Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyzer suggests a new era in laboratory medicine
Abstract
Objectives
Due to its high specificity, liquid chromatography-tandem mass spectrometry (LC-MS/MS) is considered the gold standard in diagnostic areas such as therapeutic monitoring of immunosuppressive drugs (ISDs). However, many laboratories still rely on immunoassays for ISD quantification in a tradeoff between analytical performance and the advantages of fully automated analyzers – shorter turnaround times, greater ease of use, and 24/7 availability.
Methods
The LC-MS/MS-based Thermo Scientific™ Cascadion™ SM Immunosuppressant Panel was evaluated for >6 months in the routine laboratory of a university hospital. We assessed the analytical performance of the panel and compared it to conventional LC-MS/MS as well as to immunoassays (cyclosporine A, sirolimus, tacrolimus (Siemens) and everolimus (Thermo Fisher)). In addition, both ISD panel and Cascadion analyzer were scrutinized with regards to, e.g., turnaround time, usability, and robustness.
Results
All ISDs showed high linearity and precision (CV≤6%) and a good correlation with conventional LC-MS/MS. The mean deviation to the immunoassays was 17–19% and negative for all ISDs except everolimus with a positive 19% bias. No weak points were revealed when challenging assay and system with, e.g., high haematocrit, sedimented whole blood or priority samples. The Cascadion integrated well into our 24/7 routine and could easily be operated simultaneously with several other analyzers by technical staff without LC-MS experience.
Conclusions
The ISD panel showed excellent analytical performance and demonstrated that a fully automated LC-MS-based analysis starting from primary samples is feasible, suggesting that LC-MS could become an integral part of 24/7 diagnostics in the near future.
Acknowledgments
We thank Isolde Riedlinger and Susanne Faix for excellent technical assistance.
-
Research funding: None declared.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
-
Informed consent: Informed consent was obtained from all individuals included in this study.
-
Ethical approval: The study was conducted according to the Declaration of Helsinki and approved by the local Ethics Committee of the medical faculty of Tübingen (protocol number: 113/2014BO1).
References
1. Seger, C, Salzmann, L. After another decade: LC-MS/MS became routine in clinical diagnostics. Clin Biochem 2020;82:2–11. https://doi.org/10.1016/j.clinbiochem.2020.03.004.Search in Google Scholar PubMed
2. Heaney, LM. Advancements in mass spectrometry as a tool for clinical analysis: part II. Clin Chem Lab Med 2020;58:855–7. https://doi.org/10.1515/cclm-2020-0259.Search in Google Scholar PubMed
3. Gounden, V, Soldin, SJ. Tacrolimus measurement: building a better immunoassay. Clin Chem 2014;60:575–6. https://doi.org/10.1373/clinchem.2013.220012.Search in Google Scholar PubMed
4. McShane, AJ, Bunch, DR, Wang, S. Therapeutic drug monitoring of immunosuppressants by liquid chromatography-mass spectrometry. Clin Chim Acta 2016;454:1–5. https://doi.org/10.1016/j.cca.2015.12.027.Search in Google Scholar PubMed
5. Wallemacq, PE. Therapeutic monitoring of immunosuppressant drugs. Where are we? Clin Chem Lab Med 2004;42:1204–11. https://doi.org/10.1515/cclm.2004.242.Search in Google Scholar
6. Zhang, Y, Zhang, R. Recent advances in analytical methods for the therapeutic drug monitoring of immunosuppressive drugs. Drug Test Anal 2018;10:81–94. https://doi.org/10.1002/dta.2290.Search in Google Scholar PubMed
