Home Dynamic loading leakage test of dental abutment/implant connections based upon a novel implant system with an abutment switch feature: an in vitro study
Article Open Access

Dynamic loading leakage test of dental abutment/implant connections based upon a novel implant system with an abutment switch feature: an in vitro study

  • Timea Mezey , Christoph Bourauel ORCID logo and Ludger Keilig ORCID logo EMAIL logo
Published/Copyright: May 5, 2025

Abstract

Objectives

The objective of this study was to evaluate the sealing ability of the implant-abutment interface in a specific implant system in comparison to other implant systems under conditions of increasing dynamic loading.

Methods

Three different implants and four abutment types were tested: one implant with a conical abutment connection, one implant with a flat abutment connection, and one implant with an abutment switch feature and two different abutment connections. The tests consisted of a phase of cyclic loading followed by a leakage test. The maximum loading force was increased, and the procedure was repeated, until either the implant-abutment connection failed, or a leakage was detected. Loading criteria were modified based on the ISO standard 14801:2016.

Results

The conical implant abutment connections exhibited fracture prior to leakage at varying failure forces and loading cycles, whereas the platform abutment type showed leakage or fracture at a relatively low force. The two different abutment types used on the same implant system showed extremely different results.

Conclusions

Literature shows similar high chewing forces mostly only for healthy subjects. For patients with dental implants, leakage is still preferable to fracture. In conclusion, the test showed that some implants with a conical abutment are more resistant to leakage and may be preferable if higher chewing forces are expected.

Introduction

The failure of dental implants is multifactorial and can be attributed to a number of reasons. Previous studies have addressed the question of whether the cause of implant failure is primarily mechanical, resulting from incorrect loading, or whether biological complications, such as microbial colonization, play a greater role [1]. As posited by Kim et al., the notable discrepancies between implants and natural teeth can be attributed to the absence of the periodontal ligament. This is associated with disparate tactile perception and mobility of the implants, which consequently manifests as different behaviors when subject to excessive loading [1]. The tooth mobility provided by the periodontal ligament enables the teeth to adapt to varying load conditions, including those resulting from jaw deformation during mastication. In contrast to natural teeth, dental implants lack the periodontal ligament, which transmits sensory feedback regarding the magnitude of chewing forces. Consequently, it is hypothesized that implants are subjected to considerably higher chewing forces than their natural counterparts [1],2]. It is assumed that occlusal contacts, parafunctions and extra axial loading along with occlusal forces play an important role in their survival rate [3], [4], [5]. Conversely, other studies seek to demonstrate that implant loss is primarily due to the development of peri-implantitis [6].

In relation to this matter, the configuration of the implant-abutment junction is of significance regarding the longevity of dental implants and has already been the subject of investigation in several previous studies [7], [8], [9]. Dynamic loading (e.g. during the chewing process) results in micromotion of the abutment. It has been observed that this micromotion can lead to a bacterial colonization in the interior of the implant and its associated soft tissues around two-part implant systems, despite the absence of this phenomenon in one-part implants [10], [11], [12], [13]. Conventional implant systems feature either a conical or a butt joint, more precisely platform abutment interface with or without anti-rotation protection. The use of conical abutments has become increasingly common over time, as evidenced by the fact that the majority of studies are conducted with these superstructures in mind [14]. However, depending on the initial situation or the practitioner, it is still advisable to use butt joint abutments [15]. The tioLogic® TWINFIT implant series features a newly developed “abutment switch” and permits a transition between conical and butt joint abutment configuration at any instance. According to the manufacturer Dentaurum®, this system has been designed to give the dentist maximum flexibility during implant placement.

The purpose of this investigation was to compare the sealing capability of the implant-abutment interface between the tioLogic® TWINFIT implant series and implants from other manufacturers. A dynamic loading leakage test was performed, according to which it is to be weighed whether the design can withstand intra-oral exposure and to assess the implants’ clinical behavior. The objective was to examine the impact of an increasing dynamic load on the sealing capacities of the implants.

Materials and methods

Implant systems

The implant systems used are listed in Table 1, together with information on implant dimension, abutment type, and manufacturer name.

Table 1:

The different Implants used in this study and their dimensions as well as their material.

Manufacturer Implant system Implant dimension Abutment system
Dentaurum

Ispringen, Germany
TWINFIT Implant M Ø 4.2 mm, L 15.0 mm CAD/CAM Titanium base M, platform

CAD/CAM Titanium base M, conical
Camlog

Basel, Switzerland
Conelog® Progressive Implant Ø 4.3 mm, L 16.0 mm Conelog® Ti base CAD/CAM Crown
Dentsply Sirona

North Carolina, USA
Astra Tech Osseo Speed™ EV Ø 4.2 mm, L 15.0 mm TiBase Cerec/inLab

Tiologic® TWINFIT

The tioLogic® TWINFIT implant system features an “abutment switch” that enables the selection of either a “conical” or a “platform” abutment configuration, eliminating the need for the practitioner to decide in advance which abutment type to use. Both abutments have five indexes and an integrated platform switch. It should be noted that the platform abutment offered by Dentaurum® deviates from the standard platform abutments, as it has a slight cone at the upper part (Figure 1). We utilized the TWINFIT implant (Dentaurum, Ispringen, Germany) in size “M”, which has a 4.2 mm diameter and a length of 15.0 mm.

