Home On the complexity of hemostasis and the need for harmonization of test practice
Article Publicly Available

On the complexity of hemostasis and the need for harmonization of test practice

  • Emmanuel J. Favaloro EMAIL logo and Giuseppe Lippi ORCID logo
Published/Copyright: May 1, 2018

Hemostasis=Love. Everyone talks about it, but no one understands it.

Hemostasis, a homeostatic process that is meant to prevent blood loss, is achieved by the interaction of various systems, sometimes opposing (or acting to balance pro- and antihemostatic forces) [1]. Although hemostasis is sometimes also called “coagulation” by some people, this term actually reflects only a small part of hemostasis. Hemostasis comprises four main “systems”, although it is impossible to fully separate any from another. The four systems are “primary hemostasis”, “secondary hemostasis”, “anticoagulation system” and “fibrinolytic system” [1, 2]. The “primary hemostasis” system is mostly driven by the interaction of blood platelets, adhesive plasma proteins including von Willebrand factor (VWF) and the (damaged) vessel wall comprising endothelium and sub-endothelial matrix components. The “secondary hemostasis” system is mostly driven by “sequential” activation of procoagulant proteins (the so-called “coagulation cascade”) causing generation of thrombin and eventual conversion of fibrinogen to fibrin (the so-called “clot”). The “anticoagulation system” aims to control the process of “secondary hemostasis”, so that it does not progress irrevocably, in which case a thrombosis may occur and lead (for example) to deep vein thrombosis (DVT) or pulmonary embolism (PE). The “fibrinolytic system” acts to further control “secondary hemostasis”, and also to dissolve any clot formed to enable tissue repair. A breakdown in any of these systems will predispose to either bleeding or thrombosis, depending on what element of the process has failed.

As mentioned, “coagulation” is only a part of hemostasis, and essentially only reflective of the “secondary hemostasis” system. This system comprises many coagulation proteins, perhaps most well-known being factors VIII (FVIII) and IX (FIX), the deficiency of which will, respectively, cause hemophilia A and B, thus reflecting the most common inherited bleeding disorders related to secondary hemostasis [3]. The term “coagulation” (or “coag”) laboratory usually refers to laboratories that are capable of performing (a limited number of) “routine” tests of hemostasis, as listed in Table 1.

Table 1:

A summary of hemostasis tests, variables, and why harmonization is required.

