How to handle lipemic CBC samples on Sysmex hematology analyzers?

  • Vanja Radišić Biljak ORCID logo EMAIL logo , Lucija Dolovčak , Iva Bakarić , Ana Nikler ORCID logo , Andrea Saračević ORCID logo and Marija Grdić Rajković
Published/Copyright: March 20, 2025

Abstract

Objectives

Lipemia poses a significant preanalytical problem for complete blood count (CBC) measurement due to limited and non-standardized methods for recognition and removal. We aimed to verify the optical hemoglobin (Hb-O) measurements on the Sysmex XN-1000 hematology analyzer (HA) as a possible reliable method for managing lipemic CBC samples.

Methods

Ninety CBC samples with varying Hb concentrations were gradually spiked with a lipid emulsion. Measurements were repeated and Hb-O concentrations were recorded. Spiked CBC samples were centrifuged (400 g/10 min). Plasma was carefully removed, and Hb concentration was measured. The values obtained from the lipemic samples were adjusted according to the measurements in the plasma. The removed plasma was substituted with the analyzer’s diluent, and measurements were repeated. Triglyceride concentrations were measured in lipemic plasma samples.

Results

Hb-O showed statistically insignificant and acceptable bias compared to the initial Hb measurement according to the strictest acceptability criteria (−0.4 %, 95 % CI: −1.2–0.3, p=0.2447). The observed bias did not correlate with the degree of lipemia (rho=−0.072, 95 % CI: −0.295 to 0.157, p=0.537). Hemoglobin measured in samples with lipemic plasma replaced by analyzer diluent exhibited minimal, albeit statistically significant, bias (−1.1 %, 95 % CI: −2.0 to (−0.1), p=0.025). The observed bias negatively correlated with the degree of lipemia (rho=−0.369, 95 % CI: −0.550 to (−0.155), p=0.001). The highest unacceptable bias was found in the recalculated hemoglobin values based on the measured plasma hemoglobin (−3.5 %, 95 % CI: −4.1 to (−2.9), p<0.0001).

Conclusions

Hb-O measurement is the most reliable measure of lipemia removal in CBC samples on the Sysmex XN-1000 HA.

Introduction

Lipemia, defined as sample turbidity due to the accumulation of lipoprotein particles, causes a significant problem in the clinical laboratory setting [1]. Although relatively uncommon, it poses a significant preanalytical problem for complete blood count (CBC) measurement in lipemic whole blood samples for various reasons. As most hematology analyzers (HA) measure hemoglobin spectrophotometrically on 530 nm, lipemia interferes with hemoglobin measurement, causing falsely high concentrations of hemoglobin and, consequently, mean corpuscular hemoglobin (MCH) and mean corpuscular hemoglobin concentration (MCHC) [2]. In biochemistry and endocrinology measurements hemolysis, icteria and lipemia interference is mostly identified visual inspection, by (semi)quantitative HIL measurements, or some combination of both [3]. Contrary, the most common alert that could indicate a lipemic CBC sample is the increase of MCHC above a certain threshold. Although some authors propose MCHC of 360 g/L as a lipemia threshold [4], every laboratory should have an established cut-off level of lipemia above which results should not be reported [5]. Several possible solutions to eliminate lipemia interference in CBC measurement are present in routine hematology laboratories worldwide [6]. One includes hemoglobin concentration measurement in spun lipemic plasma samples. The obtained hemoglobin value measured in lipemic plasma samples allows further whole blood hemoglobin concentration correction. The measured hemoglobin concentration from the lipemic plasma sample is subtracted from the hemoglobin concentration in the whole blood sample. After correction, MCH and MCHC values are recalculated. Another widespread lipemia removal procedure is the plasma replacement procedure. This procedure involves removing the lipemic plasma and replacing it with an appropriate volume of the analyzer’s diluent, followed by a sample re-analysis. The most practical solution for obtaining reliable CBC results from lipemic whole blood samples is employing hematology analyzers insensitive to lipemia interference. Siemens Advia 2120i HA measures cellular hemoglobin concentration mean (CHCM) – hemoglobin is measured directly in the intact erythrocytes. This approach allows for accurate hemoglobin measurement without interference from lipemia [7], [8], [9].

