Laboratory medicine as the science that underpins medicine: the “high-sensitivity” troponin paradigm

  • Simona Ferraro

    Simona Ferraro, BSc, is specialized in clinical pathology and in medical statististics. She works at the Clinical Pathology Unit of the “Luigi Sacco” University Hospital in Milan in the field of tumor, hematological, and cardiac biomarkers. She has contributed 19 articles published in peer-reviewed journals, 15 of which as first author.

    EMAIL logo
    and Mauro Panteghini

    Prof. Mauro Panteghini is Full Professor of Clinical Biochemistry and Clinical Molecular Biology and Director of the corresponding Chair at the Medical School of the University of Milan, Italy. He also covers the following institutional positions: Director of the Department of Laboratory Medicine and Director of Clinical Pathology Unit of the “Luigi Sacco” University Hospital in Milan; Director of the Centre for Metrological Traceability in Laboratory Medicine (CIRME) of the University of Milan. He has served in a number of international and national scientific activities in the field of Laboratory Medicine. He is currently President of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM). He has published 450 manuscripts (h-index: 43) and more than 400 abstracts.

Published/Copyright: October 2, 2014

Abstract

The availability of so-called high-sensitivity troponin assays (hsTn) has scored a compelling goal for laboratory medicine, allowing the safe clinical application of international recommendations for the definition of acute myocardial infarction (AMI). However, the introduction of hsTn has not been welcomed by clinicians, claiming an increase in false-positive results. Here we critically trace back the steps following the introduction of hsTn by referring to the 5-year practical experience in our academic hospital and to suitable information available in the literature. In agreement with published data, we found that hsTn introduction was associated with an increased number of AMI diagnoses, whereas the test volume, the revascularization rate, and the proportion of cases with negative angiography findings remained virtually unchanged. Fast-track protocols for ruling out AMI have been further optimized to recommend sampling at presentation and after 3 h only. We focus on a cost-effective use of hsTn that can account for all clinical variables increasing the pre-test probability in order to ensure that tests are ordered only for patients at medium to high risk for acute coronary syndrome (ACS). To guide interpretation of results, hsTn typical release patterns suggestive for AMI should be identified by evaluating the significance of concentration changes. hsTn have markedly shortened the time to rule out or rule in AMI and has the potential to improve the prognostic assessment of critical patients in clinical contexts different from ACS.

Introduction

The implementation of so-called high-sensitivity cardiac troponin (cTn) assays (hsTn) has benefited the laboratory through enhanced analytical sensitivity and precision and now its ability to fulfill international recommendations for the “universal” definition of acute myocardial infarction (AMI) that have been released over the past 14 years [1–3]. However, with the implementation of hsTn, emergency department (ED) physicians and cardiologists have been rather confused on the interpretation of marker results [4, 5]. The capability of hsTn to measure cTn concentrations >10-fold lower than those detectable by previous-generation assays, and with highest precision, has unexpectedly complicated patient triage [6, 7]. The trade-off of greater clinical sensitivity for lower clinical specificity for AMI diagnosis has sparked hot debates and discussions involving both clinicians and laboratorians [8–11]. With hsTn introduction, cardiologists claim there is an increase in false-positive results and consider hsTn to be a poorer diagnostic test than previous-generation, less-sensitive assays because hsTn does not fit with the clinical workup [6]. In reply, laboratorians have reminded cardiologists that cTn-positive results only indicate the presence of myocardial injury from various pathophysiological mechanisms and that hsTn enables more accurate detection of “microscopic zones” of myocardial injury [5, 7]. Some clinicians had initially misunderstood the interpretation and the additional diagnostic value contributed by hsTn, underestimating the importance of clinical findings and serial cTn testing [5]. Indeed, the eroded diagnostic specificity claimed by clinicians suggested that test ordering and result management needed to be better “guided” by laboratorians by utilizing the improved analytical performance of hsTn [6, 11].

It is now clear that the hsTn introduction requires changes to diagnostic rules and algorithms. In this review, we summarize the available evidence showing how broadly the replacement of previous-generation cTn assays with hsTn has changed patient clinical outcome and costs of care. Along with the retrieved evidence, we will trace the steps of hsTn introduction in our academic hospital by reporting the evolution of internal recommendations for ordering and interpreting test results as well as data retrieved by internal audits.

The prequel: before hsTn

If we trace the history of cTn since its clinical introduction, we may realize that it has gained increased relevance from its inclusion in Cardiology Society Consensus Documents for defining AMI. The further adoption of more sensitive assays has required a more detailed definition of the interpretative criteria [1–3]. In the first Consensus Document, released in 2000, cTn was indicated as the preferred test to detect myocardial necrosis, due to absolute tissue specificity and high sensitivity, and an increased value was defined “as a measurement exceeding the 99th percentile of a reference control group” [1]. AMI was tailored on cTn increase secondary to coronary ischemia. The updated 2007 redefinition of AMI confirmed the 99th percentile cutoff “as decision level for the diagnosis of AMI”, cautioning on the need to estimate the 99th percentile reference limit for each specific assay “with appropriate quality control” [2]. It is now known that the selection of “cardio-healthy” reference individuals requires partitioning according to age and gender and that sample size may markedly affect the estimate of the 99th percentile cutoff [12]. By fixing the decision limit to the 99th percentile value, one theoretically maximizes the capability of cTn to identify subjects with causes of cTn release other than physiological cardiomyocyte turnover, not to diagnose patients with AMI. More importantly, the detection limit (LoD) and the analytical performance of the assay may strongly influence the definition of the 99th percentile cutoff since this implies that cTn concentrations in the majority of subjects (if not all) should be detectable [12]. Apple and Murakami showed, however, that commercial assays were unable to detect cTn in up to 98% of apparently healthy subjects [13]. Therefore, under these conditions, a frequency distribution of cTn values in a reference population cannot be defined, and the derivation of the 99th percentile is flawed. In addition, the lack of analytical sensitivity is associated with a high measurement error at the very low cTn concentrations corresponding to the 99th percentile cutoff and implies that there is a high risk of misclassifying positive and negative results for myocardial injury [14]. The concept of limit of quantitation (LoQ), i.e., the lowest amount of cTn that can be quantitatively determined with a stated acceptable (read “clinically meaningful”) inaccuracy, is central here, as the 99th percentile limit, to be clinically usable, cannot be lower than the LoQ of the cTn assay [15]. Using conventional cTn assays, in 2004, an IFCC cooperative study showed that none of the commercially available cTn assays displayed acceptable analytical performance at low concentration values to correctly define the 99th percentile limit [16]. This implied that cTn assays used in the middle of the first decade of the new century did not fulfill the desired analytical requirements, remaining inadequate for running in clinical practice the AMI universal definition [17].

The manufacturers’ reaction: introduction of hsTn

To adjust the misfit between guidelines and the analytical performance of available cTn assays, manufacturers aimed to improve the analytical sensitivity and imprecision of their assays [16]. Accordingly, several approaches were proposed to achieve the accurate measurement of cTn concentrations, even down to the nanogram per liter expected to characterize healthy people [18]. For instance, in the Roche Diagnostics hsTn determining cardiac troponin T (hsTnT), the improvement in the analytical signal has been achieved by increasing the sample volume and decreasing background noise through optimization of buffer composition. A first prototype assay was already available in 2003, when a first multicenter pilot study was performed. However, this first version of the hsTnT assay was never marketed, even if a clinical study using it was later published [19]. Four years later, a development version of hsTnT was proposed and externally evaluated [20]. Finally, in 2008, a final hsTnT version was evaluated in a multicenter trial and released on the market 1 year later, replacing the conventional assay [21]. In our hospital, we introduced in clinical practice the hsTnT on July 2009 and performed the first internal audit 3 months later (see below). The described steps are summarized in Figure 1.