7. Qin, X, Rui, J, Xia, Y, Mu, H, Song, SH, Raja Aziddin, RE, Miles, G, et al. Multi-center performance evaluations of tacrolimus and cyclosporine electrochemiluminescence immunoassays in the Asia-Pacific region. Ann Lab Med 2018;38:85–94. https://doi.org/10.3343/alm.2018.38.2.85.Search in Google Scholar PubMed PubMed Central
8. Taguchi, K, Ohmura, T, Ohya, Y, Horio, M, Furukawa, K, Jono, H, et al. False tacrolimus concentrations measured by antibody-conjugated magnetic immunoassay in liver transplant patient: 2 case reports and literature review. Exp Clin Transplant 2014;12:474–8. https://doi.org/10.6002/ect.2013.0113.Search in Google Scholar PubMed
9. Morelle, J, Wallemacq, P, Van Caeneghem, O, Goffin, E. Clinically unexpected cyclosporine levels using the ACMIA method on the RXL dimension analyser. Nephrol Dial Transplant 2011;26:1428–31. https://doi.org/10.1093/ndt/gfr002.Search in Google Scholar PubMed
10. Peter, A, Shipkova, M, Wieland, E, Schleicher, E, Müller, I. Increased cyclosporine concentrations in the absence of cyclosporine administration. Clin Chem 2011;57:670–3. https://doi.org/10.1373/clinchem.2010.148718.Search in Google Scholar PubMed
11. Schniedewind, B, Niederlechner, S, Galinkin, J, Johnson-Davis, K, Christians, U, Meyer, E. Long-term cross-validation of everolimus therapeutic drug monitoring assays: the Zortracker study. Ther Drug Monit 2015;37:296–303. https://doi.org/10.1097/ftd.0000000000000191.Search in Google Scholar PubMed PubMed Central
12. Christians, U, Vinks, AA, Langman, LJ, Clarke, W, Wallemacq, P, van Gelder, T, et al. Impact of laboratory practices on interlaboratory variability in therapeutic drug monitoring of immunosuppressive drugs. Ther Drug Monit 2015;37:718–24. https://doi.org/10.1097/FTD.0000000000000205.Search in Google Scholar PubMed
13. Benton, SC, Tetteh, GK, Needham, S-J, Mücke, J, Sheppard, L, Alderson, S, et al. Evaluation of the 25-hydroxy vitamin D assay on a fully automated liquid chromatography mass spectrometry system, the thermo scientific cascadion SM clinical analyzer with the cascadion 25-hydroxy vitamin D assay in a routine clinical laboratory. Clin Chem Lab Med 2020;58:1010–7. https://doi.org/10.1515/cclm-2019-0834.Search in Google Scholar PubMed
14. Shipkova, M, Rapp, S, Rigo-Bonnin, R, Wieland, E, Peter, A. Therapeutic drug monitoring of everolimus: comparability of concentrations determined by 2 immunoassays and a liquid chromatography tandem mass spectrometry method. Ther Drug Monit 2017;39:102–8. https://doi.org/10.1097/ftd.0000000000000376.Search in Google Scholar
15. Salvagno, GL, Danese, E, Lippi, G. Mass spectrometry and total laboratory automation: opportunities and drawbacks. Clin Chem Lab Med 2020;58:994–1001. https://doi.org/10.1515/cclm-2019-0723.Search in Google Scholar PubMed
16. Jannetto, PJ, Fitzgerald, RL. Effective use of mass spectrometry in the clinical laboratory. Clin Chem 2016;62:92–8. https://doi.org/10.1373/clinchem.2015.248146.Search in Google Scholar PubMed
17. Vogeser, M. From therapeutic drug monitoring to total drug monitoring and drug-omics. Clin Chem Lab Med 2020;59:287–90. https://doi.org/10.1515/cclm-2020-0339.10.1515/cclm-2020-0339Search in Google Scholar PubMed
Supplementary Material
The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2020-0848).
© 2020 Walter de Gruyter GmbH, Berlin/Boston
Articles in the same Issue
- Frontmatter
- Editorial
- Home pregnancy tests: quality first
- Review
- Non-invasive determination of uric acid in human saliva in the diagnosis of serious disorders
- Opinion Papers
- Basophil counting in hematology analyzers: time to discontinue?