In 2019 The Croatian Working Group for Laboratory Hematology, on behalf of the Croatian Society of Medical Biochemistry and Laboratory Medicine, wanted to explore the background in field of laboratory hematology routine practice among Croatian laboratories in order to develop future strategies for producing national recommendations in order to promote harmonization in the field [5]. The survey showed various strategies for resolving lipemia-related issues in CBC measurements [5]. The most common strategy was the first procedure in which hemoglobin concentration was measured in spun lipemic plasma samples, which accounted for 40 % of cases [5]. However, this method requires considerable sample manipulation and necessitates recalculating results to ensure reliable hemoglobin values.

Until recently, our laboratory utilized Siemens Advia 2120i hematology analyzers. Thus, lipemia certainly was not an issue for CBC measurements. However, Advia has been replaced with Sysmex XN-1000. Therefore, in order to optimize and standardize our practice, as well as to ensure patient safety, we aimed to establish our lipemia interference protocol for reliable hemoglobin measurement. Although Sysmex recognizes this possible sample interference, the company does not provide any practical solution for lipemia removal under the System limitations subheading in the Automated Hematology Analyzer XN series (XN-1000) Instructions for Use user manual. However, based on the recent Berda-Haddad et al. study results [10], Sysmex Corporation proposes an algorithm for resolving increased MCHC measurements. This algorithm can be integrated into laboratory-extended IPU (EPU) if available. As increased sample turbidity is one of the possible causes of increased MCHC measurements, based on the proposed Sysmex algorithm, a reliable method for eliminating lipemia interference in lipemic CBC samples is the measurement of optical hemoglobin in the reticulocyte measurement mode (Hb-O) [11]. However, Hb-O is currently a research-only parameter and is not included in the quality control (QC) menu. Therefore, we wanted to explore the analytical characteristics of the new optical hemoglobin measurement and compare it with the most commonly used lipemia-eliminating protocols.

Materials and methods

The experimental study was performed in the Department of Medical Laboratory Diagnostics, University Hospital “Sveti Duh”, Zagreb, Croatia, during October and November 2024.

Precision

Before the lipemia assessment, a short precision study was performed regarding Hb-O measurements on the Sysmex XN-1000 hematology analyzer (HA) (Sysmex, Kobe, Japan). Within-day precision was explored by using patient samples in three different concentration ranges, and commercial control samples (Sysmex XN Check, lot: Level 1 42461101, Level 2 42461102, Level 3 42461103, exp: 24.11.2024.) by 20 consecutive analyses. Between-day precision was established by using commercial control samples, daily for 30 days. The coefficient of variation was calculated for within- and between-day precision study.

Accuracy

Ninety whole blood K2EDTA patient samples with varying Hb concentrations and a clinician request for CBC were analyzed on Sysmex XN-1000 HA. The initial CBC measurement was considered the true value. To ensure various Hb concentrations, samples were selected according to the distribution presented in Table 1.

Table 1:

The distribution of required whole-blood K2EDTA samples for the lipemia removement verification protocol.

Initial Hb concentration, g/L Addition of lipid emulsion, µL
20 40 60 80 100 120
50–110 5 5 5 5 5 5
110–150 5 5 5 5 5 5
>150 5 5 5 5 5 5

To simulate lipemia interference the samples were gradually spiked with a SMOFlipid 200 mg/mL Emulsion zur Infusion lipid emulsion (Fresenius Kabi, Graz, Austria, exp. 03/2025), by following the predetermined emulsion volumes (Table 1).

After the addition of lipemic emulsion CBC measurements were repeated, and standard Hb concentrations measured by spectrophotometrical method were recorded.

Verification protocol for reliable measurement of CBC parameters in lipemic whole blood samples on the Sysmex XN-1000 hematology analyzer included three different methods:

  1. Measurement of lipemic samples in reticulocyte measurement mode to obtain Hb-O and MCHC-O measurements.

  2. Hemoglobin concentration measurement in spun lipemic plasma samples.

    Spiked lipemic samples were centrifuged on 400 g for 10 min. Lipemic plasma was carefully removed in order not to disturb the cells, and standard Hb concentrations were measured in lipemic plasma samples. Consenquently, the values obtained from the lipemic samples were adjusted according to the measurements in the plasma by following the simple equation:

    Hb corrected g / L = Hb lipemic sample Hb lipemic plasma × 1  hematocrit  L / L .

