Relationship between plasma and salivary melatonin and cortisol investigated by LC-MS/MS

  • Martijn van Faassen EMAIL logo , Rainer Bischoff und Ido P. Kema
Veröffentlicht/Copyright: 20. Dezember 2016
Veröffentlichen auch Sie bei De Gruyter Brill

Abstract

Background:

Disturbance of the circadian rhythm has been associated with disease states, such as metabolic disorders, depression and cancer. Quantification of the circadian markers such as melatonin and cortisol critically depend on reliable and reproducible analytical methods. Previously, melatonin and cortisol were primarily analyzed separately, mainly using immunoassays.

Methods:

Here we describe the validation and application of a high-throughput liquid chromatography in combination with mass spectrometry (LC-MS/MS) method for the combined analysis of melatonin and cortisol in plasma and saliva. The LC-MS/MS method was validated according to international validation guidelines. We used this method to analyze total plasma, free plasma (as obtained by equilibrium dialysis) and saliva melatonin and cortisol in healthy adults.

Results:

Validation results for plasma and saliva melatonin and cortisol were well within the international validation criteria. We observed no difference between saliva collected by passive drooling or Salivette. Moreover, we noted a significant difference in saliva vs. free plasma melatonin. We observed on average 36% (95% CI: 4%–60%) higher salivary melatonin levels in comparison to free plasma melatonin, suggestive of local production of melatonin in the salivary glands.

Conclusions:

The novel outcome of this study is probably due to the high precision of our LC-MS/MS assay. These outcomes illustrate the added value of accurate and sensitive mass spectrometry based methods for the quantification of neuroendocrine biomarkers.

Introduction

There is mounting evidence that disturbance of circadian rhythm is involved in the etiology of several disease states such as depression, the metabolic syndrome and cancer [1], [2]. Typical circadian rhythm markers are melatonin and cortisol. Melatonin, a pineal hormone and metabolite of the tryptophan-serotonin-pathway is best known for its role as a signaling molecule for the length of day and night [3], [4]. Cortisol, an adrenal cortex derived glucocorticoid hormone, is at the end-point of the hypothalamic-pituitary-adrenal axis, and adapts the body to stress conditions by mobilizing energy and inhibiting non-emergency processes, such as sleep, sexual activity and growth [5]. The cortisol response is further related to awakening, as cortisol peaks after waking up [6].

Quantification of melatonin is still mainly performed by immunochemical methods, whereas in recent years cortisol determination has shifted towards liquid chromatography in combination with mass spectrometry (LC-MS/MS). This discrepancy in the methods used is probably due to the fact that cortisol is a frequently used diagnostic marker in routine patient care (for Cushing’s syndrome) [7], [8]. While a few reports have been published on melatonin analysis in saliva or plasma by LC-MS/MS, validation was not performed according to internationally accepted guidelines [9], [10], [11], [12]. When compared to immunoassays LC-MS/MS has significant advantages due its better analytical reproducibility, specificity and standardization. Immunoassays tend to lack specificity and show variability between vendors [13], [14], [15], [16]. Moreover, LC-MS/MS has the possibility of measuring multiple analytes in one assay, whereas immunoassays are mainly single-analyte assays. In circadian rhythm studies measurements of melatonin and cortisol are often performed separately, where it would be more effective to combine measurements of these analytes thereby saving sample, time and costs [17], [18], [19], [20].

Circadian rhythm studies usually have multiple sampling points on consecutive days and therefore saliva is preferred, as its sampling is noninvasive and more practical than blood sampling. Another advantage of saliva is, that it reflects the free non-protein bound fraction of plasma, which is considered the biologically active fraction [21]. For cortisol, the relationship between saliva and plasma free cortisol plasma has been extensively studied, and late night salivary cortisol is now routinely used as diagnostic test for Cushing’s syndrome [22], [23], [24]. However, to our knowledge there is only one study which describes the relationship between total plasma, free plasma, and salivary melatonin, which uses an immunoassay [25]. In this study, we investigated the correlations between total plasma, free plasma and salivary cortisol and melatonin levels by a fully validated LC-MS/MS method and show for the first time a significant difference in saliva vs. free plasma melatonin.

Participants and methods

Materials

Water, acetonitrile, methanol, formic acid and ammonium acetate were obtained from Biosolve and were ULC/MS grade. Ascorbic acid was purchased from Merck. Isopropanol, cortisol, melatonin, cortisone, prednisone and prednisolone were purchased from Sigma-Aldrich. Deuterated analogs of cortisol (d4) and melatonin (d4) were purchased from C/D/N Isotopes.

Sample collection and preparation

Morning blood and saliva samples (8:00 AM) were collected simultaneously in 10 non-smoking volunteers. Saliva was collected in two ways, either by passive drooling into polypropylene tubes or by cortisol Salivettes® (Sarstedt, Nümbrecht, Germany). Participants did not eat or drink 30 min before saliva sampling. Salivette rolls were kept in the mouth for 1 min and centrifuged at 1000 g for 2 min. One milliliter of saliva of passive drooling and Salivette was collected and stored at –20°C until analysis. Before analysis, saliva was centrifuged for 5 min at 2000 g. Blood samples were taken via venipuncture, using 4 mL Vacutainer Tubes (Becton, Dickinson and Company, Franklin Lakes, NJ, USA) containing K2-EDTA as anticoagulant. Blood samples were centrifuged at 2500 g for 15 min and plasma was stored at –20°C until analysis. Free melatonin and cortisol were determined in plasma dialysate obtained by equilibrium dialysis essentially as described by Fiers et al. for testosterone [26], with the exception that the dialysis membranes used in our study had a 10 kDa pore size instead of 25 kDa pore size, as cortisol binding globulin is significantly lower in molecular weight than sex hormone binding globulin, 50 and 90 kDa, respectively.

