To the Editor,
With interest we read the paper by Pasqualetti et al. [1] regarding the care of the laboratory community of pancreatic lipase.
The authors highlight the importance of the recipe in the standardisation of lipase assays. However, it should be noted that, from a physicochemical perspective, lipase (EC 3.1.1.3) is a unique and challenging enzyme to measure, where standardisation approaches cannot overcome all issues related to the contemporary, automated lipase assays.
First, lipase acts at the interphase area between water and the substrate [2, 3], whereby also the cuvette surface will play a role in kinetic lipase assays, especially in present-day miniaturised cuvettes. The physical properties of present-day commercial cuvettes show important differences in chemical composition, surface/volume ratio, contact angles, and total surface energy [2]. Cattoir et al. [4] demonstrated that serum lipase activities not only depend on substrate and enzyme concentrations, but also on the chemical composition of the reaction cuvette used. Since the chemical composition of the cuvette is determined by the choice of the analytical platform, inter-laboratory comparison of serum or plasma lipase results is a difficult task.
Second, Vogel et al. [5] and Oyaert et al. [6]. reported about non-linearity within the primary measurement range of different lipase assays. Consequently, even when all lipase standardisation issues would have been completely solved, it is to be expected that discrepant results will remain when lipase assays are applied on different chemistry analyzers equipped with different cuvettes and different measurement cycles. This is also illustrated by the fact that assays from vendors, making use of the same supplier of bulk lipase reagent, cannot be categorized in the same external quality control method group. The methodological variation potentially leads to incorrect interpretation of lipase test results.
Of note, serum may contain three enzymes that have lipolytic activity: L1 and L2, which are pancreatic isoenzymes or isoforms of lipase (EC 3.1.1.3), and L3, which is probably pancreatic carboxyl ester lipase, also known as cholesterol esterase (EC 3.1.1.13) [7]. The presence of non-pancreatic isoenzymes reduces the specificity of currently used lipase assays as a pancreatic test.
Although lipase immunoassays are not a perfect diagnostic solution [8], these promising ELISA assays disappeared prematurely from the market when the practical colorimetric lipase methods were introduced approximately 30 years ago. At that time, rapid immunoassays were not yet available in the routine laboratory. In contrast to some other clinically used diagnostic enzymes, macromolecular bound lipases (macro-lipases) are rare, which facilitates test interpretation [9]. Furthermore, the lipase isoforms L1 and L2 are immunochemically distinct from isoform L3, which increases the diagnostic specificity of lipase immunoassays [7].
As colorimetric lipase assays are facing numerous analytical problems which are very hard to solve, we share the view of Pasqualetti et al. [1] that we should encourage IVD manufacturers to use their modern immunochemistry technology for the development of a reliable lipase immunoassay that can provide clinicians with rapid lipase results of high quality.
-
Research funding: None declared.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
-
Informed consent: Not applicable.
-
Ethical approval: Not applicable.