- The role of laboratory hematology between technology and professionalism: the paradigm of basophil counting
- Recommendations for validation testing of home pregnancy tests (HPTs) in Europe
- General Clinical Chemistry and Laboratory Medicine
- The use of preanalytical quality indicators: a Turkish preliminary survey study
- The Italian External Quality Assessment (EQA) program on urinary sediment by microscopy examination: a 20 years journey
- Non-HDL-C/TG ratio indicates significant underestimation of calculated low-density lipoprotein cholesterol (LDL-C) better than TG level: a study on the reliability of mathematical formulas used for LDL-C estimation
- Evaluation of the protein gap for detection of abnormal serum gammaglobulin level: an imperfect predictor
- Impact of routine S100B protein assay on CT scan use in children with mild traumatic brain injury
- Using machine learning to develop an autoverification system in a clinical biochemistry laboratory
- Effect of collection matrix, platelet depletion, and storage conditions on plasma extracellular vesicles and extracellular vesicle-associated miRNAs measurements
- Pneumatic tube transportation of urine samples
- Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyzer suggests a new era in laboratory medicine
- A validated LC-MS/MS method for the simultaneous quantification of the novel combination antibiotic, ceftolozane–tazobactam, in plasma (total and unbound), CSF, urine and renal replacement therapy effluent: application to pilot pharmacokinetic studies
- Immunosuppressant quantification in intravenous microdialysate – towards novel quasi-continuous therapeutic drug monitoring in transplanted patients
- Reference Values and Biological Variations
- Reference intervals for venous blood gas measurement in adults
- Cardiovascular Diseases
- Detection and functional characterization of a novel MEF2A variation responsible for familial dilated cardiomyopathy
- Diabetes
- Evaluation of the ARKRAY HA-8190V instrument for HbA1c
- Infectious Diseases
- An original multiplex method to assess five different SARS-CoV-2 antibodies
- Evaluation of dried blood spots as alternative sampling material for serological detection of anti-SARS-CoV-2 antibodies using established ELISAs
- Variability of cycle threshold values in an external quality assessment scheme for detection of the SARS-CoV-2 virus genome by RT-PCR
- The vasoactive peptide MR-pro-adrenomedullin in COVID-19 patients: an observational study
- Corrigenda
- Corrigendum to: Understanding and managing interferences in clinical laboratory assays: the role of laboratory professionals
- Corrigendum to: Age appropriate reference intervals for eight kidney function and injury markers in infants, children and adolescents
- Letters to the Editor
- A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data on Sysmex XN-series analysers
- Letter in reply to the letter to the editor of Harte JV and Mykytiv V with the title “A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data”
- SARS-CoV-2 serologic tests: do not forget the good laboratory practice
- Long-term kinetics of anti-SARS-CoV-2 antibodies in a cohort of 197 hospitalized and non-hospitalized COVID-19 patients
- Self-sampling at home using volumetric absorptive microsampling: coupling analytical evaluation to volunteers’ perception in the context of a large scale study
- Vortex mixing to alleviate pseudothrombocytopenia in a blood specimen with platelet satellitism and platelet clumps
- Comparative evaluation of the fully automated HemosIL® AcuStar ADAMTS13 activity assay vs. ELISA: possible interference by autoantibodies different from anti ADAMTS-13
- Significant interference on specific point-of-care glucose measurements due to high dose of intravenous vitamin C therapy in critically ill patients
- As time goes by, on that you can rely … preservation of urine samples for morphological analysis of erythrocytes and casts
- Stability of control materials for α-thalassemia immunochromatographic strip test
- Reformulated Architect® cyclosporine CMIA assay: improved imprecision, worse comparability between methods
- Urine-to-plasma contamination mimicking acute kidney injury: small drops with major consequences
- Automated Mindray CL-1200i chemiluminescent assays of renin and aldosterone for the diagnosis of primary aldosteronism
- Use of common reference intervals does not necessarily allow inter-method numerical result trending
- Reply to Dr Hawkins regarding comparability of results for monitoring
Articles in the same Issue
- Frontmatter
- Editorial
- Home pregnancy tests: quality first
- Review
- Non-invasive determination of uric acid in human saliva in the diagnosis of serious disorders
- Opinion Papers
- Basophil counting in hematology analyzers: time to discontinue?