Test nameAbbreviationWhat test measuresTest variablesExamples of why standardization and harmonization may improve test reporting
Prothrombin timePTTissue factor pathway of coagulation; sensitive to factors II, V, VII, X, II, IDifferent reagents and instruments will yield different test times. Also, variable sensitivity to certain factorsWill help standardize test results between instruments and reagents and thus across different laboratories. Most common process is to create a PT ratio (e.g. as part of liver disease MELD score) or convert to an INR (for monitoring of VKA therapy). Such modifications provide a more standard means of prioritizing patients for liver transplant MELD score or reduce variability in VKA therapy (INR)
Activated partial thromboplastin timeAPTTContact factor pathway of coagulation; sensitive to factors XII, XI, IX, VIII, V, X, II, IDifferent reagents and instruments will yield different test times. Also, variable sensitivity to certain factors, heparin and lupus anticoagulant (LA)Will help standardize test results between instruments and reagents and thus across different laboratories. One means to help standardization is to create an APTT ratio. Another is to test and group reagents according to various sensitivities. For example, LA sensitive vs. LA insensitive reagents can be defined and partnered for investigation of LA
International normalized ratioINRPT conversion using mathematical formula (INR=[patient PT/MNPT]ISI)PT, mean normal PT (MNPT) and international sensitivity index (ISI)The INR system represents a harmonization process for the PT, which takes into consideration the test reagent and instrument used. However, different laboratories identify different MNPT and ISI values for the same reagent/instrument used because they use different ways of deriving these values. Thus, further harmonization of practice is needed to further reduce INR variation and improve VKA therapy
FibrinogenfibFibrinogen. Potentially sensitive to afibrinogenemia dysfibrinogenemia hypofibrinogenemiaTest methodology – von Clauss, derived or antigenFunctional assays using a von Clauss procedure are the most relevant assays, and thus harmonization to such assays is preferred. Derived assays are cheaper, and so may be offered instead. Antigenic assays only useful in select situations (e.g. to identify dysfibrinogenemia)
D-dimerD-DFibrin breakdown products containing D-D domains. Potentially applicable to investigation of DVT, PE, DICTest methodology (reagents) and reporting unitsDifferent test reagents measure D-dimers differently (i.e. variably sensitive to D-dimer fragments), thereby yielding different values for the same sample, and potentially different conclusions regarding “positive” findings (e.g. suggestive of thrombosis or not), and therefore leading to different clinical interventions. Reporting of different units complicates comparison between research studies and some clinicians may misunderstand whether a sample is positive or not. Harmonization will improve clinical management as it will decrease variability of measurement and response to test results
Thrombin timeTTFibrinogen conversion to fibrin. Potentially sensitive to afibrinogenemia dysfibrinogenemia hypofibrinogenemia and anticoagulants, including unfractionated heparin and dabigatranDifferent reagents and instruments will yield different test timesWill help standardize test results between instruments and reagents and thus across different laboratories. This will improve clinical patient management
Protein CPCCongenital deficiencies in PCFunctional clotting and/or chromogenic assays. Chromogenic assays are preferred. Clotting assays may be sensitive to factors other than PCWill help standardize test results between across different laboratories, and improve identification/exclusion of PC deficiency
Protein SPSCongenital deficiencies in PSFunctional clotting and/or antigenic assays. Antigenic assays for Free PS are preferred. Clotting assays may be sensitive to factors other than PSWill help standardize test results between across different laboratories, and improve identification/exclusion of PS deficiency
AntithrombinATCongenital deficiencies ATChromogenic or antigenic assays. Chromogenic assays are preferred. Antigenic assays may not identify functional defects in ATWill help standardize test results between across different laboratories, and improve identification/exclusion of AT deficiency
Activated protein C resistanceAPCRCongenital or acquired thrombophilia (APCR, FVL)APTT or RVVT based assays. These have different sensitivities to APCR and FVLWill help standardize test results between across different laboratories, and improve identification of thrombophilia
Anti-Factor XaAnti-XaMonitoring LMWH, or measuring the level of apixaban and rivaroxabanChromogenic assay using assay specific calibratorsWill help standardize test results between across different laboratories, and improve clinical patient management
Dilute thrombin timedTTMeasuring the level of dabigatranFunctional clotting assayWill help standardize test results between across different laboratories, and improve clinical patient management
Dilute Russell’s Viper venom time (screen and/or confirm)dRVVTsc, dRVVTconIdentification of LAFunctional clottingWill help standardize test results between across different laboratories, and improve identification of thrombophilia
Silica clotting timeSCTsc, SCTconIdentification of LAFunctional clottingWill help standardize test results between across different laboratories, and improve identification of thrombophilia
VWF:AntigenVWF:AgMeasures the quantity of VWF protein (not its functionality)ELISA, LIA, or CLIA assayWill help standardize test results between across different laboratories, and improve the diagnosis and management of VWD
VWF:Ristocetin co-factorVWF:RCoIdentification of Type 2 VWD; sensitive to the loss of the HMW forms of VWF; sensitive to VWF mutations affecting GPIb bindingAssesses the ability of VWF to bind to platelets (via GPIb) in the presence of ristocetin. Usually platelet aggregation or LIA assayWill help standardize test results between across different laboratories, and improve the diagnosis and management of VWD
VWF:Collagen bindingVWF:CBIdentification of Type 2 VWD; sensitive to the loss of the HMW forms of VWF; sensitive to VWF mutations affecting collagen bindingAssesses the ability of VWF to bind to sub-endothelial matrix proteins (namely collagen). ELISA or CLIA assayWill help standardize test results between across different laboratories, and improve the diagnosis and management of VWD
VWF:ActivityVWF:ActIdentification of Type 2 VWD; may be sensitive to loss of HMW forms of VWF; may be sensitive to various VWF mutations affecting VWF functionGeneric term for measurement of VWF “activity” that otherwise not defined by specific activity. Usually ELISA or LIA assayWill help standardize test results between across different laboratories, and improve the diagnosis and management of VWD
VWF:FVIII bindingVWF:FVIIIBIdentification or exclusion of Type 2N VWD; sensitive to loss of VWF-FVIII binding functionUsually an ELISA based assayWill help standardize test results between across different laboratories, and improve the diagnosis and management of VWD
VWF multimersNAAssesses structural composition/defects of the VWF moleculeImmunological assay (complex assay consisting of several steps). Typically gel electrophoresisWill help standardize test results between across different laboratories, and improve the diagnosis and management of VWD
Platelet function screenPFA-100/200Identification of VWD, platelet function defects or antiplatelet medicationsPFA-100 or PFA-200 instrumentWill help standardize test results between across different laboratories, and improve the identification/exclusion of VWD and platelet dysfunction
Platelet function testingNAIdentification and characterization of VWD, platelet function defects or antiplatelet medicationsLight transmission or whole blood aggregometryWill help standardize test results between across different laboratories, and improve the identification/exclusion of platelet disorders
Factor assaysFII, FV, FVII, FVIII, FIX, FX, FXI, FXIIIdentification of hemophilia (FVIII, FIX), or other specific factor deficienciesClotting assays (modification of PT or APTT assays); chromogenic assays for FVIII and FIXWill help standardize test results between across different laboratories, and improve the identification/exclusion of hemophilia and other bleeding disorders/factor deficiencies
Factor inhibitorNAFVIII or FIX (or other factor) inhibitorsPT/APTT based Bethesda or Nijmegen modification assaysWill help standardize test results between across different laboratories, and improve the identification/exclusion of acquired hemophilia
  1. CLIA, chemiluminescence assay; DIC, disseminated intravascular coagulation; DVT, deep vein thrombosis; ELISA, enzyme-linked immunosorbent assay; FVL, factor V Leiden; HMW, high molecular weight; ISI, international normalized ratio; LIA, latex immune assay; LMWH, low molecular weight heparin; MNPT, mean normal prothrombin time; PE, pulmonary embolism; UFH, unfractionated heparin; VKA, vitamin K antagonist.