    MCH and MCHC were calculated as instructed:

    MCH = Hb g / L Erc   × 10 12 / L p g

    MCHC = Hb g / L Hct   L / L g / L

  3. Replacement of the removed lipemic plasma with an adequate volume of the analyzer’s diluent.

    The removed plasma was substituted with the analyzer’s diluent (Sysmex Cellpack™ DCL). After thorough mixing, measurements were repeated and standard Hb concentrations were recorded.

To assess the degree of simulated lipemia, triglyceride concentrations, as well as HIL index, were measured in lipemic plasma samples on Abbott Alinity c analyzer (Abbott, Abbott Park, IL, USA). Tryglicerides were measured utilizing the glycerol phosphate oxidase method that produces a quinoneimine dye, where the absorbance of the dye, measured at 604 nm is proportional to the concentration of triglyceride present in the sample. HIL indices were measured spectrophotometrically using different wavelength pairs and a specific algorithm to obtain HIL index values that approximately correspond to free hemoglobin, bilirubin and triglycerides concentrations.

The whole protocol is presented in Table 2.

Table 2:

Verification protocol for reliable measurement of CBC parameters in lipemic whole blood K2EDTA samples on the Sysmex XN-1000 hematology analyzer.

Steps Analysis Additional
CBC measurement in non-lipemic native samples Hemoglobin, MCH and MCHC measured in the native samples will be considered as “true value”. Print the original CBC measurement.
Lipid emulsion addition Smoflipid lipid emulsion will be added to the selected samples. Starting with 20 µL and gradually increasing the volume. The samples will be gently mixed well to evenly distribute the lipid emulsion.
Hb-O and MCHC-O measurement Lipemic samples will be analyzed in the retic measurement mode. The Hb-O and MCHC-O measurement is located in the Lab only tab.
Lipemic sample centrifugation Lipemic samples will be centrifuged for 10 min at 400 g. Choose CSF program on the centrifuge.
Lipemic plasma removal After centrifugation, the plasma level in the CBC sample will be recorded. The lipemic plasma will be carefully separated into a clean plastic tube using an automatic pipette.
Hb measurement in lipemic plasma samples Separated lipemic plasma will be analyzed in open sampling mode on a Sysmex XN-1000 analyzer. The result of Hb in the lipemic plasma is printed. Plasma is referred to biochemistry for triglyceride and HIL index measurement.
Addition of analyzer’s diluent Analyzer diluent (Sysmex Cellpack™ DCL) will be added to the CBC sample up to the mark where the plasma has reached. The sample will be gently mixed well to evenly distribute the diluent throughout the sample.
Reanalyzing the sample The CBC samples will be reanalyzed on Sysmex XN-1000 HA. The results are printed.
Results entry All results are entered in the suitable Excel sheet. All possible problems during the implementation of the protocol are recorded.

No patient information was collected in this study as the interference study included the remnant plasma samples with CBC orders. The ethics committee ot University Hospital “Sveti Duh”, Zagreb, Croatia granted an universal approval for verification studies on remnant samples. They confirmed that patient consent in such studies were not needed.

Statistical analysis

An average bias between the measurements obtained from different lipemia removing methods, compared to the initial measurements, was calculated by employing the Bland-Altman plot. Average bias was expressed as percentage and compared to biological variability criteria for bias according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Biological Variation Database [12]. To assess the possible correlation of the calculated biases and the degree of lipemia, as well as hemoglobin concentration, rank correlation was performed [13]. Differences between three lipemia removing methods were tested by the Friedman test, as variables were not normally distributed. p values<0.05 were considered statistically significant. Statistical analysis was performed in MedCalc® Statistical Software version 23.0.2 (MedCalc Software Ltd, Ostend, Belgium).

