Abstract
Background: Many papers evaluating high sensitivity troponin assays make the diagnosis of myocardial infarction based on conventional troponin assays in clinical use at the time of recruitment. Such analyses often do not show superiority of high sensitivity assays compared with contemporary assays meeting precision guidelines.
Methods: Three hundred and twenty-two patients presenting to the emergency department between November 2006 and April 2007 for evaluation for acute coronary syndrome had serial (0 h and >6 h) bloods taken to compare troponin assays (Roche hsTnT, Abbott TnI, Roche TnT and Vitros TnI). The diagnosis of myocardial infarction was made using each troponin assay separately with which that same assay was analysed for diagnostic performance.
Results: The rate of myocardial infarction would be 38.9% using serial hsTnT, 31.3% using serial Abbott TnI, 27.1% using serial TnT and 26.4% using serial Vitros TnI. The baseline sensitivities (0 h) are 89.9% (85.2–93.3) for hsTnT, 77.9% (71.0–87.5) for Abbott TnI, 73.0% (65.6–78.7) for TnT and 86.8% (74.6–94.4%) for Vitros TnI. The specificities (peak 0 h and >6 h samples) are 93.1% (91.2–93.1) for hsTnT, 88.3% (86.5–88.3) for Abbott TnI, 92.2% (90.5–92.2) for TnT and 90.6% (70.1–90.6) for Vitros TnI.
Conclusions: hsTnT has superior sensitivity for myocardial infarction than even assays at or near guideline precision requirements (Abbott and Vitros TnI). The specificity of hsTnT assay is not as poor as previous analyses suggest.
©2012 by Walter de Gruyter Berlin Boston
Articles in the same Issue
- A sensitive quantitative test strip based point-of-care albuminuria screening assay
- Editorials
- Serum biomarkers for the non-invasive diagnosis of liver fibrosis: the importance of being validated
- Primary blood tubes mixing: time for updated recommendations
- Reviews
- 10.1515/cclm-2011-0847
- The hCG assay or pregnancy test
- Mini Reviews
- The role of mean platelet volume in predicting thrombotic events
- The role of red blood cell distribution width in cardiovascular and thrombotic disorders
- Genetics and Molecular Diagnostics
- Comparison of different collection procedures and two methods for DNA isolation from saliva
- Detection of α-globin gene deletion and duplication using quantitative multiplex PCR of short fluorescent fragments
- High-throughput detection, genotyping and quantification of the human papillomavirus using real-time PCR
- General Clinical Chemistry and Laboratory Medicine
- Quantification of human serum insulin concentrations in clinical pharmacokinetic or bioequivalence studies: what defines the “best method”?
- Serum cystatin C is an early biomarker for assessment of renal function in burn patients
- 10.1515/cclm.2011.0849
- Mid-stream vs. first-voided urine collection by using automated analyzers for particle examination in healthy subjects: an Italian multicenter study
- Percentage of hypochromic erythrocytes as a potential marker of iron availability
- Laboratory evaluation of a flow cytometric BCR-ABL immunobead assay
- Validation and Outcome Studies
- Independent validation of the Enhanced Liver Fibrosis (ELF) score in the ANRS HC EP 23 Fibrostar cohort of patients with chronic hepatitis C
- Cancer Diagnostics
- Demasking of epithelial cell adhesion molecule (EpCAM) on circulating epithelial tumor cells by Tween®20 treatment in breast cancer patients
- Anti-thyroid peroxidase antibodies are associated with the absence of distant metastases in patients with newly diagnosed breast cancer
- Blood presence of circulating oncofetal fibronectin mRNA, by RT-PCR, does not represent a useful specific marker for the management and follow-up of thyroid cancer patients
- Serum mesothelin and megakaryocyte potentiating factor in pancreatic and biliary cancers
- Cardiovascular Diseases
- The performance of high sensitivity troponin for the diagnosis of acute myocardial infarction is underestimated
- The prognostic value of N-terminal pro-brain natriuretic peptide in non-ST elevation acute coronary syndromes: a meta-analysis
- Measurement of high-molecular-weight adiponectin is not useful in assessing coronary stenosis
- Letters to the Editor
- Serum concentration of neopterin on admission does not improve the diagnostic performance of highly-sensitive troponin I
- Evaluation of the interference by homogentisic acid and other organic acids on the enzymatic and Jaffé method creatinine assay
- A comparative analysis of the Iris iQ200 with manual microscopy as a diagnostic tool for dysmorphic erythrocytes in urine
- Falsely decreased ionized calcium results due to analytical interference by teriflunomide, the active metabolite of leflunomide (Arava®)
- Is polyethylene glycol precipitation of hyperprolactinaemic samples required when using the Beckman Coulter DxI-800 prolactin assay?
