To the Editor,
Primary aldosteronism (PA) is the most common, but even the least identified cause of arterial hypertension characterized by low levels of plasma renin and high plasma levels of aldosterone (ALD). The latter are inappropriately high for the volume and blood pressure status [1]. Despite several harmful consequences on the heart, arterial wall, and its common occurrence, PA is rarely diagnosed, mainly because it mimics primary (essential) hypertension, but also for the scarce of availability of accurate and reliable laboratory tests to be performed on automated platforms. In fact, in addition to the measurement of serum ALD, the diagnosis is based on the aldosterone/renin ratio (ARR), which is the ALD concentration divided by plasma renin [2]. Until some years ago, the measurements of both ALD and plasma renin activity (PRA) have been performed by radioimmunoassays which were time consuming, affected by risk of errors related to manual procedures and medico-legal associated problems of radioactive waste.
In addition, available methods for measuring renin, including both the plasma renin activity assay and the direct active renin concentration (DRC) assays have been found to be inaccurate in the low range of renin values. The same poor analytical sensitivity issues have been reported for ALD assays [3].
More recently, more sensitive chemiluminescent immunoassays for ALD and direct active renin have been developed and automated on high-throughput platforms to provide both accurate and rapid results [4].
Here, we describe the main performances of two novel fully automated competitive binding chemiluminescent assays for ALD and renin automated on Mindray CL-1200i (Shenzhen Mindray Bio-Medical Electronics Co, Shenzhen China), with measurement execution times of 41 min for ALD and 26 min for renin.
Traceability: The traceability process for both assays is based on EN ISO 17511:2003; the measured (analyte) in the calibrators for renin is traceable to WHO Reference Material IPR68/356, while for ALD is traceable to a commercial ALD test (CLIA).
Analytical sensitivity: The ALD (CLIA) reagent kit has an analytical sensitivity of 40.2 pmol/L and a functional sensitivity of 47.9 pmol/L. Renin assay has an analytical sensitivity of 0.84 mIU/L and a functional sensitivity of 1.17 mIU/L. The linearity range for ALD spans from 40.2 to 5,540 pmol/L and for renin from 0.835 to 835 mIU/L.
Imprecision: Imprecision studies performed according to the EP5-A3, demonstrated for ALD a repeatability coefficient of variation (CV%) of 1.95% at 609.9 pmol/L and 1.54% at 4,210.9 pmol/L, respectively. The within-laboratory CV%, at the same concentrations, resulted 2.90 and 2.50%, respectively. For renin, the repeatability (CV%) was found to be 1.71% at 21.3 mIU/L and 1.60% at 87.3 mIU/L, respectively. The within-laboratory CV%, at the same concentrations, resulted 2.28 and 2.23%, respectively.
Correlation studies: The LIAISON® XL [DiaSorin, Saluggia (VC), Italy] analytical system was used for comparability study with respect to Mindray CL-1200i. Considering the Mindray CL-1200i results, correlation studies have been performed for a total of 243 plasma EDTA specimens for renin and of 177 serum specimens for ALD, collected in a dynamic range from 1.44 to 2,829.6 mIU/L for renin and from 41.2 to 9,940 pmol/L for ALD.
Bland-Altman analyses of ALD and Renin, reported in Figure 1, showed that, overall, the comparison presented a bias of 173.4 pmol/L (95% CI: 142.2–204.6) and of 9.9 mIU/L (95%: 6.2–13.6), respectively. Differently, the slopes and intercepts of Passing-Bablok regression results and their corresponding 95% CI showed that the comparison for ALD presented a constant and proportional bias, whereas for renin there is a proportional bias. In particular, for ALD the slope is 1.74 (95% CI: 1.61–1.88) and intercept is −64.47 (95% CI: −99.42–24.69); for Renin the slope is 1.16, (95% CI: 1.07 to 1.30) and intercept is 0.55 (95% CI: 0.35–1.14).

Bland-Altman analyses of Mindray CL-1200i vs. Diasorin LIAISON® XL for aldosterone (A) and renin (B) results.