Figure 1 Steps of Roche Diagnostic “high-sensitivity” cardiac troponin T assay development and its introduction in the authors’ laboratory.
Figure 1

Steps of Roche Diagnostic “high-sensitivity” cardiac troponin T assay development and its introduction in the authors’ laboratory.

The published trial derived the 99th percentile reference limit in a consistent population of healthy blood donors by setting it at a cTn T concentration of 14.2 ng/L [21]. In particular, it was observed that 32% of individuals had cTn concentrations above the assay LoD, i.e., >5.0 ng/L, as defined according to the Clinical and Laboratory Standards Institute EP17 standard [22]. At the 99th percentile concentration, comparable results across multiple platforms and laboratories were obtained, and the estimated inter-laboratory CV result was approximately 6.5% [21]. Afterward, our internal quality control data on pooled human serum with a cTn T concentration close to the 99th percentile cutoff confirmed these results, reporting an average imprecision far below the proposed minimum goal for CV of 10% [15]. External quality assessment surveys on samples with average cTn T concentrations around the 99th percentile limit confirmed that ≥99% of participating laboratories using hsTnT met the desired quality level, with a measurement total error <22.5% [23, 24].

The (quiet?) chaos[1]: increasing cTn positivity rates

The increase in positive cTn results may be considered typical of new assay generations increasing the analytical sensitivity of measurement, even when hsTn were still unavailable. For instance, Melanson et al. [25] reported that the changing generation of Siemens cTn I assay resulted in an increase in positive test results by 44.2% hospital-wide and by 114.4% in the ED, without a significant change in test volume.

More recently, others have found that 20% of consecutive patients admitted to the chest pain unit had an hsTnT elevation, and in 69% of cases, this was not due to acute coronary syndrome (ACS) [26]. Our audit reported data on the hospital-wide impact of replacing the fourth-generation cTn T assay with hsTnT showed that the absolute number of positive cTn results increased by 95%, corresponding to a relative increase by 111% of positive patients, with no significant change in test volume [27, 28]. By selecting only ED requests, the increase in positive cTn results amounted to 144%, corresponding to a relative increase of 133% in positive patients, with the rate of test ordering still significantly unchanged [27, 28].

From its initial advent, cTn, although providing higher sensitivity for small injury and virtually absolute specificity for myocardial damage when compared with creatine kinase MB (CK-MB) isoenzyme, left some aspects open to discussion [29]. At that time, some authors already underlined that the cardiac specificity and sensitivity of cTn measurement could be a two-edged sword when this marker is used in clinical practice [30]. If we focus our attention on the comparison of rates of positive cTn results between assay generations, it is apparent that lowering the cutoff at the 99th percentile limit and increasing the assay sensitivity has resulted in (a) a shift from diagnoses of unstable angina (UA) to AMI and (b) more cases of cTn increase caused by non-ischemic cardiac necrosis. The World Health Organization, during 2008–2009, reported there was a relevant shift from diagnoses of UA to AMI [31]. This demonstrated that the definition of UA is strongly influenced by the assay used to measure cTn and will be affected by the implementation of hsTn, namely by the increased capability to detect very minor myocardial necrosis [32].

Fixing the cTn cutoff at the 99th percentile: a trade-off between sensitivity and specificity?

Initially, cardiologists recommend using the 99th percentile cutoff when assays were unable to measure cTn accurately. However, currently, when the new generation of hsTn can finally fulfill the AMI recommendations reliably, cardiologists generally have not welcomed their introduction. In this regard, several authors have remarked the need to differentiate between the specificity of cTn for cardiomyocyte injury and its clinical specificity for AMI [33]. In particular, the improved sensitivity of cTn assays has reinforced the evidence that the 99th percentile limit, if applied to only one result, is not functional to the diagnosis of AMI. Only serial testing allows for the discrimination of acute from chronic pathophysiological mechanisms of cTn release. Pathophysiological release of cTn is characterized by an increasing (or decreasing) cTn pattern, indicating an acute disease process, whereas an unchanged stable cTn course indicates a chronic clinical condition (Figure 2) [34]. Indeed, the 2007 universal definition of AMI stated that “the detection of an increase and/or decrease in cTn with at least one value above the 99th percentile limit” is the biochemical criterion to diagnose AMI, together with the clinical or instrumental evidence of coronary ischemia [2].

Figure 2 Temporal patterns describing different types of cardiac troponin variation between consecutive samples in one patient.
Figure 2

Temporal patterns describing different types of cardiac troponin variation between consecutive samples in one patient.

Importantly, if we dichotomize typical/atypical cTn curves (built on two to three serial measurements), in addition to at least one cTn value above the 99th percentile limit, an additional decisional criterion, related to the entity of increasing/decreasing of cTn trend, should be defined to characterize cTn patterns typical for acute myocardial necrosis. Currently, there is no general agreement on the definition of changes (δ) suggestive for acute conditions [35]. In our daily clinical practice, we defined as “typical” an increasing or decreasing pattern showing a cTn T change between two consecutive samples exceeding 50% for increasing cTn T results and 30% for decreasing results [27, 28]. Otherwise, the cTn pattern was considered “atypical”. For the definition of these δ percentages (estimated as reference change value), we refer to the short-term biological variation for cTn I [36] because, unfortunately, published studies that tried to assess biological variability of cTn T were implausible, as the majority [37] or a significant number [38] of cTn T results in selected individuals were lower than the LoD. One should be aware, however, that cTn I and cTn T may have different biological kinetics in blood, so their biological variation might be different.

To compare the clinical impact of hsTnT, we retrieved cTn curves (at least two results during patient examination) of comparable 3-month periods before and after hsTnT implementation and selected conventional cTn T assay and hsTnT curves with typical marker release (δ>50%) [28]. Although an increase in positive individual cTn results and in positive curves was detected using hsTnT, the proportion of positive curves with δ>50% did not change (Figure 3). This was an important proof confirming that, if the scrutiny of cTn release kinetics in blood is performed, the use of hsTn does not imply lower specificity for diagnosing AMI. Meanwhile, any effort to optimize the evaluation of a single cTn value by a decisional level is diminished by the biological behavior of the marker. According to the estimation of cTn biological variation, the far lower within-person variation with respect to the inter-individual variation implies a very low index of individuality (i.e., ratio of within- to between-subject variances) [36]. Under this condition, an isolated dichotomized interpretation of cTn results, resorting to the 99th percentile limit, can be misleading, and this agrees with clinical evidence and cardiologists’ perception. As described above, the longitudinal monitoring of serial cTn changes is undoubtedly more effective in classifying patients with acute or chronic structural heart disease. Accordingly, a number of algorithms incorporating both baseline hsTn concentrations and changes in values over serial measurements have been proposed to address marker specificity concerns and to allow more accurate ruling out and ruling in of AMI [39].