Results

Precision

The precision study results were within the acceptance criteria and are presented in Table 3. The Hb-O measurement on Sysmex XN-1000 HA is currently a research-only parameter, and there are no manufacturer-defined acceptability criteria for precision. Therefore, we have used the biological variability criteria for precision.

Table 3:

Results of the precision study for measurement of optical hemoglobin on Sysmex XN-1000 hematology analyzer.

Precision, % Patient samples Control samples
Low Medium High L1 L2 L3
x CV BV x CV BV x CV BV x CV BV x CV BV x CV BV
HB-O 101 1.4 2.0 119 1.6 2.0 141 1.3 2.0 56 1.7 2.0 104 0.9 2.0 132 1.7 2.0

HB-O, measurement of optical hemoglobin on Sysmex XN-1000 hematology analyzer; CV, coefficient of variation; BV, acceptance criteria for precision derived from the EFLM biological variability database (miminum).
Within acceptance criteria

Accuracy

Fifteen samples with the highest degree of simulated lipemia (addition of 120 μL of lipid emulsion) were grossly hemolyzed and were excluded from further statistical analysis. The gradual increase in hemoglobin concentration, MCHC measurement and triglyceride concentration is presented in Table 4.

Table 4:

The gradual increase of hemoglobin and MCHC concentration after addition of lipid emulsion.

Added volume of lipid emulsion Hb native Hb+ lipid emulsion Percentage change MCHC native MCHC+ lipid emulsion Percentage change TG L index H index
20 Median 113 117 3.5 336 350 4.1 8.2 264 31
IQR 97–166 103–170 2.9–5.5 331–341 339–356 3.0–5.8 7.2–9.8 223–387 11–64
40 Median 115 124 7.1 333 361 9.1 14.7 497 19
IQR 97–156 107–162 4.8–9.7 328–339 352–373 6.8–11.7 12.8–18.6 454–622 5–42
60 Median 138 148 10.6 331 369 13.1 21.6 803 66
IQR 90–158 103–173 7.5–14.3 323–340 362–400 9.5–17.8 16.9–30.3 647–1115 27–92
80 Median 127 148 13.9 330 381 15.8 25.7 998 61
IQR 102–156 117–168 11.2–15.1 318–341 369–399 13.9–18.5 22.8–33.4 797–1181 16–91
100 Median 134 158 17.9 332 404 21.4 40.5 1539 37
IQR 94–157 125–176 12.8–27.0 314–341 389–438 13.8–36.3 32.7–45.7 1192–1712 22–94

Reference change value (RCV) for hemoglobin=6.8 %; RCV for MCHC=3.0 %.
Within RCV Greater than RCV

The results of the accuracy study are presented in Table 5. Hb-O showed statistically insignificant and acceptable bias compared to the initial Hb measurement according to the strictest acceptability criteria (−0.4 %, 95 % CI: −1.2–0.3, p=0.2447). Hemoglobin measured in samples with lipemic plasma replaced by analyzer diluent exhibited minimal, albeit statistically significant, bias (−1.1 %, 95 % CI: −2.0 – (−0.1), p=0.025). The highest unacceptable bias was found in the recalculated hemoglobin values based on the measured plasma hemoglobin (−3.5 %, 95 % CI: −4.1 – (−2.9), p<0.0001).

Table 5:

Comparison of hemoglobin, MCH and MCHC concentrations in lipemia removal methods.

Lipemia removal method Bias % 95 % CI p-Value Biological variability criteria for biasa
Optimal bias Desirable bias Minimum bias
1. Measurement of lipemic samples in reticulocyte measurement mode
Hb-O −0.4 −1.2 to 0.3 0.244 0.8 1.7 2.5
MCH (Hb-O/RBC) −1.2 / / 0.6 1.2 1.7
MCHC-O 1.3 / / 0.2 0.4 0.6
MCHC (Hb-O/HCT) −0.5 / / 0.2 0.4 0.6
2. Hemoglobin concentration measurement in spun lipemic plasma samples
Hb corrected −3.5 −4.1 to (−2.9) <0.001 0.8 1.7 2.5
MCH calculated −4.4 / / 0.6 1.2 1.7
MCHC calculated −1.5 / / 0.2 0.4 0.6
3. Replacement of the removed lipemic plasma with the analyzer’s diluent
Hb diluent −1.1 −2.0 to (−0.1) 0.025 0.8 1.7 2.5
MCH diluent 2.5 / / 0.6 1.2 1.7
MCHC diluent 2.7 / / 0.2 0.4 0.6
MCHC (Hb diluent/HCT) 0.7 / / 0.2 0.4 0.6