These comparisons were performed using a total of 243 plasma EDTA specimens for renin and of 177 serum specimens for aldosterone, collected in a dynamic range from 1.44 to 2,829.6 mIU/L for renin and from 41.2 to 9,940 pmol/L for aldosterone.
Considering the ALD/Renin ratio (ARR), the comparison between Mindray CL-1200i vs. Diasorin LIAISON® XL results were analysed by Bland-Altman and Passing-Bablok regression. Both analyses showed an absence of statistically significant bias between assays Figure 2.

Bland-Altman analysis (A) and Passing-Bablok regression (B) of the aldosterone and renin ratio (aldosterone/renin) for Mindray CL-1200i vs. Diasorin LIAISON® XL.
Recovery tests: The recovery tests were performed using two pools. The renin concentration in low (L) pool was 8.7 mIU/L and in the high (H) pool 129.7 mIU/L. A series of nine serial dilutions were made by mixing the two pools with the following ratios: 10 (L): 0 (H), 9(L): 1(H), 8(L): 2(H), 7(L): 3(H), 6(L): 4(H), 5(L): 5(H), 4(L): 6(H), 3(L): 7(H), 2(L): 8(H), 1(L): 9(H), 10 (H). The recovery test is show in Supplementary Figure 1. The percentage of recovery ranged from 98.42 to 108.16%. Similar data have been observed for ALD with percentages of recovery ranging from 97.61 to 107.8%.
The data collected highlight the satisfactory analytical sensitivity, imprecision and repeatability of both the Mindray chemiluminescent ALD and Renin assays. In addition, the recovery studies demonstrated a satisfactory accuracy of results within a wide analytical range.
While a bias between the individual assays of Mindray CL-1200i and Diasorin LIAISON® XL for both ALD and renin has been observed, no significant bias has been found for the ARR which assures the most interesting information from a clinical viewpoint, particularly at an appropriate cut-off [1], [4]. However, as a suppressed renin value can increase the ARR even when ALD is normal, to avoid overinflating the ARR value some precaution should be used, particularly in some subgroups of patients who may have low renin values [1].
In conclusion, our data confirm that chemiluminescent assays assure accurate and reliable measurement of both ALD and Renin and that these data, particularly when reported as ARR represents a valuable information for the diagnosis of primary aldosteronism in clinical practice.
-
Research funding: None declared.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
References
1. Rossi, GP. Primary aldosteronism: JACC state-of-the-art review. J Am Coll Cardiol 2019;74:2799–811.10.1016/j.jacc.2019.09.057Search in Google Scholar PubMed
2. Hiramatsu, K, Yamada, T, Yukimura, Y, Komiya, I, Ichikawa, K, Ishihara, M, et al.. A screening test to identify aldosterone-producing adenoma by measuring plasma renin activity. Results in hypertensive patients. Arch Intern Med 1981;141:1589–93.10.1001/archinte.1981.00340130033011Search in Google Scholar
3. Rossi, GP, Barisa, M, Belfiore, A, Desideri, G, Ferri, C, Letizia, C, et al.. The aldosterone-renin ratio based on the plasma renin activity and the direct renin assay for diagnosing aldosterone-producing adenoma. J Hypertens 2010;28:1892–9.10.1097/HJH.0b013e32833d2192Search in Google Scholar PubMed
4. Rossi, GP, Ceolotto, G, Rossitto, G, Seccia, TM, Maiolino, G, Berton, C, et al.. Prospective validation of an automated chemiluminescence-based assay of renin and aldosterone for the work-up of arterial hypertension. Clin Chem Lab Med 2016;54:1441–50.10.1515/cclm-2015-1094Search in Google Scholar PubMed
Supplementary Material
The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2020-1860).
© 2021 Chiara Cosma et al., published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.
Articles in the same Issue
- Frontmatter
- Editorial
- Home pregnancy tests: quality first
- Review
- Non-invasive determination of uric acid in human saliva in the diagnosis of serious disorders
- Opinion Papers
- Basophil counting in hematology analyzers: time to discontinue?
