To the Editor,
The definition and fulfillment of suitable analytical performance specifications (APS) for the measurement uncertainty (MU) is essential to ensure that laboratory measurements are clinically usable [1], [2]. There is now a global consensus that the derivation of APS for different measurands should be carried out by using the models established in 2014 by the Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) held in Milan, Italy [3]. In particular, model 1, based on the effect of analytical performance on clinical outcome, applies for measurands that have a central role in diagnosis and monitoring of a specific disease; model 2, based on the components of biological variation (BV) of the measurand, should be used for measurands under strict metabolic control; and model 3, based on the state of the art of the measurement (defined as the highest level of analytical performance technically achievable) should be used for measurands that cannot be included in the first two models [4].
C-reactive protein (CRP) is the most sensitive of the acute phase proteins and its concentrations in serum increase rapidly during inflammatory processes [5]. However, an elevation of CRP is not diagnostic of any one specific disease as it occurs in many diseases involving tissue damage or inflammation. Besides not having a central and well-defined role in the decision making of a specific disease, CRP is a biologically challenging analyte. We previously highlighted the difficulties in deriving reliable CRP BV data [6], [7]. The meta-analysed BV data made recently available on the EFLM database (https://biologicalvariation.eu/) are 33.5% for intraindividual (CVI) and 87.7% for inter-individual variability (CVG), respectively. As Franzini pointed out, the CV describing the distribution of a set of positive distributed values, as laboratory results are, cannot exceed 33.3% [8]. Higher CVI and CVG values probably represent incorrect estimates due either to a transitory illness of some subjects in the evaluated sample group or, more generally, to non-normal distribution of the CRP data in apparently healthy individuals.
Considering that neither of the first two EFLM models seems to be suitable for CRP, the model three constitutes the best option to derive APS for this measurand. According to the definition of the “state of the art” by the EFLM Strategic Conference, to derive APS for MU by using this model it is necessary to identify the highest quality of performance that is currently achievable [3]. With this aim, we compared experimentally estimated standard MU of four widely used commercial measuring systems for CRP measurement available in our laboratory: Alinity c and Architect c16000 (Abbott Diagnostics), AU680 (Beckman Coulter) and Cobas c501 (Roche Diagnostics), all ‘Conformité Européenne’/CE marked. Table 1 summarizes the characteristics of the evaluated measuring systems.
Characteristics of measuring systems for C-reactive protein measurement included in the study.
Manufacturer | Platform | Commercial name and assay principle | Manufacturer’s instruction for use version | Calibrator name | Stated traceability | Certified value of reference material ± standard uncertainty |
---|---|---|---|---|---|---|
Abbott Diagnostics | Alinity c | CRP Vario Wide Range Immunoturbidimetric method (art. 07P56) | G71182/R01,May 2017 | CRP Vario Wide Range Calibrator kit | ERM-DA 472/IFCC | 41.8 ± 1.25 mg/L |
Abbott Diagnostics | Architect c16000 | CRP Vario Wide Range Immunoturbidimetric method (art. 6K26) | 306731/R04, August 2015 | CRP Calibrator Set | ERM-DA 472/IFCC | 41.8 ± 1.25 mg/L |
Beckman Coulter | AU680 | CRP Latex Immunoturbidimetry (art. OSR6199) | BLOSR6x99.04, July 2013 | CRP Latex Calibrator Normal Set | ERM-DA 474/IFCC | 41.2 ± 1.25 mg/L |
Roche Diagnostics | Cobas c501 | CRPL3 Particle Enhanced Immunoturbidimetric Gen. 3 (art. 04956842) | V 10.0, December 2018 | C.f.a.s. Proteins | CRM 470 | 39.2 ± 0.95 mg/L |
According to recently published documents [1], [9], MU associated with laboratory results should combine two uncertainty components: the MU of calibrator value (ucal) and the MU accounting for random sources (uRw). The former should in turn combine the MU of the higher-order reference selected by the in vitro diagnostics (IVD) manufacturer for implementing traceability with the MU deriving from the process for assignment of calibrator values. uRw gives information about the stability of the measuring system over time and its variability when employed by an individual laboratory. The recommended time span for this evaluation is 6 months, as it allows covering all the significant sources of uRw [9]. We previously described in detail how correctly estimating uRw through the internal quality control (IQC) information [10]. In this study, for each measuring system, we collected the information about the metrological traceability and MU of commercial calibrators at CRP concentration of 10 mg/L, the 99th percentile limit of reference value distribution, with the exception of Roche C.f.a.s. calibrator for which only one level is available at 79.9 mg/L concentration. Then, we calculated the ucal by combining the standard MU of employed reference material, available on the corresponding certificate of analysis, with the standard MU of commercial calibrator as provided by the manufacturer. The uRw was experimentally estimated as intermediate reproducibility from 6-month consecutive measurement data of a serum pool with a selected CRP concentration near to 10 mg/L, randomly analysed daily during the ordinary laboratory activity. This material had the characteristics previously recommended for correctly deriving the MU of measuring systems due to random effects [1], [10]. In daily practice, the system alignment to the manufacturer’s specifications was checked by measuring, three times per day, i. e. every 8 h, the control materials offered by each manufacturer as part of their CE-marked measuring system and by verifying that results were within the range declared by the manufacturer. Finally, the standard MU associated to clinical samples were calculated as √(ucal 2 + uRw 2) and the system with the best analytical performance identified (Table 2).