aAarsand AK, Fernandez-Calle P, Webster C, Coskun A, Gonzales-Lao E, Diaz-Garzon J, et al. The EFLM Biological Variation Database. https://biologicalvariation.eu/. Allowed bias: <0.25 × (CVI2 + CVG2)1/2 The factor 0.25 refers to desirable Analytical Performance Specification (APS). The factor for optimum and minimum performance specifications are 0.125 and 0.375, respectively. CVI, within-subject (CVI) estimate; CVG, between-subject (CVG) estimate. Available at: Aarsand AK, Fernandez-Calle P, Webster C, Coskun A, Gonzales-Lao E, Diaz-Garzon J, et al. The EFLM Biological Variation Database. https://biologicalvariation.eu/. Hb-O, measurement of optical hemoglobin on Sysmex XN-1000 hematology analyzer; MCHC-O, measurement of optical MCHC concentration on Sysmex XN-1000 hematology analyzer; RBC, red blood cell concentration; HCT, hematocrit.
Within acceptance criteria Greater than acceptance criteria

MCHC measured with all three lipemia removal protocols showed unacceptable results according to the acceptability criteria for accuracy. However, if a mathematical correction is applied to the calculation of MCHC by using the Hb-O measurement, we obtained an acceptable bias according to the criteria of minimally acceptable accuracy compared to native MCHC.

Friedman’s test showed a statistically significant difference between the observed hemoglobin measurements (p<0.00001). Post hoc testing showed a statistically significant difference between hemoglobin measured in the native sample and hemoglobin obtained by correction for plasma hemoglobin and hemoglobin measured in samples where plasma was replaced by the analyzer’s diluent (p<0.05). No statistically significant difference was observed between initial hemoglobin measurement and Hb-O (Table 6). The same results were obtained when all included hemoglobin measurements were divided into categories: <110 g/L, 110–150 g/L, and >150 g/L, thus reflecting low, normal, and high hemoglobin concentrations. Hb-O measurements were equally reliable independently of concentration spans (Table 6).

Table 6:

Comparison of measurements by using the Friedman statistical test for paired samples.

Comparison of measurements, n=75 1. Native sample measurement 2. Optical measurement 3. Spun lipemic plasma samples correction 4. Replacement of the removed lipemic plasma with the analyzer’s diluent p-Value
Median (IQR) Median (IQR) Median (IQR) median (IQR)
Hemoglobin (all samples) 132 (101–158) 129 (102–158) 124a (97–151) 126a (98–155) <0.00001
<110 g/L 92 (83–98) 91 (83–96) 89a (83–96) 92 (86–98) 0.00484
110–150 g/L 128 (119–137) 127 (121–137) 122a (116–131) 125a (119–133) 0.00001
>150 g/L 162 (158–167) 163 (158–170) 158a (155–167) 155a (151–164) <0.00001
MCH 30.2 (28.4–31.2) 30.2 (28.3–31.2) 29.2a (27.5–30.1) 30.9a (29.3–32.6) <0.00001
MCHC 334 (323–340) 340a (324–353) 331a (319–337) 341a (331–348) <0.00001
  1. a Post-hoc statistically significant difference compared to the initial measurement in the native sample (p<0.05).

The rank correlation statistical analysis explored the possible correlations between the observed biases in Hb measurement in all three lipemia removal methods and the degree of lipemia/hemolysis (measured by triglyceride concentration and HIL indexes). The results are presented in Table 7.

Table 7:

Correlation between deviations in hemoglobin measurement in lipemic samples and the degree of lipemia and hemolysis.