The study was evaluated by the Local Medical Ethics Committee and waived, because the purpose of this study was to compare melatonin and cortisol levels in saliva to the values in plasma and the burden for the participants was considered to be very low. All participants gave written informed consent.

Analytical procedures

Stock standards for melatonin and cortisol were prepared in methanol at a concentration of 1.0 g/L (melatonin 4.31 mmol/L, cortisol 2.76 mmol/L). Eight individual calibrators were prepared fresh from the stock solution in 0.04% (v/v) ascorbic acid on the day of analysis. Calibration curves for saliva were prepared in 0.04% ascorbic acid and for plasma in blank plasma. Plasma depleted of melatonin and cortisol was prepared by extensive dialyzing plasma against PBS using a 10 kDa molecular weight cut-off membrane.

For saliva the calibrator ranges were: melatonin 15–2000 pmol/L, and cortisol 0.60–75 nmol/L. For plasma: melatonin 15–2000 pmol/L, and cortisol 40–5500 nmol/L. Internal standard working solutions for saliva (melatonin 800 pmol/L, cortisol 100 nmol/L) and plasma (melatonin 800 pmol/L, cortisol 1000 nmol/L) were prepared on the day of analysis in 0.04% (v/v) ascorbic acid. Quality control samples were prepared by pooling several saliva or plasma samples and were, when necessary, spiked with melatonin and cortisol to give three pools (low, medium and high). All samples were stored at −70°C until use.

For analysis, 250 μL saliva/dialysate, 50 μL internal standard solution and 200 μL 0.04% ascorbic acid were added to wells of a 2.0 mL polypropylene 96-deep well plate (Greiner Bio-One). The plate was sealed before being vortex mixed for 10 min. Following vortexing, the plate was transferred to the autosampler and 100 μL was injected for analysis.

For plasma analysis, 200 μL of plasma was pipetted into a 2.0 mL 96-deep well plate (Greiner Bio-One). To this, 50 μL of internal standard solution was added, and the plate was vortex mixed for 1 min. Proteins were precipitated by adding 200 μL of 0.3 M zinc sulfate in methanol (1:5 v/v) and the plate was vortexed again for 1 min. Water was added to fill up to 1.0 mL and the plate was sealed and centrifuged for 15 min at 2500× g. Following centrifugation, the plate was transferred to the autosampler and 100 μL of the supernatant was injected for analysis.

Online solid phase extraction and LC-MS/MS

Online solid phase extraction (SPE) was performed using the fully automated Symbiosis™ system (Spark Holland, Emmen, The Netherlands) in eXtraction liquid chromatography (XLC) mode as previously described [27]. Oasis HLB (Waters, Milford, USA) cartridges were used for extraction. Liquid chromatography (LC) was performed on a Phenomenex® Luna Phenyl-Hexyl 2.0×100 mm, 3 μm column (Phenomenex, Torrance, CA, USA) and total run time was 6.5 min, including online SPE cleanup. For a complete description of the online SPE and LC method see Supplemental Data. Melatonin and cortisol were analyzed in positive ionization mode on a Waters® Xevo TQ-MS triple quadrupole mass spectrometer. Mass spectrometer settings were optimized by tuning in the selective reaction monitoring mode (SRM). The following settings were applied throughout the study: capillary voltage 0.5 kV, desolvation temperature 600°C, desolvation gas flow 1000 L/h, cone gas flow 50 L/h and collision gas flow 0.15 mL/min. Cone voltage and collision energies were optimized for all transitions and are listed in Table 1. Quantifier and qualifier m/z transitions were monitored for melatonin and cortisol as well as for their internal standards melatonin-d4 and cortisol-d4. Quantitation was performed by using the peak-area response ratios of the quantifier transitions for cortisol:cortisol-d4 and melatonin:melatonin-d4 using the Targetlynx software.

Table 1:

Transitions, cone voltage and collision energy for quantifier and qualifier of each compound.

Compoundm/zCone voltage, VCollision energy, eV
Melatonin-quantifier233.2>174.11814
Melatonin-d4237.2>178.11814
Melatonin-qualifier233.2>159.11827
Melatonin-d4237.2>163.11827
Cortisol-quantifier363.3>121.12523
Cortisol-d4367.3>121.12523
Cortisol-qualifier363.3>327.32516
Cortisol-d4363.3>331.12516

Analytical validation

Method validation was performed for saliva and plasma to the same extent, unless otherwise stated and based on the guideline for bioanalytical method validation from the European Medicines Agency [28].

Linearity of the calibration curves for melatonin and cortisol was assessed by analyzing the calibration curves on 12 different days. Curves were plotted using least-squares linear regression and checked for linearity. To test whether melatonin and cortisol behaved the same in water-based calibrators as in saliva-based calibrators, we spiked calibration curves in six different saliva samples to compare the slope in comparison to water-based calibrators.

Intra-assay variation was determined by analyzing three pooled saliva or plasma samples with melatonin and cortisol at low, medium and high concentrations in 20 replicates. Inter-assay variation was determined by analyzing three pooled saliva or plasma samples on 12 different days in duplicate over an 8-week period.

Quantification limits for melatonin and cortisol were determined by serial dilution of a saliva sample (melatonin 20 pmol/L, cortisol 0.4 nmol/L,) and plasma sample (melatonin 20 pmol/L, cortisol 130 nmol/L) and analyzing the dilutions on six different days in duplicate (for plasma on three different days). LLOQ was set where the precision was <20% and the signal to noise ratio >10.