Estimated combined standard measurement uncertainty of evaluated measuring systems for C-reactive protein (CRP) measurement.
Measuring system | Intermediate reproducibility data | Combined standard uncertainty of calibrator value (ucal)a, % | Combined standard uncertainty on clinical samplesb, % | ||
---|---|---|---|---|---|
Number of measurements in the 6-month period | Mean CRP, mg/L | Standard uncertainty accounting for random sources (uRw), % | |||
Alinity c | 111 | 10.9 | 4.03 | 3.15 | 5.12 |
Architect c16000 | 147 | 10.2 | 2.06 | 3.15 | 3.76 |
AU680 | 184 | 10.5 | 2.41 | 3.16 | 3.97 |
Cobas c501 | 113 | 10.2 | 2.48 | 4.50 | 5.14 |
aCalculated by combining the standard MU of the employed reference material with the standard MU of commercial calibrator as declared by manufacturers. bEstimated as √(ucal 2 + uRw 2).
Although it is theoretically possible that there is some option on the in vitro diagnostics market performing better, our data show that the Architect c16000 performance in terms of MU on clinical samples (3.76%) may represent the state of the art of the CRP measurement to be employed for the definition of APS for MU according to the model three of the EFLM Strategic Conference. If we consider this APS as desirable, we can also modulate the quality level to minimum goal (3.76% + ½ 3.76% = 5.64%), as previously described [10]. From data in Table 2, all the evaluated systems fulfilled this minimum quality goal for CRP MU. It is however expected that involved manufacturers work for improving the quality of assay performance to move towards the desirable quality goal in the next future. In this context, it is noteworthy the worsening of the analytical performance of CRP measurement between successive generations of platforms when Abbott Diagnostics products are considered. We recommend that manufacturers, before releasing any kind of technological upgrade, should pay major attention to assuring an equivalent, if not better, analytical quality of provided measurements.
In this brief communication, we gave an example of the concepts previously mentioned for deriving APS for MU using the Milan model 3 [10]. The major components of the process are as follows: 1) assess combined standard MU for the selected measurand, based on ucal and random effects for commonly used measuring systems; 2) for random effects assessment, use material(s) other than those used to control the assay alignment and randomly spaced throughout the day; 3) identify the combined standard MU from the best performing system as the ‘desirable’ APS; and 4) establish the ‘minimum’ APS as being 50% greater than the desirable one. One potential limitation of our study was that only one level for pool material was employed, based on the CRP threshold having the major clinical impact (i. e., the upper reference limit). The need of more than one level that spans the reportable range for the test for the optimal estimate of MU has been previously underlined [1], [2]. Another study limitation is the use of only one version of each method, including the determined best system which is used to set the APS. This could not allow optimal capture of the variation in performance that would be seen in various laboratories.
Acknowledgments
We are indebted to Simona Borille, Assunta Carnevale, Concetta Sasso and Sara Zampetti (Clinical Pathology Unit, ASST Fatebenefratelli-Sacco) for collection and processing of IQC data.
-
Research funding: None declared.
-
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
-
Competing interests: Authors state no conflict of interest.
-
Ethical approval: The local Institutional Review Board deemed the study exempt from review.