Figure 1: 
Conical and platform abutment in cross-section. A comparison of the conical and platform abutment of the tioLogic® TWINFIT implant system in cross-section reveals that, although the platform abutment possesses a slight cone in the upper area, it is predominantly positioned with the flat surface on the implant shoulder. (Figure: courtesy of Dentaurum).
Figure 1:

Conical and platform abutment in cross-section. A comparison of the conical and platform abutment of the tioLogic® TWINFIT implant system in cross-section reveals that, although the platform abutment possesses a slight cone in the upper area, it is predominantly positioned with the flat surface on the implant shoulder. (Figure: courtesy of Dentaurum).

Astra Tech OsseoSpeed™ EV

According to Dentsply Sirona Inc. (North Carolina, USA), the Astra Tech OsseoSpeed™ EV series provides a wide range of implant options. For this study, we selected the OsseoSpeed™ EV implant with conical abutment geometry, featuring an upper part with a diameter of 4.2 mm and a total length of 15.0 mm. The implant body is 0.6 mm slimmer than the upper part, measuring 3.6 mm. The manufacturer claims that these implants are suitable for narrow jawbones. The Astra Tech OsseoSpeed™ EV series is equipped with seven indexes, enabling three different abutment positions: “One-position-only” using all seven indices, with six indices, or without indices.

CONELOG® Progressive

The CONELOG® implant system, manufactured by CAMLOG Vertriebs GmbH (Switzerland), is characterized by the deep internal positioning of the implant abutment connection and an integrated platform switch. According to the manufacturer, positioning the implant abutment connection further away from the marginal bone supports bone preservation. The implants feature a 7.5° conical abutment connection and three indexes. We utilized the 16.0 mm long implant with an upper diameter of 4.3 mm and an apical diameter of 2.7 mm.

Specimen preparation

To conduct the dynamic loading leak test, the implants were altered to include an apical Luer-Lock hose fitting that was linked to the pump (Figure 2). Blue water infused with standard ink (LAMY T 52 ink – blue, LAMY GmbH, Heidelberg, Germany) was then pumped into the specimen. After that, the threaded part of the implants was set in resin (PalaXpress, Haereus Kulzer GmbH, Hanau, Germany) within an 18 mm long copper tube using an embedding guide. The resin level was set to replicate a 3 mm bone loss. The abutments were fitted with a spherical load head on top, with the sphere’s center positioned 8 mm above the implant shoulder, resulting in a total height of 11 mm for the dynamic load applied (Figure 3). Before tightening the abutments with a 30 Ncm torque in accordance with the manufacturer’s instructions, the pump was activated to vent the remaining air from the tubes and the specimen, to prevent the formation of air bubbles within the hose, which could otherwise obstruct its functionality. Each specimen was subjected to an initial leak test at maximum pressure of the pump for 1 h after the screws were tightened, before undergoing the actual dynamic loading leak test.

Figure 2: 
Schematic representation of the prepared test specimen. The cross-section shows the implant with abutment, embedded in a copper tube, as well as the apical modification using a Luer connection. (Figure: courtesy of Dentaurum).
Figure 2:

Schematic representation of the prepared test specimen. The cross-section shows the implant with abutment, embedded in a copper tube, as well as the apical modification using a Luer connection. (Figure: courtesy of Dentaurum).

Figure 3: 
The four different tested abutment systems with spherical load head on top. (A) tioLogic® TWINFIT M conical specimen. (B) tioLogic® TWINFIT M platform specimen. (C) Astra Tech OsseoSpeed™ EV specimen. (D) CONELOG® specimen.
Figure 3:

The four different tested abutment systems with spherical load head on top. (A) tioLogic® TWINFIT M conical specimen. (B) tioLogic® TWINFIT M platform specimen. (C) Astra Tech OsseoSpeed™ EV specimen. (D) CONELOG® specimen.

