosurgical interventions
  29. Intraventricular flow features and cardiac mechano-energetics after mitral valve interventions – feasibility of an isolated heart model
  30. Localization of endovascular tools in X-ray images using a motorized C-arm: visualization on HoloLens
  31. Multicriterial CNN based beam generation for robotic radiosurgery of the prostate
  32. Needle placement accuracy in CT-guided robotic post mortem biopsy
  33. New insights in diagnostic laparoscopy
  34. Robotized ultrasound imaging of the peripheral arteries – a phantom study
  35. Segmentation of the distal femur in ultrasound images
  36. Shrinking tube mesh: combined mesh generation and smoothing for pathologic vessels
  37. Surgical audio information as base for haptic feedback in robotic-assisted procedures
  38. Surgical phase recognition by learning phase transitions
  39. Target tracking accuracy and latency with different 4D ultrasound systems – a robotic phantom study
  40. Towards automated correction of brain shift using deep deformable magnetic resonance imaging-intraoperative ultrasound (MRI-iUS) registration
  41. Training of patient handover in virtual reality
  42. Using formal ontology for the representation of morphological properties of anatomical structures in endoscopic surgery
  43. Using position-based dynamics to simulate deformation in aortic valve replacement procedure
  44. VertiGo – a pilot project in nystagmus detection via webcam
  45. Visual guidance for auditory brainstem implantation with modular software design
  46. Wall enhancement segmentation for intracranial aneurysm
Heruntergeladen am 12.9.2025 von https://www.degruyterbrill.com/document/doi/10.1515/cdbme-2020-0006/html
Button zum nach oben scrollen