References
1. Braga, F, Panteghini, M. The utility of measurement uncertainty in medical laboratories [published online ahead of print, 2020 Mar 3]. Clin Chem Lab Med 2020;58:1381-3.10.1515/cclm-2019-1336Search in Google Scholar
2. Braga, F, Panteghini, M. Defining permissible limits for the combined uncertainty budget in the implementation of metrological traceability. Clin Biochem 2018;57:7–11. https://doi.org/10.1016/j.clinbiochem.2018.03.007.Search in Google Scholar PubMed
3. Sandberg, S, Fraser, CG, Horvath, AR, Jansen, R, Jones, G, Oosterhuis, W, et al. Defining analytical performance specifications: consensus Statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine. Clin Chem Lab Med 2015;53:833–5. https://doi.org/10.1515/cclm-2015-0067.Search in Google Scholar PubMed
4. Ceriotti, F, Fernandez-Calle, P, Klee, GG, Nordin, G, Sandberg, S, Streichert, T, et al. Criteria for assigning laboratory measurands to models for analytical performance specifications defined in the 1st EFLM Strategic Conference. Clin Chem Lab Med 2017;55:189–94. https://doi.org/10.1515/cclm-2016-0091.Search in Google Scholar PubMed
5. Mazlam, MZ, Hodgson, HJ. Why measure C reactive protein? Gut 1994;35:5–7. https://doi.org/10.1136/gut.35.1.5.Search in Google Scholar PubMed PubMed Central
6. Braga, F, Panteghini, M. Biologic variability of C-reactive protein: is the available information reliable? Clin Chim Acta 2012;413:1179–83. https://doi.org/10.1016/j.cca.2012.04.010.Search in Google Scholar PubMed
7. Braga, F, Ferraro, S, Szöke, D, Lanzoni, M, Panteghini, M. Estimate of intraindividual variability of C-reactive protein: a challenging issue. Clin Chim Acta 2013;419:85–6. https://doi.org/10.1016/j.cca.2013.02.004.Search in Google Scholar PubMed
8. Franzini, C. Need for correct estimates of biological variation: the example of C-reactive protein. Clin Chem Lab Med 1998;36:131–2. https://doi.org/10.1515/cclm.1998.024.Search in Google Scholar
9. ISO/TS 20914:2019. Medical laboratories – Practical guidance for the estimation of measurement uncertainty, 1st ed. Geneva, Switzerland: ISO, 2019.Search in Google Scholar
10. Braga, F, Pasqualetti, S, Aloisio, E, Panteghini, M. The internal quality control in the traceability era. Clin Chem Lab Med 2020, Article Number: 20200371. Published online: 28 Apr 2020. https://doi.org/10.1515/cclm-2020-0371.Search in Google Scholar PubMed
© 2020 Federica Braga and Mauro Panteghini, published by De Gruyter, Berlin/Boston
This work is licensed under the Creative Commons Attribution 4.0 International License.
Articles in the same Issue
- Frontmatter
- Editorial
- Biotin interference in cardiac troponin immunoassay – where the wild things are?
- Review
- Laboratory-related issues in the measurement of cardiac troponins with highly sensitive assays
- Mini Review
- Chromatographic methods development for clinical practice: requirements and limitations
- Opinion Paper
- Harmonising EQA schemes the next frontier: challenging the status quo
- Genetics and Molecular Diagnostics
- Direct comparison study between droplet digital PCR and a combination of allele-specific PCR, asymmetric rapid PCR and melting curve analysis for the detection of BRAF V600E mutation in plasma from melanoma patients
- A novel mitochondrial m.14430A>G (MT-ND6, p.W82R) variant causes complex I deficiency and mitochondrial Leigh syndrome
- Obesity status modifies the association between rs7556897T>C in the intergenic region SLC19A3-CCL20 and blood pressure in French children
- General Clinical Chemistry and Laboratory Medicine
- Influence of reagent lots and multiple measuring systems on estimating the coefficient of variation from quality control data; implications for uncertainty estimation and interpretation of QC results
- Electrophoretic α1-globulin for screening of α1-antitrypsin deficient variants
- A continued method performance monitoring approach for the determination of pediatric renin samples – application within a European clinical trial
- Pilot study for cystic fibrosis neonatal screening: the Cuban experience
- Validation of the analytical performance of the NOVEOS™ System, a system which improves upon the third-generation in vitro allergy testing technology
- IgE cross-reactivity measurement of cashew nut, hazelnut and peanut using a novel IMMULITE inhibition method
- Sexual dimorphism in the cerebrospinal fluid total protein content
- Current state of the morphological assessment of urinary erythrocytes in The Netherlands: a nation-wide questionnaire
- Reference Values and Biological Variations
- Within-subject and between-subject biological variation of first morning void urine amino acids in 12 healthy subjects
- Proenkephalin as a new biomarker for pediatric acute kidney injury – reference values and performance in children under one year of age
- Hematology and Coagulation
- Quality performance for indirect Xa inhibitor monitoring in patients using international external quality data
- Cardiovascular Diseases
- Clinical risk assessment of biotin interference with a high-sensitivity cardiac troponin T assay
- Short- and long-term biological variation of cardiac troponin I in healthy individuals, and patients with end-stage renal failure requiring haemodialysis or cardiomyopathy
- Infectious Diseases
- Monocyte distribution width (MDW) as a screening tool for sepsis in the Emergency Department
- Performance of a Toxo IgM prototype assay for the diagnosis of maternal and congenital Toxoplasma infections
- Letters to the Editors
- Evaluation of an ELISA for SARS-CoV-2 antibody testing: clinical performances and correlation with plaque reduction neutralization titer
- Preliminary evaluation of Roche Cobas Elecsys Anti-SARS-CoV-2 chemiluminescence immunoassay
- Hypoalbuminemia and elevated D-dimer in COVID-19 patients: a call for result harmonization
- Total pathway to method validation
- Derivation of performance specifications for uncertainty of serum C-reactive protein measurement according to the Milan model 3 (state of the art)
- FGF23 measurement in burosumab-treated patients: an emerging treatment may induce a new analytical interference
- Use of a modified IDS-ISYS intact PTH assay for intraoperative PTH measurements
- Agreement of dried blood spot lyso-Gb3 concentrations obtained from different laboratories in patients with Fabry disease
- Influence of delayed separation of plasma from whole blood and centrifugation protocol on Zn plasma concentration
- A survey of order of draw on inpatient wards and adherence to EFLM-COLABIOCLI recommendations
- Successful implementations of automated minimum re-test intervals to overcome ferritin over-requesting in a Spanish hospital laboratory
- Remarkable pseudoleucocytosis induced by mild cryoglobulinemia
- Massive hemolysis due to Clostridium perfringens: a laboratory’s perspective
Articles in the same Issue
- Frontmatter
- Editorial
- Biotin interference in cardiac troponin immunoassay – where the wild things are?