Calculated bias

Triglycerides (mmol/L)

Lipemia degree (L index)

Hemolysis (H index)

Rho (95% CI)

p-Value

Rho (95% CI)

p-Value

Rho (95% CI)

p-Value

Hb-O

−0.072 (−0.295 to 0.157)

0.537

−0.041 (−0.266 to 0.187)

0.725

0.068 (−0.161 to 0.291)

0.559

Hb-corrected

−0.915 (−0.946 to (−0.868))

<0.001

−0.918 (−0.947 to (−0.873))

<0.001

−0.390 (−0.566 to (−0.178))

0.001

Hb-diluent

−0.369 (−0.550 to (−0.155))

0.001

−0.412 (−0.590 to (−0.196))

0.001

−0.406 (−0.585 to (−0.190))

0.001


Statistically significant correlation.

Discussion

The Sysmex XN-1000 Hb-O measurement has effectively addressed interference caused by lipemia in lipemic whole blood EDTA samples. By re-analyzing samples in the reticulocyte measurement mode, we eliminated the need for the manual procedures traditionally used in medical-biochemistry laboratories to remove lipemia interference.

While we observed a significant increase in hemoglobin (Hb) concentrations, even with the smallest addition of lipids (20 µL), this increase was not clinically significant when compared to reference change values (RCV) obtained from the EFLM database. The concentration of triglycerides in such samples could reach up to 10 mmol/L. However, further addition of lipid emulsion did result in Hb concentrations exceeding the clinically significant RCV, prompting us to reevaluate the proposed Sysmex MCHC cut-off of 365 g/L as an indicator of lipemic CBCs. To minimize unnecessary testing and, at the same time effectively identify lipemia interference, we suggest a lower MCHC threshold of 360 g/L that could enhance the management of analytical interferences, similar to recommendations made by Henry et al. [14].

Another advantage of Hb-O measurement, in contrast to the existing protocols for lipemia removal, is its independence from the levels of both lipemia and hemolysis. It is well-known that increased lipid concentration is associated with heightened hemolysis, as demonstrated in our study [15]. The typical approach for addressing lipemia-related issues in CBC measurements in Croatia – measuring hemoglobin concentration in spun lipemic plasma samples – is significantly affected by both the degree of lipemia and hemolysis, leading to unreliable results in samples with high levels of lipemia. Similar findings were reported in a study by Aruga et al., which highlighted the effectiveness of the optical method for measuring hemoglobin in chylous samples [16]; however, their research did not include a comparison with the commonly used lipemia-removal methods.

Our study does have some limitations. The results might vary with different lipid emulsions used for simulating lipemia interference. However, simulating lipemia this way is the best we can achieve in laboratories. The absence of internal quality control or Hb-O measurement presents a potential concern regarding using this research-only parameter in routine clinical assessments. Although Hb-O showed higher total imprecision compared to the spectrophotometrically measured Hb concentrations, the observed imprecision was well within the Sysmex acceptance criteria. Sysmex internal QC samples can further be validated for Hb-O measurement in various QC lots.

Despite some shortcomings, we have successfully demonstrated that Hb-O measurement using the Sysmex HA is superior in eliminating lipemia interference compared to existing lipemia removal methods used worldwide, including in Croatia. This method offers a straightforward approach to standardizing the lipemia elimination protocol for complete blood count (CBC) measurements, without necessitating any manipulation of the samples. Traditional lipemia removal techniques often involve sample manipulation, which can introduce significant bias, potentially leading to erroneous results and compromising patient safety [17], 18]. Ultimately, adopting a simple and safe procedure for lipemia removal through direct Hb-O measurement could reduce the overall workload in light of the globally recognized shortage of laboratory personnel [19].


Corresponding author: Assist. Prof. Vanja Radišić Biljak, PhD, EuSpLM, Department of Medical Laboratory Diagnostics, University Hospital “Sveti Duh”, Sveti Duh 64, 10000 Zagreb, Croatia; and Department of Sport and Exercise Medicine, Faculty of Kinesiology, University of Zagreb, Zagreb, Croatia, E-mail:

  1. Research ethics: The local Institutional Review Board deemed the study exempt from review.

  2. Informed consent: Not applicable.

  3. Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  4. Use of Large Language Models, AI and Machine Learning Tools: Grammarly for improving language.