The four hemostasis systems mentioned are not independent of each other, and all processes intertwine. For example, although VWF is primarily involved in “primary hemostasis”, because it binds to, and protects, and also delivers FVIII to sites of vascular damage, it is also integral to the success of “secondary hemostasis”. This is why some patients with deficiency or defect in VWF, thereby suffering von Willebrand disease (VWD), may experience bleeding considered due to defects in “primary hemostasis” as well as “secondary hemostasis” [4]. Similarly, fibrinogen is primarily involved in “secondary hemostasis”, generating the clot after conversion to fibrin. However, fibrinogen can also bind to activated platelets and, thus, contribute to “primary hemostasis”.

The modern hemostasis laboratory has an arsenal of tests it can perform for investigation of hemostasis (Table 1). With these tests, laboratories can help diagnose bleeding disorders such as hemophilia or VWD, explain the occurrence of familial thrombophilia (e.g. deficiency in proteins of the anticoagulant system) or manage patients under therapy (e.g. patients being treated for a DVT with vitamin K antagonist [VKA] therapy). However, the ability of laboratories to help clinicians diagnose hemostatic “disease” or better manage patients is compromised by lack of standardization and harmonization. Different laboratories may have the appearance of having the “same tests” in their repertoire, but given the high variety of test reagents and methods available for any given test, it is unlikely they will all perform these test in the same way. This variability will compromise patient management because different test results can arise from testing the same sample. As test result will typically direct further clinical management, different data may lead to different clinical actions.