- The role of laboratory hematology between technology and professionalism: the paradigm of basophil counting
- Recommendations for validation testing of home pregnancy tests (HPTs) in Europe
- General Clinical Chemistry and Laboratory Medicine
- The use of preanalytical quality indicators: a Turkish preliminary survey study
- The Italian External Quality Assessment (EQA) program on urinary sediment by microscopy examination: a 20 years journey
- Non-HDL-C/TG ratio indicates significant underestimation of calculated low-density lipoprotein cholesterol (LDL-C) better than TG level: a study on the reliability of mathematical formulas used for LDL-C estimation
- Evaluation of the protein gap for detection of abnormal serum gammaglobulin level: an imperfect predictor
- Impact of routine S100B protein assay on CT scan use in children with mild traumatic brain injury
- Using machine learning to develop an autoverification system in a clinical biochemistry laboratory
- Effect of collection matrix, platelet depletion, and storage conditions on plasma extracellular vesicles and extracellular vesicle-associated miRNAs measurements
- Pneumatic tube transportation of urine samples
- Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyzer suggests a new era in laboratory medicine
- A validated LC-MS/MS method for the simultaneous quantification of the novel combination antibiotic, ceftolozane–tazobactam, in plasma (total and unbound), CSF, urine and renal replacement therapy effluent: application to pilot pharmacokinetic studies
- Immunosuppressant quantification in intravenous microdialysate – towards novel quasi-continuous therapeutic drug monitoring in transplanted patients
- Reference Values and Biological Variations
- Reference intervals for venous blood gas measurement in adults
- Cardiovascular Diseases
- Detection and functional characterization of a novel MEF2A variation responsible for familial dilated cardiomyopathy
- Diabetes
- Evaluation of the ARKRAY HA-8190V instrument for HbA1c
- Infectious Diseases
- An original multiplex method to assess five different SARS-CoV-2 antibodies
- Evaluation of dried blood spots as alternative sampling material for serological detection of anti-SARS-CoV-2 antibodies using established ELISAs
- Variability of cycle threshold values in an external quality assessment scheme for detection of the SARS-CoV-2 virus genome by RT-PCR
- The vasoactive peptide MR-pro-adrenomedullin in COVID-19 patients: an observational study
- Corrigenda
- Corrigendum to: Understanding and managing interferences in clinical laboratory assays: the role of laboratory professionals
- Corrigendum to: Age appropriate reference intervals for eight kidney function and injury markers in infants, children and adolescents
- Letters to the Editor
- A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data on Sysmex XN-series analysers
- Letter in reply to the letter to the editor of Harte JV and Mykytiv V with the title “A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data”
- SARS-CoV-2 serologic tests: do not forget the good laboratory practice
- Long-term kinetics of anti-SARS-CoV-2 antibodies in a cohort of 197 hospitalized and non-hospitalized COVID-19 patients
- Self-sampling at home using volumetric absorptive microsampling: coupling analytical evaluation to volunteers’ perception in the context of a large scale study
- Vortex mixing to alleviate pseudothrombocytopenia in a blood specimen with platelet satellitism and platelet clumps
- Comparative evaluation of the fully automated HemosIL® AcuStar ADAMTS13 activity assay vs. ELISA: possible interference by autoantibodies different from anti ADAMTS-13
- Significant interference on specific point-of-care glucose measurements due to high dose of intravenous vitamin C therapy in critically ill patients
- As time goes by, on that you can rely … preservation of urine samples for morphological analysis of erythrocytes and casts
- Stability of control materials for α-thalassemia immunochromatographic strip test
- Reformulated Architect® cyclosporine CMIA assay: improved imprecision, worse comparability between methods
- Urine-to-plasma contamination mimicking acute kidney injury: small drops with major consequences
- Automated Mindray CL-1200i chemiluminescent assays of renin and aldosterone for the diagnosis of primary aldosteronism
- Use of common reference intervals does not necessarily allow inter-method numerical result trending
- Reply to Dr Hawkins regarding comparability of results for monitoring