Recovery was estimated by spiking melatonin and cortisol at three different levels to three different pooled saliva and plasma samples containing endogenous melatonin and cortisol at three different concentration levels. These samples were analyzed on six different days (for plasma on three different days). Recovery percentage was calculated as follows: [(final concentration – initial concentration)/added concentration]×100%. Recovery was considered acceptable between 85 and 115%.

Ion suppression was checked by performing a post-column infusion experiment [29]. Solvent, saliva and plasma samples (n=2) containing a low concentration of melatonin and cortisol were analyzed as described above with constant post-column infusion of melatonin and cortisol at a flow-rate of 10 μL/min (melatonin and melatonin-d4 at 800 pmol/L, cortisol, cortisol-d4 at 20 nmol/L). Chromatograms of the samples were compared with those of the solvent blank.

Cortisone, prednisone and prednisolone are known compounds that can interfere with cortisol analysis. These compounds were analyzed at a concentration of 100 nmol/L and checked for interference in the cortisol analysis. For melatonin, n-acetylserotonin, 6-hydroxymelatonin and 6-sulfatoxymelatonin were analyzed at a concentration of 100 nmol/L and checked for interference in the melatonin analysis.

Method comparison

Thirty-five anonymized saliva samples, which were analyzed in routine patient care for cortisol by an existing XLC-MS/MS method, were also analyzed by the new method. This routine XLC-MS/MS method was developed in our lab based on the method described by Jones et al. using C18 cartridges for online SPE and a 3.0×150 mm phenyl-hexyl column [30]. For salivary melatonin, we compared our method to a commercial immunoassay from IBL International (Hamburg, Germany) using the same saliva samples as mentioned above. Bland-Altman and Passing-Bablok regression were used for evaluation of the results using RStudio [31], [32], [33]. Ninety-five percent confidence intervals (CI) of slope and intercept were calculated to investigate if there was a significant difference (CI of slope should contain 1, CI of intercept should contain 0).

Impact of collection device and correlation between plasma total, plasma free and salivary melatonin and cortisol

Saliva was collected by passive drooling (PD) and by Salivette (SAL), and plasma from 10 healthy volunteers (six females and four males, with respective mean age of 31 and 35 years) to investigate the potential influence of the collection device on cortisol and/or melatonin. Furthermore, we analyzed total and free plasma, and salivary melatonin and cortisol to investigate the correlations in these matrices.

Statistics

Statistical analyses were performed using RStudio [34]. Non-parametric correlations (Spearman’s ρ) were calculated to investigate the relationship between total plasma free plasma, and saliva melatonin and cortisol. A two-sided p<0.05 was considered statistically significant. Passing-Bablok regression analysis was performed to examine the distribution of melatonin and cortisol between the matrices. Ninety-five percent confidence intervals (CI) of slope and intercept were calculated to investigate if there was a significant difference (CI of slope should contain 1, CI of intercept should contain 0).

Results

Assay performance

Total analysis time including automated sample extraction using the online SPE step was 6.5 min. Representative chromatograms obtained for saliva, plasma and plasma dialysate are presented in Figure 1. Chromatographic selectivity was achieved by using a phenyl-hexyl column which baseline separates cortisone from cortisol and minimizes interference from prednisone (0%) and prednisolone (0.4% at 100 nmol/L prednisolone) while maintaining short run times. The analytes n-acetylserotonin, 6-hydroxymelatonin and 6-sulfatoxymelatonin did not show any interference in the melatonin assay. Melatonin and cortisol were not baseline separated, but this causes no erroneous results as detection is based on different mass over charge ratios of the precursor ions and the product ions. Melatonin eluted first at 3.26 min, followed by cortisol at 3.37 min.

Figure 1: Representative chromatograms of cortisol and melatonin in plasma, saliva and plasma dialysate from the same individual.(A) Traces of the plasma sample (cortisol: 429 nmol/L; melatonin 84 pmol/L). (B) Traces of the saliva sample (cortisol: 8.4 nmol/L; melatonin 24 pmol/L) and panel C shows the traces of the plasma dialysate (cortisol: 23 nmol/L; melatonin 24 pmol/L).
Figure 1:

Representative chromatograms of cortisol and melatonin in plasma, saliva and plasma dialysate from the same individual.

(A) Traces of the plasma sample (cortisol: 429 nmol/L; melatonin 84 pmol/L). (B) Traces of the saliva sample (cortisol: 8.4 nmol/L; melatonin 24 pmol/L) and panel C shows the traces of the plasma dialysate (cortisol: 23 nmol/L; melatonin 24 pmol/L).

Method validation

Calibration curves were linear over the calibration range for both melatonin and cortisol over n=12 days with correlation coefficients, R2>0.99. Intra-assay CV was below 9% for cortisol and below 6% for melatonin at all three QC levels for saliva and plasma. Inter-assay CV was below 10% for cortisol and below 9% for melatonin at the three QC levels (see Table 2). Complete results for the intra- and inter-assay variation of the method for saliva and plasma are shown in Table 2. The lower limit of quantification (LLOQ) was determined to be 4.0 pmol/L for melatonin, and 0.1 nmol/L for cortisol in saliva and 10 pmol/L, 0.5 nmol/L in plasma for, respectively, melatonin and cortisol. Recoveries for melatonin at the three levels ranged from 87% to 98% and from 97% to 111% for cortisol (see Table 3). No significant ion suppression was present at the retention times where melatonin and cortisol eluted.

Table 2:

Intra- and inter-assay imprecision results for cortisol and melatonin in plasma and saliva.