Figure 3 Absolute number and percentage of (A) positive individual troponin results, (B) positive troponin curves (i.e., at least one result positive during patient examination), and (C) positive troponin curves with typical marker release (higher than the reference change value). Adapted from ref. 28.TnT, fourth-generation troponin T assay; hsTnT, high-sensitivity troponin T assay.
Figure 3

Absolute number and percentage of (A) positive individual troponin results, (B) positive troponin curves (i.e., at least one result positive during patient examination), and (C) positive troponin curves with typical marker release (higher than the reference change value). Adapted from ref. 28.

TnT, fourth-generation troponin T assay; hsTnT, high-sensitivity troponin T assay.

The major problem in the practical application of the δ approach is that most cTn tests are ordered just once without any follow-up on biomarker concentrations. In our hospital, approximately 60% of hsTnT results refer to the one test ordered at ED presentation only [28]. This clinician behavior has been confirmed by other audits in similar settings [40] and strongly suggests that laboratories should provide tools to clinicians, e.g., hsTn results with an automatic calculation of δ provided a second sample is collected within a time frame of no more than 12 h [41].

Regarding the triage of ED patients, our experience showed that the hsTnT implementation did not significantly change the rate of admissions in intensive and non-intensive care facilities despite the 2.5-fold increase in positive cTn tests and the 2-fold increase in patient hospitalization [42]. Importantly, although the rate of discharged patients with at least one hsTn-positive result increased significantly compared with the conventional cTn T assay (26.6% vs. 8.5%), the number of patient readmissions after 2 months of follow-up did not significantly increase [42]. The lack of short-term implications suggests that patients with mildly increased hsTnT concentrations and no signs of ACS do not necessarily need hospitalization, confirming that triage decisions in the ED should not only be based on hsTn results but should be interpreted together with other available clinical information [43].

The clinical context to avoid “troponin consult”

As for all the tests used in laboratory medicine, the capability of hsTn results to correctly identify or exclude an AMI [i.e., their positive (PPV) and negative (NPV) predictive values] strongly depends on the prevalence (pre-test probability) of the disease in the clinical setting in which the test is applied. To this regard, the ACS prevalence in ED amounts to a maximum of 10%–12% of admitted subjects with suspected chest pain symptoms, and it is different from the 40%–50% prevalence expected by cardiologists in their selected setting. This is relevant, for example, for a test with 90% sensitivity and specificity at a fixed cutoff point where the PPV may change from 85.7% to 50% and the NPV from 93.1% to 98.8% just by shifting the disease prevalence from 40% to 10%. Thus, by applying hsTn in a setting with low prevalence of acute ischemic damage, a positive marker result is less likely associated with the presence of AMI. The concept was originally raised in 2002 in a commentary published by a joint cardiologist-laboratory group stating that “with the purpose of limiting the number of cases with a finding of myocardial damage in the absence of an AMI, it is advisable that cTn measurement be limited to those patients with a medium to high probability of acute myocardial ischemia…indiscriminate cTn measurement in patients who are evaluated in an emergency or critical clinical setting is to be avoided” [44]. Particularly, with the introduction of hsTn, ED physicians should develop and use well-validated protocols exploiting the whole clinical information in order to increase the pre-test probability, and consequently, the test PPV, before requesting cTn. Clinical judgment becomes more important with new than with conventional cTn assays; type of symptoms, age and gender, presence of cardiovascular risk factors, history of cardiovascular disease (CVD), impaired renal function, atypical electrocardiographic changes, and so on may all significantly predispose to a higher risk for ACS and indicate (or not) the need for hsTn ordering. Very recently, an authoritative group has highlighted in a clear recommendation when hsTn should be requested (Table 1) [45].

Table 1

The clinical context in which the request of hsTn is justified. Adapted from ref. [45].

The measurement of hsTn is indicated in patients presenting to the ED with ongoing or previous chest pain.
The measurement of hsTn should be considered in patients without chest pain, but who have one of the following symptoms leading to the suspicion of ACS: sudden onset isolated dyspnea, diaphoresis, palpitations, nausea/vomiting, sudden fatigue, an acute confused state, or syncope.
In case of symptoms listed under item 2, the hsTn measurement should be requested in patients with at least one of the following conditions: previous stroke, previous heart failure, diabetes mellitus, age >75 years, female gender.

Are there better alternatives to the 99th percentile cutoff?

Thresholds alternatives to the 99th percentile limit have been proposed to limit the number of “false-positive” hsTn results. A classical approach to derive decision limits is to use receiver operator characteristic (ROC) curves to identify the cutoff representing the best compromise between test sensitivity and specificity. Using ROC curves, Reiter et al. [46], evaluating the diagnostic performance of hsTnT in >70-year-old patients presenting to the ED with acute chest pain, derived a best cutoff of 54 ng/L, which is far higher than the manufacturer-defined 99th percentile cutoff of 14 ng/L. As expected, the increase in the threshold concentration markedly increased the specificity of the test (from 49% to 96%), shifting the PPV from 38% to 86%, decreasing the NPV (from 99% to 93%), and resulting in a relevant decrease in AMI cases [46].

As emergency medicine is constantly seeking safe ruling-out processes, the minimum standard for the accuracy of a ruling-out discharge pathway for chest pain should be at least 99% NPV [47]. Body et al. [48] have achieved this NPV using the limit of blank (LoB) of hsTnT (3 ng/L) as decisional threshold at patient admission. However, the impact of this strategy has been criticized, as an undetectable hsTn at admission does not necessarily mean patient discharge and a very early rule-out concerns only a minority of subjects presenting to the ED (only 5 individuals out of 150 per day) [49].

Other studies have pointed out the relevance of achieving both rapid rule-out and rule-in of AMI. Reichlin et al. [50] have proposed a diagnostic algorithm including two serial hsTnT measurements at ED presentation and after 1 h, respectively. AMI was ruled out with an NPV of 100% if the former result was <12 ng/L and the absolute δ at 1 h was <3 ng/L. Patients with baseline values ≥52 ng/L or absolute δ≥5 ng/L were ruled in for AMI with 84% PPV. The intermediate group of patients (corresponding to 23% of tested subjects) remained under observation, further resulting in a prevalence of AMI of 8%. The lowest 30-day mortality for early discharged patients substantially supported the effectiveness of this approach. This study was designed accounting for the data by Mueller et al. [51] reporting that an absolute hsTnT δ change of 6.9 ng/L appears to be superior to the relative δ change for excluding an AMI. Currently, a multicenter clinical trial is aiming to prospectively validate the diagnostic performance of this algorithm in the early AMI diagnosis.

The sequel (toward absolution): optimizing diagnostic sensitivity and specificity

We have previously reported that when replacing conventional cTn T assay with hsTnT, the proportion of curves displaying a typical pattern of marker release did not change (Figure 3). An audit of typical positively raising hsTnT patterns with at least three cTn results showed that 87.5% of these were elevated already by the first sample, and the remaining became positive for the second sample, 6 h after collection of the first blood at admission. Conversely, by assaying cTn T using conventional assay, only 60.4% of typical raising curves were positive on the first sample, 32.1% became positive on the second at 6 h, and 7.5% on the third one collected 12 h after admission. The difference in sensitivity between cTn T and hsTnT at different sampling points was statistically significant (p<0.001) [28]. The early increase typical of hsTn was further confirmed in a specific study comparing hsTnT with the conventional cTn T assay in 150 patients admitted to ED with suspected ACS (ST-elevation AMI excluded) [52]. A 100% sensitivity for non-ST-elevation AMI (NSTEMI) diagnosis was already achieved by hsTnT within 3 h from admission, whereas the conventional cTn T assay did not reach this performance even after 6 h. By considering the admission result, diagnostic sensitivities of 87.1% and 58.1% for hsTnT and for the standard assay, respectively, were estimated [52]. Using hsTn, a typical increase was always detectable within 3 h from admission vs. 6–12 h for previous less sensitive generations of cTn assays (Figure 4).