References
1. Pasqualetti, S, Borrillo, F, Rovegno, L, Pantheghini, M. Pancreatic lipase: why laboratory community does not take enough care of this clinically important test? Clin Chem Lab Med 2021;59:1914–20.10.1515/cclm-2021-0850Search in Google Scholar
2. Tietz, NW, Shuey, DF. Lipase in serum – the elusive enzyme: an overview. Clin Chem 1993;39:746–55. https://doi.org/10.1093/clinchem/39.5.746.Search in Google Scholar
3. Sarda, L, Desnuelle, P. Action de la lipase pancréatique sur les esters en émulsion. Biochim Biophys Acta 1958;30:513–21. https://doi.org/10.1016/0006-3002(58)90097-0.Search in Google Scholar
4. Cattoir, L, Pede, V, De Cavele, A-S, Seccombe, D, Boterberg, V, Dubruel, P, et al.. Effect of cuvette material on pancreatic lipase activity. Clin Chem Lab Med 2013;51:2109–214. https://doi.org/10.1515/cclm-2013-0369.Search in Google Scholar
5. Vogel, AN, Goldsmith, BM, Marzinke, MA, Sokoll, LJ, Stickle, DF. Non-linearity within the primary measurement range of a lipase assay as the cause of a gap in the interpatient lipase results distribution. Clin Biochem 2016;49:176–9. https://doi.org/10.1016/j.clinbiochem.2015.10.002.Search in Google Scholar
6. Oyaert, M, Delanghe, J, Van den Bossche, J, Ghys, T, Stove, V. Non-linearity in commercially available lipase assays: still gaps to close. Clin Chem Lab Med 2019;58:e21–e23. https://doi.org/10.1515/cclm-2019-0587.Search in Google Scholar
7. Lott, JA, Lu, CJ. Lipase isoforms and amylase isoenzymes: assays and application in the diagnosis of acute pancreatitis. Clin Chem 1991;37:361–8. https://doi.org/10.1093/clinchem/37.3.361.Search in Google Scholar
8. Rizzotti, P, Dechecchi, C, Zanchetta, M, Zaninotto, M, Plebani, M, Burlina, A. Enzyme immunoassay for pancreatic lipase: comparison with turbidimetric method in pancreatic diseases. Clin Biochem 1985;18:230–2. https://doi.org/10.1016/s0009-9120(85)80045-x.Search in Google Scholar
9. Galasso, PJ, Litin, SC, O’Brien, JF. The macroenzymes: a clinical review. Mayo Clin Proc 1993;68:349–54. https://doi.org/10.1016/s0025-6196(12)60129-3.Search in Google Scholar
© 2022 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Obituary
- Dr Per Hyltoft Petersen: an appreciation
- Editorials
- Extrapolated normative GFR data for living kidney donation
- Effectiveness of interventions to improve test appropriateness
- Reviews
- The pathogenesis, epidemiology and biomarkers of susceptibility of pulmonary fibrosis in COVID-19 survivors
- S100B in cardiac surgery brain monitoring: friend or foe?
- Mini Review
- Blood lactate concentration in COVID-19: a systematic literature review
- Opinion Paper
- Commercial immunoassays for detection of anti-SARS-CoV-2 spike and RBD antibodies: urgent call for validation against new and highly mutated variants
- General Clinical Chemistry and Laboratory Medicine
- A randomized controlled study of biochemical tests in primary care: interventions can reduce the number of tests but usage does not become more appropriate
- A national surveillance program for evaluating new reagent lots in medical laboratories
- External quality assessment providers’ services appear to more impact the immunohaematology performance of laboratories than national regulatory and economic conditions
- International collaborative study to evaluate and calibrate two recombinant L chain Ferritin preparations for use as a WHO International Standard
- Method comparison of three serum free light chain assays on the Roche Cobas 6000 c501 chemistry analyzer
- Prospective urinary albumin/creatinine ratio for diagnosis, staging, and organ response assessment in renal AL amyloidosis: results from a large cohort of patients
- Analytical evaluation of the Nittobo Medical tartrate resistant acid phosphatase isoform 5b (TRACP-5b) EIA and comparison with IDS iSYS in different clinically defined populations
- Reference Values and Biological Variations
- Age-adapted percentiles of measured glomerular filtration in healthy individuals: extrapolation to living kidney donors over 65 years
- Indirectly determined hematology reference intervals for pediatric patients in Berlin and Brandenburg
- Hematology and Coagulation
- Monocyte distribution width (MDW): a useful biomarker to improve sepsis management in Emergency Department
- Diabetes
- A step towards optimal efficiency of HbA1c measurement as a first-line laboratory test: the TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project
- Labile glycated hemoglobin: an underestimated laboratory marker of short term glycemia
- Infectious Diseases
- Neutralizing antibody titers six months after Comirnaty vaccination: kinetics and comparison with SARS-CoV-2 immunoassays
- Letters to the Editors
- Molecular detection of SARS-CoV-2 eta VOI in Northern Italy: a case report
- Optimizing effectiveness of COVID-19 vaccination: will laboratory stewardship play a role?