Apparatus

A combined apparatus, which corresponds to previous tests, was necessary to conduct both, the dynamic fatigue loading and the leakage test [8], [16]. The “Dyna-Mess TP 5 kN HF” machine (DYNA-MESS Prüfsysteme GmbH, Aachen/Stolberg, Germany) meets the requirements specified in the DIN EN ISO 14801:2016 standard: it applies and monitors the prescribed frequency, as well as maximum and minimum load and the number of loading cycles and is able to apply the load without a deviation from ±5 % at maximum load [17]. A micropump (Typ SP 270 EC-L 12 V, Schwarzer Precision GmbH + Co. KG, Essen, Germany), powered by a commercial laboratory power supply (Voltcraft LPS 1153, Conrad Electronic, Hirschau, Germany), capable of maintaining a constant pressure of 500 mBar was used for the leakage test. The pump pushes the ink-infused water through the apical Luer-Lock into the inner cavity of the implant. If there is a gap between implant and abutment, the blue water will be pushed through this gap into the surrounding water bath, and will be detectable visibly.

Loading criteria

The objective is to analyze the sealing capability of the implant-abutment interface under extra-axial loading. The criteria for dynamic fatigue testing of endosseous dental implants are defined in the DIN EN ISO 14801 standard, which offers a solid foundation for comparing results among various specimen types [17]. A specimen holder was prepared with a central bore for the above-mentioned copper tube to mount the implant in a way that the loading direction of the apparatus is in a specific angle without any lateral restriction. Contrary to the DIN standard, the loading angle was set at an angulation of 20° relative to the implant’s long axis, to allow comparison with the test on which our setup is based [16]. Following the ISO standard, the loading force was applied via the custom designed spherical head of the abutments (see Figure 3) without constraining lateral movement. The applied force varied between 100 and 10 % of maximum loading. To conduct the leakage test, specimens were submerged in distilled water (aqua dest, Ampuwa, Fresenius Medical Care, Bad Homburg v.d. Höhe, Germany) at a temperature of 37 ± 2 °C. The overall setup including a scheme of the loading geometry is shown in Figure 4.

Figure 4: 
Schematics and photo of the overall setup. (A) Scheme of the loading geometry [17]. D: angle of the incoming force, which is changed to 20° in our setup. I: The loading is applied at a height of 11 mm above the implant shoulder. (B) Embedding apparatus with copper tube and embedded specimen. The outer frame provides space for the copper tube. An implant with the appropriate diameter fits into the inner recess so that it is held in position during the embedding process. The apical Luer-Lock-hose fitting at the bottom of the specimen is visible. (C) Specimen holder. A wedge ensured the angulation of 20°. (D) Specimen in the specimen holder mounted in the Dyna-Mess device in the water bath.
Figure 4:

Schematics and photo of the overall setup. (A) Scheme of the loading geometry [17]. D: angle of the incoming force, which is changed to 20° in our setup. I: The loading is applied at a height of 11 mm above the implant shoulder. (B) Embedding apparatus with copper tube and embedded specimen. The outer frame provides space for the copper tube. An implant with the appropriate diameter fits into the inner recess so that it is held in position during the embedding process. The apical Luer-Lock-hose fitting at the bottom of the specimen is visible. (C) Specimen holder. A wedge ensured the angulation of 20°. (D) Specimen in the specimen holder mounted in the Dyna-Mess device in the water bath.

Test procedure and failure criteria

After the specimen holder was mounted in the water bath and placed in the Dyna-Mess device (Figure 4D), the test procedure was started. A cyclic loading frequency of 2 Hz was set with an initial maximum force of 100 N, and a total of 100,000 loading cycles were performed, with the pump turned off. After this, the pump was switched on, and the specimen was submitted to another 1,000 cycles with the same force. During this loading phase the release of colored fluid from the implant-abutment interface was monitored by a camera. If the leak test was completed successfully and no colored water was released, the force was increased by 50 N for an additional 100,000 loading cycles with the pump off. Afterwards, the pump was turned on for the next 1,000 cycles and the camera recorded the results. This process was repeated until the implant-abutment interface leaked, or a fracture and plastic deformation of the implant and/or abutment was registered, respectively. Each implant-abutment combination was tested with three specimens.

Ethical approval

This in vitro study did not require any ethical approval, as no biological tissues or patient-specific data are involved in our tests.

Data analysis and statistics

In accordance with DIN EN ISO 14801, three samples for each specimen type were analyzed, and the outcome was displayed in column charts that categorize the results into average failure force, average loading cycles, and failure mode [17]. This facilitates the data comparison within one implant type. No further tests were performed due to time constraints and because there was not more than a 50 % variation in the results. However, due to the limited amount of data, it is not possible to optimally reproduce the mean value and the standard deviation. Therefore, two additional diagrams were presented to display the number of load cycles withstood and the failure force, depicting all three results. All charts and diagrams were created using Excel.

Results

Table 2 gives a full overview of all specimens and the corresponding test results from this study.

Table 2:

Detailed list of the results for each specimen tested, divided into specimen, maximum force level, survived loading cycles and comments on the failure mode.