Figure 4 Troponin sampling frequency and diagnostic anticipation by implementing highly sensitive assays.
Figure 4

Troponin sampling frequency and diagnostic anticipation by implementing highly sensitive assays.

With this diagnostic anticipation becoming evident, a change in guidelines was addressed to recommend fast-track rule-out hsTn protocols, i.e., samples drawn at patient presentation and after 3 h as well as the identification of typical cTn changes within curves with at least one value above the cutoff [53]. This optimized diagnostic approach resulted in an advantageous cost-benefit ratio. By comparing the 3-h diagnostic protocol for hsTnT with the conventional 10-h protocol using the fourth-generation cTn T assay, the incremental cost-effectiveness ratio (or cost per total quality-adjusted life years gained) was approximately 4-fold lower for hsTnT 3-h testing [54]. This result is further reinforced by the consistent reduction in morbidity and mortality due to the enhanced capability of hsTn to identify patients at high risk for short-term cardiovascular events (AMI and/or death at 3–12 months) [55].

Figure 5 summarizes the diagnostic algorithms currently employed in our hospital, which integrate all the evidence-based information reported above. Importantly, for the group of patients with baseline hsTnT concentrations >15 ng/L and δ<50% at 3 h, we suggest to consider an adverse prognosis and possibly retest hsTnT at 6 h. A question that remains is: what is the optimal treatment of the NSTEMI cases, previously reported as UA, that were only detected by hsTn? As recently suggested, this reclassification creates a need for novel prognostication models that would identify patients at different levels of risk and thus guide treatment decisions (i.e., conservative vs. invasive approach) [56]. Notably, hsTnT implementation in our hospital was surely associated with an increased number of NSTEMI diagnoses (from 8% to 11% of ED patients admitted with suspected ACS), but without a significant increase in cases with negative angiography findings, even if NSTEMI patients detected with hsTnT had less extended coronary artery disease (CAD) with respect to those diagnosed by conventional cTn T assay (unpublished data). Interestingly, the revascularization rate and the proportion of percutaneous transluminal coronary angioplasty stent and coronary artery bypass graft (CABG) procedures have remained unchanged.

Figure 5 Recommendations for appropriate use and interpretation of hsTnT results in our academic hospital. The above values represent initial troponin concentrations on suspected ACS patients admitted to the ED.AMI, acute myocardial infarction; NPV, negative predictive value; NSTEMI, non-ST-elevation myocardial infarction; PPV, positive predictive value.
Figure 5

Recommendations for appropriate use and interpretation of hsTnT results in our academic hospital. The above values represent initial troponin concentrations on suspected ACS patients admitted to the ED.

AMI, acute myocardial infarction; NPV, negative predictive value; NSTEMI, non-ST-elevation myocardial infarction; PPV, positive predictive value.

This is not the end: other pieces to the puzzle

When hsTn is used in clinical frameworks other than type 1 and type 2 AMI, specific thresholds or categories need to be established. Peri-procedural AMI is still an area of intense controversy concerning the use and interpretation of myocardial necrosis biomarkers. The 2011 guideline by American College of Cardiology Foundation/American Heart Association/Society for Cardiovascular Angiography and Interventions (SCAI) indicates that CK-MB and/or cTn should be measured in patients symptomatic for AMI during or after percutaneous coronary interventions (PCI) as well as in asymptomatic patients with persistent angiographic complication [57]. However, no interpretative indications or cutoffs are given. The universal definition of AMI, released 4 years before, indicated that, in these patients, a 3-fold increase above the 99th percentile limit of CK-MB/cTn (assuming a normal value at baseline) should, by convention, be defined as AMI (type 4a) [2]. Unfortunately, this was fully arbitrary and not supported by any scientific evidence or soundly anchored to outcome data [58]. Furthermore, the different analytical and clinical performances of biomarker assays were not considered, as it is clear that a CK-MB or a cTn 3-fold increase above the respective 99th percentile limits may reflect infarct sizes of a different degree that cannot be compared due to the huge difference in biomarker sensitivity for detecting myocardial damage [59]. Given these pitfalls, the 2007 universal definition of peri-procedural AMI was not embraced by the interventional community [60]. Recent data from the US National Cardiovascular Data Registry have reported that most institutions do not routinely check cTn post-procedure, being reluctant to include in their diagnostic armamentarium a test of unclear prognostic meaning [61]. Although limited, outcome data on cTn after PCI are, however, available. A study performed at the Karolinska Institute in Sweden reported that only cTn T concentrations >300 ng/L (i.e., 15 times the 99th percentile limit) measured at 06:00 am after PCI were associated with a significant increase in the risk of death or non-fatal AMI (hazard ratio 2.81, 95th confidence interval 1.32–6.00) [62].

In the third universal definition of AMI, the recommendations for diagnosing type 4a AMI (i.e., AMI related to PCI) have been revised, and once again in an arbitrary way, a peri-procedural AMI defined by a cTn elevation >5-fold, the 99th percentile limit in patients with normal baseline values, when associated with either clinical symptoms or signs suggestive of myocardial ischemia [3]. This was, however, again ignored in the expert consensus document by SCAI updating the definition of clinically relevant AMI after coronary revascularization [63]. CK-MB remains the recommended marker, and cTn use is suggested only in the absence of CM-MB measurements, with a cutoff (to be used within 48 h of the PCI) ≥70× the local 99th percentile limit or ≥35× when new pathological Q waves appear at electrocardiogram. These recommendations are valid after both PCI and CABG procedures; furthermore, the use of hsTn to assess post-PCI or CABG myonecrosis is specifically discouraged [63]. This position is, however, prone to criticism because both the absolute concentrations and the relative increase in cardiac biomarkers are far higher after CABG with respect to PCI [64]. Moreover, hsTn, specifically designed to detect lowest concentrations with high precision, displays higher cTn concentrations, such as those detected after CABG, results comparable to those of previous generations being therefore usable interchangeably [21, 27].