Moreover, even using the same test reagents and instruments can still lead to differences in test results. As a pragmatic example, consider the routine tests “prothrombin time” (PT) and “activated partial thromboplastin time” (APTT). The PT is used for several reasons, including evaluation of the tissue factor pathway of coagulation, for prioritizing liver transplantation in end stage liver disease (usually as a PT ratio of patient PT/normal PT) and for monitoring patients on VKA therapy (usually as an international normalized ratio [INR]) (Table 1). The APTT is also used for several reasons, including evaluation of the contact factor pathway of coagulation, as a screen for hemophilia, for identification of lupus anticoagulant (LA; a prothrombotic disorder) and for monitoring of anticoagulant therapy with unfractionated heparin. First, different reagents and different instruments, and the arising various combinations, can yield vastly different PT and APTT values among laboratories. Second, although normalizing test values by use of test ratios (as a form of test harmonization) or producing an INR will help reduce this variability, it will not completely eliminate it. Indeed, laboratories using the same instrument and the same test reagent (even the same lot of reagent) may still generate highly disparate INR values, simply because they have calculated the components of the INR equation (namely, mean normal prothrombin time [MNPT] and international sensitivity index [ISI]) in different ways [5, 6]. For patients on VKA therapy, the recommended INR is usually around 2.0–3.0 for most indications. In some cases, as identified in external quality assessment (EQA) surveys, different laboratories can yield differences in INR that can range from 2.4 to over 6.0 for the sample homogenous sample [5], as also shown in Figure 1. A value of 2.4 would be considered therapeutic, and no clinical intervention would be needed (the patient would likely continue the same dose of VKA until next testing). However, an INR value of 6 is typically considered an alert (critical) value [7] and would suggest clinical intervention – reducing dosage and possibly close monitoring for potential bleed events. Although the wide range of reported INRs is influenced by outlier INRs, most INRs fall in the region of 3.3–4.5 for this example, which may still represent different clinical responses. As explanation for the wide INR range, it can be reflected that participant reported ISI and MNPT values were also widely variable and could have theoretically led to the wide INR values, as highlighted in Figure 1. Similarly, for the APTT, a heparin therapeutic range can be exceeded or not for the same sample as tested in different laboratories [8]. Moreover, the new direct oral anticoagulants may also (variably) affect routine coagulation tests such as PT and APTT, and/or these tests may be used to help identify the presence or absence of such drugs [9, 10].

Figure 1: An example of external quality assessment (EQA) for an individual sample as assessed in a local EQA program in Australia (the RCPAQAP Hematology [7]).A wide range of INR values was reported by laboratories for the same homogeneous sample.
Figure 1:

An example of external quality assessment (EQA) for an individual sample as assessed in a local EQA program in Australia (the RCPAQAP Hematology [7]).

A wide range of INR values was reported by laboratories for the same homogeneous sample.

As another example, D-dimer testing is plagued not only by differences in test reagents and methods, leading to differences in test results for the same sample, but also by inconsistency in reporting units [11, 12]. In fact, 28 different combinations of units are feasible, thereby acting to considerably confuse clinicians who order these tests and thereafter manage patients. Notably then, based on differing reagents, the same sample may be identified as “positive” or “negative” for D-dimer based on different test values. However, even with the same reagent, the same sample may also be interpreted as “positive” or “negative” by different clinicians for the same numerical value, depending on their experience/inexperience with the reported unit [13].

Another final example is VWF testing for VWD diagnosis/exclusion. VWF is a large and complex protein with multiple functionalities, including binding to platelets, subendothelial matrix components such as collagen, and binding to FVIII (thereby protecting it from degradation and also delivering it to sites of injury). VWD can arise from defects in any of these activities. Thus, identification/exclusion/characterization of VWD requires the use of multiple assays [14]. This may not seem a problem because indeed an arsenal of potential assays exist in laboratories (Table 1). However, each of these assays measures different aspects of VWF (level/different activities), and laboratories may use many different combinations of assays as test panels or may not understand strengths and limitations of the assays they use individually or in composite, thereby leading to many (potentially avoidable) errors in VWD diagnosis [15]. Moreover, these same assays may be used to monitor therapy in VWD, with the same limitations applied in terms of current knowledge, and corresponding adverse outcomes in regards to differential therapies [4, 16].

Standardization and harmonization is intended to reduce the variability of laboratory test practice and, thus, to provide more consistent standards of health care. In this issue of the journal, there are several papers describing various initiatives that attempt to drive standardization and harmonization of hemostasis test practice at local, regional or international levels [17, 18].

Hemostasis is complex. Standardization and harmonization aims to make understanding hemostasis less complex and tests of hemostasis more useful. We hope readers enjoy this issue of the journal, and in particular, the papers related to harmonization of hemostasis test practice.


Corresponding author: Dr. Emmanuel J. Favaloro, Department of Haematology, Institute of Clinical Pathology and Medical Research (ICPMR), NSW Health Pathology, Westmead Hospital, Westmead, NSW, 2145, Australia; and Sydney Centres for Thrombosis and Haemostasis, Westmead, NSW, Australia, Phone: +(612) 8890 6618, Fax: +(612) 9689 2331

  1. Author contributions: EJF wrote the original manuscript. Both coauthors contributed to manuscript revision and have approved the final manuscript for publication. All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Disclosure: The authors state that they have no interest that might be perceived as posing a conflict or bias.