Specimen Maximum force level Survived loading cycles Comments
TWINFIT conical 1 450 N 736,146 Machine reached max. deflection, abutment and fixation screw bended, but could be removed in one piece
TWINFIT conical 2 650 N 1,188,640 Machine reached max. deflection, abutment and fixation screw broke off at implant level
TWINFIT conical 3 650 N 1,147,646 Machine reached max. deflection, abutment and fixation screw bended, but could be removed in one piece
Conelog® Progressive 1 550 N 961,483 Machine reached max. deflection, partly fractured in the neck region at resin level, abutment could not be removed
Conelog® Progressive 2 700 N 1,311,968 Machine reached max. deflection, fractured in the neck region at resin level, fixation screw could be removed from the abutment, abutment could not be removed from the fractured implant part
Conelog® Progressive 3 650 N 1,210,965 Machine reached max. deflection, fractured in the neck region at resin level, fixation screw could be removed from the abutment, abutment could not be removed from the fractured implant part
Astra Tech Osseo Speed™ 1 350 N 585,961 Machine reached max. deflection, fractured in the neck region below resin level, fixation screw could be removed from the abutment, abutment could not be removed from the fractured implant part
Astra Tech Osseo Speed™ 2 350 N 552,733 Machine reached max. deflection, fractured in the neck region below resin level, fixation screw could be removed from the abutment, abutment could not be removed from the fractured implant part
Astra Tech Osseo Speed™ 3 300 N 497,318 Machine reached max. deflection, fractured in the neck region below resin level, head of the fixation screw and abutment could be removed from the implant
TWINFIT platform 1 250 N 402,992 Leaked while dynamic loading leakage test
TWINFIT platform 2 350 N 604,985 Leaked while dynamic loading leakage test
TWINFIT platform 3 350 N 604,987 Leaked while dynamic loading leakage test

Failure force

Figure 5 presents the average failure force for the tested implant-abutment systems, along with a detailed breakdown of results for each specimen tested. The conical tioLogic® TWINFIT implant-abutment connections did not leak prior to fracture at an average force of 580 N. The machine reached maximum deflection, resulting in fracture of one abutment and fixation screw, while the other two specimens bent at the abutment level. Each specimen in the Conelog® Progressive series fractured in the neck region at the resin level before leaking, at a rather high force around 630 N. However, the Astra Tech™ implants exhibited a permanent loading fracture at a relatively small force level of 330 N, and the implants fractured below the resin level. The platform version of the tioLogic® TWINFIT implant-abutment connection was the only one to leak prior to fracture at a relatively low force of 320 N.

Figure 5: 
Failure force in Newton (N). Detailed display for every specimen of each implant type. The columns display the mean force of failure, with each point representing an individual specimen. Some test specimens reached the same maximum force, resulting in overlapping points on the diagram. Which points overlap can also be identified through the displacement of the column between two points.
Figure 5:

Failure force in Newton (N). Detailed display for every specimen of each implant type. The columns display the mean force of failure, with each point representing an individual specimen. Some test specimens reached the same maximum force, resulting in overlapping points on the diagram. Which points overlap can also be identified through the displacement of the column between two points.

Average loading cycles until failure

Considering the absolute number of loading cycles applied to the specimens as presented in Table 2, it can be observed that the Conelog® Progressive and tioLogic® TWINFIT conical implant abutment connections endured a higher number of loading cycles than the other types of specimens. The Conelog® Progressive demonstrated the highest survival rate, with an average of 1,161,472 cycles, followed by the tioLogic® TWINFIT conical specimens with 1,024,144 cycles. Despite the conical implant abutment connection, the Astra Tech™ implants survived only around 545,337 cycles, while the tioLogic® TWINFIT platform specimen failed due to leakage after 537,654 load cycles. The data on the number of loading cycles is additionally visualized in the Supplementary Material as Figure S1.

Failure mode

The failure modes manifest in fracture, permanent deformation, and leakage. According to Figure 6, the conical connection type of the tioLogic® TWINFIT series exhibits better performance than the other samples. These implants were capable of withstanding relatively higher forces and were permanently deformed. The Conelog® Progressive and Astra Tech™ implants exhibited evidence of fracture prior to leakage, whereas the tioLogic® TWINFIT platform connection types demonstrated leakage in all cases.

Figure 6: 
Different failure modes for each implant. The following tables present the different types of failure and the corresponding number of test specimens for each implant abutment connection.
Figure 6:

Different failure modes for each implant. The following tables present the different types of failure and the corresponding number of test specimens for each implant abutment connection.

Discussion

A microgap can occur between the components of a two-part implant system. Butt joint abutments are seated with a platform on the implant shoulder and feature parallel surfaces at the inside of the implant. Due to the loose fit, it is inevitable that a small gap between the longitudinal contact surfaces appears. If fixated with an abutment screw and under static conditions, this gap is sealed. A loading force higher than the introduced force onto the abutment screw can lead to a micromovement of the abutment [18]. Due to the two parallel cone surfaces, the conical connection is able to eliminate the space between the two parts. In the hypothetical event of tilting when loaded, the abutment cone’s interior portion contacts the implant’s cone surface, preventing tilting and gap formation [19].