Patients submitted to non-cardiac surgical interventions may also benefit from the improved sensitivity of cTn assays. This setting is associated with relevant cardiovascular morbidity and mortality (with major complications rate varying from 1.7% to 3.5%) [65]. Several potential mechanisms have been recognized to contribute to the pathophysiology of peri-operative AMI, resulting in type 1 or type 2 AMI, or due to stent thrombosis, which is particularly associated with oral anti-platelet therapy interruption [66]. Notably, the pre-operative risk assessment is of great clinical relevance, and over the past decade, cTn elevation has been recognized as a relevant prognostic factor for adverse short- and longer-term outcomes [67]. With the introduction of hsTn, the prognostic value of the marker in these patients has potentially increased according to the capability to detect very minor myocardiocyte damage. A recent study has evaluated if the hsTnT increasing 24 h after surgery could detect peri-operative myocardial damage in patients at high cardiovascular risk undergoing elective non-cardiac surgery [68]. A post-operative myocardial necrosis by hsTnT was identified in 22% of patients (a proportion ∼4-fold higher with respect to conventional cTn T assay), even if only 2% of patients experienced clinically apparent AMI. Interestingly, a pre-operative hsTnT elevation, detected in 31% of patients, did not predict peri-operative MI [68]. A larger study has reported that hsTnT measured 7 days prior to surgery outperforms conventional risk algorithms, including clinical variables as well as use of natriuretic peptides, in predicting major adverse events (death, AMI, acute heart failure, and cardiac arrest/resuscitation) [69]. In addition, hsTnT was reported as the strongest independent predictor for the combined endpoint, as patients with concentrations above the 99th percentile cutoff had a 2.6-fold increased risk for the occurrence of one of the defined adverse events [69].

hsTn in stable CAD as a risk biomarker: the “sound of silence”

hsTn has been additionally suggested as a potential screening tool to identify asymptomatic individuals who are at high risk for CVD, both in patients with stable CAD as well as in general population [70]. Particularly, in stable CAD:

  • Slightly elevated hsTn concentrations are relatively common (up to a third of patients with computed tomography angiography-documented CAD).

  • hsTn elevations are strongly associated with heart failure and cardiovascular death, whereas there is weaker or no association with ischemic events.

  • Other mechanisms for release than ischemic injury likely play an important role such as mechanical stress, left ventricular mass, renal dysfunction.

Patients with stable CAD and hsTnT concentrations >18.6 ng/L had a 2.6-fold increased risk for the recurrence of cardiovascular events with respect to those with concentrations <5 ng/L, when monitored during an 8-year follow-up [71]. Similar data have been published for the HOPE cohort, in which subjects with hsTnT concentrations >21 ng/L displayed a 2-fold increased probability for AMI, stroke, or cardiovascular death [72]. Interestingly, an hsTn increase in subjects with stable CAD has been linked to coronary lesion morphology. Patients with non-calcified plaque seem to display higher hsTnT concentrations vs. those with calcified lesions, and far higher hsTnT concentrations (>21 ng/L) have been characterized in patients with remodeled non-calcified plaque [73]. Because the plaques occurring in these subjects appear to lie in an active “vulnerable” state, one could reasonably speculate on the more likely acute intermittent nature of hsTnT increase with respect to the chronic stable pattern [73].

Limitations

The original clinical data reported in this article are related to our own experience in one academic hospital center. Consequently, the results may be not applicable to other different clinical settings. Furthermore, although we often used the general term “hsTn” in this review, we actually report (and discuss) data obtained in our experience with hsTnT. Because it is known that there can be differences in results found with cTn I and cTn T assays, the clinical data reported in this article cannot be directly applied to cTn I methods [74].

Conclusions

By summarizing the running evidence in the framework of AMI, we can reasonably conclude that hsTn, by markedly improving the analytical performance at low-range marker concentrations, has allowed the safe clinical application of international recommendations (i.e., the 99th percentile limit concept). The impact of the universal definition of AMI has resulted in an increase of this diagnosis by approximately 25%, thus becoming an independent predictor of mortality at 10 years [75].

hsTn have shortened the time to diagnose NSTEMI as well as the observation period of patients with suspected myocardial damage in ED from 9–12 h to 3 h. Fast-track protocols for ruling out and ruling in NSTEMI have been optimized by recommending blood sampling at patient presentation and after 3 h as well as to guide interpretation of hsTn results. ED physicians and cardiologists should be advised that a more diagnostically and cost-effective use of hsTn requires that all clinical variables should be considered in order to increase the pre-test probability of AMI/ACS and that serial hsTn testing be done to detect patients at medium to high risk for AMI/ACS. Importantly, laboratories need to get involved in communicating with clinicians through education, test interpretation, tools, and internal audits of test usage and patient outcomes. Our experience shows that ED physicians and cardiologists can positively respond to the laboratory’s change in cTn strategy and be satisfied with the education and tools offered by the laboratory specialists.

In frameworks other than spontaneous AMI, the cost-benefit of applying hsTn to critical patients for their risk assessment has recently gained relevance. The capability of accurately detecting very minor cardiac damage appears to contribute more accurate prognostic information and it is expected to improve the survival trends in clinical contexts different from ACS.

Author contributions All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

Financial support: Prof Mauro Panteghini received research grants from Roche Diagnostics.

Employment or leadership: None declared.

Honorarium: None declared.

Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.


Corresponding author: Simona Ferraro, UOC Patologia Clinica, Azienda Ospedaliera “Luigi Sacco”, Via G.B. Grassi 74, Milano, Italy, Phone: +39 2 3904 2766, Fax: +39 2 503 19835, E-mail: ; and Cattedra di Biochimica Clinica e Biologia Molecolare Clinica, Dipartimento di Scienze Biomediche e Cliniche “Luigi Sacco”, Università degli Studi, Milan, Italy

About the authors

Simona Ferraro

Simona Ferraro, BSc, is specialized in clinical pathology and in medical statististics. She works at the Clinical Pathology Unit of the “Luigi Sacco” University Hospital in Milan in the field of tumor, hematological, and cardiac biomarkers. She has contributed 19 articles published in peer-reviewed journals, 15 of which as first author.

Mauro Panteghini

Prof. Mauro Panteghini is Full Professor of Clinical Biochemistry and Clinical Molecular Biology and Director of the corresponding Chair at the Medical School of the University of Milan, Italy. He also covers the following institutional positions: Director of the Department of Laboratory Medicine and Director of Clinical Pathology Unit of the “Luigi Sacco” University Hospital in Milan; Director of the Centre for Metrological Traceability in Laboratory Medicine (CIRME) of the University of Milan. He has served in a number of international and national scientific activities in the field of Laboratory Medicine. He is currently President of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM). He has published 450 manuscripts (h-index: 43) and more than 400 abstracts.

References

1. Alpert JS, Thygesen K, Antman E, Bassand JP. Myocardial infarction redefined. A consensus document of the Joint European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction. J Am Coll Cardiol 2000;36:959–69.10.1016/S0735-1097(00)00804-4Search in Google Scholar

2. Thygesen K, Alpert JS, White HD; Joint ESC/ACCF/AHA/WHF Task Force for the Redefinition of Myocardial Infarction. Universal definition of myocardial infarction. Circulation 2007;116:2634–53.10.1161/CIRCULATIONAHA.107.187397Search in Google Scholar PubMed

3. Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD: the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction. Third universal definition of myocardial infarction. J Am Coll Cardiol 2012;60:1581–98.10.1016/j.jacc.2012.08.001Search in Google Scholar PubMed

4. Omland T. Sensitive cardiac troponin assays: sense and sensibility. Eur Heart J 2012;33:944–6.10.1093/eurheartj/ehs009Search in Google Scholar PubMed

5. Wu AH, Bolger AF, Hollander JE. Growing pains with the use of high-sensitivity cardiac troponin assays. J Am Coll Cardiol 2013;62:1250–1.10.1016/j.jacc.2013.02.075Search in Google Scholar PubMed

6. Jesse RL. On the relative value of an assay versus that of a test. A history of troponin for the diagnosis of myocardial infarction. J Am Coll Cardiol 2010;55:2025–8.10.1016/j.jacc.2010.03.014Search in Google Scholar PubMed

7. Katus HA, Giannitsis E, Jaffe AS, Thygesen K. Higher sensitivity troponin assays: quo vadis? Eur Heart J 2009;30:127–8.10.1093/eurheartj/ehn555Search in Google Scholar PubMed