- Congress Abstracts
- ISMD2012: Ninth International Symposium on Molecular Diagnostics
- ArabMedLab Marrakech 2012
- Masthead
- Masthead
Articles in the same Issue
- A sensitive quantitative test strip based point-of-care albuminuria screening assay
- Editorials
- Serum biomarkers for the non-invasive diagnosis of liver fibrosis: the importance of being validated
- Primary blood tubes mixing: time for updated recommendations
- Reviews
- 10.1515/cclm-2011-0847
- The hCG assay or pregnancy test
- Mini Reviews
- The role of mean platelet volume in predicting thrombotic events
- The role of red blood cell distribution width in cardiovascular and thrombotic disorders
- Genetics and Molecular Diagnostics
- Comparison of different collection procedures and two methods for DNA isolation from saliva
- Detection of α-globin gene deletion and duplication using quantitative multiplex PCR of short fluorescent fragments
- High-throughput detection, genotyping and quantification of the human papillomavirus using real-time PCR
- General Clinical Chemistry and Laboratory Medicine
- Quantification of human serum insulin concentrations in clinical pharmacokinetic or bioequivalence studies: what defines the “best method”?
- Serum cystatin C is an early biomarker for assessment of renal function in burn patients
- 10.1515/cclm.2011.0849
- Mid-stream vs. first-voided urine collection by using automated analyzers for particle examination in healthy subjects: an Italian multicenter study
- Percentage of hypochromic erythrocytes as a potential marker of iron availability
- Laboratory evaluation of a flow cytometric BCR-ABL immunobead assay
- Validation and Outcome Studies
- Independent validation of the Enhanced Liver Fibrosis (ELF) score in the ANRS HC EP 23 Fibrostar cohort of patients with chronic hepatitis C
- Cancer Diagnostics
- Demasking of epithelial cell adhesion molecule (EpCAM) on circulating epithelial tumor cells by Tween®20 treatment in breast cancer patients
- Anti-thyroid peroxidase antibodies are associated with the absence of distant metastases in patients with newly diagnosed breast cancer
- Blood presence of circulating oncofetal fibronectin mRNA, by RT-PCR, does not represent a useful specific marker for the management and follow-up of thyroid cancer patients
- Serum mesothelin and megakaryocyte potentiating factor in pancreatic and biliary cancers
- Cardiovascular Diseases
- The performance of high sensitivity troponin for the diagnosis of acute myocardial infarction is underestimated
- The prognostic value of N-terminal pro-brain natriuretic peptide in non-ST elevation acute coronary syndromes: a meta-analysis
- Measurement of high-molecular-weight adiponectin is not useful in assessing coronary stenosis
- Letters to the Editor
- Serum concentration of neopterin on admission does not improve the diagnostic performance of highly-sensitive troponin I
- Evaluation of the interference by homogentisic acid and other organic acids on the enzymatic and Jaffé method creatinine assay
- A comparative analysis of the Iris iQ200 with manual microscopy as a diagnostic tool for dysmorphic erythrocytes in urine
- Falsely decreased ionized calcium results due to analytical interference by teriflunomide, the active metabolite of leflunomide (Arava®)
- Is polyethylene glycol precipitation of hyperprolactinaemic samples required when using the Beckman Coulter DxI-800 prolactin assay?
- Congress Abstracts
- ISMD2012: Ninth International Symposium on Molecular Diagnostics
- ArabMedLab Marrakech 2012
- Masthead
- Masthead