- Measurement in different sample types may aid in detecting interferences and macrocomplexes affecting cardiac troponin measurements
- Which method to detect macrotroponin?
- Marked geographical variation in the prevalence of macroprolactinemia
- Pancreatic lipase assays: time for a change towards immunoassays?
- Analytical evaluation of the performances of a new procalcitonin immunoassay
- Necessity of harmonization of tissue transglutaminase IgA assays to align clinical decision making in coeliac disease
- A fortuitous but characteristic blood smear observation allowing a late diagnosis of MPS-VII
- Prognostic significance of blood-based multi-cancer detection in plasma cell-free DNA
Articles in the same Issue
- Frontmatter
- Obituary
- Dr Per Hyltoft Petersen: an appreciation
- Editorials
- Extrapolated normative GFR data for living kidney donation
- Effectiveness of interventions to improve test appropriateness
- Reviews
- The pathogenesis, epidemiology and biomarkers of susceptibility of pulmonary fibrosis in COVID-19 survivors
- S100B in cardiac surgery brain monitoring: friend or foe?
- Mini Review
- Blood lactate concentration in COVID-19: a systematic literature review
- Opinion Paper
- Commercial immunoassays for detection of anti-SARS-CoV-2 spike and RBD antibodies: urgent call for validation against new and highly mutated variants
- General Clinical Chemistry and Laboratory Medicine
- A randomized controlled study of biochemical tests in primary care: interventions can reduce the number of tests but usage does not become more appropriate
- A national surveillance program for evaluating new reagent lots in medical laboratories
- External quality assessment providers’ services appear to more impact the immunohaematology performance of laboratories than national regulatory and economic conditions
- International collaborative study to evaluate and calibrate two recombinant L chain Ferritin preparations for use as a WHO International Standard
- Method comparison of three serum free light chain assays on the Roche Cobas 6000 c501 chemistry analyzer
- Prospective urinary albumin/creatinine ratio for diagnosis, staging, and organ response assessment in renal AL amyloidosis: results from a large cohort of patients
- Analytical evaluation of the Nittobo Medical tartrate resistant acid phosphatase isoform 5b (TRACP-5b) EIA and comparison with IDS iSYS in different clinically defined populations
- Reference Values and Biological Variations
- Age-adapted percentiles of measured glomerular filtration in healthy individuals: extrapolation to living kidney donors over 65 years
- Indirectly determined hematology reference intervals for pediatric patients in Berlin and Brandenburg
- Hematology and Coagulation
- Monocyte distribution width (MDW): a useful biomarker to improve sepsis management in Emergency Department
- Diabetes
- A step towards optimal efficiency of HbA1c measurement as a first-line laboratory test: the TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project
- Labile glycated hemoglobin: an underestimated laboratory marker of short term glycemia
- Infectious Diseases
- Neutralizing antibody titers six months after Comirnaty vaccination: kinetics and comparison with SARS-CoV-2 immunoassays
- Letters to the Editors
- Molecular detection of SARS-CoV-2 eta VOI in Northern Italy: a case report
- Optimizing effectiveness of COVID-19 vaccination: will laboratory stewardship play a role?
- Measurement in different sample types may aid in detecting interferences and macrocomplexes affecting cardiac troponin measurements
- Which method to detect macrotroponin?
- Marked geographical variation in the prevalence of macroprolactinemia
- Pancreatic lipase assays: time for a change towards immunoassays?
- Analytical evaluation of the performances of a new procalcitonin immunoassay
- Necessity of harmonization of tissue transglutaminase IgA assays to align clinical decision making in coeliac disease
- A fortuitous but characteristic blood smear observation allowing a late diagnosis of MPS-VII
- Prognostic significance of blood-based multi-cancer detection in plasma cell-free DNA