8. Hollander JE. Highly sensitive troponins. The answer or just more questions? J Am Coll Cardiol 2009;54:1173–5.10.1016/j.jacc.2009.05.052Search in Google Scholar PubMed

9. de Lemos JA, Morrow DA, deFilippi CR. Highly sensitive troponin assays and the cardiology community: a love/hate relationship? Clin Chem 2011;57:826–9.10.1373/clinchem.2011.163758Search in Google Scholar PubMed

10. Daniels LB. The enemy of good? Making the most of highly sensitive troponin assays. J Am Coll Cardiol 2013;61:1914–6.10.1016/j.jacc.2013.01.065Search in Google Scholar PubMed

11. Kramer CM. Avoiding the imminent plague of troponinitis. The need for reference limits for high-sensitivity cardiac troponin T. J Am Coll Cardiol 2014;63:1449–50.10.1016/j.jacc.2013.12.031Search in Google Scholar PubMed

12. Panteghini M. A critical appraisal of experimental factors influencing the definition of the 99th percentile limit for cardiac troponins. Clin Chem Lab Med 2009;47:1179–82.10.1515/CCLM.2009.295Search in Google Scholar PubMed

13. Apple FS, Murakami MM. Serum and plasma cardiac troponin I 99th percentile reference values for 3 2nd-generation assays. Clin Chem 2007;53:1558–60.10.1373/clinchem.2007.087718Search in Google Scholar PubMed

14. Hall AS, Barth JH. Universal definition of myocardial infarction. Heart 2009;95:247–9.Search in Google Scholar

15. Panteghini M. Quality requirements for troponin assays – an overview. In: Troponin monograph 2012. Alexandria, NSW: Australasian Association of Clinical Biochemists, 2012:53–61.Search in Google Scholar

16. Panteghini M, Pagani F, Yeo KT, Apple FS, Christenson RH, Dati F, et al. Evaluation of imprecision for cardiac troponin assays at low-range concentrations. Clin Chem 2004;50:327–32.10.1373/clinchem.2003.026815Search in Google Scholar PubMed

17. Tate JR. Troponin revisited 2008: assay performance. Clin Chem Lab Med 2008;46:1489–500.10.1515/CCLM.2008.292Search in Google Scholar PubMed

18. Panteghini M. Assay-related issues in the measurement of cardiac troponins. Clin Chim Acta 2009;402:88–93.10.1016/j.cca.2008.12.037Search in Google Scholar PubMed

19. Latini R, Masson S, Anand IS, Missov E, Carlson M, Vago T, et al. Prognostic value of very low plasma concentrations of troponin T in patients with stable chronic heart failure. Circulation 2007;116:1242–9.10.1161/CIRCULATIONAHA.106.655076Search in Google Scholar PubMed

20. Jarausch J, Braun S, Dolci A, Jaffe A, Mühlbacher A, Olven C, et al. Evaluation of a development version of the Elecsys® high sensitive troponin T assay. Clin Chem 2008;54(Suppl):A91.Search in Google Scholar

21. Saenger AK, Beyrau R, Braun S, Cooray R, Dolci A, Freidank H, et al. Multicenter analytical evaluation of a high-sensitivity troponin T assay. Clin Chim Acta 2011;412:748–54.10.1016/j.cca.2010.12.034Search in Google Scholar PubMed

22. Clinical and Laboratory Standards Institute. Protocols for determination of limits of detection and limits of quantitation; Approved guideline. CLSI document EP17-A, 2004.Search in Google Scholar

23. Tate JR, Panteghini M, Koerbin G, Hickman PE, Schneider HG, Jaffe AS. Verification of the analytical characteristics of troponin assays in the laboratory – a how to guide. In: Troponin monograph 2012. Alexandria, NSW: Australasian Association of Clinical Biochemists, 2012:69–85.Search in Google Scholar

24. UKNEQAS External Quality Assessment for Cardiac Markers in Medical Practice. Available from: http://www.ukneqas-cm.org.uk/. Accessed August 2014.Search in Google Scholar

25. Melanson SE, Conrad MJ, Mosammaparast N, Jarolim P. Implementation of a highly sensitive cardiac troponin I assay: test volumes, positivity rates and interpretation of results. Clin Chem Acta 2008;395:57–61.10.1016/j.cca.2008.05.007Search in Google Scholar PubMed

26. Giannitsis E. High sensitive troponin assays in the diagnosis of ACS patients: implementation in everyday practice. In: proCardio Symposium, 8th International Symposium on Cardiac Biomarkers, August 2011, Paris, France.Search in Google Scholar

27. Dolci A, Braga F, Valente C, Guzzetti S, Panteghini M. Impact of implementation of the high-sensitivity cardiac troponin T assay in a university hospital setting. Clin Chem 2011;57:1211–2.10.1373/clinchem.2011.164426Search in Google Scholar PubMed

28. Braga F, Dolci A, Valente C, Boido A, Moretti A, Guzzetti S, et al. Impact of the implementation of highly sensitive cardiac troponin T assay in a university hospital setting. Biochim Clin 2011;35:301–6.Search in Google Scholar

29. Dolci A, Panteghini M. The exciting story of cardiac biomarkers: from retrospective detection to gold diagnostic standard for acute myocardial infarction and more. Clin Chim Acta 2006;369:179–87.10.1016/j.cca.2006.02.042Search in Google Scholar PubMed

30. Collinson PO, Stubbs PJ. Are troponins confusing? Heart 2003;89:1285–7.10.1136/heart.89.11.1285Search in Google Scholar PubMed PubMed Central

31. Mendis S, Thygesen K, Kuulasmaa K, Giampaoli S, Mahonen M, Blackett KN, et al. World Health Organization definition of myocardial infarction: 2008–09 revision. Int J Epidemiol 2011;40:139–46.10.1093/ije/dyq165Search in Google Scholar PubMed

32. Braunwald E, Morrow DA. Unstable angina is it time for a requiem? Circulation 2013;127:2452–7.10.1161/CIRCULATIONAHA.113.001258Search in Google Scholar PubMed

33. Morrow DA. Clinical application of sensitive troponin assays. N Engl J Med 2009;361:913–5.10.1056/NEJMe0905790Search in Google Scholar PubMed

34. Wu AH, Jaffe AS. The clinical need for high-sensitivity cardiac troponin assays for acute coronary syndromes and the role for serial testing. Am Heart J 2008;155:208–14.10.1016/j.ahj.2007.10.016Search in Google Scholar PubMed

35. Cullen L, Jaffe AS. The clinical evidence for ‘delta troponin’ use. In: Troponin monograph 2012. Alexandria, NSW: Australasian Association of Clinical Biochemists, 2012:45–52.Search in Google Scholar

36. Wu AH, Lu QA, Todd J, Moecks J, Wians F. Short- and long-term biological variation in cardiac troponin I measured with a high-sensitivity assay: implications for clinical practice. Clin Chem 2009;55:52–8.10.1373/clinchem.2008.107391Search in Google Scholar PubMed

37. Vasile VC, Saenger AK, Kroning JM, Jaffe AS. Biological and analytical variability of a novel high-sensitivity cardiac troponin T assay. Clin Chem 2010;56:1086–90.10.1373/clinchem.2009.140616Search in Google Scholar PubMed