PlasmaSaliva
Intra-assayInter-assayIntra-assayInter-assay
MeanCV, %MeanCV, %MeanCV, %MeanCV, %
Cortisol, nmol/L
 Low1292.31272.60.4408.60.4779.6
 Medium3901.03962.72.512.12.585.2
 High8481.38642.811.92.111.44.5
Melatonin, pmol/L
 Low20.58.921.59.520.85.120.38.9
 Medium1243.91224.160.43.859.06.8
 High3183.53004.21142.81096.0
Table 3:

Mean recovery results for cortisol and melatonin in plasma and saliva (ranges between parentheses).

PlasmaSaliva
RecoveryRecovery
Mean (range), %Mean (range), %
Cortisol, nmol/L
 Low92.7 (86–97)107 (98–116)
 Medium98.7 (96–103)100 (95–104)
 High98.8 (96–101)100 (95–105)
Melatonin, pmol/L
 Low96.7 (86–107)94.3 (87–105)
 Medium99.3 (92–107)94.3 (87–111)
 High102 (91–111)91.8 (85–112)
  1. Three different samples (native concentrations in parentheses) were spiked at three different levels with respectively cortisol and melatonin.

Method comparison

In order to verify our method for salivary cortisol, we compared it to the currently used routine LC-MS/MS method for salivary cortisol (Figure 2A and C). The developed method compared well to the routine method with a slope of 1.02 (95% CI: 1.0–1.05) and an intercept of −0.05 (95% CI: −0.09 to 0.01). Bland Altman analysis showed a bias of 0.04 nmol/L between the two methods. For melatonin, we compared the developed method to an ELISA (Figure 2B and D). The developed method showed some discrepancy with the ELISA. In the lower range (<30 pmol/L), there is reasonable agreement, but above >30 pmol/L the ELISA measures considerably lower levels than our LC-MS/MS assay. Passing-Bablok regression gave a slope of 1.23 (95% CI: 1.09–1.51) and intercept of −4.6. (95% CI: −10 to 0.2). Bland Altman analysis showed a bias of 7.9 pmol/L between the two methods.

Figure 2: Scatter and Bland-Altman plots for the method comparison of saliva cortisol with XLCMS/MS (A, C) and saliva melatonin with commercial immunoassay (B, D). Solid line in the scatterplot (A, B) represents the line of identity.Passing-Bablok regression analysis gave the following result for cortisol (A) y=1.02x−0.05 and for melatonin (B) y=1.23x−4.61. Gray area represents the 95% confidence interval. Bland-Altman plots are expressed with the method difference as absolute concentration (C, D). Dashed lines represent 95% limits of agreement.
Figure 2:

Scatter and Bland-Altman plots for the method comparison of saliva cortisol with XLCMS/MS (A, C) and saliva melatonin with commercial immunoassay (B, D). Solid line in the scatterplot (A, B) represents the line of identity.

Passing-Bablok regression analysis gave the following result for cortisol (A) y=1.02x−0.05 and for melatonin (B) y=1.23x−4.61. Gray area represents the 95% confidence interval. Bland-Altman plots are expressed with the method difference as absolute concentration (C, D). Dashed lines represent 95% limits of agreement.

Impact of collection device and correlation between total plasma, free plasma and salivary melatonin and cortisol

For saliva melatonin and cortisol no significant concentration differences were observed between samples collected by passive drooling or Salivette in the 10 samples (see Figure 3A and 4A). Passing-Bablok regression gave the following equations, for melatonin: Salivette=1.0×passive drooling –0.64; r=0.98, for cortisol: Salivette=0.999×passive drooling –0.65; r=0.95. As there was no significant difference between saliva collected by passive drooling or Salivette, we used results of saliva obtained by passive drooling in calculations presented below.

Figure 3: Cortisol: Correlation and Passing-Bablok regression between (A) passive drooling vs. Salivette, (B) total plasma vs. equilibrium dialysis, (C) free plasma vs. saliva.Gray area represents the 95% confidence interval. Dotted line is the line of identity.
Figure 3:

Cortisol: Correlation and Passing-Bablok regression between (A) passive drooling vs. Salivette, (B) total plasma vs. equilibrium dialysis, (C) free plasma vs. saliva.

Gray area represents the 95% confidence interval. Dotted line is the line of identity.

Figure 4: Melatonin: Correlation and Passing-Bablok regression between (A) passive drooling vs. Salivette, (B) total plasma vs. free plasma, (C) free plasma vs. saliva.Gray area represents the 95% confidence interval. Dotted line is the line of identity.
Figure 4:

Melatonin: Correlation and Passing-Bablok regression between (A) passive drooling vs. Salivette, (B) total plasma vs. free plasma, (C) free plasma vs. saliva.

Gray area represents the 95% confidence interval. Dotted line is the line of identity.

Levels of total plasma cortisol correlated moderately with free plasma cortisol levels, r=0.65 (Figure 3B). Free plasma cortisol levels correlated well with salivary cortisol levels, r=0.87, but concentrations in saliva were approximately 50% lower compared to free plasma cortisol (Saliva=0.51×free plasma –2.1; Figure 3C).

Total plasma melatonin correlated excellently with free plasma melatonin, r=0.98; free plasma=0.26×total plasma+1.2, indicating that approximately 75% of melatonin is protein-bound in plasma (Figure 4B). Comparison of free plasma melatonin with salivary melatonin showed that levels of melatonin in saliva were on average 36% higher when compared to levels of free plasma melatonin (95% CI=4–60); passive drooling=1.36×free plasma –1.1 (Figure 4C).

Discussion

In this study, we describe the characteristics, validation and application of a mass spectrometry based method for the simultaneous quantification of melatonin and cortisol in saliva and plasma. We demonstrated its potential by investigating the relationship between total plasma, free plasma and salivary melatonin and cortisol, and found that melatonin in saliva was on average 36% (95% CI: 4–60) higher in saliva than free plasma melatonin.