- Review
- Laboratory-related issues in the measurement of cardiac troponins with highly sensitive assays
- Mini Review
- Chromatographic methods development for clinical practice: requirements and limitations
- Opinion Paper
- Harmonising EQA schemes the next frontier: challenging the status quo
- Genetics and Molecular Diagnostics
- Direct comparison study between droplet digital PCR and a combination of allele-specific PCR, asymmetric rapid PCR and melting curve analysis for the detection of BRAF V600E mutation in plasma from melanoma patients
- A novel mitochondrial m.14430A>G (MT-ND6, p.W82R) variant causes complex I deficiency and mitochondrial Leigh syndrome
- Obesity status modifies the association between rs7556897T>C in the intergenic region SLC19A3-CCL20 and blood pressure in French children
- General Clinical Chemistry and Laboratory Medicine
- Influence of reagent lots and multiple measuring systems on estimating the coefficient of variation from quality control data; implications for uncertainty estimation and interpretation of QC results
- Electrophoretic α1-globulin for screening of α1-antitrypsin deficient variants
- A continued method performance monitoring approach for the determination of pediatric renin samples – application within a European clinical trial
- Pilot study for cystic fibrosis neonatal screening: the Cuban experience
- Validation of the analytical performance of the NOVEOS™ System, a system which improves upon the third-generation in vitro allergy testing technology
- IgE cross-reactivity measurement of cashew nut, hazelnut and peanut using a novel IMMULITE inhibition method
- Sexual dimorphism in the cerebrospinal fluid total protein content
- Current state of the morphological assessment of urinary erythrocytes in The Netherlands: a nation-wide questionnaire
- Reference Values and Biological Variations
- Within-subject and between-subject biological variation of first morning void urine amino acids in 12 healthy subjects
- Proenkephalin as a new biomarker for pediatric acute kidney injury – reference values and performance in children under one year of age
- Hematology and Coagulation
- Quality performance for indirect Xa inhibitor monitoring in patients using international external quality data
- Cardiovascular Diseases
- Clinical risk assessment of biotin interference with a high-sensitivity cardiac troponin T assay
- Short- and long-term biological variation of cardiac troponin I in healthy individuals, and patients with end-stage renal failure requiring haemodialysis or cardiomyopathy
- Infectious Diseases
- Monocyte distribution width (MDW) as a screening tool for sepsis in the Emergency Department
- Performance of a Toxo IgM prototype assay for the diagnosis of maternal and congenital Toxoplasma infections
- Letters to the Editors
- Evaluation of an ELISA for SARS-CoV-2 antibody testing: clinical performances and correlation with plaque reduction neutralization titer
- Preliminary evaluation of Roche Cobas Elecsys Anti-SARS-CoV-2 chemiluminescence immunoassay
- Hypoalbuminemia and elevated D-dimer in COVID-19 patients: a call for result harmonization
- Total pathway to method validation
- Derivation of performance specifications for uncertainty of serum C-reactive protein measurement according to the Milan model 3 (state of the art)
- FGF23 measurement in burosumab-treated patients: an emerging treatment may induce a new analytical interference
- Use of a modified IDS-ISYS intact PTH assay for intraoperative PTH measurements
- Agreement of dried blood spot lyso-Gb3 concentrations obtained from different laboratories in patients with Fabry disease
- Influence of delayed separation of plasma from whole blood and centrifugation protocol on Zn plasma concentration
- A survey of order of draw on inpatient wards and adherence to EFLM-COLABIOCLI recommendations
- Successful implementations of automated minimum re-test intervals to overcome ferritin over-requesting in a Spanish hospital laboratory
- Remarkable pseudoleucocytosis induced by mild cryoglobulinemia
- Massive hemolysis due to Clostridium perfringens: a laboratory’s perspective