References

1. Bonar RA, Lippi G, Favaloro EJ. Overview of hemostasis and thrombosis and contribution of laboratory testing to diagnosis and management of hemostasis and thrombosis disorders. Methods Mol Biol 2017;1646:3–27.10.1007/978-1-4939-7196-1_1Search in Google Scholar PubMed

2. Lippi G, Favaloro EJ. Laboratory hemostasis: from biology to the bench. Clin Chem Lab Med 2018 Feb 19; doi: 10.1515/cclm-2017-1205. [Epub ahead of print].10.1515/cclm-2017-1205Search in Google Scholar PubMed

3. Lippi G, Franchini M, Favaloro EJ. Diagnostics of inherited bleeding disorders of secondary hemostasis: an easy guide for routine clinical laboratories. Semin Thromb Hemost 2016;42:471–7.10.1055/s-0036-1571311Search in Google Scholar PubMed

4. Curnow J, Pasalic L, Favaloro EJ. Treatment of von Willebrand disease. Semin Thromb Hemost 2016;42:133–46.10.1055/s-0035-1569070Search in Google Scholar PubMed

5. Bonar R, Favaloro EJ. Explaining and reducing the variation in inter-laboratory reported values for International Normalised Ratio. Thromb Res 2017;150:22–9.10.1016/j.thromres.2016.12.007Search in Google Scholar PubMed

6. Tripodi A, Lippi G, Plebani M. How to report results of prothrombin and activated partial thromboplastin times. Clin Chem Lab Med 2016;54:215–2.10.1515/cclm-2015-0657Search in Google Scholar PubMed

7. Lippi G, Adcock D, Simundic AM, Tripodi A, Favaloro EJ. Critical laboratory values in hemostasis: toward consensus. Ann Med 2017;49:455–61.10.1080/07853890.2016.1278303Search in Google Scholar PubMed

8. Baluwala I, Favaloro EJ, Pasalic L. Therapeutic monitoring of unfractionated heparin – trials and tribulations. Expert Rev Hematol 2017;10:595–605.10.1080/17474086.2017.1345306Search in Google Scholar PubMed

9. Favaloro EJ, Pasalic L, Curnow J, Lippi G. Laboratory monitoring or measurement of direct oral anticoagulants (DOACs): advantages, limitations and future challenges. Curr Drug Metab 2017;18:598–608.10.2174/1389200218666170417124035Search in Google Scholar PubMed

10. Favaloro EJ, Pasalic L, Lippi G. Replacing warfarin therapy with the newer direct oral anticoagulants, or simply a growth in anticoagulation therapy? Implications for Pathology testing. Pathology 2017;49:639–43.10.1016/j.pathol.2017.04.011Search in Google Scholar PubMed

11. Thachil J, Lippi G, Favaloro EJ. D-Dimer testing: laboratory aspects and current issues. Methods Mol Biol 2017;1646:91–104.10.1007/978-1-4939-7196-1_7Search in Google Scholar PubMed

12. Longstaff C, Adcock D, Olson JD, Jennings I, Kitchen S, Mutch N, et al. Harmonisation of D-dimer – A call for action. Thromb Res 2016;137:219–20.10.1016/j.thromres.2015.11.031Search in Google Scholar PubMed

13. Lippi G, Tripodi A, Simundic AM, Favaloro EJ. International survey on D-dimer test reporting: a call for standardization. Semin Thromb Hemost 2015;41:287–93.10.1055/s-0035-1549092Search in Google Scholar PubMed

14. Favaloro EJ, Pasalic L, Curnow J. Laboratory tests used to help diagnose von Willebrand disease: an update. Pathology 2016;48:303–18.10.1016/j.pathol.2016.03.001Search in Google Scholar PubMed

15. Favaloro EJ, Bonar RA, Meiring M, Duncan E, Mohammed S, Sioufi J, et al. Evaluating errors in the laboratory identification of von Willebrand disease in the real world. Thromb Res 2014;134:393–403.10.1016/j.thromres.2014.05.020Search in Google Scholar PubMed