As melatonin and cortisol are frequently analyzed in circadian rhythm studies, combined analysis of these two neuroendocrine markers is a valuable asset. Quantification of these markers requires sensitive and accurate analytical methods that are routinely applicable. Despite the fact that immunoassays have provided a wealth of information on biochemical aspects of low molecular weight neuroendocrine biomarkers over the past decades, these techniques are rapidly being replaced by mass spectrometry based methods because of their superior accuracy [35]. This is especially true in the field of routinely applied biomarkers such as steroids, vitamins, and biogenic amines. In the field of chronobiology research still mainly relies on immunoassays, despite the fact that the limitations of these assays are becoming clear now. Melatonin is a (protein bound) low molecular weight biomarker that is present at very low concentrations amidst numerous potentially interfering structure analogs such as serotonin, N-acetylserotonin and kynurenamines. The same holds true for cortisol. This makes quantification of these markers by immunoassays vulnerable to variation in binding protein concentration, matrix and structure analog interferences [34], [36], [37].

Recently, two LC-MS/MS methods have been described in the literature for the combined analysis of cortisol and melatonin in saliva, but none for plasma [10], [11]. The two described methods for saliva have certain drawbacks in comparison to our method, such as manual sample preparation (liquid-liquid extraction) [10] and a significantly higher LLOQ (0.5 nmol/L) for cortisol, which is insufficient to quantify late night saliva cortisol levels [11]. Moreover, in comparison to the other two published methods for cortisol and melatonin in saliva, our method is also suitable for plasma analysis. Our method enables efficient high throughput analyses of melatonin and cortisol, and was recently used in a large longitudinal study in which depressed and non-depressed individuals collected saliva over 30 days, thrice a day [38], [39].

In several studies in which immunoassays were used, it was found that the device for saliva collection can influence the results for melatonin and cortisol [40], [41], [42]. We therefore investigated to what extend sample collection methods influenced results of our LC-MS/MS assay. We found that results for melatonin and cortisol in saliva collected by either passive drooling or Salivette were similar. This finding is in line with previous results obtained by LC-MS [11]. A possible explanation for this observation could be that immunoassays generally are more sensitive to interfering substances as could be coming from the Salivette, whereas LC-MS/MS generally is less matrix-dependent [43]. We compared salivary melatonin results from our LC-MS/MS assay to a commercial immunoassay designed for use in saliva. The comparison showed reasonable agreement, with more scatter in the higher calibrator range. A difference in calibration between the melatonin immunoassay and our LC-MS/MS assay was excluded as explanation for the discrepancy, as calibrators were cross-checked (data not shown). The difference in results between IA and MS for salivary melatonin is in agreement with a previous publication from Jensen et al. which also reports a similar discrepancy between LC-MS/MS and an immunoassay [44].

In line with previous studies, we found that cortisol in saliva correlated well with plasma free cortisol, but not with total cortisol in plasma [45], [46]. Levels of cortisol in saliva were approximately half of that of plasma free cortisol. One contributing factor to this difference is probably the conversion of cortisol to cortisone by 11β-hydroxysteroid dehydrogenase in the salivary glands [47]. Levels of plasma free melatonin were 26% (95% CI: 22 –30%) of total melatonin in plasma, which corresponds well with values reported by Kennaway and Voultsios. (mean 23%, range 15–39%) [25]. However, salivary melatonin concentrations were on average 36% (95% CI: 4–60) higher than the corresponding plasma free melatonin concentrations, which could indicate local production of melatonin in the salivary glands. Recently, Shimozuma et al. reported the presence of the melatonin-synthesizing enzymes arylalkylamine N-acetyltransferase (AANAT) and hydroxyindole-O-methyltransferase (HIOMT) in epithelial cells of human submandibular glands [48]. Local synthesis of melatonin in salivary glands could explain the large variation found in the relation of salivary melatonin and plasma concentrations [49], [50]. This is to our knowledge the first study that investigated the relation between saliva and plasma dialysate by LC-MS/MS and shows a higher concentration of salivary melatonin vs. free plasma melatonin obtained by equilibrium dialysis. A limitation of this observation is that we only measured samples from 10 different individuals. To confirm local synthesis of melatonin in the salivary glands, more samples from different individuals should be included.

In summary, we showed that our mass spectrometry based method enables straightforward, reproducible and sensitive quantification of cortisol and melatonin in human saliva and plasma and illustrates the added value of reliable mass spectrometry based methods for the quantification of circadian rhythm biomarkers.


Corresponding author: Martijn van Faassen, BASc, Department of Laboratory Medicine, University of Groningen, University Medical Center Groningen, EA61, P.O. Box 30.001, 9700 RB Groningen, The Netherlands, Phone: +31 50 361 03 99, Fax: +31 50 361 91 91

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission. M.v.F. carried out all experimental work and performed the data analysis. I.P.K. made the concept and design of the study, interpretation of data, and drafted together with M.v.F. the manuscript. I.P.K. together with R.B revised and approved the final version of the manuscript.

  2. Research funding: None declared.

  3. Employment or leadership: None declared.

  4. Honorarium: None declared.

  5. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.