In several studies the implant-abutment interface in various two-part implant systems has been investigated. Due to differing testing methods, the exact location of the measured gap varies, thus making the results difficult to compare, but in general, the presence of a microgap and bacterial contamination of the internal implant surface and the abutment screw have been demonstrated [20], [21], [22]. Such contamination was not observed for single-piece implant systems [10].

Dynamic loading simulators have previously been utilized and validated to provide evidence of the presence of a microgap [23]. Suttin et al. utilized a specific method to examine the sealing capacity of the implant abutment interface [16]. The method described in our study adapted the dynamic leak test from Suttin et al. to adhere to the DIN EN ISO 14801 standard where applicable. Implants with an angle that deviates by more than 15° from the original tooth axis require pre-angulated abutments, hence, the angulation criteria were modified to a 20° axis instead of the original 30 ± 2°. In our study, we exclusively examined straight abutments.

Suttin et al. analyzed four different types of implants, including two platform abutment types (Thommen SPI® Element and 3i T3® with DCD®) and two conical abutment types (Astra Tech™ Osseo Speed™ and Straumann® Bone Level). In their study, the platform implant abutment connections showed leakage prior to failure and the conical abutment specimens exhibited different failure modes [16]. The earliest failure in our study was the tioLogic® TWINFIT sample, which exhibited a leaking failure mode at an average force of 320 N. This is slightly above to the earliest failure from Suttin et al. (Thommen SPI® Element implants, which all failed due to leakage at a mean force of 230 N). In contrast, their Astra Tech™ Osseo Speed™ and Straumann® Bone Level implants failed with an average load of 520 and 570 N, respectively. Our conical tioLogic® TWINFIT sample sustained an average load of 580 N before fracturing, indicating performance similar to that of the Straumann® Bone Level implants. Nevertheless, the Astra Tech™ Osseo Speed™ sample behaved differently in both investigations. In the aforementioned investigation, Astra Tech™ implants leaked after surviving a mean loading of 520 N. In our study, similar specimens fractured at a meagre loading of 330 N. Suttins 3i T3® with DCD® implant specimens, which are equipped with a platform abutment connection, survived the highest loading forces with an average of 740 N. However, there is a distinction to our specimen: the Thommen, Astra Tech™, and Straumann® abutments were fitted with titanium alloy screws just like our specimen, whereas the 3i T3® abutments were fitted with a Certain® Gold-Tite® screw [16].

In another study, Suttin and Towse [24] compared the 3i Certain® Connection with both the Gold-Tite® screw and the Ti alloy screw in a slightly modified test protocol that involved replacing the screws during testing. They conclude that their findings demonstrate the importance of the screw material on the sealing capability of the implant-abutment connection, given that the sample with the Ti alloy screw failed at an average force of only 500 N, whereas when switched to the gold screw an average sealing capacity up to 780 N could be reached [24]. Future investigations with our specimens could explore whether changing the abutment screw would impact the sealing capability of the tioLogic® TWINFIT platform abutment type. As for the Astra Tech™ Osseo Speed™ sample, the test could be redone with a lower level of simulated bone loss to investigate whether the early fracture is caused by the narrower implant neck.

Most importantly, screw loosening or fractures are key contributors to implant-abutment system failure [25]. Over time, it becomes more likely for implants to fail under increased loads and unfavorable extra axial forces. In 2015, Gupta et al. published a review of causes of implant failure, such as screw loosening or fracture, and how to prevent or manage these complications [4]. Some studies have suggested a relationship between microleakage and varying closing torques. Specifically, microleakage decreases as the closing torque increases [2]. According to Zipprich et al. [18], forces from different directions affect dental implant crowns during chewing. These include axial, mesiodistal, and orovestibular forces, which result in extra axial forces. Additionally, the authors noted the presence of a torque moment, in which a 100 N loading force at a 30° angle to the implant axis induces a 40 Ncm torque, resulting in abutment screw loosening and micromovements. Therefore, it is crucial to adhere to the correct fixation torques for the abutment screw [12], [26]. Additionally, it is recommended to retighten the screw after a few minutes [4]. Although we followed this procedure in our approach, it did not prevent the platform implant-abutment connection from leaking.