- The role of laboratory hematology between technology and professionalism: the paradigm of basophil counting
- Recommendations for validation testing of home pregnancy tests (HPTs) in Europe
- General Clinical Chemistry and Laboratory Medicine
- The use of preanalytical quality indicators: a Turkish preliminary survey study
- The Italian External Quality Assessment (EQA) program on urinary sediment by microscopy examination: a 20 years journey
- Non-HDL-C/TG ratio indicates significant underestimation of calculated low-density lipoprotein cholesterol (LDL-C) better than TG level: a study on the reliability of mathematical formulas used for LDL-C estimation
- Evaluation of the protein gap for detection of abnormal serum gammaglobulin level: an imperfect predictor
- Impact of routine S100B protein assay on CT scan use in children with mild traumatic brain injury
- Using machine learning to develop an autoverification system in a clinical biochemistry laboratory
- Effect of collection matrix, platelet depletion, and storage conditions on plasma extracellular vesicles and extracellular vesicle-associated miRNAs measurements
- Pneumatic tube transportation of urine samples
- Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyzer suggests a new era in laboratory medicine
- A validated LC-MS/MS method for the simultaneous quantification of the novel combination antibiotic, ceftolozane–tazobactam, in plasma (total and unbound), CSF, urine and renal replacement therapy effluent: application to pilot pharmacokinetic studies
- Immunosuppressant quantification in intravenous microdialysate – towards novel quasi-continuous therapeutic drug monitoring in transplanted patients
- Reference Values and Biological Variations
- Reference intervals for venous blood gas measurement in adults
- Cardiovascular Diseases
- Detection and functional characterization of a novel MEF2A variation responsible for familial dilated cardiomyopathy
- Diabetes
- Evaluation of the ARKRAY HA-8190V instrument for HbA1c
- Infectious Diseases
- An original multiplex method to assess five different SARS-CoV-2 antibodies
- Evaluation of dried blood spots as alternative sampling material for serological detection of anti-SARS-CoV-2 antibodies using established ELISAs
- Variability of cycle threshold values in an external quality assessment scheme for detection of the SARS-CoV-2 virus genome by RT-PCR
- The vasoactive peptide MR-pro-adrenomedullin in COVID-19 patients: an observational study
- Corrigenda
- Corrigendum to: Understanding and managing interferences in clinical laboratory assays: the role of laboratory professionals
- Corrigendum to: Age appropriate reference intervals for eight kidney function and injury markers in infants, children and adolescents
- Letters to the Editor
- A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data on Sysmex XN-series analysers
- Letter in reply to the letter to the editor of Harte JV and Mykytiv V with the title “A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data”
- SARS-CoV-2 serologic tests: do not forget the good laboratory practice
- Long-term kinetics of anti-SARS-CoV-2 antibodies in a cohort of 197 hospitalized and non-hospitalized COVID-19 patients
- Self-sampling at home using volumetric absorptive microsampling: coupling analytical evaluation to volunteers’ perception in the context of a large scale study
- Vortex mixing to alleviate pseudothrombocytopenia in a blood specimen with platelet satellitism and platelet clumps
- Comparative evaluation of the fully automated HemosIL® AcuStar ADAMTS13 activity assay vs. ELISA: possible interference by autoantibodies different from anti ADAMTS-13
- Significant interference on specific point-of-care glucose measurements due to high dose of intravenous vitamin C therapy in critically ill patients
- As time goes by, on that you can rely … preservation of urine samples for morphological analysis of erythrocytes and casts
- Stability of control materials for α-thalassemia immunochromatographic strip test
- Reformulated Architect® cyclosporine CMIA assay: improved imprecision, worse comparability between methods
- Urine-to-plasma contamination mimicking acute kidney injury: small drops with major consequences
- Automated Mindray CL-1200i chemiluminescent assays of renin and aldosterone for the diagnosis of primary aldosteronism
- Use of common reference intervals does not necessarily allow inter-method numerical result trending
- Reply to Dr Hawkins regarding comparability of results for monitoring
Articles in the same Issue
- Frontmatter
- Editorial
- Home pregnancy tests: quality first
- Review
- Non-invasive determination of uric acid in human saliva in the diagnosis of serious disorders
- Opinion Papers
- Basophil counting in hematology analyzers: time to discontinue?