References

1. Rajaratnam SM, Arendt J. Health in a 24-h society. Lancet 2001;358:999–1005.10.1016/S0140-6736(01)06108-6Suche in Google Scholar

2. Gamble KL, Berry R, Frank SJ, Young ME. Circadian clock control of endocrine factors. Nat Rev Endocrinol 2014;10:466–75.10.1038/nrendo.2014.78Suche in Google Scholar

3. Arendt J. Importance and relevance of melatonin to human biological rhythms. J Neuroendocrinol 2003;15:427–31.10.1046/j.1365-2826.2003.00987.xSuche in Google Scholar

4. Hardeland R, Cardinali DP, Srinivasan V, Spence DW, Brown GM, Pandi-Perumal SR. Melatonin–a pleiotropic, orchestrating regulator molecule. Prog Neurobiol 2011;93:350–84.10.1016/j.pneurobio.2010.12.004Suche in Google Scholar

5. Chrousos GP. Stress and disorders of the stress system. Nat Rev Endocrinol 2009;5:374–81.10.1038/nrendo.2009.106Suche in Google Scholar

6. Pruessner JC, Wolf OT, Hellhammer DH, Buske-Kirschbaum A, von Auer K, Jobst S, et al. Free cortisol levels after awakening: a reliable biological marker for the assessment of adrenocortical activity. Life Sci 1997;61:2539–49.10.1016/S0024-3205(97)01008-4Suche in Google Scholar

7. de Almeida EA, Di Mascio P, Harumi T, Spence DW, Moscovitch A, Hardeland R, et al. Measurement of melatonin in body fluids: standards, protocols and procedures. Childs Nerv Syst 2011;27:879–91.10.1007/s00381-010-1278-8Suche in Google Scholar PubMed PubMed Central

8. Turpeinen U, Hämäläinen E. Determination of cortisol in serum, saliva and urine. Best Pract Res Clin Endocrinol Metab 2013;27:795–801.10.1016/j.beem.2013.10.008Suche in Google Scholar PubMed

9. Wang A-Q, Wei B-P, Zhang Y, Wang Y-J, Xu L, Lan K. An ultra-high sensitive bioanalytical method for plasma melatonin by liquid chromatography-tandem mass spectrometry using water as calibration matrix. J Chromatogr B Analyt Technol Biomed Life Sci 2011;879:2259–64.10.1016/j.jchromb.2011.06.010Suche in Google Scholar PubMed

10. Jensen MA, Hansen AM, Abrahamsson P, Nørgaard AW. Development and evaluation of a liquid chromatography tandem mass spectrometry method for simultaneous determination of salivary melatonin, cortisol and testosterone. J Chromatogr B Analyt Technol Biomed Life Sci 2011;879:2527–32.10.1016/j.jchromb.2011.07.005Suche in Google Scholar PubMed

11. Fustinoni S, Polledri E, Mercadante R. High-throughput determination of cortisol, cortisone, and melatonin in oral fluid by on-line turbulent flow liquid chromatography interfaced with liquid chromatography/tandem mass spectrometry. Rapid Commun Mass Spectrom 2013;27:1450–60.10.1002/rcm.6601Suche in Google Scholar

12. Carter MD, Calcutt MW, Malow BA, Rose KL, Hachey DL. Quantitation of melatonin and n-acetylserotonin in human plasma by nanoflow LC-MS/MS and electrospray LC-MS/MS. J Mass Spectrom 2012;47:277–85.10.1002/jms.2051Suche in Google Scholar

13. Taylor AE, Keevil B, Huhtaniemi IT. Mass spectrometry and immunoassay: how to measure steroid hormones today and tomorrow. Eur J Endocrinol 2015;173:D1-12.10.1530/EJE-15-0338Suche in Google Scholar

14. Lemaitre BJ, Hartmann L. Preparation of anti-melatonin antibodies and antigenic properties of the molecule. J Immunol Methods 1980;32:339–47.10.1016/0022-1759(80)90026-5Suche in Google Scholar

15. Miller R, Plessow F, Rauh M, Gröschl M, Kirschbaum C. Comparison of salivary cortisol as measured by different immunoassays and tandem mass spectrometry. Psychoneuroendocrinology 2013;38:50–7.10.1016/j.psyneuen.2012.04.019Suche in Google Scholar PubMed

16. Ju Bae Y, Gaudl A, Jaeger S, Stadelmann S, Hiemisch A, Kiess W, et al. Immunoassay or LC-MS/MS for the measurement of salivary cortisol in children? Clin Chem Lab Med 2016;54:811–22.10.1515/cclm-2015-0412Suche in Google Scholar PubMed

17. Hofstra WA, de Weerd AW. How to assess circadian rhythm in humans: a review of literature. Epilepsy Behav 2008;13:438–44.10.1016/j.yebeh.2008.06.002Suche in Google Scholar PubMed

18. Waller KL, Mortensen EL, Avlund K, Fagerlund B, Lauritzen M, Gammeltoft S, et al. Melatonin and cortisol profiles in late midlife and their association with age-related changes in cognition. Nat Sci Sleep 2016;8:47–53.10.2147/NSS.S75946Suche in Google Scholar PubMed PubMed Central

19. Gunn PJ, Middleton B, Davies SK, Revell VL, Skene DJ. Sex differences in the circadian profiles of melatonin and cortisol in plasma and urine matrices under constant routine conditions. Chronobiol Int 2016;33:39–50.10.3109/07420528.2015.1112396Suche in Google Scholar PubMed PubMed Central

20. Fatima G, Sharma VP, Verma NS. Circadian variations in melatonin and cortisol in patients with cervical spinal cord injury. Spinal Cord 2016;54:364–7.10.1038/sc.2015.176Suche in Google Scholar PubMed

21. Gröschl M. Current status of salivary hormone analysis. Clin Chem 2008;54:1759–69.10.1373/clinchem.2008.108910Suche in Google Scholar

22. Umeda T, Hiramatsu R, Iwaoka T, Shimada T, Miura F, Sato T. Use of saliva for monitoring unbound free cortisol levels in serum. Clin Chim Acta 1981;110:245–53.10.1016/0009-8981(81)90353-3Suche in Google Scholar