The objective of the fatigue load tests is to determine the forces that result in failure of the implant-abutment connection. In clinical conditions, the applied forces vary over a large range, depending on the muscles involved as well as on the characteristics of the food being processed. These factors, in turn, influence the number of chewing cycles required for mastication [27]. Food consistency changes during the process, making it challenging to replicate the outcome and causing difficulties in measuring the exact chewing forces [28]. In several studies the maximum chewing forces of patients were investigated. In some cases, different regions of the jaw were examined using different methods, making it difficult to compare the results. The average force achieved by healthy young participants in one study was 597 N for female participants, while the male participants achieved an average force of 847 N [29]. Other studies have shown average bite forces of 372 ± 133 N in patients with normal occlusion [30]. The effect of masticatory forces on natural teeth and implants differs, as the natural teeth are anchored to the bone via the periodontium, while implants are in contrast firmly fused to the bone [28]. Other studies investigated older, previously edentulous patients who are then rehabilitated with a prosthesis supported by two implants. Although higher bite forces were achieved after treatment compared to the edentulous situation, these forces were still in the low force range well below 100 N [31], [32]. Van der Built et al. conducted a study on edentulous patients who had undergone the insertion of two implants in the mandible. The maximum bite force of the patients was measured prior to treatment, after 5 and 8–14 months, and after 10 years. The initial decrease in maximum bite force from an average of 183 N before treatment to 162 N after 5 months was followed by an increase to 337 N and 341 N at the subsequent measurements, which is almost equivalent to a doubling of the force [33].

In summary, the majority of studies demonstrated that rehabilitated patients with implant-supported prostheses exhibited lower bite forces in comparison to dentate patients. The load limits of several test specimens fell within this range, while others exceeded it. However, it should be noted that implants lack the periodontal ligament and, consequently, the sensory feedback present in natural teeth [1],2]. This suggests that implants may, in fact, be exposed to very high loads, which may be further exacerbated by incorrect loading.

As this study constitutes a worst-case scenario, with the primary focus being on the identification of leakage, the maximum bite force achieved is of secondary importance. Nevertheless it is important to emphasize that, in real-life situations, appropriate occlusal contacts, elimination of bad habits or parafunctions such as bruxism, and reduction of extra axial loading can help prevent material fatigue [4], [5].

Conclusions

This research investigated the ability of the tioLogic® TWINFIT implant system to withstand intraoral exposure. It conducted a comparative analysis between the two abutment types and those of other well-established implant systems under conditions of a worst-case scenario. The results of our study confirm those of previous investigations indicating that cylindrical implant-abutment connections are more susceptible to microgap formation, thereby facilitating the penetration of fluids and bacteria into the interior of the implant. It is important to mention that the tioLogic® TWINFIT platform abutment is distinct from a conventional platform connection. Consequently, the outcomes of this study may not be directly comparable to those of previous research in this field. Furthermore, additional research approaches could facilitate further investigations into the possibility of prolonging the lifespan of the tioLogic® TWINFIT platform abutments. As the reasons for implant failure are multifactorial, the use of the tioLogic® TWINFIT implant system remains a viable recommendation. In the final analysis, the choice of abutment connection ultimately falls to the practitioner. However, in terms of the range of materials and equipment required for different implant systems, the tioLogic® TWINFIT system provides great flexibility while maintaining a manageable number of tools.


Corresponding author: Ludger Keilig, Oral Technology, Dental School, University Hospital Bonn, Bonn, Germany; and Department of Prosthetic Dentistry, Preclinical Education and Materials Science, Dental School, University Hospital Bonn, Bonn, Germany, E-mail:

Acknowledgments

We would like to thank the company Dentaurum (Ispringen, Germany) for providing the implants and preparing the Luer-Lock-hose fittings.

  1. Research ethics: Not applicable.

  2. Informed consent: Not applicable.

  3. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission. Conceptualization: TM, CB, LK; methodology: TM, CB, LK; investigation: TM; supervision: CB, LK; resources: CB, LK; formal analysis: TM, LK; writing – original draft: TM; writing – review and editing: CB, LK

  4. Use of Large Language Models, AI and Machine Learning Tools: None declared.

  5. Conflict of interest: All other authors state no conflict of interest.

  6. Research funding: None declared.

  7. Data availability: The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Appendix: Complications

During the experimentation, a minor issue arose with cyclic loading in the lower force range. As per the DIN EN ISO 14801 standard, the load deviation should not surpass 5 %. The Dyna-Mess identified particular issues when dealing with lower force levels and certain specimen types. The loading cycles experienced repeated interruptions due to the machine’s inability to regulate the force within the range of 100–200 N. These difficulties were only evident in the Conelog® Progressive and Astra Tech OsseoSpeed™ EV test series. Whenever the force or deflection tolerance was surpassed, the machine ceased and necessitated a restart. This did not affect the overall loading process or the validity of the experiment, as the number of loading cycles was accurately documented. It should be explicitly stated that this was merely an issue with the loading device in the lower force regime and did not extend to the regime of higher forces, which are pertinent to the leakage test.