38. Frankenstein L, Wu AH, Hallermayer K, Wians FH, Giannitsis E, Katus HA. Biological variation and reference change value of high-sensitivity troponin T in healthy individuals during short and intermediate follow-up periods. Clin Chem 2011;57:1068–71.10.1373/clinchem.2010.158964Search in Google Scholar PubMed

39. De Lemos JA. Increasingly sensitive assays for cardiac troponins. A review. JAMA 2013;309:2262–9.10.1001/jama.2013.5809Search in Google Scholar PubMed

40. Supak Smolcic V, Drvar V, Hrabric Vlah S, Bilic-Zulle L. The issue of rational ordering of high sensitive cardiac troponin T test in outpatients’ emergency room. Clin Chem Lab Med 2014;52:S1142.Search in Google Scholar

41. Wu AH. Interpretation of high sensitivity cardiac troponin I results: reference to biological variability in patients who present to the emergency room with chest pain: case report series. Clin Chim Acta 2009;401:170–4.10.1016/j.cca.2008.12.004Search in Google Scholar PubMed

42. Dolci A, Braga F, Valente C, Boido A, Moretti A, Guzzetti S, et al. Impact of the highly sensitive cardiac troponin (hsTnT) assay on the triage of emergency department (ED) patients. Clin Chem Lab Med 2011;49:S292.Search in Google Scholar

43. Meune C, Reichlin T, Irfan A, Schaub N, Twerenbold R, Meissner J, et al. How safe is the outpatient management of patients with acute chest pain and mildly increased cardiac troponin concentrations? Clin Chem 2012;58:916–28.10.1373/clinchem.2011.178053Search in Google Scholar PubMed

44. Galvani M, Panteghini M, Ottani F, Cappelletti P, Chiarella F, Chiariello M, et al. The new definition of myocardial infarction: analysis of the ESC/ACC consensus document and reflections on its applicability to the Italian Health System. Ital Heart J 2002;3:543–57.Search in Google Scholar

45. Casagranda I, Cavazza M, Clerico A, Galvani M, Ottani F, Zaninotto M, et al. Proposal for the use in emergency departments of cardiac troponins measured with the latest generation methods in patients with suspected acute coronary syndrome without persistent ST-segment elevation. Clin Chem Lab Med 2013;51:1727–37.10.1515/cclm-2013-0423Search in Google Scholar PubMed

46. Reiter M, Twerenbold R, Reichlin T, Haaf P, Peter F, Meissner J, et al. Early diagnosis of acute myocardial infarction in the elderly using more sensitive cardiac troponin assays. Eur Heart J 2011;32:1379–89.10.1093/eurheartj/ehr033Search in Google Scholar PubMed

47. Than M, Cullen L. The challenges of sensitive and highly sensitive troponin assay use in the emergency department – the emergency physician’s perspective. In: Troponin monograph 2012. Alexandria, NSW: Australasian Association of Clinical Biochemists, 2012:27–32.Search in Google Scholar

48. Body R, Carley S, McDowell G, Jaffe AS, France M, Cruickshank K, et al. Rapid exclusion of acute myocardial infarction in patients with undetectable troponin using a high-sensitivity assay. J Am Coll Cardiol 2011;58:1332–9.10.1016/j.jacc.2011.06.026Search in Google Scholar PubMed

49. Ray P. Role of the high sensitive troponin assays in the universal definition of AMI. The emergency doctor’s perspective. In: proCardio Symposium, 8th International Symposium on Cardiac Biomarkers, August 2011, Paris, France.Search in Google Scholar

50. Reichlin T, Schindler C, Drexler B, Twerenbold R, Reiter M, Zellweger C, et al. One-hour rule-out and rule-in of acute myocardial infarction using high-sensitivity cardiac troponin T. JAMA 2012;172:1211–8.Search in Google Scholar

51. Mueller M, Biener M, Vafaie M, Doerr S, Keller T, Blankenberg S, et al. Absolute and relative kinetic changes of high-sensitivity cardiac troponin T in acute coronary syndrome and in patients with increased troponin in the absence of acute coronary syndrome. Clin Chem 2012;58:209–18.10.1373/clinchem.2011.171827Search in Google Scholar PubMed

52. Braga F, Dolci A, Cavallero A, Ghezzi A, Infusino I, Milano M, et al. Evaluation of the sensitivity of two highly sensitive troponin assays for early detection of non ST-elevation myocardial infarction (NSTEMI). Biochim Clin 2011;35:186–9.Search in Google Scholar

53. Hamm CW, Bassand JP, Agewall S, Bax J, Boersma E, Bueno H, et al. ESC Guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation. The Task Force for the management of acute coronary syndromes (ACS) in patients presenting without persistent ST-segment elevation of the European Society of Cardiology (ESC). Eur Heart J 2011;32:2999–3054.10.1093/eurheartj/ehr236Search in Google Scholar PubMed

54. Thokala P, Goodacre SW, O Collinson PO, Stevens JW, Mills NL, Newby DE, et al. Cost-effectiveness of presentation versus delayed troponin testing for acute myocardial infarction. Heart 2012;98:1498–503.10.1136/heartjnl-2012-302188Search in Google Scholar PubMed

55. Mills NL, Churchhouse AM, Lee KK, Anand A, Gamble D, Shah AS, et al. Implementation of a sensitive troponin I assay and risk of recurrent myocardial infarction and death in patients with suspected acute coronary syndrome. JAMA 2011:305; 1210–6.10.1001/jama.2011.338Search in Google Scholar PubMed

56. Filippatos G, Farmakis D, Parissis J. Novel biomarkers in acute coronary syndromes. New molecules, new concepts, but what about new treatment strategies? J Am Coll Cardiol 2014;63:1654–6.10.1016/j.jacc.2013.11.055Search in Google Scholar PubMed

57. Levine GN, Bates ER, Blankenship JC, Bailey SR, Bitt JA, Cercek B, et al. 2011 ACCF/AHA/SCAI Guideline for Percutaneous Coronary Intervention: executive summary. A report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Society for Cardiovascular Angiography and Interventions. J Am Coll Cardiol 2011;58:2550–83.10.1016/j.jacc.2011.08.006Search in Google Scholar

58. Panteghini M. The new definition of myocardial infarction and the impact of troponin determination on clinical practice. Int J Cardiol 2006;106:298–306.10.1016/j.ijcard.2005.01.046Search in Google Scholar PubMed

59. Valgimigli M, Leonardi S. Peri-procedural myocardial infarction and troponism: a quick journey through the land of confusion. Heart 2012;98:1400–1.10.1136/heartjnl-2012-302612Search in Google Scholar PubMed

60. White HD, Thygesen K, Alpert JS, Jaffe AS. Ischaemic heart disease. Clinical implications of the Third Universal Definition of Myocardial Infarction. Heart 2014;100:424–32.10.1136/heartjnl-2012-302976Search in Google Scholar PubMed

61. Masoudi FA, Ponirakis A, Yeh RW, Maddox TM, Beachy J, Casale PN, et al. Cardiovascular Care Facts. A report from the National Cardiovascular Data Registry: 2011. J Am Coll Cardiol 2013;62:1931–47.10.1016/j.jacc.2013.05.099Search in Google Scholar PubMed

62. Saleh N, Svane B, Hansson LO, Jensen J, Nilsson T, Danielsson O, et al. Response of serum C reactive protein to percutaneous coronary intervention has prognostic value. Clin Chem 2005;51:2124–30.10.1373/clinchem.2005.048082Search in Google Scholar PubMed