23. Raff H, Raff JL, Findling JW. Late-night salivary cortisol as a screening test for Cushing’s syndrome. J Clin Endocrinol Metab 1998;83:2681–6.10.1210/jc.83.8.2681Suche in Google Scholar

24. Antonelli G, Ceccato F, Artusi C, Marinova M, Plebani M. Salivary cortisol and cortisone by LC-MS/MS: validation, reference intervals and diagnostic accuracy in Cushing’s syndrome. Clin Chim Acta 2015;451:247–51.10.1016/j.cca.2015.10.004Suche in Google Scholar PubMed

25. Kennaway DJ, Voultsios a. Circadian rhythm of free melatonin in human plasma. J Clin Endocrinol Metab 1998;83:1013–5.10.1210/jc.83.3.1013Suche in Google Scholar

26. Fiers T, Delanghe J, T’Sjoen G, Van Caenegem E, Wierckx K, Kaufman J-M. A critical evaluation of salivary testosterone as a method for the assessment of serum testosterone. Steroids 2014;86:5–9.10.1016/j.steroids.2014.04.013Suche in Google Scholar PubMed

27. de Jong WHA, Graham KS, van der Molen JC, Links TP, Morris MR, Ross HA, et al. Plasma free metanephrine measurement using automated online solid-phase extraction HPLC tandem mass spectrometry. Clin Chem 2007;53:1684–93.10.1373/clinchem.2007.087114Suche in Google Scholar PubMed

28. Agency EM. Guideline on bioanalytical method validation [Internet] 2011. Available at: http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2011/08/WC500109686.pdf. Accessed 2005 Apr 20.Suche in Google Scholar

29. Annesley TM. Ion suppression in mass spectrometry. Clin Chem 2003;49:1041–4.10.1373/49.7.1041Suche in Google Scholar PubMed

30. Jones RL, Owen LJ, Adaway JE, Keevil BG. Simultaneous analysis of cortisol and cortisone in saliva using XLC-MS/MS for fully automated online solid phase extraction. J Chromatogr B Analyt Technol Biomed Life Sci 2012;881–882:42–8.10.1016/j.jchromb.2011.11.036Suche in Google Scholar PubMed

31. Bablok W, Passing H, Bender R, Schneider B. A general regression procedure for method transformation. Application of linear regression procedures for method comparison studies in clinical chemistry, Part III. J Clin Chem Clin Biochem 1988;26:783–90.10.1515/cclm.1988.26.11.783Suche in Google Scholar PubMed

32. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307–10.10.1016/j.ijnurstu.2009.10.001Suche in Google Scholar

33. R Core Team. R: A language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing, 2013. Available at: http://www.R-project.org/.Suche in Google Scholar

34. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105–20.Suche in Google Scholar

35. Jannetto PJ, Fitzgerald RL. Effective use of mass spectrometry in the clinical laboratory. Clin Chem 2016;62:92–8.10.1373/clinchem.2015.248146Suche in Google Scholar

36. Heijboer AC, Blankenstein MA, Kema IP, Buijs MM. Accuracy of 6 routine 25-hydroxyvitamin D assays: influence of vitamin D binding protein concentration. Clin Chem 2012;58:543–8.10.1373/clinchem.2011.176545Suche in Google Scholar

37. Hawley JM, Owen LJ, Lockhart SJ, Monaghan PJ, Armston A, Chadwick CA, et al. Serum cortisol: an up-to-date assessment of routine assay performance. Clin Chem 2016;62:1220–9.10.1373/clinchem.2016.255034Suche in Google Scholar

38. Booij SH, Bos EH, Bouwmans MEJ, van Faassen M, Kema IP, Oldehinkel AJ, et al. Cortisol and alpha-Amylase Secretion Patterns between and within Depressed and Non-Depressed Individuals. PLoS One 2015;10:e0131002.10.1371/journal.pone.0131002Suche in Google Scholar

39. Bouwmans MEJ, Bos EH, Booij SH, van Faassen M, Oldehinkel AJ, de Jonge P. Intra- and inter-individual variability of longitudinal daytime melatonin secretion patterns in depressed and non-depressed individuals. Chronobiol Int 2015;32:441–6.10.3109/07420528.2014.973114Suche in Google Scholar

40. Gröschl M, Köhler H, Topf H-G, Rupprecht T, Rauh M. Evaluation of saliva collection devices for the analysis of steroids, peptides and therapeutic drugs. J Pharm Biomed Anal 2008;47:478–86.10.1016/j.jpba.2008.01.033Suche in Google Scholar

41. Kozaki T, Hashiguchi N, Kaji Y, Yasukouchi A, Tochihara Y. Effects of saliva collection using cotton swab on cortisol enzyme immunoassay. Eur J Appl Physiol 2009;107:743–6.10.1007/s00421-009-1178-3Suche in Google Scholar

42. Kozaki T, Lee S, Nishimura T, Katsuura T, Yasukouchi A. Effects of saliva collection using cotton swabs on melatonin enzyme immunoassay. J Circadian Rhythms 2011;9:1.10.1186/1740-3391-9-1Suche in Google Scholar

43. Shirtcliff EA, Granger DA, Schwartz E, Curran MJ. Use of salivary biomarkers in biobehavioral research: cotton-based sample collection methods can interfere with salivary immunoassay results. Psychoneuroendocrinology 2001;26:165–73.10.1016/S0306-4530(00)00042-1Suche in Google Scholar

44. Jensen MA, Mortier L, Koh E, Keevil B, Hyttinen S, Hansen ÅM. An interlaboratory comparison between similar methods for determination of melatonin, cortisol and testosterone in saliva. Scand J Clin Lab Invest 2014;74:454–61.10.3109/00365513.2014.900693Suche in Google Scholar PubMed