References

1. Kim, Y, Oh, T-J, Misch, CE, Wang, H-L. Occlusal considerations in implant therapy: clinical guidelines with biomechanical rationale. Clin Oral Implants Res 2004;16:26–35. https://doi.org/10.1111/j.1600-0501.2004.01067.x.Search in Google Scholar PubMed

2. Jacobs, R, van Steenberghe, D. Comparison between implant-supported prostheses and teeth regarding passive threshold level. Int J Oral Maxillofac Implants 1993;8:549–54. PMID: 8112795.Search in Google Scholar

3. Goldstein, G, Goodacre, C, Taylor, T. Occlusal schemes for implant restorations: best evidence consensus statement. J Prosthodont 2021;30:84–90. https://doi.org/10.1111/jopr.13319.Search in Google Scholar PubMed

4. Gupta, S, Gupta, H, Tandan, A. Technical complications of implant-causes and management: a comprehensive review. Natl J Maxillofac Surg 2015;6:3–8. https://doi.org/10.4103/0975-5950.168233.Search in Google Scholar PubMed PubMed Central

5. Yao, KT, Kao, HC, Cheng, CK, Fang, HW, Huang, CH, Hsu, ML. Mechanical performance of conical implant-abutment connections under different cyclic loading conditions. J Mech Behav Biomed Mater 2019;90:426–32. https://doi.org/10.1016/j.jmbbm.2018.10.039.Search in Google Scholar PubMed

6. D’Ercole, S, D’Addazio, G, Di Lodovico, S, Traini, T, Di Giulio, M, Sinjari, B. Porphyromonas Gingivalis load is balanced by 0.20% chlorhexidine gel. A randomized, double-blind, controlled, microbiological and immunohistochemical human study. J Clin Med 2020;9:284. https://doi.org/10.3390/jcm9010284.Search in Google Scholar PubMed PubMed Central

7. Klinsmann, HM, Seufert, A, Wiegner, U, Kreher, T, Gowosdz, F, Wiegner, JU. Der Einfluss der Implantat-Abutment-Verbindung auf das marginale Knochenniveau. Zahnärztl Impl 2016;32:297–303.Search in Google Scholar

8. Gross, M, Abramovich, I, Weiss, EI. Microleakage at the abutment-implant interface of osseo integrated implants: a comparative study. Int J Oral Maxillofac Implants 1999;14:94–100.Search in Google Scholar

9. Jansen, VK, Conrads, G, Richter, EJ. Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants 1997;12:527–40.Search in Google Scholar

10. Broggini, N, McManus, LM, Hermann, JS, Medina, RU, Oates, TW, Schenk, RK, et al.. Persistent acute inflammation at the implant-abutment interface. J Dent Res 2003;82:232–7. https://doi.org/10.1177/154405910308200316.Search in Google Scholar PubMed

11. do Nascimento, C, Miani, PK, Pedrazzi, V, Gonçalves, RB, Ribeiro, RF, Faria, AC, et al.. Leakage of saliva through the implant-abutment interface: in vitro evaluation of three different implant connections under unloaded and loaded conditions. Int J Oral Maxillofac Implants 2012;27:551–60.Search in Google Scholar

12. Mishra, SK, Chowdhary, R, Kumari, S. Microleakage at the different implant abutment interface: a systematic review. J Clin Diagn Res 2017;11:10–15. https://doi.org/10.7860/JCDR/2017/28951.10054.Search in Google Scholar PubMed PubMed Central

13. Yamanishi, Y, Yamaguchi, S, Imazato, S, Nakano, T, Yatani, H. Influences of implant neck design and implant-abutment joint type on peri-implant bone stress and abutment micromovement: three-dimensional finite element analysis. Dent Mater 2012;28:1126–33. https://doi.org/10.1016/j.dental.2012.07.160.Search in Google Scholar PubMed

14. Merz, BR, Hunenbart, S, Belser, UC. Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000;15:519–26. PMID: 10960985.Search in Google Scholar

15. Kieschnick, A. Im Gespräch mit Ztm. Dirk Bachmann und Ztm. Carina Krämer Ein Implantat, Zwei Anschlussgeometrien. Dent Dialog 2019;7:48–53.Search in Google Scholar

16. Suttin, Z, Towse, R. Dynamic loading fluid leakage characterization of dental implant systems. ART1205EU BIOMET 3i White Paper. BIOMET 3i, Palm Beach Gardens, Florida, USA; 2013.Search in Google Scholar

17. International Organization for Standardization (ISO). DE EN ISO 14801:2016: Dentistry – Implants – Dynamic loading test for endosseous dental implants, Geneva, Switzerland; 2016.Search in Google Scholar

18. Zipprich, H, Weigl, P, Lange, B, Lauer, HC. Erfassung, Ursachen und Folgen von Mikrobewegungen am Implantat-Abutment-Interface. Implantologie 2007;15:31–46.Search in Google Scholar