63. Moussa ID1, Klein LW, Shah B, Mehran R, Mack MJ, Brilakis ES, et al. Consideration of a new definition of clinically relevant myocardial infarction after coronary revascularization: an expert consensus document from the Society for Cardiovascular Angiography and Interventions (SCAI). J Am Coll Cardiol 2013;62:1563–70.10.1016/j.jacc.2013.08.720Search in Google Scholar PubMed PubMed Central

64. Panteghini M. Cardiac marker testing following surgery. Ann Clin Biochem 2001;38:287–8.Search in Google Scholar

65. Boersma E, Kertai MD, Schouten O, Bax JJ, Noordzij P, Steyerberg EW, et al. Perioperative cardiovascular mortality in noncardiac surgery: validation of the Lee cardiac risk index. Am J Med 2005;118:1134–41.10.1016/j.amjmed.2005.01.064Search in Google Scholar PubMed

66. Landesberg G, Beattie WS, Mosseri M, Jaffe AS, Alpert JS. Perioperative myocardial infarction. Circulation 2009;119: 2936–44.10.1161/CIRCULATIONAHA.108.828228Search in Google Scholar PubMed

67. Bursi F, Babuin L, Barbieri A, Politi L, Zennaro M, Grimaldi T, et al. Vascular surgery patients: perioperative and long-term risk according to the ACC/AHA guidelines, the additive role of post-operative troponin elevation. Eur Heart J 2005;26:2448–56.10.1093/eurheartj/ehi430Search in Google Scholar PubMed

68. Alcock RF, Kouzios D, Naoum C, Hillis GS, Brieger BD. Perioperative myocardial necrosis in patients at high cardiovascular risk undergoing elective non-cardiac surgery. Heart 2012;98:792–8.10.1136/heartjnl-2011-301577Search in Google Scholar PubMed

69. Weber M, Luchner A, Manfred S, Mueller C, Liebetrau C, Schlitt A, et al. Incremental value of high-sensitive troponin T in addition to the revised cardiac index for perioperative risk stratification in non-cardiac surgery. Eur Heart J 2013;34:853–62.10.1093/eurheartj/ehs445Search in Google Scholar PubMed

70. Wijns W, Crea F. Elevated troponin in stable coronary artery disease: the sound of silence. Heart 2011;97:785–6.10.1136/hrt.2010.207035Search in Google Scholar PubMed

71. Koenig W, Breitling LP, Hahmann H, Wusten B, Brenner H, Rothenbacher D. Cardiac troponin T measured by a high-sensitivity assay predicts recurrent cardiovascular events in stable coronary heart disease patients with 8-year follow-up. Clin Chem 2012;58:1215–24.10.1373/clinchem.2012.183319Search in Google Scholar PubMed

72. McQueen MJ, Kavsak PA, Xu L, Shestakovska O, Yusuf S. Predicting myocardial infarction and other serious cardiac outcomes using high-sensitivity cardiac troponin T in a high-risk stable population. Clin Biochem 2013;46:5–9.10.1016/j.clinbiochem.2012.10.003Search in Google Scholar PubMed

73. Korosoglou G, Lehrke S, Mueller D, Hosch W, Kauczor HU, Humpert PM, et al. Determinants of troponin release in patients with stable coronary artery disease: insights from CT angiography characteristics of atherosclerotic plaque. Heart 2011;97:823–31.10.1136/hrt.2010.193201Search in Google Scholar PubMed

74. Lee KK, Shah AS, Mills NL. Diagnosis of myocardial infarction: cardiac troponin I or troponin T? Clin Biochem 2014;47: 319–20.10.1016/j.clinbiochem.2014.03.002Search in Google Scholar PubMed

75. Costa FM, Ferreira J, Aguiar C, Dores H, Figueira J, Mendes M. Impact of ESC/ACCF/AHA/WHF universal definition of myocardial infarction on mortality at 10 years. Eur Heart J 2012;33:2544–50.10.1093/eurheartj/ehs311Search in Google Scholar PubMed

Received: 2014-8-13
Accepted: 2014-9-3
Published Online: 2014-10-2
Published in Print: 2015-4-1

©2015 by De Gruyter

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. High-sensitivity assays for cardiac troponins
  4. Reviews
  5. High sensitivity cardiac troponin assays in the clinical laboratories
  6. Laboratory medicine as the science that underpins medicine: the “high-sensitivity” troponin paradigm
  7. Cardiac troponin assays: a review of quantitative point-of-care devices and their efficacy in the diagnosis of myocardial infarction
  8. Cardiovascular Diseases
  9. Evaluation of standardization capability of current cardiac troponin I assays by a correlation study: results of an IFCC pilot project
  10. Estimation of age- and comorbidities-adjusted percentiles of high-sensitivity cardiac troponin T levels in the elderly
  11. High-sensitivity cardiac troponin I in the general population – defining reference populations for the determination of the 99th percentile in the Gutenberg Health Study
  12. Kinetics of high-sensitivity cardiac troponin T or troponin I compared to creatine kinase in patients with revascularized acute myocardial infarction
  13. Biological variation of high sensitivity cardiac troponin-T in stable dialysis patients: implications for clinical practice
  14. Diagnosis of acute myocardial infarction in patients with renal insufficiency using high-sensitivity troponin T
  15. General Clinical Chemistry and Laboratory Medicine
  16. Active intervention in hospital test request panels pays
  17. An approach to establish the uncertainty budget of catalytic activity concentration measurements in a reference laboratory
  18. Uric acid: a potential biomarker of multiple sclerosis and of its disability
  19. Experience with the first fully automated chemiluminescence immunoassay for the quantification of 1α, 25-dihydroxy-vitamin D
  20. I-FABP and L-FABP are early markers for abdominal injury with limited prognostic value for secondary organ failures in the post-traumatic course
  21. Quantification of piperacillin, tazobactam, cefepime, meropenem, ciprofloxacin and linezolid in serum using an isotope dilution UHPLC-MS/MS method with semi-automated sample preparation
  22. Plasma visfatin/nicotinamide phosphoribosyltransferase (visfatin/NAMPT) concentration is not related to kidney function in elderly subjects
  23. Diagnostic performance study of an antigen microarray for the detection of antiphospholipid antibodies in human serum
  24. Dry ice exposure of plasma samples influences pH and lupus anticoagulant analysis
  25. Reference Values and Biological Variations
  26. A new robust statistical model for interpretation of differences in serial test results from an individual
  27. Cancer Diagnostics
  28. The value of red blood cell distribution width in endometrial cancer
  29. Letters to the Editors
  30. Serum high-sensitivity troponin concentrations in a multi-ethnic Asian population of stable chronic kidney disease patients
  31. Comparison between BNP values measured in capillary blood samples with a POCT method and those measured in plasma venous samples with an automated platform
  32. Reply to the article entitled “Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D” by Farrell et al., Clin Chem Lab Med 2014;52:1579–87
  33. Reply. Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D. Authors’ response to the Letter to the Editor by Donnelly et al.
  34. Therapeutic drug monitoring of voriconazole: validation of a novel ARK™ immunoassay and comparison with ultra-high performance liquid chromatography
  35. Interference of C-reactive protein with clotting times
Downloaded on 12.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2014-0812/html
Scroll to top button