45. Perogamvros I, Keevil BG, Ray DW, Trainer PJ. Salivary cortisone is a potential biomarker for serum free cortisol. J Clin Endocrinol Metab 2010;95:4951–8.10.1210/jc.2010-1215Suche in Google Scholar PubMed

46. Vining RF, McGinley RA, Maksvytis JJ, Ho KY. Salivary cortisol: a better measure of adrenal cortical function than serum cortisol. Ann Clin Biochem 1983;20(Pt 6):329–35.10.1177/000456328302000601Suche in Google Scholar PubMed

47. Vining RF, McGinley RA, Symons RG. Hormones in saliva: mode of entry and consequent implications for clinical interpretation. Clin Chem 1983;29:1752–6.10.1093/clinchem/29.10.1752Suche in Google Scholar

48. Shimozuma M, Tokuyama R, Tatehara S, Umeki H, Ide S, Mishima K, et al. Expression and cellular localizaion of melatonin-synthesizing enzymes in rat and human salivary glands. Histochem Cell Biol 2011;135:389–96.10.1007/s00418-011-0800-8Suche in Google Scholar PubMed

49. McIntyre IM, Norman TR, Burrows GD, Armstrong SM. Melatonin rhythm in human plasma and saliva. J Pineal Res 1987;4:177–83.10.1111/j.1600-079X.1987.tb00854.xSuche in Google Scholar PubMed

50. Voultsios A, Kennaway DJ, Dawson D. Salivary melatonin as a circadian phase marker: validation and comparison to plasma melatonin. J Biol Rhythms 1997;12:457–66.10.1177/074873049701200507Suche in Google Scholar PubMed


Supplemental Material:

The online version of this article offers supplementary material (DOI: https://doi.org/10.1515/cclm-2016-0817).


Received: 2016-9-12
Accepted: 2016-10-31
Published Online: 2016-12-20
Published in Print: 2017-8-28

©2017 Walter de Gruyter GmbH, Berlin/Boston

Artikel in diesem Heft

  1. Frontmatter
  2. Editorial
  3. Mass spectrometry or immunoassay: est modus in rebus
  4. Reviews
  5. The use of liquid chromatography-tandem mass spectrometry for therapeutic drug monitoring of antibiotics in cancer patients
  6. Tackling serum folate test in European countries within the health technology assessment paradigm: request appropriateness, assays and health outcomes
  7. Genetics and Molecular Diagnostics
  8. Genetic diagnosis of α1-antitrypsin deficiency using DNA from buccal swab and serum samples
  9. General Clinical Chemistry and Laboratory Medicine
  10. Serum triglyceride measurements: the commutability of reference materials and the accuracy of results
  11. Variant peptide detection utilizing mass spectrometry: laying the foundations for proteogenomic identification and validation
  12. Evaluation of two fully automated immunoassay based tests for the measurement of 1α,25-dihydroxyvitamin D in human serum and comparison with LC-MS/MS
  13. Parallel diurnal fluctuation of testosterone, androstenedione, dehydroepiandrosterone and 17OHprogesterone as assessed in serum and saliva: validation of a novel liquid chromatography-tandem mass spectrometry method for salivary steroid profiling
  14. Determination of oxycodone and its major metabolites noroxycodone and oxymorphone by ultra-high-performance liquid chromatography tandem mass spectrometry in plasma and urine: application to real cases
  15. Identification and quantitation of phosphatidylethanols in oral fluid by liquid chromatography-tandem mass spectrometry
  16. Relationship between plasma and salivary melatonin and cortisol investigated by LC-MS/MS
  17. Paramagnetic micro-particles as a tool for rapid quantification of apixaban, dabigatran, edoxaban and rivaroxaban in human plasma by UHPLC-MS/MS
  18. Measurements of serum non-ceruloplasmin copper by a direct fluorescent method specific to Cu(II)
  19. The serum concentrations of leptin and MCP-1 independently predict low back pain duration
  20. Immunoassay screening in urine for synthetic cannabinoids – an evaluation of the diagnostic efficiency
  21. Cancer Diagnostics
  22. Study of kallikrein-related peptidase 6 (KLK6) and its complex with α1-antitrypsin in biological fluids
  23. Cardiovascular Diseases
  24. A candidate liquid chromatography mass spectrometry reference method for the quantification of the cardiac marker 1-32 B-type natriuretic peptide
  25. The natriuretic peptide MR-proANP predicts all-cause mortality and adverse outcome in community patients: a 10-year follow-up study
  26. CASZ1 loss-of-function mutation contributes to familial dilated cardiomyopathy
  27. Diabetes
  28. Evaluating new HbA1c methods for adoption by the IFCC and NGSP reference networks using international quality targets
  29. Infectious Diseases
  30. Analytical and diagnostic performance of two automated fecal calprotectin immunoassays for detection of inflammatory bowel disease
  31. Letters to the Editor
  32. Is fasting necessary for lipid profile determinations? Some considerations from the perspective of the clinical laboratory
  33. Precision of nonfasting lipid profiles should focus on clinical relevance rather than necessarily obtaining the least variation
  34. Triglyceride concentrations should be measured after elimination of free glycerol to exclude interindividual variations due to adiposity and fasting status
  35. Estimation of the reference interval for serum folate measured with assays traceable to the WHO International Standard
  36. Implausible elevation of peripheral thyroid hormones during therapy with a protein supplement
  37. Interference in Na+ measurements on the Siemens RAPIDPoint® 500 after nortriptyline intoxication: a case report
  38. Usefulness of maternal red cell antibodies to predict hemolytic disease of the fetus and newborn and significant neonatal hyperbilirubinemia: a retrospective study