- The role of laboratory hematology between technology and professionalism: the paradigm of basophil counting
- Recommendations for validation testing of home pregnancy tests (HPTs) in Europe
- General Clinical Chemistry and Laboratory Medicine
- The use of preanalytical quality indicators: a Turkish preliminary survey study
- The Italian External Quality Assessment (EQA) program on urinary sediment by microscopy examination: a 20 years journey
- Non-HDL-C/TG ratio indicates significant underestimation of calculated low-density lipoprotein cholesterol (LDL-C) better than TG level: a study on the reliability of mathematical formulas used for LDL-C estimation
- Evaluation of the protein gap for detection of abnormal serum gammaglobulin level: an imperfect predictor
- Impact of routine S100B protein assay on CT scan use in children with mild traumatic brain injury
- Using machine learning to develop an autoverification system in a clinical biochemistry laboratory
- Effect of collection matrix, platelet depletion, and storage conditions on plasma extracellular vesicles and extracellular vesicle-associated miRNAs measurements
- Pneumatic tube transportation of urine samples
- Evaluation of the first immunosuppressive drug assay available on a fully automated LC-MS/MS-based clinical analyzer suggests a new era in laboratory medicine
- A validated LC-MS/MS method for the simultaneous quantification of the novel combination antibiotic, ceftolozane–tazobactam, in plasma (total and unbound), CSF, urine and renal replacement therapy effluent: application to pilot pharmacokinetic studies
- Immunosuppressant quantification in intravenous microdialysate – towards novel quasi-continuous therapeutic drug monitoring in transplanted patients
- Reference Values and Biological Variations
- Reference intervals for venous blood gas measurement in adults
- Cardiovascular Diseases
- Detection and functional characterization of a novel MEF2A variation responsible for familial dilated cardiomyopathy
- Diabetes
- Evaluation of the ARKRAY HA-8190V instrument for HbA1c
- Infectious Diseases
- An original multiplex method to assess five different SARS-CoV-2 antibodies
- Evaluation of dried blood spots as alternative sampling material for serological detection of anti-SARS-CoV-2 antibodies using established ELISAs
- Variability of cycle threshold values in an external quality assessment scheme for detection of the SARS-CoV-2 virus genome by RT-PCR
- The vasoactive peptide MR-pro-adrenomedullin in COVID-19 patients: an observational study
- Corrigenda
- Corrigendum to: Understanding and managing interferences in clinical laboratory assays: the role of laboratory professionals
- Corrigendum to: Age appropriate reference intervals for eight kidney function and injury markers in infants, children and adolescents
- Letters to the Editor
- A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data on Sysmex XN-series analysers
- Letter in reply to the letter to the editor of Harte JV and Mykytiv V with the title “A panhaemocytometric approach to COVID-19: a retrospective study on the importance of monocyte and neutrophil population data”
- SARS-CoV-2 serologic tests: do not forget the good laboratory practice
- Long-term kinetics of anti-SARS-CoV-2 antibodies in a cohort of 197 hospitalized and non-hospitalized COVID-19 patients
- Self-sampling at home using volumetric absorptive microsampling: coupling analytical evaluation to volunteers’ perception in the context of a large scale study
- Vortex mixing to alleviate pseudothrombocytopenia in a blood specimen with platelet satellitism and platelet clumps
- Comparative evaluation of the fully automated HemosIL® AcuStar ADAMTS13 activity assay vs. ELISA: possible interference by autoantibodies different from anti ADAMTS-13
- Significant interference on specific point-of-care glucose measurements due to high dose of intravenous vitamin C therapy in critically ill patients
- As time goes by, on that you can rely … preservation of urine samples for morphological analysis of erythrocytes and casts
- Stability of control materials for α-thalassemia immunochromatographic strip test
- Reformulated Architect® cyclosporine CMIA assay: improved imprecision, worse comparability between methods
- Urine-to-plasma contamination mimicking acute kidney injury: small drops with major consequences
- Automated Mindray CL-1200i chemiluminescent assays of renin and aldosterone for the diagnosis of primary aldosteronism
- Use of common reference intervals does not necessarily allow inter-method numerical result trending
- Reply to Dr Hawkins regarding comparability of results for monitoring