19. Wylezalek, M. Numerische Untersuchungen der Spaltbildung und Mikrobewegung zwischen Implantat und Abutment an verschiedenen Interface-Geometrien [Dr.med.dent thesis]. Bonn, Germany: Medical Faculty, University of Bonn; 2020.Search in Google Scholar

20. Quirynen, M, van Steenberghe, D. Bacterial colonization of the internal part of two-stage implants. An in vivo study. Clin Oral Implants Res 1993;4:158–61. https://doi.org/10.1034/j.1600-0501.1993.040307.x.Search in Google Scholar PubMed

21. Candotto, V, Gabrione, F, Oberti, L, Lento, D, Severino, M. The role of implant-abutment connection in preventing bacterial leakage: a review. J Biol Regul Homeost Agents 2019;33:129–34.Search in Google Scholar

22. Rack, A, Rack, T, Stiller, M, Riesemeier, H, Zabler, S, Nelson, K. In vitro synchrotron-based radiography of micro-gap formation at the implant-abutment interface of two-piece dental implants. J Synchrotron Radiat 2010;17:289–94. https://doi.org/10.1107/S0909049510001834.Search in Google Scholar PubMed PubMed Central

23. Steinebrunner, L, Wolfart, S, Bössmann, K, Kern, M. In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems. Int J Oral Maxillofac Implants 2005;20:875–81.Search in Google Scholar

24. Suttin, Z, Towse, R. Effect of abutment screw design on implant system seal performance. Presented at the European Association for Osseointegration, 20th Annual Scientific Meeting; October 2012; Copenhagen, Denmark.Search in Google Scholar

25. Alsubaiy, EF. Abutment screw loosening in implants: a literature review. J Fam Med Prim Care 2020;9:5490–4. https://doi.org/10.4103/jfmpc.jfmpc_1343_20.Search in Google Scholar PubMed PubMed Central

26. Quek, HC, Tan, KB, Nicholls, JI. Load fatigue performance of four implant-abutment interface designs: effect of torque level and implant system. Int J Oral Maxillofac Implants 2008;23:253–62.Search in Google Scholar

27. van der Bilt, A, Engelen, L, Pereira, LJ, van der Glas, HW, Abbink, JH. Oral physiology and mastication. Physiol Behav 2006;89:22–7. https://doi.org/10.1016/j.physbeh.2006.01.025.Search in Google Scholar PubMed

28. Kohyama, K, Hatakeyama, E, Sasaki, T, Dan, H, Azuma, T, Karita, K. Effects of sample hardness on human chewing force: a model study using silicone rubber. Arch Oral Biol 2004;49:805–16. https://doi.org/10.1016/j.archoralbio.2004.04.006.Search in Google Scholar PubMed

29. Waltimo, A, Könönen, M. A novel bite force recorder and maximal isometric bite force values for healthy young adults. Scand J Dent Res 1993;101:171–5. https://doi.org/10.1111/j.1600-0722.1993.tb01658.x.Search in Google Scholar PubMed

30. Alam, MK, Alfawzan, AA. Maximum voluntary molar bite force in subjects with malocclusion: multifactor analysis. J Int Med Res 2020;48:300060520962943. https://doi.org/10.1177/0300060520962943.Search in Google Scholar PubMed PubMed Central

31. da Silva, RJ, Issa, JP, Semprini, M, da Silva, CH, de Vasconcelos, PB, Celino, CA, et al.. Clinical feasibility of mandibular implant overdenture retainers submitted to immediate load. Gerodontology 2011;28:227–32. https://doi.org/10.1111/j.1741-2358.2010.00374.x.Search in Google Scholar PubMed

32. Müller, F, Duvernay, E, Loup, A, Vazquez, L, Herrmann, FR, Schimmel, M. Implant-supported mandibular overdentures in very old adults: a randomized controlled trial. J Dent Res 2013;92:154S–60S. https://doi.org/10.1177/0022034513509630.Search in Google Scholar PubMed PubMed Central

33. van der Bilt, A, Burgers, M, van Kampen, FM, Cune, MS. Mandibular implant-supported overdentures and oral function. Clin Oral Implants Res 2010;21:1209–13. https://doi.org/10.1111/j.1600-0501.2010.01915.x.Search in Google Scholar PubMed


Supplementary Material

This article contains supplementary material (https://doi.org/10.1515/bmt-2025-0141).


Received: 2025-04-07
Accepted: 2025-04-11
Published Online: 2025-05-05
Published in Print: 2025-10-27

© 2025 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Downloaded on 5.11.2025 from https://www.degruyterbrill.com/document/doi/10.1515/bmt-2025-0141/html?lang=en&srsltid=AfmBOoohZuTXHMDc5EwS-3lrXjJQKQknZxaPiHAR56EpLrVn5msHKn2m
Scroll to top button