  5. Conflict of interest: The authors state no conflict of interest.

  6. Research funding: None declared.

  7. Data availability: The raw data can be obtained on request from the corresponding author.

  8. Article Note: A translation of this article can be found here: https://doi.org/10.1515/almed-2025-0037.

References

1. Nikolac, N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med 2014;24:57–67. https://doi.org/10.11613/bm.2014.008.Search in Google Scholar

2. Gulati, G, Uppal, G, Gong, J. Unreliable automated complete blood count results: causes, recognition, and resolution. Ann Lab Med 2022;42:515–30. https://doi.org/10.3343/alm.2022.42.5.515.Search in Google Scholar PubMed PubMed Central

3. Cadamuro, J, Lippi, G, von Meyer, A, Ibarz, M, van Dongen-Lases, E, Cornes, M, et al.. European survey on preanalytical sample handling – Part 2: Practices of European laboratories on monitoring and processing haemolytic, icteric and lipemic samples. On behalf of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for the Preanalytical Phase (WG-PRE). Biochem Med 2019;29:020705. https://doi.org/10.11613/BM.2019.020705.Search in Google Scholar PubMed PubMed Central

4. Fernández Prendes, C, Castro Castro, MJ, Sánchez Navarro, L, Rapún Mas, L, Morales Indiano, C, Arrobas Velilla, T. Handling of lipemic samples in the clinical laboratory. Adv Lab Med 2023;4:5–15. https://doi.org/10.1515/almed-2023-0003.Search in Google Scholar PubMed PubMed Central

5. Radišić Biljak, V, Lapić, I, Vidranski, V, Herceg, I, Tomić, F, Šimac, B, et al.. Policies and practices in the field of laboratory hematology in Croatia – a current overview and a call for improvement. Clin Chem Lab Med 2021;60:271–82.10.1515/cclm-2021-1027Search in Google Scholar PubMed

6. Zandecki, M, Genevieve, F, Gerard, J, Godon, A. Spurious counts and spurious results on haematology analysers: a review. Part II: White blood cells, red blood cells, haemoglobin, red cell indices and reticulocytes. Int J Lab Hem 2007;29:21–41. https://doi.org/10.1111/j.1365-2257.2006.00871.x.Search in Google Scholar PubMed

7. Zeng, S-G, Zeng, T-T, Jiang, H, Wang, L-L, Tang, S-Q, Sun, Y-M, et al.. A simple, fast correction method of triglyceride interference in blood hemoglobin automated measurement. J Clin Lab Anal 2013;27:341–5. https://doi.org/10.1002/jcla.21568.Search in Google Scholar PubMed PubMed Central

8. Harris, N, Kunicka, J, Kratz, A. The ADVIA 2120 hematology system: flow cytometry-based analysis of blood and body fluid sin the routine hematology laboratory. Lab Hematol 2005;11:47–61. https://doi.org/10.1532/lh96.04075.Search in Google Scholar PubMed

9. Choi, BH, Cho, SR, Park, IJ. Cellular hemoglobin is a useful alternative parameter when lipemia interferes with hemoglobin measurement. J lab Med Qual Assur 2017;39:9–15. https://doi.org/10.15263/jlmqa.2017.39.1.9.Search in Google Scholar

10. Berda-Haddad, Y, Faure, C, Boubaya, M, Arpin, M, Cointe, S, Frankel, D, et al.. Increased mean corpuscular haemoglobin concentration: artefact or pathological condition? Int J Lab Hematol 2017;39:32–41. https://doi.org/10.1111/ijlh.12565.Search in Google Scholar PubMed

11. CBC–O – Discover the truth about increased MCHC. https://www.sysmex-europe.com/academy/library/documents [Accessed 15 Dec 2024].Search in Google Scholar

12. Aarsand, AK, Fernandez-Calle, P, Webster, C, Coskun, A, Gonzales-Lao, E, Diaz-Garzon, J, et al.. The EFLM Biological Variation Database. https://biologicalvariation.eu/ [Accessed 15 Dec 2024].Search in Google Scholar