16. Favaloro EJ, Pasalic L, Curnow J. Monitoring therapy during treatment of von Willebrand disease. Semin Thromb Hemost 2017;43:338–54.10.1055/s-0036-1585080Search in Google Scholar PubMed

17. Favaloro EJ, Jennings I, Olson J, Van Cott EM, Bonar R, Gosselin R, et al. Towards harmonization of external quality assessment/proficiency testing in hemostasis. Clin Chem Lab Med 2018.10.1515/cclm-2018-0077Search in Google Scholar PubMed

18. Favaloro EJ, Gosselin R, Olson J, Jennings I, Lippi G. Recent initiatives in harmonization of hemostasis practice. Clin Chem Lab Med 2018;56:1608–19.10.1515/cclm-2018-0082Search in Google Scholar PubMed

Published Online: 2018-05-01
Published in Print: 2018-09-25

©2018 Walter de Gruyter GmbH, Berlin/Boston

Articles in the same Issue

  1. Frontmatter
  2. Editorials
  3. Harmonization in laboratory medicine: Blowin’ in the wind
  4. Standardization and harmonization of autoimmune diagnostics
  5. On the complexity of hemostasis and the need for harmonization of test practice
  6. Harmonization of laboratory hematology: a long and winding journey
  7. Section 1: Current Harmonization Activities at Global Level
  8. Harmonization in laboratory medicine: more than clinical chemistry?
  9. Harmonization of External Quality Assessment Schemes and their role – clinical chemistry and beyond
  10. An overview of EFLM harmonization activities in Europe
  11. Metrological traceability and harmonization of medical tests: a quantum leap forward is needed to keep pace with globalization and stringent IVD-regulations in the 21st century!
  12. Assessment of bone turnover in osteoporosis: harmonization of the total testing process
  13. Recent initiatives in harmonization of hemostasis practice
  14. EASI – European Autoimmunity Standardisation Initiative: facing the challenges of diagnostics in autoimmunity
  15. Harmonization of microbiology processes and standards: work in progress
  16. Harmonization initiatives in the generation, reporting and application of biological variation data
  17. Harmonization of accreditation to ISO15189
  18. External quality assessment programs in the context of ISO 15189 accreditation
  19. Section 2: Pre-Pre and Pre-Analytical Phase
  20. Laboratory testing in the emergency department: an Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC) and Academy of Emergency Medicine and Care (AcEMC) consensus report
  21. The EFLM strategy for harmonization of the preanalytical phase
  22. Section 3: The Analytical Phase
  23. The roadmap for harmonization: status of the International Consortium for Harmonization of Clinical Laboratory Results
  24. The quest for equivalence of test results: the pilgrimage of the Dutch Calibration 2.000 program for metrological traceability
  25. Current state and recommendations for harmonization of serum/plasma 17-hydroxyprogesterone mass spectrometry methods
  26. International normalized ratio (INR) testing in Europe: between-laboratory comparability of test results obtained by Quick and Owren reagents
  27. Detecting molecular forms of antithrombin by LC-MRM-MS: defining the measurands
  28. A design for external quality assessment for the analysis of thiopurine drugs: pitfalls and opportunities
  29. Harmonization of PCR-based detection of intestinal pathogens: experiences from the Dutch external quality assessment scheme on molecular diagnosis of protozoa in stool samples
  30. Harmonization of urine albumin/creatinine ratio (ACR) results: a study based on an external quality assessment program in Polish laboratories
  31. Standardization of autoimmune testing – is it feasible?
  32. Diagnostic laboratory tests for systemic autoimmune rheumatic diseases: unmet needs towards harmonization
  33. Clinically relevant discrepancies between different rheumatoid factor assays
  34. An international survey on anti-neutrophil cytoplasmic antibodies (ANCA) testing in daily clinical practice
  35. Predictive autoimmunity using autoantibodies: screening for anti-nuclear antibodies
  36. Harmonization in autoimmune thyroid disease diagnostics
  37. International consensus on antinuclear antibody patterns: definition of the AC-29 pattern associated with antibodies to DNA topoisomerase I
  38. Reference standards for the detection of anti-mitochondrial and anti-rods/rings autoantibodies
  39. International Consensus on Antinuclear Antibody Patterns: defining negative results and reporting unidentified patterns
Downloaded on 10.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2018-0174/html
Scroll to top button