13. Udovičić, M, Baždarić, K, Bilić-Zulle, L, Petrovečki, M. What we need to know when calculating the coefficient of correlation? Biochem Med 2007;17:10–15. https://doi.org/10.11613/bm.2007.002.Search in Google Scholar

14. Henry, S, Gérard, D, Salignac, S, Perrin, J. Optimizing the management of analytical interferences affecting red blood cells on XN-10 (Sysmex®). Int J Lab Hematol 2022;44:1068–77. https://doi.org/10.1111/ijlh.13951.Search in Google Scholar PubMed PubMed Central

15. Dimeski, G, Mollee, P, Carter, A. Increased lipid concentration is associated with increased hemolysis. Clin Chem 2005;51:2425. https://doi.org/10.1373/clinchem.2005.058644.Search in Google Scholar PubMed

16. Aruga, Y, Ikeda, C, Hanai, A, Yoshimura, S, Kito, M, Miyaki, S, et al.. Convenience of Hgb-O detected by optical method in XN-series hematology analyzers in evaluating hemoglobin concentration in samples with chylous turbidity. Nat Sci Rep 2021;11:14978. https://doi.org/10.1038/s41598-021-94394-z.Search in Google Scholar PubMed PubMed Central

17. Plebani, M. Errors in laboratory medicine and patient safety: the road ahead. Clin Chem Lab Med 2007;45:700–7. https://doi.org/10.1515/cclm.2007.170.Search in Google Scholar

18. Plebani, M, Aita, A, Sciacovelli, L, Donaldson, L, Ricciardi, W, Sheridan, S, et al.. Patient safety in laboratory medicine. In: Textbook of patient safety and clinical risk management [Internet]. Cham (CH): Springer; 2021. Chapter 24.2020 Dec 15.10.1007/978-3-030-59403-9_24Search in Google Scholar PubMed

19. Louzan, MJ, Martin, RM. Addressing the laboratorian shortage. JALM 2024;9:172174.10.1093/jalm/jfad080Search in Google Scholar PubMed

Received: 2024-12-23
Accepted: 2025-02-02
Published Online: 2025-03-20

© 2025 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Review / Artículo de Revisión
  3. Implementation of circulating cell-free DNA screening for fetal aneuploidies
  4. Implementación del ADN libre circulante para la detección de aneuploidías fetales
  5. Mini Review / Mini Revisión
  6. Data science applied to the assessment of biological variation estimates
  7. Ciencia de datos aplicada a la obtención de estimados de variación biológica
  8. Original Article / Artículo Original
  9. How to handle lipemic CBC samples on Sysmex hematology analyzers?
  10. Cómo manejar las muestras lipémicas para realización de contaje de células sanguíneas (CBC) en los analizadores hematológicos Sysmex
  11. Use of artificial intelligence to assess genetic predisposition to develop critical COVID-19 disease: a comparative study of machine learning models
  12. Uso de inteligencia artificial en la predisposición genética a enfermedad crítica por COVID-19: evaluación comparativa de modelos de aprendizaje automático
  13. Use of external quality assessment in extra-analytical phases in clinical laboratories in Spain: a survey by the Spanish Society of Laboratory Medicine (SEQCML)
  14. Uso de la evaluación externa de la calidad en las fases extraanalíticas de los laboratorios clínicos españoles: una encuesta de la Sociedad Española de Medicina de Laboratorio (SEQCML)
  15. Case Report / Caso Clínico
  16. Diagnosis of hemogobinopathies in the clinical laboratory: an occult Hofu hemoglobin on HPLC
  17. Diagnóstico de hemoglobinopatías en el laboratorio clínico: hallazgo de una hemoglobina hofu oculta en HPLC
  18. Letter to the Editor / Carta al Editor
  19. Biomarker repurposing: opportunities and challenges in the evolving field of laboratory medicine
  20. La reutilización de biomarcadores para otros fines: oportunidades y dificultades en el cambiante campo de la medicina de laboratorio
Downloaded on 8.9.2025 from https://www.degruyterbrill.com/document/doi/10.1515/almed-2024-0206/html
Scroll to top button